Professional Documents
Culture Documents
Treatment Planning Lab Paper
Treatment Planning Lab Paper
The patient that was selected had a lung tumor that was located in the center of the left
lung and the surrounding structures that were contoured were the left and right lungs, spinal
cord, and heart. The treatment plan was created with an AP/PA arrangement and the beam
weighting was equal for both plans. A 6MV beam was selected for the plan as well as a 1 cm
margin around the Planning Target Volume (PTV) for both beams. The plans were normalized to
isocenter which were placed in the center of the target volume. The 100% line is represented by
the yellow line, the 95% line is the green and the 90% line is the blue.
Figure 5 - Dose values for each structure for plan with heterogeneity correction on
Figure 11 - Dose values for each structure for plan with heterogeneity correction off
When treating/planning lung treatments, there are many aspects to consider for the plan.
One of the biggest differences between lung planning versus any other type of planning
throughout the body is the tissue density. While the Hounsfield Unit value of muscle is around
40 and water is around 0, lung has a value of around -800. This means that the dose distribution
will be much different through this area of the body as compared to the rest of the body. The
lower density that a lung has, the higher the dose will be in and through the rest of the lung.1
As shown in figures above for my patient, the monitor units and the hot spot are greater for the
plan that has the heterogeneity turned off. This goes to show that when the density is less, the
Since lungs are a structure that is very low density, there are other problems that can arise
in treatment planning. One of these problems is the energy selection that is used to treat the
tumor. Since the tumor is oftentimes located deep in the lung, the thought would be to use a
higher energy beam. However, the higher the beam energy when treating through a low density
material like lung, the more lateral electron scatter one will have.1Therefore, the plans that were
created above both used the lowest energy that available for the treatment plan. The lateral
electron scatter also plays a big role when there is a change in heterogeneity. In the first plan
when the heterogeneity is on, the dose distribution builds up in an odd way with the 100% line
covering the body upon entrance. The dose then scatters a bit more with the low density material
and then builds back up when it hits the tumor. In the other plan when the heterogeneity was
turned off, the isodose lines have no problem staying consistent through the patient as well as
One major aspect between the two plans is the difference in the body inhomogeneities. In
the first plan when the heterogeneity is on, the beam first has to travel through soft tissue and
some bone and then has to traverse through the lung tissue. When the heterogeneity is off in the
second plan, there is no difference to how the beam is traversing between any of the tissue. This
difference in body inhomogeneity plays a huge role in how the dose will actually get to the target
volume and how the beam will be absorbed throughout the tissue.1 This is also one of the two
categories that can be an effect of tissue inhomogeneities. The other being the secondary electron
fluence changes. As shown between the two plans, the monitor units are a bit higher for the plan
that has the heterogeneity turned off. Since there is no difference in the tissue the beam is
traveling though, there will be an overestimation of the amount of dose the tumor is actually
getting. The same thing can be said for when higher energy beams are used. The increase in
energy causes more lateral scatter which also causes more dose and a lower amount of coverage
to the actual tumor.2 The difference in body inhomogeneitiescan also cause complications to the
surrounding tissue. If the value of the tissue is not correct, then the surrounding tissue could be
Dose accuracy is an important aspect of any treatment plan. One of the biggest
differences in heterogeneity in general is when metal artifact is in the path of the beam. Metal
artifact is usually set to an electron density of around 1.8 which is similar to the density of bone.3
This is a much higher electron density than soft tissue which means that it needs to be accounted
for when creating a treatment plan. The monitor units can change in a plan by up to 3MU for
each fraction that the metal artifact is not corrected for.3 This means that the patient could be
receiving more dose than what was actually prescribed. The same idea can be applied to lungs.
The electron density is much lower than soft tissue and can cause a big difference in the dose
distribution. If the heterogeneity is turned off, the beam has only one electron density to travel
through versus the multiple densities it travels through when the heterogeneity is turned on. This
can cause a change in the monitor units that are used as well as the total dose that the patient is
receiving.
This same principle can also be applied to when there is metal streaking from a prosthetic
or other metal object within the body. Metal streaking can cause the scanner to pick up a higher
density in an area than what is actually there. This is oftentimes corrected to a general value that
resembles the tissue around it. This can also be related to the two different plans above. There are
times that the lung is corrected or overridden because there could be a change in density or to
make the dose to the area more accurate to what it will actually be.
When it comes to planning lungs, I do not correct for any difference in tissue densities.
When the heterogeneity correction is turned on, it gives a more accurate plan as well as an
accurate representation of how the dose adds up for the other structures. The differences in the
two plans above show how important it is to make sure all tissue densities are correct. Having the
correct monitor units to make sure the dose is what is expected is vital as well as ensuring that
the coverage on the tumor is what is expected. All in all, heterogeneity correction is important
regardless of where the target volume is, but it is especially important when it comes to lung
1. Khan FM, Gibbons JP.Khan’s the Physics of RadiationTherapy.5th ed. Philadelphia,
2. Wang L, Yorke E, Desobry G, et al. Dosimetric advantage of using 6 MV over 15 MV
photons in conformal therapy of lung cancer: Monte Carlo studies in patient geometries. J
3. Banaee N, Barough MS, Asgari S, Hosseinzadeh E, Salimi E. Evaluating the effects of
metal artifacts on dose distribution of the pelvic region.J Cancer Res Ther.
2021;17(2):450-454. doi:10.4103/jcrt.JCRT_786_19