v4.0 Monitor Service Protocol EN
v4.0 Monitor Service Protocol EN
Service protocol
Recurrent test Test after repair
Result o.k.
1. Visual inspection N/A Yes No
Power supply cord is without any visible cracks, kinks or other damage
Readability of all safety related labels, tags and identifications
Equipment in good condition, clean and without any visible indication of damage
2. Electrical safety test
2.1 Protective earth resistance (Test current ≥ 0,2A) Measured Result o.k.
Threshold
according IEC62353 value Yes No
Protective earth resistance, only equipment Rmax = 0,2 Ω Ω
Protective earth resistance, only power supply cord Rmax = 0,1 Ω Ω
2.2 Insulation resistance (Test voltage 500V d.c.) according Measured Result o.k.
Threshold
IEC62353 value Yes No
Between mains part and enclosure Rmin = 2 MΩ MΩ
Measured Result o.k.
2.3 Leakage current measurements according IEC62353 Threshold
value Yes No
LF-Equipment leakage current (Alternative method) Imax = 1 mA mA
Result o.k.
3. Function check Yes No
No safety or functional deficiencies were detected
4. Test results
Yes No
The device shows no relevant deviations and can be used without any restrictions
Comments: