Professional Documents
Culture Documents
X Ray Diffraction and Clay Mineral Identifica. Moore Reynolds
X Ray Diffraction and Clay Mineral Identifica. Moore Reynolds
Quantitative Analysis
1. The sample must be longer than the spread of the incident beam at the
lowest diffraction angles used. Therefore, two quantities need attention:
(1) the sample length, and (2) the angular divergence of the beam or
size of the divergence stit.
2. The sample must be infinitely thick at the highest diffraction angles
used. -
- The sample must be mounted in the diffractometer so that, for all
w
diffraction angles, the angle between the sample surface and the °
incident beam is equal to the angle between the sample surface and the
diffracted beam.
4. There must be no particle-size gradient between the top and bottom of
the sample. -
These factors can be difficult to control, but their effects on the accuracy
and precision of analyses are minimized if you adhere to two important
principles. -
- Sample Length
Diffraction intensity is proportional to the volume of
the sample irradiated,
The goniometer geometry fixes the theta-independent
width of the sample
area irradiated. The irradiated length of the sample
is proportional to Using, - -
. . so the irradiated area is equal to a
/'Bººmd'"fgº"ºº constant divided by sin0. The
- - - integrated thickness (depth)
. exposed to the incident beam is
proportional to sin8, Volume is
Sample, high 20 equal to area times thickness, so the
product of these two cancels the
Sample, low 20 angle-dependent character of each
if, and only if, the sample is
infinitely thick and longer than the
spread of the beam spot at the
Lost intensity lowest diffraction angles recorded.
atlow 26 The sketch in Fig. 9.1 shows a
Fig. 9.1.” Incident intensity loss at low diffraction sample at high and low diffraction
angles. angles. The sample does not
intercept the entire spread of the
incident beam at the low angle shown, and some diffracted
intensity will be
lost. If Rg is the goniometer radius in centimeters and
o is the angular
aperture of the divergence slit in degrees, then the length L of the area
irradiated is
L= Rotano/sing on
Table 9.1. Sample shape and size and the loss of diffraction intensity as a
function of 26 -
Intensity. rectangular — Intensity. circular
*28 - sample (4 cm long) sample (2.5 cm diam.)
2 0.20 0.10
4 0.40 0.20
6 0.60 0.29
8 0.80 0.39
10 1.002 0.49
15 1.00 0.73
20 1.00 0.88
25 1.00 0.95
30 1.00 0.95
Goniometer radius = 20 cm; incident beam slit = 1°. Fuil intensity.
“footprint” is unlikely. You could try to maintain al! samples at some, fixcd
length and standardize procedures with standard minera! slides of the same
length, keeping the divergence slit fixed, of course. This is fine for your
instrument, but you will not be able to use values published by others for
mineral intensity parameters unless the published values have been measured
with identical geometrical conditions. The best solution is to use fine slits at
low diffraction angles, where the intensity loss due to such slits is
unimportant, and lacger slits, consistent with the relations of Eq. (9.1), for the
higher angles. Record overlapping angular ranges with the two-slit systems
and then normalize one pattern to the same intensity basc as the other. This
procedure wilt produce a single diffractogram that contains no intensity
distortions due to sample length. Still another alternative is to avoid the use of
low-angle reflections for quantitative analysis.
Sample Thickness
“A different error affects high-angle peaks that can be serious and difficult to
control. The error arises from samples that are too thin, and, as you”will
discover with a little laboratory experience, it is very difficult to make
samples thick enough if you use procedures that are optimum in other
respects. .
