Professional Documents
Culture Documents
SEP-21S Plus Viltechmeda BS023004EBUKL Service
SEP-21S Plus Viltechmeda BS023004EBUKL Service
DOUBLE SYRINGE
INFUSION PUMP
SERVICE MANUAL
BS023004E-P01
CONTENTS
1. INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2. TECHNICAL DATA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
7
3. TECHNICAL D ESCRI PTI O N . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.1 Electrical schematic diagram of the pump . . . . . . . . . . . . . . . . . . . . . . . . 7
3.2 Electrical schematic diagram of electronic board B7030 . . . . . . . . . . . 8
3.3 Electrical schematic diagram of control unit . . . . . . . . . . . . . . . . . . . . . 11
4. PARAMETER CHECKI NG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.1 Means required for parameter checking . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.2 Checking infusion volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.3 Checking infusion rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.4 Checking bolus volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.5 Checking occlusion pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
5. CHECKING ALARM SI G NAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
6. TROUBLESHOO TI NG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
26
7. PUMP REPAIR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7.1 Disassembling the pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
7.2 Replacing the internal batteries GB-A, GB-B . . . . . . . . . . . . . . . . 26
7.3 Replacing the rubber bellows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
7.4 Replacing the optical sensors TAX-A,TAX-B . . . . . . . . . . . . . . . . . . . . 27
7.5 Replacing the end switch S1-A, S1-B . . . . . . . . . . . . . . . . . . . . . . 27
7.6 Replacing the syringe sensor S2-A, S2-B . . . . . . . . . . . . . . . . . . . . 27
7.7 Replacing the pre-alarm switch S3-A, S3-B . . . . . . . . . . . . . . . . . . . . . . 28
7.8 Replacing elements the electronic board B7030 . . . . . . . . . . . . . . . . . . 28
7.9 Replacing the membrane keyboard K1 . . . . . . . . . . . . . . . . . . . . . 28
7.10 Replacing microprocessor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7.11 Replacing buzzer unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7.12 Replacing power supply inlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7.13 Replacing battery fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
2
SEP-21S Plus double syringe infusion pump
ANNEX G
50 3
SEP-21S Plus double syringe infusion pump Service manual
1. INTRODUCTION ANNEX F
NOTE:
Service Alarm signal Comment
The technical servicing of the pump may be carried Manual
out by the manufacturer’s authorised personnel only! section
5(1) OCCLUSION!!!
5(2) OCCLUSION OR END
5(3) NO MAINS!!!
5(4) LOW BATTERY!
5(5) VERY LOW BATTERY
5(6) 5 min. PREALARM!
5(7) END of INFUSION!
5(8) KOR X.X ml/h
5(9) Display backlighting blinks
Pump owner/ward . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Tests performed by (name, surname, signature) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Tests performed (date) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4 49
SEP-21S Plus double syringe infusion pump Service manual
ANNEX E 2. TECHNICAL DATA
6 47
SEP-21S Plus double syringe infusion pump Service manual
ANNEX D 3. TECHNICAL DESCRIPTION
u 3.1.2. Keyboard
The keyboard contains the two groups of switches which
are identical and alarm switch off key which is common for
each channel.
Each channel of the keyboard is composed of 8 keys in form
of a matrix having 8 inputs and 1 output. Scanning of the
keyboard and reading of information from it are performed
by electronic board B7030. The keys ON/OFF control the
circuits of electronic power supply switches located on the
electronic board B7030.
u 3.1.3. Drives
The stepper motor SM-A (SM-B) rotates the lead screw with
a nut on it converting rotating movement of the motor to
linear movement of the tube. The tube supported by four
guiding bearings is connected to the syringe pusher. Marginal
positions of the tube with syringe pusher are limited by:
w limit switch S1-A (S1-B) at the right extreme position
which is actuated by the end of the tube;
w The pre-alarm switch S3-A, S3-B is activated when the
distance between syringe finger flange and syringe thumb
rest is equal to 45 mm. When the pre-alarm switch has
been activated, the pump calculates distance to the end
46 7
SEP-21S Plus double syringe infusion pump Service manual
of appropriate syringe. Based on this there are calculated
and activated the following warning messages:
Xmin. PREALARM!
where X ≤ 5 min.
and
SYRINGE EMPTY!
10 43
SEP-21S Plus double syringe infusion pump Service manual
ANNEX C 3.3. ELECTRICAL SCHEMATIC DIAGRAM
OF CONTROL UNIT
ASSEMBLY DRAWING OF ELECTRONIC The electrical schematic diagram of control unit is shown
BOARD B7030 in the annex D.
It is composed of the following main parts:
w display circuits;
w stepper motor control circuits.
