Cal Lab: The International Journal of Metrology 18 Apr • May • Jun 2012
METROLOGY 101
Figure 1. Small shorting devices are shown here. Specially made circuit boards, supplied with Fluke’s 8508A DMM, provide both
ohmic and thermal zeros for quick stabilization.
Apr • May • Jun 2012 19 Cal Lab: The International Journal of Metrology
METROLOGY 101
resistance ‘zero.’ You see the measurement on the DMM Some manufacturers and experienced measurement
shows a number much different than the PCB. This is due technicians find that simple copper wire is very satisfactory.
to the fact the material on the banana plugs is comprised of This is often called bell wire (a simple wire that has
nickel. It also is at a different temperature. This will create a historically been used for telephone connections). When
thermal EMF that is often on the order of hundreds of counts. a small solid copper wire is connected to the terminals,
you will see this measured zero is very consistent with the
performance of the shorting PCB.
Cal Lab: The International Journal of Metrology 20 Apr • May • Jun 2012
METROLOGY 101
Summary
Thermal EMFs can add unwanted
errors to sensitive voltage and
resistance measurements. Proper
application practices can be used to
minimize these errors:
• Use low thermal EMF cables
& connections.
• Use proper zero techniques
on measurement instruments.
• Use reversal techniques
whenever possible.
• Take advantage of EMF
limiting techniques found in
precision instrumentation.
Diagram 2.
References
Another best measurement practice has an internal precision current
to minimize thermal EMFs and various source which is used to stimulate the [1] Jack Somppi, “How to Avoid Sur-
offsets is to utilize instrument specific resistance being measured and uses
prising Errors from Thermal EMFs,”
EMF reduction techniques. When its voltmeter capability to measure
measuring resistance with a high the voltage drop across the test Fluke Calibration How To Seminar.
sensitivity, or at a low value, the four- resistor. In two wire connection
wire ohms function is a technique that mode, the additional resistance of the
eliminates the unwanted effect of lead test leads can cause an error in the
resistance and causing measurement measurement. Making the measuring Martin L. Kidd, Fluke Corporation,
error. Considering how a resistance connection separate from the current Everett, WA, martin.kidd@flukecal.
measurement is done, a DMM usually source connection eliminates the lead com.