Professional Documents
Culture Documents
Digital Oscilloscope Measurements in High Frequency Switching Power Electronic
Digital Oscilloscope Measurements in High Frequency Switching Power Electronic
net/publication/3088054
CITATIONS READS
27 1,837
2 authors:
All content following this page was uploaded by Jean-Pierre Keradec on 07 February 2019.
856 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 41, NO. 6, DECEMBER 1992
I. INTRODUCTION
11. ANALOGREQUIREMENTS
A. Bandwidth Requirements
Fig. 2. Typical voltage waveform.
The bandwidth of each measurement channel affects the
observation of the switching time. Instead of obtaining
the real switching time t,, tu is measured by the voltage pends on resistances only, whereas the high-frequency
measurement channel, and tc is measured by the current gain G H F is fully determined by capacitances. If the probe
measurement channel. For the moment we assume that is perfectly compensated, its voltage gain remains con-
the time delay td between voltage and current measure- stant across the whole bandwidth. If not, G L F and G H F
ment is 0. In order to obtain an overall 5 % accuracy, the may differ 1-3% [4].The crossover occurs at a few kHz.
budget for this error contribution is set to 2 % . The values The collector emitter voltage VcE of the electronic
of tu and t, are independent. From (2) it follows that tu switch is shown in Fig. 2. The wave form contains a dc
and tc must be both less than 1.02 times the real switching component of approximately 200 V and an ac component
time t,. which is slightly less than the dc component, in this case
In first approximation, input amplifiers and probes act 198 V. The dc component is subject to the dc voltage gain
as low-pass filters. Rise time t, and bandwidth B are re- G L F and the ac component to the high-frequency gain G H F .
lated by: A 0.05% difference between G L F and G H F causes a per-
0.35 ceptible 0.1 V or 5 % inaccuracy in the minimum value of
t, = 7 and to = (3) vCE*
This level of performance is not readily available. The
where to and t, represent, respectively, output and input mismatch between G L F and G H F can be measured [ 5 ] .
rise times. In (3), tmis substituted for ti, and tu or t, for to Digital filter techniques are then used to correct the results
as applicable. for VcE. However, it is felt that the desired accuracy per-
From (2) and (3) it follows that the bandwidth should formance has not yet been fully achieved.
be at least 70 MHz to satisfy the requirement. Using a The low-frequency response problems are not limited
round number, this turns into a 100 MHz bandwidth re- to those of the probes. The internal attenuators of the os-
quirement for an oscilloscope input channel. In reality, cilloscope can cause the same kind of problems. Careful
the bandwidth is limited by the performance of the probes. observation of the square wave without the use of a probe
Probes with higher attenuation exhibit lower bandwidth is helpful to evaluate gain flatness of the oscilloscope.
performance. Fortunately, only slower switch power de- Digital filtering techniques are also used to correct the
vices are capable of higher stand-off voltages. Current distortion caused by high-pass filtering of the passive cur-
probes exhibit the same characteristic. Note also that high- rent probes [5].
voltage probes are passive, requiring a 1 MQ oscilloscope
input impedance. D. Overdrive Voltage Limits and Recovery Times
The sensitivity of the oscilloscope may be increased to
B. Differential Time Delay Between Voltage and better observe the lower portion of the wave form and
Current Measurement improve resolution. However, this causes overload of the
The difference td between time delays in the measure- input section of the oscilloscope when the signal is near
ment of instantaneous voltage and current also causes an its maximum value. Two voltage limits can be distin-
error in the computation of power. From (2), with tu = tc guished. The full-scale value of the analog-to-digital con-
= rm (infinite bandwidths), it can be observed that td = verter (ADC) generally corresponds to 5 divisions; see
t,/lOO results in a 2 % error contribution in P,,,. Fig. 3. The dynamic range of the amplifier is always
Most of the time lag is caused by the probes, and it higher. It depends very much on the particular model used
often reaches 20 ns. Compensation is necessary. Some and sensitivity selected.
oscilloscopes have provisions built in; in other cases, The ADC recovers quickly when the overload condi-
computer compensation must occur, and the time lag tion is removed, in most cases within several micro-
should be known with an uncertainty better than tm/ 100 seconds, down to approximately 50 ns [ 6 ] .Recovery from
= 250 ps. a saturated input amplifier is considerably longer, typi-
cally more than 100 ps. The fact that temporary overdrive
C. Low-Frequency Response conditions cause significant errors is generally known:
A passive voltage divider probe contains a capacitance Yet, only few manufacturers publish specifications for this
adjustment to obtain the best match between the DC and behavior, and no testing methods appear to have been es-
the low-frequency response. The dc voltage gain G L F de- tablished.
