Professional Documents
Culture Documents
STACK - Procal 2000 Operating Manual
STACK - Procal 2000 Operating Manual
STACK - Procal 2000 Operating Manual
Analyser Unit
Operating Manual
PROCAL 2000 Operating Manual
All possible care has been taken in the preparation of this publication, but
Kittiwake Procal Ltd its agents and distributors, accept no liability for any
inaccuracies that may be found.
Kittiwake Procal Limited reserves the right to make changes without notice both
to this publication and the products which it describes.
In-service history
This manual is intended to assist the user in the safe and efficient installation,
operation and maintenance of the Procal 2000 series of analyser units. It has
been written in accordance with the requirements of British Standard BS4884
Technical manuals Part 1: 1992.
Principles behind the system are described in the section entitled Technical
description, together with descriptions of the different component parts that
make up the system. Every system including a PROCAL 2000 Analyser Unit
also incorporates an Analyser Controller. An In-Situ Heater (ISH) is available
as an option, as an Input Output Unit (IOU).
The Procal 1100 (Analyser Control for Windows Network software only for
installation on customer support PC) which can control up to Six Analyser Units,
is supplied with its own Operating Manual. Information relating to additional
equipment that can be used in the system is given in this manual and in the
Procal 1200 manual.
An optical diagram of the PROCAL 2000 system is shown in Figure 1-1 below.
Filtered IR radiation passes from the Analyser Unit into the sample cell, which is
fitted into the process duct. This sample cell has sintered panels, so allowing the
gas under study to permeate through the sample cell. IR radiation passes
through the sample gas, is reflected by a retro-reflector, and passes through the
sample gas once more back to the Analyser Unit. A PT100 platinum resistance
thermometer is situated in the sample cell. This provides temperature
information to the Procal 1000 series, which acts as the basis for the temperature
compensation.
The Procal 2000 has a built in Auto Verification Unit (AVU) which enables both
automatic and manual purging of the In-Situ Sample Cell.
Procal 1200
Stack
Procal 2000
Optional Outputs
110V / 240V
Local PSU
110V / 240V
Calibration Gas
The Procal 1200 is an Analyser Control Unit which is supplied with Procal 1100 (ACW) software
installed and optionally an internal Input/Output Unit Integrally mounted.
Documentation conventions
Abbreviations
Within this manual, the following abbreviations are used:
IR Infra-red
AU Analyser Unit
Lists
Often in this manual, instructions or information are presented in list form. Use
of black squares indicates that there is no special order to the instructions
or information. However, when instructions are numbered, it is important that
the individual instructions or points are followed sequentially.
Figures
Figures in the text are always numbered in the form Figure X-Y, where X is the
section number, and Y is the sequential figure number within that section. For
example, Figure 3-2 is the second figure in section 3. When a figure reference is
in brackets (Figure 9-4), this refers you to that figure, usually to confirm the
location of a component, control or indicator.
Screen displays
Screen displays on the Procal 1000 are presented in this manual as
approximations. When values are shown on screens, it should be remembered
that these appear by way of example only. It is highly unlikely that the values
displayed on your screen will be the same as in this manual.
Italics
Paragraphs in italics usually indicate background information which may be of
benefit to the reader. Groups of words in italics are usually cross-referring the
reader to a section or sub-section by name.
Dimensions
All dimensions in this Operating Manual are in mm (millimetres) unless
otherwise indicated.
Analyser Identification
Instrument Identification
Fig 1-3
In all correspondence associated with the analyser please use the serial number.
Further information
This Operating Manual, when used in conjunction with the Operating Manual
for the Procal 1000 series, should give you all the information you require to
install, operate and dispose of your PROCAL 2000 system. If you require any
further information regarding the system or its use, you should contact Kittiwake
Procal Ltd., or your Procal distributor or agent as shown in the panel below:
e-mail: post@procal.com
internet: www.procal.com
Introduction
This section describes how to install the standard PROCAL 2000 Analyser.
WARNINGS
YOU MUST NOT ATTEMPT TO INSTALL THIS SYSTEM UNLESS YOU ARE
QUALIFIED, COMPETENT AND AUTHORIZED TO WORK ON ELECTRICAL
EQUIPMENT OPERATING AT YOUR LOCAL MAINS ELECTRICAL SUPPLY
VOLTAGE.
READ THIS SECTION IN ITS ENTIRETY BEFORE ATTEMPTING TO
INSTALL ANY PART OF THE SYSTEM. IF THERE IS ANYTHING YOU DO
NOT UNDERSTAND, OR YOU DO NOT FEEL CONFIDENT OF YOUR
ABILITY TO FOLLOW THE INSTALLATION INSTRUCTIONS, DO NOT
PROCEED. CONTACT KITTIWAKE PROCAL OR YOUR PROCAL-
AUTHORISED DISTRIBUTOR.
ALWAYS ENSURE THAT YOU HAVE A VALID PERMIT TO WORK ON THE
PIPEWORK OR DUCTING TO WHICH THE PROCAL 2000 ANALYSER UNIT
IS TO BE FITTED.
Caution
The PROCAL 2000 Analyser Unit contains factory-set optical components. No
alignment or focusing of these components is required during installation or
operation. Adjustment of any mirrors, lenses or filters may invalidate the
warranty and will almost certainly prevent the system from functioning properly.
Pre-installation information
Before installing the system, you must confirm that:
Using the instrument serial number check the analyser is being installed in
the correct location.
Sufficient working space for two people to work is available around the
point at which each Analyser Unit is to be fitted. The working platform
should be adequately fenced
Clean and dry instrument air, (Procal Standard ISO 8573-1 1 2 1 see 7-5823-
00 in Appendix C) at up to 10 l/min (intermittent) and 6.0 barG maximum
(nominal inlet pressure 2.5 to 3 barG) is available adjacent to where the AU
will be fitted
The cable route from Analyser Unit to Procal 1000 is not more than 1,000
m (greater distances may be possible: please consult Procal)
200mm
Instrument Identification (8”)
and serial number Minimum
Stack
Flange
Stack
Insulation
M16 Studs
50mm
Insulation
Un‐acceptable
2. Remove and retain the Four 5 mm pozi drive screws securing the
cover of the Junction Box on the side of the GRP cover.
Figure 2‐3
4. Locate the relevant cable as identified in the wiring tables. Gently feed this
end of the cable through the cable gland underneath the AU junction box.
You should leave a 'loop' in the cable so that the AU can be withdrawn from
the process duct without disconnecting the cable.
5. Repeat this procedure for all cables as indicated in the wiring table relevant
to the installation.
