Download as pdf or txt
Download as pdf or txt
You are on page 1of 18

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles

for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

Designation: C1871 − 22

Standard Test Method for


Determination of Uranium Isotopic Composition by the
Double Spike Method Using a Thermal Ionization Mass
Spectrometer1
This standard is issued under the fixed designation C1871; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.

1. Scope ments of low enriched or natural uranium isotopic reference


1.1 This test method describes the determination of the materials on a regular basis.
isotope amount ratios of uranium material as nitrate solutions 1.6 The DS method can only be applied to uranium samples
by the double spike (DS) method using a thermal ionization with relative isotope abundances 233U/U below 10–5 and
236
mass spectrometer (TIMS) instrument. U/U below 5 × 10–4, the DS method is therefore mainly used
1.2 The analytical performance in the determination of the for low enriched or close to natural uranium samples.
235
U/238U major isotope amount ratio by the DS method is five 1.7 Units—The values stated in SI units are to be regarded
to ten times better in terms of the internal and external as the standard. When no SI units are provided, the values are
reproducibility compared to the (“classical”) total evaporation for information only.
(TE) method as described in Test Method C1672 and the 1.8 This standard does not purport to address all of the
“modified total evaporation” (MTE) as described in Test safety concerns, if any, associated with its use. It is the
Method C1832. This is due to the use of an internal rather than responsibility of the user of this standard to establish appro-
external mass fractionation correction by using a double spike priate safety, health, and environmental practices and deter-
material with a known or certified 233U/236U isotope ratio, mine the applicability of regulatory limitations prior to use.
which is mixed with the sample prior to the measurement, 1.9 This international standard was developed in accor-
either during the sample preparation or directly on the TIMS dance with internationally recognized principles on standard-
filament. ization established in the Decision on Principles for the
1.3 The DS method cannot be applied for the determination Development of International Standards, Guides and Recom-
of the 236U/238U minor isotope amount ratio, and is also not mendations issued by the World Trade Organization Technical
recommended for the determination of the 234U/238U minor Barriers to Trade (TBT) Committee.
isotope amount ratio.
2. Referenced Documents
1.4 In case the uranium amount concentration of the double
spike is known or certified, the uranium amount concentration 2.1 ASTM Standards:2
of the sample can be determined using the isotope dilution C753 Specification for Nuclear-Grade, Sinterable Uranium
mass spectrometry (IDMS) method as described in Test Dioxide Powder
Method C1672, by blending the sample gravimetrically with C776 Specification for Sintered Uranium Dioxide Pellets for
the double spike and performing a DS measurement. Light Water Reactors
C787 Specification for Uranium Hexafluoride for Enrich-
1.5 An external mass fractionation correction by measure-
ment
ments of a certified reference material loaded on different
C833 Specification for Sintered (Uranium-Plutonium) Diox-
filaments and measured in the same measurement sequence, as
ide Pellets for Light Water Reactors
recommended for TE and required for MTE measurements, is
C859 Terminology Relating to Nuclear Materials
not necessary for the DS method. However, for quality control
C967 Specification for Uranium Ore Concentrate
(QC) purposes it is recommended to perform DS measure-
C996 Specification for Uranium Hexafluoride Enriched to
Less Than 5 % 235U
1
This test method is under the jurisdiction of ASTM Committee C26 on Nuclear
Fuel Cycle and is the direct responsibility of Subcommittee C26.05 on Methods of
2
Test. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved Feb. 1, 2022. Published March 2022. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 2018. Last previous edition approved in 2018 as C1871 – 18a. DOI: Standards volume information, refer to the standard’s Document Summary page on
10.1520/C1871-22. the ASTM website.

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG )UL-DQ*07


1
'RZQORDGHGSULQWHGE\
8&5 8QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
C1871 − 22
C1008 Specification for Sintered (Uranium-Plutonium) 3.3 Acronyms:
DioxidePellets—Fast Reactor Fuel (Withdrawn 2014)3 3.3.1 CRM—certified reference material
C1068 Guide for Qualification of Measurement Methods by 3.3.2 DS—double spike
a Laboratory Within the Nuclear Industry
C1128 Guide for Preparation of Working Reference Materi- 3.3.3 DU—depleted uranium
als for Use in Analysis of Nuclear Fuel Cycle Materials 3.3.4 EU—European Union
C1156 Guide for Establishing Calibration for a Measure- 3.3.5 FAR—Faraday Cup
ment Method Used to Analyze Nuclear Fuel Cycle Mate-
3.3.6 HEU—high enriched uranium
rials
C1347 Practice for Preparation and Dissolution of Uranium 3.3.7 IAEA—International Atomic Energy Agency
Materials for Analysis 3.3.8 ICPMS—inductively coupled mass spectrometry
C1411 Practice for The Ion Exchange Separation of Ura- 3.3.9 IRMM—Institute for Reference Materials and Mea-
nium and Plutonium Prior to Isotopic Analysis surements (since 1 July 2016 called JRC-Geel, the only unit
C1672 Test Method for Determination of Uranium or Pluto- working with nuclear materials at JRC-Geel is JRC-G.2)
nium Isotopic Composition or Concentration by the Total
Evaporation Method Using a Thermal Ionization Mass 3.3.10 ITU—Institute for Transuranium Elements (since 1
Spectrometer July 2016 called JRC-Karlsruhe, the only unit involved with
C1832 Test Method for Determination of Uranium Isotopic thermal ionization mass spectrometry measurements of nuclear
Composition by Modified Total Evaporation (MTE) materials at JRC-Karlsruhe is JRC-G.II.6)
Method Using Thermal Ionization Mass Spectrometer 3.3.11 JRC—Joint Research Centre
D1193 Specification for Reagent Water 3.3.12 LEU—low enriched uranium
E2586 Practice for Calculating and Using Basic Statistics
E2655 Guide for Reporting Uncertainty of Test Results and 3.3.13 MTE—modified total evaporation
Use of the Term Measurement Uncertainty in ASTM Test 3.3.14 NBL—New Brunswick Laboratory (since 15 May
Methods 2016 called NBL-Program Office)
3.3.15 NML—Nuclear Material Laboratory (part of the
3. Terminology IAEA)
3.1 Terminology C859 contains terms, definitions, descrip- 3.3.16 QC—quality control
tions of terms, nomenclature, and explanations of acronyms
and symbols specifically associated with standards under the 3.3.17 RSD—relative standard deviation—SD (see below)
jurisdiction of Committee C26 on Nuclear Fuel Cycle. divided by the mean value of the observations in repeated
sampling.
3.2 Definitions:
3.2.1 abundance sensitivity, n—in isotope amount ratio 3.3.18 RSE—relative standard error—SE (see below) di-
measurements, the ratio of the measured intensity of an ion vided by the mean value of the observations in repeated
beam at a mass, m, to the measured intensity from the same sampling.
isotope measured at one mass unit difference (for example, m 3.3.19 SD—standard deviation—according to Practice
6 1). E2586, 3.1.30: The square root of the sum of the squared
3.2.1.1 Discussion—Abundance sensitivity is a measure of deviations of the observed values in the sample divided by the
the magnitude of the peak tailing correction. For measuring sample size minus 1.
uranium on thermal ionization mass spectrometer (TIMS) and 3.3.20 SE—standard error—according to Practice E2586,
inductively coupled plasma mass spectrometry (ICP-MS) 3.1.29: Standard deviation of the population of values of a
instruments, the abundance sensitivity is typically calculated as sample statistic (that is, the mean value) in repeated
the ratio of the measured signal intensities at masses 237 and measurements, or an estimate of it.
238 using a suitable uranium sample. 3.3.20.1 Discussion—According to Practice E2586, 3.1.30:
3.2.2 modified total evaporation, MTE, n—analytical If the standard error (SE, see above) of a statistic is estimated,
method for determination of isotope amount ratios of uranium, it will itself be a statistic with some variance that depends on
as described in Test Method C1832. the sample size, that is, the number of observed values in the
3.2.3 total evaporation, TE, n—analytical method for deter- sample (Practice E2586, 3.1.26).
mination of isotope amount ratios of uranium or plutonium, as 3.3.20.2 Discussion—According to Practice E2655, 5.8.4.1:
described in Test Method C1672, also called “classical” total From statistical theory, a 95 % confidence interval for the mean
evaporation in this test method. of a normal distribution, given n independent observations x1,
x2, ..., xn drawn from the distribution, is x̄ 6 t × SD / √n, where
3.2.4 turret, n—holder for sample filaments.
x̄ is the sample mean, SD is the standard deviation of the
3.2.4.1 Discussion—Alternate names for turret are carousel,
observations (see above), and t is the 0.975 percentile of the
magazine, and wheel.
Student’s t distribution with n-1 degrees of freedom. Because
Student’s t distribution approaches the normal as n increases,
3
The last approved version of this historical standard is referenced on the value of t approaches 1.96 as n increases. This is the basis
www.astm.org. for using the (coverage) factor 2 for expanded uncertainty. The

