Professional Documents
Culture Documents
1 s2.0 S0263224122004638 Main
1 s2.0 S0263224122004638 Main
Measurement
journal homepage: www.elsevier.com/locate/measurement
A R T I C L E I N F O A B S T R A C T
Keywords: In this work, a non-invasive heart rate (HR) and blood pressure (BP) monitoring system using a lead zirconate
Blood pressure monitoring titanate (PZT) piezoelectric and photoplethysmographic (PPG) sensor has, preliminarily, been investigated. A
Modified pulse transit time modified pulse transit time (MPTT) is applied to demodulate HR, systolic blood pressure (SBP), and diastolic
Non-invasive health monitoring
blood pressure (DBP) from obtained BP waveforms. In addition, an engineering application software written by
Piezoelectric sensor
Photoplethysmographic sensor
LabView programming is, also, developed for dynamic signal monitoring and processing of desired parameters.
Consequently, HR and BP measurements on 15 volunteers are performed and validated against a reference digital
sphygmomanometer. The results show equivalent HR from both sensors with a mean absolute difference (MAD)
of 1.78 beats per minute. Meanwhile, MADs of SBP and DBP measurements are 2.62 and 1.36 mmHg, complying
with AAMI-BHS accuracy of “Grade A”, respectively. Therefore, PZT and PPG-based monitoring system could
lead to development of accurate, non-invasive, and low-cost HR and BP devices.
* Corresponding author.
E-mail address: saroj@su.ac.th (S. Pullteap).
https://doi.org/10.1016/j.measurement.2022.111211
Received 30 September 2021; Received in revised form 8 April 2022; Accepted 15 April 2022
Available online 19 April 2022
0263-2241/© 2022 Elsevier Ltd. All rights reserved.
P. Samartkit et al. Measurement 196 (2022) 111211
2. Related theories Generally, non-invasive heart rate (HR) and blood pressure (BP)
measurements rely on the same principle: the acquisition of arterial
2.1. Brief of photoplethysmographic sensor pulse waveform (APW). As blood propagation corresponds to the cardiac
cycle, APW also indicates important cardiovascular parameters of a
The principle of the photoplethysmographic (PPG) sensor is the person, as shown in Fig. 3.
detection of light absorption by the tissue. Particularly, the light source From the figure, the APW can be categorized into 2 separate phases:
illuminates the desired body parts, where the target biocomponents (e. systole and diastole. During the former period, the heart contracts to
g., blood cells) would absorb the light [10,14]. Here, illumination using inject blood and BP rapidly rises until reaching the systolic peak, and the
green light (~530 nm of wavelength) causes optical backscattering BP at this point is defined as SBP. Afterwards, the heart enters the
which could be observed via a photodetector [21]. The operating prin diastole or relaxation phase with a noticeable decrease in BP. However,
ciple of this type of PPG sensor, also known as “reflectance mode PPG” the remaining energy within the arterial walls causes a subsequent in
[10], could be illustrated in Fig. 1. crease in BP, indicating a dicrotic notch between the systole and diastole
Incident light from the light source is absorbed by the blood cells period. BP later reaches the dicrotic peak and then reduces to a baseline
while the rest is backscattered to the PPG sensor. Since the blood volume value of DBP before the next cardiac cycle occurs [3].
fluctuates according to the cardiac cycle, the absorption of light depends In the case of HR measurement, the systolic peak is counted as it
on the amount of blood propagating through the illuminated area. In this corresponds to the heart pulse. Basically, counting the systolic peak over
aspect, the back-scattered light intensity is inversely proportional to the 60 s would give HR in beats per minute (bpm). Alternatively, this
blood volume, as expressed by the Beer-Lambert law [22,23]: operation could be completed in a shorter time by mathematical
2
P. Samartkit et al. Measurement 196 (2022) 111211
where l indicates the length between the 2 arterial sites, E corresponds to 3.1. Designing and configuration of developed system
the Young’s modulus of the arterial wall, h is the wall thickness, ρ rep
resents the blood density, r is the vessel radius, E0 indicates E at zero BP The proposed system for HR and BP monitoring is comprised of a PZT
value, while γ is the vessel coefficient. For simplicity, rearranging (4) sensor from Murata model: 7BB-20-3 with 3.6 kHz resonant frequency
gives a general BP calculation in the form of: and a PPG sensor connected to their respective signal conditioning cir
( ) cuit. A microcontroller (Arduino UNO) is, consequently, utilized to ac
2 lρr
BP = ln (5) quire and digitize the conditioned signals with 10 bits of analog-to-
γ E0 hPTT
digital converter (ADC) at 100 samples/seconds. These data are,
finally, transferred into the dedicated computer and then processed with
an engineering application software developed via LabView program
ming, which allows the PPG and PZT waveforms to be recorded for
subsequent analyses. The configuration of the monitoring system is
illustrated in Fig. 5.
