A C Q U I S I TI O N I N F O R M A TI O N
In the Noise and Drift qualification, mobile phase (90% Water : 10% Acetonitrile) is pumped at 0.6 mL/min.
A blank chromatogram at 254 nm is collected at 3.6 nm resolution, 2 points per second and a 1 second
time constant for 15 minutes.
The average peak to peak noise is calculated based on a 30 second noise interval from 3 to 15 minutes.
The average peak to peak noise in microAU is reported.
The qualification passes if the Noise is < or = 80 microAU.
Drift Result
PDA 254.0 nm
Noise Chromatogram
0.00004
0.00003
0.00002
AU
0.00001
0.00000
-0.00001
3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00 15.00
Minutes
Channel Description PDA Spectrum (210-400)nm; Processed Channel Descr. PDA 254.0 nm