Professional Documents
Culture Documents
Nakul Aquaronne2003
Nakul Aquaronne2003
The Sysmex XE-2100s (Sysmex Corp. optical microscopy and the XE-2100s were
Kobe, Japan) is a latest-generation hema- surprising, and favorable to the XE-2100s.
tology analyzer. Its optical and electrical This analyzer provides the user with an
measuring technology is improved by the undeniable feeling of security concerning its
addition of flux cytometry, fluorescence, reliability in detecting and identifying anoma-
and differential lysis. Its analytical perfor- lies in the automated leukocyte differential
mance in terms of precision, reproducibility, count. With a sensitivity of 96%, a negative
linearity, carryover, and time stability was predictive value (NPV) of 98%, and a false-
found to be entirely satisfactory. In addition, negative (FN) rate of 4%, the XE-2100s has
the results of 500 complete blood counts perhaps reached the technological limits for a
and differentials correlated perfectly with machine performing morphological recogni-
those obtained by the Coulter STKSs tion of normal and pathological blood cells.
(Beckman Coulter, Villapointe, France). J. Clin. Lab. Anal. 17:113–123, 2003. c 2003
The comparison of 500 leukocyte differen- Wiley-Liss, Inc.
tial count results analyzed in parallel with
Key words: hematology analyzer; Sysmex XE-2100s; Coulter STKSs; performance
evaluation
analyzer uses two fundamental analytical methods: Sysmex XE-2100s Analytical Performance
electrical and optical.
Precision studies
The electrical method combines two sources of
electric current: low-frequency direct current, which The objective of these tests was to determine whether
measures impedance and provides information on the analyzer was capable of reproducing the same results
cellular volume, and high-frequency alternating current for a given specimen when the analysis was done under
(radiofrequency), which supplies information on cell standard conditions, that is, by the same technician, at
structure and density. the same time of the day, and with the same reagents for
In the optical method, the origin of the light source is the count parameters. For specimen selection, specific
a semiconductor laser. The cells that are stimulated are criteria were chosen as follows:
differentiated using three methods: 1) forward-scattered
light, 2) side-scattered light (901), and 3) fluorescent * Leukocytes (four categories): o1 109/L, 1 109/L
emission from cells previously marked with a fluoro- to 4 109/L, 4 109/L to 10 109/L, and 420 109/
chrome (polymethine), which forms unstable complexes L.
with the ribosomal RNA and the DNA. * RBCs (three categories): hematocrit (Ht) o0.20 L/L,
Data on the analyzer operative characteristics for 0.37–0.42 L/L, and 40.50 L/L.
WBCs, WBC differential, red blood cells (RBCs), * Platelets (five categories): 10 109/L to 20 109/L,
nucleated RBCs (NRBCs), reticulocytes, and platelet 20 109/L to 50 109/L, 50 109/L to 100 109/L,
counts can be found in Refs. 1 and 2. 150 109/L to 400 109/L, and o400 109/L.
Table 1 summarizes in chart format the different
techniques used for the parameters studied. One specimen in each of the ranges (total of 12
samples) was tested a maximum of 20 consecutive times
in manual mode on the XE-2100s. The mean, standard
Specimens
deviation (SD), and variation coefficient were calculated
Venous blood samples in collection tubes containing a for each precision test.
tripotassium EDTA-type anticoagulant (Vacuettes,
Greiner Labortechnik) were used throughout the
Leukocyte and platelet linearity tests
evaluation for a period of 3 weeks. All of the specimens
(from Nice University Hospital, Nice, France) were These tests enabled validity thresholds to be estab-
analyzed within 6 hr after blood collection, except for lished for the linear relationship that exists between
those used in the time stability tests. theoretical and observed values.