According to the configuration, the proximal PZT sensor (a) is placed
and then strapped to the user’s wrist without any underlying elements.
This ensures the arterial distension would properly be induced onto the
sensor by tightening the wrist strap. Once correctly fastened, the
circumference of the strap is marked using its scale for repeatable ex
amination. Therefore, equivalent tightening force can be repeated for
Fig. 4. Basic principle of pulse transit time (PTT) technique. subsequent examination. Meanwhile, a handheld probe is made with an
3
P. Samartkit et al. Measurement 196 (2022) 111211
embedded distal PPG sensor (b), which allows firm placement of the (equivalent to 1 Hz), another high-pass filtering circuit of 0.3 Hz cutoff
thumb during the monitoring. Note that an additional circuit for the PZT frequency is, therefore, implemented to filter out leftover signal drift
sensor (c) is separated from other components to initially amplify the due to minor motion artifacts (e.g. breathing and slight muscle
PZT output and also filter the signal noise. Other signal conditioning contraction). Once the signal conditioning process is completed, the
circuits (d) are, subsequently, implemented to pre-process the raw PZT microcontroller digitizes the PZT and PPG signals to be displayed using a
and PPG signals. Ultimately, the microcontroller (e) is utilized to digitize LabView-based engineering application software. An example of its user
and transfer the pulse waveforms for monitoring and further process by interface is shown in Fig. 7.
the computer (f). To emphasize signal and data processing, the purpose In particular, the engineering application software is developed with
of the aforementioned circuits is, however, expanded upon in the next 5 main interfaces as follow: (a) run or stop function, (b) configure the
section. period in which the signals would be displayed and recorded, (c) export
the APWs to spreadsheet files (.xlsx) for subsequent analyses, (d) real-
time display of PZT and PPG signals, and (e) recorded waveforms over
3.2. Data processing for heart rate and blood pressure demodulation a configured period from 0 to 60 s. Note that after monitoring the signals
for a set period of time, HR could be automatically indicated. To prevent
The overall signal processing of the monitoring system from the potential errors, both PZT and PPG signals are applied for the HR
signal conditioning circuits to the digitized data could be shown ac calculation using (3), which allows the user to validate the results.
cording to Fig. 6. Meanwhile, PTTs and PTTd are determined elsewhere via the exported
Although both PZT and PPG sensors could detect the APW of data. This gives more flexibility to the analysis, since parameters other
humans, the retrieved signal strength is very small and prone to noises. than the mentioned PTT could be investigated as well. A moving average
Therefore, the signal conditioning circuits are, next, employed. As function and also a peak detection algorithm, included in the developed
mentioned before, the PZT sensor has an additional circuit, which am software, have been employed to eliminate undesirable signal noise, and
plifies the sensing signal and also removes noise before further opera indicate the essential systolic peaks and dicrotic notches, respectively.
tions. Nonetheless, the rest of the signal conditioning of both PZT and Afterwards, the pre-calibration of the monitoring system could be per
PPG signals are as follows: first, a high-pass filtering circuit with a 0.17 formed, leading to the calculation of the user’s SBP and DBP. In this case,
Hz cutoff frequency eliminates potential signal drift in the system. the method of pre-calibration would be detailed in the experimentation
Consequently, an approximately 52 dB signal amplification increases the part.
signal strength before a low-pass filtering circuit removes possible noises
of above 28 Hz. Since the human heart rate is approximately 60 bpm
3.3. Performance validation
• The APW acquired through PZT and PPG sensors are applied for PTTs
and PTTd calculations. Then, the sphygmomanometer measures the
SBP, DBP, and HR of the volunteers.
• Afterwards, all determined parameters are applied in (6)–(7), which
consequently determines the pre-calibration values of a and b of the
Fig. 6. Signal processing procedures for HR and BP measurements. volunteer.
