Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

Design of a high sensitivity MEMS piezoresistive

intracranial pressure sensor using three turns meander


shaped piezoresistors
Mazita Mohamada,d, Norhayati Soina,c,l, and Fatimah Ibrahimb,c
aOepartment of Electrical Engineering, bOepartment of Biomedical Enginering,
Faculty of Engineering, University of Malaya,
Kuala Lumpur, Malaysia
Inorhayatisoin@um.edu.my

CCenter for Innovation in Medical Engineering (CIME), University of Malaya,


Kuala Lumpur, Malaysia
dOepartment of Electronic Engineering,
Faculty of Electrical and Electronic Engineering, Universiti Tun Hussein Onn Malaysia (UTHM),
Batu Pahat, Johor, Malaysia

Abstract--- The advancement in MEMS technology has Ideally, the ICP pressure sensor should be durable, bio-
triggered the development of small sensing devices such as compatible and inexpensive. As ICP monitoring deals with
intracranial pressure sensor. As the intracranial signal is a the well-being, therefore accuracy and sensitivity of the ICP
pulsating type, the sensor must be sensitive enough to detect these pressure sensors are very crucial. The sensitivity of the
changes. Hence, the important parameters of intracranial commercially available ICP microtransducer is 5 fl VN ImmHg
pressure sensor are the sensitivity and linearity. At present, the (Gaeltec & Raumedic). Based on the academic researches, the
sensitivities were within the range of 2 I1V/v/mmHg to 0.17 sensitivities of the piezoresistive sensors were within the range
mV/v/mmHg with the smallest sensor size of 500 11m x 500 11m
of 2 fl VN ImmHg to 0.17 mVN ImmHg [3-10]. To obtain the
(for adult). This paper presents the improved design of MEMS
accurate readings, several factors need to be considered such
piezoresistive pressure which aims at achieving higher sensitivity
as the shapes of piezoresistors and the dimensions of the
for subarachnoid space microsensor by using the three turns
meander shaped piezoresistors. A square diaphragm of200 11m x
diaphragm. For the ICP application, its critical point is the
size of the sensor itself. It needs to be small enough to fit into
200 11m x 2 11m is set for the sensor. Different piezoresistors
configurations were simulated using COMSOL Multiphysics to
the probe. The smallest size of ICP pressure sensor for the
determine the better sensing outputs. The position of each adult is 500 flm x 500 flm with the sensitivity of 2
piezoresistor was based on von Mises stress. By choosing the fl VN ImmHg, but it is for the lumbar area [10].
shape of piezoresistor that occupies most of the high stress This paper presents the design of an ICP piezoresistive
region, the change in resistance per pressure will be higher. The
pressure sensor using three turns meander shaped
design which consolidates the three turns and zero turn meander
piezoresistors. Thus far, only the one turn meander shaped
shaped piezoresistors was found to be the best with the sensitivity
of 0.125 mV/v/mmHg and linearity of 99%. This design has
piezoresistors have been applied in the designs ofICP pressure
fulfilled the specification ofthe intracranial pressure sensor. sensor [5, 7, 9, 10] while [3] have used the three turns
meander shaped piezoresistors on the cantilever, not on the
Keywords- intracranial pressure sensor; MEMS; piezoresistive diaphragm like this work. The size of diaphragm used is 200
flm x 200 flm x 2 flm using n-type silicon. The targeted
I. INTRODUCTION location of ICP pressure sensor is the subarachnoid space with
The sensors developed using the MEMS technology have the measurable pressure within the range of 0 - 100 mmHg.
tremendously shrunk in size make them ideal for biomedical
II. PIEZORESISTIVE PRESSURE SENSOR
application like intracranial monitoring devices. Using a small
pressure sensor requires a small intubation, makes the healing A piezoresistive pressure sensor is normally consists of 4
time shorter and minimizes the risk of infection. Based on [1], piezoresistors arranged in Wheatstone bridge configuration as
the most common location for intracranial pressure (lCP) illustrated in Fig. I. When no pressure is applied, the output
monitoring is cerebral parenchymal and contemporary voltage, Vaut is zero as all the piezoresistors are equal,
intracranial pressure sensor of that location is a piezoresistive therefore the bridge is in a balanced state. When the
pressure sensor. Although the piezoresistive output is diaphragm deflects due to the applied pressure, these
sensitive to temperature changes, it could be corrected easily piezoresistors' values change accordingly. For the p-type
by using a simple compensation circuit, namely Wheatstone piezoresistors oriented at <110> on an-type (100) silicon
bridge [2]. substrate, change in resistance value is expressed in (I) [2].