Figure 2-4
Wiring Table
Option 1 - Analyser → Procal 1200 (ACU MK 4) with Remote Power Supply Unit (PSU) (see drawing 7-2498)
Procal 1200 (ACU MK 4 Cable Specification
Function Analyser Terminal Box
+ IOU)
Data Return B TB1 – 1B TS6 - 1 2 Twisted pair with O/All screened Znom
50-120ohms Belden 9842 or equivalent
Data Return A TB1 – 2B TS6 - 2
Data Out B TB1 – 3B TS6 - 3
Data Out A TB1 –4B TS6 - 4
Screen Not Connected TS6 - 7
Option 2 - Analyser → Procal 1100 (ACWn) with Remote Power Supply Unit (PSU) (see drawing 7-2103)
Procal 1100 (ACWn) (1- Cable Specification
Function Analyser Terminal Box
2773) (TX100I)
Data Return B TB1 – 1B RX-(A) O/All screened Znom 50-120ohms Belden
9842 or equivalent
Data Return A TB1 – 2B RX-(B)
Option 3 Analyser → Power Supply Unit (PSU) 1-1971 → Procal 1200 (ACU MK 4)
Analyser Terminal Box Power Supply Unit (PSU) 1-1971 Procal 1200 (ACU MK 4)
Cable 1 Cable 1 Cable 2 A/B Cable 2A Cable 2B
Function ACU ACU
AU IN Mk4 + Mk4 2 Twisted
1OU (A53) pair with
O/All
Data Return B TB1 – 1B TB1 – 7B 6B TS6 - 1 TB-4
screened
Data Return A TB1 – 2B TB1 – 10B 9B TS6 - 2 TB-3 Znom 50-
120ohms
Data Out B TB1 – 3B TB1 – 13B 12B TS6 - 3 TB-2 Belden
9842 or
Data Out A TB1 – 4B TB1 – 16B 15B TS6 - 4 TB-1 equivalent
Screen Not Connected TB1-3B N/C TS6 - 7 TB-5
+24 V TB1 – 26B TB1 – 5B N/A N/A 2 core
1.5mm2
0V TB1 – 29B TB1 – 4B N/A N/A unscreened
Option 4 Analyser → Power Supply Unit (PSU) 1-1971 → Procal 1100 (ACWn)
Power Supply Unit (PSU) 1- Procal 1100 Cable Specification
Analyser Terminal
Function 1971 (ACWn) (See 7-
Box
AU IN 2103) (TC100I)
Data Return 2 Twisted pair with
TB1 – 1B TB1 – 7B TB1 – 6B RX-A
B O/All screened Znom
Data Return 50-120ohms Belden
A
TB1 – 2B TB1–10B TB1 – 9B RX-B 9842 or equivalent
Data Out B TB1 – 3B TB1–13B TB1 – 12B RX-A
Data Out A TB1 – 4B TB1– 16B TB1 – 15B RX-B
Screen Not Connected TB1– 3B Not Connected SGND
+24 V TB1 – 26B TB1 – 5B N/A 2 core 1.5mm2
unscreened
0V TB1 – 29B TB1 – 4B N/A
Table 2-1
All unused glands must be sealed before commissioning the analyser
Analyser Covers
Procal analyser’s are fitted with two basic types of Cooler / Heater however there are the following
exceptions:
This type can be identified by the presence of a 100mm spigot protruding from
the base of the analyser cover.
Spigot
In‐Situ Probe
Under Side of Fan Cooled Electrically Heated PROCAL 2000 AC/AVU-F-24V Analyser Cover
Figure 2-5
The cover is fitted with an electrical heater and fan which enables the analyser
to operate in ambient temperatures of -30 ºC to +50 ºC. The heater / cooler
function is controlled by the AU, the set point is factory.
Forced air from an integral low voltage electrical fan is channelled past the AU
to draw heat away from the analyser. The Control of the fan is by the Analyser
Unit and cooling capacity is dependent on the ambient air temperature.
Services required – None additional to an AU. Note the Fan requires an
additional 0.8 amp allowance on the 24 Volt power supply. The Heater requires
an additional 2.5 amp allowance on the 24 Volt power supply, Fan and Heater
are not on at the same time.
As the Cover insulates the AU from the ambient air, then if the Fan is disabled
the AU can self-heat to a higher than normal temperature. It is advisable to
configure the system to have a high alarm on the head temperature of +55ºC
and to remove the AU power if this limit is exceeded.
The cover is fitted with a 100mm spigot to enable a flexible hose to be
connected allowing cool air to be ducted to the analyser if installed in a high
ambient temperature environment. Note the analyser will be delivered with a
300mm length of hose if an extended hose is not to be fitted then this should be
attached.
The cover also incorporates as standard an Auto Verification Block internally
piped and wired. The function of Auto Verification Block is described later.
In‐Situ Probe
Under Side of Vortex Cooled Electrically Heated PROCAL 2000 AC/AVU-V-24V Analyser Cover
Figure 2-6
Figure 2-7
Both types of covers are fitted with wired and piped Auto Zero / Verification Units to facilitate
automatic zero and verification / calibration routines. It is mandatory that the analyers is configured to
instigate automatic zero’s every 12 or 24 hours depending on application.
Note: The following services are required:
Clean dry instrument air (Procal Standard ISO 8573-1 1 2 1 see 7-5823-00
in Section 7. page 7-2) at pressure of 1 barG above the process gas
pressure, up to a maximum 6.8 barG. The flow-rate available should be
between 0.5 l/min (constant) and up to 10 l/min (intermittent during auto-
zero/auto-purge)
Note the air supply fitted with an appropriate regulator, filter and isolation valve should be
located within 1 meter of the AU.
100mm
Spigot
Figure 2-8
The Auto Zero / Verification function is achieved by activating a series of solenoid valves mounted on
a manifold block. The block is mounted on an access plate which is secured to the Cover by four M4
socket headed screws.
The zero (Instrument Air) and test gas flow rates are controlled by a need valve mounted on the
manifold block. The correct flow is set during the analyser manufacture and therefore there should be
no need to readjust.
The slotted screw flow valve is located on the Auto Verification Unit manifold accessible by removing
the AU outer case.
Figure 2-9
Note: It is important during commissioning that the Zero / Cal Gas flow rate is verified as
process conditions have to be taken into account. To verify that the flow rate is correct the
pressure signal during zero should be approximately 20mbar above process conditions.
If required the flow rate can be adjusted by turning the needle valve mounted within the manifold.
Void Purge
The space in the reflector tube referred to as the void is continuously purge,
the flow rate is controlled using a orifice therefore no adjustment required.
Analyser Purge
The cast aluminium enclosure containing the main optical components is also
continuously purged, the flow rate is controlled using a orifice therefore no
adjustment required.
Introduction
Introduction
This section is intended for those people who need or would like to know how the
Procal 2000 analyser works as a measuring instrument. Understanding the
technical aspects of the analyser and the theory behind its operation are not
necessary for installation and routine operation.
(CM-1)
5000 2500 2000 1500 1000 700
100 0
80 .1
.2
Transmission%
60
.3
40 .4
.5
Carbon .6
20 Dioxide .8
CO2 1.0
1.5
2 4 6 8 10 12 14
Wavelength(μm)
200
160
Absorbance mAu
120
80
40
0 5 10 15 20
CO2 (%)
Most process gas analysers - including the PROCAL 2000 - operate at a number
of fixed wavelengths, which are chosen according to the gases to be measured.
Within the analyser is an IR source which emits broadband IR radiation. An IR
detector detects the radiation which has passed through the sample gas. With a
single gas, it should be a straightforward matter to measure absorbance, identify
the gas and calculate concentration. In practice, however, there are usually
several gases. More than one gas may exhibit absorption at a given wavelength.
This effect is termed 'cross-sensitivity', and may be compensated for such that
accurate concentrations of individual gases may still be displayed.
In the PROCAL 2000, interference filters are placed in the path of the broadband
IR radiation. Two filters are used per component. One selects a measuring
wavelength, while the other selects a reference wavelength. The filters are
mounted on a rotating disc immediately in front of the IR source. This
arrangement gives rise to an IR beam alternating rapidly between measuring
wavelength and reference wavelength for each component. This alternating
beam passes through the gas to be measured, eventually striking the IR detector.
