Professional Documents
Culture Documents
Mphya6 000037 000724 - 1
Mphya6 000037 000724 - 1
Mphya6 000037 000724 - 1
Key words: modulation transfer function 共MTF兲, resolution, digital radiography, linear systems
analysis, noise power spectrum 共NPS兲, noise, edge, image quality
I. INTRODUCTION function 共LSF兲, taking the Fourier transform of the LSF, and
use of various noise reduction and fitting
The modulation transfer function 共MTF兲 is the most widely techniques.6,11,12,15–20
accepted measure of the spatial resolution response of medi- The slit11 and edge12 response methods are the two most
cal x-ray imaging detectors and has proved to be a valuable commonly used and accepted techniques for measurement of
tool for determining and comparing detector performance.1–4 the MTF, with the edge-response method being preferred
Although the methods used to measure the MTF have 共with adoption by the IEC兲 共Ref. 21兲 for various reasons,
changed slightly over the past several decades,5–14 the under- including simpler construction and less sensitivity to
lying premise has remained a constant: A precisely machined misalignment.12 However, it is well documented that the
object or test device is imaged and the subsequent response edge-response method is also vulnerable to a host of poten-
of the detector is used to obtain the MTF using several steps, tial problems that could influence the measurement and re-
which generally include determination of the line spread sult in inaccuracies. Inaccuracies may result from errors in
724 Med. Phys. 37 „2…, February 2010 0094-2405/2010/37„2…/724/12/$30.00 © 2010 Am. Assoc. Phys. Med. 724
725 Kuhls-Gilcrist et al.: Accurate MTF measurement using noise response 725
the calculated edge angle,12,22,23 noise,12,23,24 influences of phors are implemented, resulting from the emission and re-
scattered radiation,24,25 use of finite-element absorption of secondary x rays.38,46 Parallel processes also
differentiation,10,22 profile misregistration and phasing occur in photoconductive 共direct兲 x-ray detectors.47 Cas-
errors,22,23,26 truncation of the LSF tails and incorrect caded linear systems theory has been an extremely useful
normalization,27 and windowing and processing.12,20,22 Diffi- tool for characterizing and optimizing detector performance
culties also arise when comparing measurements with pub- and has been used to accurately predict the signal and noise
lished results of other investigators, as slight differences in transfer properties of a wide variety of digital x-ray imaging
object, acquisition, and/or processing techniques could easily technologies.33–40 Excellent agreement has been observed
alter the results.23,24,27–32 Significant variations have even between theoretical calculations and measured results, even
been shown to occur when different individuals measured the with the required estimation of many system performance
MTF of the same data set.32 The aim of this work is to parameters.
develop a new method for measurement of the MTF of digi- Parallel cascade theory was used to generalize the re-
tal radiographic systems that is less susceptible to such in- sponse from digital x-ray detectors, which were considered
fluences, by using the intrinsic noise response of the detector. to consist of four fundamental stages: X-ray absorption,
Cascaded linear systems methods33 have been used to ac- x-ray-to-light conversion, light-to-electron conversion, and
curately predict the signal and noise transfer properties of a digitization. Each stage consists of several interaction sub-
wide variety of digital x-ray imaging technologies including stages, where the quanta undergo one of three processes:48
x-ray image intensifiers,34 direct and indirect flat panel de- Amplification, blurring, or addition of noise. A brief sum-
tectors 共FPD兲,35,36 and CCD/EMCCD-based detectors.37–40 mary of the parallel cascade model analysis is presented be-
Careful inspection of the resultant noise response of such low. Additional details can be seen in a number of
systems from a theoretical framework, as described by the works.33–40
two-dimensional noise power spectrum 共NPS兲, indicates that Amplification processes may involve binary selection or
the resolution response 共i.e., the MTF兲 is inherently included conversion of quanta from one form to another. The signal at
in such information.41,42 In this work, we present an exact the output of stage i of an amplification process is described
relationship using a generalized cascaded linear systems by
analysis. This relationship was then utilized to measure the
presampled detector MTF from the noise response alone, ⌽i共u, v兲 = gi⌽i−1共u, v兲, 共1a兲
without use of a test object.