4
P. Samartkit et al. Measurement 196 (2022) 111211
Fig. 7. User interface of HR and BP monitoring system with dedicated functions: (a) run/stop button, (b) display period setting, (c) data export option, (d) real-time
display, and (e) record of acquired waveforms over a set period.
• The above processes are repeated for 3 times to compensate for any differs from the BHS, the lower grade is suggested for the proposed BP
BP and PTT changes during the pre-calibration. Subsequently, an device. These BP validation protocols are considered for several BP
average value of the coefficients a and b are acquired for each measuring devices [35–37]. Important to note is that the AAMI-BHS
volunteer. standard requires at least 85 subjects with varying BP levels for the
device validation. In this work, 15 volunteers have, however, been
The continuous HR and BP monitoring are performed for a duration recruited to preliminarily investigate the developed PZT and PPG-based
of 60 s at a time, sequentially followed by the operation of the reference system before future clinical trials. To achieve this, the system should
device. This procedure is carried out for 3 times of repeatability, thereby attain at least a “Grade B” accuracy for the passing criteria.
a total of 45 datasets are obtained. To compensate for the arterial oc Regardless, an additional experiment is performed to validate the
clusion, reference readings adjacent to each continuous monitoring repeatability of the PZT and PPG sensor-based monitoring system in
result are averaged for the validation. The results from both measuring daily usage. In this case, 3 volunteers have participated for 7 days long
instruments are later compared against one another to investigate the term examination, where their HR and BP are measured according to the
performance of the proposed system. Here, the mean absolute difference previously described procedure. However, the pre-calibrations of co
(MAD) is applied to assess the accuracy of the developed system by using efficients a and b are obtained through 3 different methods: determined
the expression of [32]: only once on the first day, pre-calibrated every day throughout the 7
days, and averaged once the examination is over.
1∑ M
MAD = |χ − χ t | (8)
n n=1 m 4. Experimental results and discussion
where χ m is the BP value measured by the developed system, χ t is the BP During the experiment, the proposed monitoring system can,
given by the reference device, n represents the test number, and M in continuously, detect and record the APW obtained from both PZT and
dicates the total number of measurements. Consequently, the MAD re PPG sensors in real-time. Examples of these are illustrated in Fig. 8. Note
sults are categorized using the grading criteria set by the Association for that the amplitude of the PPG signal is inverted to show its similarity to
the Advancement of Medical Instrument (AAMI) and the British Society the actual APW, which eases the detection of important APW charac
of Hypertension (BHS), as displayed in Table 1 [33,34]. teristics such as the systolic peaks and dicrotic notches.
From the table, the grading is achieved when the resulting MAD and From Fig. 8(a), the first peaks of each PZT and PPG signal denote the
cumulative percentage of absolute difference are within the specified systolic peaks, which are considered as the number of heart pulse (N)
values. Here, an instrument which passes the AAMI criteria must, also, and parameters for PTTs measurement. In Fig. 8(b), the PTTd is
have MAD and standard deviation (S.D.) of lower than 5 mmHg and 8 measured as the time difference between the dicrotic notches or the
mmHg, respectively. To pass the BHS criteria, however, the accuracy of inflections between the first and second signal peaks. In the case of the
at least “Grade B” must be fulfilled [35]. In case the grading from AAMI examples, PTTs and PTTd are found to be ~0.11 and ~0.12 s, respec
tively. These PTT parameters are, consequently, applied for the pre-
calibration process and later the subsequent SBP and DBP de
Table 1 modulations. This also proves that PZT and PPG sensors can detect APW
AAMI-BHS grading criteria for BP measuring devices [33,34]. and obtain vital cardiovascular parameters for further analyses.
Grade AAMI criteria BHS criteria It should be noted that the PZT sensor utilizes less energy during the
Criteria MAD within certain range Cumulative percentage of absolute difference measurement, as no power supply is required. This also lowers accidents
within:
from sensor overheating over long periods of operation, improving the
<5 mmHg <10 mmHg <15 mmHg overall safety of using the system. This encourages the adoption of the
A ≤4 mmHg ≥60% ≥80% ≥95% PZT sensor for energy-efficient HR measurement applications. Here, an
B 4–6 mmHg ≥50% ≥75% ≥90% experiment using a PZT sensor instead of the PPG at the thumb has been
C 6–7 mmHg ≥40% ≥65% ≥85% carried out. It is found that the APW can also be continuously detected
D ≥7 mmHg Worse than C
comparable to the PPG signal. This, thus, implies that the use of dual PZT
5
P. Samartkit et al. Measurement 196 (2022) 111211
Fig. 8. Examples of PZT and PPG signals for (a) PTTs measurement and (b) PTTd measurement.
