Plomo en Grafito

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 7

Spectrochimica Acta Part B: Atomic Spectroscopy 190 (2022) 106324

Contents lists available at ScienceDirect

Spectrochimica Acta Part B: Atomic Spectroscopy


journal homepage: www.elsevier.com/locate/sab

Determination of lead in blood by graphite furnace atomic absorption


spectrometry with Zeeman background correction: Improving a
well-established method to support a lower blood lead reference value
for children
Emily J. Pacer a, b, Christopher D. Palmer a, b, Patrick J. Parsons a, b, *
a
Laboratory of Inorganic and Nuclear Chemistry, Division of Environmental Health Sciences, Wadsworth Center, New York State Department of Health, Empire State
Plaza, Albany, NY 12237, USA
b
Department of Environmental Health Sciences, University at Albany, Rensselaer, NY 12144, USA

A R T I C L E I N F O A B S T R A C T

Keywords: Lead is ubiquitous in the environment and exposure can have adverse health effects even at low concentrations,
Graphite furnace particularly in young children. Blood Pb levels are the best indicator of recent Pb exposure. The Centers for
Permanent modifier Disease Control and Prevention (CDC) recently lowered the blood Pb reference value from 5 μg/dL to 3.5 μg/dL,
Childhood lead poisoning, blood lead reference
based on data calculated from the 97.5th percentile of children 1–5 years old in the National Health and
value
Nutrition Examination Survey (NHANES). The established method for determining Pb in blood by GFAAS has an
estimated detection limit (LOD) of ≈1 μg/dL. This presents an analytical challenge for its clinical use in the US,
given that the current population mean blood Pb level is 0.82 μg/dL. Accuracy and precision may be insufficient
to quantitate low blood Pb levels in the 1–5 μg/dL range, which is essential now that the CDC blood Pb reference
value is 3.5 μg/dL.
The original GFAAS method for blood Pb published in 1993 has been significantly improved to achieve a lower
LOD of 0.2 μg/dL, with the ability to quantitate blood Pb levels down to 1 μg/dL. This method is fit for the
purpose of detecting blood Pb levels currently, with excellent accuracy and precision in the 1–5 μg/dL range.
Results from analyzing SRMs, PT materials and human blood samples were in good agreement between two
different GFAAS platforms, as well as with results obtained by ICP-MS. The improvements provide evidence of an
GFAAS method capable of achieving lower LODs with improved accuracy and precision at lower blood Pb levels,
which can support the lower blood Pb reference value for children at 3.5 μg/dL.

1. Introduction assessing exposure to lead. At that time, a biochemical test based on


erythrocyte protoporphyrin was widely used to screen children for Pb
In 1993, when our rapid method for the determination of Pb in blood exposure. However, this biomarker of Pb exposure was not sufficiently
by graphite furnace atomic absorption spectrometry (GFAAS) with sensitive to identify children with blood Pb concentrations exceeding 10
Zeeman background correction was first published in this journal [1] the μg/dL [3]. Consequently, direct measurements for blood Pb became
US Centers for Disease Control and Prevention (CDC) had just recom­ necessary for both screening and diagnostic purposes, and on small
mended lowering the blood Pb action level for young children from 25 blood volumes obtained from a capillary fingerstick. The GFAAS method
μg/dL (1.21 μmol/L) to 10 μg/dL1 (0.48 μmol/L) [2]. This landmark published in this journal in 1993 was developed for a transversely-
change in environmental health policy was adopted in other countries. heated graphite atomizer (THGA) equipped with longitudinal Zeeman
Many laboratories were forced to reconsider their routine method for background correction and had an estimated limit of detection (LOD) of

* Corresponding author at: Laboratory of Inorganic and Nuclear Chemistry, Division of Environmental Health Sciences, Wadsworth Center, New York State
Department of Health, Empire State Plaza, Albany, NY 12237, USA.
E-mail address: patrick.parsons@health.ny.gov (P.J. Parsons).
1
In the US and elsewhere, blood Pb levels are typically reported in units of micrograms per deciliter (μg/dL). To convert μg/dL to μmol/L divide the former by
20.72.

https://doi.org/10.1016/j.sab.2021.106324
Received 10 September 2021; Received in revised form 5 November 2021; Accepted 7 November 2021
Available online 9 November 2021
0584-8547/© 2021 Wadsworth Center, NYS DOH. Published by Elsevier B.V. This is an open access article under the CC BY-NC-ND license
(http://creativecommons.org/licenses/by-nc-nd/4.0/).
E.J. Pacer et al. Spectrochimica Acta Part B: Atomic Spectroscopy 190 (2022) 106324