where gi is the average gain of stage i and ⌽i−1共u , v兲 is the
The accuracy of the noise-response method was investi-
signal in the spatial frequency coordinates 共u , v兲 at stage
gated using image simulations of a modeled detector system,
i − 1 共i.e., at the input of stage i兲. The NPS at stage i is given
in which the “true” MTF was known exactly. Measurements
by48
were also performed on several different clinical imaging
technologies, including an x-ray image intensifier 共XII兲 and a NPSi共u, v兲 = gi2NPSi−1共u, v兲 + gi2⌽i−1 共1b兲
FPD, as well as a custom-built solid-state x-ray image inten-
sifier 共SSXII兲,40,43–45 which is an electron-multiplying CCD where gi2 is the gain variance of stage i, ⌽i−1 is the average
共EMCCD兲-based detector with very high-resolution capabili- signal at stage i − 1, and NPSi−1共u , v兲 is the NPS at stage
ties 共greater than 10 lp/mm兲. Results from the noise-response i − 1 in the spatial frequency coordinates 共u , v兲. It follows
method were compared with those obtained using the stan- that consecutive binomial processes can be combined into a
dard edge-response method. single binomial process with a probability equal to that of the
Potential benefits of this method, including elimination of product of the probabilities of each individual stage.37 Blur-
the requirement for a test object 共which requires careful con- ring processes may be either stochastic 共caused by the spatial
struction and precise alignment兲 and the ability to acquire the spreading or scattering of stochastic quanta兲 or deterministic
MTF in all directions simultaneously, will be discussed. Ef- 共e.g., pixelization resulting from digital sampling兲. The sig-
fects of potential noise sources that have not been considered nal at the output of stage i of a blurring process is given by
in the theoretical development of the noise-response method
and aliasing will also be discussed. ⌽i共u, v兲 = Ti共u, v兲⌽i−1共u, v兲, 共2a兲
II. METHODS where Ti is the MTF at stage i. The NPS after a stochastic
blurring stage can be written as48
II.A. Theoretical development of the noise-response
method NPSi共u, v兲 = 关NPSi−1共u, v兲 − ⌽i−1兴兩Ti共u, v兲兩2 + ⌽i−1 共2b兲
Cascaded linear systems theory models the overall re- and for a deterministic blurring stage
sponse of linear and shift-invariant detector systems as a
contributing sum of individual stages, each of which consists NPSi共u, v兲 = NPSi−1共u, v兲兩Ti共u, v兲兩2 . 共2c兲
of a single interaction process.33 Progression of quanta
through the imaging chain is generally serial in nature, in An additive noise stage does not change the signal as the
which the output of one stage is the subsequent input of the images are assumed to have been offset corrected. The NPS
next. Parallel processes occur when x-ray converting phos- after an additive noise stage is given by
NPSi共u, v兲 = NPSi−1共u, v兲 + NPSAdd共u, v兲. 共3兲 gPE = W−1兵E − Ek Pkk关1 − f k兴其, 共6兲
Using Eqs. 共1a兲, 共1b兲, 共2a兲–共2c兲, and 共3兲 at each appropriate
substage of the imaging chain enables calculation of the TPE共u, v兲 = 兵E − Ek Pkk关1 − f kTk共u, v兲兴其
frequency-dependent signal and noise at the detector output. ⫻兵E − Ek Pkk关1 − f k兴其−1 , 共7兲
The NPS for a generic digital x-ray detector 共the stages of
which are consistent with a wide range of detector technolo- AS共u, v兲 = 兵E − Ek Pkk关1 − f kTk共u, v兲兴其2
gies兲 can then be written at each stage as described below.
⫻兵E2 − 2EEk Pkk关1 − f kTk共u, v兲兴
冉 冊
⬁ ⬁
n m
+ 兺 兺 NPSPRE u⫾
⌬x
,v ⫾
⌬y
NPS4共u, v兲 = NPSA共u, v兲
冉 冊
m=1 n=1
⬁ ⬁
n m
+ NPSAdd共u, v兲, 共11b兲 + 兺 兺 NPSA
m=1 n=1
u⫾
⌬x
,v ⫾
⌬y
38
where
+ NPSB共u, v兲
NPSPRE共u, v兲 = ⌬x2⌬y 2g24g23gPE
2
g1⌽0A−1
S 共u, v兲T4共u, v兲
2
冉 冊
⬁ ⬁
n m
⫻T23共u, v兲TPE 共u, v兲T22共u, v兲 兺 兺 NPSB
2
+ u⫾ ,v ⫾
m=1 n=1 ⌬x ⌬y
+ ⌬x 2
⌬y 2g24g3gPEg1⌽0T24共u, v兲 共11c兲
+ NPSAdd共u, v兲, 共14b兲
is the presampled NPS, ⌬x and ⌬y are the pixel width in the
horizontal and vertical directions, n and m are the number of where
replicates, g4 is the electron-to-digital-number conversion
factor which is assumed to have no variation,51 T4共u , v兲 is the NPSA共u, v兲 = ⌬x⌬yg̃A−1
S 共u, v兲TSys共u, v兲⌽4 共15a兲
2
aperture MTF associated with integration of the signal over a
finite region 共e.g., the sinc function of the detector element and
size兲, and NPSAdd共u , v兲 is the additive electronic noise.
The system MTF 共TSys兲 and the effective system gain 共g̃兲 NPSB共u, v兲 = ⌬x⌬yg4T24共u, v兲⌽4 共15b兲
in units of DN per absorbed x-ray photon can be written as
are the correlated and uncorrelated noise components, re-
TSys共u, v兲 = T4共u, v兲T3共u, v兲TPE共u, v兲T2共u, v兲 共12兲
spectively, and ⌽4 is the average digital detector value for an
and input of ⌽0 quanta/ mm2.
