Download as pdf or txt
Download as pdf or txt
You are on page 1of 32

HEALTHCARE Chapter 6

Imaging Services Repair


Document No: DD+DIS101.11E

DX-D 600

► Purpose of this Document

This document contains all the information required for repair.

► Document History

Edition. Release Changes


Revision Date compared to previous version 1.0
1.1 02-2012  Updated information about DR Detector in section 2.
 Added information about Wireless DR Detector in
section 3.
 Added information about DX-D FLFS Stand
in section 7.

► Referenced Documents

Document Title
For the lists of referenced documents, refer to following pages.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1
02-2012 Printed in Germany Document Node ID: 33248749
so_standard_e_template_v08
Agfa Company Confidential Copyright © 2012 Agfa HealthCare N.V.
DD+DIS101.11E Repair

► Manufacturer
Agfa HealthCare N.V.

Published by
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
Copyright  2012 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
Agfa and the Agfa rhombus are trademarks of Agfa-Gevaert N.V., Belgium,
or its affiliates.

WARNING:
Improper operation or service activities may cause damage or injuries.

INSTRUCTION:
(1) Read the "Generic Safety Directions" document
(see Agfa HealthCare Library > General Info > Agfa HealthCare > Publications >
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.

IMPORTANT:
The installation and service of the product(s) described herein is to be performed
by qualified personnel who are employed by Agfa HealthCare or one of its affiliates
or who are otherwise authorized by Agfa HealthCare or one of its affiliates to provide
such services.

NOTE:
To verify the latest version of a chapter of the Service Manual refer to the “Checklist for
Completeness” in the Agfa HealthCare Library.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 2 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

LIST OF CONTENTS

1 X-RAY SYSTEM .......................................................................................................................4


1.1 Ceiling Support..........................................................................................................................4

1.2 Wall Stand .................................................................................................................................4

1.3 X-ray Generator ........................................................................................................................4

1.4 RAD Table.................................................................................................................................4

2 FIXED AND PORTABLE DR DETECTOR ...............................................................................5


2.1 Required Information in Case of Image Quality Problems and Exchange of Fixed or Portable
Detector.....................................................................................................................................5

2.2 Collecting the relevant Files ......................................................................................................7

2.3 Fixed or Portable DR Detector Replacement Instructions ......................................................10

2.4 Fixed or Portable DR Detector Return Procedure ..................................................................11

2.5 Acceptance Test after Fixed or Portable DR Detector Replacement .....................................11

3 WIRELESS DR DETECTOR ..................................................................................................12


3.1 Required Information in Case of Image Quality Problems and Exchange of the Wireless DR
Detector...................................................................................................................................12

3.2 Wireless DR Detector Replacement Instructions....................................................................14

3.3 Disassembly and Re-assembly of the Wireless DR Detector System....................................14

3.3.1 X-ray Interface Box .................................................................................................................15

3.3.2 PCA-FE Board Firmware and FPGA Updating and Recovery ...............................................28

3.3.3 PCA-XIF Board Firmware Updating and Recovery ................................................................16

3.3.4 PCA-XIF Board Firmware Recovery .......................................................................................21

3.3.5 Other Reference Information ..................................................................................................25

3.3.6 Work Accompanying Replacement of Parts ...........................................................................30

3.4 Wireless DR Detector Return Procedure ................................................................................30

3.5 Acceptance Test after Wireless DR Detector Replacement ...................................................31

4 XRDI .......................................................................................................................................32
5 NX WORKSTATION ...............................................................................................................32
6 AGFA DIGITIZER (OPTIONAL)..............................................................................................32
7 DX-D FLFS STAND (OPTIONAL) ..........................................................................................32

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 3 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

1 X-ray System

Refer to relevant service documentation from the DX-D 600 partner company Sedecal
for repair descriptions.