Filters have been selected such that if any of the gas to be measured is present,
then a significant amount of radiation will be absorbed at the measuring
wavelength, with very little or none at the reference wavelength.
The quantity of radiation per unit time is a function of two factors: (1) the
number of molecules within the optical path and (2) the extent to which these
molecules absorb radiation, expressed as the absorption coefficient.
It is these two signal strengths which are shown on the TEST screen under the
titles measured value and reference value.
Cross-sensitivity
The technique effectively increases the resolution of the interference filters used,
which typically have a bandwidth of 2% of their peak transmission wavelength.
In terms of spectroscopy, this is quite broad. Figure 4-3 shows the structured
spectrum of HCl with sharp, narrow bands which are too narrow to be well
matched to the interference filter. The absorption, which is sensed when a filter
on its own is used, is the average absorption rather than the much higher peak
absorption of the HCl gas.
One way of increasing the selectivity of gases when using interference filters is
to use sealed cells filled with the gas under study. Kittiwake Procal has
produced sealed cells filled with HCl, NO, Methane and CO, and is likely to add
other gases to this range. To show how these sealed cells work, refer to Figure
4-4. A matched pair of filters is used with both filters being as similar as
possible in their transmission and wavelength characteristics.
If they were used individually, they would produce identical reference and
measurement pulses at the IR detector, and both would react to the presence of
the gas being measured. The example given is for hydrogen chloride, HCl.
However, if a gas cell filled with HCl is put in series with one of the filters, the
pulses become unbalanced and the pulse caused by the reference filter becomes
weaker. When HCl gas enters the sample tube of the PROCAL analyser, the
measurement pulse falls much more than the reference pulse. The reason for this
is that the reference pulse has been filtered by the sealed cell, so the sharp
narrow bands of wavelength at which HCl absorbs are absent. Therefore, further
absorption by the HCl is small. The measurement pulse, on the other hand,
reacts to HCl in the normal way. Therefore, the introduction of the sealed gas
cell has sensitised the pair of matched filters to HCl.
When interfering gases - such as water vapour - come into the sample cell, both
pulses reduce in almost the same proportion. This is because the water vapour
absorption is broadband and does not follow the same pattern as the HCl
absorption. It therefore causes absorption of both pulses. The method of signal
processing divides the intensity of the measurement pulse by the intensity of the
reference pulse before calculating the absorbance. Therefore, equal,
proportionate changes which occur to both pulses are cancelled out. This shows
that substantial immunity to water vapour and other interfering gases can be
achieved in this elegant fashion.
There are further advantages in the use of gas filter correlation, as opposed to the
conventional use of reference wavelengths. Since the measurement and reference
wavelengths are the same, and as they come from matched filters and only differ
by the filtering caused by the gas in the sealed cell, any effect due to temperature
change of the source also cancels out. When different wavelengths are used
there can still be some effect due to temperature fluctuations in the infra-red
source. The signal-noise ratio has been raised by the use of gas filter correlation,
permitting the measurement of smaller amounts of gas. Kittiwake Procal uses
gas filter correlation in combination with software, such as PROCxsens and
matrix algebra, which enable complete correction to cross-sensitivity to be
achieved, even eliminating the small cross-sensitivities that remain after gas
filter correlation cells are used. In practice, an improvement in the immunity to
water vapour of more than a order of magnitude is achieved. In the case of
measuring HCl - one of the most demanding analyses - the main improvement is
in the immunity to the interference effects of hydrocarbons.
R M
Matched
Filters
Detector
signal
R M
HCl
Gas filled
cell
Detector
signal with
zero gas R M
Detector
signal
H20 M
R
introduced
Detector
signal
HCl
introduced R M
Analyser Unit
The Analyser Unit is shown, with its cover removed, in Figure 4-5 below. All
the principal components are as notated on the illustration, and described in text.
Figure 4-5 PROCAL 2000 Analyser Unit with inner and outer cover removed
Once the IR beam has passed through the filter, it passes through the head exit
lens and into a void that is purged with a non-absorbent gas. This void acts as a
spacer between the AU and the sample cell so that the latter can be inserted fully
into the duct, thus avoiding stratification of the gas which can occur near the
wall of the duct. The lens is normally made of calcium fluoride. For
measurement of longer wavelengths, a zinc selenide lens is used. Lens materials
are selected for their good IR transmission characteristics and their inert nature.
When the IR beam has passed through the sample gas and been returned to the
AU, it is incident on the detector. This detector is a high-performance solid state
device working on the pyro-electric principle. Radiation striking the detector
causes minute temperature changes which are translated into electrical signals.
These signals are proportional to the level of irradiation, and are largely
independent of wavelength.
Within the AU is the intelligent head pcb. This has a number of functions. It
amplifies the signal from the IR detector, and passes this to the ACU. It supplies
and regulates DC power to all parts of the AU. Lastly, it amplifies and transmits
the temperature signals from the two temperature sensors in the sample cell and
the AU respectively.
The integral Auto Verification Unit has two primary & three secondary
functions.
Primary Functions
During this Auto-zero procedure, all displays and outputs are frozen
at the last 'on line' levels before commencement of the Auto-zero
routine. This includes the time required for the sample cell to purge
2 In the event of a power failure to the system, the In-Situ Heater element (when
fitted) and the sample cell will cool, possibly to below the dewpoint.
Condensation could then occur on the process side of the sample lens. The de-
energised solenoid valve on the Auto Verification Unit opens and floods the
sample cell with dry instrument air, thus removing the possibility of
condensation. The unit will continue to purge the cell until the fault condition is
rectified.
Secondary Functions
1 Optionally if the system is set up for Auto Calibration the Auto Verification
Unit can switch on 1 of 2 Calibration Gas lines in the required sequence under
control of the Procal 1000 to facilitate Span Calibration or verification.
2 when the sample temperature falls below a preset level, typically 120°C.
Introduction
Caution
While the instrument is accurately aligned and calibrated as part of the
standard production procedure, it is possible that components may
become misaligned due to prolonged vibration in operation. If
misalignment is suspected, the user should not attempt to re-align the
instrument. Incorrect realignment will invalidate the calibration. Call
Kittiwake Procal or your Procal distributor.
If the user does not have a service contract and decides to undertake routine
preventative maintenance in-house, the following schedules should be strictly
followed.
Please note that if the system is not properly maintained, the manufacturer's
guarantee may be voided or limited.
access the TEST screen on the Control Unit, and check all operational data
displayed
keep a record of the signal readings so that any deterioration due to lens
clouding or electrical faults can be quickly identified.
visual inspection of all cables and gas connections for integrity, and for cracks,
corrosion or any other signs of physical damage.
calibration (see under the heading Field calibration check within this section);
note that calibration may
need to be performed more frequently if required by regulatory authorities
Annual schedule
internal visual inspection of Analyser Unit for signs of ingress of dust, moisture
or process fluids and/or gases
check that the source is emitting and that its lens is clean
check the filter wheel motor (when unplugged) for ease of rotation, and replace
the motor if any friction is apparent
visual inspection of all cables and gas connections for integrity, and for cracks,
corrosion or any other signs of physical damage.
If the results of any of the above checks cause you any concern, contact
Kittiwake Procal, or your Procal distributor or agent, immediately.
Cleaning
Please do not attempt to clean any of the optical surfaces within the instrument.