Other investigators have shown that sum of uncorrelated
g̃ = g4g3gPE . 共13兲
noise and its alias produces a white spectrum in the fre-
Combining Eqs. 共11a兲–共11c兲, 共12兲, and 共13兲 we can then sim- quency domain.52,53 Hence, the digital output NPS can be
plify to the following expressions: written as52
冉 冊
冤冢 冣 冥
n m
⬁ ⬁
2
TSys u⫾ ,v ⫾
T2 共u, v兲 ⌬x ⌬y g4
+兺兺
冉 冊
NPS4共u, v兲 = ⌬x⌬y g̃ Sys + ⌽4 + NPSAdd共u, v兲, 共16兲
AS共u, v兲 m=1 n=1 n m FP
AS u ⫾ ,v ⫾
⌬x ⌬y
where F P is the pixel fill factor. If the aliasing of the corre- term which scales proportionally to ⌽42 should also be in-
lated noise is negligible, or, similarly, if cluded, as a result of variations in detector sensitivity. How-
ever, all images used in this analysis were corrected for gain
2
TSys 共f N兲 Ⰶ 1, 共17兲
variations using an average of 30 flat-field images at each
which is a reasonable assumption for indirect detectors,54,55 exposure level, resulting in a negligible structure noise com-
Eq. 共16兲 simplifies to53 ponent. Hence, this term was omitted. The slope of a linear
NPS4共u, v兲 = ⌬x⌬y g̃ 冋 2
TSys 共u, v兲 g4
+
AS共u, v兲 F P
册
⌽4 + NPSAdd共u, v兲.
fit of the NPS plotted versus mean signal 共⌽4兲 then goes as
II.B. Experimental procedure whereas the quantum noise scales proportionally with the
detector entrance exposure. A linear regression fitted to a plot
The procedure used for determining the presampling MTF
of NPS versus signal was used to separate the two noise
from the detector noise response is described below.
components 关Eq. 共18兲兴 with the slope representing the quan-
tum noise per unit signal. To provide a good estimate of
II.B.1. Step 1: Preparation of the imaging setup regression coefficients, the NPS was measured across a range
A standardized x-ray spectrum56 was used to facilitate of exposures, which encompassed a majority of the dynamic
comparisons as the response of digital x-ray detectors has range of the detector.
been shown to have a significant dependence on the energy
of the incident x-ray beam.57 The desired spectra were
achieved using a specified thickness of added aluminum fil- II.B.4. Step 4: Determination of the MTF
tration 共Alloy 1100兲 共Ref. 58兲 and by adjusting the kVp of The system MTF 关TSys共u , v兲兴 was then determined from
the generator to achieve a specific half value layer. A stan- the quantum noise component of the NPS, using Eq. 共19兲. A
dard measurement geometry was also used, as defined by the functional form of TSys共u , v兲, based on underlying physical
IEC 共Ref. 21兲 to minimize the effects of scatter, with as large principles, was used in the fit of the slope or quantum noise
a source-to-image distance as the imaging system would al- per unit signal as a function of spatial frequency. Fitting
low 共greater than 100 cm兲 and with the added filtration functions used for this analysis were based on a Gaussian
placed as close as possible to the source. To simplify the mixture model60 and the complementary error function
measurement, the x-ray scatter reduction grid was removed 共ERFC兲 共Ref. 61兲 and are shown below, in one dimension for
共when appropriate兲. Clinical digital x-ray detectors typically simplicity
implement highly nonlinear image processing in an attempt
to improve perceived image quality. This image processing
could potentially affect the measurement and was disabled to
FGaussian Mixture共f兲 =
h1
AS共f兲
冋 N=1,2, or 3
兺 h 2i
冊册
i=1
冉
avoid any complications.
共f − h3i兲2 2
⫻exp − + h5 共21兲
II.B.2. Step 2: Measurement of the NPS h 4i
NPS measurements were also done as prescribed by the and
冋 冉 冉 冊冊册
IEC.21 Flat-field images were acquired using the standard 2
spectrum and geometry with no test object in the beam. Re- h1 1 f
FERFC共f兲 = ERFC h2 log + h5 , 共22兲
sponses of the detectors used in this study were linear rela- AS共f兲 2 h3
tive to the number of input quanta, so the implementation of where As共f兲 is the calculated frequency-dependent Swank
a conversion function 共for linearization兲 was unnecessary. A factor, hi corresponds to the ith fitting parameter, and N is the
central region of interest 共ROI兲 was selected in the image for number of Gaussian forms in the Gaussian mixture model.