1.1 Ceiling Support

Ceiling Support - Service Documentation Links for Offline Navigation


NOVA Automatic Ceiling Suspension
Click here
Service Manual; section 11
Gas Springs Replacement Procedure Click here

1.2 Wall Stand

Wall Stand - Service Documentation Links for Offline Navigation


NBS2100 Vertical Bucky Stand Service Manual;
Click here
section 8 and section 9

1.3 X-ray Generator

X-ray Generator - Service Documentation Links for Offline Navigation


Renewal Parts Click here
Disassembly / Reassembly Procedures Click here
Calibration Click here
Configuration Click here

1.4 RAD Table

Radiographic Table - Service Documentation Links for Offline Navigation


NET X-Ray Table Service Manual; section 8 and
Click here
section 9

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 4 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

2 Fixed and Portable DR Detector

NOTE:
For the DX-D 600 are following separate spare parts list available:
 DD+DIS273.10E, Spare parts list for Portable DR Detector
 DD+DIS171.09E, Spare parts list for Fixed DR Detector

Fixed DR Detector – Spare Parts List Links for Offline Navigation


Spare Parts List Click here

Portable DR Detector – Spare Parts List Links for Offline Navigation


Spare Parts List Click here

NOTE:
The least replaceable unit (LRU) is the Receptor Assembly.
For more details about the DR Detector return procedure, refer to section 0.

2.1 Required Information in Case of Image Quality Problems and


Exchange of Fixed or Portable Detector

NOTE:
When dealing with image quality or bad pixel mapping problems of the flat panel the
FSE must get in touch with the RSN/GSC.
Only when the information listed in this section will be made available to the RSN/GSC,
the advice and counsel elaborated at the RSN/GSC will really help in troubleshooting.

Information needed in following cases:


 Perceived problems with the flat panel strongly indicate a possible necessity
for replacement / exchange due to image quality problems.
 Troubleshooting on a DR Detector when failing the bad pixel mapping DMAT tool

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 5 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

The following data are required in the GSC case:


 Detailed problem description
 Actions performed
 Serial number of the DR Detector
 Installation date of the DR Detector
 Exchange date of the DR Detector
 Last calibration date of the DR Detector

Relevant files to be posted on an FTP server in Mortsel:


Following document describes how to access and upload files on the AGFA FTP
server. It is available from the Agfa HealthCare Library:
General Info > Agfa HealthCare > Publications > Service
Bulletin> ”General info - Service Bulletin - Temporary AGFA FTP Account”.

Temporary AGFA FTP Account Links for Offline Navigation


Service Bulletin Click here

(1) Create a folder In the directory Varian with Customer name and GSC Case #
(2) Upload a zip file with the following files*:
 For fixed DR Detectors:
o defect_map002.bin
o gain_img002.viv
o ofst_img002.viv

 For portable DR Detectors:


o defect_map008.bin
o gain_img008.viv
o ofst_img008.viv

 mode_0_defect_analysis.txt
 Flatfield exposure with the ViVA tool / save function
 Fault images
* Refer to next section for detailed description about file locations.

(3) Inform the RSN/GSC when the upload is finished.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 6 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

2.2 Collecting the relevant Files

(1) Go to C:\IMAGERs\<detector serialnumber>\00_4pfG2550msSCTMSR4


and copy the following files to an USB Memory Stick:
 For fixed DR Detectors:
o defect_map002.bin
o gain_img002.viv
o ofst_img002.viv
o mode_0_defect_analysis.txt

 For portable DR Detectors:


o defect_map008.bin
o gain_img008.viv
o ofst_img008.viv
o mode_0_defect_analysis.txt

(2) Perform step (1) also for C:\IMAGERs\<detector serialnumber>\


00_4pfG21000msSCTMSR4

NOTE:
Make sure that the permissions of the Folder C:\IMAGERs are set correctly before
copy the files to an USB Memory Stick, see Chapter 2 - Installation and Set-up.

(3) Stop the NX Workstation. Go to: <Start  Agfa  DR Calibration


Tools  Stop NX>
(4) Open the ViVA tool.