or PPG sensors could be implemented. However, the thumb must be been performed on 15 volunteers to allow further demodulation. A total
tightly strapped to the PZT sensor, as the finger compression is explicitly of 45 SBP and DBP calculations using (6)–(7) have, consequently, been
revealed. This would eventually cause discomfort to the subject and carried out, with the results summarized in Fig. 10.
leads to unintentional movements, resulting in significant measurement Fig. 10(a) shows the overall BP measurement results, while the ac
errors. Therefore, the PPG sensor might be more suitable for APW curacy validations of SBP and DBP have been illustrated in Fig. 10(b)–
monitoring at the thumb and fingertips, due to its non-occlusive design. (c), respectively. Consequently, it found that the developed monitoring
On the other hand, the signal conditioning circuits utilized for the PZT system has highly accurate SBP and DBP measurement capabilities with
and PPG sensors are similar. This is to prevent any potential signal de MAD ± S.D. of 2.62 ± 1.92 mmHg and 1.36 ± 1.05 mmHg, respectively.
lays, which would consequently cause an error to the PTTs and PTTd Furthermore, the percentages of absolute difference in SBP under 5, 10,
measurements. Nevertheless, the pre-calibration of PTT using a and 15 mmHg are 86.67%, 100.00%, and 100.00%, respectively.
population-based BP might compensate for this bias, thus would not
result in erroneous BP readings. This topic is of interest for the future.
After monitoring all volunteers, their HR, SBP, and DBP are
demodulated. In the former measurement, the results given by the PZT
and PPG sensors have been compared with one another and also with the
reference sphygmomanometer. These are summarized in Fig. 9.
Incidentally, the HR obtained from the PZT and PPG signals are
equivalent, implying that the proposed monitoring system could utilize
any sensors for HR monitoring. Once compared to the reference device,
the mean absolute difference (MAD) in measurement is 1.78 bpm with a
standard deviation (S.D.) of 1.98 bpm, respectively. Additionally,
93.33% of all acquired HR through both PZT and PPG sensors are within
the accuracy range of the sphygmomanometer, shown by the error bars
[38]. In this case, the difference in HR values is due to the different
techniques employed in the measurements. Since the digital sphygmo
manometer operates no longer than 30 s, the given HR is estimated via a
built-in algorithm, whereas the PZT and PPG sensor-based monitoring
system performs the measurement over 60 s to give an actual HR of a
person. The latter, however, does not require an inflatable cuff and can
also display the dynamic APW of a person in test. This means the
developed system can continuously measure HR without arterial oc
clusion, in contrast to the digital sphygmomanometer. Since HR
measured by the PZT and PPG-based system is obtained via systolic peak
detection, the developed system could, therefore, monitor APW with
high accuracy for subsequent BP measurement.
In BP measurement, the pre-calibration of the developed system has
Fig. 10. BP validation results: (a) overall BP measured from reference and
proposed system, (b) absolute difference in SBP readings, and (c) absolute
Fig. 9. Summary of HR measurement results. difference in DBP results.
6
P. Samartkit et al. Measurement 196 (2022) 111211
SBP results MAD (mmHg) 9.76 2.53 CRediT authorship contribution statement
S.D. (mmHg) 9.25 2.38
DBP results MAD (mmHg) 8.81 2.56 Piyawat Samartkit: Investigation, Validation, Writing – original
S.D. (mmHg) 8.52 2.34
draft. Saroj Pullteap: Conceptualization, Methodology, Software,
7
P. Samartkit et al. Measurement 196 (2022) 111211
Writing – review & editing, Supervision. Olivier Bernal: Resources, [18] K.-H. Huang, F. Tan, T.-D. Wang, Y.-J. Yang, A Highly Sensitive Pressure-Sensing
Array for Blood Pressure Estimation Assisted by Machine-Learning Techniques,
Methodology, Validation.
Sensors 19 (4) (2019) 848.
[19] D. Zambrana-Vinaroz, J.M. Vicente-Samper, C.G. Juan, V. Esteve-Sala, J.
Declaration of Competing Interest M. Sabater-Navarro, Non-Invasive Device for Blood Pressure Wave Acquisition by
Means of Mechanical Transducer, Sensors 19 (19) (2019) 4311.