42 11
SEP-21S Plus double syringe infusion pump Service manual
4. PARAMETER CHECKING
NOTE:
Programming of infusion parameters, filling of the
syringe and the extension tube with drug solution,
inserting the syringe into the pump and starting
infusion are described in the OPERATING IN-
STRUCTIONS of the pump.
OCCLUSION!!!
OCCLUSION!!!
the pressure gauge shall read the pressure 120 kPa ± 20%.
NOTE:
If measured values of pump parameters are out of
ranges permitted, contact the Manufacturer.
38 15
SEP-21S Plus double syringe infusion pump Service manual
5. CHECKING ALARM SIGNALS
12. OCCLUSION or END The syringe pusher stopped before the very
end of infusion.
16. X min. PREALARM! Not more then 5 min. left to the end of
infusion.
18. KOR X.X ml/h 1. Infusion was stopped and not resumed in
2 minutes.
16 37
SEP-21S Plus double syringe infusion pump Service manual
ANNEX B All alarm signals are accompanied by the audible signal.
Alarm signals listed in the Table are checked in following
ways:
ELECTRONIC BOARD B7030
SCHEMATIC DIAGRAM OF
OCCLUSION or END
* In case the pump was used for more than 2 years, this time
may be shorter.
18 35
SEP-21S Plus double syringe infusion pump Service manual
ANNEX A u 5. Checking of this alarm signal shall be carried out imme-
diately after the “Low Battery” check. Continue procedure
described in the step 4 and measure the time interval be-
SCHEMATIC DIAGRAM OF SEP-21S PLUS
DOUBLE SYRINGE INFUSION PUMP
LOW BATTERY
and
5 min PREALARM
END of Infusion
KOR XX ml/h
and
END of Infusion
Press the key PROG and restart the new infusion with pa- PROG
rameters set in step 6. Press the RUN/STOP key. 2 minutes
later there shall be heard the audible signal and there shall RUN/STOP
be alternatively displayed the messages
KOR XX ml/h
and
STOP XX.X
SYRINGE EMPTY!
Fig. 2
1. Cover
2. Screws M3x8
3. Plate 145
4. Screws M3x16
5. Rubber bellows
6. Tube
7. Support
8. Housing
9. Plate 146
10. Syringe clamp
32 21
SEP-21S Plus double syringe infusion pump Service manual
6. TROUBLESHOOTING
All failures are listed in the Table:
1 Message in the display 1. External fuses are blown. 1. Check the external fuses
„NO MAINS!!!” 2. Faulty mains cord. and replace with identical
although the pump is ones, if required. In case
connected to the mains they are blown again, con-
(the green indicator sult the Manufacturer.
MAINS isn’t glowing). 2. Check the mains cord
and replace it, if necessary.
2 Short cordless work 1. Fuse F1, F2 is blown in 1. Open the pump as de-
time or no cordless the electronic board B7030. scribed in section 7.1.
work at all. 2. Decreased capacity of the Check and replace the fuse
battery. with identical one accord-
ing to section 7.12. In case
it is blown again, consult
the Manufacturer.
2. Connect the pump to the
mains for 24 hours in or-
der to charge the internal
battery. Disassemble the
pump as pointed out in sec-
tion 7.1. Connect the resis-
tor of 12Ω/15 W to the bat-
tery and measure discharge Fig. 1
time to the moment the
battery voltage drops to 8V. 1. Screw M4x60
In case this time is shorter 2. Battery
than 1 hour (for 1300 mAh
capacity battery), replace 3. Buzzer
the battery as pointed out 4. Earthing screws
in section 7.2. 5. Self-tapping screws
6. Plastic cap
7. Cross knob
3 Rubber bellows Rubber ageing. Impermis- Replace the rubber bellows
8. Screw M4x60 with nut M4
cracked. sible chemical substances as described in sections 7.1
were used for disinfecting and 7.3. 9. Power supply
and cleaning. 10. Screws M3x10 with nuts M3
22 31
SEP-21S Plus double syringe infusion pump Service manual
7.13. REPLACING BATTERY FUSES No SYMPTOM CAUSE CORRECTIVE ACTIONS
Disassemble the pump according to the section 7.1. Check
the fuses F1, F2 in the electronic board B7030 and replace 4 Without syringe, the 1. Faulty optical sensor 1. Disassemble the pump as
with identical ones. syringe pusher doesn’t (Fig. 3 pos. 1) or dirty described in section 7.1.
Reassemble the pump in reverse order to the section 7.1. move when pressing raster in the coupling Clean the coupling raster
, keys. (Fig. 3 pos. 3). with a piece of soft cloth.
When starting Switch on the pump press-
infusion without the ing the ON/OFF key.