More than 4,000 samples were studied in parallel on Two specimens were selected: one with leukocytes
the XE-2100s and the STKSs by determining a 450 109/L, and one with platelets >800 109/L. Two
complete blood count. In addition, 500 of the samples dilution ranges (1/2, 1/4, 1/8, 1/20, 1/50, 1/100) were
that generated no qualitative flag for the differential carried out in a diluent (Cell Packs, Sysmex) for the
count were used for correlation studies. Analysis of the same sample. Each range (containing seven points,
leukocyte differential count, combined with a systematic including the pure specimen) was tested three consecu-
blood smear stained using the May Grünwald Giemsa tive times in manual mode on the XE-2100s. The
procedure, was carried out for 500 randomly chosen relationship between the values of the range measured
specimens. These samples were taken from patients in and the values of the theoretical range was studied by
the intensive care and clinical hematology units of Nice determining the correlation coefficient and the regres-
University Hospital. sion line equation using the least-squares method. The
s
TABLE 1. Techniques used according to cell type Sysmex XE-2100
Cell type Forward scattered light Side scattered light Fluorescence Radio frequency Impedance
significance of the difference between these two ranges This procedure was carried out a maximum of 20 times
was analyzed by the paired Student’s t-test (3). in the same day in automatic mode on the XE-2100s.
For each reproducibility test, the mean, SD, and
Carryover studies variation coefficient were calculated.
– ‘‘Variant lymphs’’: atypical lymphocytes. except for the low values for WBCs (o1 109/L) and
– ‘‘Platelet clumps’’: platelet clumps. platelets (10 109/L to 20 109/L) (Tables 2–4).
– ‘‘NRBCs’’: nucleated red blood cells. The precision threshold of the analyzer corresponded
to these lower threshold values.
Each differential not given by the two analyzers was
considered to be pathological. The positivity threshold Leukocyte and platelet linearity tests
for the leukocytic elements not usually present in the
differential count was set at Z1% with regard to the For each sample (WBC average=358.69 109/L,
benchmark method. platelet average=1,468.00 109/L), two ranges of iden-
The use of optical microscopy enabled us to determine tical dilutions were prepared. The linearity tests were
the percentages of the normal and pathological differ- performed on these four ranges. The results were
entials (i.e., the true negatives (TNs) and true positives excellent and showed positive correlations. The correla-
(TPs)). For each analyzer, the rejection rates corre- tion coefficients found between the theoretical and
sponding to the ratio of the number of differentials with observed values for the leukocytes and platelets were
flags and the number of specimens examined expressed >0.99 (Figs. 1 and 2).
as a percentage were determined. The significance of the
difference in rejection rates was examined using the w2 Carryover studies
test. Then the number of false positives (FPsFthe
The carryover study procedure was carried out four
number of differentials with flags that were normal on
times in succession. A specimen with a high WBC value
optical microscopy) and the number of false negatives
(average=42.68 109/L) and another with a low WBC
(FNs–the number of differentials that had no flags but
value (average=0.27 109/L) were employed. The
were pathological on optical microscopy) were calcu-
average of the carryover percentages was 0.05%.
lated. The sensitivity, or the proportion of positive
For the platelets, two samples (one with a high value
results found by the analyzers among the truly
(average=1,702.33 109/L), and one with a low value
pathological differentials, was determined by the ratio
(average=4.33 109/L)) were chosen for this test.
TP/TP+FN. The specificity, or the proportion of
negative results found by the analyzers among the truly
normal differentials, was calculated using the ratio TN/
TABLE s2. Precision tests on the leukocytes (109/L) Sysmex
TN+FP. Finally, the positive (PPV) and negative XE-2100
(NPV) predictive values were determined by the ratios
WBC range Mean SD Variation coefficient
TP/TP+FP and TN/TN+FN, respectively. These
values correspond to the capacity of an analyzer to WBCo1 0.51 0.04 6.91
provide a truly positive result compared to all the 1oWBCo4 2.58 0.07 2.81
pathological differentials, or a truly negative result 4oWBCo10 8.29 0.09 1.10
WBC420 34.60 0.40 1.16
compared to all the normal differentials. All of the
results are expressed as a percentage. In addition, the
sensitivity, specificity, and PPVs and NPVs were also
studied for each of these flags. When two or more flags TABLE
s
3. Precision tests on the hematocrit (L/L) Sysmex XE-
were found to coexist, the nature of the flag retained as a 2100
priority was chosen using the following rule: blasts > Hematocrit range Mean SD Variation coefficient
[atypical or abnormal lymphocytes or lymphoblasts] >
Hto0.20 0.176 0.002 1.004
IGss> others [NRBC, platelet clumps, WBC distribu- 0.37oHto0.42 0.409 0.002 0.515
tion anomaly]. The set [atypical or abnormal lympho- Ht40.50 0.531 0.003 0.507
cytes or lymphoblasts] is represented by the global
‘‘atypical lymphocytes’’ flag for both systems.
