Service protocol
Recurrent test Test after repair

55" 4K UHD 3D Monitor (PV015)


Manufacturer: Aesculap AG, Am Aesculap-Platz, 78532 Tuttlingen / Germany
Serial-No.: Inventory-No.: Applied part type: 0
Class of protection: l Detachable power supply cord
Responsible organization:

Measurement equipment Tests Typ Test equipment number


Device tester 2.1 , 2.3 Secutest SIII
Insulation tester 2.2 Megger MIT410

Result o.k.
1. Visual inspection N/A Yes No
 Power supply cord is without any visible cracks, kinks or other damage
 Readability of all safety related labels, tags and identifications
 Equipment in good condition, clean and without any visible indication of damage
2. Electrical safety test
2.1 Protective earth resistance (Test current ≥ 0,2A) Measured Result o.k.
Threshold
according IEC62353 value Yes No
 Protective earth resistance, only equipment Rmax = 0,2 Ω Ω
 Protective earth resistance, only power supply cord Rmax = 0,1 Ω Ω
2.2 Insulation resistance (Test voltage 500V d.c.) according Measured Result o.k.
Threshold
IEC62353 value Yes No
 Between mains part and enclosure Rmin = 2 MΩ MΩ
Measured Result o.k.
2.3 Leakage current measurements according IEC62353 Threshold
value Yes No
 LF-Equipment leakage current (Alternative method) Imax = 1 mA mA
Result o.k.
3. Function check Yes No
 No safety or functional deficiencies were detected
4. Test results
Yes No
The device shows no relevant deviations and can be used without any restrictions
Comments:

Place / Date: Name & signature service technician:

Effective

Document No.: NA - Version: 4.0 - Document ID: AAG-Service Documentation-000788 Print Date: 2023-01-23 22:46 (CET)