Professional Documents
Culture Documents
EC8073 ME 2M QB IQ - by WWW - EasyEngineering.net 2M QB IQ 1
EC8073 ME 2M QB IQ - by WWW - EasyEngineering.net 2M QB IQ 1
net
ww
w.E
asy
E ngi
nee
rin
g.n
et
ww
• EEG (Electroencephalogram)
• EOG (Electrooculogram)
w.E
• EMG (Electromyogram)
• PCG (Phonocardiogram)
3.
asy
What is Resting potentials? Or What is called as resting membrane
potential?
En
The membrane of excitable cells readily permits the entry of K+ ions and
Cl- ions, while it effectively blocks the entry of Na+ ions. Therefore the
gin
concentration of Na+ ions inside the cell becomes much lower than that outside
the cell. Since the Na+ ions are positive, the outside cells are more positive than
eer
the inside. Thus the charge balance is not achieved. Thus a potential difference
i
is developed across the membrane. This membrane potential caused by the
ng.
different concentration of ions is called the resting potential of the cell.
7.
ww
What is meant by depolarization and repolarization of a cell?
DEPOLARIZATION:-
w.E
When the impulse reaches the muscle, the polarized condition (-90mv)
is altered. i.e., the resting membrane potential is abolished. The interior of the
muscle becomes positive and outside becomes negative. This condition is called
as depolarization.
asy (Or)
En
The process of changing from resting state to the action potential state
is called as depolarization.
REPOLARIZATION:-
gin
eer
With in a short period, the muscles obtain the resting electrical
i
potential once again. Interior of the muscle becomes negative and outside
ng.
becomes positive. So, the polarized state of the muscle is re-established. This
process is called as repolarization.
(Or)
The process of changing from action state to the resting potential state
is called as repolarization.
net
8. What is meant by Depolarization?
The process of changing from resting state to the action potential state
is called as depolarization.
ww
electrodes resting on the skin is measured.
In unipolar electrode, the reference surface electrode is placed on the
w.E
skin and the needle electrode which acts as active electrode, is inserted into
the muscle.
asy
14. List the types of Biopotential Electrode. (Or) What are the different types
of electrodes used in bipolar measurement? (Or) Classify Bipotential
electrodes
En
The types of biopotential electrodes are,
a. Surface Electrode
b. Micro Electrode and
gin
c. Needle Electrode
eer
i
15. Draw the electrical equivalent circuit of a surface electrode
ng.
net
17. Name the electrodes used for recording EMG and ECG.
The electrodes used for recording EMG are,
• Surface electrode - Metal Disc electrode, Disposable electrode
• Needle electrode - Unipolar and Bi polar electrode
18. What are the different types of electrodes used in bipolar measurement?
The types of biopotential electrodes are,
a. Surface Electrode
b. Micro Electrode and
c. Needle Electrode
ww
19. What is half cell potential?
w.E
The voltage developed at an electrode-electrolyte interface is called as
half cell potential or electrode potential.
asy
20. Draw a typical ECG waveform.
En
gin
eer
i ng.
net
ww
i)
w.E
24. Mention the various lead systems used in ECG recording
Bipolar limb leads or Standard Leads or Einthoven lead system
asy
1. Lead I
2. Lead II
ii)
3. Lead III
En
Unipolar limb leads or Wilson Lead System
a) Augmented unipolar limb lead
1. aVR
gin
2. aVL
3. aVF
eer
b) Unipolar chest leads
i ng.