NOTE: syringe (the cylinder By means of an oscilloscope
of 20 mm x 20 mm is observe the signal at pin 2
1. Make sure before closing the pump, all the con- of the chip U3A (U17A)
inserted under the
nectors are reconnected and the earthing wires is re- syringe sensor), there is (rotate the coupling slowly
connected by means of its screw. the message by hand) in the electronic
„OCCLUSION!!!” board B7030.The low level
2. All screws, bolts and nuts which were covered with is displayed. of the signal shall be not
enamel paint before disassembling, shall again be cov- more than 1 V and the high
ered with enamel paint after reassembling. level — not less than 3.8 V
during the full revolution of
the motor. Otherwise, re-
place the optical sensor (sec-
tion 7.4).
2. Broken connecting wires 2. Disconnect the connec-
to the stepper motor. tor P1 (P2) of the step mo-
tor from the control unit
and check the wires to the
step motor for continuity
according to the electrical
schematic diagram by
means of an ohmmeter. If
the wires are broken, re-sol-
der them. Re-connect the
motor connector to the
control unit.
5 In the syringe inser- Faulty end switch Check the end switch and
tion mode, the syringe (Fig. 3 pos. 4). replace, if required, as de-
pusher stops with scribed in the sections 7.1
noise when it reaches and 7.5.
its right hand side
position.
30 23
SEP-21S Plus double syringe infusion pump Service manual
No SYMPTOM CAUSE CORRECTIVE ACTIONS
pressing the keyboard with a piece of soft cloth. Reconnect
keyboard connectors to the electronic board B7030. Reas-
semble the pump in reverse order to the section 7.1.
6 The pump doesn’t dis- Faulty pre-alarm switch Check the pre-alarm switch
play the messages (Fig. 3 pos. 6). and replace, if required, as
„X min. PREALARM!” described in the sections 7.10. REPLACING MICROPROCESSOR
and 7.1 and 7.7.
„SYRINGE EMPTY!” Disassemble the pump according to the section 7.1. Note
although the infusion the orientation of the key on the body of microprocessor
with the volume greater U4(U18) on the electronic board B7030. Lift carefully the
than that of the syringe
and the syringe is filled microprocessor U4(U18) from its socket on the electronic
with more than 50% of board B7030.
its nominal volume. Insert the microprocessor into the socket on the electronic
board B7030 so as to have the same orientation of the key
of its body as it was on the old microprocessor. Reassemble
7 In insertion mode, when Faulty syringe sensor (Fig. Check the syringe sensor
inserting the syringe 3 pos. 5) and replace, if required, ac- the pump in reverse order to the section 7.1.
and depressing the YES cording to the paragraphs
key, the message 7.1 and 7.6
„INSERT SYRINGE” 7.11. REPLACING BUZZER UNIT
isn’t replaced with the (Fig. 1)
message „READY to
run” Disassemble the pump according to the section 7.1. Dis-
connect the buzzer connector P5 from the connector JP1
on the electronic board B7030. Detach the buzzer (3) from
8 One or several keys Faulty keys. Check and replace the key-
fail to operate. board, if required, as de- the housing. Glue the new buzzer to the housing in that
scribed in sections 7.1 and same location. Reconnect buzzer connector P5 to the con-
7.8. nector JP1 on the electronic board B7030.
Reassemble the pump in reverse order to the section 7.1.
24 29
SEP-21S Plus double syringe infusion pump Service manual
7.7. REPLACING THE PRE-ALARM SWITCH No SYMPTOM CAUSE CORRECTIVE ACTIONS
S3-A, S3-B (Fig. 3)
Disassemble the pump according to the section 7.1. Unsolder 11 The pump fails to Faulty electronic board Replace the electronic board
two conductors from the pre-alarm switch (6). Replace the operate when switched B7030. B7030 as described in the
pre-alarm switch (6) and re-solder the conductors to the on with ON/OFF key. sections 7.1 and 7.7.
same terminals. Pull insulation tubes onto the soldering No message or the
places. Fasten the pre-alarm switch (6) in such a position as message
„INTERNAL ERROR”
to be activated when the distance between the housing sur- is displayed. Audible
face supporting finger flanges of the syringe and the syringe alarm is activated. No
pusher surface supporting thumb rest of the syringe is equal key operates except the
to 45 mm. Reassemble the pump in reverse order to the key ON/OFF.
section 7.1.