TABLE
s
4. Precision tests on the platelets (109/L) Sysmex XE-
RESULTS 2100
Platelet range Mean SD Variation coefficient
Analytical Performances
Precision studies 10oPLTo20 12.81 1.52 11.83
20oPLTo50 24.88 1.22 4.90
Considering that the maximum acceptable variation 50oPLTo100 62.00 2.10 3.39
coefficient was 5%, the Sysmex XE-2100s precision test 150oPLTo400 262.60 5.34 2.04
PLT4400 588.29 9.18 1.56
showed excellent results for the parameters analyzed,
Sysmex XE-2100s Analyzer Evaluation 117
Reproducibility tests
In the precision study, the maximum limit selected for
Fig. 1. Sysmex XE-2100 s
leukocyte linearity test. the variation coefficient was 5%. All of the results were
excellent, with the exception of the platelet values, which
were 10 109/L to 20 109/L (Tables 5–7). For a
platelet value o20 109/L, the Sysmex XE-2100s did
not provide reproducible results.
Correlation Studies
For each parameter studied, there existed a positive
correlation between the Sysmex XE-2100s and the
TABLE 5. Reproducibility
s
tests on the leukocytes (109/L)
Sysmex XE-2100
WBC range Mean SD Variation coefficient
TABLE 6. Reproducibility
s
tests on the hematocrit (L/L)
Fig. 2. Sysmex XE-2100s platelet linearity test. Sysmex XE-2100
Hematocrit range Mean SD Variation coefficient
The average of the carryover percentages calculated Hto0.20 0.185 0.002 0.851
0.37oHto0.42 0.412 0.003 0.708
over five tests was 0.04%.
Ht40.50 0.543 0.003 0.625
The carryover percentages found were very satisfac-
tory. The interspecimen carryover was practically
nonexistent.
TABLE 7.s
Reproducibility tests on the platelets (109/L) Sysmex
XE-2100
Time stability tests Platelet range Mean SD Variation coefficient
These tests took place over a period of 4 days. All 10oPLTo20 16.64 1.15 6.91
specimens stored at room temperature (average of 251C) 20oPLTo50 41.29 1.59 3.85
showed an increase in MCV from 12 hr (minimum of 50oPLTo100 70.80 1.74 2.46
150oPLTo400 345.73 3.58 1.03
4 U), shown by a rise in the Ht and a reduction in the
PLT4400 629.36 8.77 1.39
MCHC. This phenomenon continued over time. Also,
118 Nakul-Aquaronne et al.
the CBC + differential count. The rejection rate of the Optical microscopy
flag (FNs) for six differentials (4%) and the STKSs Fig. 7. Numbers of Sysmex XE-2100s and Coulter STKSs
ignored 48 (33%) pathological differentials (Table 10). pathological differentials.