[20] P. Samartkit, S. Pullteap, H.C. Seat, Validation of Fiber Optic-Based Fabry-Perot
The authors declare that they have no known competing financial Interferometer for Simultaneous Heart Rate and Pulse Pressure Measurements,
interests or personal relationships that could have appeared to influence IEEE Sens. J. 21 (5) (2021) 6195–6201.
the work reported in this paper. [21] N. Sviridova, T. Zhao, K. Aihara, K. Nakamura, A. Nakano, Photoplethysmogram at
Green Light: Where does Chaos Arise From? Chaos, Solitons Fractals 116 (2018)
157–165.
References [22] M.P.D. Pont, J.L.B. Marques, Reflective Photoplethysmography Acquisition
Platform With Monitoring Modules and Noninvasive Blood Pressure Calculation,
[1] H.-G. Kim, E.-J. Cheon, D.-S. Bai, Y.H. Lee, B.-H. Koo, Stress and Heart Rate IEEE Trans. Instrum. Meas. 69 (8) (2020) 5649–5657.
Variability: A Meta-Analysis and Review of the Literature, Psychiatry Investig. 15 [23] X. Ding, B.P. Yan, Y.-T. Zhang, et al., Pulse Transit Time Based Continuous Cuffless
(3) (2018) 235–245. Blood Pressure Estimation: A New Extension and A Comprehensive Evaluation, Sci.
[2] A. Temko, Accurate Heart Rate Monitoring During Physical Exercises Using PPG, Rep. 7 (1) (2017) 11554.
IEEE Trans. Biomed. Eng. 64 (9) (2017) 2016–2024. [24] Y. Chen, Y. Zhu, H.T. Ma, H. Huang, A study of photoplethysmography intensity
[3] R.E. Klabunde, Cardiovascular Physiology Concepts, second ed., Lippincott ratio in hypertension, Proceedings of the 2016 IEEE International Conference on
Williams & Wilkins, China, 2012. Real-time Computing and Robotics (RCAR), 6–10 June 2016, pp. 317–320, doi:
[4] B. Williams, G. Mancia, W. Spiering, et al., 2018 ESC/ESH Guidelines for The 10.1109/RCAR.2016.7784046.
Management of Arterial Hypertension: The Task Force for the Management of [25] A. Panahi, A. Hassanzadeh, A. Moulavi, Design of a Low Cost, Double Triangle,
Arterial Hypertension of the European Society of Cardiology (ESC) and the Piezoelectric Sensor for Respiratory Monitoring Applications, Sens. Bio-Sens. Res.
European Society of Hypertension (ESH), Eur. Heart J. 39 (33) (2018) 3021–3104. 30 (2020) 100378.
[5] A.S. Alghamdi, K. Polat, A. Alghoson, A.A. Alshdadi, A.A. Abd El-Latif, A Novel [26] D.S. Aulakh, S. Bhalla, 3D torsional experimental strain modal analysis for
Blood Pressure Estimation Method based on the Classification of Oscillometric structural health monitoring using piezoelectric sensors, Measurement 180 (2021)
Waveforms using Machine-Learning Methods, Appl. Acoust. 164 (2020). 109476.
[6] C. Peng, M. Chen, H.K. Sim, Y. Zhu, X. Jiang, Noninvasive and Nonocclusive Blood [27] C.W.D. Silva, Sensor System: Fundamentals and Applications, 25th ed., CRC Press,
Pressure Monitoring via a Flexible Piezo-Composite Ultrasonic Sensor, IEEE Sens. Boca Raton, FL, 2017.
J. 21 (3) (2021) 2642–2650. [28] J.-J. Huang, H.-Y. Syu, Z.-L. Cai, A.R. See, Development of a long term dynamic
[7] P. Palatini, R. Asmar, Cuff Challenges in Blood Pressure Measurement, J. Clin. blood pressure monitoring system using cuff-less method and pulse transit time,
Hypertens. 20 (7) (2018) 1100–1103. Measurement 124 (2018) 309–317.
[8] R. Kumar, P.K. Dubey, A. Zafer, A. Kumar, S. Yadav, Past, present and future of [29] C. Yang, N. Tavassolian, Pulse Transit Time Measurement Using
blood pressure measuring instruments and their calibration, Measurement 172 Seismocardiogram, Photoplethysmogram, and Acoustic Recordings: Evaluation
(2021) 108845. and Comparison, IEEE J. Biomed. Health. Inf. 22 (3) (2018) 733–740.