Reproducible measurements of diffraction intensity from thin preparations
require that all samples have exactly the same thicknesses and identical mass
absorption coefficients. The mass absorption coefficient of a mixture depends
on its composition, which is the unknown quantity here, so these requirements
are manifestly impossible. The mass absorption coefficient is controiled by
the chemical composition of the material, so you might suggest obtaining
separate chemical analyses for your unknowns and using these data as the
basis for intensity corrections. Fair enough, but you still are left with the
difficult or impossible problem of producing uniform thicknesses for all the
302 X-Ray Diffraction and the Identifi
cation and Analysis of Clay Minerals
Consider the small imaginary crystallite nearest to the sample surface. The
diffracted radiation from this crystallite has been diminished twice by
absorption: once by absorption over the path of the incident beam, and once by
the path length out of the sample. We will denote the depth of this increment
not by distance but by g, the mass of material (expressed as grams per square
centimeter) that lies above it. The sum of the path lengths in and out of the
sample is proportional to 2g/sin0. So the intensity contributed by this
increment is described by the equation //1) = expC-21*g/sin0). Here, we
intend to derive the total integrated diffraction energy that leaves the sample
and is recorded as intensity. Pretty clearly, we need to integrate all increments
x of diffraction such as those shown by Fig. 9.3. The limit of the integration
must extend from x = 0 (the sample surface) to x = g (the total mass of the
sample) where Az is the incremental mass.
Remember, intensity of diffraction is proportional to the volume of the
sample irradiated. We showed before (Fig. 9.1) that the area of the sample that
is bathed in the incident beam is proportional to 1/sinf. Equation (9.3) gives
- the diffraction due to the integrated depth of irradiation, and you can see that
this quantity is proportional to sin0 (the term 1/24* is constant with respect to
6). Therefore, the volume, which is the product of area times depth, is
independent of 20. This is the great advantage of the diffraction geometry of
the flat-specimen diffractometer, namely, it produces no 28-dependent
absorption effect. This advantage is lost, however, if the sample is too short or
if a theta-compensating slitis used (i.e., a slit that automatically changes as 0
changes).
o (9.2)
The solution ofthe integral is
I _ sin®
T 2E
diffraction intensities
The formulation given here shows that the measured
not canceled by the product of
are inversely proportional to u* because * is
p* is the mean
the quantities that make up the irradiated volume. Here,
to the mineral mass -
sample mass- absorption cosfficient and is identical
lic.
absorption coefficient only if the sample is monominera
thicknesses (g) and
Equation (9.3) has been solved for different sample
Table 9.2, demonstrate the
diffraction angles (26). The results, shown in
clay mineral sample
effects of thin samples on high-angle peak intensity. A
r 6) covers about 11 cm?,
on a standard 47-mm membrane filter (see Chapte
is by no means an
and 5 mg/em? of sample amounts to 55 mg. This
intensity loss is 12% in
exceptionally thin sample. Note that at 25% 26,
material is decreased to
comparison with the lower'angles. If the amount of
peaks are used that cover a
33 mg, the errors become serious if analytical
these calculations is a good
large range of 20. The value of 4* (45) used for
average for illites. If a sample consists of mostly montmo rillonite or kaolinite,
severe becaus e these minera ls have ¡t* values
the thickness problem is more
-
of about 30 to 35.
mental arrang ement that can be used to correct
Figure 9.4 shows the experi
by thin samples. We use the Drever (1973)
for the intensity loss caused
sample substrate is a 40 x 50
method to prepare samples (see-Chapter 6). The able
(0.2 mm) to allow measur
mm thin-section cover glass and is thin enough
a polycrystalline quartz
transmission of the incident X-ray beam. Mount
angle at the position of
aggregate in the diffractometer, and set the diffraction
to produce a strong
the strongest quartz peak (26.65° 20) in order
absorp tion measur ements . First the substrate and
monochromatic beam for the
A measurement of their
then the sample plus substrate are treated as filters.
the transmission through the
filter factors will give us what we need. Measure
l layer and measure the
cover glass (7o) first, then add the clay minera a
9.4), thus obtaining
transmission of the quartz-diffracted beam again (Fig.