25. Mention the important bands of frequencies in EEG and their importance
Alpha waves (8-13)Hz – to monitor the level of consciousness
Beta waves (13-30)Hz – to monitor cerebral and mental activity
net
Theta waves (4-8)Hz – to analyse the emotional stress in adults
Delta waves (0.5-4)Hz – to study sleep disorders and brain tumours
28. Give the EMG signal characteristics. (Or) Mention the normal amplitude
and frequency of EMG signal
The EMG signal ranges from 0.1mV to 0.5mV. The frequency
components of the EMG signal vary from 20Hz to 10 KHz and they are
restricted to the frequency range of 20Hz to 200Hz for clinical purpose using a
low pass filter.
w.E
stimulus and response.
asy
The electrodes used for recording EMG are,
Surface electrode
En
Metal Disc electrode, Disposable electrode
Needle electrode
Unipolar and Bi polar electrode
gin
31. What is the purpose of electrode paste?
eer
i
The outer skin of the body is highly conductive and it will not establish
ng.
a good electrical contact with an electrode so, some of the outer cells of the
skin is removed and the electrode paste is applied to reduce the contact
impedance and to avoid the movement artifacts.
net
PART-B
1. Explain the sources of Biopotential in detail. (Ans: Unit 1 Page No. 2-3)
2. Explain the origin of biopotential. (Ans: Unit 1 Page No. 3-8)
3. Draw the action potential waveform and explain the following terms. Resting
potential; Action potential; Absolute refractory period and Relative refractory
period. (Ans: Unit 1-Page No. 6-8)
4. Draw equivalent circuit of a biopotential electrode interface (Ans: Unit 1 Page
No. 10).
5. Discuss about the different types of electrode used in bio potential
measurement. (Or) Discuss in detail about various types of biopotential
electrodes. (Ans: Unit 1 Page No. 12-21)
6. Draw a typical ECG waveform and mark the important features and the
associated function of the heart. (Ans: Unit 1 Page No. 28-30)
7. With neat diagrams explain the 12 lead system in ECG measurement. (Or)
Draw and explain the different lead configuration and its significances in ECG.
(Ans: Unit 1 Page No. 30-33)
8. Explain the working principle of a ECG machine with a neat block diagram.(
Ans: Unit 1 Page No. 33-35)
9. Give the origin of brain waves and describe the 10-20 electrode system used in
EEG. (Ans: Unit 1 Page No. 35-38)
10. With a neat block schematic diagram, describe the principles involved in a
EEG recorder. (Ans: Unit 1 Page No. 39-40)
11. Explain the measurement of EMG. (Ans: Unit 1 Page No. 41-43)
ww PART-C
1. What should be the characteristics of biopotential amplifier? Explain with proper
w.E
justification. (Ans: Unit 1 Page No. 22-23)
2. Design a suitable amplifier that can be used in the front end of an ECG
machine. Justify by specifying the features of the selected amplifier. (Ans: Unit 1
Page No. 22-23)
asy
3. With neat diagrams, explain the schematic diagram of EEG machine. Also, show
En
the recording method of unipolar and bipolar EEGs. (Ans: Unit 1 Page No. 41-
43).
gin
eer
i ng.
net
3.
ww
What is a calorimeter?
Colorimeter is used to measure the transmitted and absorbed light as it
w.E
passes through a sample. The basic principle behind the colorimeter is that
many chemical compounds in solution appear coloured with the saturation of
asy
the colour depending on the concentration of the compound. By analyzing the
transmitted light through the sample or emitted light by the sample, the
En
concentration of the substance can be determined.
4.
S.No Colorimeter gin
Differentiate colorimeter and spectrophotometer
Spectrophotometer
1
Colorimeter uses filter as
eer
spectrophotometer uses
2
wavelength selector
It measures transmittance i
monochromator as wavelength selector.
It measures absorbance
ng.
5. State Beer and Lamberts Law
net
Beer and Lamberts law states that the absorption of light transmitted
through a medium is directly proportional to the concentration and length of
the medium.
w.E
11. Mention the application of auto analyzer
Auto analyzer is used to measure dozens of fluid samples an hour for a
variety of markers such as cholesterol, phosphate levels and proteins.
asy
12. What is the use of Electromagnetic Blood flow meter?
En
Electromagnetic Blood flow meter is used for measuring the flow
through blood vessels within the body
gin
13. What are the various types of blood flow meters?
1. Electromagnetic blood flow meter
eer
2. Ultrasonic (Doppler shift) blood flow meter
3. NMR blood flow meter and
4. Laser Doppler blood flow meter
i ng.
