C DJ-002
El presente documento tiene vigencia por 6 meses o en caso haya alguna actualización de los
lineamientos de Parque Arauco y/o cambio del representante legal de la contratista.
TELMAAK EXFO Platinum
Telmark Peru SAC
Plaza 27 de
Noviembre 450 1 of 3
San Isidro
Lima - Peru
CALIBRATION CERTIFICATE
Customer: LEVEL 7 S.A.C.
As found
New unit (initial Within Within Outside Outside Defective
calibration) specifications (i) specifications* () specifications* (i) specifications (v) operation
*See results page for details on "As found" status
Actlon taken
No adjustment was made Adjustments were made Repair was performed
As left
Within specifications () Within specifications (i)Outside specifications* (i) Outside specifications(iv)
Calibration conditions
Fiber connector FCIAPC
Fiber type 9/125 um, N.A. = 0.13 (except 850 nm/1300 nm: 62.5/125 um, N.A. = 0.27).
Relative humidity 50 % ±10 %
Temperature 23°C ±1°C
Standards used to establish traceability
Description Serial number Calibration date Calibration validity
GP-121392: 2.3 km Singlemode Distance Reference Std LRF2016.06 2016-10-27 Indefinite
Optical ports are always cleaned before calibration
EXFO certifies that the unit has been calibrated using standards traceable to a national metrology
physical constants or using ratio measurements. NIST is the National Institute of Standards and institute(NIST, NPL, NRC, METAS or other), natura
Laboratory in the UK, NRC is the National Research Council in Canada and METAS is the SwissTechnology
in the
Federal Office for USA, NPL 0s the National Physical
Metrology.
All uncertainties are reported with a level of confidence of95%. Calibration is based on the ISOEC 17025 standard.
The certificate shall not be reproduced, except in full, without the written approOval of
EXFO.
OTDR L Test
Port Wavelength Non Linearity Test limit Verification
(nm) (dB) (dB)
As found
NIA NIA NIA NJA NIA
As left
1 1310 0.006 0.05 Pass
Test Parameters at 1310 nm: pulse 10000 ns; averaging time 45 s tested at one
wavelength only. Linearity test range 13.7 dB.
OTDR D0stance Calibration
Port Wavelength Measured Reference Deviation Uncertainty Specification Conformance limit(m)
distance distance Verification
(nm) (m) (m) (m) (m) (m) 1 2 3
As found
NIA NIA NIA NIA N/A NIA N/A NIA N/A NIA NJA
NIA NIA NIA NIA NIA N/A NIA NIÂ NIA NIA NIA
As lef
1310 2339.482 2339,350 0.132 0.058 0.848 0.801 0.848 0.896
1550 2340.444 2340.291 0.154 0.275 0.849 0.622 0.849 1.075
Test Parameters at 1310nm: pulse 30 ns; averaging 45traces; range 2500 m; high
Test Parameters at 1550 nm: pulse 30 ns; averaging 45 traces; range 2500 m; resolution mode.
high resolution mode.
Optical Time Domain Reflectometer
ELMARK Model: MAX-715B-M1-El
3 of 3
EXFO |o Serial Number: 1011420
Calibration date: 2023-09-28
Remarks:
Optical Ports are always clean before calibration; PASS
This instrument calibrated for this report has been calibrated in accordance with calibration procedure;
This calibration meets specification as outlined in ISO9001.