ed by substit uting these values into a
value for 7. The product p*g is obtain
rearranged form of the Lambert law:
w* g =-In(7/"0) D (9.4)
diffraction pattern on
Now that we know [L*g, we can proceed to run the
correct each of them,
our unknown, measure the required peak intensities, and
following example uses
at their appropriate 20 positions, by Eq. (9.3). The
the strong quartz peak
fictitious but reasonable numbers. The intensity of
is 10,000 cps. The clay layer is
transmitted through the clean glass (Fig. 9.4)
ion gives 6,000 cps. By the
added to the substrate and the same quartz reflect
Lambert law (Eq. [9.4)),
o
Thickness -. 28 = 10° 20=15° 20=25° 20=35°
(mg/em2) B -
1 0.64 0.50 0.34 0.26
0.95 0.87 071 0.59
5 0.99 0.97 0.88 0.78
10 100 - 1.00 0.98 0.95
15 . —1.00 1.00 1.00 099
The-correction factor for the intensity of a reflection at 50° 26 is obtained
by Eq. (9.3) (0.423 is the sine of 50%2 or 25%):
Sample Position
The sample must be correctly mounted in the sample holder assembly of the
diffractometer. Displacement of the sample surface away from the goniometer
axis is the largest source of error in peak position or d. But such errors have
líttie effect on diffraction peak intensity, and they are neglected here. The 2:1
alignment is another matter. By 2:1 alignment we mean positioning the -
sample so that the angle between the incident beam and the sample surface is
identical to the angle between the sample surface and the diffracted beam.
Both of these angles are called 8, and each must be exactly half of 20, which
is the angle between the undeviated incident beam and its diffracted
counterpart (see Fig. 3.7,p. 69).
The 2:1 alignment of the goniometer is carried out with precision slits and
plates and is so accurate in a properly aligned instrument that it contributes no
errors to peak intensity measurements. Clay mineral samples, however, are
not flat and smooth. There may be a ridge or high spot of material on one end
of the sample, or mineral debris on the reference pins or knife edges of the
sample holder assembly. Displacements caused by these irregularities result
in sample misalignment. Figure 9.5 shows the geometry of the 2:1 alignment
error that we call @. If o is a constant positive value, in the sense shown by
Fig. 9.5, intensity is diminished with decreasing 20. If @ is negative, intensity
is enhanced with decreasing 28. Equation (9.5) is the mathematical expression
that governs this effect:
1 — (l + sin(e—m)> )
Iy 2 sin(0 + 9) (9.5)
from different minerals are used to produce results normalized to 100%. But
absorption processes can cause errors in quantitative analysis for some highly
absorptive minerals, particularly if they are presentas large particles. *
As it is used in Eq. (9.6), 4* accurately describes absorption conditions
— only if all the grains in a sample are so small that absorption in a single grain
is no more than about 1% of the energy incident on that grain. If the grains or -
crystallites absorb more than this, microabsorption causes peak intensities to
become partiy dependent on particle size and on the individual absorption
coefficients of each of the constituents (Brindley, 1945). Bulk rock powders
whose grain sizes are approximately 10 ¡m surely produce poor results for
minerals with high absorption coefficients (for CuKa) like pyrite, siderite,
and iron oxides (Brindley, 1961). The highly absorbing minerals will be
underestimated, and for pyrite (compared to, say, quariz) the error is
approximately a factor of two (for CuKa radiation). Most clay minerals,
irradiated with CuKo. X-rays, are unaffected by microabsorption because their
crystallites are so thin. But Fe-rich chlorite and glauconite are heavy
absorbers, and their concentrations may be underestimated unless particle
thicknesses are equal to or thinner than 0.1 to 0.2 pm (Brindiey, 1961).
Sample-to-sample variations in the structure factor F can be'a major
source of error in quantitative analysis. F 'is controlled by the chemical
composition of the mineral and the atomic positions in the unit cell—the
crystal structure (see Chapter 3). Variations in atomic position are small for a
given mineral species, and have little effect on F at the moderate to low
diffraction angles where useful analytical reflections are found. Variations in
chemical composition, however, can cause large changes in F. The magnitude
of F is the sum of the atomic scattering amplitudes of the atoms within the
unit cell, the phase of each ray modified with respect to all others. The atomic
scattering factor, at a given diffraction angle, is proportional to the number of
electrons in an atom. Consequently, little change in F is caused by Mg-Al or
Si-Al substitutions because all the fully ionized forms of these atoms have the
same number of electrons. On the other hand, Fe substitution produces major
changes in F because Fe2+ has 24 electrons, whereas Al3+ or Mg2*, the usual
ions replaced, have only 10 electrons. Similarly, because of the large atomic
number of K, and therefore a high atomic scattering factor, the number of K+
ions in illite and glauconite has a major control on F.
