Professional Documents
Culture Documents
GIW WI-HQ1-026 - ASTM D5885 HPOIT Testing
GIW WI-HQ1-026 - ASTM D5885 HPOIT Testing
026-2
WORK INSTRUCTIONS PAGE: 1 of 9
TITLE: HIGH-PRESSURE OXIDATIVE-INDUCTION TIME DATE: 01/05
AUTHOR: MM
REVISED BY:
1. PURPOSE
2. SCOPE
3. RELATED DOCUMENTS
3.1. ASTM D5885-97, Standard Test Method for Oxidative Induction Time of
Polyolefin Geosynthetics by High-Pressure Differential Scanning Calorimetry.
4. DEFINITIONS
4.2. Oxidative-Induction Time (OIT) – the time elapsed between the introduction
of a material to an oxidizing gas and the oxidation of the material under
isothermal conditions.
UNCONTROLLED IF PRINTED
Geosynthetics Industrial Works DOCUMENT NO:WI-HQ1-
026-2
WORK INSTRUCTIONS PAGE: 2 of 9
TITLE: HIGH-PRESSURE OXIDATIVE-INDUCTION TIME DATE: 01/05
6.1.2. DO NOT open the DSC Cell when the temperature is over 40ºC.
6.1.3. NEVER touch the DSC Cell with your fingers or forceps. It is
extremely sensitive and it will scratch or break.
6.1.4. The DSC should be calibrated every month. See WI-Q1-058 for
Indium calibration instructions.
6.2.1.2. Write the roll number, lot number and masterbatch number
on the back of the sample using a correction pen.
6.2.2.2. Remove the heat shield and place it gently next to the cell.
6.2.2.3. Remove the cell lid and place it gently next to the cell.
UNCONTROLLED IF PRINTED
Geosynthetics Industrial Works DOCUMENT NO:WI-HQ1-
026-2
WORK INSTRUCTIONS PAGE: 3 of 9
TITLE: HIGH-PRESSURE OXIDATIVE-INDUCTION TIME DATE: 01/05
6.2.2.5. Leave the empty DSC aluminum pan in the back of the cell
for reference. It should be centered on the back circle of the
cell.
6.2.2.6. Place a new DSC aluminum pan in the front of the cell. It
should be centered on the front circle of the cell.
6.2.2.7. Place the specimen into the front DSC aluminum pan. The
specimen should be centered in the pan.
6.2.2.9. Once all the internal lids have been replaced put the top
cover on the DSC Cell.
6.2.2.12. Hand tighten the 3 screws the rest of the way. The heads of
the screws should be approximately 1 thread above the cell.
6.2.3.2. Close the Nitrogen and Oxygen valves on the wall next to the
DSC completely. Turn the 2 black knobs on the gas lines in
a clockwise manner until hand tight.
6.2.3.6. Close the pressure release valve to the left of the cell until it
is hand tight.
6.2.3.7. On the front of the cell there are 2 knobs: an IN valve and an
OUT valve. Close both of these valves until they are hand
tight (clockwise).
UNCONTROLLED IF PRINTED
Geosynthetics Industrial Works DOCUMENT NO:WI-HQ1-
026-2
WORK INSTRUCTIONS PAGE: 4 of 9
TITLE: HIGH-PRESSURE OXIDATIVE-INDUCTION TIME DATE: 01/05
6.2.4.1. If the screen is black, simply move the mouse around and
the screen should appear.
6.2.4.2.5. On the screen you will see icons on the left side
and a long rectangular box on the right. In this
rectangle
you will see a box marked with TA with a bar
that has an arrow. Click on this bar
6.2.4.5. Click on the space that says SAMPLE SIZE Enter the
weight in mg.
UNCONTROLLED IF PRINTED
Geosynthetics Industrial Works DOCUMENT NO:WI-HQ1-
026-2
WORK INSTRUCTIONS PAGE: 5 of 9
TITLE: HIGH-PRESSURE OXIDATIVE-INDUCTION TIME DATE: 01/05
6.2.4.9. Clear the purge gases. (Remove the X’s from the boxes by
clicking on them).
