You are on page 1of 49

5E-C5808

Calorimeter

Service Manual

KY2.789.0

Changsha Kaiyuan Instruments Co., Ltd.


Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual
CONTENTS

1. Instructions ................................................................................................................... 1
1.1 Safety Precautions................................................................................................. 1
1.1.1 Safety of Electricity ...................................................................................... 1
1.1.2 Safety of Gas Supply ................................................................................... 1
1.1.3 Safety of Equipment .................................................................................... 2
1.2 Caution Symbols .................................................................................................... 3
1.3 Package ................................................................................................................. 3
2. Introduction ................................................................................................................... 4
2.1 Features ................................................................................................................. 4
2.2 Specifications ......................................................................................................... 4
2.3 Working Environment ............................................................................................ 5
2.4 Structure ................................................................................................................ 6
2.5 Electric Circuit ........................................................................................................ 8
2.6 Software ............................................................................................................... 17
3. Installation and Commissioning ............................................................................... 19
3.1 Industrial Tablet Installation and Setting ............................................................. 19
3.2 Settings before APP Installation .......................................................................... 26
3.3 APP Installation ................................................................................................... 27
3.4 APP Automatic Startup ........................................................................................ 31
3.5 Shutdown Settings ............................................................................................... 31
3.6 APP Configuration Modification ........................................................................... 31
3.7 APP Multilingual Settings .................................................................................... 32
3.8 Network Connection and Settings ....................................................................... 32
3.9 Balance Data Management ................................................................................. 32
3.10 Dual-Control ....................................................................................................... 33
3.11 Refrigeration Sheet Installation ......................................................................... 36
3.12 Heating Sheet Installation.................................................................................. 38
3.13 Requirements on Sealant and Silicone Grease Coating ................................... 39
3.14 Exchange Laser Ignition and Common Ignition ................................................ 41
3.15 Focusing for Laser Ignition ................................................................................ 43
3.16 Oxygen Filling and Discharging Head Maintenance ......................................... 45
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

1. Instructions

1.1 Safety Precautions


The Service Manual introduces the installation and diagnostics of this instrument.
Maintenance personnel should read and fully understand this manual and follow all
the precautions and warnings in this Manual. CKIC assumes no responsibility for any
accident or damage caused by failure, or use in disregard of instructions in this
Manual.
Maintenance personnel shall have professional technical training by CKIC and have
the skill to complete the relevant operations under the guidance of the Manual.

1.1.1 Safety of Electricity


 Ensure the equipment is well grounded. Do use three-phase plug and three-phase
socket. Do NOT change the original standard power cord.

 The thunder and lightning may cause a risk on you. Do not operate this equipment
during thunderstorm and it must be disconnect from facility power and external
internet wire.

 Do NOT operate the equipment together with high-power electrical equipment by one
power socket, lest it may result in unstable voltage.

 Do NOT plug out any component or cord before power-off.

 Regularly check the power cord and power plug and make sure they are in a good
condition and pressed by other components.

1.1.2 Safety of Gas Supply


 The gas required should be 99.5% pure oxygen. No electrolytic oxygen is allowed.

 Regularly check the gas regulator. The pressure in cylinder should not be less than 4
Mpa.

 Pressure in charging oxygen bomb should be not more than 3.2 MPa. Perform a gas
check to oxygen bomb and cylinder.

1
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

1.1.3 Safety of Equipment


 New oxygen bomb or the one with replaced components (bomb, bomb head,
closure ring) can only be used after 20.0 MPa hydraulic proving test.
 Regularly check the strengthening components of oxygen bomb, including thread
between closure ring and bomb, connecting parts of inlet valve, exhaust valve,
electrodes and bomb. Repair the oxygen bomb and perform hydraulic proving test
when visible wearing or loosening occurs.

Perform 20.0 MPa hydraulic pressure test to oxygen bomb periodically. To ensure safe
operation, please stop using the oxygen bomb and have hydraulic pressure test as well
as thread test when any of the following circumstances occurs.
1. Used for less than 4 years and it’s been over 2 years since previous test.
2. Used for over 4 years and it’s been more than a year since previous test.
3. Used for more than 5000 times since previous test.
4. Visible wearing closure ring.
When closure ring is placed right above the bomb, the closure ring falls more than two
threads without clockwise rotation.

Note:
When the same type of oxygen bombs is used but in different design, each bomb should
be used as a complete unit. The exchange of oxygen bomb components may result in
severe accident in operation.

2
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

1.2 Caution Symbols


The following "WARNING", "CAUTION" symbols will be applied to the instrument or its
components to ensure the safety of the instrument. In this Manual, the use of these
symbols and words will highlight the important safety information. Operators and
maintenance personnel must follow these safety instructions to prevent personal injury
and equipment damage.

This symbol indicates a caution.


(Refer to the manual for specific instructions.)

This symbol indicates a high temperature surface. Do


not touch it with hand. If needed, refer to the specific
instructions and operate in compliance with this Manual

This symbol indicates a risk of electric shock in case


you touch here with hand when the instrument is
working. Operate with caution. Refer to the specific
instructions and operate in compliance with this
Manual

This symbol indicates risk of squeezing hands.