14. Name any four physical principles based on which blood flow meters are
constructed
net
Blood flow meters are constructed based on the following principles
a. Electromagnetic blood flow meter
b. Ultrasonic (Doppler shift) blood flow meter
c. NMR blood flow meter and
d. Laser Doppler blood flow meter
• Fick’s method
• Dye dilution method
• Thermal dilution
• Impedance technique
17. Find the cardiac output of a patient whose heart rate is 80BPM and a
stroke volume of 80 millilitres per beat.
The cardiac output = heart rate x stroke volume
= 80 beats / min x 80 ml/beat
= 6.4 litres / minute
w.E
19. Calculate the stroke volume in millilitres if the cardiac output is 5.2
litres/minute and heart rate is 76 beats/minute
Q = 5.2 litres/minute;
asy
HR = 76 beats/minute
Stroke volume = 𝑸𝑸/𝑯𝑯𝑯𝑯 = (5.2 𝑥𝑥 1000)/76 = 68.42 ml
En
gin
20. What are plethysmographs and plethysmography
Plethysmography is the process used to measure the volume changes in
eer
any part of the body that result from the pulsations of blood occurring with
each heart beat. These measurements are useful in the diagnosis of arterial
ng.
Plethysmograph produces a waveform that is similar to the arterial
pressure waveform.
10
ww
26. What is spirometer?
spirometer is a device used to measure the respiratory volume
w.E
measurements. By using this device, lung volume and capacities can be
determined by measuring the amount of gas inspired or expired under a given
set of conditions or during a given time interval.
asy
27. Name any two methods of respiration rate measurement
En
The methods used to measure respiration rate are,
• Thermistor method
• Impedance pneumography gin
• CO 2 method of respiration rate measurement
eer
28. How is the respiration rate measured?
Respiration rate is measured by one of the method
• Thermistor method
i ng.
• Impedance pneumography
• CO 2 method of respiration rate measurement
net
29. Name the parts of sphygmomanometer
Sphygmomanometer consists of inflatable rubber bladder which is
known as cuff, rubber squeeze-ball pump and valve assembly.
w.E
finger tip will increase, will reduce the amount of light fall on LDR and during
relaxation the amount of light will increase. This change in resistance per
minute will be measured as pulse rate.
asy
35. How is the pulse rate measured?
En
The pulse rate is measured using one of the following methods:
• Electrical impedance method
• Strain gauge method
gin
• Photoelectric method
• Microphone method eer
36. Which transducer is used for measuring temperature?Why?
Thermister, High sensitivity.
i ng.
37. What is tachycardia and bradycardia?
Bradycardia: The heart beats slowly i.e., less than 60 beats per minute
net
Tachycardia: The heart beats fastly i.e., more than 100 beats per minute
38. List the functions of blood cells. (Or) Give the functions of Blood cell
Red Blood Cell (RBC): It is used for the transport of oxygen and carbon
dioxide.
White Blood Cell (WBC): It defenses against infections and foreign
substances.
40. What is the basic principle behind the working of laser based blood cell
counter?
The basic principle behind the laser based blood cell counter is that the
angle of scattered light intensity is different for different sized particles.
PART-B
1. Describe the measurement of pH and pO 2 in blood. (Ans: Unit 2 - Page No. 4 -
6)
2. Discuss the working principle of colorimeter with a neat block diagram. (Ans:
Unit 2 - Page No. 7)
3. Explain the working principle of spectrophotometer with a neat diagram. (Ans:
Unit 2 - Page No. 8)
4. Explain the Blood flow measurement using following technique. (i)
Electromagnetic principle (ii) Thermo dilution iii) ultrasonic principle. (Ans:
5. ww
Unit 2 - Page No. 9-13)
Explain the working principle of electromagnetic blood flow meter. What are its
6. w.E
advantages and disadvantages?(Ans: Unit 2 - Page No. 9-11)
Define cardiac output. Discuss various techniques to determine cardiac
output. (Ans: Unit 2 - Page No. 14-18)
7.
asy
Explain the Direct and Indirect method of Blood pressure measurement (Ans:
Unit 2 - Page No. 26-29)
8.