(5) Click <Receptor Setup…> from


the menu bar under Acquisition.

Figure 1

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 7 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

(6) Select the specific receptor from the


Receptor Serial # drop down.
(7) Click: <OK>

Figure 2

(8) Click <Acquisition  Open


Link> to the faulty panel.

Figure 3

(9) Select: <Acquire Image tab>

Figure 4

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 8 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

(10) Wait until the VIVA is ready to


capture exposure.

Figure 5

(11) Select on the X-ray generator console the dedicated acquisition system.
(12) Change exposure settings manually to the settings used during calibration
and expose.

(13) When the image is displayed


in the ViVA application, in the
pop up window label the image
with the exposure settings used
during calibration (kVp, mA, SID).

Figure 6

(14) Save the image.

Figure 7

(15) Copy the image file to the USB stick (checked to be virus free).

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 9 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

NOTE only applicable for systems with multiple Detectors:


 If the communication to the Detector with the VIVA application fails,
repeat step (4) to (7).
 If still not able to connect switch off the second Detector or remove
the Ethernet CAT 6 cable at NX side.

NOTE:
When you encounter only intermittent image quality problems export those faulty
images on NX in DICOM format.
Post these images also on the AGFA FTP Server in the Varian directory.
Anonymize images for patient data protection.
How to access this temporary AGFA FTP Server is described in following Service
Bulletin:

Temporary AGFA FTP Account Links for Offline Navigation


Service Bulletin Click here

2.3 Fixed or Portable DR Detector Replacement Instructions

IMPORTANT:
In case of a DR Detector replacement update the serial numbers of the
Fixed or Portable DR Detectors:

After replacement of the DR Detectors, the serial numbers of the DR Detectors are
not accessible without re-opening the Bucky’s.

(1) Replace the DR Detectors.


(2) Update the DX-D 600 Serial Number tracking document and insert the new serial
numbers and installation date of the replaced DR Detectors (DR Wall Stand and
RAD Table).

Refer to the DX-D 600 system Serial Number Tracking document.

DX-D 600 - Service Documentation Links for Offline Navigation


DX-D 600 System Serial Number Tracking Click here

(3) Send the document via E-mail to: dr_start_up_team@agfa.com

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 10 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

2.4 Fixed or Portable DR Detector Return Procedure

This section defines the roles and responsibilities between VARIAN and AGFA when
it comes to necessary exchanges of the VARIAN flat panel detector due to failure.

The customer generally contacts the local AGFA NSO, when a problem with the AGFA
DR-System is encountered.

The local AGFA NSO sends a Field Service Engineer (FSE) to analyze and possibly fix
the problem. In case of a DR detector failure the Agfa FSE always needs to escalate to
Agfa Level 2 (Tier 1) support for confirmation (GSC case).

For failure assessment the data listed in section 10 are required in the GSC case.
If failure is confirmed then Level 2/Level 3 support will supply the FSE with A Return
Material Authorization (RMA), further instructions to order a new DR Detector and to
return the defect one.

IMPORTANT:
An immediate return of the DR Detector to the AGFA spare parts logistic center is
necessary, to make optimal use of the warranty period.
For fast processing of the DR Detector exchange it is essential to make sure that
all data mentioned above is available at time of escalation.

2.5 Acceptance Test after Fixed or Portable DR Detector Replacement

NOTE:
In case of a DR Detector replacement refer to Chapter 11 - Acceptance Test after
DR Detector Replacement

Fixed DR Detector Links for Offline Navigation


Acceptance Test after DR Detector Replacement Click here

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 11 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

3 Wireless DR Detector

NOTE:
For the Wireless DR Detector, a separate spare parts list is available:

Wireless DR Detector - Spare Parts List Links for Offline Navigation


Spare Parts List Click here

NOTE:
The least replaceable unit (LRU) is the Receptor Assembly.
For more details about the DR Detector return procedure, refer to section 0.