(Refer to the manual for specific instructions.)

1.3 Package
After receiving instrument, please check the packing and items in your package. For any
damage or lack of components, please contact forwarder and CKIC to confirm
responsibility. The packing list for C5808 is as follows.
Note: All components listed below are subject to change without prior notice.
Please refer to the packing list received with instrument.

3
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

2. Introduction

2.1 Features
The main technical features and unique functions of 5E-C5808/5E-C5808J are as
follows:

 Automatic oxygen charging and discharging, air pressure examination, and oxygen
vessel leakage detection.

 Automatic oxygen vessel recognition and anti-forgery detection.

 Automatic water volume measurement and leakage detection.

 Integrated structure.

 Two-way temperature control. Able to realize cooling and heating.

 Set temperature control target automatically.

 Electric leakage and overheating protection.

 Only 8 min for single sample analysis.

 Filtering capability for the water of the whole system.

 Equipped with laser ignition function.

 Touch screen with the brand new user-friendly operation interface.

2.2 Specifications
1. Power supply: 220V±10V/50Hz±1Hz

2. Power: less than 800W

3. Air source: 99.5% cylinder oxygen( Electrolytic oxygen is strictly prohibited).

4. Air pressure: 2.8MPa-3.2MPa.

5. Temperature resolution: 0.00001℃

6. Ambient temperature: 15℃-30℃.

7. Ambient humidity: 20%-80%.

8. Single analysis time: 8 min

4
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual
9. Jacket temperature precision: <0.05℃

10. Calorific capacity RSD: ≤0.05%

11. Calorific capacity stability: ≤0.2%

12. Analysis range: 10000J-40000J

13. Accuracy: conforms to JJG 672, GB/T213, ISO1928, ASTM D5865, GB 384.

14. Dimension(L*W*H): 560*680*400(mm)

15. Weight: 65kg

2.3 Working Environment


1.Ambient temperature: 15~30°C; Relative humidity ≤80%.
2.No intense vibration, dust, strong electromagnetic interference or corrosive gas from
ambient.
3.Keep a separate room for analysis. Do NOT conduct other tests in the same room.
4.Keep room temperature as constant as possible. During an analysis, the temperature
variations cannot exceed 1°C and no strong air convection, heat source or fan is allowed.
5.The test room preferably faces north lest it should be exposed to sunshine. Or keep the
calorimeter away from direct sunshine.
6.A stable power source (220±22V/50±1Hz) should be equipped for the analysis room.
Or prepare with a good UPS or regulated power source with power not lower than 1000
W.

5
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

2.4 Structure

Figure 2.1 Front View

Figure 2.2 Back View

The structure of the instrument is as shown in figure 2.1 and 2.2, including the following
function modules:
1. Oxygen vessel recognition area: where to identify the vessel with the chip produced
by CKIC. The oxygen vessel not produced by the CKIC shall be unable to identify.
2. Measuring cup: it is inside the measuring cup cover and used to fix the volume of the
water in can.

6
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual
3. Water filling area: where to add the water to the instrument. Distilled or deionized
water is required. The maximum amount of added water is about 9 litres, which is about
110%. Please add it according to the actual water shortage of the instrument. Under
normal circumstances, when the water is more than 50%, analysis can be conducted.
4. Filter: it is inside the water tank and able to filter impurities in the system waterways.
Generally, the filter should be cleaned after the use of 6 months.
5. Water level window: it is used to observe the water level of the instrument. If the water
level exceeds the visual range of the window, the water level is about 100%, and the
water level is about 50% if it is in the middle of the water level window.
6. Power button: it is the self-reset switch. In the shutdown status, press for about 1
second to start. In the boot status, press for about 3 seconds to force shutdown. General
shutdown can be performed on the software operating interface.
7. Power inlet: connect the three-core power cord. The power supply shall be AC 220V
and shall be well grounded.
8. Exhaust port: it is used to discharge the exhaust from the oxygen vessel and shall be
connected with the exhaust pipe. Combined with the exhaust treatment device or
collector. It can also be exhausted to the outdoor if allowed.
9. Air source joint: it is the inlet of high pressure oxygen. Please use the special oxygen
pipe and joint provided by CKIC.
10. Jacket exhaust port: when it needs to drain the system water, the jacket water is
excluded through this port. Please use the special drainage connector assembly
provided by CKIC.
11. Tank exhaust port: when it needs to drain the system water, the tank water is
excluded through this port. Please use the special drainage connector assembly
provided by CKIC.

Cover
Intermedium ring
Piston
Thimble
Cylinder block
Pressure
gauge

Figure 2.3 Gas Circuit Diagram

7
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual
The gas circuit of C5808/C5808J calorimeter is consistent with that of 5E-CYZ. It consists
of 5 air valves, 1 oxygen charging head with oxygen filling and exhaust function, and air
pressure sensor as shown in Figure 2.3.

Figure 2.4 Water Circuit Diagram

The water circuit of 5E-C5808/C5808J as shown in Figure 2.4 includes jacket, can, cover,
tank, measuring cup, jacket inlet valve, measuring cup inlet valve, can inlet valve, jacket
and can inlet pump, jacket circulation pump, can outlet pump. The voltage of jacket and
can inlet pump and jacket circulation pump is 220V, and that of others is 24V.
The volume of tank is 9L, and the volume of jacket is 1L.

