【 How to use the oxygen generator 】 6.

6. Press the timing button , each time, the timing time starts to accumulate timing with
10 minutes as the minimum unit; when the timing time exceeds 60 minutes, the timing time 【 WORKING PRINCIPLE 】 【 PACKING LIST 】
starts to accumulate timing with 1 hour as the minimum unit (Figure 5).
Table7
7. When the timing time is up or press the power off button, the oxygen concentrator will
Compressor
enter the power off mode. The screen will continue to light for 15 seconds, and the service No. Name Units Quantity
Scavenging
time will be displayed at the same time. 1 Oxygen concentrator host set 1
valve
SL-3C Series Oxygen 8. The machine shuts down, the machine automatically enters the standby state, and the power 2 Oxygen inhalator tube piece 2
indicator continues to light; Molecular 3 Manual instruction piece 1
Concentrator User Manual 9. If the external power supply is cut off during work, the machine will cut off and alarm
sieve tank 4 Air inlet filter cotton piece 1
5 Nebulizer mask piece 1
10、If the machine fails or the following code is displayed, please contact our after-sale service:
6 Oxygen tube piece 2
Reducing 7 humidification bottle piece 2
1. Put half a cup of pure water in the humidification bottle (the liquid level should be between 1). "E2", the compressor open circuit fault; valve

the highest and lowest water levels) and screw the lid tightly to prevent air leakage, otherwise 2). "E3", the compressor short circuit fault;
the oxygen flow at the outlet will become small or no oxygen;(See Figure 2) 3). "E7", a high-temperature fault inside the machine. Figure 18
2. Insert the oxygen suction tube into the oxygen outlet (see Figure 3); When flows to A, the molecular sieve will absorb nitrogen (N ) of the air while the oxygen
2

3. Connect the power plug, turn on the oxygen generator "power switch" (Figure 4), "O" will flow through A, then high oxygen concentration will output for the patient under the
control of the reducing valve. When bottle A is full of nitrogen (N2 ), the switch valve will
means off, "I" means on, the machine into standby state, the display screen shows the
compress the air to B, at the same time, bottle A will release nitrogen (N ).Bottle B absorbs
2

cumulative running time; the nitrogen (N ) while output the oxygen to the oxygen pot.
2

Repeat above operation, the oxygen concentrator outputs the oxygen continuously .
4. Press the local switch button , the machine starts to run and produce oxygen; Oxygen needs to be filtered, flow control, humidified and compressed before use to provide
5. Rotate the oxygen flow adjusting knob (see Figure 1) : 5L model to 5L, 8L model to 8L, to the patient with safe and comfortable using experience. (See figure18 )
10L model to 10L (take the center position of the black ball as the standard)
Figure 8 Figure 9
Figure 6 Figure 7
【 MEDICAL OXYGEN CONCENTRATOR FLOW CHART 】
Note: When the output flow rate is greater than the rated flow rate of the concentrator,
the oxygen concentration will decrease until an alarm is given. Air flow into the filter compressor condenser switch valve

oxygen pot reducing valve flowmeter humidifybottle oxygen out put molecular sieve

Figure 19

Alarm Performance and working status


Figure 10 Figure 11 Figure 12
Figure 5

6 7
14 15

11. How to use the Nebulizer When the compressor of the oxygen concentrator stops running during operation, error code
1) Take out the nebulizer mask from the box of the oxygen concentrator. Open the nebulizer "E2" will be displayed on the screen of the oxygen concentrator, the red light will flash and
mask (see Figure 6); system will make a sound to remind the user.
【SL-3C Series Models】
2) PPlace the mask on the table, install the nebulizer filter (Figure 7), pour the drug into the
nebulizer mask under the advice of the doctor, please do not exceed the limit line. 3、 Internal high temperature fault prompt of oxygen generator (example figure 12)
Basic performance Table1
Catalogue 3)Tighten the top and bottom of the nebulizer mask (see Figure 8); When the internal temperature of the oxygen generator box lasts for 1 minute above 50℃,
Series No. Model Oxygen Nebulizer Scr een 4) Put the nebulizer mask on the nebulizer lid (see Figure 9); the screen of the oxygen generator displays "E7" oxygen generator prompt light flashing
1. Important Tips-------------------------------------------------------------------1 5) Connect the bottom of the nebulizer mask to the medical tube(see Figure 10); red light and sending out sound to remind.
2. Caution---------------------------------------------------------------------------2 SL-3C-510 LED
6) Loosen the plug counterclockwise, the atomization appears;
3. Why your physician prescribed oxygen supplement ----------------------3 2 SL-3C-550 LED 7) Insert the other end of the medical tube into the outlet of the oxygen concentrator
4. Working principle of SL-3C Oxygen Concentrator-----------------------3
(see Figure 11);
5. The important components of the Oxygen Concentrator -----------------4
6. Precautions for use---------------------------------------------------------4 Basic specification Table2
【 Maintenance 】
7. How to use the oxygen generator---------------------------------------------6 Oxygen Specification of nebulizer Table5
Oxygen Oxygen Caution: Do not use any oily lubricant on any part of the oxygen concentrator as it may
Series No.