We conclude that two common ions exert important controls on the values
of F for the clay minerals-Fe in octahedral coordination, and K in the
interlayer. Review in your mind the 2:1 clay mineral structures described in
Fig. 3.20 and Table 3.1, p. 100, and you will recall that both of these ions
occupy sites that lie on centers of symmetry on projection to Z. Conseguently,
the full complement of the scattering from such sites is added to or subtracted
from F, that is, their cosine terms in the structure factor calculation are equal
to"+1 or -1. The sensitivity of F to the composition of octahedral and
interlayer sites thus arises not only from the possible variations in electron
Chapter 9 Quantitative Analysis” — 311
Af
.
It = KW U (L) 0.8)
Where p* refers to the mixture. A simil
ar equation is derived for mineral 2,
and these two equations are divided
to get
10 = MA VBO09 ©9)
in which K and u* have canceled, or
I/l = ME ey
Fig. 9.6 Clay mineral mixture of illte, kaolinits, and EG-smectite. Background drawn
under $ 002 and K 001.
Chapter 9 - Quantitative Analysis 315
emphasize again that for this procedure to be valid, 0* must be the same for
the minerals grains in a given sample. If we cannot assume this, we cannot do
quantitative analysis. (This same system has been used for many of the
commón minerals in the JCPDS database using the strongest peak of
corundum as the reference intensity. Why wouldn't this work as a reference
for clay minerals?)
Application of the equations developed above is itlustrated by the
following hypothetical but realistic quantitative analysis. Figure 9.6 shows a
calculated diffraction pattern of an oriented clay mineral aggregate that has
been treated with ethylene glycol. Using criteria discussed in Chapter 7, we
have decided that the sample contains kaolinite, ¡llite, and dioctahedral
smectite. We must choose the peaksto use. None should be from the 1ow-
angle region, and all should be close together. Unfortunately, the strong illite -
003 reflection is almost perfectly superimposed on the smectite 005, so
neither of these is useful. Good choices are the smectite 003, illite 002, and
kaolinite 001 or 002 peaks. MIF values for these must have been measured or
calculated, and for the present, assume that they are accurately represented by
the data in Table 9.4. The MIF values have been normalized so that the
intensity of the illite 003 is taken as unity. The illite 003 reflection cannot be
used for the analysis we describe here because of interference problems, but
the 002 peak is in the clear and it nominally has about half the intensity of the
003, so we will use the illite 002 reflection and assign it the value of 0.51.
Table 9.4 contains the integrated peak areas or intensities (cps) for the
relevant peaks in Fig. 9.6 and demonstrates how the quantitative analysis is
computed. Peak areas are the product of the peak height times peak width at
half-height (sce Figs. 9.7 and 9.8).
Two examples are given in Tuble 9.4: one that uses the kaolinite 001
reflection, and one that uses the kaolinite 002 reflection. The results are not
”
316 X-Ray Diffraction and the Identification and Analysis of Clay Minerals ;
Y = (20)(MIF,/1)
where MIFp and / refer, respectively, to the pyrophyilite mineral intensity
factor and the intensity óf the standard pyrophyilite reflection for this
hypothetical analysis. The quotients //MIF for all the other minerals are
multiplied by Y to provide values for their weight percent abundance in the
sample. If the entire procedure is error-free, and if there is no amorphous
material present, the sum of the unknown minerals will add to 80%. For a real
analysis, sums between 70% and 90% should be considered “good.” The
analysis can be recast to a pyrophyllite-free format by multiplying all the
mineral percentages by 100/80.
There are two advantages to the orienting internal standard method. OF
most importance is the analysis total. This provides a figure of merit for the
analysis. If it is near 100%, then large errors are probably absent and minerals
have neither been missed nor misidentified. Canceling errors could be
present, and if you suspect such a situation, you will have to devise other tests
that are beyond the scope of this discussion. Another advantage of this
method lies in its suitability for analyses of only one or two minerals in a
mixture with others, Polycomponent mixtures may present such severe
interference problems that some minerals cannot be determined. For example,
smectite would be easy to measure in a mixture of smectite, chlorite, and
kaolinite, but chlorite and kaolinite mixtures may present insurmountable
difficuities. If you are lucky, perhaps all you really need to know-for your
application is the amount of smectite in cach sample from a suite. If this is the
case, the orienting internal standard method is the best procedure.