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG )UL-DQ*07


2
'RZQORDGHGSULQWHGE\
8&5 8QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
C1871 − 22
standard error (SE) of the mean value of a series of n This measurement can be performed using the TE or MTE
independent repeated measurements can be derived from that methods (Test Methods C1672 and C1832, respectively).
by using t = 1, so the standard error (SE) is given by SD / √n.
5. Significance and Use
3.3.21 TIMS—thermal ionization mass spectrometry
5.1 Uranium material is used as a fuel in certain types of
3.3.22 WRM—working reference material
nuclear reactors. To be suitable for use as nuclear fuel, the
4. Summary of Test Method starting material shall meet certain specifications such as those
described in Specifications C753, C776, C787, C833, C967,
4.1 The double spike method has been developed with the C996, and C1008, or as specified by the purchaser. The
intention to improve the precision and decrease the uncertain- 235
U/238U isotope amount ratios and the amount content of
ties for 235U/238U major isotope ratio measurements compared uranium material can be measured by mass spectrometry
to the known methods such as the “classical” total evaporation following this test method to ensure that they meet the
technique (1-4),4 also described in Test Method C1672, and the specification.
modified total evaporation technique (5 and 6), also described
in Test Method C1832. For the double spike method the mass 5.2 The double spike method has been used for studies of
fractionation correction for the 235U/238U ratio is performed uranium fractionation effects in isotope geochemistry and
internally throughout the measurement rather than externally, cosmochemistry, for uranium source attribution in nuclear
by using the mass fractionation observed for a double spike forensics and for investigation of conversion or sampling
material with a known or certified 233U/236U isotope ratio (also processes in nuclear industry and nuclear safeguards (7-11).
spanning three mass units), which is mixed with the sample Most recently, the double spike method has been used for the
prior to the measurement, either during the sample preparation validation of the Cristallini sampling method of UF6 (12 and
or directly on the TIMS filament. If necessary, uranium is 13). The double spike method can be used for a wide range of
separated from plutonium and other elements (to eliminate sample sizes even in samples containing as low as 50 µg of
isobaric interferences) by selective extraction, anion exchange uranium. The concentration of the loading solution for the DS
(see Practice C1411), or extraction chromatography. The puri- method has to be in the range of 1 to 6 mg/g to allow a sample
fied uranium fraction as nitrate solution is loaded onto an loading of 4 to 6 µg of uranium. A minimum loading of 4 µg
evaporation filament (made of metals such as rhenium, zone- uranium per filament is recommended.
refined rhenium, or tungsten with high evaporation 5.3 The measurement of 236U/238U ratios using this method
temperature), and blended with an appropriate amount of is not possible due to the large isobaric interference from the
236
double spike solution, and converted to an oxide by controlled U ion beam of the double spike onto the 236U ion beam from
heating of the filament under atmospheric conditions. In case the sample (>50.000 times for close to natural material, for
only the 235U/238U ratio of the sample has to be determined, it example, like IRMM-184).
is recommended to mix the sample with the double spike 5.4 The application of the double spike method for mea-
during the loading process on the filament. In case the uranium surements of 235U/238U ratio is limited by the isobaric inter-
amount concentration of the sample has to be determined, the ference between the 236U from the double spike material and
sample solution has to be blended gravimetrically with the the 236U contained in the sample. As a consequence, the
double spike solution prior to filament loading, for which method is not suitable for samples which contain significant
weighable amounts have to be used. amounts of 236U due to prior neutron capture from 235U in the
4.2 The sample amount to be loaded for DS analyses is predecessor materials. For samples with 236U/238U ratios
within a range of about 4 to 6 µg to achieve ion beam signals higher than about 10–6, the double spike method should be
of about 20 to 30 V for the major isotope 238U for DU, NU, and applied with care for the isobaric correction. For an appropriate
LEU samples. isobaric correction, the 236U/238U ratios of the samples should
4.3 The 235U/238U isotope amount ratios are corrected for be determined separately using a suitable measurement
mass fractionation for each integration step individually. This method, for example, the modified total evaporation MTE
is accomplished in an internal manner, the magnitude of the method (Test Method C1832, Ref (5) and (6)).
mass fractionation is calculated from the measured mass 5.5 The measurement of 234U/238U ratios using this method
fractionation of the 233U/236U ratio. The peak tailing contribu- is very limited in the analytical performance due to the isobaric
tions are determined at two mass positions, 0.5 mass units interference of the 234U from the double spike with the 234U
below and 0.5 mass units above the isotope masses of interest. from the sample (range from 5 to 15 %). The correction
4.4 For the correction of isobaric interferences, a separate algorithms are presented in 14.3, but statements for precision
measurement of the isotopic composition of the (unspiked) and bias are not given. Other methods like MTE (Test Method
sample is required, unless this information is already available. C1832, Ref (5) and (6)) are better suited and more reliable for
measurements of 234U/238U ratios.
5.6 The DS method described here can also be extended to
4
The boldface numbers in parentheses refer to a list of references at the end of measurement of elements other than uranium, if a suitable
this standard. double spike material is available.

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG )UL-DQ*07


3
'RZQORDGHGSULQWHGE\
8&5 8QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
C1871 − 22
6. Interferences 7.1.3 A multiple Faraday collector system to allow simulta-
238
6.1 Isobaric nuclides such as Pu interfere in the uranium neous detection of isotope beams from m/z = 233 to 238 for U+
measurements. The removal of interferences is generally ac- ions;
complished by chemical separation leading to ionization of 7.1.4 For the Faraday cups used to measure the major ion
uranium only and improved precision of measured isotope beams of 235U and 238U, there shall be current amplifiers
amount ratios. equipped with 1011 Ω resistors and, for the ion beams of 233U
and 236U, there shall be current amplifiers equipped with at
6.2 For the isotopes 234U, 235U, 236U, and 238U isobaric least 1011 Ω, and preferably 1012 Ω resistors to improve the
interferences between the sample and the double spike occur. signal-to-noise ratio. In case the ion beam of 235U is below the
The accurate correction of isobaric interferences is a prerequi- ion beams of 233U and 236U, the use of an amplifier equipped
site for obtaining results for the DS method with satisfactory with a 1012 Ω resistor is recommended, if available.
precision and accuracy. The corrections depend strongly on the 7.1.5 A sample turret to allow automatic measurement
mixing proportions between the sample and the double spike. sequences of several replicate filament loadings per sample and
Usually the sample/spike amount ratio is adjusted in such a per quality control standard (preferentially a CRM);
manner that the ion beam intensity at mass m/z=236 is lower 7.1.6 A pumping system that is able to attain a vacuum of
compared to that at mass m/z=235, in order to reduce the <4.0 × 10–5 Pa (3 × 10–7 torr) in the ion source, the analyzer,
influences from low-mass tailings, that is, abundance and the detector is required. Tailing corrections are dependent
sensitivities, at one mass unit below and beyond the U peaks. on the vacuum levels inside the mass spectrometer. Analyzer
Therefore a sample-to-spike amount ratio of >20 is recom- pressures below 7.0 × 10–7 Pa (5 × 10–9 torr) are preferred;
mended. The correction of isobaric interferences is explained 7.1.7 A mechanism to scan masses by varying the magnetic
in detail in Section 14 (“calculations”) of this test method. field or the accelerating voltage or both;
6.3 It has to be ensured that samples are not contaminated 7.1.8 A computer for control of the data acquisition accord-
by environmental uranium. The level of effort required to ing to a predefined sequence.
minimize contamination shall be based upon the sample size 7.2 Special MTE Capabilities—It is recommended to have a
and the levels of contamination present in the analytical mass spectrometer software to be flexible enough to implement
facility. For extremely small samples or extremely low 236U a user-defined filament-heating program as for MTE.5
abundances, residual uranium from chemicals used for sample
dissolution and sample preparation is possible source for bias 7.3 A separate filament degassing device for cleaning of the
in the isotopic data. filaments before sample loading can be used.

6.4 Samples shall be chemically purified to assure reliable 7.4 A pipette or microsyringe to transfer microliter volumes
analyses by TIMS. Impurities, especially alkali elements, of solution.
produce unstable ion emission leading to poor precision in the 7.5 A separate filament heating device for drying and
isotope amount ratios. Organic contaminants or oxide layers on oxidizing the sample on the filament ribbon after loading.
the filaments also adversely influence TIMS analyses. Isobaric
interferences, if not removed, will bias the isotope amount 8. Reagents and Materials
ratios. Contaminants in reagents, lab ware, or filament material 8.1 Purity of Reagents—Ultra-high purity reagents shall be
are also sources for bias in the isotope amount ratios. used for processing small sample amounts or samples with
6.5 The performance of the instrument can be adversely extremely small isotope amount ratios. The level of uranium
affected by changes in the environmental conditions of the contamination from chemicals, water, and the sample handling
laboratory, that is, temperature and humidity. For this reason, environments shall be determined to ensure that the materials
controlled laboratory environmental conditions should be and the environment are sufficiently pure for the samples being
maintained (within the manufacturer’s specifications) during analyzed.
instrument operation. 8.2 Nitric Acid (HNO3, 15.8 M)—Concentrated nitric acid.
7. Apparatus 8.3 Nitric Acid (HNO3, 1 M)—One volume of concentrated
nitric acid (HNO3, 15.8 M) brought to 15.8 volumes with
7.1 The suitability of the mass spectrometer for carrying out water.
measurements by the DS method shall be evaluated by means
of performance tests. The relevant instrument characteristics 8.4 Purity of Water—Unless otherwise indicated, references
are as follows: to water shall be understood to mean laboratory accepted
7.1.1 A thermal ionization source for using single, double, demineralized or deionized water as described by Type I of
or triple filament assemblies with rhenium or tungsten Specification D1193.
filaments, or both;
7.1.2 A mass analyzer sufficient to resolve adjacent masses
5
in the mass-to-charge range being studied, m/z = 233 to 238 for The sole source of supply of the apparatus known to the committee at this time
is Thermo Fisher Scientific Inc., 81 Wyman St., Waltham, MA 02451. If you are
U+. Resolution shall be greater than 350 (full width at 1 % of
aware of alternative suppliers, please provide this information to ASTM Interna-
peak height) and the abundance sensitivity at mass 237 for ions tional Headquarters. Your comments will receive careful consideration at a meeting
of 238U less than 8 × 10–6; of the responsible technical committee,1 which you may attend.

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG )UL-DQ*07


4
'RZQORDGHGSULQWHGE\
8&5 8QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
C1871 − 22
8.5 Filaments—Filaments with sufficient purity, for 11. Preparation of Apparatus
example, zone refined or high purity Re or W, should be used.
11.1 Filament Degassing—Filaments can be degassed be-
The size and configuration of the filament is instrument
fore using them for DS measurements. This is recommended in
dependent.
case the heating of the filaments during the mass spectrometric
8.6 Liquid Nitrogen—Liquid nitrogen can be filled into the measurement is causing the vacuum pressure in the ion source
cold trap of a mass spectrometer to improve the vacuum to increase beyond 1.3 × 10–4 Pa (1 × 10–6 torr) is reached.
pressure. Recommended filament currents for degassing are in the range
4.0 to 5.0 A, corresponding to temperatures in the range of
9. Hazards 1700 to 2000 °C. Perform the degassing for a duration of at
9.1 TIMS instruments operate at 8 to 10 kV electrical least 30 min.
potential. Ensure that the high-voltage is switched off before 11.2 “Initialization” of the Sample Turret—Adjust, if
insertion or removal of the sample turret into/from the needed, the position of the sample turret in the ion source and
instrument, and working with the ion source, or accessing other verify proper electrical connections for both the evaporation
electronic components. and ionization filament for each sample loading position. It
9.2 The filaments reach temperatures in excess of 2000 °C. shall be ensured that the electrical contact is not interrupted in
The filament holders, sample wheel, and ion sources parts are case the turret is slightly moved for the purpose of ion beam
expected to be hot. Ensure that a sufficient time has lapsed focusing.
since the last filament heating before accessing the filaments,
sample turrets, and ion source. 11.3 Electronics Test—Modern mass spectrometric instru-
ments offer an automated routine for testing the stability and
9.3 Wear eye protection and suitable gloves when filling performance of the electronic systems (for example, Faraday
cold traps with liquid nitrogen. Protect hands, torso, and feet in cup amplifier baselines and gains, high-voltage unit, and
the event of splashing or spilling of the liquid nitrogen. magnet current supply units). A report is produced flagging
9.4 Handle radioactive materials with appropriate attention systems or components that are out of specification. Users of
to radiological safety. the instruments should perform routine checks of the perfor-
9.5 Handle chemical hazards and toxins like uranium with mance of the electronic systems and ensure that the perfor-
appropriate care. mance is within manufacturer’s specifications. The frequency
for this test shall be established by the user based on manu-
10. Sampling, Test Specimens, and Test Units facturer recommendations or as specified in the user’s quality
assurance program.
10.1 Isotope Reference Materials—Uranium reference ma-
terials used in the analysis should be prepared from certified 11.4 Amplifier Signal Decay Adjustment—Adjust the signal
reference materials (CRMs) traceable to SI units. Examples decay characteristics of the current amplifiers of the Faraday
include the Double Spikes IRMM-3636 (1 mg U/g), IRMM- cups. This is important for measuring isotope amount ratios
3636a (0.1 mg U/g), and IRMM-3636b (0.01 mg U/g), which with a large dynamic range, high precision and accuracy, or
all have the same isotopic composition (uncertainties with both. Depending on the combination of the capacitance and
coverage factor k = 2, 95 % confidence level): resistance of the current amplifier, the response time for a
233
U/236U = 1.01906(16),
sudden change in the ion beam signal to the Faraday cup can
234
U/236U = 0.00036606(48), reach up to 5 s. For amplifier resistances higher than the 1011
235
238
U/236U = 0.000045480(74), and Ω, longer response times of about 15 s can be expected. The
U/236U = 0.00023481(38).
DS method has to be designed by taking into consideration the
In addition, various in-house double spike materials are required response times. The amplifier response can be
being used at numerous facilities, which have been calibrated checked either using a custom-made software module within
using isotope reference materials, like NBL CRM U500 the operating software (when available), or “manually” by
(235U/238U > 1) or IRMM-184 (close to natural U). For the means of a large ion beam that is abruptly directed into a
preparation of in-house double spike reference materials, see Faraday cup to check the signal ingrowth time, or away from
Guide C1128 for additional guidance on preparation of trace- a Faraday cup (for example, by closing a valve between ion
able working reference materials. source and analyzer) to check the signal decay time.
10.2 Uranium samples to be measured and isotope reference 11.5 Ion Source and Analyzer Pressure—It is important to
materials used for quality control purposes shall be in the same achieve a certain level of vacuum before the isotope amount
medium, same concentration and in the same oxidation state. ratio measurements can be started; see 7.1.6 for the recom-
The solutions for loading onto the filaments are usually 1 to 5 mended pressure. The peak tailing depends strongly on the
M nitric acid solutions. The loading and drying sequence of the vacuum pressure in the detector system since the number of ion
filaments shall also be similar. collisions with gas molecules inside the mass spectrometer is a
10.3 In the case of characterization studies of test materials, direct function of the ambient pressure. An increase of the
possible inhomogeneity between test units shall be evaluated pressure within the ion source caused by the heating of the
statistically and included in the uncertainty calculations of the ionization and evaporation filaments can be subdued, to a
isotope amount ratios assigned. certain extent, by using a cold trap filled with liquid nitrogen.

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG )UL-DQ*07


5
'RZQORDGHGSULQWHGE\
8&5 8QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
C1871 − 22
12. Calibration and Standardization performed prior to the Faraday cup efficiency test. The Faraday
NOTE 1—The measurement method may be qualified following Guide cup efficiency test can be performed in several ways, as
C1068 and calibrated following Guide C1156. described in 12.5.1 – 12.5.4.
12.1 Mass Calibration—The relationship between the 12.5.1 The calibration may be performed by switching a
known atomic masses and the magnetic field necessary to stable ion beam of 187Re (from a blank filament) between each
direct the isotope beams into the detectors shall be updated on Faraday cup and a reference Faraday cup. In case a relative
a regular basis. Mass calibration shall be performed at intervals efficiency between the detectors is significantly different from
specified by the manufacturer or the user’s quality assurance unity, this result can be used to correct for differences in the
program. detector response. This procedure can be performed with a
relative uncertainty at the level of <0.1 %.
12.2 Peak Centering—The peak centering routine is used as
a fine adjustment to ensure that the ion beam is centered within 12.5.2 A series of peak-jumping measurements between all
the detector. Peak centering usually occurs via fine adjustments Faraday cups and a reference cup to be checked can also be
of the accelerating voltage, and any difference between the performed using a sufficiently large uranium sample and one
value optimized during peak centering from the default accel- large stable ion beam, for example, a 10 to 20 V ion beam of
238
erating voltage requires a readjustment of the mass calibration U from a LEU or natural uranium sample. The drift of the
curve. Peak centering shall be performed for at least at three signal intensity shall be corrected for using the operating
uranium masses as part of the mass calibration before the start software. This procedure can be performed with a relative
of each DS measurement sequence. During the DS uncertainty at the level of <0.01 %.
measurement, peak centering is performed on a regular basis. 12.5.3 A series of comparative neodymium (Nd) isotope
amount of ratio measurements can be performed in two
12.3 Amplifier Baseline Calibration—The baselines of the different modes such as the multi-dynamic mode and the static
Faraday cup amplifiers, that is, the amplifier responses without mode with “amplifier rotation” (only for TRITONTM TIMS,
incoming ion beam to the cup, shall be measured on a regular also called “virtual amplifier”: each Faraday cup is connected
basis and checked for stability. During the DS measurements, to each amplifier for regular time intervals during the measure-
baseline measurements are performed on a regular basis. Note ment). This procedure can be performed with a relative
that the integration time for the baseline measurement has a uncertainty at the level of few ppm (5). It shall be repeated
significant influence on the uncertainty of Faraday cup until all Faraday cups of interest for DS measurements have
measurements, particularly at lower ion beam intensities. been included.
Therefore, the integration time of the baseline (within a 12.5.4 A series of static measurements can be performed
measurement) shall be comparable to the integration time of using special “multi-isotope” reference materials, such as
the actual ion beam signal integration. The long-term historical IRMM-3100a (233U/235U/236U/238U=1/1/1/1), IRMM-072/1,
baseline data shall be regularly reviewed by the user to assure IRMM-074/1, or IRMM-199 (233U/235U/238U=1/1/1), to in-
that the system performance is within manufacturer specifica- clude all Faraday cups. This procedure can be performed with
tions and quality system requirements. relative uncertainties of about 0.03 %.
12.4 Amplifier Gain Calibration—The stability and re- 12.6 Linearity Test—There are various procedures to check
sponse of individual Faraday detector amplifiers shall be the linearity of an isotope mass spectrometer detection system.
measured and differences between amplifiers corrected for via The procedures described in 12.6.1 and 12.6.2 are mainly
the amplifier gain calibration. Gain calibration is normally applicable for Faraday multi-collector systems.
performed by sequentially applying a stable calibration current 12.6.1 The linearity of the mass spectrometer is determined
to the input of each Faraday cup amplifier and the output is over the working range of the Faraday cups by measuring the
then normalized to a reference value to generate a gain 235
U/238U ratios of various reference materials under identical
calibration factor for each amplifier. A gain calibration shall be conditions. The mass spectrometer system is linear if the K
performed prior to each automatic DS sequence. Historical factor, that is, the ratio of the certified 235U/238U ratio to the
gain calibration data can be used to evaluate the stability of the measured 235U/238U ratio, is independent of the isotopic
amplifiers. composition of the material. For this procedure, the NBL
12.5 Faraday Cup Effıciency Test—The response of indi- U-series of reference materials (U005a to U970, 0.5 to 97 % of
235
vidual Faraday cups depends on several factors, for example, U) is ideal and can be combined with the IRMM-183-187
extent of usage, manufacturing variability, and can also be series (0.3 to 5 % of 235U) and the IRMM 019-029 series (0.17
affected by an insufficient electron suppression voltage. The to 5 % of 235U, to be converted from UF6). This procedure shall
relative response of the Faraday cups, therefore, shall be be performed sequentially for all Faraday cups of the multi-
determined periodically. Usually, the Faraday cups of a multi- collector system needed for the DS analyses.
collector system are only intercalibrated for the current ampli- 12.6.2 The IRMM-072 and IRMM-074 series of reference
fiers connected to them (see 12.4) but not for the differences in materials are characterized by 238U/235U ratios of ≈1 and
233
the efficiencies of the Faraday cups themselves. The efficien- U/235U ratios ranging from ≈1 down to ≈10–6 for the 15 or
cies of the Faraday cups are expected to be similar to each 10 units, respectively, of the used series. For each unit, the bias
other, which means that the relative efficiencies (relative to one of the measured 238U/235U ratios from the certified ones can be
reference cup) are normally close to unity. Note that an used for internal mass fractionation correction of the measured
233
(electronic) amplifier gain calibration (see 12.4) shall be U/235U ratios. The comparison of the corrected 233U/235U

