Professional Documents
Culture Documents
Astm C1871-22
Astm C1871-22
for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: C1871 − 22
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
6.4 Samples shall be chemically purified to assure reliable 7.4 A pipette or microsyringe to transfer microliter volumes
analyses by TIMS. Impurities, especially alkali elements, of solution.
produce unstable ion emission leading to poor precision in the 7.5 A separate filament heating device for drying and
isotope amount ratios. Organic contaminants or oxide layers on oxidizing the sample on the filament ribbon after loading.
the filaments also adversely influence TIMS analyses. Isobaric
interferences, if not removed, will bias the isotope amount 8. Reagents and Materials
ratios. Contaminants in reagents, lab ware, or filament material 8.1 Purity of Reagents—Ultra-high purity reagents shall be
are also sources for bias in the isotope amount ratios. used for processing small sample amounts or samples with
6.5 The performance of the instrument can be adversely extremely small isotope amount ratios. The level of uranium
affected by changes in the environmental conditions of the contamination from chemicals, water, and the sample handling
laboratory, that is, temperature and humidity. For this reason, environments shall be determined to ensure that the materials
controlled laboratory environmental conditions should be and the environment are sufficiently pure for the samples being
maintained (within the manufacturer’s specifications) during analyzed.
instrument operation. 8.2 Nitric Acid (HNO3, 15.8 M)—Concentrated nitric acid.
7. Apparatus 8.3 Nitric Acid (HNO3, 1 M)—One volume of concentrated
nitric acid (HNO3, 15.8 M) brought to 15.8 volumes with
7.1 The suitability of the mass spectrometer for carrying out water.
measurements by the DS method shall be evaluated by means
of performance tests. The relevant instrument characteristics 8.4 Purity of Water—Unless otherwise indicated, references
are as follows: to water shall be understood to mean laboratory accepted
7.1.1 A thermal ionization source for using single, double, demineralized or deionized water as described by Type I of
or triple filament assemblies with rhenium or tungsten Specification D1193.
filaments, or both;
7.1.2 A mass analyzer sufficient to resolve adjacent masses
5
in the mass-to-charge range being studied, m/z = 233 to 238 for The sole source of supply of the apparatus known to the committee at this time
is Thermo Fisher Scientific Inc., 81 Wyman St., Waltham, MA 02451. If you are
U+. Resolution shall be greater than 350 (full width at 1 % of
aware of alternative suppliers, please provide this information to ASTM Interna-
peak height) and the abundance sensitivity at mass 237 for ions tional Headquarters. Your comments will receive careful consideration at a meeting
of 238U less than 8 × 10–6; of the responsible technical committee,1 which you may attend.
236
U
U DS
(2)
at the very end of each block. This is needed to determine the
ion beam intensity before the next heating step (see 13.9).
S D
233
236
U
U
ME
238
U
U ME
1 2
232.55
233
U D ME
238
U
5 K R235 S D 235
U
238 (3)
using
U CO
U ME
S D 238
U
where:
S D 238
U
5
236
U DS
(7)
CO = corrected ratio, and
ln~ 235 ⁄ 238!