Contact Kittiwake Procal, or your Procal distributor or agent, for a cleaning
service. If you attempt to clean optical surfaces and cause damage, the warranty
on the PROCAL 2000 system may be voided or limited.
Introduction
The gases used in the original factory calibration are of known accuracy and the
calibration achieved using them should not be discarded lightly.
Note that the Routine maintenance schedule within this section calls for an
annual Field Calibration Check.
Calibration Equipment
The calibration gases must be in cylinders that can be located close to the AU.
Each cylinder must be fitted with a regulator. It is very important that each
cylinder has a supplier certificate of calibration, since experience has shown that
large variations in actual concentration can be encountered.
It is important that the person carrying out the calibration should be aware that
he/she is dealing with toxic gases and carries out all safety procedures associated
with the handling of high-pressure gas cylinders.
pressure gauge with 1/8 BSP taper male fitting (Pipe thread EN 10226 R1/8);
scaled 1 barG FSD
3 Unscrew the sintered muffler fitted at the Void purge outlet. Screw to the Void
purge outlet the pressure gauge with the 1/8 BSP taper male fitting, sealed with
PTFE tape.
4 Identify the void air feed connection on the auto verification unit and disconnect.
Ensuring the restrictor is still connected to the auto verification unit. Insert the ball
valve between the auto verification unit void air feed and the void pipe ensuring the
ball valve is in the off position.
6 Open the ball valve until the gauge reads approximately 1 bar, and then turn the ball
valve to the off position.
7 After 10 minutes check that the gauge reading has dropped by less than 5%.
If the drop has been greater than this, check that the pressure gauge fitting is
not leaking and then repeat the test. If the test fails again it is likely that the
sealing of the centre lens is defective. In this case, the analyser will have to
be dismantled and the centre lens re-sealed.
9 Remove the gauge from the Void Purge port and refit the sintered muffler.
Remove the ball valve and reconnect the void purge pipe.
10 Ensure all connections disturbed during the test are correctly fitted.
Observe the stack pressure under normal operation and note the reading (this is
found on the TEST SCREEN in the Analogue and derived Channels Table and
is labelled as P SAMPLE). From the AUTOZERO drop down menu select
MANUAL AIR PURGE, wait for the pressure to stabilize and then note the
reading. If the increase in pressure is greater than 50mbar (0.72 psi), the sintered
panel is partially blocked.
Gas Calibration
1) On the Test Screen of ACWn select START AUTOZERO NOW from the
AUTOZERO drop down menu, and allow the analyser to go through the whole
cycle, noting pressure, abs and concs readings. Upon completion of Auto-zero,
select MANUAL AIR PURGE from the same drop down menu, wait for the
readings to stabilise and then record the results. Readings should be as close to
zero as possible, to provide a baseline for calibration.
2) Connect the outlet of the gas cylinder via the necessary fittings and regulator to
the AVU CAL port to be used located on the underside of the AU outer cover
(Usually Port 4 – CAL gas 1). From the AUTOZERO menu (fig.1) select MAN
CAL GAS 1, open gas regulator and set pressure to approx 1.9barG/27psiG.
3) To ensure that all water vapour and auto-zeroing gas has been driven out of the
sample cell wait for at least three minutes before taking any readings. Observe the
TEST screen and note the reading for water vapour, which should have reduced to
a level of less than 2% of the abs reading during normal operation (e.g. if abs
during normal operation is 100, reading should be <= 2). If it is higher, slightly
increase the gas flow.
4) After allowing enough time for the calibration gas in question to stabilise (around
3 minutes) record the readings. If the displayed concentration reading is within
±2% of actual bottle value no calibration is necessary and the user can go to Step
6.
5) If the display shows a calibration reading that is >±2% of the gas bottle,
recalibration is required using steps 7 to 9.
6) Turn off the gas regulator and gas bottle and allow regulator to drop to 0, then
select MANUAL AIR PURGE from the AUTOZERO menu (fig.1). Repeat steps 2
to 5 AFTER readings on TEST screen have stabilised around 0.
9) The software automatically recalculates the coefficients for the given gas so that
returning to the TEST screen should give the correct reading. An indication in the
information window will display a successful calibration. To return to the TEST
screen simply select the EXIT button shown in figure 3. Figure 4 shows the
selection of Hydrogen Chloride from the drop down menu, and a successful
calibration.
10) When all necessary channels have been calibrated disconnect the cylinder from
CAL Gas port, and select MAN AIR PURGE from AUTOZERO menu (fig.1).
After taking zero readings select MANUAL AIR PURGE again from
AUTOZERO menu to turn off air (MANUAL AIR PURGE should NOT have a
tick next to it). This returns the analyser to the normal process operation.
A water vapour calibration should only be performed in the field if there is good
reason to doubt the readings obtained. This is because laboratory conditions
CANNOT be repeated in the field. A calibration carried out in the field IS NOT as
accurate as one done in the factory.
2) Using the ACU options menu, or ACWn Main screen selection turn off the Auto
Zero function
3) Set the Calibrator to give a level of 80% of the FULL RANGE water vapour.
4) Monitor the H2O readings and wait for them to settle (approx 5-10 minutes)
6) If the displayed concentration reading shows a deviation of >±5% from the set
water vapour concentration recalibration should be carried out by following steps
7 to 10.
7) If operating with ACU use the SERVICE menu (described earlier in the manual)
to select Calibration option, if using ACWn, select MANUAL CALIBRATION
from the TEST SCREEN menu, and then CALIBRATE COMPONENTS (fig.2)
8) The analyser will average the water vapour reading for 1 minute and then display
it as shown in the calibration graph. If the set concentration figure is significantly
different from the previously observed average, then it is likely that one of the
surges occurred during the 1 minute averaging time. In this case, repeat steps 3 to
8 of this procedure.
9) On the ACU adjust the required concentration to read the same as the value set on
the calibrator, with ACWn check the Measured value against the Expected value
10) On ACU select calibrate option and apply, on ACWn click the Adjust button. The
analyser will immediately recalculate the calibration table values for water vapour
so that on return to the TEST screens the correct value should be displayed
11) Disconnect the KITTIWAKE PROCAL 9000 Field Calibrator and replace the
“Hot Access” blanking plug
12) Using ACU turn on the Auto Zero via the options selection, with ACWn turn on
Auto Zero via AutoZero selection (fig.1)
Fault-finding
WARNING
You must not attempt to FAULT-FIND OR REPAIR this system unless
you are qualified, competent and authorized to work on electrical
equipment operating at your local mains electrical supply voltage.
2 If the TEST screen shows low readings, check for condensation on, or
contamination of lenses and mirrors.
3 The quality of the IR signal may be poor. Check with an oscilloscope that the
waveform shown in Figure 5-2 (a) appears at TP1 on the OHU pcb (with ground
connection to E). It is important that the observed waveform is as smooth as in
Figure 5-2.
+5V DC
0 volts DC
-4V DC
16.7 ms
133 ms
5V–
0V–
Repair policy
Kittiwake Procal recommends that repairs to the PROCAL 2000 system are only
made by its own trained support staff, or by those of its distributors world-wide.
If you have the necessary technical qualifications, training and experience you
may wish to make straightforward repairs in-house. A spares list is given for
this purpose. However, you should note that if a repair is incorrectly carried out,
this may void or limit the warranty on the system. You should also note that this
manual is not intended to describe fault-finding or repair down to component
level.
