Download as doc, pdf, or txt
Download as doc, pdf, or txt
You are on page 1of 2

TESTEX INSTRUMENT LTD

NO.13, WanJiang Industrial Street, WanJiang, Dongguan, Guangdong China


www.TestexTextile.com
+ 86 - 769 - 2329 4842

CERTIFICATE OF CALIBRATION No. CC23011503

Model TN142

Article Simulation wear pretreatment tester

Serial No. BR22120206XS

Standards EN 149 8.3.1

Manu. Date 2022- 12 - 02

Conclusion Calibrated items meet all requirements

Cali. Date 2023 - 01 - 15

Approved Signatory

Inspected by

Calibrated by

1. Major standards of measurement used in the calibration

Equipment Specification Range

Our Instruments 1/2 Your Success


TESTEX INSTRUMENT LTD
NO.13, WanJiang Industrial Street, WanJiang, Dongguan, Guangdong China
www.TestexTextile.com
+ 86 - 769 - 2329 4842

Differential pressure gauge 1 Pa/cm2 2000 Pa/cm2

Flow-meter 0.5 L/min 200 L/min

2. Place and environmental conditions of the calibration

Place Temp. R. H.

Lab of TESTEX 20.0 ± 3.0 ℃ 65 ± 3 %

3. Result of calibration

3.1 General Inspection

Item Goal Result

Appearance Complete, no scratches and varnish OK

Screw tightness Locked OK

Electrical parts Grounding, no damage OK

Controller and monitor Running well without any error OK

Leakage No gas leakage OK

3.2 Physical / Dimensional calibration

Item Goal Result

Flow-meter range 0L / min ~ 100L / min,


OK
accuracy is ± 2%

Ventilation volume (85 ± 1) L / min 85 L / min

Micro-pressure measuring range -1000Pa ~ 1000Pa, accuracy is 1Pa OK

The suction volume of the


≥100L / min OK
suction pump

Statements

1. The results of calibration relate only to the items calibrated.

2. Without the written permission from the calibration side, the certificate shall not be partially duplicated

Our Instruments 2/2 Your Success

You might also like