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG )UL-DQ*07


6
'RZQORDGHGSULQWHGE\
8&5 8QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
C1871 − 22
ratios with the certified ones allows the linearity of the 13.1.3 Sample Purification—Use Practice C1411 or similar
detection system to be checked over a dynamic range of six procedure to separate uranium from plutonium and other
orders of magnitude for the ion beam intensity. A detailed impurities, if necessary.
description of the procedure is given in (14). This procedure 13.2 Sample Loading and Conditioning—Samples for the
shall be performed sequentially for all Faraday cups needed for DS method are usually directly loaded on the filament by drop
the DS analyses. deposition. Samples and QC materials shall be prepared for
12.7 Peak Overlap—When a Faraday multi-collector sys- analysis by the same method at similar mass loadings. Drop
tem for the simultaneous detection of several masses is used, it deposition onto the filament can be accomplished with the use
needs to be ensured that the peak overlap is acceptable. A mass of a microsyringe or pipette fitted with a plastic tip. Change the
scan, usually by scanning the magnetic field, shall be per- tip between sample loadings to prevent cross contamination. In
formed by which all ion beams are simultaneously moved particular for loading and mixing a sample solution and the
through the respective cups. The measured intensities for all double spike solution on the same filament, cross contamina-
detectors shall be plotted versus the mass of a reference tion between these solutions has to be prevented by changing
detector to make the peak overlap visible. All peaks shall have pipette tips. It is recommended to load the double spike
a symmetric shape with a common flat region in the center, solution first and the sample afterwards in order to minimize
with the peak centers reasonably close together, as specified by the risk of contaminating the double spike solution with any
the manufacturer or the user quality system. After a satisfactory sample material. For filaments loaded by drop deposition, the
peak overlap is realized (by moving cups relative to one solution shall be evaporated by passing sufficient electrical
another if necessary), the positions of all detectors shall be current through the filament to cause gentle drying without
saved, for example, as a Faraday cup configuration file. The boiling. Samples for DS are usually prepared in a 1 to 5 M
positions shall be checked and possibly readjusted, manually or nitric acid with a uranium concentration between 1 to 6 mg/g,
using stepping motors, as needed before a new automatic which is equivalent to 1 to 6 µg/µL. Depending on the uranium
measurement sequence. amount to be loaded, more than 1 µL of the sample solution
12.8 Mass Fractionation Correction: may be needed. The recommended uranium amount for load-
12.8.1 For the double spike method the mass fractionation ing is 4 to 6 µg. Deposit drops very carefully and slowly. It is
correction for the 235U/238U ratio is performed for each recommended to keep the filament heated by passing 0.5 to
integration step (called “mass cycle,” see below) internally 0.7 A current and depositing 1-µL-drops at a time. After all
throughout the measurement rather than externally, by using drops are loaded, the solution on the filament is heated until
the mass fractionation observed for a double spike material dryness, for at least one more minute, and then heated for
with a known or certified 233U/236U isotope ratio (also span- several seconds at a higher current of 1.5 to 2.0 A for
ning three mass units). conditioning. Alternatively, a stepped-heating program can be
12.8.2 The mass fractionation correction factor, K, is calcu- used to condition samples, that is, to convert samples to
lated as follows: suitable chemical forms. Avoid quick evaporation of the
sample or melting the filament. At different facilities, different
K 5 ~ R c ⁄ R m! (1)
loading and conditioning procedures have been established and
where: validated. Each procedure shall be applied in a consistent
K = mass fractionation correction factor, manner for all samples, and quality control samples.
Rm = measured 233U/236U isotope amount ratio for the 13.3 Mount all sample filaments and ionization filaments on
double spike, and a sample turret and insert the sample turret into the ion source
Rc = known or certified 233U/236U isotope amount ratio of the mass spectrometer.
value for the double spike.
13.4 Close the source and start evacuating.
Details about the mass fractionation and further correction
algorithms are presented in 14.2. 13.5 Perform adjustment (also called initialization) of the
sample turret if needed. In case of problems with electrical
13. Procedure connections, the source might have to be vented again to
13.1 Sample Preparation: resolve the problem.
13.1.1 Sample Dissolution—Dissolve an appropriate sample 13.6 Evacuate ion source to the manufacturer’s recom-
amount to obtain the desired filament loading solution for the mended minimum pressure or according to the user’s proce-
mass spectrometric analysis. See Practice C1347 for the dure.
dissolution of uranium.
13.7 Add liquid nitrogen to the cold trap if desired.
13.1.2 Prepare the sample and any reference material solu-
tions as purified nitrates, using identical chemical preparation 13.8 Isotope Amount Ratio Measurement—The following
and handling steps. For uranium samples hydrolyzed from steps 13.8.1 – 13.10 are typically performed automatically
uranium hexafluoride, it is recommended that the samples are under computer control depending upon the instrument.
converted to U3O8 before dissolution in nitric acid and analy- 13.8.1 Perform an amplifier gain calibration for each new
sis. The solution concentrations shall be chosen to allow for a automatic sequence.
convenient filament loading (for example, a 2-mg U/mL 13.8.2 Measure the baseline during the course of each
solution yields 2 µg of uranium per µL, see also 13.2). measurement of a sample and QC standard.