WARNING
You must not attempt to FAULT-FIND OR REPAIR this system unless
you are qualified, competent and authorized to work on electrical
equipment operating at your local mains electrical supply voltage.
Spares list
The figure in the right-hand column is the quantity of that part fitted to the
analyser.
If you wish to order a spare part, please contact Kittiwake Procal, or your Procal
distributor or agent.
If you require a spare part that is not listed here, please contact Kittiwake Procal,
or your Procal distributor or agent.
When ordering spare parts, please quote both the Part Number and the serial
number of the part of your system concerned - AU, Procal 1000, or ISH. This
will provide you with a quick and efficient service.
Kittiwake Procal Ltd. reserves the right to change the prices, specifications or
manufacture of any component parts at any time.
Certain components can be changed by the user within the PROCAL 2000
system, when necessary.
The IR source has a replaceable filament, part number 1-2716J, that is the only
part which needs replacing. Provided the protective air purge is maintained
through the analyser during the time that it is switched off, it is possible to
change an IR source filament without withdrawing the AU from the process
duct.
1. Switch off the mains supply to the local PSU. Check that the air purge
has automatically switched on.
3. Remove and retain the eight 5 mm socket head screws securing the
inner cover of the AU to the AU case.
6. Unplug the two-pin (Berg-type) pcb connector on the end of the cable
from PL1 on the optical head pcb. Figure 5-4 shows the position of
PL1.
PL 1 PL 2 PL3
Detector Not Used Sample TS1
Temp. 1
2
PL 4 3
Filter Wheel 4
PL1 5
Source 6
PL 4 7
Transition 8
PCB 9
10
11
12
13
14
15
16
7. Loosen the two pan head M3 screws at the top of the integral source
that hold the filament in position until it is possible to rotate the
filament through 90o and withdraw it. Note the orientation of the
filament flange as it is withdrawn.
8. Taking great care not to damage it, insert the new filament carefully
into the hole in the top of the Integral source observing the same
orientation as when the old filament was withdrawn.
9. Rotate the inserted filament through 90o so that the elongated portion of
the flange is positioned under the M3 screws, and then tighten them.
10. Restore mains power to the Local PSU and check that the filament can
be seen glowing inside the Integral source.
11. Replace the AU inner cover and tighten the eight fixing screws.
Note: Select manual air purge. When readings have stabilised de-select manual air
purge, the perform Auto zero now.
3 Remove and retain the eight M6 socket head screws securing the inner
cover of the AU to the AU case.
6 Unplug the FE PCB, remove and retain the two M3 socket head screws
and washers securing the filter wheel pcb to the bracket, then place the
pcb to one side (taking care not to strain the ribbon cable connected to the
pcb).
7 Note the connector to which the motor cable is connected, and the
orientation, then disconnect the cable.
8 Remove and retain the two M3 counter sink screws that secure the filter
wheel bracket to the base plate then withdraw the assembly from the AU.
9 Using a 1.5 mm Allen key, loosen the screw holding the filter wheel on to
the motor spindle and withdraw the filter wheel. This contains a number
of fragile optical filters so should be placed where it is protected from
damage.
10 Locate the three small pan head motor fixing screws on the newly-
exposed face of the bracket. Remove and retain these screws, then
remove the motor.
11 Fit the new motor to the bracket using the three small pan head screws
and tighten them.
12 Fit the filter wheel on to the motor spindle with the wheel boss facing the
motor. Leave the centre fixing screw fairly loose at this stage.
13 Screw the bracket to the base plate with the two retained sets of screws
and washers. Rotate the filter wheel by hand and check that it rotates
freely, tighten shaft grub screw.
14 Fit the Filter Wheel PCB to the top of the bracket using the retained
screws and washers.
Introduction
It is important that, at the end of their working lives, all parts of the PROCAL
2000 are disposed of in accordance with all local environmental statutes and
regulations which have the force of law.
If you are ever in any doubt about how to dispose of the PROCAL 2000, you
may return it to Kittiwake Procal in the United Kingdom, carriage-paid.
Kittiwake Procal will dispose of the system for you. There will be no charge for
this service, but neither will you receive any payment or allowance for any
components or materials recovered.
This manual is an integral part of the PROCAL 2000 analyser system. If you
pass the system on other than for final disposal, you should always pass this
manual on with it.
WARNING
YOU MAY ONLY DISPOSE OF ZINC SELENIDE OPTICS BY
DELIVERING THEM TO AN ENVIRONMENTAL WASTE DISPOSAL
AGENCY THAT ACCEPTS SUCH MATERIAL OR BY RETURNING THE
OPTICS TO KITTIWAKE PROCAL IN STRONG PROTECTIVE
PACKAGING.
DISPOSAL BY ANY OTHER MEANS MAY LEAD TO DAMAGE TO
HUMAN HEALTH.
Introduction
This section contains summary installation diagrams and information specific to the
PROCAL 2000 series system with which you have been supplied. The following
standard drawings of the version current at date of publication are included as a
guide only.
Drawing Title
For full installation details and for installation drawing versions applying to your
specific delivery refer to the Drawing Package referenced below as delivered with
the equipment.
Introduction
To achieve this, the equipment has undergone rigorous tests at the Baseefa
authority against the British and European Standards BS EN 60079-0:2004 and BS
EN 60079-1:2004
These tests and QA in production ensure that the equipment cannot cause an
explosion when operated correctly in the stated hazardous areas.
To comply with the requirements of the approval, several steps have to occur
during production under the terms of the licence M1655 held by Kittiwake Procal
Ltd.
(a) The materials used in safety critical parts are supplied with Certificates of
Conformity
(b) Certain mechanical parts have to be inspected to drawing to ensure that critical
dimensions are correct and records are kept.
(c) The optical head casting is subject to a pressure of 10.65 bar (g) for a minimum of
10 minutes without distortion.
(d) The thermopocket of the reflector tube is subject to a pressure test of 47.5 Bar
for 1 minute in line with the certified product requirements. No leaks, distortion
or deformation are accepted.
(e) The completed analyser is subject to final inspection against the overall approval
drawing.
Allowed environments
The standards above detail the gases, and environments in with which
the ATEX approved equipments are allowed to be used. Reference
may be made to them.
The PROCAL 2000 (Formally Pulsi 200)) may be used for potentially
explosive gas atmospheres classified as IIB
The PROCAL 2000 may be used for the ambient temperature range
–20°C to +40 °C
Note the Procal 2000 is fitted with a Vortex cooler maximum surface
temperature is 135ºc therefore classification is T4.
When ordering spare parts for the PROCAL 2000, please state to Kittiwake
Procal or their distributor its serial number and whether it is for use in an area
with an explosive gas atmosphere.
All glands and hole blanking plugs including any replacement must have the
same thread form as the original items, and bear the ATEX mark for flameproof
enclosure use.
The joint surfaces of the enclosure and its cover shall be wired clean before re-
assembly. If there is any damage to these surfaces consult Kittiwake Procal Ltd
or their distributor. All 8 stainless steel fixing bolts size M6 X 16 must be used
and they shall have a mark A2-70. Missing or damaged ones must only be
replaced by screws ordered from Kittiwake Procal Ltd.
Laws and regulations about flammable atmospheres that apply to the customer
should be observed in addition to these ATEX requirements.
Introduction
Input Output Units can be added to the Procal Instrument Serial Cable to provide
addition inputs, current outputs and relay outputs to any system.
