This form may be copied provided that the copyright notice is retained. Tube Assembly Leakage Complies (≤ 1 mGy)? q Max measured leakage = … µGy.h-1 at 1m Maximum leakage from:-
This form may be copied provided that the copyright notice is retained. TS 41 – APPENDIX B ADDITIONAL IMAGE INTENSIFIER rh2006/03
PRACTICE… GENERATOR… ROOM…
Image Intensifier? q DR detector? q Low contrast detectability …... mm
Manufacturer …. Complies (≤ 1 mm at10%)? q Model …. Limiting resolution ....... lp.mm-1 Field sizes … cm … cm … cm Complies (See Table 5)? q Grid? q Maximum Fluoroscopic Dose Rates Ratio… Focus … Line/cm… Normal fluoroscopic mode Image and X-ray Beam Congruency Max free-in-air .... mGy.min-1 Beam ¨ or ¡? Area ratio (sect 5.2)=… using …. kVp …. mA Complies? (Ratio = 1.05) q Intensifier Input Dose Rates Complies (≤ 100 mGy.min-1 )? q Field size … cm .... µGy.min-1 Typical with PEP .... mGy.min-1 using … kVp …. mA using …. kVp …. mA -1 Field size ... cm .... µGy.min Complies (≤ 15 mGy.min-1 )? q using … kVp …. mA High dose rate mode? q Field size … cm .... µGy.min-1 Max free-in-air .... mGy.min-1 using … kVp …. mA using …. kVp …. mA Field size … cm .... µGy.min-1 using … kVp …. mA Complies (≤ 150 mGy.min-1 )? q Image Quality ( …...... cm field) Separate audible or visual signal? q Mode selector uses deadman switch? q Image distortion visible? q If þ, unit complies. Low contrast threshold ……. % If ý, mode deselects automatically - a) after a fixed time if unused? q Complies (≤ 5% at 10 mm)? q b) at machine power off? q Complies? (both a & b þ) q Comments