HEALTHCARE Enclosure

Imaging Services

Document No: DD+DIS139.10E


This number is needed for production line only:
F7.5175.9004.1 CR 30-X
Type 5175 /100/110

Input Rollers Replacement Instructions

► Purpose of this Document

This enclosure is part of the spare part "Input Rollers",


spare part number*: CM+9 5175 9030 4

It describes all necessary steps for the replacement.

* The last digit in the spare part number indicates the spare part revision at release of
this document. When ordering, the actual revision of the spare part is delivered.

► Referenced Documents

Document Title
Not applicable

NOTE:
Dispose the enclosure document after finishing your task
(observe the confidentiality note below).

CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0
04-2010 printed in Germany Document Node ID: 29939235
Enclosure_e_template_v07
Agfa Company Confidential Copyright © 2010 Agfa HealthCare N.V.
DD+DIS139.10E Enclosure

► Manufacturer
Agfa HealthCare N.V.

Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
Copyright  2010 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.

WARNING:
Improper operation or service activities may cause damage or injuries.

INSTRUCTION:
(1) Read the "Generic Safety Directions" document
(see MEDNET GSO => General Info => Agfa HealthCare => Publications =>
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X Page 2 of 12


04-2010 Type 5175 /100/110 Agfa Company Confidential
DD+DIS139.10E Enclosure

LIST OF CONTENTS

1 REPLACEMENT OF THE ROLLERS.......................................................................................5


2 STALL CALIBRATION............................................................................................................12
3 VERIFICATION.......................................................................................................................12

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X Page 3 of 12


04-2010 Type 5175 /100/110 Agfa Company Confidential
DD+DIS139.10E Enclosure

Subject Input Rollers Replacement Instructions

IMPORTANT:
This spare part kit contains all 4 input 3
rollers.
1
It is required to replace always all 4
rollers.
The two upper rollers 1 and 2 are
identical. 4
The segmented roller always has to be 2
mounted at position 4 .

Figure 1

SCOPE OF DELIVERY:

Roller 1 HEALTH CARE Enclosure

Document No: DD+DIS139.10E


This number is needed for production line only:
F7.5175.9004.1 CR 30-X
Type 5175 /100/110

Roller 2 Input Rollers Replacement Instructions

Roller 3 ? Purpose of this Document

This enclosure is part of the spare part "Input Rollers",


spare part number*: CM+9 5175 9030 4

It describes all necessary steps for the replacement.

Segmented roller 4 * The last digit in the spare part number indicates the spare part revision at release of
this document. When ordering, the actual revision of the spare part is delivered.

? Referenced Documents

Document Title
Not applicable Not applicable

NOTE:
Dispose the enclosure document after finishing your task

Cable ties (observe the confidentiality note below).

CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0
04-2010 printed in Germany Document Node ID: 29939235
Enclosure_e_template_v07
Agfa Company Confidential Copyright © 2010 Agfa HealthCare N.V.

REQUIRED TOOLS:

Medium size 3 mm,


with ball-shaped head

REQUIRED TIME:
Approximately 1 hour

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X Page 4 of 12


04-2010 Type 5175 /100/110 Agfa Company Confidential
DD+DIS139.10E Enclosure

1 Replacement of the Rollers

(1) Switch off the digitizer.


(2) Disconnect the power cable.
(3) Simultaneously press the
two buttons (A) positioned
underneath the feed table.
(4) Pull out cassette unit (B).
B
A
A

Figure 2

(5) Remove 3 upper screws at front


cover.

Figure 3

(6) Remove 3 lower screws at front


cover.

Figure 4

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X Page 5 of 12


04-2010 Type 5175 /100/110 Agfa Company Confidential
DD+DIS139.10E Enclosure

(7) Press two handles and pull out front


cover approx. 5 cm
(8) Disconnect cable at user interface
board. See Figure 5.
(9) Remove front cover

Figure 5

(10) Disconnect ground cable at left


hand side.