3.1 Required Information in Case of Image Quality Problems and


Exchange of the Wireless DR Detector

NOTE:
When dealing with image quality or bad pixel mapping problems of the detector the
FSE must get in touch with the RSN/GSC.
Only when the information listed in this section will be made available to the RSN/GSC,
the advice and counsel elaborated at the RSN/GSC will really help in troubleshooting.

Information needed in following cases:


Perceived problems with the flat panel strongly indicate a possible necessity
for replacement / exchange due to image quality problems.

The following data are required in the GSC case:


 Detailed problem description
 Actions performed
 Serial number of the Wireless DR Detector
 Installation date of the Wireless DR Detector
 Exchange date of the Wireless DR Detector
 Last calibration date of the Wireless DR Detector

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 12 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

Relevant files to be posted on an FTP server in Mortsel:


Following document describes how to access and upload files on the AGFA FTP
server. It is available from the Agfa HealthCare Library:
General Info > Agfa HealthCare > Publications > Service
Bulletin> ”General info - Service Bulletin - Temporary AGFA FTP Account”.

Temporary AGFA FTP Account Links for Offline Navigation


Service Bulletin Click here

(1) Create a folder in the directory Canon with Customer name and GSC Case #

(2) Go to XRDI\Panels\CDI\Configuration or
XRDI\Panels\CDI\Bin and upload a zip file with the following files:
 NX Snapshot
 All files in QCOutputData dir:
Contains the calibrate /self diagnosis files off the Canon DR Detector
 All files in Log dir: Contains the Canon logging

(3) Inform the RSN/GSC when the upload is finished.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 13 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

3.2 Wireless DR Detector Replacement Instructions

IMPORTANT:
In case of a Wireless DR Detector replacement update the serial numbers of the
Wireless DR Detector:

(1) Replace the Wireless DR Detector.


(2) Update the DX-D 600 Wireless Serial Number tracking document and insert the
new serial numbers and installation date of the replaced DR Detector.

Refer to the DX-D 600 Wireless Serial Number Tracking document.

DX-D 600 Wireless - Service Documentation Links for Offline Navigation


DX-D 600 Wireless Serial Number Tracking Click here

(3) Send the document via E-mail to: dr_start_up_team@agfa.com

3.3 Disassembly and Re-assembly of the Wireless DR Detector System

IMPORTANT:
Before starting with the disassembly and re-assembly of the Wireless DR Detector
refer to chapter 3 - Safety Instructions for safety descriptions.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 14 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

3.3.1 X-ray Interface Box

3.3.1.1 Replacing the PCA-XIF Board

(1) Turn off the power of the X-ray interface box (by disconnecting the AC adapter).
(2) Disconnect the LAN cable and grounding wire.

NOTE:
Do not pull the cables free in order to disconnect the connectors.
When re-assembling the parts, follow the procedure for their
disassembly in reverse unless otherwise specified.

(3) Remove the two screws on the bottom panel of the X-ray interface box.
(4) Remove the six crews on the side panels.
(5) Slide the top cover of the X-ray interface box toward the side where the
connectors of the box are connected, and lift to remove it.

NOTE:
Do not incline the top cover too far since doing so may damage the
inside of the X-ray interface box.

Figure 8

(6) Remove the eight screws used to secure the PCA-XIF board.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 15 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

(7) Lift the PCA-XIF board in the direction shown by the red arrows.

Figure 9

(8) Mount the PCA-XIF Board in reverse order.

3.3.2 PCA-XIF Board Firmware Updating and Recovery

The PCA-XIF board has two code areas where the firmware is stored:
 One area for backup code positioned at the head of the ROM
 One area for the normal codes stored in an intermediate area of the ROM

When the power is turned on, the backup code initializing routine boots, and the
backup code body boots when the normal codes have been erased for some reason or
when the BOOTSEL switch on the board has been pressed.
Unless other specific action is taken, this routine boots the normal code body.
For regular updating only the normal codes are updated.
This has the effect of preventing the board from failing to boot since the backup codes
at the ROM head are preserved even if a power failure should occur during updating,
for example.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 16 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

3.3.2.1 Accessing the Web Browser Tool

For the connection between the PCA-XIF board and the NX, a crossover cable must be
used.