En
Explain the working of a blood cell counter with a neat block diagram.
gin
(Or)Explain the principle of operation of coulter counter. What is its
application? (Ans: Unit 2 - Page No. 36-39)
9.
20-21) eer
Explain CO 2 method of Respiration rate measurement (Ans: Unit 2 - Page No.
10.
11.
Page No. 21-22) i
Explain spirometer method of Respiration rate measurement (Ans: Unit 2 -
PART-C
1. What are the different types of ultrasonic blood flow meter? Explain each in
detail. (Ans: Unit 2 - Page No. 11-13)
2. Explain the working principle and calibration procedure followed in colorimeter
with neat diagram. (Ans: Unit 2 - Page No. 7), also Refer videos
3. Elucidate the detailed procedure to detect blood pressure using
Sphygmomanometer and ultrasonic method. (Ans: Unit 2 - Page No. 28),(
Ans: Unit 2 - Page No. 11-13)
4. Explain in detail about thermo dilution and dye dilution of cardiac output
measurement technique. (Ans: Unit 2 - Page No. 16-17)
13
ww i) R-wave Synchronized
3.
w.E
Which type of electrode is applied in the case of external stimulation and
what is the current range?
The paddle shaped electrodes are applied on the surface of the chest
asy
and the current range is 20-150mA.
4.
En
When is internal stimulation employed?
gin
Internal stimulation is employed in cases requiring long term pacing
because of permanent damage that prevents normal self triggering of heart.
5. eer
Which type of electrode is applied in the case of internal stimulation and
what is the current range?
i ng.
The electrodes in the form of fine wires of teflon coated stainless steel,
spoon like electrodes are used. The current range is 2 - 15 mA.
14
9. What is defibrillator?
w.E
fibrillation or atrial fibrillation
En
which this necessary synchronism is lost is known as fibrillation.
wwi) Large currents used in external defibrillations not only produces heart
muscle contraction but also produces burning of skin under the electrodes.
w.E
ii) Atrium fibrillation is introduced while arresting ventricular fibrillation
asy
charged to a potential of 5000V (dc)
Given: C = 16µF; V = 5000V
1 1
En
E = 2 CV2 = 2 x 16 x 10−6 x (5000)2 = 200 Joules.
gin
19. Classify the defibrillator based on applied voltage.
eer
Based on the nature of the voltage applied, the defibrillators can be
classified into 6 types.
• A.C Defibrillator
• D.C Defibrillator i ng.
• Synchronized D.C Defibrillator
• Square pulse Defibrillator
• Double square pulse Defibrillator
net
• Biphasic D.C Defibrillator
20. Draw the defibrillator output waveform and indicate the output energy
level. (Or) Draw the circuit of DC defibrillator and give its output
specifications.
16
ww
similar to the normal levels of electrolyte found in ECF Dialysate solution is
almost isotonic solution. Dialysate solution commonly contains 6 electrolytes.
w.E
they are, Sodium (Na+), Magnesium (Mg++), Chloride (Cl-), Potasium (K+),
Calcium (Ca++) and bicarbonate (HCO3-). the 7th component, non-electrolyte
glucose or dextrose is invariably present in the dialysis.
asy
24. What are the membrane used for Haemodialysis?
En
Synthetic and natural membranes are commonly used for filtration
applications. Membrane materials most often used include regenerated
gin
cellulose, cellulose acetate, polysulfone, polycarbonate, polyethylene,
polyolefin, polypropylene, and polyvinylidene fluoride.
17
ww 1
as artificial kidney. Blood is
removed from the body. Waste
patient acts as semi permeable
membrane. The dialysate solution
w.E
products diffuse through the
semi – permeable membrane.