analysis and broken into 25 half-overlapping 256 Derivation and calculation of AS共u , v兲 has been described by
⫻ 256 pixel regions. To improve statistics of the measure- Hadjok et al.,49,62,63 demonstrating good agreement with
ment, 30 flat field images 共providing more than 17⫻ 106 in- Monte Carlo and measured results. A total of four different
dependent image pixels兲 were used for analysis. The NPS fitting functions 关Eq. 共21兲 with N = 1, 2, and 3 and Eq. 共22兲兴
was then calculated using59
冏兺 兺
were used and the one providing the best fit was used for
M 256 256 subsequent analysis. Different fitting functions were used to
⌬x⌬y
NPS共un, vk兲 = 兺
M · 256 · 256 m=1
关I共xi,y j兲 ensure very high precision in the fitting technique. Frequen-
i=1 j=1 cies were sampled at an interval spacing of f N / 128. The
冉 冉 冊冊
signal at pixel location 共xi , y j兲, Ī共xi , y j兲 is the average signal
1 f
at pixel location 共xi , y j兲, and un and vk are the spatial fre- TSys共f兲 = ERFC h2 log 共23兲
quency values in the horizontal and vertical directions, re- 2 h3
spectively, which are sampled at an interval of f N / 128. but could also be determined indirectly using the slope data
and fit parameters h1 and h5
冑
II.B.3. Step 3: Separation of additive
AS共f兲共slope共f兲 − h5兲
instrumentation noise TSys共f兲 = , 共24兲
h1
Additive instrumentation noise is generally “white” 共i.e., a
constant as a function of spatial frequency兲 and does not which is more robust in allowing the data to deviate some-
contain information of the detector MTF. The additive instru- what from the fitted result. If results provided by Eqs. 共23兲
mentation noise is constant as a function of exposure, and 共24兲 were to show a significant discrepancy, it would be
再 冎
For the noise-response analysis, 30 flat-field images were
1 acquired at six different mA s values spanning a majority of
T“True”共u, v兲 = exp − 关共2⌬xu兲2 + 共2⌬y v兲2兴
2 the 12 bit dynamic range of the detectors. Images were also
acquired of an opaque edge, slightly angulated at 1°–3° rela-
sin共⌬xu兲 sin共⌬y v兲
⫻ , 共25兲 tive to both the horizontal and vertical directions of the pixel
⌬xu ⌬y v matrix. All images were gain and offset corrected. MTF mea-
where is the standard deviation of the Gaussian blurring surements for both the noise-response and edge-response
function. methods were done on three separate occasions to gauge the
Edge images were also simulated for the detector model uncertainty of the measurements.
described above. Images of an edge test object were gener-
ated in a manner similar to that described by Carton et al.24
The edge was centered in the image, oriented at an angle of III. RESULTS AND DISCUSSION
2° relative to the pixel rows, and assumed to absorb 95% of III.A. Image simulations
the incident x rays. An average of ten frames was used to
The NPS was determined for the simulated flat-field im-
reduce the noise content in the images.
ages in a central 768⫻ 768 region, providing 25 overlapping
ROIs per image or 750 overlapping ROIs for the entire 30
II.D. Experimental measurements
image sequence. The combined horizontal and vertical re-
For further validation of the method, a comparison was sponse was taken by averaging along the zero frequency axes
made between the MTFs determined for digital x-ray detec- and the three adjacent rows/columns. This averaged NPS is
tors by the noise-response method and an established mea- shown in Fig. 1 for the six different incident x-ray exposures
surement technique. Images were acquired for three different used.
TABLE I. Summary of the goodness of fit of the parameters of the noise-response method for each of the three detector systems. The average 95% confidence
intervals for the regression coefficients, obtained by fitting the NPS versus signal to a linear curve at each spatial frequency, are given in the second column.
Four different fitting functions were then used to fit the resulting slope data as a function of spatial frequency and the function giving the best fit that was used
to provide the presampled MTF is indicated in the third column. The average percent deviation between the slope data and the best fit function is given in the
last four columns for four frequency intervals.
Fourier transform兲20 and is indicative of the inherent error in assuming the same functional form 关e.g., using only Eq. 共22兲
the edge-response MTF procedure used for this analysis.22 as the fitting function兴 for each detector. As a result, errors
due to the uncertainties in the linear regressions and fitting
III.B. Experimental measurements technique were negligible. The phosphor thickness was var-
ied by ⫾100 m from the manufacturer specified value in
A similar analysis was also done for the three actual de-
order to assess potential errors resulting from the frequency-
tector systems as that described in Secs. II B and III A. Zero-
dependent Swank factor estimate, which is dependent on the
frequency NPS values were omitted from the fitting proce-
phosphor thickness. Results suggest that the corresponding
dure due to the potential for a sharp increase at very low
error has less than a 3% effect on the measured MTF at the
spatial frequencies as a result of artifacts.64 The combined
Nyquist frequency 共even less at lower frequencies兲, which is
horizontal and vertical response was taken by averaging
in agreement with the assertion that the Swank factor is
along the zero frequency axes and the three adjacent rows/
largely insensitive to converter thickness for quantum effi-
columns. For each detector, the quantum NPS per unit signal
ciency values greater than 0.5.62 Others have shown that dif-
was fitted with Eqs. 共21兲 and 共22兲. The fitting function that
ferent methodologies for measuring the NPS 共differing in the
provided the best fit was used for further analysis. Careful
number of ROIs, ROI sizes, detrending techniques, and
attention was paid to ensure that a good fit was obtained at
methods used for extraction of 1D data from the 2D NPS兲
all spatial frequency values 共which were sampled at an inter-
provide consistent results with less than a 2% variability,65
val of f N / 128兲 in order to avoid potential systematic errors
suggesting that the slightly different techniques used for
resulting from poor fitting techniques. Table I summarizes
measuring the NPS would not significantly affect the results.