[9] M. Elgendi, R. Fletcher, Y. Liang, N. Howard, N.H. Lovell, D. Abbott, K. Lim, [30] M. Sharma, K. Barbosa, V. Ho, D. Griggs, T. Ghirmai, S. Krishnan, T. Hsiai, J.-
R. Ward, The Use of Photoplethysmography for Assessing Hypertension, npj Digital C. Chiao, H. Cao, Cuff-Less and Continuous Blood Pressure Monitoring: A
Med. 2 (1) (2019). Methodological Review, Technologies 5 (2) (2017) 21.
[10] G. Wang, M. Atef, Y. Lian, Towards a Continuous Non-Invasive Cuffless Blood [31] Y. Ma, J. Choi, A. Hourlier-Fargette, et al., Relation between Blood Pressure and
Pressure Monitoring System Using PPG: Systems and Circuits Review, IEEE Circuits Pulse Wave Velocity for Human Arteries, Proc. Natl. Acad. Sci. 115 (44) (2018)
Syst. Mag. 18 (3) (2018) 6–26. 11144–11149.
[11] X.R. Ding, Y.T. Zhang, J. Liu, W.X. Dai, H.K. Tsang, Continuous Cuffless Blood [32] J. Qi, J. Du, S.M. Siniscalchi, X. Ma, C.H. Lee, On Mean Absolute Error for Deep
Pressure Estimation Using Pulse Transit Time and Photoplethysmogram Intensity Neural Network Based Vector-to-Vector Regression, IEEE Signal Process Lett. 27
Ratio, IEEE Trans. Biomed. Eng. 63 (5) (2016) 964–972. (2020) 1485–1489.
[12] R. Lazazzera, Y. Belhaj, G. Carrault, A New Wearable Device for Blood Pressure [33] IEEE Standard for Wearable, Cuffless Blood Pressure Measuring Devices -
Estimation Using Photoplethysmogram, Sensors 19 (11) (2019) 2557. Amendment 1, in IEEE Std 1708a-2019 (Amendment to IEEE Std 1708-2014),
[13] Y.-H. Kao, P.C.P. Chao, C.-L. Wey, Towards Maximizing the Sensing Accuracy of an 2019, pp. 1–35, doi: 10.1109/IEEESTD.2019.8859685.
Cuffless, Optical Blood Pressure Sensor using a High-order Front-end Filter, [34] P. Li, T.M. Laleg-Kirati, Central Blood Pressure Estimation From Distal PPG
Microsyst. Technol. 24 (11) (2018) 4621–4630. Measurement Using Semiclassical Signal Analysis Features, IEEE Access 9 (2021)
[14] G. Slapničar, N. Mlakar, M. Luštrek, Blood Pressure Estimation from 44963–44973.
Photoplethysmogram Using a Spectro-Temporal Deep Neural Network, Sensors 19 [35] S. Baek, J. Jang, S. Yoon, End-to-End Blood Pressure Prediction via Fully
(15) (2019) 3420. Convolutional Networks, IEEE Access 7 (2019) 185458–185468.
[15] S.S. Srikant, P.K. Chaturvedi, Basic Electronics Engineering: Including Laboratory [36] H.P. Friz, V. Punzi, F. Petri, et al., Validation of the Suntech® CT40™ Blood
Manual, Springer Singapore, Germany, 2020. Pressure Measurement Device by the BHS Protocol and the AAMI/ISO 8160–2:
[16] N. Isakadze, S.S. Martin, How useful is the smartwatch ECG? Trends Cardiovasc. 2013 Standard, J. Hypertens. 35 (2017) e164.
Med. 30 (7) (2020) 442–448. [37] J. Topouchian, Z. Hakobyan, J. Asmar, et al., Clinical Accuracy of the Omron M3
[17] D. Wang, J. Shen, L. Mei, et al., Performance Investigation of a Wearable Comfort® and the Omron Evolv® for Self-Blood Pressure Measurements in
Distributed-Deflection Sensor in Arterial Pulse Waveform Measurement, IEEE Sens. Pregnancy and Pre-eclampsia - Validation according to the Universal Standard
J. 17 (13) (2017) 3994–4004. Protocol, Vasc. Health Risk Manag. 14 (2018) 189–197.
[38] Automatic Blood Pressure Monitor Model HEM-7130 Instruction Manual, Omron
Healthcare Co. Ltd., 2013.