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG )UL-DQ*07


7
'RZQORDGHGSULQWHGE\
8&5 8QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
C1871 − 22
13.8.3 Usually under computer control, the ionization fila- 13.9 Interblock Actions and Filament Heating for DS:
ment is heated up to a temperature of about 1800 to 1950 °C. 13.9.1 Peak Centering and Baseline—Each measurement
For automatic sequences, usually the magnitude of the 187Re block usually consists of a maximum of five mass cycles, with
ion beam is used as the regulated quantity instead of the each cycle having a duration of about 60 to 90 s. Before each
filament temperature by a pyrometer because only the narrow block, a peak centering is performed using the 238U isotope.
sides of the filaments are oriented towards the pyrometer. Additionally, an ion source focusing is performed before each
Typical 187Re ion beams of about 300 to 600 mV (on a 1011 Ω block. Typically, every five blocks, the electronic baselines of
resistor) are used for non-zone refined filaments and lower all the Faraday cup amplifiers are re-measured. In case 1012 Ω
values of 150 to 400 mV are used for zone-refined filaments, resistors are used in some of the amplifiers, the idle times for
depending on the thickness and brand. the baseline measurements have to be sufficiently long, for
13.8.4 The 187Re ion beam is peak centered and focused. If example, at least 15 s.
necessary, the 187Re ion beam size is readjusted towards the 13.9.2 DS Interblock Heating—The DS interblock heating
target value, which shall be similar (within about 20 %) for is recommended to be performed similar to the MTE interblock
each measurement within an automatic sequence, by changing heating as described in Test Method C1832. The filament
the filament current. heating is controlled by a special program script executed after
13.8.5 The sample filament is heated to a temperature each measurement block. This program script first measures
sufficient to yield and ion beam sum intensity from 235U and the actual sum intensity (mainly from 235U and 238U) and
238
U of about 1 to 4 V (on an amplifier with a 1011 Ω resistor). compares it with the predefined (initial) target sum intensity.
The beam is focused and peak centered using the largest U ion This predefined target sum intensity is typically in the range of
beam, which is usually the 238U beam. 20 to 30 V, selected by the operator depending on the sample
13.8.6 Data acquisition is started. The data acquisition is amount loaded, the experience with the ion source and the
performed on a per block basis in which each block consists of instrument transmission. Based on the comparison of the
a minimum of two and a maximum of five mass cycles and actually measured sum intensity with the target sum intensity,
each mass cycle of (usually) four integration steps. The data the following filament heating step is calculated. A detailed
acquisition is usually continued until the whole sample is description of the interblock heating script is given in Test
evaporated; see 13.10. Method C1832.
13.8.7 DS Mass Cycle—The DS mass cycle consists of four 13.10 Termination of DS Measurement—The DS measure-
steps as shown in Table 1. Depending on the ion beam ment is terminated when the sum intensity is below a user-
intensities observed the use of amplifiers with 1012 Ω resistors defined limit of 1 to 2 V.
is recommended. For the recommended sample-to-spike
amount ratios >20, the 233U and 236U intensities coming from 14. Calculations
the double spike shall be detected using amplifiers with 1012 Ω 14.1 Within this section, the DS data evaluation and calcu-
resistors, if available, in case of depleted samples with low lation of the various correction factors is discussed. The mass
235
U ion beams this applies as well. fractionation correction for the major ratios 235U/238U is
13.8.7.1 The first step is the main integration for all iso- performed internally for each mass cycle throughout the
topes; the integration time is usually about 16 s. measurement based on the results of the 233U/236U ratio using
13.8.7.2 In Steps 2 and 3 of the DS mass cycle, the the exponential fractionation law. The differences between
background intensities at the low and high mass side of all different fractionation laws, for example, power law, exponen-
uranium isotopes of interest are measured. The mutual back- tial law, linear law, or Rayleigh law are at the level of few parts
grounds for 235U and 236U are the most critical ones and per million for heavy elements like uranium, which is insig-
depend a lot on the sample to spike ratio. Due to the differences nificant for the DS measurements.
in the tailing effect between the low and high mass sides, an 14.2 Evaluation for Major Ratio 235U/238U:
interpolation for the isotope ratios shall be performed, using a 14.2.1 The K factor is derived from the 233U/236U ratio for
linear approximation. each mass cycle:
13.8.7.3 Step 4 is an optional “dummy” step, which might
be needed to move the mass setting back to the main cup
configuration (as in step 1) for checking the ion beam signals K5
S D
233

236
U
U DS
(2)
at the very end of each block. This is needed to determine the
ion beam intensity before the next heating step (see 13.9).
S D
233

236
U
U
ME

TABLE 1 Mass Cycle for DS


A A
Cup L2 L1 C (FAR)A H1A H2A Idle Time/
Measurement Time
234 235 236 238
Step 1 U U U 237 u U 12 s/16 s
234 235 236 238
Step 2 U – 0.5 u U – 0.5 u U – 0.5 U – 0.5 u 5 s/8 s
234 235 236 238
Step 3 U + 0.5 u U + 0.5 u U + 0.5 u U + 0.5 u 5 s/8 s
234 235 236 238
Step 4 U U U 237 u U 3 s/1 s
A
The acronyms L2, L1, C (FAR), H1, and H2 indicate the names of the five Faraday cup detectors of the Faraday multi-collector system, which are typically used for DS
measurements.

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG )UL-DQ*07


8
'RZQORDGHGSULQWHGE\
8&5 8QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
C1871 − 22
where:
ME = measured ratio (not corrected for background and
S D S D S D S SS
238
Ui
238
U ME
5
238
U
233
U DS
233

238
U
U ME
1 2
232.55
233
U D ME

isobaric interferences, for details about the corrections


see below), and S D DD 1
233.55
233
U ME
⁄2 (6)
DS = known ratio of the double spike, preferentially a
certified reference material. Where (238U/233U)DS is the 238U/233U ratio of the double
spike, which can be calculated from the known or certified
14.2.2 The K factor is applied in the following way: (238U/236U)DS and (233U/236U)DS ratios given on the certificate
S D
235

238
U
5 K R235 S D 235
U
238 (3)
using
U CO
U ME
S D 238
U
where:
S D 238
U
5
236
U DS
(7)
CO = corrected ratio, and
ln~ 235 ⁄ 238!
S D
233
U DS
233

236
U
U
DS
the exponent R235 is given by R2355 ln 233 ⁄ 236
~ ! 14.3.4 The relative isobaric contribution (235Ui/235U)ME
>0.991543, according the exponential fractionation law. from the double spike is given by
14.2.3 The exponent R235 is given by R235 = 1 in case the
linear fractionation law is preferred as an approximation. S DS D 235
U 233
U

S D
235
Ui
5
233
U DS
238
U ME
(8)
14.3 Background and Isobaric Corrections—All back-
ground and isobaric corrections of the isotope ratios are
235
U
S D ME
235

238
U
U
ME
performed internally on a per mass cycle basis for each 235 233 235 233
measurement. Where ( U/ U)DS is the U/ U ratio of the double
14.3.1 The measured 233U/236U ratio is corrected for back- spike, which can be calculated from the certified (235U/236U)DS
ground and the relative isobaric contribution (236Ui/236U) from and (233U/236U)DS ratios given on the certificate using:
the sample:
S D 235
U

S D S D
233
U 233
U S D 235
U
5
236
U DS
(9)
236
U CO
5 236
U ME
·
S D
233
U DS
233

236
U
U

S SS D S D D D 1 2
232.55
233
U ME
1
233.55
233
U ME
⁄2
14.3.5 The relative isobaric contribution (234Ui/234U)ME
DS

S SS D S D D S D D
1 2
235.55
236 1
236.55
236 ⁄2 2
236

236
Ui from the double spike is given by
U ME
U ME
U ME
S DS D 234
U 233
U
(4)
S D
234
Ui
5
233
U DS
238
U ME
(10)
where:
ME = measured ratio, and
234
U
S D ME
234

238
U
U
ME

CO = corrected ratio. 234 233 234 233


Where ( U/ U)DS is the U/ U ratio of the double
14.3.2 The relative isobaric contribution (236Ui/236U)ME spike, which can be calculated from the certified (234U/236U)DS
from the sample is given by and (233U/236U)DS ratios given on the certificate using

S D S D 236
U 233
U
S D 234
U

S D
236
Ui 238
U Sample
236
U ME
S D 234
U
5
236
U DS
(11)
236
U
S DME
5 233

238
U
U
ME
S D
233
U DS
233

236
U
U
DS

S SS D S D D S D D
1 2
237.55
238
U ME
1
238.55
238
U ME
⁄2 2
238

238
Ui
U ME
14.3.6 The K factor for a double spike measurement is
derived from the certified (233U/236U)DS ratio for each mass
S SS D S D D D 1 2
235.55
236
U ME
1
236.55
236
U ME
⁄2
cycle according to Eq 2 by using the (233U/236U)CO ratio from
Eq 4 which is corrected for background and isobaric interfer-
(5) ences:
Where (236U/238U)Sample is the 236U/238U ratio of the sample,
which should be approximately known from separate mea-
surements (for example, using the MTE method, Test K5
S D 233

236
U
U DS
(12)
Method C1832); and (238Ui/238U)ME is the relative isobaric
contribution from the double spike.
S D 233

236
U
U
CO

238 238
14.3.3 The relative isobaric contribution ( Ui/ U)ME 14.3.7 The K factor is applied in the following way to obtain
from the double spike is given by the corrected 235U/238U ratio of the sample:

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG )UL-DQ*07


9
'RZQORDGHGSULQWHGE\
8&5 8QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
C1871 − 22

S D 235
U
238
U CO
S D
5 K R235
235

238
U
U ME
Expression of Uncertainty in Measurements (GUM) (15). For
DS measurements, the uncertainty evaluation steps are ex-

S SS D
1 2
234.55
235
U ME
1 S 235.55
235
U D D S D D
ME
⁄2 2
Ui
U
235

235
ME
plained in the following.
235
14.4.2 For the uncertainty of the corrected major ratio
U/238U calculated according to Eq 12 and Eq 13, the relative
S SS D
1 2
237.55
238
U ME
1 S 238.55
238
U D D S D D
ME
238
⁄2 2 238
Ui
U ME
uncertainties of the measured 233U/236U ratios of the double
(13) spike material used for mass fractionation correction and the
measured 235U/238U ratios of the sample, as well as the relative
234 238
14.3.8 The corrected U/ U ratio of the sample can be uncertainty of the known or certified 233U/236U ratio of the
obtained in the following way: double spike, are added in quadrature to obtain the relative
uncertainty of the correct 235U/238U ratio of the sample.
S D 234
U
238
U CO
S D
5 K R234
234