the goodness of fits for each of the three detectors. The 95%
Figures 5–7 show the measured presampled MTF using
confidence intervals of the regression coefficients, taken as a
the noise-response and edge-response methods for the XII,
percentage of the regression coefficients, were averaged for
FPD, and SSXII, respectively, plotted up to the Nyquist fre-
all spatial frequencies, and the average was less than 4% for
quency, which was different for the three detectors. Error
all three detectors. The “best fit” fitting function matched the
bars are representative of the standard deviation of the three
data within 1% across all spatial frequencies for each of the
measurements. Uncertainty in the edge-response method was
three detectors studied, which was a slight improvement over
FIG. 5. The XII presampled MTF measured using the noise and edge- FIG. 6. The FPD presampled MTF measured using the noise and edge-
response methods plotted up to the Nyquist frequency of 4.2 cycles/mm. response methods plotted up to the Nyquist frequency of 2.5 cycles/mm.
FGaussian Mixture共f兲 =
h1
AS共f兲
冋兺 100
h2
冊册
n=−100
冉
2
共兩f − 2nf N兩 − h3兲2
⫻exp − + h5 ,
h4
共26兲
Accurate MTF measurements were obtained using the an XII, FPD, and SSXII兲 demonstrated agreement between
noise-response method on three very different detector tech- the noise-response and edge-response methods within ex-
nologies, using the same generalized cascaded-systems perimental uncertainty, with discrepancies likely resulting
model, supporting the notion that the generalized form of the from errors inherent in the edge-response MTF procedure.
NPS, as given in Eq. 共14b兲, accurately describes a wide Compared to current measurement methods, the noise-
range of detectors. Cascaded-systems models by others, for a response method simplifies the MTF determination by elimi-
wide variety of detector technologies 共including direct detec- nating the need for manufacture and alignment of precisely
tors兲, also indicate general agreement with Eq. 共14b兲. There- machined test objects, thereby eliminating inaccuracies that
fore, verification of Eq. 共14b兲 appears to be unnecessary for result from the use of such objects and subsequent analysis
well-behaved systems. In each instance, the NPS contains a of the resulting images. Further, the two-dimensional MTF is
correlated and uncorrelated component which scales propor- readily obtained with the noise-response method, whereas
tionally with exposure, with the correlated component being traditional edge and slit methods are inherently one-
passed by the system MTF. Appropriate isolation of this dimensional.
component enables recovery of the system MTF from the
NPS. Admittedly, it is difficult to foresee all physical pro-
cesses and to include each in the model. Other potential ef- ACKNOWLEDGMENTS
fects that have not been explicitly included, such as lag or
This work was supported in part by the National Institutes
detector multiplicative noise, would scale the NPS equally at
of Health under Grant Nos. RO1-EB002873 and RO1-
all spatial frequencies,66 and therefore would not affect the
EB008425 and an equipment grant from Toshiba Medical
measured MTF, as they would get absorbed in the fitting
Systems Corp.
coefficients h1 and h5 in Eqs. 共21兲 and 共22兲. Afterglow should
have the same frequency components as the light 共or elec- a兲
Electronic mail: atkuhls@buffalo.edu
1
trons兲 emitted “immediately” after x-ray absorption and K. Rossmann, “Point spread-function, line spread-function, and modula-
therefore would also not affect the measurement. Only noise tion transfer function, tools for the study of imaging systems,” Radiology
93, 257–272 共1969兲.
sources that were to vary with spatial frequency would affect 2
I. A. Cunningham, “Applied linear-systems theory,” in Handbook of
the measurement, and it is difficult to envision any such pro- Medical Imaging: Physics and Psychophysics, edited by J. Beutel, H. L.
cesses. Kundel, and R. L. V. Metter 共SPIE, Bellingham, 2000兲, Vol. 1, pp. 82–
The two-dimensional NPS provides the noise response in 156.
3
E. Samei and M. J. Flynn, “An experimental comparison of detector
all directions. As such, the MTF measured using the noise- performance for computer radiography systems,” Med. Phys. 29, 447–
response method can be determined not only in the horizon- 459 共2002兲.
tal and vertical directions 共an average of which was used in 4
B. Lazzari, G. Belli, C. Gori, and M. R. D. Turco, “Physical characteris-
this analysis for comparison with the edge-response method兲, tics of five clinical systems for digital mammography,” Med. Phys. 34,
2730–2743 共2007兲.
but at any arbitrary angle relative to the pixel matrix. This is 5
J. W. Coltman, “The specification of imaging properties by response to a
contrary to standard methods, which are inherently one di- sine wave input,” J. Opt. Soc. Am. 44, 468–469 共1954兲.