2.5 Electric Circuit


Circuit connection starts from the main board, and the core is in the main board (which is
divided into bottom plate and core plate), as shown in Figure 2.5. This main board is an
universal board,which could be applied to other instruments produced by CKIC.
C5808/C5808J only adopts the partial function of the main board, so the interfaces are
not all used. The interfaces used by C5808/C5808J is as shown in Table 2.1.
Take the J01 row interfaces in the top left corner of main board for example (as shown in
Figure 2.5), C5808/C5808J uses PWM1, PWM3 and PWM5, which are connected to
lifting motor driver U5-PUL-, stirring motor U7-blue PWM and lifting motor driver U5-
DIR-, respectively.
Mark Purpose Mark Purpose Remarks
Motor driver Piston reset
J01-+5v J04-Q13
U5-PUL+ valve FA1
Oxygen
Motor driver
J01-+5v J04-Q14 charging valve
U5-DIR+
FA2

8
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual
Temperature
Thimble valve
J01-+5v box fan U6- J04-Q15
FA3
5V
Thimble
Motor driver
J01-PWM1 J04-Q16 decompression
U5-PUL-
valve FA4
Stirring motor Exhaust valve
J01-PWM3 J04-Q17
U7-blue PWM FA5
Control jacket
Motor driver
J01-PWM5 J04-Q18 AC SSR U1- circulation
U5-DIR-
pump BE2
Measuring cup Control tank
J03-Q1 inlet valve J04-Q19 AC SSR U2- circulation
FA6 pump BE2
Control jacket
Jacket inlet
J03-Q2 J04-Q20 AC SSR U3- heating sheet
valve FA7
JR1
Can inlet Relay JDQ1- Main power
J03-Q3 J04-Q21
valve FA8 13 control
Diaphragm Relay JDQ2- Ignition wire
J03-Q4 J04-Q22
pump BE1+ 13 control

Laser power
Optical slot Upper optical
J03-Q6 supply U9- JI4-I1
WD1-black slot
VIN
Temperature Optical slot Lower optical
J03-Q7 JI4-I2
box U6-24V WD2-black slot
Temperature Stirring motor Stirring motor
J03-Q8 JI4-I3
box U6-JR U7-yellow -FG
Cooling sheet Ignition wire
J03-Q11 JI4-I4 Relay JDQ2-4
U4+ detection
Cooling sheet Power button
J03-Q12 JI4-I5
U4+ U10-2-A
Laser oxygen
JI4-I6 vessel
detection YDJ
Baroceptor
JI1-AI0
U11-white
Baroceptor
JI2-5V
U11-red
Baroceptor
JI2-AGND
U11-black

9
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual
Table 2.1 Electric Circuit Wiring Marks and Instruments

The wiring of power module is independent from the main board as shown in Figure2.6.
Starting from the socket, enter the leakage protection switch first, then the power button,
and enter the relay JDQ1 in parallel. The relay is controlled by the main board, and then
enter the switching power supply. When turning the power supply on by power button,
the switching power supply is energized and the main board is activated. After that, the
relay JDQ1 will be controlled to close. Therefore, the power button looses and the
switching power supply will not be cut off. The connection mode realizes the function of
cutting off power supply by software control.

10
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual
At the same time, the power button also provides a signal to the software to request
power off. The signal interface used is the power button's JI4-I5 (see Figure 2.6).

Figure 2.7 Wiring diagram of AC solid state relay

There are three AC solid state relays, which are connected to the jacket circulation
pump, the water tank circulation pump (the jacket and can inlet pump) and the jacket
heating sheet as shown in Figure 2.7. The relay is dc 24V controlling ac 220V.

11
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

Figure 2.8 Wiring diagram of stepping motor and optical slot

The lifting stepping motor driver is consistent with C5508, and the connection mode is also
consistent as in Figure 2.8. However, for accurate positioning of C5808/C5808J, optical
slot positioning is used instead of the microswitch of C5508.

Figure 2.9 Wiring diagram of guide rail

The function of the guide rail module is to extend the wiring, as shown in Figure 2.9. For
example, there are more than ten connectors for grounding. If one wire is tied to more
than ten wires, it is obviously unprofessional and inconvenient to unplug, so the guide rail
is used for extension.

12
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

Figure 2.10 Wiring diagram of air valve

Figure 2.11 Wiring diagram of lifting stepping motor and optical slot

Both air and water valves use dc 24V, as shown in figure 2.10 and figure 2.11. The water
inlet valve of the can is a ball valve.

13
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

Figure 2.12 Wiring diagram of radio card reader

Radio card reader module, as shown in figure 2.12. Direct access to the main board.

Figure 2.13 Wiring diagram of cooling sheet

The refrigeration module, as shown in figure 2.12, is directly connected to the main board.
The cooling sheet has a word side for the cold end, no word side for the hot end. If it
cannot be confirmed, a dry battery can be used to connect the positive and negative poles
of the refrigeration chip to the red line and the black line (cannot be reversed), and touch
the two sides of the refrigeration chip by hand. If the refrigeration chip is not damaged,
there will be one cold side and one hot side, then the cold end can be determined.
The installation of the cooling sheet should be coated with thermal conductive silicon
grease, as thin and uniform as possible. The pressure of screw tightening refrigeration
piece is moderate, and is about 1.5Mpa (1~2Mpa), the whole surface pressure of
refrigeration piece is about 240kg force.