Each Unit can provide up to 32 Input or Output Modules (mixed allocation), has up
to 32 Relays and provides 16 Digital Inputs.
The controller software can allocate any channel to any output, any alarm to any
relay and configure any input as an additional channel.
All Input Modules and Output Modules are fully isolated. The Relay outputs are
Voltage Free and the Digital Inputs are isolated in voltage or current mode.
The addressing of the Units by the controller does not affect the system ability to
address Analyser Units.
The IOU is mains (line) powered and can be used in place of a Local Power
Supply in certain configurations.
Services required
WARNINGS
YOU MUST NOT ATTEMPT TO INSTALL THIS SYSTEM UNLESS YOU
ARE QUALIFIED, COMPETENT AND AUTHORIZED TO WORK ON
ELECTRICAL EQUIPMENT OPERATING AT YOUR LOCAL MAINS
ELECTRICAL SUPPLY VOLTAGE.
READ THIS SECTION IN ITS ENTIRETY BEFORE ATTEMPTING TO
INSTALL ANY PART OF THE SYSTEM. IF THERE IS ANYTHING YOU
DO NOT UNDERSTAND, OR YOU DO NOT FEEL CONFIDENT OF
YOUR ABILITY TO FOLLOW THE INSTALLATION INSTRUCTIONS,
DO NOT PROCEED. CONTACT KITTIWAKE PROCAL OR YOUR
PROCAL-AUTHORISED DISTRIBUTOR.
If this has not already been done, unpack the system components and check that
they correspond to the units ordered and listed on the accompanying packing
note. If there is any discrepancy, or any damage is apparent, do not attempt to
install the system. Contact Kittiwake Procal, or your Procal distributor.
Mount the IOU in the control room, shelter or other designated location securely
mounted by its locating brackets. Mounting details are shown in Fig. 1. The
IOU weighs 11kg (24lb). The IOU should be mounted in the chosen
configuration before wiring is attempted.
WARNING
DO NOT CONNECT THE MAINS CABLE (POWER CORD) TO THE
ELECTRICAL SUPPLY UNTIL ALL CONNECTIONS WITHIN THE IOU
HAVE BEEN MADE. FAILURE TO FOLLOW THIS WARNING MAY
RESULT IN SERIOUS OR FATAL INJURY FROM ELECTRIC SHOCK.
The typical cables used to connect other system components to the IOU are
shown in Figure 4a and 4b.
When the IOU is wall mounted, cable glands are recommended to be on the
underside of the unit.
110 V ac nominal (low limit 90 V, high limit 132 V), 47-63 Hz.
230 V ac nominal (low limit 180 V, high limit 264 V), 47-63 Hz.
1 Ensure that all wiring between the IOU, AUs, Control Unit and ancillary
equipment is correct. To connect the above power supply the Terminal Block
Cover will need to be removed, ensure this is replaced before continuing to
power the unit.
2 Connect the free end of the mains cable to a suitable ac supply. Follow local
regulations for permanently installed equipment. It is recommended as good
practice that a means of disconnection is provided within 2 meters and an RCD
or ELB is also fitted.
The installation is now complete and the system is ready for operation.
System configuration
The System Database provided will be configured to suit the Inputs and Outputs
configurations ordered. If changes need to be made consult the relevant section
in the Controller Unit Manual.
Module configuration
See IOU Configuration Sheet (7-7806) form entries for this delivery Module
setup details.
See IOU Customer Connections drawing 7-4384 for Module locations, Relay
Links and Digital Input Link locations.
Operation
CPU Board 2-1117 has LEDs which indicate power on and program running.
LED2 and LED3 are blank, used by Procal in programming the board.
AU Power connections
The system communications connectors TS5 and TS6 .can also, by use of the
Links 1,2,3,4,5 and 6, either supply 24V or be used to feed an existing 24V
supply to the next unit on the cable.
Links 1 and 2 IN will connect the IOU 24V power supply to TS5 pins 5 and 6.
If the Links 1,2,5,6 are removed and Links 3 and 4 are fitted then TS5 pins 5 and
6 are connected to TS6 pins 5 and 6 only, allowing an external supply to be
connected through the unit.
The IOU can supply 3.5A at 24V for powering Procal AU instruments only.
Introduction
It is essential that the air used for optical purge, void purge and auto zero is of the
required standard. This section describes the quality of air supplied. If in doubt it
essential that the Quality of Air is tested to confirm that it meets the required
standard as stated below.
If the air is shown to be inadequate please install an Air Preparation Panel (see
appendix L).
7-5823-00 issue 03
Quality Standard for Instrument Air Supplied to Proca Analysers
The quality of instrument air should meet the Class 1 standard given in ISO 8573-1
Lubricant content should in no circumstances exceed 1 part per million by weight or volume. – Very
approximately ISO Class 4. (< 5mg/m3 or < 0.006 ppm) being the maximum class size quoted
Hence to achieve the ISO standard additional air conditioning equipment is required.
Introduction
Some PROCAL 2000 systems are supplied with an optional In-Situ Heater
(ISH). This Appendix describes this option, and details how to install it.
An optional In-Situ Heater (ISH) may be fitted around the sample cell, and is
recommended for applications involving flue gases at temperatures below
120°C, or on applications with high acid dew point or where monitoring is
required when process is off. The ISH contains an electrical heater, which has
the effect of keeping the temperature of the sample cell well above the dew
point, thus preventing condensation onto any optics in contact with the process
gas.
WARNING
DO NOT CONNECT ANY MAINS CABLE (LINE CORD) TO THE
ELECTRICAL SUPPLY UNTIL ALL CONNECTIONS WITHIN THE ISH
HAVE BEEN MADE AND THE UNITS ARE SECURELY CLOSED ONCE
MORE. FAILURE TO FOLLOW THIS WARNING MAY RESULT IN
SERIOUS OR FATAL INJURY FROM ELECTRIC SHOCK.
Caution
Do not attempt to connect any mains cable (line cord) to the ISH if the
voltage configuration of the ISH differs from your local electrical supply
voltage. Contact Kittiwake Procal, or your Procal-authorised distributor.
Using an incorrect supply voltage will result in the In-Situ Heater failing to
function properly.
Sample Inlet
Sample Outlets
1. Determine voltage of supply 110V or 240V (Others consult Kittiwake Procal Ltd or
Distributor)
2. Remove and retain the four screws securing the cover of the ISH junction box.
3. Feed through one end of the ISH power cable through the cable gland in the ISH
junction box.
4. Connect the power cable: 110V (Note colours may change depending on local code)
Brown (live) Terminal 1 on SSR
Blue (neutral) TB 2 (Retain link)
Green/yellow (earth) to the earth stud
5. Connect the power cable: 240V (Note colours may change depending on local code)
Brown (live) Terminal 1 on SSR
Blue (neutral) TB 1 (Discard link)
Green/yellow (earth) to the earth stud
7. Connect the AU control to the ISH SSR wiring the control cable through the other
gland.
a. SSR + to AU Terminal Box T1 – 13
b. SSR - to AU Terminal Box T1 – 15
8. Ensure the control cable is securely connected and with adequate clearance from the
mains supply terminations.
9. Tighten all glands and replace the cover securely, ensuring the cover earth wire is
fitted.
During installation of the In-Situ Heater, you may need to refer to the relevant
installation drawing in the Order-specific information section of this manual,
particularly if you will be making up your own interconnecting cables.