6
mensional and rely on slight angulations relative to the pixel P. F. Judy, “The line spread function and modulation transfer function of
a computer tomographic scanner,” Med. Phys. 3, 233–236 共1976兲.
rows or columns to provide a “finely sampled” response, 7
N. J. Schneiders and S. C. Bushong, “Single-step calculation of the MTF
hence limiting them to the orthogonal or near-orthogonal from the ERF,” Med. Phys. 5, 31–33 共1978兲.
directions.11,21,23 Accurate characterization of the two- 8
R. A. Sones and G. T. Barnes, “A method to measure the MTF of digital
dimensional MTF could prove useful for improved system x-ray systems,” Med. Phys. 11, 166–171 共1984兲.
9
R. T. Droege and M. S. Rzeszotarski, “An MTF method immune to alias-
evaluation and observer performance models.14 ing,” Med. Phys. 12, 721–725 共1985兲.
10
I. A. Cunningham and A. Fenster, “A method for modulation transfer
function determination from edge profiles with correction for finite-
element differentiation,” Med. Phys. 14, 533–537 共1987兲.
11
H. Fujita, D. Y. Tsai, T. Itoh, K. Doi, J. Morishita, K. Ueda, and A.
Ohtsuka, “A simple method for determining the modulation transfer func-
IV. CONCLUSIONS tion in digital radiography,” IEEE Trans. Med. Imaging 11, 34–39 共1992兲.
12
E. Samei, M. J. Flynn, and D. A. Reimann, “A method for measuring the
A new method for determination of the presampled MTF, presampled MTF of digital radiographic systems using an edge test de-
the “noise-response method,” has been described and evalu- vice,” Med. Phys. 25, 102–113 共1998兲.
13
C. D. Bradford, W. W. Peppler, and J. M. Waidelich, “Use of a slit camera
ated. The accuracy of this method was demonstrated using for MTF measurements,” Med. Phys. 26, 2286–2294 共1999兲.
both simulated and experimental data sets. For the simulated 14
K. A. Fetterly, N. J. Hangiandreou, B. A. Schueler, and E. R. Ritenour,
image set which used a simple detector model for which the “Measurement of the presampled two-dimensional modulation transfer
function of digital imaging systems,” Med. Phys. 29, 913–921 共2002兲.
true MTF was known exactly, excellent agreement was ob- 15
K. Doi, K. Strubler, and K. Rossmann, “Truncation errors in calculating
tained with the MTF determined using the noise-response the MTF of radiographic screen-film systems from the line spread func-
method, with a maximum deviation of 1.1%. Comparison tion,” Phys. Med. Biol. 17, 241–250 共1972兲.
16
measurements were also made on this simulated data set with F. F. Yin, M. L. Giger, and K. Doi, “Measurement of the presampling
modulation transfer function of film digitizers using a curve fitting tech-
the established edge-response method, and these showed de- nique,” Med. Phys. 17, 962–966 共1990兲.
viations greater than 35% from the true MTF. Experimental 17
S. E. Reichenbach, S. K. Park, and R. Narayanswamy, “Characterizing
measurements on a range of detector technologies 共including digital image acquisition devices,” Opt. Eng. 30, 170–177 共1991兲.
18 40
J. M. Boone and J. A. Seibert, “An analytical edge spread function model A. Kuhls-Gilcrist, G. Yadava, V. Patel, A. Jain, D. R. Bednarek, and S.
for computer fitting and subsequent calculation of the LSF and MTF,” Rudin, “The solid state x-ray image intensifier 共SSXII兲: An EMCCD-
Med. Phys. 21, 1541–1545 共1994兲. based x-ray detector,” Proc. SPIE 6913, 69130K-1–69130K-10 共2008兲.
19 41
A. Stenman and F. Gustafsson, “Adaptive smoothing methods for K. Rossmann, “Modulation transfer function of radiographic systems us-
frequency-function estimation,” Automatica 37, 675–685 共2001兲. ing fluorescent screens,” J. Opt. Soc. Am. 52, 774–777 共1962兲.
20 42
A. D. Maidment and M. Albert, “Conditioning data for calculation of the G. Lubberts, “Random noise produced by x-ray fluorescent screens,” J.
modulation transfer function,” Med. Phys. 30, 248–253 共2003兲. Opt. Soc. Am. 58, 1475–1482 共1968兲.
21 43
International Electrotechnical Commission, “Medical electrical equip- S. Rudin, A. T. Kuhls, G. K. Yadava, G. C. Josan, Y. Wu, R. N. Chityala,
ment: Characteristics of digital x-ray imaging devices–Part 1: Determina- H. S. Rangwala, C. N. Ionita, K. R. Hoffmann, and D. R. Bednarek, “New
tion of the detective quantum efficiency,” IEC Report No. 62220-1 共In- light-amplifier-based detector designs for high spatial resolution and high
ternational Electrotechnical Commission, Geneva, Switzerland, 2003兲. sensitivity CBCT mammography,” Proc. SPIE 6142, 614263 共2006兲.