There are also some disadvantages to this method: It is difficult to grind
pyrophyllite to < 24m (reducing its megascopic crystals to the clay-sized
range requires extensive wet grinding and ultrasonic treatment); once
pyrophyllite has been mixed into a sample, it can not be removed; and the
minerals composing a natural sample are probably better and more uniformly
oriented than laboratory mixtures,
Compositional parameters
Mineral Reflection _ Intensity 4(cps) - per (SLA
O N1
OH)>
Montmorillonite 003 3110 - 0.1 Fe: ethylene elyco]
Montmorillonite 005 4070 0.1 Fe; ethylene glycol
Saponite - 003 1920 0.0 Fe; ethylene glycol
Saponite 005 4950 —.0 Fe; ethylene glycol
Saponite 003 620 1.0 Fe; ethylenc glycol
Saponite 005 6070 1.0 Fe; ethylene glycol
Tilite 002 1900 0.75 K; 0.1 Fe
Hlite 003 3940 0.75 K; 0.1 Fe
Chlorite 003 4430 1.0 Fe 2:1 tayer and
- “1.0 Fe intertayer
Vermiculite 005 4140 0.0 Fe; Mg-wtr interlayer
Vermiculite 005 5000 1.0 Fe; Mg-wrr interlayer
Kaolinite 002 8610 Composition can't be changed -+*
Kaolinite 003 830 Composition cant be changed *
“The constant in Eq. (9.6) has been adjusted to provide 3940 cps for the illiic 003 peak
assuming conditions of u* = 45, 0* = 12°, and CuKet radiation.
O-compensating slit causes an intensity variation that is proportiona! to sin6,
Consequently, valid reference intensities should be obtained by maltiplying
each of the intensities of Tables 9.5 and 9.6 by an appropriate value of sin6.
Mixed-layered clay minerals are often difficult to deal with because of peak
interferences from 'other clay minerals. For example, illite/smectite is-often -
mixed with discrete itlite, and if the composition of the illite/smectite is near
the illite end-member, resolution of the requisite reflections: may be
impossible by ordinary XRD methods. Similarly, corrensite usually contains
discrete chlorite that causes the same kinds of interference problems.
Several methods can be used to determine the percent illite in
illite/smectite, and we describe two of them that are based on the ethylene
glycol-solvated condition (Table 8.3). The intensity of the illite/smectite
003/005 mixed-layered peak'is almost independent of composition (see Table
9.6) and ordering type. This peak is the best to use if no interference is present
due'to discrete illite, which, unfortunately, is very common. Quartz also has a
significant interference on this peak, but that can be eliminated by measuring
the integrated area óf the quartz 101 at 20.85° 26, multiplying by 4.3, and
subtracting that quantity from the integrated area of the clay mineral
reflection near 26.65%. If both illite (or smectite) and illite/smectite are
present, the illite/smectite 002/003 reflection can be used, although its
i|
|
320 X-Ray Diffraction and the Identification and Analysis of Clay Minerals
of the low-angle
half of the peak (sce Fig. 9.9). This proce
dure also
minimizes interference from the low-angle
tail of the discrete ¡lite 002
reflection, although this interference makes this metho
d unwor
kable if the
illite/smectite is rich in'the illite end-member
. IF discrete illite and discrete
- smectite are not present, the best procedure
is to use the illite/smectite
0037005 values shown in Table 9.6.
322 X-Ray Diffraction and the Identification and Analysis of Clay Minerals
intersecting arcs, and the arc portions near the center of the image are much
broader than they are at the extremities of the picture. You have convolved a
complicated two-dimensional distorted function that is characteristic of your
poor lens onto the image of a sharp orthogonal grid. Every part of the image
of the sharp grid passed through every part of your cheap lens to produce the
final convolution. Then you photograph a friend ard obtain a strange portrait
indeed. Can it be cleaned up? Yes, by the process of deconvolution. If the
original grid is very sharp and strictly orthogonal, the poor image of that grid
is solely a signature of the lens. That distorted signature can be removed from
the friend’s portrait by deconvolution (not simple subtraction) to produce an
image that, ideally, is what the picture would have looked like had the lens
been distortion-free. The word “ideally” is used here because the.
mathematical process of deconvolution produces intensity ripples that
combine to produce a nonuniform intensity background and can blur the
edges of some portions of the sharp image.