S D
233
U DS
233
236
U
U
DS
the exponent R235 is given by R2355 ln 233 ⁄ 236
~ ! 14.3.4 The relative isobaric contribution (235Ui/235U)ME
>0.991543, according the exponential fractionation law. from the double spike is given by
14.2.3 The exponent R235 is given by R235 = 1 in case the
linear fractionation law is preferred as an approximation. S DS D 235
U 233
U
S D
235
Ui
5
233
U DS
238
U ME
(8)
14.3 Background and Isobaric Corrections—All back-
ground and isobaric corrections of the isotope ratios are
235
U
S D ME
235
238
U
U
ME
performed internally on a per mass cycle basis for each 235 233 235 233
measurement. Where ( U/ U)DS is the U/ U ratio of the double
14.3.1 The measured 233U/236U ratio is corrected for back- spike, which can be calculated from the certified (235U/236U)DS
ground and the relative isobaric contribution (236Ui/236U) from and (233U/236U)DS ratios given on the certificate using:
the sample:
S D 235
U
S D S D
233
U 233
U S D 235
U
5
236
U DS
(9)
236
U CO
5 236
U ME
·
S D
233
U DS
233
236
U
U
S SS D S D D D 1 2
232.55
233
U ME
1
233.55
233
U ME
⁄2
14.3.5 The relative isobaric contribution (234Ui/234U)ME
DS
S SS D S D D S D D
1 2
235.55
236 1
236.55
236 ⁄2 2
236
236
Ui from the double spike is given by
U ME
U ME
U ME
S DS D 234
U 233
U
(4)
S D
234
Ui
5
233
U DS
238
U ME
(10)
where:
ME = measured ratio, and
234
U
S D ME
234
238
U
U
ME
S D S D 236
U 233
U
S D 234
U
S D
236
Ui 238
U Sample
236
U ME
S D 234
U
5
236
U DS
(11)
236
U
S DME
5 233
238
U
U
ME
S D
233
U DS
233
236
U
U
DS
S SS D S D D S D D
1 2
237.55
238
U ME
1
238.55
238
U ME
⁄2 2
238
238
Ui
U ME
14.3.6 The K factor for a double spike measurement is
derived from the certified (233U/236U)DS ratio for each mass
S SS D S D D D 1 2
235.55
236
U ME
1
236.55
236
U ME
⁄2
cycle according to Eq 2 by using the (233U/236U)CO ratio from
Eq 4 which is corrected for background and isobaric interfer-
(5) ences:
Where (236U/238U)Sample is the 236U/238U ratio of the sample,
which should be approximately known from separate mea-
surements (for example, using the MTE method, Test K5
S D 233
236
U
U DS
(12)
Method C1832); and (238Ui/238U)ME is the relative isobaric
contribution from the double spike.
S D 233
236
U
U
CO
238 238
14.3.3 The relative isobaric contribution ( Ui/ U)ME 14.3.7 The K factor is applied in the following way to obtain
from the double spike is given by the corrected 235U/238U ratio of the sample:
S D 235
U
238
U CO
S D
5 K R235
235
238
U
U ME
Expression of Uncertainty in Measurements (GUM) (15). For
DS measurements, the uncertainty evaluation steps are ex-
S SS D
1 2
234.55
235
U ME
1 S 235.55
235
U D D S D D
ME
⁄2 2
Ui
U
235
235
ME
plained in the following.
235
14.4.2 For the uncertainty of the corrected major ratio
U/238U calculated according to Eq 12 and Eq 13, the relative
S SS D
1 2
237.55
238
U ME
1 S 238.55
238
U D D S D D
ME
238
⁄2 2 238
Ui
U ME
uncertainties of the measured 233U/236U ratios of the double
(13) spike material used for mass fractionation correction and the
measured 235U/238U ratios of the sample, as well as the relative
234 238
14.3.8 The corrected U/ U ratio of the sample can be uncertainty of the known or certified 233U/236U ratio of the
obtained in the following way: double spike, are added in quadrature to obtain the relative
uncertainty of the correct 235U/238U ratio of the sample.
S D 234
U
238
U CO
S D
5 K R234
234
238
U
U ME
Because of the good repeatability of DS measurements (about
0.003 to 0.006 % RSD, only slightly dependent of the value of
S SS D
1 2
233.55
235
U ME
1 S 234.55
235
U D D S D D
ME
⁄2 2
Ui
U
234
234
ME
the ratio), this uncertainty is largely dominated by the uncer-
tainty of the known or certified ratio of the double spike
S SS D
1 2
237.55
238
U ME
1 S 238.55
238
U D D S D D
ME
238
⁄2 2 238
Ui
U ME
(typically in the range of 0.015 to 0.03 %). Typically values for
the expanded uncertainty (coverage factor k = 2) are about
(14)
ln~ 234 ⁄ 238! 0.016 to 0.020 % (see 16.1 for additional details).
The exponent R234 is given by R2345 ln 233 ⁄ 236 >1.32487, 14.4.3 The uncertainty contributions of the corrected ratio
~ ! 235
which is close to 4⁄3 . U/238U ratio arising from the background and isobaric
corrections are negligible, if recommended sample/spike
14.3.9 Quality Control for the Corrected Ratio (233U/
236 amount ratios of >20 are applied.
U)CO—In case the entire sample is evaporated and used for
14.4.4 For the uncertainty of the corrected ratio 234U/238U
the data acquisition using an MTE heating script (see 4.1 and
calculated according to Eq 14, the uncertainty from the isobaric
13.9.2), it is recommended to calculate a weighted mean of the
correction due to 234U stemming from the double spike
corrected (233U/236U)CO ratios derived using Eq 4, as usually
material may become the dominant source of uncertainty.
done according to the total evaporation principle (see Test
Methods C1672 and C1832): 14.5 Calculation of Isotope Amount Fractions (“Atom
Percent”), Isotope Mass Fractions (“Weight Percent”), and
( FS D233
U
· I ~ 236U ! G Molar Mass (“Atomic Weight”):
S D
233
236
U
U
CO2TE
5
ALL_INTEGRATIONS
(
236
U CO
I~ 236
U!
(15) 14.5.1 The isotope amount fractions (also called “atom
percent”), isotope mass fractions (also called “weight
ALL_INTEGRATIONS
percent”), and the molar mass can only be calculated if all
where: isotope ratios 234U/238U, 235U/238U, and 236U/238U of the
I(236U) = signal intensity (in V) for 236U measured sample are known. The 234U/238U and 236U/238U ratios of the
on a Faraday cup, corrected for the sample have to be measured using a method other than the DS
Faraday cup amplifier gain and baseline, method, for example, by TE (Test Method C1672) or MTE
and (Test Method C1832).
(233U/236U)CO-TE = mean for the (233U/236U)CO ratios ac- 14.5.2 Isotope Amount Fraction for Isotope i (i = 234, 235,
quired for all mass cycles during a DS 236, 238):
measurement, weighted by I(236U).
n ~ iU ! n ~ iU ! Ri
The results for (233U/236U)CO-TE are expected to be in mutual A ~ iU ! 5 5 5 (16)
agreement for all measurements on individual filaments as well
n~U! (
j5234, 235, 236, 238
n ~ jU ! (
j5234, 235, 236, 238
Rj
m(U) = mass for all isotopes, usually 234U, 235U, 236U, and sions obtained in DS measurements is shown in A1.1.3 as well
238 as by the QC charts in A1.1.4 – A1.1.7. In contrast to the
U;
Ri = isotope amount ratio (i = 234, 235, 236, and 238) internal measurement precision and the within-turret precision,
relative to one (no matter which) of the isotopes; for the between-turret precision the decrease with increasing
and enrichment of 235U towards the sample with 50 % enrichment
Mi = molar mass for isotope i. is not observed. The reason for the absence of a trend is not
known. Among the used QC materials, apparently the between-
15. Evaluation and Statements for Precision and Bias turret precision is best for NBL CRM 112A. This might be due
15.1 A total of two thermal ionization mass spectrometers at to the fact that the loading solution for this material is the only
JRC-G.2 (formerly IRMM), and the NML (IAEA) were used to one prepared from directly dissolving a uranium metal instead
determine the precision and bias for uranium isotope amount of an oxide. A similar observation has been made for the
ratio measurements via the DS method. Only TIMS instru- precision for MTE measurements of samples such as the metal
ments of the type TRITONTM were used for the analyses. The reference materials NBL CRM 115 and NBL CRM 116A (see
following uranium certified reference materials were used to Test Method C1832).
evaluate the precision and bias of DS measurements: 15.3 Bias for DS Measurements of the Major Ratio, 235U/
15.1.1 From NBL (U.S. DOE), the CRM 112A was used as 238
U:
a QC material. It is also known with the name NBS SRM 960.
15.1.2 From JRC-G.2, the 235U/238U base material from the 15.3.1 No significant biases (that is, deviations of measured
IRMM-074 series (labelled unofficially IRMM-3050, not for and corrected ratios from the known or certified ratios) for the
sale), IRMM-184 and the converted UF6 materials IRMM-012, major ratios, 235U/238U, were observed for the CRM 112A
IRMM-22, and IRMM-023 were applied, which were used as from NBL. The deviations shown in Fig. A1.1 (in A1.1.4) for
well for the validation of the Cristallini sampling method for the measurements of NBL CRM 112A were calculated based
UF6 (see (12) and (13)). Only measurements of IRMM-021, on the new average value of 0.00725495(80) which was
IRMM-22, and IRMM-023 samples prepared by direct hydro- measured in a collaborative effort by several geochemistry
lysis are considered for the evaluation of precision and bias laboratories and JRC-G.2 (11). This new average value has
within this test method. been acquired using the double spike IRMM-3636a, but also
15.1.3 The used materials cover the range of 235U enrich- other double spike materials, combined with variety of instru-
ments from 0.2 to 50 %. The reference materials used for mentation types (for example, TIMS and MC-ICPMS) and new
quality control of the DS method are not identical between measurement methods. This “inter-laboratory comparison”
JRC-G.2 and NML. Therefore, common general observations (11) also showed that the result for NBL CRM 112A using the
and some examples for precision and bias are presented within method described in this test method is in good agreement (at
this test method. a level of better than 0.002 %) with all results from other
15.1.4 For the double spike material, only the IRMM-3636a laboratories (other than JRC-G.2 and NML), who were also
reference material was used by both JRC-G.2 and NML. using other DS materials and other techniques like MC-
15.2 Precision for DS Measurements of the Major Ratio ICPMS. Further confirmatory results for NBL CRM 112A can
235
U/238U: be found in (8). This new consensus value is about 0.031 %
15.2.1 For the corrected major ratio, 235U/238U, the internal higher than the old consensus value (1/137.88=0.0072527), but
measurement precision varies within a range of 0.0015 % and in agreement with the new certified value 0.0072543(40) from
0.0035 % relative standard error (RSE) for all mass cycles NBL. Moreover, it is traceable to the SI and an uncertainty
during the measurement. The precision slightly depends on the statement is provided according to the Guide to the Expression
235
U enrichment. For depleted samples the 235U ion beams are of Uncertainty in Measurements (15). As an alternative to NBL
smaller causing slightly inferior precision due to the inferior CRM 112A, the (close to) natural reference material IRMM-
signal-to-amplifier noise ratio. Therefore the use of amplifiers 184 has been re-measured as well by several laboratories using
with 1012 Ω resistors for detecting 235U in depleted samples is the double spike IRMM-184 has been re-measured as well by
recommended. An example for internal measurement preci- several laboratories using the double spike IRMM-3636, re-
sions obtained in DS measurements is shown in A1.1.1. The sulting in a 235U/238U value of 0.0072631(11), which is in
precision is at the same level as reported for measurements agreement with the certified value of 0.0072623(22). But NBL
within the scientific community, see references (8-11, 16-19). CRM 112A still remains the primary choice for quality control,
15.2.2 For the corrected major ratio, 235U/238U, the within- as a reference material for external normalization for high
turret precision varies within a range of 0.0015 % and precision 235U/238U measurements and even for calibration of
0.0050 % relative standard deviation (RSD) for replicate (>5) non-certified in-house double spike WRMs (see also Guide
filament loadings of reference materials analyzed as reported in C1128). This popularity is due to its widespread 236U relative
(13). The precision slightly depends on the 235U enrichment. abundance, which is below the detection limit of 6 × 10–10 of
An example for within-turret precisions obtained in DS mea- the MTE method (5, 6), compared to IRMM-184 (1.2 × 10–7)
surements is shown in A1.1.2. and IRMM-3050 (3.8 × 10–5). The metallic form of the original
15.2.3 For the corrected major ratio, 235U/238U, the CRM 112A reference material was also found to be an
between-turret precision varies between 0.0015 % and 0.006 % advantage for high precision measurements using the DS and
RSD as reported in (13). Examples for between-turret preci- MTE techniques.
ANNEX
(Mandatory Information)
A1.1 A discussion and statements regarding precision and TABLE A1.2 Within-Turret Precision (% RSD) for DS
bias for DS measurements are given in Section 15. Associated Measurements of the Major Ratio 235U/238U
235
tables and figures are presented in A1.1.1 – A1.1.9. CRM U/238U 235
U/U, % Average RSD, %
Data from (13, 20)
A1.1.1 See Table A1.1. IRMM-020 0.00210 0.2 0.0047 %
IRMM-021 0.00441 0.4 0.0041 %
A1.1.2 See Table A1.2. IRMM-022, IRMM-184 0.00726 0.7 0.0025 %
IRMM-023 0.03388 3.3 0.0027 %
A1.1.3 See Table A1.3. IRMM-3050 (-074) 1.00026 50.0 0.0018 %
A1.1.4 QC chart to show bias and between turret precision
for 235U/238U measurements of NBL CRM 112A performed at
JRC-G.2, taken from (13). See Fig. A1.1. A1.1.5 QC chart to show bias and between turret precision
for 235U/238U measurements of IRMM-3050 (IRMM-074 se-
TABLE A1.1 Internal Measurement Precision (% RSE) for DS
Measurements of the Major Ratio 235U/238U ries) performed at JRC-G.2, taken from (13). See Fig. A1.2.