22 44
P. B. Greer and T. van Doorn, “Evaluation of an algorithm for the assess- A. Kuhls-Gilcrist, G. Yadava, V. Patel, D. R. Bednarek, and S. Rudin,
ment of the MTF using an edge method,” Med. Phys. 27, 2048–2059 “Progress in electron-multiplying CCD 共EMCCD兲 based, high-resolution,
共2000兲. high-sensitivity x-ray detector for fluoroscopy and radiography,” Proc.
23
E. Buhr, S. Gunther-Kohfahl, and U. Neitzel, “Accuracy of a simple SPIE 6510, 651047 共2007兲.
45
method for deriving the presampled modulation transfer function of a A. Kuhls-Gilcrist, D. R. Bednarek, and S. Rudin, “Component analysis of
digital radiographic system from an edge image,” Med. Phys. 30, 2323– a new solid state x-ray image intensifier 共SSXII兲 using photon transfer
2331 共2003兲. and instrumentation noise equivalent exposure 共INEE兲 measurements,”
24
A. K. Carton, D. Vandenbroucke, L. Struye, A. D. Maidment, Y. H. Kao, Proc. SPIE 7258, 725842 共2009兲.
46
M. Albert, H. Bosmans, and G. Marchal, “Validation of MTF measure- J. Yao and I. A. Cunningham, “Parallel cascades: New ways to describe
ment for digital mammography quality control,” Med. Phys. 32, 1684– noise transfer in medical imaging systems,” Med. Phys. 28, 2020–2038
1695 共2005兲. 共2001兲.
25 47
U. Neitzel, E. Buhr, G. Hilgers, and P. R. Granfors, “Determination of the W. Zhao, W. G. Ji, and J. A. Rowlands, “Effects of characteristic x rays
modulation transfer function using the edge method: Influence of scat- on the noise power spectra and detective quantum efficiency of photocon-
tered radiation,” Med. Phys. 31, 3485–3491 共2004兲. ductive x-ray detectors,” Med. Phys. 28, 2039–2049 共2001兲.
26 48
M. C. Steckner, D. J. Drost, and F. S. Prato, “A cosine modulation artifact M. Rabbani, R. Shaw, and R. V. Metter, “Detective quantum efficiency of
in modulation transfer function computations caused by the misregistra- imaging systems with amplifying and scattering mechanisms,” J. Opt.
tion of line spread profiles,” Med. Phys. 20, 469–473 共1993兲. Soc. Am. 4, 895–901 共1987兲.
27 49
S. N. Friedman and I. A. Cunningham, “Normalization of the modulation G. Hajdok, J. Yao, J. J. Battista, and I. A. Cunningham, “Signal and noise
transfer function: The open-field approach,” Med. Phys. 35, 4443–4449 transfer properties of photoelectric interactions in diagnostic x-ray imag-
共2008兲. ing detectors,” Med. Phys. 33, 3601–3620 共2006兲.
28 50
I. A. Cunningham and B. K. Reid, “Signal and noise in modulation trans- M. L. Giger, K. Doi, and C. E. Metz, “Investigation of basic imaging
fer function determinations using the slit, wire, and edge techniques,” properties in digital radiography. 2. Noise Wiener spectrum,” Med. Phys.
Med. Phys. 19, 1037–1044 共1992兲. 11, 797–805 共1984兲.
29 51
J. Morishita, K. Doi, R. Bollen, P. C. Bunch, D. Hoeschen, G. Sirand-rey, J. R. Janesick, Scientific Charge-Coupled Devices 共SPIE, Bellingham,
and Y. Sukenobu, “Comparison of two methods for accurate measurement 2001兲.
52
of modulation transfer functions of screen-film systems,” Med. Phys. 22, W. Zhao and J. A. Rowlands, “Digital radiology using active matrix read-
193–200 共1995兲. out of amorphous selenium: Theoretical analysis of detective quantum
30
F. Rogge, H. Bosmans, and G. Marchal, “Practical MTF calculation in efficiency,” Med. Phys. 24, 1819–1833 共1997兲.
53
digital mammography: A multicenter study,” Proc. SPIE 5368, 761–769 A. Mackenzie and I. D. Honey, “Characterization of noise sources for two
共2004兲. generations of computed radiography systems using powder and crystal-
31
E. Samei, N. T. Ranger, J. T. Dobbins III, and Y. Chen, “Intercomparison line photostimulable phosphors,” Med. Phys. 34, 3345–3357 共2007兲.
54
of methods for image quality characterization. I. Modulation transfer W. Zhao, G. Ristic, and J. A. Rowlands, “X-ray imaging performance of
function,” Med. Phys. 33, 1454–1465 共2006兲. structured cesium iodide scintillators,” Med. Phys. 31, 2594–2605 共2004兲.