No diffraction goniometer is perfectly focused. A profile of a diffraction
peak that is infinitely sharp (from an infinitely thick perfect crystal) will show
CuKa-CuK& asymmetrical broadening, and asymmetrical and symmetricai
broadening caused by a number of instrumental factors (see Klug and
Alexander, 1974). Your instrument has convolved these onto the diffracted
intensity. But that experimental reflection (which is your instrument's
signature) can be deconvolved from any portion of another experimental
diffraction pattern to produce the patiern from a perfect diffractometer. This
Tllite 002 - Chlorite 003
1 !
- 16 18
20
Fig. 9.7, Peak arcas for 67% ilfite and 33% chlorite. Vertical lines represent peak
height above background. Horizontal lines, labeled f, depict the peak width at half-
height. -
324 X-Ray Diffraction and the Identification and Analysis of Clay Minerals
and are”
symmetrical, the simplest way to measure their areas is to integrat
e from the
tail of a reflection to the midpoint of the saddle between thera.
This is shown
—
Chlorite 003
; L— *
ur Tilite 002 K
T T I
19 -
26
Fig. 9.8, A mixture of 75% chlorite and 25% iite. Vertical lines represent
above background. Horizontal lines, labeled B, which peak height
depict the peak width at half-height. is equival ent 1o FWHM and
-
by the two areas, one of which is shaded. The result
s from this integration
are
shown in Table 9.8. The vertical lines throu
gh the peaks represent their
heights,-and the horizontal lines through them,
labeled B, mark the width at
half-height. Relative areas were obtained by multi
plying the widths by the
heights. Table 9.8 gives á comparison between the
correct relative peak areas
and those obtained by the two different merho
ds described.
The results are poor for integrated area probably
because. for this case,
both of the peaks are asymmetrical toward the low-a
ngle direction. Note that
the peak-height-times-B results are better. This
may seem to be a “rough-and-
ready” method, but, as we will see later, it
often produces the best results of
any of the techniques save those that require
sophisticated computer
techniques of curve-fitting.
Figure 9.8 dépicts a more difficult analysis. The calcul
ated diffraction
pattern is based on 75% chlorite and 25% illite.
The illite 002 reflection is
weak and distorted by the tail of the iniense chlori
te peak. The peak areas
were isolated by projecting the low-angle tail of the
chlorite reflection under
the illite peak to produce the background for the
illite reflection. Integrated
peak areas were measured by planimeter, divide
d by the appropriate mineral
reference intensities (Table 9.5), and normalized
(Table 9.4), yielding chlorite
70% and ¡Hite 30%. The composition of the simulated mixture also was
determined by multiplying peak heights by B. Clearly, the illite 002 peak is
too distorted to allow a direct measurement of B, so a value was obtained for
B/2 (see Fig. 9.8) and doubled to give B. The composition based on this
method is 73% chlorite and 27% illite, which is closer to the correct value
than the result based on integrated arcas. Again
we see the superiority of the
326 X-Ray D¿Ú'ractí¿n and the Identification and Analysis
f Clay Minerals
—
and illite diminish as the
illite/smectite becomes more expandable because the
002/003 reflection is
displaced farther from the illite 002. But as you can
see from Fig. 9.9,
illite/smectite minerals containing more than approximately
=
80% illite cannot be “
accurately determined in the presence of moderate to large
amounts of discrete
illite. .
We have seen that superior results are obtained for measu
==
rements of peak
area by the use of the peak-height-half-width method. Use
this technique
whenever you can: However; it is not always applicable.
For example, it
Tilite 002 — B—
/S 002/003
S
T B—
—
16 29 18 20
==
of the peak itself and interference from the discrete ¡llite peak.
Chapter9 — Quantitative Analysis 327 —
cannot be used for a peak that lies on or near tiie saddle between
two large
reflections because the central peak will be distorted on both
sides and thus
provide no opportunity to use the half-width (B/2) as we did for our
examples.