6.2.5.2. Very SLOWLY open the IN valve on the front of the DSC.
6.2.5.3. Watch the gage on the front of the DSC. When it reads 200
psi, close the IN valve hand tight.
6.2.5.5. When there is just a little bit of gas left in the cell, close the
pressure release valve hand tight.
6.2.5.6. Again, SLOWLY open the IN valve on the front of the DSC.
6.2.5.7. Again, watch the gage on the front of the DSC. When the
gage reads 200 psi, close the IN valve hand tight.
6.2.5.9. When there is just a little bit of gas still in the cell, close the
pressure release valve hand tight.
6.2.6.1. SLOWLY open the IN valve on the front of the DSC. Watch
the computer screen. When the pressure reaches 3500 Kpa
to 3800 Kpa, close the IN valve.
UNCONTROLLED IF PRINTED
Geosynthetics Industrial Works DOCUMENT NO:WI-HQ1-
026-2
WORK INSTRUCTIONS PAGE: 6 of 9
TITLE: HIGH-PRESSURE OXIDATIVE-INDUCTION TIME DATE: 01/05
6.2.6.2. Begin the test by moving the mouse to the green arrow and
clicking. The green arrow is located in the upper left-hand
corner of the screen. It is the first button.
6.2.6.3. Close the Oxygen line on the wall by turning the black arrow
knob to point up.
6.2.6.4.2. Start the test as usual and when the method gets
to the isothermal segment (approx. 10 min. after
starting), you can lengthen the test time.
6.2.6.6. To stop the test, move the mouse to the red square in the
upper left-hand corner. This is next to your green arrow that
you used to start the test.
6.2.6.7. A message will appear on the screen and ask if you would
like to terminate the test. Click YES, to continue and
terminate the test.
UNCONTROLLED IF PRINTED
Geosynthetics Industrial Works DOCUMENT NO:WI-HQ1-
026-2
WORK INSTRUCTIONS PAGE: 7 of 9
TITLE: HIGH-PRESSURE OXIDATIVE-INDUCTION TIME DATE: 01/05
6.2.6.8. Open the pressure release valve SLOWLY, allowing the gas
to GENTLY escape.
6.2.7.4. From the menu on the bottom of the screen, select FILE.
6.2.7.5. Click on the next empty row. If all rows are full, click on row
1 to reset the table.
6.2.7.15. Look at the graph. You may want to adjust the area of the
graph, removing the area after the “hump”.
UNCONTROLLED IF PRINTED
Geosynthetics Industrial Works DOCUMENT NO:WI-HQ1-
026-2
WORK INSTRUCTIONS PAGE: 8 of 9
TITLE: HIGH-PRESSURE OXIDATIVE-INDUCTION TIME DATE: 01/05
like this:
6.2.11.1. With the mouse, select points (1) and (2) on the curve.
Move the mouse to the desired spot and click. These should
be on the flattest points of the horizontal portion of the curve.
6.2.11.2. With the mouse, select points (3) and (4) on the curve.
Move the mouse to the desired spot and click. These should
be on the flattest points of the vertical portion of the curve.
UNCONTROLLED IF PRINTED
Geosynthetics Industrial Works DOCUMENT NO:WI-HQ1-
026-2
WORK INSTRUCTIONS PAGE: 9 of 9
TITLE: HIGH-PRESSURE OXIDATIVE-INDUCTION TIME DATE: 01/05
6.2.12.8. With the mouse, move the cross arrows to the upper left
quadrant of the screen. This will determine where the label
will be on the graph.
6.2.13.4. Once the graph is finished printing, one copy goes into the
Formulation file and one goes into the Formulation Data
Binder.
6.2.13.5. Percent OIT are calculated by dividing the aged sample OIT
by the unaged sample OIT.
UNCONTROLLED IF PRINTED