1 μg/dL. The method was modified slightly for a conventional GFAAS modifier consists of depositing 250 μg of W and 200 μg of Rh on
system equipped with continuum background correction [4] and was pyrolytically-coated graphite tubes in a series of drying steps, which
also successfully transferred to other manufacturers’ platforms. For takes approximately 45 min to perform [13].
almost 30 years, this method was widely recommended as the estab­ Samples are prepared in 0.5 mL acid-washed autosampler sample
lished GFAAS protocol in clinical laboratory medicine [5,6]. cups with a diluent that consists of 0.5% (v/v) Triton X-100 and 0.2%
Over the past 5 decades, extensive efforts have succeeded in reducing (v/v) HNO3. All reference materials, standard reference materials
Pb exposures and further help lower population blood Pb levels. Envi­ (SRMs), quality control materials (QCs), proficiency test (PT) samples,
ronmental regulations and public health interventions have resulted in and human specimens are treated in a same manner, i.e., diluted 1 + 9
the US geometric mean blood Pb level dropping from 12.8 μg/dL in with the Triton X-100/HNO3 diluent. A 10 mg/L intermediate Pb stock
1976–1980 to 0.82 μg/dL in 2015–2016 [7,8]. This trend has been solution is prepared to make weekly stock solutions, by pipetting 10 mL
observed in other countries too. Despite the drop in population blood Pb of a 1000 μg/mL Pb standard solution (High-Purity Standards, North
levels across the globe, Pb still remains a significant environmental risk Charleston, SC, USA) in 1% HNO3. The Pb standard solution used should
to young children and there is increasing evidence that adverse cogni­ be traceable to SI units via a certificate issued by a National Metrological
tive effects are associated with even lower blood Pb levels than previ­ Institute, for example the National Institute of Standards and Technol­
ously realized [9–11]. In 2012 after more than 20 years, the CDC ogy (NIST, Gaithersburg, MD USA).
recommended revisiting the definition of an elevated blood Pb level In contrast to the original published method, matrix matched (i.e.,
preferring to use a “reference value” that was based on the 97.5th blood-based) calibration standards are now mandatory for reliable
percentile of the distribution of US children aged 1 to 5 years old. In work. They are prepared daily as a 1 + 9 dilution of a 50 μL Pb cali­
2012, the blood Pb reference value was established as 5 μg/dL, and has bration working stock solution with 50 μL of “base” blood, and 400 μL of
since been adopted by many public health agencies in the US and else­ Triton X-100/HNO3 diluent. Ideally, a “base” blood should be confirmed
where. However in late 2021, the CDC lowered the blood lead reference as having no more than 0.5 μg/dL endogenous Pb. For this study, the
value yet further, from 5 μg/dL to 3.5 μg/dL (0.17 μmol/L) [12] based base blood (BB005) was obtained from a caprine source available to the
on the most up to date 97.5th percentile of the distribution of US chil­ New York State Department of Health (NYS DOH), but any whole blood
dren’s blood Pb levels [8]. This lower blood lead reference value now source (human, caprine or bovine) should suffice.
raises the question, is the GFAAS method developed in the 1990s for an
elevated blood Pb level at 10 μg/dL still fit for purpose in 2021 to sup­ 2.3. Certified reference materials, proficiency testing samples and human
port measurements at 3.5 μg/dL? blood specimens
In this report, we describe our efforts to re-visit and improve the
original GFAAS method proposed in 1993. We explore various ap­ Standard reference materials (SRM) 955c Toxic Metals in Caprine
proaches to achieving a lower LOD, and therefore a lower limit of Blood and 955d Toxic Metals and Metabolites in Frozen Human Blood
quantitation (LOQ) to support reporting blood Pb measurements below were obtained from NIST. The certified Pb values of these two SRMs
3.5 μg/dL. The improved method also takes into account the diminishing were established by Isotope Dilution Inductively Coupled Plasma Mass
prevalence of blood Pb levels above 20 μg/dL, which allows us to focus Spectrometry (ID-ICP-MS). Archived blood Pb proficiency test (PT)
on better performance at lower concentrations. New validation data for samples were obtained from the NYS DOH Biomonitoring PT Program
the improved method are included and are more focused on perfor­ and from the Centre de Toxicologie du Québec (CTQ), which operates
mance at lower blood Pb levels. As the focus shifts toward blood Pb the Québec Multielement External Quality Assessment Scheme (QME­
concentrations at 3.5 μg/dL and below, the contribution of background QAS). The PT material target values were assigned by the respective PT
contamination becomes much more significant and needs special program providers after the conclusion of each event. De-identified
attention. human blood samples (n = 30) were available from the Wadsworth
Center’s Trace Elements Laboratory. The archived human samples had
2. Methods been previously analyzed by ICP-MS and the range of blood Pb con­
centrations varied from 1 μg/dL to 22 μg/dL [15].
2.1. Instrumentation
3. Results and discussion
Two atomic absorption instruments were used throughout this study:
the Perkin Elmer (PE) AAnalyst 600 and the PE PinAAcle 900Z (Perki­ 3.1. Method optimization studies
nElmer Inc., Shelton, CT, USA). Both instruments are equipped with a
THGA design, longitudinal Zeeman background correction, and facili­ 3.1.1. Calibration
tate the implementation of methods based on STPF technology. The The method established previously recommended calibration stan­
basic instrumental parameters (line source, analytical wavelength, etc.) dards in the range of 10 μg/L to 60 μg/L, equivalent to a blood Pb range
were adapted from the original method published in 1993. Sample ali­ from 10 μg/dL to 60 μg/dL after correcting for the dilution factor [1].
quots were delivered using an AS-800/900 autosampler. Given that the current (2015–2016) geometric mean blood Pb level of
the US population has fallen to 0.82 μg/dL, and the blood Pb reference
2.2. Reagents and consumables value has been lowered to 3.5 μg/dL in 2021, the calibration range was
adjusted to fit the current population blood Pb levels. The dilutions the
In the original method, a conventional modifier was used consisting weekly stock solutions from the intermediate Pb stock solution are
of a diluent reagent comprising 0.5% (w/w) ammonium dihydrogen described in Table S1 (Appendix).
phosphate, 0.2% (v/v) HNO3 and 0.5% (v/v) Triton X-100 [1]. In this The daily calibration standards ranged from 1 to 40 μg/L, (equivalent
study, we also use a W-Rh permanent modifier, reagent solutions and the to 1–40 μg/dL blood Pb after correcting for the dilution factor) consis­
coating process that are described in detail elsewhere [13,14]. In brief, tent with CLSI recommendations [6]. Shifting the lowest calibrator to 1
the W solution was prepared by dissolving 0.1 g of Na2WO4• 2H2O (Alfa μg/L more appropriately reflects the current geometric mean blood Pb
Aesar®, Ward Hill, MA, USA) in 100 mL of double deionized (DI) water. level. Given the levels in the current population, a top calibrator of 40
The Rh solution was prepared by dissolving 0.1 g of (NH4)2[RhCl6] • μg/L is better suited for clinical assessment and biomonitoring purposes
xH2O (Alfa Aesar®, Ward Hill, MA, USA) in 10 mL of 10% HCl and 90 [6].
mL of DI water, which gives an orange-red colour to the solution. The While aqueous Pb calibration standards were used in the original
procedure for coating the L’vov platform with a permanent W-Rh method [1], matrix matched calibration was mandatory as the

2
E.J. Pacer et al. Spectrochimica Acta Part B: Atomic Spectroscopy 190 (2022) 106324

atomization profiles of aqueous standards consistently showed double volumes of 16 μL, 18 μL, and 24 μL for 1 + 9 diluted blood, thus
peaks with a W-Rh permanent modifier. Double or multiple peaks may increasing the amount of whole blood deposited up to 2.4 μL. Fig. 2a
be due to some of the analyte being transferred to the tube walls during shows the atomization profiles obtained for different injection volumes
the drying steps. After visual inspection of the sample during the drying of a matrix-matched calibration standard diluted 1 + 9, i.e., to 40-μg/L.
steps and increasing the drying time by >30 s, double peaks were still As expected, as more sample is deposited on the platform, the
observed. Differences in atomization profile for (a) aqueous standards recorded Ai increases too: 0.0653 s, 0.0810 s, 0.0862 s, and 0.1127 s, for
compared to a blood (b) matrix-matched Pb standard are shown in injection volumes of 12 μL, 16 μL, 18 μL, and 24 μL, respectively.
Fig. 1. The calibration of Pb in both standards was adjusted to 40 μg/L, However, the magnitude of the increase in Ai is not proportionate to the
equivalent to 40 μg/dL blood Pb. volume (i.e., analyte mass) deposited. The observed increase in Ai was
Significant differences in atomization profile, i.e., peak shapes can be less than expected: ≈93% for 16 μL; ≈88% for 18 μL; and ≈86% for 24
observed for the two calibrators when using a permanent W-Rh modi­ μL. Clearly, atomization efficiency (and therefore sensitivity) decreases
fier. A previous study noted a shoulder on the peak profiles of Pb when as the volume of diluted blood deposited increases. This could be due to
W was used as the permanent modifier, but in the presence of Rh, well- the effect of the increased blood matrix on atomization. The absolute
defined peaks were always observed for water-based reference materials sensitivity or characteristic mass (m0) for volume deposited was 32.3 pg
[13]. Other authors have reported good experience for blood Pb using a for 12 μL, 34.8 pg for 16 μL, 36.7 pg for 18 μL, and 37.5 pg for 24 μL.
Mo-coated L’vov platform [16]. Resano et al. successfully used a Pt While increased sensitivity is desirable, increasing the amount of
modifier with solid-sampling GFAAS to analyze dried blood spots for Pb (blood) matrix on the platform can adversely affect measurement pre­
content [17] and consistent with our studies, they reported the necessity cision, and thus the LOD. Increasing the injection volume requires an
to use matrix-matched calibration standards. Tsalev et al. reported that adjustment to drying, pyrolysis, and clean-out times. For example, a 24-
one of the limitations of using permanent modifiers can be the presence μL injection volume required an additional 25 s per replicate to achieve
of double or multiple peaks [18]. In our study, double peaks were complete drying. LODs were calculated by analyzing NIST SRM 955c
consistently observed with aqueous calibration standards when using W Level 1 over 7 independent runs (3SD). Increasing the injection volume
with Rh. Since matrix-matched standards showed no double peaks with from 12 μL to 16 μL lowered the LOD from 0.27 μg/dL to 0.16 μg/dL, and
W-Rh permanent modifier, their use was considered mandatory. 16 μL was then adopted.

3.1.2. Injection volume 3.1.3. Blood dilution factor


Increasing the volume of diluted blood deposited on the platform An alternative approach to depositing more analyte on the platform
was explored as this increases the amount of analyte available for at­ is to change the blood sample dilution factor. The established method
omization, but it also increases the amount of blood matrix that should used a 1 + 9 dilution factor, wherein 50 μL of whole blood was diluted
be removed during pyrolysis. Any improvement in performance would with 450 μL of the Triton X-100/HNO3 diluent with phosphate modifier
depend on how much the signal increases relative to the noise from the [1]. In this study, we explored using a 1 + 4 dilution factor while
matrix [19]. There is a limit to the maximum sample volume that can be maintaining the injection volume of 12 μL, equivalent to doubling the
deposited onto the L’vov platform, typically, 50 μL or less [20]. Other amount of analyte deposited. For a 40-μg/L Pb standard, the mass of
methods that describe the determination of Pb in biological matrices analyte deposited onto the platform increased from 480 pg to 960 pg.
have recommended injection volumes ranging from 10 μL to 20 μL Fig. 2b shows the comparison of atomization profiles for 12 μL of a 40-
[21,22]. The 1993 blood Pb method recommended an injection volume μg/L matrix-matched calibration standard diluted 1 + 9 (red) and one
of 12 μL of 1 + 9 diluted blood equivalent to 1.2 μL of whole blood diluted 1 + 4 (black).
deposited on a L’vov platform [1]. In this study, we compared injection The 1 + 9 dilution yielded an Ai of 0.0653 s, while the Ai for 1 + 4

0.15

0.10 (b) Matrix-Matched Pb Standard-AA


Absorbance

(b) Matrix-Matched Pb Standard-BG

0.05 (a) Aqueous Pb Standard-AA


(a) Aqueous Pb Standard-BG

0.00
0 1 2 3 4
Time (s)
Fig. 1. Absorption profile from (a) aqueous Pb standard (blue, Ai = 0.0636 s) and (b) blood matrix-matched Pb standard (red, Ai = 0.0653 s), equivalent to 40 μg/L
using the W-Rh permanent modifier. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)

3
E.J. Pacer et al. Spectrochimica Acta Part B: Atomic Spectroscopy 190 (2022) 106324

0.3 modifier, m0 ranged from 35 to 40 pg [1]. As previously found the


optimal m0 for Pb in blood using a permanent W-Rh modifier was 31 pg
(a) 24 µL-AA [14]. We found m0 ranged from 32 to 36 pg using the W-Rh modifier,
24 µL-BG similar to values previously reported.
The change from a conventional phosphate modifier to a permanent
18 µL-AA
0.2 W-Rh modifier also necessitated calibration using blood-based, i.e.,
18 µL-BG
matrix-matched Pb standards. The calibration range was limited to
Absorbance

16 µL-AA 1–40 μg/L, which better reflects current population blood Pb levels. Fig.
16 µL-BG S1 (Appendix) shows the typical calibration curves using the final
12 µL-AA optimized parameters for the two different THGA platforms.
0.1 12 µL-BG The LOD for the improved blood Pb method on two THGA platforms
was calculated based on the IUPAC/ISO harmonized protocol, i.e., 3 x
SD for n = 7 measurements, i.e., collected over 7 independent analytical
runs, and using NIST SRM 955c Level 1 [24]. The LOQ was calculated as
10 x SD per laboratory policy. The LOD with the method improvements
0.0 on the PE AAnalyst 600 was calculated as 0.16 μg/dL and the LOQ as 0.5
0 1 2 3 4
μg/dL. On the PE PinAAcle 900Z, the LOD was calculated as 0.21 μg/dL
Time (s)
and the LOQ as 0.7 μg/dL.
0.3
A lower LOD of ≈0.2 μg/dL was achieved on the two THGA platforms
(b) using the improved blood Pb method with a W-Rh permanent modifier
and matrix matched calibration. This also provides a lower blood Pb
LOQ of ≈0.6 μg/dL, which is certainly fit-for-purpose. Now that a lower
12 µL (1+4)-AA CDC blood Pb reference value of 3.5 μg/dL has been adopted, this is
0.2 12 µL (1+4)-BG uncomfortably close to the previously established method LOQ of 2.5–3
Absorbance

μg/dL, such that quantitating blood Pb levels below 2–3 μg/dL are
subject to larger degree of relative uncertainty. After implementing the
method improvements, blood Pb quantitation down to 1 μg/dL and even
12 µL (1+9)-AA
12 µL (1+9)-BG
lower is now feasible. The improved blood Pb method would extend the
0.1 application of GFAAS technology within small laboratory settings where
ICP-MS is neither cost effective nor feasible.
The W-Rh coating is stable for approximately 350 firings, approxi­
mately equivalent to an 8-h workday, consistent with that reported in
other studies [13,14]. The W-Rh can be recoated to extend THGA tube
0.0 lifetime up to 1000 firings [13,14], compared to using the established
0 1 2 3 4
GFAAS method in which, the tube lifetime using the conventional
Time (s)
modifier was approximately 350 firings [1]. Biological matrices such as
Fig. 2. (a) Typical atomization profiles for injection volumes of 12 μL (red, Ai
blood, or high concentrations of Triton X-100, have been known to cause
= 0.0653 s), 16 μL (blue, Ai = 0.0810 s), 18 uL (black, Ai = 0.0862 s) and 24 μL carbonaceous residue build-up with a decrease in graphite tube life [25].
(green, Ai = 0.1127 s) of a matrix-matched calibration standard equivalent to After 250 firings, carbonaceous residues were visible on the L’vov
40 μg/L. (b) Comparison of atomization profiles of a 12 μL injection of a 40-μg/ platform. This is consistent with other findings, that found an excessive
L matrix-matched Pb standard diluted 1 + 9 (red, Ai = 0.0653 s) and diluted 1 build-up after 250 firings using the W-Rh modifier compared to the
+ 4 (black, Ai = 0.1080 s). (For interpretation of the references to colour in this conventional phosphate modifier [14]. However, the carbonaceous
figure legend, the reader is referred to the web version of this article.) build up did not obstruct the optical beam, therefore measurements
were unaffected, but it is possible that this could lead to higher back­
dilution only increased 0.1080 s, or ≈82.7% of the expected increase. ground levels and low accuracy over multiple firings. One study found
The corresponding sensitivities (m0) were 32.3 pg for a 1 + 9 dilution that the use of hydrogen peroxide could reduce the carbonaceous res­
and 39.1 pg for a 1 + 4 dilution, i.e., ≈21% poorer sensitivity. An idue build up from analyzing biological matrices [26].
additional complication was observed with 1 + 4 diluted blood wherein
it was more difficult to clean the autosampler tip between injections.
3.3. Validation studies
The poorer sensitivity and increased maintenance issues with 1 + 4
diluted blood indicated a 1 + 9 dilution was preferable.
Numerous human and caprine blood certified reference materials
were analyzed to assess the method accuracy and precision on two
3.2. Final optimized parameters THGA platforms. The results are shown in Table 1.
The mean and expanded uncertainty (U) were calculated from n = 12
The optimized THGA heating program shown in Table S2 (Appendix) measurements. Uncertainties of the measured SRMs were calculated
was modified from the original method, which called for a conventional according to the GUM and Eurolab guidelines [27,28]. The expanded
modifier. The updated method calls for a permanent W-Rh modifier, uncertainty (U) was calculated using a coverage factor of 2:
which is applied as a coating to the L’vov platform prior to calibration
and sample analysis. Injection aliquots of 16 μL with a 1 + 9 dilution UMEAS = 2uc (1)
were selected as optimal based on the development studies reported in
For each certified reference material, the combined standard un­
Sections 3.1.2 and 3.1.3 above.
certainty (uc) was defined as:
The characteristic mass (m0), or absolute sensitivity is a fundamental
√̅̅̅̅̅̅̅̅̅̅̅̅̅̅
parameter in GFAAS that is useful as a day-to-day indicator of method uc = s2 + b2 (2)
performance, and is defined as the mass of analyte that absorbs 1% of
transmitted light i.e., equivalent to an integrated absorbance value of where s = standard deviation for each respective THGA platform and b
0.0044 s [23]. For the established GFAAS method using a phosphate was defined as:

4
E.J. Pacer et al. Spectrochimica Acta Part B: Atomic Spectroscopy 190 (2022) 106324

Table 1
Analysis of NIST SRMs for blood Pb on the PE AAnalyst 600 and PE PinAAcle 900Z.
Standard reference material Certified value (μg/dL ± USRM)a AAnalyst 600 mean (μg/dL ± UMEAS)b PinAAcle 900Z mean (μg/dL ± UMEAS)b

NIST 955c- Level 1 0.424 ± 0.0011 0.4 ± 0.3 0.5 ± 0.2


NIST 955c- Level 2 13.950 ± 0.080 13.8 ± 1.0 14.1 ± 0.9
NIST 955c- Level 3 27.76 ± 0.16 26.8 ± 2.5 27.4 ± 2.0
NIST 955d- Level 1 1.480 ± 0.026 1.4 ± 0.3 1.6 ± 0.4
NIST 955d- Level 2 4.947 ± 0.085 4.7 ± 0.6 5.1 ± 0.4
a
USRM = expanded measurement uncertainty from the SRM certificate (k = 2).
b
UMEAS = expanded measurement uncertainty of the measured SRM (k = 2).

√̅̅̅̅̅̅
b= ∆2 + U 2SRM (3) 3.4. Human samples

Eq. (3) defines b as the square root of the sum of squares of both ∆ Human blood samples were also analyzed using the improved
representing the bias between the certified and found values, and USRM method as part of the assessment. The results of the 30 samples were
is the uncertainty of the SRM, given on the SRM certificate. The results compared with those obtained using a well- established ICP-MS method
with the method improvements are in good agreement between both [15]. Results in Fig. 3 show the blood Pb data obtained from the
THGA platforms, as well as with the certified values provided by NIST. AAnalyst 600 and PinAAcle 900Z, plotted against values obtained by the
Participation in PT programs provides beneficial information ICP-MS method. For comparison purposes, the solid line represents y = x
regarding laboratory performance via circulation of well characterized and the dashed lines represents ±2 μg/dL.
blood Pb samples [29]. The current United States PT regulatory limits The results indicate excellent agreement between the GFAAS and
for blood Pb were set more than 25 years ago, at ±4 μg/dL or 10% of the ICP-MS method over the 1–5 μg/dL blood Pb concentration range. All
target value, whichever is greater [30]. Most laboratories can achieve a the human blood Pb values measured by GFAAS were within ±1 μg/dL
performance of at least ±2 μg/dL at low blood Pb levels [31], and there of the ICP-MS value, providing further validation of the proposed
have been calls over many years to tighten these PT limits. The current method.
US criteria mean a blood Pb PT sample of 5 μg/dL could be reported
within a range from 1 to 9 μg/dL, and still be considered acceptable. This
3.5. Commutability
makes interpreting patient blood Pb test results uncertain at low levels.
In addition to the NIST SRMs, archived blood Pb PT materials from the
A reference material that is commutable will have the same numeric
NYS DOH Biomonitoring PT program and the CTQ QMEQAS programs
result that is expected for an equivalent human sample containing the
were also analyzed on the two THGA platforms (Table 2).
same quantity of analyte [32]. Since caprine SRMs were used in method
The mean and standard uncertainty (u) were calculated based on the
development process to assess the improvements for blood Pb mea­
analysis of 12 measurements. Standard uncertainty (uMEAS) of the
surements, it is important to address whether the caprine blood refer­
measured PT samples were calculated according to the GUM and
ence materials are commutable to human samples. While using caprine
Eurolab guidelines [27,28], as:
blood may have advantages, such as being able to endogenously dose
s
uMEAS = √̅̅̅ (4) animals with desired levels of Pb, it may be less likely to carry human
n pathogens [33]. Non-commutability for a reference material is not a
deficiency of the method, especially those that produce acceptable re­
where s = standard deviation and n = number of analyses for each
sults for samples. Rather, it suggests that a specific reference material
respective THGA platform.
may not be appropriate for use to calibrate or verify accuracy for that
The results obtained with the improved GFAAS method are in good
particular measurement due to matrix effects [34].
agreement between both instrument platforms, as well as with the
To assess commutability, human samples and variety of SRMs and PT
assigned target values from the PT program provider. All values were
program materials discussed previously in Section 2.3 were analyzed for
within ±2 μg/dL of the assigned value, the NYS DOH program’s
Pb. All of the blood materials analyzed were compared using the
acceptable range. All but one of the values were within the CTQ pro­
improved GFAAS method with the target values from ICP-MS instru­
gram’s z-score defined range. The materials that were available pre­
mentation or certified values from the NIST certificate. Results are
cluded blind analysis.
shown in Fig. 4 for the AAnalyst 600 and PinAAcle 900Z, plotted against
values found with ICP-MS. The SRMs and PT program are portrayed by
different symbols for each respective material, while the human samples
are shown as black circles, as described in the figure legend. For

Table 2
Analysis of archived blood Pb proficiency testing materials on the PE AAnalyst 600 and PE PinAAcle 900Z.
PT samples Target value Acceptable range for program AAnalyst 600 mean (μg/dL ± uMEAS)b PinAAcle 900Z mean (μg/dL ± uMEAS)b
(μg/dL ± uPT)a

CTQ QMEQAS
QM-B-Q2010 1.36 ± 0.017 1.05–1.67 1.4 ± 0.03 1.8 ± 0.1
QM-B-Q2011 10.6 ± 0.110 9.08–12.1 11.2 ± 0.1 11.3 ± 0.1
QM-B-Q2012 6.64 ± 0.067 5.64–7.64 6.8 ± 0.05 7.2 ± 0.1
NYS Biomonitoring
BE20-12 1.94 ± 0.05 0.0–3.94 1.8 ± 0.04 2.2 ± 0.03
BE20-13 8.4 ± 0.2 6.4–10.4 8.9 ± 0.1 9.1 ± 0.1
BE20-14 1.19 ± 0.03 0.0–3.19 1.2 ± 0.03 1.4 ± 0.03
BE20-15 14.0 ± 0.3 12.0–16.0 14.5 ± 0.1 15.0 ± 0.2
a
uPT = standard measurement uncertainty from the PT provider report.
b
uMEAS = standard measurement uncertainty of the measured PT sample.

5
E.J. Pacer et al. Spectrochimica Acta Part B: Atomic Spectroscopy 190 (2022) 106324

25 30
AAnalyst 600
AAnalyst 600
25
20
GFAAS, µg/dL

20

GFAAS, µg/dL
15

15
10

10

5
5

0
0 5 10 15 20 25 0
0 5 10 15 20 25 30
ICP-MS, µg/dL
ICP-MS, µg/dL
30
25
PinAAcle 900Z
PinAAcle 900Z
25
20

20

GFAAS, µg/dL
GFAAS, µg/dL

15
15

10
10

5 5

0 0
0 5 10 15 20 25 0 5 10 15 20 25 30

ICP-MS, µg/dL ICP-MS, µg/dL

NIST 955c CTQ QMEQAS Human Samples


Fig. 3. Analysis of 30 human blood samples for Pb using the improved GFAAS
method: (top panel) AAnalyst 600 and (bottom panel) PinAAcle 900Z. All NIST 955d NYS PT
values are plotted against values obtained using an established ICP-MS method
[15]. For comparison purposes, the solid line represents y = x and the dashed Fig. 4. Analysis of 30 human blood samples (black circles) and for Pb using the
lines represents ±2 μg/dL. proposed GFAAS method improvements on the (top panel) AAnalyst 600 and
(bottom panel) PinAAcle 900Z. Caprine and human SRMs and PT materials are
represented by different shapes indicated in the legend. All values are plotted
comparison purposed, the solid line represents y = x and the dashed
against target or certified values established by ICP-MS. The solid line y = x is
lines represent the limits ±2 μg/dL. shown for comparison purposes and the dashed lines represent ±2 μg/dL.
The caprine and human SRMs and PT materials align with the human
samples within acceptable criteria on both GFAAS platforms. This sug­
gests the caprine blood materials are commutable to human blood ma­
terials. The patient samples were limited to a small range with the dL. Newer instrumentation, increased sample injection volume and
highest concentration at 22 μg/dL. Therefore, the SRM material above different calibration strategies were all combined to improve the
the highest patient sample could be commutable, but with less confi­ method. The improved GFAAS method achieved a detection limit of 0.2
dence than those below the highest human sample. μg/dL, that can quantitate blood Pb levels at 1 μg/dL. This provides
justification of an GFAAS method that could feasibly operate with a
4. Conclusion lower blood Pb reference value, and improved accuracy and precision of
measurements in the 1–5 μg/dL concentration range. Additionally,
Throughout the past 5 decades, extensive efforts have been suc­ analysis of all SRMs, PT materials, and human samples were all well
cessful in reducing Pb exposures to further help lower population blood within ±2 μg/dL of the values established by ICP-MS instrumentation.
Pb levels. Even though blood Pb levels have fallen dramatically, Pb still Therefore, the improved GFAAS method could also operate within
remains a significant environmental risk to young children. A new narrower PT acceptable limits of ±2 μg/dL, compared to the current US
reference value of 3.5 μg/dL was adopted by the CDC in 2021, based on federal limit of ±4 μg/dL.
the 97.5th percentile of the distribution of US children’s blood Pb levels. As reliability of blood Pb measurements improve, fewer false posi­
The GFAAS method developed in the 1993 for a blood Pb action level at tives and false negatives of patient samples are an added benefit. This
10 μg/dL has now been improved to ensure it is fit for purpose for could ease the burden of reporting patient blood Pb results, and benefit
current population blood Pb levels. public health officials who initiate expensive environmental in­
We explored various approaches to achieving a lower LOD, and vestigations at higher blood Pb levels. The improved method could
therefore a lower LOQ to support blood Pb measurements below 3.5 μg/ benefit hospitals and smaller laboratories in terms of keeping pace with

6
E.J. Pacer et al. Spectrochimica Acta Part B: Atomic Spectroscopy 190 (2022) 106324

the goals of detecting and quantifying lower blood Pb levels using below 10 μg per deciliter, N. Engl. J. Med. 348 (2003) 1517–1526, https://doi.org/
10.1056/NEJMoa022848.
GFAAS instrumentation, especially if an ICP-MS instrument is not
[11] J. Schwartz, Low-level lead exposure and children’s IQ: a metaanalysis and search
available. The method improvements discussed provide evidence that for a threshold, Environ. Res. 65 (1994) 42–55, https://doi.org/10.1006/
mature techniques, such as GFAAS, can still provide results that are fit enrs.1994.1020.
for purpose given the current population blood Pb levels. [12] P.Z. Ruckart, R.L. Jones, J.G. Courtney, T.T. LeBlanc, W. Jackson, M.P. Karwowski,
P.Y. Cheng, P. Allwood, E.R. Svendsen, P.N. Breysse, Update of the blood lead
reference value - United States, 2021, MMWR Morb. Mortal. Wkly Rep. 70 (2021)
Funding 1509–1512.
[13] É.C. Lima, F.J. Krug, K.W. Jackson, Evaluation of tungsten–rhodium coating on an
integrated platform as a permanent chemical modifier for cadmium, lead, and
This work was supported in part by funding provided by the Centers selenium determination by electrothermal atomic absorption spectrometry,
for Disease Control and Prevention (5 NU88EH001329-03-00) and by Spectrochim. Acta Part B 53 (1998) 1791–1804, https://doi.org/10.1016/S0584-
the Wadsworth Center, New York State Department of Health. 8547(98)00227-4.
[14] Y. Zhou, R.A. Zanão, F. Barbosa Jr., P.J. Parsons, F.J. Krug, Investigations of a W-
Rh permanent modifier for the determination of Pb in blood by electrothermal
Author statement atomic absorption spectrometry, Spectrochim. Acta Part B 57 (2002) 1291–1300,
https://doi.org/10.1016/S0584-8547(02)00085-X.
[15] C.D. Palmer, M.E. Lewis, C.M. Geraghty, F. Barbosa, P.J. Parsons, Determination of
This statement certifies that all authors have seen and approved the lead, cadmium and mercury in blood for assessment of environmental exposure: a
final version of the manuscript being submitted. The authors warrant comparison between inductively coupled plasma–mass spectrometry and atomic
that the article is the authors’ original work, hasn’t received prior absorption spectrometry, Spectrochim. Acta Part B 61 (2006) 980–990, https://
doi.org/10.1016/j.sab.2006.09.001.
publication and isn’t under consideration for publication elsewhere.
[16] P.C. D’Haese, L.V. Lamberts, L. Liang, F.L. Van de Vyver, M.E. De Broe, Elimination
of matrix and spectral interferences in the measurement of lead and cadmium in
Declaration of Competing Interest urine and blood by electrothermal atomic absorption spectrometry with deuterium
background correction, Clin. Chem. 37 (1991) 1583–1588.
[17] M. Resano, L. Rello, E. García-Ruiz, M.A. Belarra, Minimally-invasive filter paper
The authors declare that they have no known competing financial test in combination with solid sampling-graphite furnace atomic absorption
interests or personal relationships that could have appeared to influence spectrometry for Pb determination in whole blood, J. Anal. Atom. Spectrometry 22
the work reported in this paper. (2007) 1250–1259.
[18] D.L. Tsalev, V.I. Slaveykova, L. Lampugnani, A. D’Ulivo, R. Georgieva, Permanent
modification in electrothermal atomic absorption spectrometry — advances,
Acknowledgment anticipations and reality, Spectrochim. Acta Part B 55 (2000) 473–490, https://doi.
org/10.1016/S0584-8547(00)00194-4.
[19] B. Welz, M. Sperling, Atomic Absorption Spectrometry, third ed., Wiley-VCH, New
This paper is dedicated to the memory of Walter Slavin (1927–2021), York, 1998.
a former editor of Spectrochimica Acta, Part B Atomic Spectroscopy, and a [20] R.D. Beaty, Concepts, Instrumentation and Techniques in Atomic Absorption
leader in the design and development of atomic absorption spectrome­ Spectrophotometry, The Perkin-Elmer Corporation, 1988.
[21] D. Andrada, F.G. Pinto, C.G. Magalhaes, B.R. Nunes, M.B. Franco, J.B.B. da Silva,
ters. Walter was a mentor to many of us, and his numerous contributions Direct determination of lead in human urine and serum samples by electrothermal
to graphite furnace atomic absorption spectrometry included the atomic absorption spectrometry and permanent modifiers, J. Braz. Chem. Soc. 17
determination of lead in blood. (2006) 328–332, https://doi.org/10.1590/S0103-50532006000200016.
[22] É.C. Lima, F. Barbosa Jr., F.J. Krug, Lead determination in slurries of biological
materials by ETAAS using a W-Rh permanent modifier, Fresenius J. Anal. Chem.
Appendix A. Supplementary data 369 (2001) 496–501, https://doi.org/10.1007/s002160000667.
[23] B.V. L’Vov, V.G. Nikolaev, E.A. Norman, L.K. Polzik, M. Mojica, Theoretical
Supplementary data to this article can be found online at https://doi. calculation of the characteristic mass in graphite furnace atomic absorption
spectrometry, Spectrochim. Acta Part B 41 (1986) 1043–1053, https://doi.org/
org/10.1016/j.sab.2021.106324. 10.1016/0584-8547(86)80125-2.
[24] L.A. Currie, Nomenclature in evaluation of analytical methods including detection
References and quantification capabilities (IUPAC recommendations 1995), Anal. Chim. Acta
391 (1999) 105–126, https://doi.org/10.1016/S0003-2670(99)00104-X.
[25] K.S. Subramanian, Determination of metals in biofluids and tissues: sample
[1] P.J. Parsons, W. Slavin, A rapid Zeeman graphite furnace atomic absorption preparation methods for atomic spectroscopic techniques, Spectrochim. Acta Part B
spectrometric method for the determination of lead in blood, Spectrochim. Acta 51 (1996) 291–319, https://doi.org/10.1016/0584-8547(95)01425-X.
Part B 48 (1993) 925–939, https://doi.org/10.1016/0584-8547(93)80094-B. [26] J.L. Burguera, C. Rondón, M. Burguera, M.E. Roa, Y. Petit de Peña, Electrothermal
[2] CDC, Preventing Lead Poisoning in Young Children, U.S. Dept. of Health and atomic absorption spectrometry determination of molybdenum in whole blood,
Human Services, Center for Disease Control and Prevention, Atlanta, GA, 1991. Spectrochim. Acta Part B 57 (2002) 561–569, https://doi.org/10.1016/S0584-
[3] P.J. Parsons, A.A. Reilly, A. Hussain, Observational study of erythrocyte 8547(01)00377-9.
protoporphyrin screening test for detecting low lead exposure in children: impact [27] JCGM, Evaluation of measurement data—Guide to the expression of uncertainty in
of lowering the blood lead action threshold, Clin. Chem. 37 (1991) 216–225. measurement (GUM), in: Joint Committee for Guides in Metrology 100, 2008.
[4] H. Qiao, P.J. Parsons, W. Slavin, Transferability of blood lead determinations by [28] Eurolab, Measurement uncertainty revisited: Alternative approaches to uncertainty
furnace atomic absorption spectrophotometry and continuum background evaluation, in: Technical Report, 2007.
correction, Clin. Chem. 41 (1995) 1451–1454. [29] P.J. Parsons, F. Barbosa, Atomic spectrometry and trends in clinical laboratory
[5] D.L. Tsalev, Z.K. Zaprianov, Atomic Absorption Spectrometry in Occupational and medicine, Spectrochim. Acta Part B 62 (2007) 992–1003, https://doi.org/
Environmental Health Practice, CRC Press, Boca Raton, FL, 1995. 10.1016/j.sab.2007.03.007.
[6] CLSI, Measurement Procedures for the Determination of Lead in Whole Blood; [30] P.J. Parsons, K.G. McIntosh, Human exposure to lead and new evidence of adverse
Approved Guideline—Second Edition CLSI document C40-A2, Clinical and health effects: implications for analytical measurements, Powder Diffract. 25
Laboratory Standards Institute, Wayne, PA, 2013. (2010) 175–181, https://doi.org/10.1154/1.3402340.
[7] T. Dignam, R.B. Kaufmann, L. LeStourgeon, M.J. Brown, Control of lead sources in [31] Council on Environmental Health, Prevention of Childhood Lead Toxicity,
the United States, 1970–2017: public health progress and current challenges to Pediatrics 138, 2016, p. e20161493, https://doi.org/10.1542/peds.2016-1493.
eliminating lead exposure, J. Public Health Manag. Pract. 25 (2019) S13–S22, [32] W.G. Miller, G.R.D. Jones, G.L. Horowitz, C. Weykamp, Proficiency testing/
https://doi.org/10.1097/PHH.0000000000000889. external quality assessment: current challenges and future directions, Clin. Chem.
[8] K.L. Caldwell, P.Y. Cheng, J.M. Jarrett, A. Makhmudov, K. Vance, C.D. Ward, R. 57 (2011) 1670–1680, https://doi.org/10.1373/clinchem.2011.168641.
L. Jones, M.E. Mortensen, Measurement challenges at low blood lead levels, [33] L.L. Yu, G.C. Caceres, S.E. Long, K.E. Murphy, C.E. Bryan Sallee, T.W. Vetter,
Pediatrics 140 (2017), e20170272, https://doi.org/10.1542/peds.2017-0272. C. Hagwood, M.A. Franklin, J. Castro Georgi, J.M. Jarrett, D.M. Jones, R.L. Jones,
[9] B.P. Lanphear, R. Hornung, J. Khoury, K. Yolton, P. Baghurst, D.C. Bellinger, R. Z. Li, Y.L. Sommer, D.S. Tevis, K. Wallon, C.D. Ward, M. Maras, S. Erdahl,
L. Canfield, K.N. Dietrich, R. Bornschein, T. Greene, S.J. Rothenberg, H. C. Gilmer, M. Wermers, H. Woldeysus, K.F. Mehigan, M. Morrissette, C.D. Palmer,
L. Needleman, L. Schnaas, G. Wasserman, J. Graziano, R. Roberts, Low-level P.J. Parsons, Certification of Standard Reference Material® 955d, Toxic Elements
environmental lead exposure and children’s intellectual function: an international and Metabolites in Frozen Human Blood, National Institute of Standards and
pooled analysis, Environ. Health Perspect. 113 (2005) 894–899, https://doi.org/ Technology (NIST), 2020, https://doi.org/10.6028/NIST.SP.260-206.
10.1289/ehp.7688. [34] H.W. Vesper, W.G. Miller, G.L. Myers, Reference materials and commutability,
[10] R.L. Canfield, C.R. Henderson, D.A. Cory-Slechta, C. Cox, T.A. Jusko, B. Clin. Biochem. Rev. 28 (2007) 139–147.
P. Lanphear, Intellectual impairment in children with blood lead concentrations

You might also like