The default IP address of the board is: 192.168.100.96/255.255.255.0


IP address of the PC used to communicate with the PCA-XIF board must have the
IP address that belongs to the same network.
(Recommended address: 192.168.100.10/255.255.255.0)

(1) Enter the IP address of the PCA-XIF board in the URL field of the web browser.
Default IP address as factory: 192.168.100.96

If the IP address has been changed and no longer known refer to


section 3.3.3.1 Forced backup boot operation
(2) Input the user name and password, as described in the training.

3.3.2.2 Checking the Firmware Version

(1) Click <status view> to display


information on the actual firmware
version, boot reason and network
configuration.

Figure 10: Status View screen

(2) Click <return to top> to open


the top menu screen.

Figure 11: Status View screen

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 17 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

3.3.2.3 Updating the normal Codes

(1) Click <firmware update>.

Figure 12: Top Menu screen

NOTE:
Do not make any changes in the “ROM
Area” input field.

(2) Click <Browse> to select the


firmware file.
Te firmware used has the
extension: .ehex Figure 13: XIF firmware update screen

(3) Click: <UPDATE>

It takes 10 to 20 seconds until the screen appears:


XIF firmware update was successfully done

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 18 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

3.3.2.4 Update Verification

(1) Return to the top menu screen.


(2) Click <status view>.

Figure 14: Top Menu screen

(3) Check the firmware version (running).


(4) Check that the version in the normal area (Flash #1) has been updated.

NOTE:
To activate the updated firmware press the RESET switch on the PCA-XIF board.

3.3.2.5 Updating the Backup Codes

IMPORTANT:
The backup codes should not be updated unless otherwise specified such as when a
fatal bug is found in the backup codes.

The procedure to update the backup codes is virtually the same as the procedure for
the normal codes.
To update the backup codes, except for inputting “flash0” (half-size lower-case
characters with no spaces) in the “ROM Area” input field, follow the same procedure as
for the normal codes.
When the updating has been completed successfully, check that the version in the
backup area (Flash #0) has been updated.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 19 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

3.3.2.6 Telnet Connections

Subsequent to Windows Vista, telnet which is the PC application for logging in to the
PCA-XIF board is not installed under the standard setup. However, a telnet connection
with the PCA-XIF can still be established by using TeraTerm.

Perform following steps when TeraTerm is used:

(1) Start up TeraTerm.


(2) On the setting screen, set:
 TCP/IP as the connection method
 The IP address of the PCA-IF board as the host address
 Telnet (change from the default of SSH) as the Service
 23 (default) as the port number
(3) Click <OK> to initiate the telnet connection.
Perform following steps when the telnet client is used:

(1) Select: <Control Panel  Programs  Windows functions enable


or disable>
(2) Enable: <telnet client>
(3) Start the command prompt, input the “IP address of the telnet PCA-XIF board”,
and establish the telnet connection.
(4) The login prompt can be input simply by pressing <return> without entering the
login name or password.
(5) When XIF_GUEST appears, login is completed.
(6) To log out, use the <bye> command.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 20 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

3.3.3 PCA-XIF Board Firmware Recovery

It is possible for the firmware in the normal area to be erased or damaged if the
updating has failed because a communication error occurred while the normal area
firmware was being updated or because the power of the X-ray interface box was
turned off.
If the normal area firmware has been erased, the backup code body can be booted by
the backup code initializing routine when the power is turned off and then back on so
the firmware can be updated again.
However, if the normal area firmware has been damaged, the backup code body will
not be booted simply by turning the power off and then back on, and it will be
necessary to forcibly boot the backup codes.
This section describes how to perform the forced backup booting operation and also
the precautions to be heeded.