14
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

Figure 2.14 Wiring diagram of stirring module

The stirring motor is a new brushless dc motor with feedback. The wiring is shown in figure
2.14.

Figure 2.15 Wiring diagram of laser ignition module (optional)

The wiring of the laser ignition module is shown in figure 2.15. For the details for focusing,
please refer to 3.9.1.

15
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

Figure 2.16 Wiring diagram of ignition wire module (optional)

The wiring of ignition wire ignition module (normal ignition) is shown in figure 2.16. Normal
ignition has the ignition wire detection function. If the function is complete, after debugging,
you should change the keyword ListenBombPlace to 1 to enable the function. This feature
allows the instrument to identify if the ignition wire is properly mounted (which can only be
identified when drop in place, thus it is not significant).

16
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

2.6 Software

The initial interface is as shown above, including the main barrel menu (as shown in Figure
2.17) and sub-barrel menu (as shown in Figure 2.18). The contents of those menus are
basically the same, which includes Test, Set, and Debug. The only difference is that the
main barrel menu includes Data function. In fact, the data function includes test data for
the entire instrument, not just the main or sub-barrel, so multiple menus are not required.

17
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

Figure 2.17 Main Barrel Menu

Figure 2.18 Sub-barrel Menu

The menu area contains the function menu of the system, in which the user can select
the required functions. When the user selects each function menu, the corresponding
contents will be displayed in the main display area of the main (sub) barrel for the
convenience of the user to view and operate. Please refer to the following related
sections for detailed instructions.
The system status area is located at the bottom of the main interface (as shown in figure
2.19) and is used to display relevant system information and system time.

Figure 2.19 System Status Area

For other operations, please refer to Instruction Manual.

18
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

3. Installation and Commissioning

3.1 Industrial Tablet Installation and Setting


Industrial tablet will automatically start up when plugged in (this function is set by the
manufacturer, otherwise it needs to be set in the BIOS Set).

1. Installation of the Industrial Tablet


This step could be skipped if the software has been installed when leaving factory.
The first time you turn on the tablet, you will see the following drawing, which
indicates that the device is a new machine. We need a series of settings for the
device to work properly and comfortably.
As shown in figure 3.1, we need to select the contents in the red boxes. Generally,
Set the country and the language as needed. Of course, it can be adjusted later.

Figure 3.1 Boot Screen for the First Time

Click Next to show the interface as in Figure 3.2.

19
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

Figure 3.2

Click “Use Express Setting” to accept update.

Figure 3.3 Input Product Key

This device does not directly provide win10 system legitimate registration code. It needs
network automatic registration. Therefore, to activate the win10 system, it requires
networking. Wifi is recommended. If there is no connection at present, click "Do this later"
and deal with the problem later.
Then, accept the agreement.
Then, the interface requiring the account name and password will appear. Fill in the
account name with "csky" and password shall not be input.

20
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual
Wait a few minutes, then enter the common win10 interface.
2. Settings
The system shall sleep in 10 minutes in default and requires a mouse or keyboard to
wake up, but there is no mouse keyboard! So you should first set it to never sleep
(otherwise you'll just have to unplug and reboot). The operation is shown in figure 3.4:

.
Figure 34 Setting

Enter Windows Settings, select Power & sleep, and select Never as shown in the red box.
Set the portrait screen in the same interface as shown in the Figure 3.5.

21
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

Figure 3.5 Portrait Screen

The USB flash drive auto-play function is turned off to avoid the pop-up box caused by
inserting USB flash drive.

22
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

Figure 3.7

In order to control user operations, you need to disable the Win10 tablet edge sliding
gesture. The operation is as follows:
1. Enter regedit in the Cortana search bar and press enter to enter the registry editor.
2. Locate to
HKEY_LOCAL_MACHINE\SOFTWARE\Policies\Microsoft\Windows\EdgeUI
Note: users without EdgeUI items need to manually create, you can click the right mouse
button on Windows, select new → item, and then rename to EdgeUI.
Create a new DWORD (32-bit) value in EdgeUI, rename it AllowEdgeSwipe, and keep the
numeric data as the default "0".

23
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

24
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

Figure 3.8 Disable Edge Sliding

This change takes effect after the computer is rebooted. To restore the edge slide
gesture, change the AllowEdgeSwipe data to "1" and restart the computer.

25
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

Figure 3.9 Tablet Mode

Set the startup screen in tablet mode as in figure 3.9.

3.2 Settings before APP Installation


In order to increase users' touch screen experience, this software is APP mode, that is,
smart phone or Pad mode. Installation is different from desktop software.
First, you need to set the system to developer mode (because the software is not in
Microsoft store) before the software can be installed as in figure 3.9. Select Developer
Mode. To confirm, select Yes.

26
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

Figure 3.9 Developer Mode

If the tablet system is not activated, some functions, such as interface diagram changes,
touch keyboard pop-ups, may not be used.