238
U
U ME
Because of the good repeatability of DS measurements (about
0.003 to 0.006 % RSD, only slightly dependent of the value of
S SS D
1 2
233.55
235
U ME
1 S 234.55
235
U D D S D D
ME
⁄2 2
Ui
U
234

234
ME
the ratio), this uncertainty is largely dominated by the uncer-
tainty of the known or certified ratio of the double spike
S SS D
1 2
237.55
238
U ME
1 S 238.55
238
U D D S D D
ME
238
⁄2 2 238
Ui
U ME
(typically in the range of 0.015 to 0.03 %). Typically values for
the expanded uncertainty (coverage factor k = 2) are about
(14)
ln~ 234 ⁄ 238! 0.016 to 0.020 % (see 16.1 for additional details).
The exponent R234 is given by R2345 ln 233 ⁄ 236 >1.32487, 14.4.3 The uncertainty contributions of the corrected ratio
~ ! 235
which is close to 4⁄3 . U/238U ratio arising from the background and isobaric
corrections are negligible, if recommended sample/spike
14.3.9 Quality Control for the Corrected Ratio (233U/
236 amount ratios of >20 are applied.
U)CO—In case the entire sample is evaporated and used for
14.4.4 For the uncertainty of the corrected ratio 234U/238U
the data acquisition using an MTE heating script (see 4.1 and
calculated according to Eq 14, the uncertainty from the isobaric
13.9.2), it is recommended to calculate a weighted mean of the
correction due to 234U stemming from the double spike
corrected (233U/236U)CO ratios derived using Eq 4, as usually
material may become the dominant source of uncertainty.
done according to the total evaporation principle (see Test
Methods C1672 and C1832): 14.5 Calculation of Isotope Amount Fractions (“Atom
Percent”), Isotope Mass Fractions (“Weight Percent”), and
( FS D233
U
· I ~ 236U ! G Molar Mass (“Atomic Weight”):
S D
233

236
U
U
CO2TE
5
ALL_INTEGRATIONS

(
236
U CO

I~ 236
U!
(15) 14.5.1 The isotope amount fractions (also called “atom
percent”), isotope mass fractions (also called “weight
ALL_INTEGRATIONS
percent”), and the molar mass can only be calculated if all
where: isotope ratios 234U/238U, 235U/238U, and 236U/238U of the
I(236U) = signal intensity (in V) for 236U measured sample are known. The 234U/238U and 236U/238U ratios of the
on a Faraday cup, corrected for the sample have to be measured using a method other than the DS
Faraday cup amplifier gain and baseline, method, for example, by TE (Test Method C1672) or MTE
and (Test Method C1832).
(233U/236U)CO-TE = mean for the (233U/236U)CO ratios ac- 14.5.2 Isotope Amount Fraction for Isotope i (i = 234, 235,
quired for all mass cycles during a DS 236, 238):
measurement, weighted by I(236U).
n ~ iU ! n ~ iU ! Ri
The results for (233U/236U)CO-TE are expected to be in mutual A ~ iU ! 5 5 5 (16)
agreement for all measurements on individual filaments as well
n~U! (
j5234, 235, 236, 238
n ~ jU ! (
j5234, 235, 236, 238
Rj

as for all different samples within the same sequence, because


14.5.3 Isotope Mass Fraction for Isotope i (i = 234, 235,
they should be independent on the sample/spike ratio on
236, or 238):
individual filaments and also on the isotopic composition of the
different samples. If this is not the case, there could be a m ~ iU ! R iM i
W ~ iU ! 5 5 (17)
mistake in the application of some of the isobaric corrections m~U! (
j5234, 235, 236, 238
R jM j
according to Eq 4-13. Therefore the results for (233U/
236 14.5.4 Molar Mass:
U)CO-TE serves as quality control for the isobaric correc-
tions. But it is possible that all the results for (233U/236U)CO-TE
consistently deviate (usually less than 0.05 %) from the known (
j5234, 235, 236, 238
R jM j
M~U! 5 (18)
or certified 233U/236U ratio of the double spike, due a possible ( Rj
j5234, 235, 236, 238
overall mass bias effect (13). This does not have an influence
on the (average) internally corrected (235U/238U)CO ratios where:
calculated according to Eq 13. n(iU) = number of moles for isotope i;
234
14.4 Uncertainty Estimation: n(U) = number of moles for all isotopes, usually U,
235
14.4.1 The uncertainties for DS measurements are estimated U, 236U, and 238U;
according to the principles described in the Guide to the m(iU) = mass for isotope i;

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG )UL-DQ*07


10
'RZQORDGHGSULQWHGE\
8&5 8QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
C1871 − 22

m(U) = mass for all isotopes, usually 234U, 235U, 236U, and sions obtained in DS measurements is shown in A1.1.3 as well
238 as by the QC charts in A1.1.4 – A1.1.7. In contrast to the
U;
Ri = isotope amount ratio (i = 234, 235, 236, and 238) internal measurement precision and the within-turret precision,
relative to one (no matter which) of the isotopes; for the between-turret precision the decrease with increasing
and enrichment of 235U towards the sample with 50 % enrichment
Mi = molar mass for isotope i. is not observed. The reason for the absence of a trend is not
known. Among the used QC materials, apparently the between-
15. Evaluation and Statements for Precision and Bias turret precision is best for NBL CRM 112A. This might be due
15.1 A total of two thermal ionization mass spectrometers at to the fact that the loading solution for this material is the only
JRC-G.2 (formerly IRMM), and the NML (IAEA) were used to one prepared from directly dissolving a uranium metal instead
determine the precision and bias for uranium isotope amount of an oxide. A similar observation has been made for the
ratio measurements via the DS method. Only TIMS instru- precision for MTE measurements of samples such as the metal
ments of the type TRITONTM were used for the analyses. The reference materials NBL CRM 115 and NBL CRM 116A (see
following uranium certified reference materials were used to Test Method C1832).
evaluate the precision and bias of DS measurements: 15.3 Bias for DS Measurements of the Major Ratio, 235U/
15.1.1 From NBL (U.S. DOE), the CRM 112A was used as 238
U:
a QC material. It is also known with the name NBS SRM 960.
15.1.2 From JRC-G.2, the 235U/238U base material from the 15.3.1 No significant biases (that is, deviations of measured
IRMM-074 series (labelled unofficially IRMM-3050, not for and corrected ratios from the known or certified ratios) for the
sale), IRMM-184 and the converted UF6 materials IRMM-012, major ratios, 235U/238U, were observed for the CRM 112A
IRMM-22, and IRMM-023 were applied, which were used as from NBL. The deviations shown in Fig. A1.1 (in A1.1.4) for
well for the validation of the Cristallini sampling method for the measurements of NBL CRM 112A were calculated based
UF6 (see (12) and (13)). Only measurements of IRMM-021, on the new average value of 0.00725495(80) which was
IRMM-22, and IRMM-023 samples prepared by direct hydro- measured in a collaborative effort by several geochemistry
lysis are considered for the evaluation of precision and bias laboratories and JRC-G.2 (11). This new average value has
within this test method. been acquired using the double spike IRMM-3636a, but also
15.1.3 The used materials cover the range of 235U enrich- other double spike materials, combined with variety of instru-
ments from 0.2 to 50 %. The reference materials used for mentation types (for example, TIMS and MC-ICPMS) and new
quality control of the DS method are not identical between measurement methods. This “inter-laboratory comparison”
JRC-G.2 and NML. Therefore, common general observations (11) also showed that the result for NBL CRM 112A using the
and some examples for precision and bias are presented within method described in this test method is in good agreement (at
this test method. a level of better than 0.002 %) with all results from other
15.1.4 For the double spike material, only the IRMM-3636a laboratories (other than JRC-G.2 and NML), who were also
reference material was used by both JRC-G.2 and NML. using other DS materials and other techniques like MC-
15.2 Precision for DS Measurements of the Major Ratio ICPMS. Further confirmatory results for NBL CRM 112A can
235
U/238U: be found in (8). This new consensus value is about 0.031 %
15.2.1 For the corrected major ratio, 235U/238U, the internal higher than the old consensus value (1/137.88=0.0072527), but
measurement precision varies within a range of 0.0015 % and in agreement with the new certified value 0.0072543(40) from
0.0035 % relative standard error (RSE) for all mass cycles NBL. Moreover, it is traceable to the SI and an uncertainty
during the measurement. The precision slightly depends on the statement is provided according to the Guide to the Expression
235
U enrichment. For depleted samples the 235U ion beams are of Uncertainty in Measurements (15). As an alternative to NBL
smaller causing slightly inferior precision due to the inferior CRM 112A, the (close to) natural reference material IRMM-
signal-to-amplifier noise ratio. Therefore the use of amplifiers 184 has been re-measured as well by several laboratories using
with 1012 Ω resistors for detecting 235U in depleted samples is the double spike IRMM-184 has been re-measured as well by
recommended. An example for internal measurement preci- several laboratories using the double spike IRMM-3636, re-
sions obtained in DS measurements is shown in A1.1.1. The sulting in a 235U/238U value of 0.0072631(11), which is in
precision is at the same level as reported for measurements agreement with the certified value of 0.0072623(22). But NBL
within the scientific community, see references (8-11, 16-19). CRM 112A still remains the primary choice for quality control,
15.2.2 For the corrected major ratio, 235U/238U, the within- as a reference material for external normalization for high
turret precision varies within a range of 0.0015 % and precision 235U/238U measurements and even for calibration of
0.0050 % relative standard deviation (RSD) for replicate (>5) non-certified in-house double spike WRMs (see also Guide
filament loadings of reference materials analyzed as reported in C1128). This popularity is due to its widespread 236U relative
(13). The precision slightly depends on the 235U enrichment. abundance, which is below the detection limit of 6 × 10–10 of
An example for within-turret precisions obtained in DS mea- the MTE method (5, 6), compared to IRMM-184 (1.2 × 10–7)
surements is shown in A1.1.2. and IRMM-3050 (3.8 × 10–5). The metallic form of the original
15.2.3 For the corrected major ratio, 235U/238U, the CRM 112A reference material was also found to be an
between-turret precision varies between 0.0015 % and 0.006 % advantage for high precision measurements using the DS and
RSD as reported in (13). Examples for between-turret preci- MTE techniques.