They are continuously rinsed
passes through it to remove the waste
from the blood.
asy
with dialysate solution.
It is more efficiently used for It is comparatively less efficiently
2
waste removal.
En used.
3
Since blood is taken out from the
body, the process is little gin
It does not involve removal of blood,
The time taken for dialysis is 3 to The time taken for dialysis is 9 to 12
4
6 hours. hours.
i
Dialysis can be performed at Dialysis can be performed only in
ng.
5
6
home dialysis centres
Dialysis must be done every day Dialysis is done 3 times in a week
net
30. List the Advantages and disadvantages of AV fistula
The AV fistula is considered to be the best choice for vascular access.
Advantages:
1. An AV fistula uses your own arteries and veins without the need for artificial
material.
2. The risk of clotting or infection is significantly lower than with other forms of
vascular accesses.
3. AV fistula surgery is usually done on an outpatient basis, under local
anesthetic, allowing for a rapid and easy recovery.
4. AV fistulas last years longer than other forms of vascular access and can last
for decades.
5. AV fistulas are less expensive to maintain then AV Grafts or Venus catheters.
18
Disadvantages:
1. The AV fistula requires several weeks to months to mature before it can be
used.
2. AV fistulas may not be suitable for people with small or weak veins.
3. Strengthening the AV fistula requires daily exercises.
4. Some AV fistulas fail to mature and process must be repeated.
5. AV Fistulas are visible on the forearm.
ww Disadvantages:
1.
2.
3.
w.E
Use of synthetic material in the body.
With AV graft there is an increased risk of blood clotting, and infections.
AV graft tends to close more quickly than the fistula.
4.
5. asy
AV graft needs constant attention and upkeep.
AV graft does not last as long as a fistula and will probably need to be
replaced eventually.
En
32. What is ventilator?
gin
A ventilator is a machine that supports breathing
Disadvantages:
1. Ultrasonic imaging is expensive than X-ray
2. Bone blocks ultrasound
3. Artifact is produced.
19
38. What are the various modes of display in ultrasound imaging system?
The various modes of display are,
1. A-mode
2. B-mode
3. M-mode Or T-M mode
ww
39. Compare the various probes used in ultrasonic imaging system
The various probes are, Linear, Curvilinear, Phased array and
w.E
Endocavitary Probes
gin
4. Diagnose developmental joint abnormalities in children
5. Diagnose sports related injuries
6. Detect hidden tumour or infection in joint
eer
7. Detect bone cancer
8. Diagnose problems with the pituitary gland and brainstemi
9. Determine the condition of nerve tissue within the spinal cord ng.
10. Detect brain abnormalities like cysts, tumours, bleeding, swelling
11. Detect damage to the brain caused by an injury or a stroke net
41. List the components of MRI scanner
A large magnet, 3 Gradient coils, a RF coils, a patient table and a computer.
20
2. MRI gives clear and detailed images of Soft tissues structures of the body
such as the heart, liver, brain, ligaments and cartilage.
3. MRI can cover large portion of the body
4. MRI can provide information about blood movement inside blood vessels.
Disadvantages of MRI:
1. MRI is expensive
2. Patient must remain still in an enclosed machine, which may be a problem
for claustrophobic patients.
3. A metal implant in a patient's body may be affected by the strong magnet of
the MRI unit
4. MRI scan does not show Bone and calcium. so, disease such as osteoporosis
cannot be detected using MRI scanning.
ww
44. Give difference between closed and opened MRI .
S.No Closed MRI Open MRI
1
2
w.E
High Field (Typical 1.5 T or 3T)
High SNR
Low Field (Typical 0.2 T – 0.4T)
Low SNR
3
asy
High image quality Low image quality
4
5
Fast imaging
Advanced Applications En Slow imaging
Limited Applications
6 Increased patient anxiety gin Less patient anxiety
7 Claustrophobic patient problems
eer
Claustrophobic patient handling
8 High acoustic noise levels
i
Lower acoustic noise levels
ng.
1.