3.3 APP Installation


Use Windows PowerShell, the new super terminal of Windows 10, to install the program.
Put the installation package into C:\user\. Then long press the file path and copy the
path (useful later).

27
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

1 App Installation Software License

Figure 3.10 Installation License


Right click the App installation file c5808_app2_1.0.*.0_x86.appxbundle, right click (long
press for right click) to select properties, select digital signature, select "kevy8" in the
signature list, click details, view the certificate, and click install certificate. Select the local
computer, pop up a confirmation, select put all certificates in the following storage, click
browse, and select the second line "trusted root certificate authority" as shown in Figure
3.10 and 3.11.

28
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

Figure 3.11 Installation License


Click Next after confirmation to complete the certificate installation.

The installation license is installed successfully.

2 APP Installation
Open Windows PowerShell.

29
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

Open the touch keyboard and select English. Click Windows PowerShell to bring the focus
back to Windows PowerShell. Type add-appxpackage (add one space), long press
Windows PowerShell, paste C:\Users\csky\ c5808_app2_1.0.*.0_test \Dependencies\x86
path.
First, install Microsoft.NET.Native.Framework.1.3.appx ;
Then, install Microsoft.NET.Native.Runtime.1.4.appx;
Finally, install Microsoft.NET.VCLibs.x86.14.00.appx.
The installation files are entered as follows:
add-appxpackage C:\Users\csky\C5808_App2_1.0.*.0_Test\
Dependencies\x86\Microsoft.NET.VCLibs.x86.14.00.appx
"Add-appxpackage" is the installation keyword.
C:\Users\csky\C5808_App2_1.0.*.0_Test\Dependencies\x86\Microsoft.NET.VCLibs.x86.
14.00.appx is the installation package path.
The above 3 files only need to be installed once in the tablet (software version upgrade
does not need to be reinstalled).
Then input:
add-appxpackage
C: \Users\csky\C5808_App2_1.0.*.0_Test\C5808_App2_1.0.*.0_x86.appxbundle
Where * represents the version number. Use the Tab key when installing. The Tab key
can be typed with only the first letter to produce the full name. As shown in the picture.
Click Enter to install.

Figure 3.15 Install the program on the installation terminal

If you have installed different versions of c5808_app2_1.0.*.0_x86.appxbundle, there


may be conflicts and the installation fails. The installation failures are as follows:
1) the higher version has been installed, then install the lower version.

30
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual
2) installation file c5808_app2_1.0.*.0_x86.appxbundle is not licensed.

3.4 APP Automatic Startup


Put wumin.bat in
C:\Users\csky\AppData\Roaming\Microsoft\Windows\Start Menu\Programs\Startup
If you don't have a wumin.bat batch file, make one. Create a text file called wumin.txt
and copy the following:
@echo off
start "C:\Windows\System32\cmd.exe"
start myapp2:
taskkill /f /im cmd.exe
Exit
Then save as wumin.bat
If it is not visible in AppData, check the hidden folder in the folder menu -> view.

3.5 Shutdown Settings


You can turn off the tablet in software interface, but must install a driver to fulfill it.
Click vc_redist. exe (provide another exe installation file). After installation, you will have
the Microsoft Visual C++ 2015 RC Redistributable(x86) item in the program.

3.6 APP Configuration Modification


The configuration path is in:
C:\Users\csky\AppData\Local\Packages\csky_cr98wsyddebhr\LocalState\Setparak.ini
The name of the csky_cr98wsyddebhr folder is automatically generated by compilation,
and the names of the latter half may differ between compilers.
If it is not visible in AppData, check the hidden folder in the folder menu -> view.
Refer to Configuration File Parameters Description for configuration file parameters, in
which:
debug_permission as 0 is in non-debugging state. You cannot debug ignition, cannot
debug isothermal surface closure. If debugging is needed, it should be changed to 1.
After debugging, it should be changed to 0 again.
ListenBomb 0 is listening for oxygen bomb identification. If it cannot be identified,
analysis shall not be conducted. If the oxygen bomb identification function is complete,
the default value should be 0, otherwise 1.
The ListenBombPlace default 0 means not listening for the bomb's location. The function
is currently only applicable to common oxygen bombs. If the C5808 with ignition wire
and this function is complete, the configuration should be changed to 1.
Factory_Number is the instrument number. After debugging, the instrument number
needs to be filled in when the nameplate is printed.

31
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

3.7 APP Multilingual Settings


In C:\Users\csky\AppData\Local\Packages\csky_cr98wsyddebhr\LocalState\
Create Translate.txt. It contains all the interface character variables. Set the
corresponding character of the variable to any language character, and then select the
language "Other" in the program interface -> setting. Please refer to "multi-language
configuration parameters instructions" for details.
Note that the text length and font size selection after the translation, the language
characters must be text inputable.

3.8 Network Connection and Settings


This tablet controls the salve computer by network. There are the following situations
that need to change the tablet system IP, or the lower computer motherboard IP:
1. This instrument is required to be connected to the user network printer.
2. This instrument is required to access the user's data upload management system.
3. This instrument is needed to access the user's balance data management system.