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG )UL-DQ*07


11
'RZQORDGHGSULQWHGE\
8&5 8QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
C1871 − 22
15.3.2 For the IRMM reference materials, IRMM-3050, the (0.0025 %) higher than the reference value (11). This could in
gravimetrically prepared 235U/238U-base material from the principle be due to slight differences in Faraday cup efficien-
IRMM-074 series (not for sale, (14)) was used at JRC-G.2 only cies between the two TIMS instruments used at JRC-G.2 and
(see A1.1.5, Fig. A1.2), and the IRMM-184 was used at NML. As a confirmation for this assumption, measurements
JRC-G.2 (see A1.1.6, Fig. A1.3) and NML (see A1.1.7, Fig. performed using both TIMS instruments on the same samples
A1.4). No significant biases for the major ratios, 235U/238U for the validation of the Cristallini samples (IRMM-020,
were observed for measurements of these materials, as shown IRMM-022, and IRMM-023, see (13) showed the same trend).
in A1.1.8. Therefore, it is recommended to normalize all DS results to the
15.3.3 Additionally, Table A1.5 and Fig. A1.5 show that the same common standard reference material, for which the
result for the major ratio of NBL CRM 112A, expressed as NBL-CRM 112A is recommended. However, if all uncertainty
238
U/235U here, obtained at JRC-G.2 using the method de- contributions are considered, there was no significant bias
scribed in this test method, is in good agreement with numer- observed for any of the used QC materials.
ous results from other laboratories within the nuclear safe-
guards and scientific community, where not only IRMM-3636 16. Measurement Uncertainty
but also other DS materials and other mass spectrometric 16.1 Major Isotope Amount Ratios 235U/238U—The propor-
techniques like MC-ICPMS were used. Further detailed dis- tions of the uncertainty contributions (see 14.4.2) depend on
cussions about this matter can be found in (17). the precision and number of replicates of the DS
15.3.4 Among the used QC materials, only IRMM-3050 is measurements, as well as on the uncertainty of the 233U/236U
certified based on gravimetric preparation, like the double isotope amount ratio of the double spike material. The com-
spike reference materials IRMM-3636, IRMM-3636a, and bined uncertainty is usually largely dominated by the uncer-
IRMM-3636b are as well. The agreement of the measured ratio tainty of the known or certified 233U/236U ratio of the double
with the certified one also confirms the accuracy of the DS spike. Typical values for the expanded uncertainty (coverage
method, however with a combined expanded uncertainty of factor k = 2) are about 0.016 to 0.020 % independent of the
0.022 % which is dominated by the uncertainties of the DS and ratios.
IRMM-3050 themselves. For NBL CRM 112A, the bias is
insignificant over the 9 year period. For IRMM-184 the very 17. Keywords
early results from 2009 at JRC-G.2 seem to be lower (about 17.1 certified reference materials; double spike; isotope
–0.005 %) than the reference value (11) but the recent results in amount ratios; mass fractionation correction; modified total
2016 agree better. At the NML the DS method was imple- evaporation; thermal ionization mass spectrometry; total
mented in 2016, and the results for IRMM-184 are slightly evaporation; uranium

ANNEX

(Mandatory Information)

A1. PRECISION AND BIAS FOR DS MEASUREMENTS

A1.1 A discussion and statements regarding precision and TABLE A1.2 Within-Turret Precision (% RSD) for DS
bias for DS measurements are given in Section 15. Associated Measurements of the Major Ratio 235U/238U
235
tables and figures are presented in A1.1.1 – A1.1.9. CRM U/238U 235
U/U, % Average RSD, %
Data from (13, 20)
A1.1.1 See Table A1.1. IRMM-020 0.00210 0.2 0.0047 %
IRMM-021 0.00441 0.4 0.0041 %
A1.1.2 See Table A1.2. IRMM-022, IRMM-184 0.00726 0.7 0.0025 %
IRMM-023 0.03388 3.3 0.0027 %
A1.1.3 See Table A1.3. IRMM-3050 (-074) 1.00026 50.0 0.0018 %
A1.1.4 QC chart to show bias and between turret precision
for 235U/238U measurements of NBL CRM 112A performed at
JRC-G.2, taken from (13). See Fig. A1.1. A1.1.5 QC chart to show bias and between turret precision
for 235U/238U measurements of IRMM-3050 (IRMM-074 se-
TABLE A1.1 Internal Measurement Precision (% RSE) for DS
Measurements of the Major Ratio 235U/238U ries) performed at JRC-G.2, taken from (13). See Fig. A1.2.
CRM 235
U/238U 235
U/U, % Average RSE, % A1.1.6 QC chart to show bias and between turret precision
Data from (13, 20) for 235U/238U measurements of IRMM-184 performed at JRC-
IRMM-020 0.00210 0.2 0.0036 %
IRMM-021 0.00441 0.4 0.0027 %
G.2, taken from (13). See Fig. A1.3.
IRMM-022, IRMM-184 0.00726 0.7 0.0022 % A1.1.7 QC chart to show bias and between turret precision
IRMM-023 0.03388 3.3 0.0015 %
IRMM-3050 (-074) 1.00026 50.0 0.0016 % for 235U/238U measurements of IRMM-184 performed at NML,
taken from (13). See Fig. A1.4.

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG )UL-DQ*07


12
'RZQORDGHGSULQWHGE\
8&5 8QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
C1871 − 22
TABLE A1.3 Between-Turret Precision (% RSD) for DS A1.1.8 See Table A1.4.
Measurements of Major Ratio 235U/238U
CRM 235
U/238U 235
U/U, % Average RSD, % A1.1.9 See Table A1.5 and Fig. A1.5.
Data from (13, 20)
IRMM-020 0.00210 0.2 0.0057 %
IRMM-184 (natural) 0.00726 0.7 0.0029 %
IRMM-112A (natural) 0.00726 0.7 0.0015 %
IRMM-3050 (-074) 1.00026 50.0 0.0045 %

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG )UL-DQ*07


13
'RZQORDGHGSULQWHGE\
8&5 8QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
C1871 − 22

FIG. A1.1 QC Chart NBL CRM 112A, Measurements Performed at JRC-G.2

FIG. A1.2 QC Chart IRMM-3050, Measurements Performed at JRC-G.2

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG )UL-DQ*07


14
'RZQORDGHGSULQWHGE\
8&5 8QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
C1871 − 22

FIG. A1.3 QC Chart IRMM-184, Measurements Performed at JRC-G.2

FIG. A1.4 QC Chart IRMM-184, Measurements Performed at NML

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG )UL-DQ*07


15
'RZQORDGHGSULQWHGE\
8&5 8QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
C1871 − 22
235
TABLE A1.4 Bias (% RD) for DS Measurements of the Major Ratio U/238U
235 238 235
CRM U/ U U/U, % RD (%) Uncertainty (w/o DS) Uncertainty (all contr.)
Data from (13)
IRMM-184 (natural), IAEA, only 2016 0.00726 0.7 0.0025 % 0.0016 % 0.015 %
IRMM-184 (natural), JRC-G.2 0.00726 0.7 –0.0024 % 0.0023 % 0.015 %
IRMM-184 (natural), JRC-G.2, only 2009 0.00726 0.7 –0.0052 % 0.0023 % 0.015 %
IRMM-184 (natural), JRC-G.2, only 2016 0.00726 0.7 0.0000 % 0.0030 % 0.015 %
IRMM-112A (natural), JRC-G.2 0.00726 0.7 –0.0013 % 0.0015 % 0.011 %
IRMM-3050 (-074), JRC-G.2 1.00026 50.0 –0.0012 % 0.0027 % 0.022 %