12 Indicator MAINS Fuse F3 blown in the elec- Open the pump as described
7.8. REPLACING THE ELECTRONIC BOARD glows, but the pump tronic board B7030 (Annex in section 7.1. Check the
fails to operate. C). fuse and replace it with
B7030 (Fig. 3) identical one, if required. In
Disassemble the pump according to the section 7.1. Dis- case it is blown again, con-
sult the Manufacturer.
connect the control unit connectors JP8—JP10, the buzzer
connector P5, battery connector P11, power supply con-
nectors P6, P7 and keyboard connectors from the elec- 13 When fixing the pump The plastic cap (Fig. 1 Remove the cross knob of
tronic board B7030. Remove six M3 screws (put several to a stand, the stand pos. 6) of the cross knob the mounting clamp rotat-
drops of acetone onto nuts to soften enamel paint) and surface is damaged is cracked. ing it unti-clockwize until
remove carefully the electronic board B7030. Install the the cap falls down. Rotate
the cross knob clockwize
new electronic board B7030. Reconnect all connectors in with the cylinder of 20mm
reverse order. Reinstall six screws. Cover the nuts with x 20mm inserted and press-
enamel paint. Reassemble the pump in reverse order to ing the new cap between the
the section 7.1. cross knob screw and the
cylinder until the cap seats
on the screw.
7.9. REPLACING THE MEMBRANE
KEYBOARD K1
14 Burned and cracked The power supply inlet was Disassemble the pump as
Disassemble the pump according to the section 7.1. Dis- power supply inlet flushed with liquid and af- described in the section 7.1.
ter connection to the mains Replace the power supply
connect the keyboard’s connectors from the electronic board 220-230 VAC there has oc- inlet as described in the sec-
B7030. Detach the faulty keyboard from the housing. Glue curred short connection. tion 7.11.
the new keyboard taking openings for the display and for
LEDs MAIN as a reference. Glue from the left to the right
28 25
SEP-21S Plus double syringe infusion pump Service manual
7. PUMP REPAIR the support and the housing (8). Fasten the screws (4).
Replace and fasten the plate (3) to the end of the tube (6)
with screws (2) clamping the end of the rubber bellows (5).
7.1. DISASSEMBLING THE PUMP Glue the cover (1) to the plate (3).
(Fig. 1) Reassemble the pump in reverse order to the section 7.1.
The syringe pusher shall be moved to its end right hand
side extreme position. 7.4. REPLACING THE OPTICAL SENSOR
Remove four self-tapping screws (5) located at the corners TAX-A, TAX-B (Fig. 3)
of the pump bottom. Carefully lift the pump housing (note Disassemble the pump according to the section 7.1. Remove
the number of washers at the corners) and disconnect the the screw (2) and the remove the optical sensor (1). Note
power connectors P11 from the element block B7030. the order of conductor connection and unsolder them.
Resolder the conductors to the new optical sensor (1) in the
7.2. REPLACING THE INTERNAL BATTERIES same order. Pull insulation tubes onto the soldering places.
GB-A, GB-B (Fig. 1) Spread an universal glue onto the optical sensor (1). Fasten
the optical sensor (1) by means of the screw (2) so as to
Disassemble the pump according to the section 7.1. Remove
maintain the distance between the optical sensor (1) and
the screw (1) and the nut (8) fastening the batteries (2).
the coupling (3) equal to 4 ± 0.1 mm. Reassemble the pump
Unsolder four conductors and note their polarity.
in reverse order to the section 7.1.
The new batteries of that same type shall be connected ex-
actly in that same manner. Replace the screw (1) and the
nut (8). Reassemble the pump in reverse order to the sec- 7.5. REPLACING THE END SWITCH S1-A, S1-B
tion 7.1. (Fig. 3)
Disassemble the pump according to the section 7.1. Remove
7.3. REPLACING THE RUBBER BELLOWS the end switch (4) and unsolder two wires. Replace the end
(Fig. 2) switch (4) and resolder the wires to the same terminals. Pull
insulation tubings onto the soldering places. Place the end
Disassemble the pump according to the section 7.1. switch (4) and reassemble the pump in reverse order to the
Using the sharp screwdriver or the knife, remove the cover section 7.1.
(1) (insert carefully the tool between the cover and the metal
part). Remove three screws (2) fastening the plate (3) to the
7.6. REPLACING THE SYRINGE SENSOR
tube (6) and remove the plate.
S2-A, S2-B (Fig. 3)
Remove the screws (4). Remove the rubber bellows (5) and
the support (7) from the tube (6). Disassemble the pump according to the section 7.1. Unsolder
The new rubber bellows (5) is inserted with its narrower end two wires from the syringe sensor (5). Replace the syringe
towards the notch on the tube (6). Spread the silicon sealer sensor (5) and resolder the wires to the same terminals. Pull
on the housing-side of the support (7) and install it in such a insulation tubes onto the soldering places. Reassemble the
way as to clamp the end of the rubber bellows (5) between pump in reverse order to the section 7.1.
26 27
SEP-21S Plus double syringe infusion pump Service manual