CRM 235
U/238U 235
U/U, % Average RSE, % A1.1.6 QC chart to show bias and between turret precision
Data from (13, 20) for 235U/238U measurements of IRMM-184 performed at JRC-
IRMM-020 0.00210 0.2 0.0036 %
IRMM-021 0.00441 0.4 0.0027 %
G.2, taken from (13). See Fig. A1.3.
IRMM-022, IRMM-184 0.00726 0.7 0.0022 % A1.1.7 QC chart to show bias and between turret precision
IRMM-023 0.03388 3.3 0.0015 %
IRMM-3050 (-074) 1.00026 50.0 0.0016 % for 235U/238U measurements of IRMM-184 performed at NML,
taken from (13). See Fig. A1.4.
238
TABLE A1.5 Results for DS Measurements of the Major Ratio U/235U for NBL CRM 112A
Laboratory, Reference Instrumentation Reference Materials used for Instrument Calibration NBL 112A 238U/235U
Uncertainty, k=2
NBL New Certified Value (21) Triton TIMS NBL-U030A, by MTE/TIMS (see Test Method C1832) 137.849(76)
“New Average Value” as introduced in (11) Triton TIMS, Neptune MC-ICPMS Double Spike IRMM-3636, and other Double Spikes 137.837(15)
calibrated using CRMs such as IRMM-074, IRMM-184
233
JRC-G.2/Geel (11) Triton TIMS U/236U Double Spike IRMM-3636, 137.836(23)
233
U/236U=1.01906(16)
233
NIGL Isotope Geoscience Laboratory (2010) Triton TIMS U/236U Double Spike IRMM-3636, 137.844(25)
233
(11) U/236U=1.01906(16)
233
NERC Isotope Geoscience Laboratory Triton TIMS U/236U Double Spike IRMM-3636, 137.832(22)
233
(2012) (16) U/236U=1.01906(16)
233
University of Frankfurt (11) Neptune MC-ICPMS U/236U Double Spike IRMM-3636, 137.833(28)
233
U/236U=1.01906(16)
233
Thermo Fisher Scientific (11) Neptune MC-ICPMS U/236U Double Spike IRMM-3636, 137.836(23)
233
U/236U=1.01906(16)
233
NERC Isotope Geoscience Laboratory Neptune MC-ICPMS U/236U Double Spike IRMM-3636, 137.829(22)
233
(2012) (16) U/236U=1.01906(16)
233
National Taiwan University, NTU, (11) Neptune MC-ICPMS U/236U Double Spike made by UMN, calibrated using 137.834(21)
IRMM-074 ⁄10, 238U/235U=0.99974(15)
233
University of Minnesota, UMN, (11) Neptune MC-ICPMS U/236U Double Spike made by UMN, calibrated using 137.830(25)
IRMM-074 ⁄10, 238U/235U=0.99974(15)
233
Lawrence Livermore National Laboratory Triton TIMS U/236U Double Spike made by LLNL, calibrated using 137.823(42)
LLNL, (11) IRMM-184, 238U/235U=137.697(41)
233
Arizona State University (18) Neptune MC-ICPMS U/236U Double Spike IRMM-3636, 137.843(23)
233
U/236U=1.01906(16)
233
Enrico Fermi Institute (19) Neptune MC-ICPMS U/236U Double Spike IRMM-3636, 137.842(23)
233
U/236U=1.01906(16)
238
FIG. A1.5 Results for DS Measurements for the Major Ratio U/235U for NBL CRM 112A
REFERENCES
(1) Callis, E. L. and Abernathy, R. M., “High-Precision Isotopic Analyses (9) Keegan, E., Richter, S., Kelly, I., Wong, H., Gadd, P., Kuehn, H., and
of Uranium and Plutonium by Total Sample Volatilization and Signal Alonso, A., “The Provenance of Australian Uranium Ore Concentrates
Integration,” Int. J. Mass Spectrom. Ion Proc., Vol 103, 1991, pp. by Elemental and Isotopic Analysis,” Applied Geochemistry, Vol 23,
93–105. 2008, pp. 765–777.