flow Oxygen Input


8. Technical alarm conditions---------------------------------------------------------- 8 output
Model testing capacity pressure purity power Noise Weight Dimension
range NO The performance parameters contaminate oxygen
9. Maintenance --------------------------------------------------------------------9
1)Inlet filter-----------------------------------------------------------------10 1 Nebulization rate ≥0.2 ml/min (2% normal saline) Caution
0.045~ 395 350
2)Humidifier bottle (reusable)-------------------------------------------10 SL-3C-510 0-5 5 390 60dB 25
0.08 620 2 Nebulization workflow≥6 LPM 1. When cleaning, please cut off the power.
3)Nebulization mask and tube-------------------------------------------10 0.045~ 395 350
2 SL-3C-550 0-10 10
0.08
630 60dB 27
620 3 Working Nebulization pressure 12Psi±1.5Psi 2. A well-packaged oxygen concentrator should be placed in a ventilated room with a
4)Air filter and oxygen outlet connector-------------------------------10
temperature of -40°C to 55°C, a relative humidity of less than 80% RH, and no corrosive gas;
The life span for molecular sieve oxygen concentrator is 8 years (In normal working condition) 4 Nebulization load noise≤65db(A)
10. Performance Parameter-----------------------------------------------------------11
3. It is forbidden to place the oxygen concentrator upside down or lying down. Handle with care.
11. Symbol definition-------------------------------------------------------------12
4. Choose certified safe and qualified. extension socket when cord extension is needed.
12. Ordinary malfunction and removal----------------------------------------13 【 Important Tips 】 【 Technical alarm conditions 】
13. Working principle------------------------------------------------------------14 Please read it carefully before using SL-3C series oxygen generator.The terms in this
manual contain important information. 1、 Power off prompt
14. Medical oxygen concentrator flow chart ---------------------------------14 【Inlet filter 】
When the power supply to the oxygen generator is disconnected, the oxygen generator will make
15. Packing list -------------------------------------------------------------15 a sound to remind the user.
16. Medical oxygen concentrator circuit charm ---------------------------15
2、 Compressor failure prompt
DANGER Urgent safety information for hazards that will cause serious injury or death.
WARNING Important safety information for hazards that might cause serious injury
CAUTION Information for preventing damage to the product
NOTE Information which you should pay special attention

Please pay special attention to all safety information indicated in this manual.
Figure 13 Figure 14 Figure 15
1 8 9
1