Connection Operation:
1. If it is a single-control instrument, directly connect to the laboratory LAN through the
network cable from the back cover, such as switch or router, and then change the
settings of the external network card IP of the tablet system and the gateway set to
adapt to the laboratory LAN.
2. If it is a dual-control device, it is necessary to insert a network cable into each device,
and then connect it to the user's LAN, such as switch or router, and then change the
setting of the external network card IP of the tablet system, and reset the IP of the sub-
bucket lower machine.

3.9 Balance Data Management


After the connection with the laboratory LAN is completed, copy the plug-in program to
C:\Users\csky (no installation of green program). Open the configuration of the plug-in
and fill in the database path of the configuration with:
C:\Users\csky\AppData\Local\Packages\csky_cr98wsyddebhr\LocalState\C5808_input.d
b
The plug-in program is a customized program, which can be connected according to the
customer's needs (charge is required here, but if it is the latest balance management
system of CKIC, there is a ready-made plug-in program, no charge is allowed).
Plugins can be placed in any folder, generally recommended under C:\Users\csky. In
general, you need to set the boot start, create a shortcut to the. exe file
(c5808_datatrans.exe), and cut the shortcut to:
C:\Users\csky\AppData\Roaming\Miscrosoft\Windows\Start Menu\Programe\Statup

32
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

3.10 Dual-Control
Remove the industrial tablet from the sub-barrel and cover it with a plastic tablet model.
Open the left cover plate, remove the main board network cable, insert the network
cable extension line into the main board, change the left cover plate, complete the sub-
barrel hardware state.
Connect the network cable from the back of the main barrel to the back of the sub barrel
network port, or plug into the customer's router or switch, and then access the sub barrel
back network port.
When starting up, first open the sub barrel, then open the main barrel (note: dual control
startup must first open the sub barrel and then open the main barrel), enter the
configuration file:
C:\Users\csky\AppData\Local\Packages\csky_cr98wsyddebhr\LocalState\ Setparak.ini
You need to go to the desktop first. To enter the desktop, follow these steps, as shown
in figure 3.16.

Figure 3.16 Back to Desktop

Sliding the lower edge (only the lower edge is not prohibited). A response button along the
right side appears, as shown in the red box.
Click the button in the red box and it pops up:

Figure 3.17 Back to Desktop

Click Pad mode to enter desktop mode as shown in figure 3.17.

33
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

Figure 3.18 Desktop Mode

Click "main barrel debugging" and "sub-barrel debugging" to prepare to close the program
illegally (if you can't click if it turns gray, you may not click). At this time, all parts of the
lower machine are closed.Click the x in the red box to close the program illegally.
At this point, you can enter the corresponding folder to change the configuration, or set
the balance connection, tablet IP.
Set up the tablet IP first. In the setting, Ethernet will have two network cards, among which
Ethernet2 IP, the built-in network card of industrial tablet, is generally set as
192.168.18.253, the submask is 255.255.255.0, and the gateway is 192.168.18.254.
External network card, that is, Ethernet card connected to sub-barrel or external LAN,
should be provided according to the customer LAN Settings (if there is a customer LAN,
at least two IP's are available, one for the sub barrel slave computer, and one for the
external network card for the industrial tablet). For example, if the customer LAN segment
is 172.17.21.*** and 172.17.21.253 can be allocated, the external Ethernet card is set to

34
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual
172.17.21.253, the submask is 255.255.255.0, and the gateway is 172.17.21.254. As
shown in figure 3.19, click the red box to enter the normal IP Settings

Figure 3.19 Network Card IP Setting

Now enter
C:\Users\csky\AppData\Local\Packages\csky_cr98wsyddebhr\LocalState\Setparak.ini
Open Setparak.ini
According to the configuration specification, the keyword "Reser_Server_Permit" is
checked to indicate whether to change the IP of the slave computer. 0 is not changed by
default, 1 is changed (the change here is only once, after which the keyword will be
restored to 0). If you want to change, set it to 1.
Since the default IP of the lower computer is "192.168.18.101", it is generally necessary
to change different IP to achieve dual control, such as "192.168.18.102", but if you need
to connect to the customer's local area network at the same time, you need to set the IP
according to the local area network. The keyword "Reser_Server_IP" is the IP number to
set the lower computer. For example, the LAN segment of the customer balance data
management system is 172.17.21.***, which can be assigned to IP172.17.21.102 (note
that IP cannot be assigned automatically in the LAN, otherwise the main bucket cannot
find the sub bucket), set the keyword "Reser_Server_IP" to "172.17.21.102".
Also, set the keyword ShuangTong to 1 to indicate that dual control is enabled. Save and
close as in figure 3.20.

35
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

Figure 3.20 Configuration Modification

Restart the program C5808_App2, and the sub-barrel status bar will display the successful
setting of the lower machine IP.
If you need to turn off the sub and main barrels, just press the power switch (the lower
machine needs to be restarted, and the setting is valid).
Turn on the sub barrel power supply first, then the main barrel, you can achieve dual
control. It should be noted that for the dual control startup, the connection time of the lower
computer is longer.