Figure 6

(11) Disconnect ground cable at right


hand side.
(12) Remove front cover.

Figure 7

(13) Take out cleaning brush holder.

Figure 8

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X Page 6 of 12


04-2010 Type 5175 /100/110 Agfa Company Confidential
DD+DIS139.10E Enclosure

(14) Disconnect two cables at erasure


unit.
(15) Remove screw at erasure unit.
(16) Take out erasure unit.

Figure 9

CAUTION:
IP jams possible if IP guide plate is deformed.
Do not bend IP guide plate at removal.

(17) Slide IP guide plate to the left and


lift it out.

Figure 10

(18) Disconnect cable at roller IP


sensor.

Figure 11

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X Page 7 of 12


04-2010 Type 5175 /100/110 Agfa Company Confidential
DD+DIS139.10E Enclosure

(19) Open three cable holders at right


hand side.

Figure 12

(20) Open four cable holders at left


hand side.

Figure 13

(21) Cut two cable ties at upper part


transport unit left hand side.

Figure 14

(22) Cut one cable tie at upper part


transport unit right hand side.

Figure 15

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X Page 8 of 12


04-2010 Type 5175 /100/110 Agfa Company Confidential
DD+DIS139.10E Enclosure

(23) Unplug P4, P17, P25 and P29 at IP


handling board.

Figure 16

(24) Remove 2 screws of upper part


transport unit right hand side.

Figure 17

(25) Remove 2 screws of upper part


transport unit left hand side.

Figure 18

(26) Take off the upper part of transport


rollers.

Figure 19

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X Page 9 of 12


04-2010 Type 5175 /100/110 Agfa Company Confidential
DD+DIS139.10E Enclosure

(27) With a screwdriver remove two ball


bearings at transport unit left hand
side.

Figure 20

(28) Remove two ball bearings at


transport unit right hand side.

Figure 21

(29) Take off the complete lower


transport unit.

Figure 22

(30) Remove two c-clips of upper


transport rollers left and right
(Figure 23 only shows two c-clips
at left hand side).

Figure 23

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X Page 10 of 12


04-2010 Type 5175 /100/110 Agfa Company Confidential
DD+DIS139.10E Enclosure

(31) Remove gear wheel at roller 2 if 2


present . See arrow.
It is not used anymore.
Figure 24

NOTE:
Keep gear wheel with c-clips and pin as spare parts in case a part gets
lost during replacement of a roller.

(32) Mount the two new rollers 1 and 1


2 at the upper side plates.

Figure 25

(33) Remove the c-clips at the lower


rollers right hand side.

Figure 26

(34) Remove the c-clip at the gear of


the segmented roller.
(35) Remove the gear: Take care not
to loose the pin. See arrow in
Figure 27.
(36) Remove the lower rollers.

NOTE:
At the segmented roller
4 a 2nd, inner gear
wheel (not shown in
Figure 27) may be
mounted. This is not
used anymore. Figure 27

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X Page 11 of 12


04-2010 Type 5175 /100/110 Agfa Company Confidential
DD+DIS139.10E Enclosure

ASSEMBLY:
(1) Mount in reverse order.

2 Stall Calibration

Perform a stall calibration for the Roller Clamping Stepper Motor (Menu "Calibrate
Roller Clamper ") as described in the Service Manual CR 30-X (DD+DIS150.06E),
Chapter 3.6, Adjustments and Calibrations.

3 Verification

Verification: (1) Switch off/on the digitizer.


(2) Wait till the status indicator at the digitizer switches to green
(takes approximately 1,5 minutes).
(3) Insert cassette with unexposed IP in the digitizer.
(4) At the processing station select examination type “System diagnosis – Flatfield”.
(5) Select “ID” to identify the image: Scanning starts.
(6) Confirm that the image arrives at the NX processing station.
(7) Reject the image.
(8) Repeat these steps with all available cassette formats.

Result: The Input Rollers are replaced.