3.3.3.1 Forced Backup Boot Operation

If the normal area firmware has been damaged, the firmware in the backup area can be
forcibly booted by operating the switches on the board.

(1) Start pressing the RESET switch and BOOTSEL switch on the PCA-XIF board at
the same time.
(2) With the BOOTSEL switch still held down, release the RESET switch first.
(3) Wait for one second, and then release the BOOTSEL switch.

Figure 15

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 21 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

When the firmware has been booted as a result of taking the steps above, the
IP address and MAC address are set to the default values.

Default settings:

 IP address: 192.168.100.96
 Subnet mask: 255.255.255.0
 Default gateway: 0.0.0.0
 MAC address: 00:1E:8F:CA:87:80

NOTE:
 Since the IP address and MAC address (separate values different from
00:1E:8F:CA:87:80) written in the flash memory are displayed as the IP address
and MAC address indicated in the network settings on the Status View screen for
the Web browser, the default values will not be displayed even if forced backup
start is implemented.
 Even after forced backup start has been implemented, the original IP address
and separate MAC address are stored on the flash memory so the original
addresses will be restored when normal start is initiated.

Since a separate MAC address is written for the PCA-XIF board, care is required
when proceeding with forced backup booting.

NOTE:
Since the operating system of the PC remembers the separate MAC address of the
PCA-XIF board, the PC tries communicating with the default MAC address of the
PCA-XIF board while the separate MAC address remains valid, and communication
may fail.

For details on how to avoid this kind of situation, refer to section 3.3.3.2 Concerning
ARP clear for the PC.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 22 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

3.3.3.2 Concerning ARP Clear for the PC

On the PC, open the command prompt, and execute the following command:

arp – d *

The above arp command has the effect of clearing the separate MAC address of the
PCA-XIF board stored in the memory of the PC’s operating system and of re-initiating
the inquiry.
Unless this step is taken, even though the MAC address of the board is switched to the
default value by the forced backup boot operation, the PC will still try communicating to
the separate MAC address used before the switch, and no further communication may
be possible by the telnet terminal or web browser.

NOTE:
The operating system of the PC stores the IP address and MAC address
correspondence table in the memory, and reset is initiated when the PC is rebooted or
there has been no communication for a certain period of time.
This means that it is not always necessary to execute the ARP command.
It is necessary only when communication is to be performed immediately to a different
MAC address using the same IP address.
It may also be necessary after the PCA-XIF board has been replaced when the IP
address matches and communication is to be tried immediately.

In a PC using Windows Vista or a later operating system, an error results when


the above command is simply executed even if the user is the administrator.
To execute the command correctly, the command prompt for executing the arp
command must be booted in the Run as administrator mode.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 23 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

To boot the command prompt in the <Run as administrator> mode,


proceed as follows:

(1) Select: <All Programs  Accessories>


(2) Right-click on <Command Prompt> and select <Run as administrator>.

Figure 16

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 24 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

3.3.4 Additional Information

3.3.4.1 Changing the IP Address of the PCA-XIF Board

The IP address of the PCA-XIF board can be changed or checked using the ipadr
command in the telnet connection status.

Example 1: When changing the IP address from “192.168.100.96” to “192.168.100.97”


XIF_GUEST>ipadr –ip 192.168.100.97

NOTE:
 The power must be turned off and turned back on for the new setting to be
reflected.
 To set the IP address of the PCA-XIF board into the detector, start the AP/LNK
configuration tool of Service Tool, and take the following steps:

(1) Input the IP address (192.168.100.97 in this case) of the PCA-XIF board
which has been changed to the X-ray interface address on the Configuration
tab screen.
(2) Press: <Save>
(3) Bring the infrared communication area of the Wireless DR Detector close to the
infrared communication unit, and proceed with the LNK operation.

