* The last digit in the spare part number indicates the spare part revision at release of
this document. When ordering, the actual revision of the spare part is delivered.
► Referenced Documents
Document Title
Not applicable
NOTE:
Dispose the enclosure document after finishing your task
(observe the confidentiality note below).
CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Revision 0
04-2010 printed in Germany Document Node ID: 29939235
Enclosure_e_template_v07
Agfa Company Confidential Copyright © 2010 Agfa HealthCare N.V.
DD+DIS139.10E Enclosure
► Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
Copyright 2010 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1) Read the "Generic Safety Directions" document
(see MEDNET GSO => General Info => Agfa HealthCare => Publications =>
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
LIST OF CONTENTS
IMPORTANT:
This spare part kit contains all 4 input 3
rollers.
1
It is required to replace always all 4
rollers.
The two upper rollers 1 and 2 are
identical. 4
The segmented roller always has to be 2
mounted at position 4 .
Figure 1
SCOPE OF DELIVERY:
Segmented roller 4 * The last digit in the spare part number indicates the spare part revision at release of
this document. When ordering, the actual revision of the spare part is delivered.
? Referenced Documents
Document Title
Not applicable Not applicable
NOTE:
Dispose the enclosure document after finishing your task
CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Revision 0
04-2010 printed in Germany Document Node ID: 29939235
Enclosure_e_template_v07
Agfa Company Confidential Copyright © 2010 Agfa HealthCare N.V.
REQUIRED TOOLS:
REQUIRED TIME:
Approximately 1 hour
Figure 2
Figure 3
Figure 4
Figure 5
Figure 6
Figure 7
Figure 8
Figure 9
CAUTION:
IP jams possible if IP guide plate is deformed.
Do not bend IP guide plate at removal.
Figure 10
Figure 11
Figure 12
Figure 13
Figure 14
Figure 15
Figure 16
Figure 17
Figure 18
Figure 19
Figure 20
Figure 21
Figure 22
Figure 23
NOTE:
Keep gear wheel with c-clips and pin as spare parts in case a part gets
lost during replacement of a roller.
Figure 25
Figure 26
NOTE:
At the segmented roller
4 a 2nd, inner gear
wheel (not shown in
Figure 27) may be
mounted. This is not
used anymore. Figure 27
ASSEMBLY:
(1) Mount in reverse order.
2 Stall Calibration
Perform a stall calibration for the Roller Clamping Stepper Motor (Menu "Calibrate
Roller Clamper ") as described in the Service Manual CR 30-X (DD+DIS150.06E),
Chapter 3.6, Adjustments and Calibrations.
3 Verification