9、Please ask for medical assistance immediately to avoid injury if you feel unwell or are Caution:
DANGER---No Smoking DANGER---No open flame
experiencing a medical emergency while receiving oxygen therapy. To avoid electric shock, do not disassemble the oxygen concentrator. The machine must be
INTRODUCTION 10、Additional monitoring and or distributed alarm systems shall be equipped for the elder, disassembled by manufacturer’s qualified technician. Do not smear liquid directly on the
child or any other patient who are unable to express their discomfort, so as to convey t machine, and do not use any petroleum-based solvents or cleaning agents.
Before using the equipment, please read this manual and pay special attention to all
he information about discomfort and medical emergencies to the related caregivers to
safety information indicated in this manual. Shall you have any question, please contact
avoid injury. 【Performance Parameter 】
your supplier or the manufacturer. Figure 16 Figure 17
11、Smoking during oxygen therapy is dangerous and may cause facial burns or death. It is
forbidden to smoke in the room where the oxygen concentrator or accessories with oxygen NO Item Rating
Caution are located. If you want to smoke, turn off the oxygen concentrator, remove the cannula, Replace primary filter cotton and HEPA filter
1) Take off the air intake window from the back of the machine, remove the primary filter 0.2 -5 L/min
and leave the room where the cannula or mask or oxygen concentrator is. Shall you cannot Delivery Rate 0.2 -8 L/min
1
1、For your safety, the oxygen concentrator must be used according to the doctor's leave the room, please turn off the oxygen concentrator for 10 minutes before smoking. cotton on the rear side of the air intake window (see Figure 14), clean it with detergent (organic 0.2 -10 L/min
prescription. 12、During oxygen therapy, open flames are dangerous and may cause fire or death. No open solvents such as gasoline and alcohol are prohibited) or replace with new cotton (see Figure 15);
2 Outlet Pressure 7.252 psi ( 50kpa )
2、During oxygen therapy, fire risk may exist because of oxygen enrichment. Do not use flame is allowed within 2 meters of the oxygen concentrator or any accessories with oxygen. 2) Take off the air intake window at the back of the machine (see Figure 16), take out HEPA
oxygen concentrators or accessories near sparks or open flames. 13、Oxygen causes rapid combustion. Do not smoke while the oxygen concentrator is working filter and replace it (see Figure 17). It is recommended that the HEPA filter should be replaced 3W-1020≤610VA
3 Power Consumption 3W- 820≤520VA
or when a person who is receiving oxygen therapy nearby. Keep away from heat and fire sources. once every 1000 hours.
3、To ensure you getting the delivery of the oxygen according to the medical condition, the 3W- 520≤420VA
oxygen concentrator SL-3C-510, shall: (Note: The machine has been running for 1,000 hours, and the reminding is shown on the screen
to replace HEPA filter (see Figure 18). After HEPA filter is replaced, plug in the power after 4 Electrical Rating 250 V, 50Hz, 2.3A (Max.)
--Be used only when individually set up for you or specific activity level assigned to you; 【Why your physician prescribed oxygen supplement 】 unplugging, and then press "Mute" button for 3 seconds to eliminate the text reminder)
--Be used together with specific parts and accessories that meet the specifications of the 1~5 LPM 93%±3%
Nowadays, many people are suffering from heart, lung and other respiratory diseases. 5 Oxygen Percentage 1~8LPM 93%±3%
concentrator manufacturer, when settings for you are confirmed.
Many people can benefit from supplemental oxygen therapy. Body needs a steady supply of 1~10LPM 93%±3%
4、Before and during oxygen therapy, only oxygen-containing aqueous lotions or ointments 【 Humidifier bottle (reusable) 】
oxygen to function properly. Sound Level
can be used. Do not use petroleum or petroleum-based lotions or ointments to avoid the The physician has prescribed oxygen supplement because you cannot get enough oxygen 1) Rotates humidifier bttle clockwise; 6 ≤60dB
( ISO 8359:1996 from front )
risk of fire and burns. from the air in the room. Oxygen supplement will help increasing the amount of oxygen your 2) Pour pure water into the humidifier, the water level is lower than the top line of the calibration
Temperature -25℃ to 55℃
5、Do not lubricate the accessories, connections, oil pipes or other accessories of the oxygen body absorbs. Oxygen supplement is not addictive. The doctor prescribes a specific flow of line (see Figure 2);
7 Storage Conditions Humidity 15% t0 85%
concentrator to avoid fire and burns. oxygen to improve symptoms such as headache, drowsiness, confusion, fatigue, or irritability. 3) Tighten the humidifier counterclockwise, replace pure water for the humidifier every day, Atmospheric pressure 70~106kpa
6、Please use spare parts recommended by the manufacturer to ensure proper function and If these symptoms persist after you get oxygen supplement, please consult your doctor.
and clean humidifier each week. Pay special attention to clean the water filter of the humidifier
As consumable material,
avoid the risk of fire and burns. to prevent clogging. 8 molecular sieve the molecular should be replaced
7、The flow rate and the concentration of oxygen are expected to be affected adversely if the 【Working principle of SL-3C Oxygen Concentrator 】 according to its usage
device is used under the following situations: altitude above 1000 meters; external temperature 【Nebulization mask and tube 】
The oxygen concentrator is currently the most reliable, efficient and convenient source to Table6
exceeds 5-40ºC or more; the relative humidity higher than 90%, and the treatment may be It is suggested to be disposed after used.
supplement oxygen, and it is electric. This device separates oxygen from air, and delivers high-purity 【 Symbol definitionL 】
affected consequently.
supplemental oxygen to you through the oxygen outlet. Although the concentrator separates the 【Air filter and oxygen outlet connector 】
8、Do not leave the nasal cannula or mask on the bed cover or chair cushion when the oxygen General warning, warning, danger
oxygen in the room, it will not affect the normal amount of oxygen in your room.
concentrator is on but not in use. Turn off the oxygen concentrator in time when not in use to The air filter and connector shall be cleaned at least once a week. Please read the instructions carefully before using this product.
prevent high oxygen concentration, which makes fires more likely to occur and spread. CAUTION: Do not operate the concentrator without installing filter or while filter is wet. These
actions could permanently damage the concentrator.
2 3
10 11