978-1-5090-4568-6/16/$31.00 ©2016 IEEE


Rj diaphragm
r - - - -0 - -'~-J""-l--------
: --I
I R2 R31

b d
[J \J
I 1
1 R,
1
~ ___ CJ ____ ~

(a) (b) (c)

Fig. l. (a) Schematic ofpiezoresistive pressure sensor (top view). (b) Pressure sensor (side view). (c) The equivalent Wheatstone configuration.[ll]

TABLE I. DESIGN SPECIFIC AnONS OF ICP PRESSURE SENSOR


(1)
Parameter Value
lr, and lrt are piezoresistive coefficients of the material Pressure range [13] 0-100 mmHg
along longitudinal and transverse directions and their Size dimension [10]:
expressions are indicated in (2) and (3). 0', and O't are the Sensor 500 11m x 500 11m
stresses along longitudinal and transverse directions [2]. Diaphragm 200 11m x 200 11m
Sensitivity [10] >0.01 mV/v/mmHg
(2) Linearity [3-10] >90%
Input voltage, Vin [13] IV
(3)
tebal(4) =2e-6 Suriace: von Mises stress (MPa) Slice: x-coordinate (11m)
Slice: y-coordinate (I-Im)
Parallel piezoresistors will decrease and the perpendicular y
~ 35.2

ones will increase [II]. Due to the symmetrical designs of 35


these piezoresistors, the changes in resistances are equal
30
depending on the arrangement, i.e. I1R, = I1Rt and I1R2 = I1RJ .
Thus, the output voltage will also change as expressed in (4) 25

where R is the value of resistance when no pressure is applied. 20

(I1RI +I1R 2 )
V =
Ollt 2R+I1R2 -I1RI
·v
111
~)
15

10

III. DESIGN METHODOLOGY


The sensors were designed based on the specifications y
summarized in Table I. The size of the sensor was set in Lx
order to fit into 4FR probe. The sensitivity of greater than
0.01 mV/V/mmHg was set as the previous research for the
same sensor size is 2 f..lV/V/mmHg [10]. The design stage was
segregated into two, i.e. determine (a) the location for the
piezoresistors on the diaphragm size of 200 f..lm x 200 f..lm x 2
/lm and (b) the best piezoresistors configuration. A square
diaphragm is selected because it has 60 % higher mechanical
stress instead of circular one for the same thickness [7].
A. Selection ofPiezoresistors Placement, Size and Shape
P-type piezoresistors (with doping concentration of 10 15
atomlcm3 ) oriented at <110> on an-type (100) silicon
substrate were chosen in the designs. These piezoresistors Fig. 2. Surface von-Mises stress of the diaphragm at pressure of 100 mmHg.
need to be placed in the high stress region [2, 11, 12]. Hence,
the von Mises stress result from the COMSOL Multiphysics more mechanical stress than the central point [2]. As the 2 /lm
simulation was used as illustrated in Fig. 2. The high stress thickness is considered a thin diaphragm, the high stress
region was found within the diaphragm and near the center of region is not extended into the bulk silicon area as mentioned
the edges, i.e. the red region. This is in agreement with the in [II, 12]. Therefore, the four piezoresistors were positioned
theory which states that the center of edges are experiencing at the center of the diaphragm's edges.
Three types of piezoresistors were used in the design as in resistance values while the perpendicular piezoresistors (R2
depicted in Fig. 3. The piezoresistors were comprised of and R3 ) increase in resistance values as the pressure increases.
piezoresistive bars (p-type) and connecting bars (metal). The The 3-turns meander shaped piezoresistor located
connecting bars will not contribute to the piezoresistive effect perpendicular to the diaphragm experiences the most change,
[12]. The piezoresistors sizes were designed to fit into the +/l..R/R = 1.6E-3 per 10 mmHg. [t is followed by the I-turn
high stress regions [II]. Hence, the dimension of O-turn piezoresistor placed perpendicular to the diaphragm in which
meander shaped piezoresistor was set to 80 /lm x 5 /lm. To the change in resistance, +/l..R/R = 1.1 E-3 per 10 mmHg. The
ensure the equal resistances at the no load condition, the total results concur with [II, 12] that the sensitivity will be
length of the piezoresistive bar must be equal [11]. Therefore, improved as the number of vertical bar increases. It is due to
the piezoresistive bars of I-turn and 3-turns meander shaped the maximum distribution area of the piezoresistor on the
piezoresistors were set to be fractions of 80 /lm as shown in high-stress region in comparison to the other types of
Table 2. piezoresistors [II].
B. Selection ofDifierent Piezoresistors Configurations However, the placement of piezoresistor also plays an
Four models of different piezoresistors sensors were important factor [2, 14] as illustrated in Fig. 6. The 3-turns
selected as shown in Fig. 4 in accomplishing the high
sensitivity sensor. The models were selected based on the r--=---, r - - c::==::::J'. - -,