(2) Fiedler, R. et al., “Report on Preliminary Experience with Total (10) Cheng, H., Edwardsa, R. L., Shenc, C.-C., Polyak, V. J., Asmerom,
Evaporation Measurements in Thermal Ionization Mass Y., Woodhead, J., Hellstrom, J., Wang, Y., Kong, X., Spötl, C., Wang,
Spectrometry,” Int. J. Mass Spectrom. Ion Proc., Vol 132, 1994, pp. X., and Alexander Jr., E. C., “Improvements in 230Th Dating, 230Th
207–215. and 234U Half-life Values, and U-Th Isotopic Measurements by
(3) Fiedler, R., “Total Evaporation Measurements—Experience with Multi-collector Inductively Coupled Plasma Mass Spectrometry,”
Multi-Collector Instruments and a Thermal Ionization Quadrupole Earth and Planetary Science Letters, Vol 371–372, 2013, pp. 82–91.
Mass-Spectrometer,” International Journal of Mass Spectrometry, Vol (11) Richter, S., Eykens, R., Kühn, H., Aregbe, Y., Verbruggen, A., and
146, 1995, pp. 91–97. Weyer, S., “New Average Values for the n(238U)/n(235U) Isotope
(4) Mathew, K. J., O’Connor, G., Hasozbek, A., and Kraiem, M., “Total Ratios of Natural Uranium Standards,” International Journal of
Evaporation Method for Uranium Isotope-amount Ratio Mass Spectrometry, Vol 295, 2010, pp. 94–97.
Measurements,” Journal of Anal. At. Spectrom., Vol 28, 2013, pp. (12) Galdoz, E., Esteban, A., Cristallini, O., and Perrotta, J., “UF6
866–876. Sampling Method Using Alumina,” 49th INMM Annual Meeting,
(5) Richter, S. and Goldberg, S. A., “Improved Techniques for High USA, 2008.
Accuracy Isotope Ratio Measurements of Nuclear Materials using (13) Richter, S., Hiess, J., and Jacobsson, U., “Validation of Cristallini
Thermal Ionization Mass Spectrometry,” International Journal of Sampling, Method for UF6 by High Precision, Double-Spike
Mass Spectrometry, Vol 229, 2003, pp. 181–197. Measurements,” 2016, JRC-103668, https://ec.europa.eu/jrc/en/
(6) Richter, S., Kühn, H., Aregbe, Y., Hedberg, M., Horta-Domenech, J., news/validation-cristallini-method-sampling-uranium-hexafluoride-
et al, “Improvements in Routine Uranium Isotope Ratio Measure- uf6).
ments using the Modified Total Evaporation Method for Multi- (14) Richter, S., Alonso, A., Aregbe, Y., Eykens, R., Kehoe, F., et al, “A
Collector Thermal Ionization,” Mass Spectrometry Journal of Anal. New Series of Uranium Isotope Reference Materials for Investigat-
At. Spectrom., Vol 26, 2011, pp. 550–564. ing the Linearity of Secondary Electron Multipliers in Isotope Mass
(7) Richter, S., Alonso, A., Eykens, R., Jacobsson, U., Kühn, H., Spectrometry,” International Journal of Mass Spectrometry, Vol 281,
Verbruggen, A., Aregbe, Y., Wellum, R., and Keegan, E., “The 2009, pp. 115–125.
Isotopic Composition of Natural Uranium Samples—Measurements (15) BIPM, “Joint Committee for Guides in Metrology, Evaluation of
using the New n(233U)/n(236U) Double Spike IRMM-3636,” Interna- Measurement Data—Guide to the Expression of Uncertainty in
tional Journal of Mass Spectrometry, Vol 269, 2008, pp. 145–148. Measurement,” JCGM, Vol 100, 2008.
(8) Brennecka, G. A., Wasylenki, E., Bargar, J. R., Weyer, S., and Anbar, (16) Hiess, J., Condon, D. J., McLean, N., and Noble, S. R., “Systematics
A. D., “Uranium Isotope Fractionation during Adsorption to Mn- in Terrestrial Uranium-Bearing Minerals,” Science, Vol 315, 2012,
Oxyhydroxides,” Environ. Sci. Technol., Vol 45, 2011, pp. 1370–1375. pp. 1610–1614.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/