1 Fit the supplied circular gasket to the four M16 bolts on the vertical flange
already fitted to the process gas duct.
2 Offer up the ISH to the vertical flange. Introduce the heater gently into the duct
and fit the ISH flange onto the four M16 bolts.
3 If the AU is not to be fitted immediately, fit nuts to the four bolts, and finger-
tighten to retain the ISH safely.
4 Once the In-Situ Heater has been fitted, you may go on to fit the Analyser Unit.
If the set point shown on the Procal 1000 TEST screen is consistently below
120oC, then it is likely that the ISH is not working properly. Take the following
actions:
2. Measure the voltage between TB1-13B and TB1-15B on the AU junction box.
This should show a steady or pulsing 24 V dc.
2 Check that all the connections between the AU and the Control Unit are properly
made.
3 Confirm that a steady or slowly switching mains voltage appears between the
L and N terminals in the ISH mains connection box.
Introduction
To satisfy the environmental regulations in the USA and some other countries, it is
necessary to both check and demonstrate that the analyser is within its quoted
calibration accuracy. This requires that the Autozero facility should be augmented
by an automatic check of the span calibration of the analyser.
Once a day the analyser zero and span accuracy is to be checked and recorded.
This record is maintained on a rolling 7-day basis, and the record for the previous
seven days should be available for viewing and printing at any time on demand.
This appendix describes the implementation of US EPA 40 CFR Part 75 in the
Procal 2000. It must be read in conjunction with the main technical manual.
Services
ZERO GAS Clean, dry instrument air at a continuous 0.5 litre/min for the
analyser void purge and 5 litres/min for 4 minutes. The zero
gas solenoid is energised-closed so that in the event of a power
failure the analyser sample cell is filled with zero gas, thus
preventing damage due to acidic condensation of the sample
gas in the sample cell.
SPAN GAS This should be a gas mixture containing all the gases for which
verification is required. The concentration of each gas should
be between 80% and 100% of the full-scale sensitivity of its
measured channel. Consumption is 5 litres/min for 4 minutes
every 24 hours. NO, CO, SO2 etc. should be mixed with N2.
NO2 should be mixed with pure air. The span gas
concentration (from the certificate) must be entered at the
menu item ‘Calibration verification’ on the Control Unit each
time a new gas bottle is installed. For US EPA compliance
purposes it should be EPA Protocol 1, calibration gas, vendor
certificated.
Operation
1) If not already on the TEST screen select the EYE symbol to navigate there.
2) Select AUTOZERO from the menu, this will give a drop down menu giving the
serial number of the Analyser head ACWn is looking at. Ensure the serial
number is for the head requiring the Auto-Zero.
The above steps assume that MANUAL AIR PURGE is NOT on. An Auto-Zero
cannot be carried out with manual air on, so this should be turned off before
hand. This is done by following the above steps but selecting MANUAL AIR
PURGE instead of Start AutoZero Now.
During the first Auto-Zero cycle after midnight a calibration routine is included
in the operation, this data will be recorded once per day over a 7-day period.
This data can be inspected and printed as described in Viewing the results.
During calibration phase the introduction of the gases causes the pressure in the
cell to increase slightly. To allow for both variations in the process pressure and
in the pressure during auto calibration, a transducer in the analyser is used to
provide continuous pressure compensation. In addition, to ensure that
calibration gas is present, the calibration gas cylinder pressure is sensed by a
pressure switch (set to operate at 100psi) integral with its regulator. An alarm is
set up within ACWn that can be set to trigger if the gas pressure is low.
A further function is for the remote start of an Autozero cycle, which overrides
the internal automatic timer. This facility is activated by shorting 2 inputs on the
Analyser Unit with the volt-free contact.
Figure 2 shows a table giving the phases of the calibration cycle including Auto-Zero,
their default timings and the states of the gas solenoids during the automatic calibration.
Note that the Zero gas solenoid valve is energized-closed so that in the event of a power
failure the valve opens, flooding the sample cell with dry air and excluding low
temperature sample gas that could cause acidic condensation and hence corrosion.
If a calibration error occurs, the action required depends on the scale of the
error:
If the error is less than 2.5% of full scale, no action is required as the calibration
is within specification
for a change in calibration of between 2.5% and 5% of full scale, the analyser
calibration is automatically adjusted to give the correct reading and the calibration
is OK
if an error is greater than 5% error or if the gas cylinder pressure falls below 100
psi, the analyser is considered to be out of calibration; no re-calibration action is
taken and the readings on the MAIN screen are shown thus:
The results of this are accessed via the MENU item “Calibration verification”.
The following items are shown on the screen with some example values:
Channel NOx
Print this day
Print last 7 days
Date 1st Jan
Time 00:00
(status) OK
Cal Gas conc (PPM) 300
Zero conc (PPM) 0
Zero error (%) 0.0
Span conc (PPM) 300
Span error (%) 0.0
If ‘Print this day’ is selected then the report for that day is printed. Similarly, if
‘Print last 7 days’ is selected then the reports for that period are printed.
Electrical connections
The electrical connections to enable the Calibration Gas controls are detailed on
the Installation Drawing 7-2054
Introduction
The Hot Access Port has been added to the reflector tube on the PROCAL 2000
analyser (See Figure G2 for location) to enable water vapour injection while
minimising condensation in the analyser probe. This will hence enable water
vapour based re-calibration on site from a suitably accurate water vapour
generator. The Procal 9000 Field Calibrator, including Water Vapour control is
available, see Procal Data Sheet 7-3057 for details.
Warning
AS THIS PORT IS CONNECTED TO THE INSTRUMENT MEASURING
CELL IT MUST BE CLOSED GAS TIGHT TO AVOID LEAKAGE OUT OR IN
WHEN NOT IN USE. A LEAK WILL INVALIDATE BOTH THE AUTO
ZERO / AUTOCAL RESULT AND CONTINUOUS MEASUREMENT
RESULTS.
This port is on the main stack flange; an optional insulation jacket is available, to
raise the port’s temperature whilst allowing access. See Figure G1 – Insulation
Jacket for details.
Operation
For most installations the use of the hot access port for water vapour injection
will require the fitting of an insulated jacket (Procal Part No 4-5455) around the
port to reduce the risk of condensation. See Fig G1 for fitting details. When the
jacket is fitted the system temperature should be allowed to stabilise before
continuing with any measurements. This may take up to 30 minutes. The jacket
allows for connection or disconnection without removal of the insulation as
shown on Fig.G1.
Before using the Hot Access Port for calibration first carry out an AutoZero with
the Hot Access Port closed off.
Carefully remove the plug with a 5mm Allen Key (Procal part 6-9901-87) and
place in a secure location for re-installing later. See Figure G1 – Insulation
Jacket and Figure G2 – Reflector Tube for details. Note that gases from the
process may escape from this port when it is open. Suitable precautions must be
taken, ie avoid breathing the hot escaping gas.
When applying water vapour a heated line or a purpose designed close coupled
evaporator should be used, and any exposed pipe or fittings containing hot
vapour should be insulated. See Figure G3 – Water Vapour Line Fitting
PROCAL 2000. On completion of tests remove the connecting line and replace it
with the hot access port plug, a 1/8 BSP taper plug, Procal part 6-6060-20 using
PTFE tape (Procal part 9-2900-01) to seal the thread. Ensure this is tightened to
form a gas tight seal.