researches by [11, 12]. In order to test their performances, the ,


I

simulations were done by using COM SOL Multiphysics


through Structural Mechanics Module - Piezoresistivity,
Boundary Currents. Combinations of tetrahedral mesh (for
,,
I

.c::==::::J
,
__ J
~
: I
W ~
, R,
L-. _ ..... _ ,.C::==::::J. _ _ J ~---l:J---i
diaphragm and substrate) and quadrilateral mesh (for circuit
elements) were used in this project. Modell Model 2
:R2 ~;
D
,
D I

IV. RESULTS AND DISCUSSIONS : ___ D~ ____ :


~ ~
~--~---i
To ensure the linearity and sensitivity performance of the , ,
pressure sensor, several parameters need to be considered.
These critical key factors include the shape and location of
piezoresistors.
i ,

~ __-.aJ___ ~
I I

A. Analysis ofSensitivity PerfOrmance ofthe Sensors


Model 3 Model 4
Fig. 5 shows the output voltage of each model. All the Fig. 4. Four models of different piezoresistors configurations.
models achieve the targeted specification for sensitivity, i.e.
above 0.01 mVIV/mmHg. Model 2, which comprised of 0- 12
turn and 3-turns piezoresistors gives the highest sensitivity of
0.125 mVIV /mmHg, followed closely by Model 3 which is 10

composed of I-turn and 3-turns meander shaped piezoresistors .....


with the sensitivity of 0.122 mVIV/mmHg. Meanwhile, :;
Modell gives the lowest sensitivity of 0.101 mVIV/mmHg. oS 6
"5
g 4
The results are supported with the graph of change in
resistance of different piezoresistor shapes, illustrated in Fig.
6. Parallel piezoresistors (R 1 and R4 of each model) decrease

20 40 60 80 100
connecting bar piezoresistive b a w
~(==l~Pb~:j~~i~
Iii Wpb~rl 5 ~
Applied pressure (mmHg)

I:
./ gap Fig. 5. Output voltage of each model at different pressure settings.
~
Web
(a) (b) (c)
Fig. 3. Meander shaped piezoresistors (a) zero turn (O-turn) (b) one turn
(l-turn) (c) three turns (3-turns).