But why is the height-width method less sensitive to errors in
selecting
background than the mere straightforward measurements of
integrated area?
To answer this question, we need to understand that a very large portion
of
the area of a peak lies in its “tails,” which are just the regions where
the most
uncertainty will occur in fitting a profile to its underlying backgro
und. Errors
in height or f give proportional errors in their product, but errors in selectin
g
background produce exponentially enlarged errors in integrated arca.
1. The sample must be loag enough for the divérgence slit used and
for .-7 *
the lowest diffraction angle recorded.
10. You can obtain calibration factors for relative mineral intensities
by
measuring peak intensities from artificial mixtures. But why,
when
calculated calibration factors are more widely applicable, easier
to
obtain, and just as good or perhaps better?
Chapter9 — Quantitative Analysis . 329
REFERENCES
Brindley, G. W. (1945) Effect of grain or particle size on X-ray reflections from mixed
powder: Phil. Mag . 36, 347-69. .
Brindley, G. W, (1961) Quantitativo analysis of clay mixtures: in Brown, G, ed., The X-Ray
Identification and Crystal Structures of Clay Minerals: Mineralogical Sociely, London,
489-516. -
Brindley, G. W. (1980) Quantitative X-ray mineral analysis of clays: in Brindley G. W., and
Brown, G., editors, Crystal Structures of Clay Minerals and Their X-Ray Identification:
Monograph 5, Mineratogical Society, London, 411-38.
Chung, F. H. (1974) Quantitative interpretation of X-ray diffraction patterns of mixtures. L
Matrix flushing method for quantitative multicomponent analysis: J. Appl. Crystallogr.
7,519-25.
Cody, R. D., and Thompson, G. L. (1976) Quantitative X-ray powder diffraction analysis of
clays using an orienting internal standard and pressed disks of bulk clay samples: Clays
and Clay Mincrals 24, 224-31. -
Drever, J. 1. (1973) The preparation of ociented clay mineral specimens for X-ray diffraction
analysis by a filter-membrane peel technique: Ámier, Minerl. 58, 553-54.
Griffiths, J. C. (1967) Scientific Method in Analysis of Sediments: McGraw-Hill, New York,
Hughes, R. E., Moore, D. M., and Glass, H. D. (1994) Qualitative and quantitative analysis of
clay minerals in soils: in Amonette, J. E., and Zelazny, L. W., editors, Quantirarive
Methods in Soil Mineratogy: SSSA Miscellaneous Publications, Soil Science Society of
America, Madison, Wis., 330-59.
James, R. W. (1965) The Optical Principles of the Diffraction of X-Rays: Comell University
Press, Ithaca, New York, 664 pp.
Klug, H. P. and Alexander, L. E. (1974) X-Ray Diffraction Procedures: Wiley, New York,
96 pp.
Mossman, M. H., Freas, D. H., and Bailey, 5. W. (1967) Orienting internal standard method
for clay mineral X-ray analysis: in Clays and Clay Minerals, Proc. 15th National. Conf..
Pittsburgh, Pa., Pergamon Press, New York, 441-53.
Reynolds, R. C., Jr. (1980) Interstratified clay minerals: in Brindiey, G. W., and Brown. G.,
editors, Crystal Structures of Clay minerals and Their X-Ray Identification: Monograph
5, Mineralogical Socicty, London, 249-303.
Reynolds. R. C., Jr. (1985) NEWMODO, a Computer Program for the Calculation of One-
Dimensional Diffraction Patierns of Mixed-Layered Clays: R. C. Reynolds, Jr., § Brook
Rd., Hanover, NH. * - -
Reynolds, R. C., J£. (1986) The Lorentz factor and preferred orientation in oriented clay
aggregates: Clays and Clay Minerals 34, 359-67. -
Reynolds, R. C.. Jr. (1989) Principles and techniques of quantitative analysis of clay minerals
by X-ray powder diffraction: in Pevear, D. R., Mumpton, E. A., cditors, Quantitative
Mineral Analysis of Clays: Werkshop Lectures, Vol. 1, Clay Minerals Society, Boulder,
CO,4-36.
Schultz, L. G. (1960) Quantitative determination of some aluminous minerals in rocks: Clays
and Clay iinerals 7, 216-24.