Enclosure - Drystar 5302 Axys - Pickup Module
Enclosure - Drystar 5302 Axys - Pickup Module
Imaging Services
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1) Read the "Generic Safety Directions" document
(see MEDNET GSO => General Info => Agfa HealthCare => Publications =>
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and
on the product.
Required Spare Part, Tools & Time Note: Spare Pickup Module can be used
as Upper or Lower Pickup Module
Upper Lower
Pickup Module Pickup Module
Additional
roller pair
Spare part = upper pickup module
Upper pickup module can be used
as lower pickup module, too.
60 min.
Do not remove additional roller pair
5366QRM15pm 5366QRM157
1 Remove Transport Protection Locks 2 Pull out upper or lower tray and lift the
tray to take it out
1 2
3 2
1
5366QRM12 5366QRM02
3 Pull out pickup module 4 Insert new Pickup Module, insert and
close tray
1 1
5366QRM12 5366QRM14
5366QRM07
5366QRM08