The type and frequency of detection errors that
resulted from fault or excess were analyzed. Figure 8
shows that for the XE-2100s, three factors of almost be noted that in the present study, neither of the two
equal frequency (corresponding to atypical lymphocytes analyzers ever generated flags for circulating blasts
(35%), IGs (31%), and other flags (NRBCs, platelet (Fig. 9). The results shown in Fig. 10 confirm that on a
clumps, and WBC distribution anomaly) (31%) were day-to-day basis, circulating IGs were the most fre-
responsible for the errors caused by excess. For the quently occurring anomaly. In our study, IGs repre-
STKSs, the causal factor was the IG flag (64%) (Fig. 8). sented 75% of the pathological differentials. The IG flag
For the FNs, it was principally IGs that were not was displayed in only 53% of the cases for the XE-
detected; the XE-2100s was in error 49% of the time 2100s and in 51% for the STKSs. On the other hand,
and the STKSs was in error 88% of the time. It should the flag for blasts was overused for the two analyzers.
120 Nakul-Aquaronne et al.
TABLE 9. Results of the leukocyte differential count evaluation Sysmex XE-2100 : number of False Negatives = 6
4% 11%
Sysmex XE-2100 : number of False Positives = 70 Blasts (16/147)
10%
3%
Atypical lymphocytes (15/147)
64%
Blasts (35/139)
6%
Fig. 8. Types of Sysmex XE-2100s and Coulter STKSs false Atypical lymphocytes (9/139)
positives.
51% Immature granulocytes (70/139)
Others (25/139)
The last part of the evaluation was devoted to Fig. 10. Type and frequency of leukocyte differential count
assigning values to the sensitivity and specificity, and anomalies.
the PPVs and NPVs for the two analyzers’ major flags
for the presence of blasts, atypical lymphocytes, and excellent. The sensitivities and specificities were 490%.
circulating IGs (Table 11). The relevance of the XE- In terms of PPV for these three types of pathologies, the
2100s flags for these three types of atypical cells was XE-2100s was no more successful than the STKSs for
Sysmex XE-2100s Analyzer Evaluation 121
TABLE 11. Flag Epidemiological Parameters For Sysmex XE-2100s and Coulter STKSs
Positive Negative
Sensitivity in % Specificity in % predictive value in % predictive value in %
the blasts and the atypical lymphocytes. On the other The linearity study was very convincing. The paired
hand, for the IGs, the PPV of 79.5% obtained by the Student’s t-test produced probabilities that exceeded the
XE-2100s was distinctly higher than that of the STKSs threshold for statistical significance (0.05). Thus, there
(55.7%). For the NPVs, the two analyzers obtained was no significant difference between the theoretical
excellent scores, both exceeding 90%. values and the observed values. As such, we recommend
that samples not be diluted up to the values of the
DISCUSSION specimens analyzed for the test (300 109/L for
leukocytes and 1,400 109/L for platelets).
Quantitative Analytical Aspects of the Sysmex XE-
The almost complete absence of intersample carry-
2100s
over is a valuable feature, especially for laboratories
In order to determine the analytical performance and working with samples from diverse sources such as
the advantages and limits of the Sysmex XE-2100s, we emergency rooms, cancer and hematology clinics, and
followed the evaluation protocol proposed by the intensive care units. A high level of leukocytes or
International Committee for Standardization in Hae- platelets in one specimen does not call the results from
matology (ICSH) (4). the specimen previously tested by the analyzer into
Precision tests showed excellent results, except for the question.
leukocyte low values (o1 109/L) and platelet low The results of the time stability tests confirm the need
values (10 109/L to 20 109/L). The same precision to keep the blood at 41C to maintain the stability of the
problems have been encountered with other analyzers differential count parameters (5,6,8,9). The quantitative
currently on the market, such as the Abbott Cell-Dyn aspect is thus stored up to 72 hr, and qualitative drifts
4000s (5) and the Sysmex SE 9500s (6). These weak are possible after 24 hr.