-
858 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 41, NO. 6. DECEMBER 1992
B. Sampling Rate
5 div Based on Nyquist criteria and the analog bandwidth re-
0 div quirements that were established in the previous section,
it follows that the sampling rate of the digital storage os-
cilloscope should be at least 200 Ms/s. However, the fre-
quency content of the signal includes components higher
than 100 MHz. As a minimum, the sampling rate should
be at least 2.5 times higher. In that case, 12 to 13 samples
of the wave form are available to cover the 25 ns switch-
ing interval.
We consider it to be desirable to obtain 50 samples dur-
-4div
ing the switching interval to obtain a smooth representa-
tion of instantaneous power wave forms. This corre-
.J div
sponds to a sampling rate of 2 G s / s . When considering
compensation of different delays of the voltage and cur-
rent measurements, it appears to be desirable to have 250
Fig. 3. Overdrive limits. Whereas the dynamic range of the ADC is almost
always five divisions, the input amplifier limits depend on the instrument ps resolution; this would mean a sampling rate of 4 G s / s .
used and its sensitivity setting. In reality, the 500 Ms/s rate is sufficient when interpo-
lation techniques are used.
The interest for such specifications is illustrated by the
C. Record Length
following remark. For the Tektronix model 11A32 plug-
in, the input amplifier is not driven into saturation by a The frequency of operation that is selected for static
0.4 V signal with 1 mV/div [ 6 ] .When applying a 4 V converters is related to the switching time of the semicon-
signal, the sensitivity must be set to 100 mV/div to avoid ductors used. Practically, the period is between 100 and
overdriving the input amplifier. This suggests the use of 400 times the switching time. Based on 12 samples during
a 1000 : 1 rather than a 100 : 1 probe for measurement of the switching time, it follows that 10 000 to 30 000 sam-
the lower portion of the Vc, curve in Fig. 2. The overall ples are sufficient to record two to five signal periods.
sensitivity with the first probe is 1 V/div. With the sec-
ond probe the overall sensitivity is only 10 V/div. Keep- IV . SIGNAL-PROCESSING CAPABILITIES
ing in mind these behaviors, probe adjustment can be en-
hanced by using a positive square wave which does not So far, it has been established that a 12-bit digitizer
saturate the input amplifier and observing the low portion with a sampling rate of at least 500 Ms/s is needed. No
of the wave form with increased sensitivity [ 5 ] , [lo]. instrument is capable of delivering this level of perfor-
mance. However, since the signals are periodic in nature,
the required information can be accumulated during sev-
111. DIGITALREQUIREMENTS eral periods. Individual measurements can be averaged,
A . Resolution and equivalent time-sampling techniques can be used.
The resolution required to measure VcE is 0.1 V. The
maximum voltage that is expected for VcE is 400 V. This A. Averaging
means 1 part in 4000, i.e., 12 bits. The only way to re- Averaging techniques are used to improve the signal-
duce this requirement is to increase sensitivity to observe to-noise ratio (SNR). Averaging 4"readings improves this
the bottom of the signal. As discussed in the previous sec- ratio by a factor 2". It should be noted that SNR improve-
tion, this is not really feasible since a recovery time of 50 ments can only be obtained if the noise is random. Syn-
ns or more is not acceptable. chronous noise and constant offset voltages can be elim-
Practically, the top of the wave form is measured with inated by subtracting two successive acquisitions [ 11. The
low sensitivity (e.g., 50 V/div). The bottom portion of first measurement includes signal, noise and offset, and
the wave form is measured with a sensitivity whereby the the second only noise and offset. Also, the resolution is
input amplifier of the oscilloscope is not overdriven, but not necessarily increased by averaging. Reference [8] de-
the ADC is. Maximum advantage is taken of the quick scribes a simple means to increase resolution using the
recovery time of the ADC. Thereafter the top portion of average mode.
the VcE wave form is juxtaposed with the bottom portion The impact of the quantization error caused by the ADC
(that has been obtained with increased sensitivity) to ob- disappears from the averaged result as the rms input noise
tain the overall result. 3
exceeds of the least significant bit (LSB) [lo].
Use of clamping circuits between the connections to the The major remaining problem is the differential nonlin-
electronic switch under test and the input of the oscillo- earity of the ADC. A method to reduce this error is dis-
scope has not yielded satisfactory results or results that cussed in [8]. A simple test illustrates the impact of these
are any better than what we obtained [7]. small but troublesome errors. A triangular signal with 0.2
CAUFFET AND KERADEC: DIGITAL OSCILLOSCOPE MEASUREMENTS 859