3.11 Refrigeration Sheet Installation


The difficulty in the installation of refrigeration sheet is to evenly apply thermal
conductive silicon grease. The method is as follows:
Open the right cover (if not installed can be omitted), as shown in the figure

36
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

Water Cooler

Refrigeration
Sheet Mounting
Groove

Figure 3.21 Install Refrigeration Sheet

Apply a layer of thermal conductive silicon grease to the refrigeration sheet mounting
groove (the new aluminum cooling block has only one refrigeration sheet mounting
groove). The thickness of thermal conductive silicone grease is about 0.5mm.
Put the word side (cold end) of the refrigeration piece into the slot, and the lead wire shall
follow the direction (generally facing down) of the electronic design. Press the cooling
sheet lightly, and move it up and down from left to right, making the thermal conductive
silicon grease spread evenly (it is not easy to take the cooling sheet again, the cooling
sheet will be damaged). Thermal conductive silicone grease should not overflow the
groove in large quantities, if overflow, it should be erased.
Then apply thermal conductive silicon grease to the non-word (hot end) of the refrigeration
sheet, which is also required to be uniform, with a thickness of about 0.5mm.
The heat transfer surface of the water cooler is pressed against the surface of non-word
(hot end) of the refrigeration sheet, and the upper and lower left and right micro
displacement is carried out to spread the thermal conductive silicon grease evenly.
Tighten the spring screw.
The best compression strength of the refrigeration sheet is 1.5mpa ± 0.5mpa, and the
pressure can be properly adjusted with a pressure wrench.

37
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

3.12 Heating Sheet Installation


The difficulty in the installation of the heating sheet is to evenly attach the heating sheet
to the jacket cavity. The non-adhesive parts of the heating sheet shall be coated with
thermal conductive silicon grease (some of the heating sheets are completely adhesive,
in this case, the thermal conductive silicon grease shall be ignored), which shall not
produce obvious bubbling, especially the adhesive parts.
Finally, lay the insulating sponge and fix it with the fastening sleeve.

Figure 3.22 Install Heating Sheet

The lay direction and lead are shown in the figure, and the lead direction is based on the
standard that no pulling force is generated.Then apply insulated cotton, as shown in figure
3.23.

38
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

Figure 3.23 Install Insulated Cotton

After laying adiabatic cotton, use fastening ring or glue to tighten.

3.13 Requirements on Sealant and Silicone Grease Coating


C5808 calorimeter has several requirements for sealing glue and silicone grease. The
purpose is fastening, waterproofing, lubrication. The following pink represents the
sealing glue, light blue represents the silicon grease.
1. Laser Oxygen Vessel

Sealing glue on Silicone grease is


top of lens holder applied to the
sealing ring of
oxygen vessel.

Sealing glue on radio


frequency chip of
oxygen vessel.

39
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual
Figure 3.24 Sealant and Silicone Grease Coating for Laser Oxygen Vessel
The laser oxygen vessel lens holder needs sealing glue. It is required that the
transmittance plate be fixed tightly and sealed horizontally (oxygen bomb placed
horizontally). The key is not to contaminate the transmittance plate center. Use
transparent or translucent silicone. Execute according to the requirements of the
drawing.
For oxygen vessel RF chip sealing, fix the RF chip tightly and in the right direction. Use
transparent or translucent silicone. Execute according to the requirements of the
drawing.
The oxygen vessel sealing ring is coated with silicone grease for lubrication, which shall
be thin and uniform. It is forbidden to be coated with silicone grease beyond the sealing
ring range. Use high temperature silicone grease. Execute according to the
requirements of the drawing.

2. Cover Buffer Sponge

Buffer sponge sealant

Figure 3.25 Cover Buffer Sponge Sealant

Buffer sponge is a ring sponge cut by tooling. It needs to be fixed to the tank of the
cover. Use transparent or translucent silica gel to coat the upper surface of the buffer
sponge with 0.5~1mm thickness, and stick it tightly. Execute according to the
requirements of the drawing

40
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual
3. Can Assembly Sealing Adhesive

Fixed ring seal ring is


coated with silicone
grease
Outer ring sealing
adhesive
Inner ring sealing
adhesive

Figure 3.26 Can Assembly Sealing Adhesive

The fixing ring and can assembly is a key component that requires sealing and aesthetics.
The sealing glue of the outer ring is required to form an chamfered angle with a uniform
surface. Inner ring seal requires seamless treatment and even surface; The inside also
needs to be glued tight. The sealing ring shall not be coated with silicone grease beyond
the sealing ring position. Use transparent or translucent silicone. Execute according to the
requirements of the drawing.

3.14 Exchange Laser Ignition and Common Ignition


Although laser ignition and common ignition are two different models, respectively 5E-
C5808J and 5E-C5808. But they do a lot of compatible design.
Only the two cover assemblies are mechanically inconsistent.
Only transformers are different on electronic circuits, as shown in FIG. 2.15 and FIG.
2.16 in chapter 2.5.

41
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

Figure 3.27 Laser Ignition Cover C5808J-06

Figure 3.28 Common Ignition Cover C5808-06 (Ignition Wire)

When replacing, replace the cover assembly, and then reconnect according to FIG. 2.15
and FIG. 2.16 in chapter 2.5.