LOCATION
【 The important components of the Oxygen Concentrator】 1. It should be placed in a ventilated and lightless room, at least 10cm from the surrounding FRAGILE
walls. Safety sign "No smoking"
Please take time to familiarize yourself with SL-3C Series Oxygen Concentrator before Fragile, Handle with care.
2. Keep away from heat and fire sources. Do not place it in the place of humidity or high
operating (Figure1)
temperature and outdoors. Safety sign “No open flames” THIS WAY UP
Before using this product, make sure you have the following items:
3. Do not put stuff, water and oil tanks on the oxygen concentrator. Indicates correct upright position of the transport package
--User Manual 4. Do not put anything at the bottom of the oxygen concentrator to avoid blocking the entry
--1 Oxygen Concentrator set “ON” for part of the device
and exit. KEEP AWAY FROM RAIN
--1 Oxygen Outlet Connector NOTE Transport package shall be kept away from rain
"Turn off" for part of the devices
1. The environment will affect the oxygen concentration and oxygen flow Since the nasal STACKING LIMIT BY NUMBER
oxygen tube is an open oxygen inhalation, when it inhales oxygen (90% V/V). At the same Alternating current Maximum number of identical packages which may be stacked on one anther
time, the surrounding air will also be inhaled ( 21% oxygen concentration). So the actual where”six”is the Limiting number
oxygen concentration inhaled into the lungs is the concentration mixed with the surrounding BF application part
KEEP AWAY FORM SUNLIGHT
Compressor intake air. When the oxygen concentration is less than 25%V/V, the oxygen concentration is close to Transport package shall not be exposed to sunlight
cover
the oxygen concentration in the air, that is, when the inhaled oxygen flow is less than 1L/min, Sitting prohibited
there is no therapeutic effect. In the process of oxygen therapy, please pay attention to oxygen
Intake louver
concentration and oxygen flow. IP21 Dust level 2, 【 ORDINARY MALFUNCTION AND REMOVAL】
Table3 Protected against solid foreign objects of 12.5mmØ and greater
Waterproof level 1 1.Unable to power on
Machine master
Oxygen flow 0 1 2 3 4 5
overcurrent switch Protection against vertically falling water drops 1)Check the plug to make sure the other connected machines to the plug works well.
protector
Oxygen
24.49 31.33 34.75 38.84 2)The overcurrent protector restoration, press restoration button and re-start machine.
20.9 27.89
concentration Class II equipment
2.No oxygen output after power on
Figure 1 1)Check the humidify bottle to make sure it is fastened, if not,
2. Under different pressure, Oxygen concentration will change accordingly In the sea areas, the air Refer to instruction manual
pressure is 101.3Kpa/cm2, in dry air, the Oxygen takes proportion of 20.4%, then the pO2 (Partial the oxygen will leak out from the gap to cause none-oxygen output.
【 PRECAUTIONS FOR USE 】 2)Rotate the flow adjustment knob to make sure that it is adjusted.
Pressure of Oxygen) is 21.15Kpa; it means that the Oxygen proportion in the air is not affected in Production serial number
--Replace damaged cables or plugs immediately; plateau area. When the air pressure decreases with the altitude, the O2 will decrease accordingly. 3.Water drop in the oxygen inhalator tube
--Using the firebreak connector to prevent the oxygen flow in the event of a fire; Authorized representative in the European community 1)Check the water level in the humidify bottle to make sure it is in the limits.
The following table is the data of Oxygen Concentration under different air pressure conditions
--Switch off the concentrator when not in use,; 2)Low temperature in the room will facilitate the oxygen with rich water content when
--Keep oxygen concentrators, cylinders and oil pipes away from anything that may Manufacturing date
The altitude it is humidified; if the temperature becomes lower that rich water content will condense
0 1000 2000 3000 4000
cause sparks or flames, such as cigarettes, candles, lanterns, portable heaters and (meter)
Manufacturer to water drops again, in the case, you need to adjust the temperature.
stove appliances; Air Pressure 101.3 89.87 79.5 70.11 61.64
--Do not use oil, grease or petroleum products near the oxygen concentrators, as these (kpa) 4.Low in oxygen concentration output
Do not re-use
increase the risk of fire and explosion; Concentration 1) Check the filter sponge to make sure it is clean, if not, please clean and dry fully to reuse.
--Place the oxygen concentrator on a flat surface to prevent accidents or damage to the under the rated 90.00 79.85 70.63 62.29 54.76 2) Normal aging, in different using environment, the oxygen concentrator has big
Batch code
compressor. flow(v/v)
difference service life; in rich humid using environment, the service life of the oxygen
Table4 According to Waste Electrical and Electronic Equipment Directive 2002/96/EC concentrator will greatly reduce. Please keep the oxygen concentrator away from the
(WEEE Directive) , the machine should be disposed of in accordance with local
humid environment.
laws and regulations if the duration of validity lasts for more than 3 years, so as
4 5 to avoid hurting users and polluting the environment.