NOTE:
After ipadr-ip 192.168.100.97 has been input, the “New setting takes effect after
power cycling” message is displayed.
Before turning the power back on, input ipadr and when the current status is checked,
192.168.100.96 will be displayed for current and 192.168.100.97 for the flash memory.
At this point in time, it is clear that only 192.168.100.97 has been written in the flash
memory and that the new setting has not been reflected.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 25 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

Figure 17: Screen on which to change and check the IP addresses

When the PCA-XIF board is now accessed from the Web browser and the details of the
network settings are checked, a display similar to the one shown below appears, and it
is clear that the setting in the flash memory has been changed, and the display now
reflects this change.

Figure 18: Screen on which to change and check the IP addresses

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 26 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

Example 2: When changing the IP address to the default “192.168.100.96”

XIF_GUEST>ipadr –default

NOTE
 The power must be turned off and turned back on for the new setting to be
reflected.
 To set the IP address of the PCA-XIF board into the detector, start the AP/LNK
configuration tool of Service Tool, and perform the following steps:

(1) Input the IP address (192.168.100.96 in this case) of the PCA-XIF board which
has been changed to the X-ray interface address on the Configuration tab
screen.
(2) Press: <Save>
(3) Bring the infrared communication area of the Wireless DR Detector close to the
infrared communication unit, and proceed with the LINK operation.

3.3.4.2 Obtaining the PCA-XIF Board Log

The PCA-XIF board is provided with a 3 MB log memory. In situations where trouble
has occurred, for instance, the log of the PCA-XIF board can be obtained by clicking
on the <log download> link on the web browser top screen of the PCA-XIF board.

NOTE :
When the <log download> link is clicked, the Open/Save dialog box is displayed.
If too much time is now taken to input the storage location, only the initial part of the log
details and not enough information may be obtained. The time permitted for the input
operation is about 30 seconds.
If the amount of the log information which has been saved is correct, the file size
indicated on the Open/Save dialog box and the size of the file actually saved will
match. Therefore, when obtaining the log, it must be checked whether these two file
sizes match.

Reference:
When the log memory becomes full, the log information is successively erased starting
with the oldest, and the latest information is written in its place. This means that the log
memory can always be expected to be nearly full (2.8 to 3 MB) with the log of the PCA-
XIF board log which is actually used.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 27 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

3.3.5 PCA-FE Board Firmware and FPGA Updating and Recovery

The operations for the PCA-FE board firmware and FPGA updating can be performed
using the tool for upgrading firmware and FPGA.
The updating is performed while power is supplied using the battery unit.
Since the pre-updated program will continue to run even after the updating, booting is
initiated using the new program either by removing the battery and re-installing it or by
turning the power off and then back on.

3.3.5.1 PCA-FE Board Firmware and FPGA Recovery

A boot switch is provided near the battery connector of the Wireless DR Detector for
recovery purposes when the data cannot be written correctly for some reason during
the updating of the PCA-FE board firmware or FPGA, resulting in trouble with the board
operation.

Perform the following steps to boot operation in the backup mode:

(1) Remove the battery.


(2) Set the boot switch to the backup mode.
(3) Install a battery which has been charged.

NOTE:
Updating is possible only when the remaining battery charge level is
60 % or more. Updating is recommended using a battery with a full
charge.

(4) Press the power switch of the Wireless DR Detector for at least a second to turn
on the unit’s power.

Operation now starts up using the backup firmware and backup FPGA.
When the power is turned on with the boot switch at the backup mode setting, the
backup code firmware is loaded into the SDRAM, and operations are performed.
With the backup firmware, the FPGA config data for backup is written into the FPGA
config ROM, and operation starts up.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 28 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

NOTE:
While the backup codes are being copied, the LINK LED (blue-green) blinks while its
light quantity changes gradually.
Since writing is completed when the LED stops blinking, it is necessary to wait at least
90 seconds.

The firmware or FPGA updating can be performed using the service tool while in the
backup mode boot status.
After updating has been completed in the backup mode boot status, follow the steps
below for booting in the normal mode.