42
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

3.15 Focusing for Laser Ignition


Laser Source

Laser Oxygen Vessel

Flow Baffle

Crucible

Laser Holder

Laser Tube

Figure 3.29 Laser Ignition


Focusing Structure
Component

Fastening
Component

Figure 3.30 Laser Source Structure

43
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

Adjust Current Increase


Clockwise
Output+

Input+
Output-
Input-

Display
Switch Voltage
Increase
Switch Set
Voltage
Decrease

Figure 3.31 Constant Current Source (on the cover)

Laser ignition includes laser source, constant current source and laser oxygen bomb .
Laser tube power is not very high, and parallel light and astigmatism are impossible to
achieve ignition. Therefore, a focusing element is used to focus the laser of the laser tube.
Focusing requires the focus to be near the crucible. See below for details.
[Commissioning]
1. Take a debugged C5808 calorimeter, turn it on to the configuration file and set the user
mode to debug mode, namely debug_permission=1, open the cover, find the constant
current source, and be ready (you can wear sunglasses). In the absence of power, use a
special screwdriver(-), counterclockwise "adjust the current" knob, adjust to the minimum
value (pay attention not to use a strong force, in case of damage).
2. Loosen the laser source fastener by hand; Use special 1mm hexagon screw driver to
loosen the top wire. Hang up the laser oxygen vessel cap and put it into the crucible.
3. Enter the debugging interface and turn on the ignition debugging. At this time, the
software keeps the ignition on (it should be noted that if the constant current source is not
adjusted to the minimum value, or before unwinding, the ignition debugging cannot be
turned on, otherwise the laser tube will burn out).
4. With a special screwdriver(-), turn clockwise the "adjust current" knob, slowly increase
the current (pay attention not to use the strong force, in case of damage), until it displays
as about 0.A and laser lights on.
5. Slowly twist the focusing part by hand and observe whether the focus falls into the
crucible. The closer the focus is to the center of the crucible, the better. The focal point
should be 1mm strip spot with a slightly larger center. The spot is the laser source diagram.
6. Turn off ignition debugging. With a special screwdriver(-), adjust the "adjust current"
knob clockwise to the maximum value (be careful not to use strong force to prevent

44
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual
damage). Use special 1mm inner hexagon screw driver to tighten the top wire. Tighten the
laser source fastener by hand. Install oxygen bomb positioning device.
7. Put a special ignition piece in the crucible and click ignition in debugging interface.

[Judgement Standard]
The closer the focus is to the center of the crucible, the better. The focal point should be
1mm strip spot with a slightly larger center. Transparent.

3.16 Oxygen Filling and Discharging Head Maintenance


The core component of automatic oxygen charger is the oxygen filling and discharging
head, which is a precision element. It is responsible for discharging exhaust gas which
contains dust, high corrosive gas, and moisture. Those ingredients affect cylinder cavity
and exhaust vent greatly. Thus, it is necessary to maintain oxygen filling and discharging
head regularly.
The structure of oxygen filling and discharging head is shown in Fig 3.32:

Fig 3.32 Oxygen Filling and Discharging Head

The major daily maintenance parts are exhaust joint together with air tube, air cavity
inside cylinder and equipped o-rings 18*1.8 and 2*1.8, thimble and equipped o-rings
16*1.8 and 13.2*1.8, and piston and equipped o-ring 6.7*1.8.

45
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual

Maintenance Procedures:
Open the cover plate of oxygen filling and discharging head, unscrew three screws, and
disassemble cylinder cavity, middle ring, piston, thimble in sequence. No need to
disconnect the air tube. The disassembly is shown in Fig 3.33.

Fig 3.33. Disassembly of the Oxygen Filling and Discharging Head

Clean the dust in cylinder cavity, piston, and middle ring with wet towel or tissue, and then
dry them with tissue or dry towel. Then, replace the damaged o-rings. Apply special high
temperature silicone provided by CKIC to the interior of the cylinder, outer ring of the piston,
inside of middle ring, and the outer ring of thimble. The coating should be as even and thin
as possible. The silicone is used to lubricate and seal, ensuring smooth operation of the
oxygen filling and discharging head.

46
Changsha Kaiyuan Instruments Co., Ltd C5808 Service Manual
Attention: to avoid falling or deformation, please gently disassemble and install for the
oxygen filling and discharging head is a precision element. No scratches on cylinder or
piston by sharp objects in case of sealing problems.

Vulnerable Parts:
No. ERP No. Name Specifications Remarks

1 213038100 Nut KY/QT05-003 General

2 213038099 Clamp KY/QT05-002 General

3 3090500003 High pressure pipe PX1025P04-02 General

Original
O-ring-16×1.8- 17×1.8
4 3090101260 GB/T3452.1-2005
fluororubber
(3090101130)

For piston
O-ring-13.2×1.8-
5 3090101199 GB/T3452.1-2005 and oxygen
fluororubber
vessel YD50

O-ring-6.7×1.8-
6 3090101173 GB/T3452.1-2005 For thimble
fluororubber

O-ring-18×1.8-
7 3090101148 GB/T3452.1-2005 For cylinder
fluororubber

O-ring-2×1.8-
8 3090101175 GB/T3452.1-2005 For air vent
fluororubber

47