concentrations in WBCs and platelets constitute the On the quantitative side, all the correlations were
limits beyond which no current analyzer is capable of positive and significant, even though the correlation
counting with reproducibility. In contrast to the Cobas coefficients for basophils and MCHCs (r=0.558 and
Vega ABXs, the variation coefficient for platelet values 0.546, respectively) were not as good as the others. All
of 50 109/L to 100 109/L for the Sysmex XE-2100s calculated probabilities associated with the paired
is very good (7.70% vs. 3.39%) (7). Student’s t-test were o0.01, except for the lymphocytes
The reproducibility tests also yielded excellent results, (P=0.57) and basophils (P=0.20). With those two
except for the platelet rates of 10 109/L to 20 109/L. exceptions, there was a highly significant difference
The same situation is encountered with the Bayer Advia between the two analyzers for all the parameters for the
120s (8), the Abbott Cell-Dyn 4000s (5), and the CBC + differentials, which can be explained by the very
Sysmex SE 9500s (6). On the other hand, for platelet dissimilar technologies of these two devices. The
values of 20 109/L to 50 109/L, the Sysmex XE- quantifications of the lymphocytes and basophils by
2100s has a variation coefficient that is entirely the Sysmex XE-2100s and the Coulter STKSs were not
acceptable (3.85%), unlike the Sysmex SE 9500s (6) statistically significantly different. For those two types of
and the Bayer Advia 120s (8) (6.62% and 7%, cells, the analyzers provided identical feedback even
respectively). though the counting technologies were dissimilar. A
To compensate for the underperformance of the recent study of 544 specimens (1), in which results from
Sysmex XE-2100s, we propose the use of a measuring optical microscopy were compared to those from the
chamber control with optical microscopy (particularly Sysmex XE-2100s, showed a good correlation for the
phase contrast microscopy) for every leukopenia polynuclear basophils (r=0.756) and an excellent
o1 109/L and every thrombocytopenia o20 109/L. correlation for the lymphocytes (r=0.922). Briggs et
122 Nakul-Aquaronne et al.
al. (2) also found that the NRBC count was highly sensitivities and specificities 490% for major flags for
correlated (r=0.97) with the manual reference count. the automatic differential (blasts, atypical lymphocytes,
For platelet counts o100 109/L, the optical method and IGs). However, even considering its exceptional FN
showed excellent correlation with the reference flow rate, microscopeless cytology with the XE-2100s still
cytometry (r=0.97). The optical and impedance platelet has room for improvement. While the consequences of
counts were also well correlated (r=0.89) (2). However, errors due to excess are less dramatic than errors due to
the precision of the impedance method was somewhat a fault, the biologist is still required to intervene. The
better than that of the optical method (10). advantage of adding a complete blood count validation
algorithm, such as the MPL system, is that these errors
can then be managed in an effective way.
Analytic Aspects of the Sysmex XE-2100s
Leukocyte Differential Count
CONCLUSIONS
The XE-2100s is currently the top-of-the-line analy-
The XE-2100s is Sysmex’s top-of-the-line hematol-
zer on the market. Its performance in terms of the
ogy analyzer. Its performance and ease of use were
leukocyte differential count was eagerly anticipated,
surprising, especially because the Sysmex brand had not
given the new technological approach announced by the
attained this level of success in the past. Although its
Sysmex company.
analytic performance (precision, reproducibility, linear-
In our comparative study, the proportion of truly
ity, and carryover) did not differentiate it from other
pathological differentials was 29%. For the same
analyzers currently on the market, the results for the
samples, the XE-2100s counted 42% of differentials
leukocyte differential count were certainly remarkable.
with flags; almost every second differential was con-
The advantage lies in the use of fluorescence together
sidered pathological. This rejection rate can be ex-
with flow cytometry in the differential classification of
plained by a very high flag-triggering sensitivity. This
blood cells. According to our evaluation, no other
hypersensitivity is shown in the results by a lower PPV
device has obtained such a respectable NPV (98%) in
compared to that for the STKSs (67% vs. 71%), and in
determining the differentials. The level of FNs is so low
practice, it increases the number of manual checks
(4%) that it almost equals the diagnostic efficiency of an
required due to an increased rate of FPs. These errors in
experienced cytologist.
excess are not due to the IG flag (as is the case for the
The XE-2100s is a fully functional analyzer: the
STKSs (64%)), but are shared among the three
circulating NRBC count is subtracted automatically
principal flags (other than blasts) (Fig. 8). The specificity
from the leukocytes, and the reticulocytes are classified
of the XE-2100s is satisfactory, although it is lower
according to their stage of maturity. The analyzer
than that of the STKSs (80% vs. 89%, respectively).
introduces new parameters for erythrocytic anisocytosis
Moreover, Ruzicka et al. (1) found that the efficiency
and platelet indices. It expresses the IG rate as both a
rates of flagging for the presence of Z1% abnormal
percentage and an absolute number. Its throughput of
WBCs were 83% for the XE-2100s.
150 CBC + differentials per hour is particularly
Roche, the French distributor of the Sysmex brand,
impressive.
has produced software for the analytic management of
The XE-2100s is the first analyzer on the market to
results, called the MPL system. When connected to the
offer an exceptional quality/price ratio. When used
XE-2100s, this system acts as a mediator between the
together with validation software, it delivers perfor-
analyzers and the laboratory information system.
mance worthy of a true top-of-the-line analyzer.
Because appropriate parameters were set up for the
algorithm used to validate the CBC + differential
results, the rejection rate of the XE-2100s fell from 42% REFERENCES
to 19% in our laboratory. We recommend that future 1. Ruzicka K, Veitl M, Thalhammer-Scherrer R, et al. The new
users of the XE-2100s purchase such a filtering system hematology analyzer Sysmex XE-2100: performance evaluation of
to reduce the cost of having a technician verify the a novel white blood cell differential technology. Arch Pathol Lab
Med 2001;125:391–396.
differentials with flags. 2. Briggs C, Harrison P, Grant D, et al. New quantitative parameters
The XE-2100s represents a direct and major im- on a recently introduced automated blood cell counter–the XE
provement in terms of the security and credibility of the 2100. Clin Lab Haematol 2000;22:345–350.
differential results, as shown by its very high NPV rate 3. Schwartz D. Méthode statistique à l’usage des médecins et des
biologistes. Paris: Flammarion Médecine Science; 1986; p 307.
(98%). This rate also remains very high for each type of
4. England JM, Rowan RM, Van Assendelft OW, et al. Protocol for
major flag (Table 11). Of the current analyzers evaluated evaluation of automated blood cell counters. International
in our laboratory (5–8), the XE-2100s was the only one Committee for Standardization in Haematology (ICSH). Clin
that provided an exceptionally low (4%) FN rate, with Lab Haematol 1984;6:69–84.
Sysmex XE-2100s Analyzer Evaluation 123
5. Ferrero-Vacher C, Maerfeld K, Fraye M, et al. Évaluation du 8. Maerfeld K. Évaluation comparative des automates haut de
Cell-Dyn 4000s Abbott automate d’hématologie. Rev Fr Lab gamme d’hématologie au Centre Hospitalier Universitaire de Nice.
2002;340:53–64. Ph.D. thesis, University of Nice, Nice, France; 1999.
6. Ollier L, Maerfeld K, Ferrero-Vacher C, et al. Évaluation de 9. Erwa W, Bauer FR, Etschmaier R, et al. Analysis of aged samples
l’automate d’hématologie Sysmex SE 9500s. Rev Fr Lab with the Abbott CD4000 hematology analyzer. Eur J Lab Med
2001;333:33–42. 1998;6:4–15.
7. Ferrero-Vacher C, Sudaka I, Jambou D, et al. Evaluation of the 10. Sandhaus LM, Osei ES, Agrawal NN, et al. Platelet counting by
ABX Cobas Vega automated hematology analyzer and the Coulter LH 750, Sysmex XE 2100, and Advia 120:
comparison with the Coulter STKS. Hematol Cell Ther a comparative analysis using the RBC/platelet ratio reference
1997;39:149–158. method. Am J Clin Pathol 2002;118:235–241.