PART-B
How pacemakers are classified based on the modes of operation? Explain its
net
working principle. (Ans: Unit 3 - Page No.3-6)
2. With neat diagram, Describe the function of ventricular inhibited pacemaker.
(Ans: Unit 3 - Page No. 10-11)
3. What are the different types of DC defibrillators? Explain the principle of DC
Defibrillator with neat diagram (Ans: Unit 3 - Page No.16, 17-19 )
4. With a neat diagram, illustrate the working of DC defibrillator. (Ans: Unit 3 -
Page No. 20-21 )
5. With a neat diagram, illustrate the working of Biphasic DC defibrillator. (Ans:
Unit 3 - Page No. 24-25 )
6. Discuss how the image is constructed using ultrasound with neat diagrams.
(Ans: Unit 3 - Page No. 17-19 )
7. Draw a block diagram of MRI system and explain the image reconstruction
using it. (Ans: Unit 3 - Page No. 22-26 )
21
PART-C
1. With a neat block diagram explain the principle of operation of hemo dialyzer
machine. (Ans: Unit 3 - Page No. 8-9)
2. Explain the function of synchronized DC Defibrillator with neat block diagram.
(Ans: Unit 3 - Page No.20-21 )
3. Explain positive pressure ventilator with neat diagram. (Ans: Unit 3 - Page No.
14-15 )
ww
w.E
asy
En
gin
eer
i ng.
net
22
3.
w.E
What are the benefits of diathermy
Treating injuries with heat can increase blood flow and make connective
asy
tissue more flexible. It can also help minimize inflammation and reduce the
incidence of edema, or fluid retention.
En
By increasing blood flow to the site of an injury, the deep heat generated
with diathermy can accelerate healing.
4. gin
Compare shortwave, microwave and ultrasound diathermy.
S. eer
No
Shortwave diathermy
uses
Microwave diathermy
ng.
uses sound waves to heat
1 electromagnetic energy generate heat in the
to generate heat body
it can’t penetrate deep
deep tissues
net
muscles, it’s best suited
it can penetrate deep it can penetrate deep
2 for areas that are closer
muscles muscles
to the skin, such as the
shoulders
27.12 MHz at 11m 2450 MHz at 12.25cm 1 MHz or 3 MHz at 300m
3
wavelength wavelength or 100 wavelength
Using capacitive plate
Microwave is produced Ultrasound is produced
4 and cable
using Magnetron using Piezo electric cyrstal
electromagnetic energy
Maximum power Maximum power
5 3W/cm2
delivered is 500W delivered is 100W
Gel is applied to the
6 No gel is applied No gel is applied
affected area of the body
23
wwDisadvantages:
Though most short-wave diathermy machines have output power
w.E
control, there is no indication of the amount converted and absorbed heat
within the body tissues. Therefore intensity of treatment depends on the
subjective sensation of warmth felt by the patient.
7. asy
Give the applications of ultrasonic diathermy
En
It is used where shortwave treatment failed and in cases where
localizing of heat is required. It is very useful in curing of diseases of peripheral
gin
nervous system like neuritis, skeletal muscle system like arthritis and skin like
ulcers
24
10. What is the frequency of currents used in surgical diathermy units? Why?
The frequency of currents used in surgical diathermy units is in the
range of 1–3 MHz. This frequency is quite high in comparison with that of the
50 Hz mains supply. This is necessary to avoid the intense muscle activity and
the electrocution hazards which occur if lower frequencies are employed
ww
12. What is electrosurgical diathermy?
w.E
An Electrosurgical diathermy is a generator capable of producing a
cutting and/or coagulating clinical effect on tissue by the use of alternating
current at a high frequency (RF - radio frequency, also known as radio
surgery).
asy
En
13. what are the various modes of electrosurgery?
gin
The different modes of electrosurgery that can be performed using
electrosurgical unit are,
1. Cutting (Or) Electrotomy
2. Coagulation eer
a. Fulguration
b. Desiccation
3. Blending (or) Haemostasis
i ng.