(1) Hold down the power switch of the Wireless DR Detector for at least 5 seconds
to turn the unit’s power OFF.
(2) Remove the battery.
(3) Return the boot switch to the normal mode setting.
(4) Connect a battery with a full charge again.
(5) Press the power switch of the detector for at least a second to turn on the
unit’s power.

When the power is turned on with the boot switch at the normal mode setting, the
normal code firmware is loaded into the SDRAM, and operations are performed.

NOTE:
 However, with backup mode booting, the FPGA data is overwritten by the backup
codes, and the normal codes are lost. Therefore, when the backup codes and
normal codes are different, the normal codes of the FPGA must be updated.
 When booting the normal mode after the backup mode booting, check the
versions of the firmware and FPGA from the tool for upgrading firmware and
FPGA and perform update to have the right combination of them.

See also Figure 19 on next page.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 29 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

Figure 19

3.3.6 Additional Steps after Replacement of Parts

When the Wireless DR Detector has been replaced:


(1) Recognize the Wireless DR Detector by infrared communication
(using the service tools).
(2) Delete the registration of the Wireless DR Detector which has just been replaced
if necessary (using the service tools).
(3) Check the firmware and FPGA versions (and update them if necessary).
(4) Check that booting is performed correctly.
(5) Proceed with calibration.
(6) Conduct self-diagnosis.
(7) Carry out the performance test.
(8) Check the image quality.

When the X-ray interface box and PCA-XIF board have been replaced:
(1) Change the IP address (but only when it has been changed from the
default setting).
(2) Check the firmware versions (and update them if necessary).
(3) Check that booting is performed correctly.
(4) Check that the operations for taking and obtaining the images and for transferring
the data can be performed normally.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 30 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

When the Wireless DR Detector cable has been replaced:


(1) Check that booting is performed correctly with the wired connection.
(2) Check that the operations for taking and obtaining the images and
for transferring the data can be performed normally.

3.4 Wireless DR Detector Return Procedure

This section defines the roles and responsibilities when it comes to necessary
exchanges of the flat panel detector due to failure.

The customer always contacts the local AGFA NSO, when he encounters a problem
with the AGFA DR-System.

AGFA representative will send a Field Service Engineer (FSE) to look at the problem.
For DR detector failure confirmation the Agfa FSE always needs to escalate to
Agfa Level 2 (Tier 1) support (GSC case). AGFA representative will analyze the
problem.

For failure assessment the data listed in section 2.1 are required in the GSC case:

If failure is confirmed then Level 2/Level 3 support will supply the FSE with:
 Canon Medical Equipment Inquiry Form
 Further instructions to order a new detector and to return the defect one

IMPORTANT:
An immediate return of the DR Detector to the AGFA spare parts logistic center is
necessary, to make optimal use of the warranty period.
For fast processing of the DR Detector exchange it is essential to make sure that
all data mentioned above is available at time of escalation.

3.5 Acceptance Test after Wireless DR Detector Replacement

NOTE:
In case of a Wireless DR Detector replacement refer to Chapter 11 - Acceptance Test
after DR Detector Replacement.

DX-D 600 Wireless - Service Documentation Links for Offline Navigation


Acceptance Test after DR Detector Replacement Click here

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 31 of 32


02-2012 Agfa Company Confidential
DD+DIS101.11E Repair

4 XRDI

No repair documentation available.

5 NX Workstation

Refer to NX workstation Service Manual for repair descriptions.

NX Workstation - Service Documentation Links for Offline Navigation


Clean Install Procedure Click here

6 Agfa Digitizer (Optional)

Refer to the respective Agfa Digitizer Service Documentation for repair descriptions.

7 DX-D FLFS Stand (Optional)

NOTE:
For the DX-D FLFS Stand is a separate spare parts list available.

DX-D FLFS Stand – Spare Parts List Links for Offline Navigation
Spare Parts List Click here

The repair descriptions are delivered with the respective spare part.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 DX-D 600 Chapter 6 / Page 32 of 32


02-2012 Agfa Company Confidential

You might also like