Drawings
Introduction
Some PROCAL 2000 systems are supplied with an optional By-Pass Assembly
Where the process temperature is > 350°C and < 800°C (> 660°F and
< 1112°F)
This Appendix describes this option, and details how to install it.
By-Pass description
1-2060 Heated By-Pass - main wetted parts Hastelloy C276 / Heater Assy
Inconel 825
Other order specific By-Pass assemblies may be supplied, contact Kittiwake Procal Ltd or
the local distributor to obtain the appropriate installation drawing
Mounting Flanges
The By-Pass is mounted on two 4” 150lb flat face ANSI Flanges with (Not normally
supplied by Kittiwake Procal Ltd). The flanges are welded to the stack using, it is important
the weld is gas tight. Note the correct orientation of the flange is observed.
24mm NOM
Note Flange Orientation
50mm
136.0mm
The scoop should be attached to the appropriate branch before installation of the By-Pass.
The is supplied with graphite gaskets, these should be inspected before installation. It is
important the supplied gaskets are used in the installation
Secure the By-Pass to the stack mounted flanges using 8 x M16 nuts & washes per flange.
Heater Assembly
The By-Pass is supplied with the Heater Assembly fitted, it can be removed for inspection
and cleaning. If removed a new gasket should be used on re installation.
With the Heater disconnected remove the four M16 nuts which are
recessed in the horizontal By-Pass Flange.
Before reinstalling ensure all particulates have been removed from the
Heater Assembly and By-Pass.
Caution
Do not attempt to connect any mains cable (line cord) to the Heater if the
voltage configuration of the Heater differs from your local electrical supply
voltage. Contact Kittiwake Procal, or your Procal-authorised distributor.
Using an incorrect supply voltage will result in the In-Situ Heater failing to
function properly.
Earth Lead
Earth Connection Power to Lid
(Earth MUST be
)
Heater
Junction Box
By-Pass
1. Determine voltage of supply 110V or 240V (Others consult Kittiwake Procal Ltd or
Distributor)
2. Remove and retain the four screws securing the cover of the Heater junction box.
3. Feed through one end of the Heater power cable through the cable gland in the
Heater junction box.
4. Connect the power cable: 110V (Note colours may change depending on local code)
Brown (live) Terminal 1 on SSR
Blue (neutral) TB 2 (Retain link)
Green/yellow (earth) to the earth stud
5. Connect the power cable: 240V (Note colours may change depending on local code)
Brown (live) Terminal 1 on SSR
Blue (neutral) TB 1 (Discard link)
Green/yellow (earth) to the earth stud
7. Connect the AU control to the Heater SSR wiring the control cable through the
other gland.
a. SSR + to AU Terminal Box T1 – 13B
b. SSR - to AU Terminal Box T1 – 15B
8. Ensure the control cable is securely connected and with adequate clearance from the
mains supply terminations.
9. Tighten all glands and replace the cover securely, ensuring the cover earth wire is
fitted.
If the sample temperature shown on the Procal 1000 TEST screen is consistently
below 120oC, then it is likely that the HEATER is not working properly. Take
the following actions:
1 Measure the voltage between TB1-13B and TB1-15B on the AU junction box.
This should show a steady or pulsing 24 V dc.
2 Check that all the connections between the AU and the Control Unit are properly
made.
3 Confirm that a steady or slowly switching mains voltage appears between the
L and N terminals in the Heater mains connection box.
Over Heating
If the sample temperature is higher than the Heater set point it is
necessary to reduce the flow through the By-Pass this can be achieved
by introducing a Orifice plate in the branch which does not have the
scoop attached.
Consult Kittiwake Procal Ltd or the local distributor for advice.
Introduction
Some PROCAL 2000 systems are supplied with an optional In-Situ Steam
Heater (ISSH). This Appendix describes this option, and details how to install it.
An optional In-Situ Steam Heater (ISSH) may be fitted around the sample cell,
and is recommended for applications involving flue gases at temperatures below
120°C. The ISH contains an electrical heater, which has the effect of keeping
the temperature of the sample cell well above the dew point, thus preventing
condensation onto any optics in contact with the process gas.
Regulation of the ISH is controlled by a Steam Regulator and steam trap not
supplied with the ISH
1 Fit the supplied circular gasket to the four M16 bolts on the vertical flange already fitted to the process gas
2 Offer up the ISSH to the vertical flange. Introduce the heater gently into the duct and fit the ISSH flange o
Sample Gas Inlet Sample gas outlet (Must be
in stack
3 If the AU is not to be fitted gas flow)fit nuts to the four bolts, and finger-tighten to retain the ISSH saf
immediately,
4 Once the ISSH has been fitted, you may go on to fit the Analyser Unit.
Steam Inlet and Outlet
10mm Tube
1 Fit the supplied circular gasket to the four M16 bolts on the vertical flange
already fitted to the process gas duct.
2 Offer up the ISSH to the vertical flange. Introduce the heater gently into
the duct and fit the ISSH flange onto the four M16 bolts.
3 If the AU is not to be fitted immediately, fit nuts to the four bolts, and
finger-tighten to retain the ISSH safely.
4 Once the In-Situ Heater has been fitted, you may go on to fit the Analyser
Unit.
5 Connect the steam supply to the Inlet and Outlet 10mm OD Tubes using
appropriate pipefittings.
If the sample temperature shown on the Procal 1000 TEST screen is consistently
below T Cal, then it is likely that the ISSH is not working properly. Take the
following actions:
Introduction
Some PROCAL 2000 systems are supplied with an optional Steam Heated By-
Pass Assembly
This Appendix describes this option, and details how to install it.
By-Pass description
Other order specific By-Pas assemblies may be supplied, contact Kittiwake Procal Ltd or
the local distributor to obtain the appropriate installation drawing
The By-Pass is mounted on two 4” 150lb flat face ANSI Flanges with (Not normally
supplied by Kittiwake Procal Ltd). The flanges are wielded to the stack using, it is
important the wield is gas tight. Note the correct orientation of the flange is observed.
24mm NOM
Note Flange Orientation
50mm
136.0mm
The scoop should be attached to the appropriate branch before installation of the By-Pass.
The is supplied with graphite gaskets, these should be inspected before installation. It is
important the supplied gaskets are used in the installation
Secure the By-Pass to the stack mounted flanges using 8 x M16 nuts & washes per flange.
Connecting Steam
1. Connect the steam supply to the Inlet and Outlet 1 Steam Inlet /
Outlet
½ in 150lb Raised Face Flange
Over Heating
If the sample temperature is higher than the Heater set point it is
necessary to reduce the flow through the By-Pass this can be achieved
by introducing a Orifice plate in the branch which does not have the
scoop attached.
Consult Kittiwake Procal Ltd or the local distributor for advice.
Introduction
As stated on page 7-1 instrument supplied to the analyser must conform to the
specification as laid in document:
On installations where the quality of instrument air does not meet the required
specification it is strongly recommended that the optional Air Preparation Panel Part
Number 1-2938 is installed.
Item 1
Item 5
360mm (14.2”)
Item 4
Item 2
Item 3
Component Identification
2 Particulate Filter
Maintenance
The Air Preparation Panel should be inspected at regular intervals for visible signs of
moisture / particulate or oil contamination in the filter housings. The frequency of filter
element will depend on the quality of the air supply however as a minimum the filters
should be changed at least annually.
Filter Elements
Housing Comment
Part No.