14. Give difference between cutting and coagulation in diathermy.
net
S.No Cutting Coagulation
1 Constant waveform is used Pulsed waveform is used
Current flows for 100% of the
2 Current flows only for 6% of the cycle
cycle
3 Low voltage (1300 – 2300)V High voltage (3500 – 9000)V
4 Produces intense heat Produces less heat
Cutting electrodes like needle,
5 Ball, Bipolar electrode is used
wire loop electrode is used
16. Define Cutting and Coagulation (Or) What is cutting and Coagulation in
Electrosurgical diathermy?
Cutting:
When high frequency current flows through the sharp edge of a wire
loop or band loop or the point of a needle into the tissue, there is a high
concentration of current at this point. The tissue is heated to such an extent
that the cells which are immediately under the electrode, are torn apart by the
boiling of the cell fluid. The indifferent electrode establishes a large area
contact with the patient.
Coagulation:
Coagulation of tissue is caused by the high frequency current flowing
through the tissue and heating it locally so that it coagulates from inside. The
coagulation process is accompanied by a grayish-white discoloration of the
ww
tissue at the edge of the electrode.
There are two types of coagulation: 1) Desiccation (Or Pinpoint) and 2)
w.E
Fulguration (Or Spray).
17. Bring out the need for patient plate in surgical diathermy.
asy
The patient electrode or indifferent electrode establishes a large area
contact with the patient and the RF current is therefore, dispersed so that very
En
little heat is developed at this electrode.
eer
then kept steady. Depending upon the intensity and duration of the current, a
high local increase in heat will be obtained. The tissue changes due to drying
and limited coagulation.
i ng.
The concurrent use of continuous radio-frequency current for cutting
and a burst wave radiofrequency for coagulation is called Haemostasis mode.
26
ww
• Monitoring the health of astronauts in space.
• Patient monitoring in an ambulance and other locations away from hospital.
w.E
• Research on unanaesthetized animals.
asy
23. What are the various modulation systems used in biotelemetry?
a. Double modulation : either AM/AM , AM/FM, FM/FM, FM/AM - to avoid
loading effect.
En
b. Pulse width modulation.: More than one bio signal can be transmitted
and recorded
gin
eer
24. What is the modulation techniques used for biotelemetry? Mention the
reason for adopting that modulation scheme.
i
• Double modulation : either AM/AM , AM/FM, FM/FM, FM/AM - to avoid
loading effect. ng.
• Pulse width modulation.: More than one bio signal can be transmitted and
recorded
net
Loading effect is avoided in double modulation.
More than one bio signal can be transmitted and recorded using pulse width
modulation.
PART-B
1. Explain the simplified circuit diagram of a microwave diathermy machine. (Ans:
Unit 4 - Page No. 14-17)
2. Draw the block diagram of short wave diathermy unit and explain it. (Ans: Unit
4 - Page No. 4-7 )
3. Explain the basic principle of operation of an ultrasonic diathermy unit. List out
its applications. (Ans: Unit 4 - Page No. 12-14 )
4. Draw the block diagram of shortwave and microwave diathermy. Explain in
detail. (Ans: Unit 4 - Page No. 4-7, 14-17)
5. Discuss the working of a surgical diathermy unit with a neat block diagram
(Ans: Unit 4 - Page No. )
6. Explain the working of single channel ECG telemetry system. (Ans: Unit 4 -
Page No. )
7. Explain the working of a Multichannel Telemetry system with a neat block
ww
diagram. (Ans: Unit 4 - Page No. )
1.
w.E PART-C
With a neat block diagram, show the operation of a combined single channel
asy
telemetry system for ECG signal and respiration rate. (Ans: Unit 4 - Page
No.33-34 )
2.
En
Explain in detail the design requirements of an ECG telemetry receiver. Also,
gin
mention the role of IF amplifier in the receiver. use suitable illustration (Ans:
Unit 4 - Page No. 36-39 )
3.
eer
Draw the typical block diagram of electrosurgical unit and explain its
functioning. Mention the hazards that commonly occur in electrosurgical unit.