32 55
U. Neitzel, S. Gunther-Kohfahl, G. Borasi, and E. Samei, “Determination J. H. Siewerdsen, L. E. Antonuk, Y. el-Mohri, J. Yorkston, W. Huang, and
of the detective quantum efficiency of a digital x-ray detector: Compari- I. A. Cunningham, “Signal, noise power spectrum, and detective quantum
son of three evaluations using a common image data set,” Med. Phys. 31, efficiency of indirect-detection flat-panel imagers for diagnostic radiol-
2205–2211 共2004兲. ogy,” Med. Phys. 25, 614–628 共1998兲.
33 56
I. A. Cunningham, M. S. Westmore, and A. Fenster, “A spatial-frequency International Electrotechnical Commission, “Medical diagnostic x-ray
dependent quantum accounting diagram and detective quantum efficiency equipment: Radiation conditions for use in the determination of charac-
model of signal and noise propagation in cascaded imaging systems,” teristics,” IEC Report No. 61267 共International Electrotechnical Commis-
Med. Phys. 21, 417–427 共1994兲. sion, Geneva, Switzerland, 1997兲.
34 57
J. A. Rowlands and K. W. Taylor, “Absorption and noise in cesium iodide K. A. Fetterly and N. J. Hangiandreou, “Effects of x-ray spectra on the
x-ray image intensifiers,” Med. Phys. 10, 786–795 共1983兲. DQE of a computed radiography system,” Med. Phys. 28, 241–249
35
W. Zhao, W. G. Ji, A. Debrie, and J. A. Rowlands, “Imaging performance 共2001兲.
58
of amorphous selenium based flat-panel detectors for digital mammogra- N. T. Ranger, E. Samei, J. T. Dobbins III, and C. E. Ravin, “Measurement
phy: Characterization of a small area prototype detector,” Med. Phys. 30, of the detective quantum efficiency in digital detectors consistent with the
254–263 共2003兲. IEC 62220-1 standard: Practical considerations regarding the choice of
36
J. H. Siewerdsen, L. E. Antonuk, Y. el-Mohri, J. Yorkston, W. Huang, J. filter material,” Med. Phys. 32, 2305–2311 共2005兲.
59
M. Boudry, and I. A. Cunningham, “Empirical and theoretical investiga- J. T. Dobbins, “Image quality metrics for digital systems,” in Handbook
tion of the noise performance of indirect detection, active matrix flat- of Medical Imaging: Physics and Psychophysics, edited by J. Beutel, H.
panel imagers 共AMFPIs兲 for diagnostic radiology,” Med. Phys. 24, 71–89 L. Kundel, and R. L. V. Metter 共SPIE, Bellingham, 2000兲, Vol. 1, pp.
共1997兲. 161–222.
37 60
M. B. Williams, P. U. Simoni, L. Smilowitz, M. Stanton, W. Phillips, and I. S. Kyprianou, A. Badano, B. D. Gallas, and K. J. Myers, “Singular
A. Stewart, “Analysis of the detective quantum efficiency of a develop- value description of a digital radiographic detector: Theory and measure-
mental detector for digital mammography,” Med. Phys. 26, 2273–2285 ments,” Med. Phys. 35, 4744–4756 共2008兲.
共1999兲. 61
A. E. Burgess, “An empirical equation for screen MTFs,” Med. Phys. 5,
38
A. Ganguly, S. Rudin, D. R. Bednarek, and K. R. Hoffmann, “Micro- 199–204 共1978兲.
62
angiography for neuro-vascular imaging. II. Cascade model analysis,” G. Hajdok, J. J. Battista, and I. A. Cunningham, “Fundamental x-ray
Med. Phys. 30, 3029–3039 共2003兲. interaction limits in diagnostic imaging detectors: Spatial resolution,”
39
S. Vedantham, A. Karellas, and S. Suryanarayanan, “Solid-state fluoro- Med. Phys. 35, 3180–3193 共2008兲.
63
scopic imager for high-resolution angiography: Parallel-cascaded linear G. Hajdok, J. J. Battista, and I. A. Cunningham, “Fundamental x-ray
systems analysis,” Med. Phys. 31, 1258–1268 共2004兲. interaction limits in diagnostic imaging detectors: Frequency-dependent
Swank noise,” Med. Phys. 35, 3194–3204 共2008兲. of methods for image quality characterization. II. Noise power spectrum,”
64
M. B. Williams, P. A. Mangiafico, and P. U. Simoni, “Noise power spec- Med. Phys. 33, 1466–1475 共2006兲.
66
tra of images from digital mammography detectors,” Med. Phys. 26, S. N. Friedman and I. A. Cunningham, “A moving slanted-edge method
1279–1293 共1999兲. to measure the temporal modulation transfer function of fluoroscopic sys-
65
J. T. Dobbins III, E. Samei, N. T. Ranger, and Y. Chen, “Intercomparison tems,” Med. Phys. 35, 2473–2484 共2008兲.