238
TABLE A1.5 Results for DS Measurements of the Major Ratio U/235U for NBL CRM 112A
Laboratory, Reference Instrumentation Reference Materials used for Instrument Calibration NBL 112A 238U/235U
Uncertainty, k=2
NBL New Certified Value (21) Triton TIMS NBL-U030A, by MTE/TIMS (see Test Method C1832) 137.849(76)
“New Average Value” as introduced in (11) Triton TIMS, Neptune MC-ICPMS Double Spike IRMM-3636, and other Double Spikes 137.837(15)
calibrated using CRMs such as IRMM-074, IRMM-184
233
JRC-G.2/Geel (11) Triton TIMS U/236U Double Spike IRMM-3636, 137.836(23)
233
U/236U=1.01906(16)
233
NIGL Isotope Geoscience Laboratory (2010) Triton TIMS U/236U Double Spike IRMM-3636, 137.844(25)
233
(11) U/236U=1.01906(16)
233
NERC Isotope Geoscience Laboratory Triton TIMS U/236U Double Spike IRMM-3636, 137.832(22)
233
(2012) (16) U/236U=1.01906(16)
233
University of Frankfurt (11) Neptune MC-ICPMS U/236U Double Spike IRMM-3636, 137.833(28)
233
U/236U=1.01906(16)
233
Thermo Fisher Scientific (11) Neptune MC-ICPMS U/236U Double Spike IRMM-3636, 137.836(23)
233
U/236U=1.01906(16)
233
NERC Isotope Geoscience Laboratory Neptune MC-ICPMS U/236U Double Spike IRMM-3636, 137.829(22)
233
(2012) (16) U/236U=1.01906(16)
233
National Taiwan University, NTU, (11) Neptune MC-ICPMS U/236U Double Spike made by UMN, calibrated using 137.834(21)
IRMM-074 ⁄10, 238U/235U=0.99974(15)
233
University of Minnesota, UMN, (11) Neptune MC-ICPMS U/236U Double Spike made by UMN, calibrated using 137.830(25)
IRMM-074 ⁄10, 238U/235U=0.99974(15)
233
Lawrence Livermore National Laboratory Triton TIMS U/236U Double Spike made by LLNL, calibrated using 137.823(42)
LLNL, (11) IRMM-184, 238U/235U=137.697(41)
233
Arizona State University (18) Neptune MC-ICPMS U/236U Double Spike IRMM-3636, 137.843(23)
233
U/236U=1.01906(16)
233
Enrico Fermi Institute (19) Neptune MC-ICPMS U/236U Double Spike IRMM-3636, 137.842(23)
233
U/236U=1.01906(16)

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG )UL-DQ*07


16
'RZQORDGHGSULQWHGE\
8&5 8QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
C1871 − 22

238
FIG. A1.5 Results for DS Measurements for the Major Ratio U/235U for NBL CRM 112A

REFERENCES

(1) Callis, E. L. and Abernathy, R. M., “High-Precision Isotopic Analyses (9) Keegan, E., Richter, S., Kelly, I., Wong, H., Gadd, P., Kuehn, H., and
of Uranium and Plutonium by Total Sample Volatilization and Signal Alonso, A., “The Provenance of Australian Uranium Ore Concentrates
Integration,” Int. J. Mass Spectrom. Ion Proc., Vol 103, 1991, pp. by Elemental and Isotopic Analysis,” Applied Geochemistry, Vol 23,
93–105. 2008, pp. 765–777.
(2) Fiedler, R. et al., “Report on Preliminary Experience with Total (10) Cheng, H., Edwardsa, R. L., Shenc, C.-C., Polyak, V. J., Asmerom,
Evaporation Measurements in Thermal Ionization Mass Y., Woodhead, J., Hellstrom, J., Wang, Y., Kong, X., Spötl, C., Wang,
Spectrometry,” Int. J. Mass Spectrom. Ion Proc., Vol 132, 1994, pp. X., and Alexander Jr., E. C., “Improvements in 230Th Dating, 230Th
207–215. and 234U Half-life Values, and U-Th Isotopic Measurements by
(3) Fiedler, R., “Total Evaporation Measurements—Experience with Multi-collector Inductively Coupled Plasma Mass Spectrometry,”
Multi-Collector Instruments and a Thermal Ionization Quadrupole Earth and Planetary Science Letters, Vol 371–372, 2013, pp. 82–91.
Mass-Spectrometer,” International Journal of Mass Spectrometry, Vol (11) Richter, S., Eykens, R., Kühn, H., Aregbe, Y., Verbruggen, A., and
146, 1995, pp. 91–97. Weyer, S., “New Average Values for the n(238U)/n(235U) Isotope
(4) Mathew, K. J., O’Connor, G., Hasozbek, A., and Kraiem, M., “Total Ratios of Natural Uranium Standards,” International Journal of
Evaporation Method for Uranium Isotope-amount Ratio Mass Spectrometry, Vol 295, 2010, pp. 94–97.
Measurements,” Journal of Anal. At. Spectrom., Vol 28, 2013, pp. (12) Galdoz, E., Esteban, A., Cristallini, O., and Perrotta, J., “UF6
866–876. Sampling Method Using Alumina,” 49th INMM Annual Meeting,
(5) Richter, S. and Goldberg, S. A., “Improved Techniques for High USA, 2008.
Accuracy Isotope Ratio Measurements of Nuclear Materials using (13) Richter, S., Hiess, J., and Jacobsson, U., “Validation of Cristallini
Thermal Ionization Mass Spectrometry,” International Journal of Sampling, Method for UF6 by High Precision, Double-Spike
Mass Spectrometry, Vol 229, 2003, pp. 181–197. Measurements,” 2016, JRC-103668, https://ec.europa.eu/jrc/en/
(6) Richter, S., Kühn, H., Aregbe, Y., Hedberg, M., Horta-Domenech, J., news/validation-cristallini-method-sampling-uranium-hexafluoride-
et al, “Improvements in Routine Uranium Isotope Ratio Measure- uf6).
ments using the Modified Total Evaporation Method for Multi- (14) Richter, S., Alonso, A., Aregbe, Y., Eykens, R., Kehoe, F., et al, “A
Collector Thermal Ionization,” Mass Spectrometry Journal of Anal. New Series of Uranium Isotope Reference Materials for Investigat-
At. Spectrom., Vol 26, 2011, pp. 550–564. ing the Linearity of Secondary Electron Multipliers in Isotope Mass
(7) Richter, S., Alonso, A., Eykens, R., Jacobsson, U., Kühn, H., Spectrometry,” International Journal of Mass Spectrometry, Vol 281,
Verbruggen, A., Aregbe, Y., Wellum, R., and Keegan, E., “The 2009, pp. 115–125.
Isotopic Composition of Natural Uranium Samples—Measurements (15) BIPM, “Joint Committee for Guides in Metrology, Evaluation of
using the New n(233U)/n(236U) Double Spike IRMM-3636,” Interna- Measurement Data—Guide to the Expression of Uncertainty in
tional Journal of Mass Spectrometry, Vol 269, 2008, pp. 145–148. Measurement,” JCGM, Vol 100, 2008.
(8) Brennecka, G. A., Wasylenki, E., Bargar, J. R., Weyer, S., and Anbar, (16) Hiess, J., Condon, D. J., McLean, N., and Noble, S. R., “Systematics
A. D., “Uranium Isotope Fractionation during Adsorption to Mn- in Terrestrial Uranium-Bearing Minerals,” Science, Vol 315, 2012,
Oxyhydroxides,” Environ. Sci. Technol., Vol 45, 2011, pp. 1370–1375. pp. 1610–1614.

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG )UL-DQ*07


17
'RZQORDGHGSULQWHGE\
8&5 8QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
C1871 − 22
(17) Andersen, M. B., Stirling, C. H., and Weyer, S., “Uranium Isotope (20) Mialle, S., Richter, S., Hennessy, C., Truyens, J., Jacobsson, U., et al,
Fractionation,” Reviews in Mineralogy & Geochemistry, Vol 82, “Certifications of Uranium Hexafluoride Reference Materials for
2017, pp. 799–850. Isotopic Composition,” 8th International Conference on Isotopes,
(18) Goldmann, A., Brennecka, J., Noordmann, J., Weyer, S., and Chicago, Aug. 24–29, 2014, J Radioanal Nucl Chem, Vol 305, 2015,
Wadhwa, M., “The Uranium Isotopic Composition of the Earth and pp. 255–266.
the Solar System,” Geochim Cosmochim Acta, Vol 148, 2015, pp. (21) Mathew, K., Mason, P., Voeks., and Narayanan, U., “Uranium
145–158. Isotope Abundance Ratios in Natural Uranium Metal Certified
(19) Tissot, F. L. H. and Dauphas, N., “Uranium Isotopic Compositions of Reference Material 112-A, International Journal of Mass
the Crust and Ocean: Age Corrections, U Budget and Global Extent Spectrometry, Vol 315, 2012, pp. 8–14.
of Modern Anoxia,” Geochim Cosmochim Acta, Vol 167, 2015, pp.
113–143.

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG )UL-DQ*07


18
'RZQORDGHGSULQWHGE\
8&5 8QLYHUVLGDGGH&RVWD5LFDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG

You might also like