(Ans: Unit 4 - Page No. 17-26)
i ng.
net
28
3. ww
Define telemedicine
w.E
• It is the application of tele communications and computer technology to
deliver health care from one location to other.
asy
• It involves the use of modern information to deliver timely health service to
those in need by the electronics transmission.
4. En
Explain the principle of telemedicine
gin
Telemedicine is a rapidly developing application of clinical medicine
where medical information is transferred via telephone, the internet or other
eer
networks for the purpose of consulting and sometimes remote medical
procedures or examinations.
29
9. What is insulin?
Insulin is a hormone created by pancreas that controls the amount of
glucose in the bloodstream at any given moment. It also helps to store glucose
in liver, fat, and muscles. Finally, it regulates body’s metabolism of
carbohydrates, fats, and proteins.
ww
11. What causes someone to be prescribed insulin?
w.E
• If the body doesn’t make insulin or doesn’t make enough, then the person
is diagnosed with type 1 diabetes.
• On the other hand, if the body doesn’t use insulin properly, then the
asy
person is diagnosed with type 2 diabetes.
En
12. What are the different types of insulin?
insulin gin
Rapid-acting, Short-acting, Intermediate-acting, Long-acting, Premixed
30
16. What are the drawbacks to insulin treatment for diabetes? (Or) What are
the disadvantages of taking insulin?
The more common side effects that occur with insulin regular
(human) include:
• Swelling of your arms and legs.
• Weight gain.
• Low blood sugar (hypoglycemia). This needs to be treated. ...
• Injection site reactions. ...
• Skin changes at the injection site (lipodystrophy)
ww
glucose level trends over time, are visible on a built-in device screen
18.
w.E
Is insulin pump better than injections?
In the largest and longest study ever of an insulin pump with a
continuous glucose sensor, patients who used the device achieved better
asy
control of their blood sugar than patients taking insulin injections
19.
En
What are the side effects of insulin?
Insulin regular (human) side effects
• sweating.
• dizziness or lightheadedness.
gin
• shakiness.
eer
• hunger.
• fast heart rate.
• tingling in your hands, feet, lips, or tongue.
i ng.
• trouble concentrating or confusion.
• blurred vision. net
20. What is the difference between injecting insulin using a syringe vs an
insulin pump?
Injections are cheaper and take less training to use than insulin pumps.
A patient will have to test blood sugar levels before every injection. There is
the possibility a patient can develop resistant areas if injections are done too
frequently in the same spot.
ww
25. What are brain computer interfaces used for?
w.E
Brain-computer interfaces (BCIs) allow their users to communicate
or control external devices using brain signals rather than the brain's
normal output pathways of peripheral nerves and muscles.
asy
26. What is brain computer interface (BCI)?
En
Brain-computer interface (BCI) is a collaboration between a brain and
a device that enables signals from the brain to direct some external activity,
gin
such as control of a cursor or a prosthetic limb. The interface enables a direct
communications pathway between the brain and the object to be controlled.
eer
PART-B i ng.
1. Explain how telemedicine helps the patients and medical practitioners. (Ans:
Unit 5 - Page No. 1-10)
2. Construct and discuss the working of endomicroscopy unit in detail (Ans: Unit 5
- Page No. 26-34)
net
3. Explain how insulin pump works in detail with suitable diagram. (Ans: Unit 5 -
Page No. 10-18)
4. What is radio pill? Explain in detail. (Ans: Unit 5 - Page No. 19-25 )
5. Explain in detail about Brain machine interface (Ans: Unit 5 - Page No. 35-48 )
6. Explain how Lab on chip (LOC) integrates several laboratory functions on a
single integrated chip. (Ans: Unit 5 - Page No. 35-48)
PART-C
1. Explain about the evolution and technologies involved in telemedicine and
discuss the application areas of telemedicine. (Ans: Unit 5 - Page No. 1-10 )