TABLE II. GEOMETRIC PARAMETERS OF P1EZORES1STORS

Geometric Parameter Zero turn One turn Three turns


lpb 80 40 20
Wpb 5 5 5
Web - 10 10
g,p - 5 5 Applied pressure (mmHg)

(All values are III J.lm) Fig. 6. Change in resistance for diflerent shape ofpiezoresistors.
piezoresistor placed in parallel experiences the least change in from 0 to 100 mmHg. However, the best linearity
resistance, -I'!.RlR = 0.70E-3 per 10 mmHg. This is why performance is by Modell which composed of the O-turn and
Model 4 does not give the highest sensitivity. For 2-turns meander shaped piezoresistors. This shows that the
piezoresistor placed in parallel, the most change happen in the sensitivity and linearity ofthe sensor are very much influenced
O-turn piezoresistor, i.e. -I'!.RlR = 0.92E-3 per 10 mmHg. by the shapes of piezoresistors and their placements on the
Hence, Model 2 which composed of O-turn piezoresistors diaphragm. The 3-turns meander shaped piezoresistor
(placed in parallel) and 3-turns piezoresistors (placed in positioned perpendicular to the diaphragm gives the highest
perpendicular) gives the best sensitivity of all. change in resistance which results in the highest sensitivity.
B. Analysis ofLinearity PerfOrmance ofthe Sensors ACKNOWLEDGEMENT

The non-linearity of the sensors is calculated using The work is supported by the University Malaya Research
expression (5) [12] where A and Pmax are the calibrated and Grant UMRG-AET (Grant Number RPOI4B-13AET). The
maximum pressure respectively while VoutCA) and VoutCAnax) authors would also like to express their gratitude to the Micro
refer to the output voltage at these pressures. Table 3 shows Electronic Laboratory, Faculty of Engineering for facilitating
the non-linearity and sensitivity of each model. the research work.
REFERENCES
v:out (p.) v,;,APmaJ p
1 1

NL,= Pm" .100% (5) [1] G. Citerio and P. 1. D. Andrews, "Intracranial pressure part two: clinical
~lf{(Pmax) applications and technology," in Applied Physiology in Intensive Care
Medicine. ed: Springer, 2009, pp. 109-112.
Overall, the linearity performances of the sensors fulfill the
[2] S. S. Kumar and B. Pant, "Design principles and considerations for the
specification as all the models' linearity was above 90%. The 'ideal'silicon piezoresistive pressure sensor: a focused review,"
worst linearity performance (in which the non-linearity of Microsystem technologies. vol. 20, pp. 1213-1247, 2014.
1.I2E-3%) was by Model 4 which comprises of three turns [3] I. Clausen, S. Moe, and A. Vogl, "Design and processing of a cost-
piezoresistors only. This is in trend with the results obtained effective piezoresistive MEMS cantilever sensor for medical and
by [11, 12]. The more number of turns ofpiezoresistor causes biomedical use," Journal ofMicromechanics and lvlicroengineering. vol.
the distinct variation of resistance between the perpendicular 22,p.074008,2012.

and parallel bars which contributes to worsen linearity [11]. [4] W. Hasenkamp, D. Forchelet, K. Pataky, 1. Villard, H. Van Lintel, A.
Bertsch. et aI, "Polyimide/SU-8 catheter-tip MEMS gauge pressure
sensor," Biomedical microdevices. vol. 14, pp. 819-828,2012.
TABLE III. SUMMARY OF EACH MODEL'S CHARACTERISTICS
[5] L.-A. Liew and V. M. Bright, "Disposable CMOS catheter-tip pressure
Parameter Modell Model 2 Model 3 Model 4 sensor for intracranial pressure measurement," in Microtechnologies in
Medicine and Biology, 1st Annual International. Conference On. 2000,
Sensitivity
0.l01 0.125 0.l22 0.114 2000, pp. 130-135.
(mV/v/mmHg)
[6] X. Liu, Y. Yao, J. Ma, Y. Zhang, Q. Wang, Z. Zhang, et aI, "Micro
Non-linearity (%) 2.40E-04 8.82E-04 9.11E-04 1.12E-03
packaged MEMS pressure sensor for intracranial pressure
measurement," Journal ofSemiconductors. vol. 36,2015.
C Comparison with the Existing MEMS Piezoresistive ICP [7] X. Meng and Y. Zhao, "Packaging a piezoresistive pressure sensor for
Sensors intracranial pressure monitoring," in SENSORS, 2014 fEEE, 2014, pp.
1827-1830.
Table 4 shows the comparison of Model 2 with the [8] B. Pang, Z.-H. Zhang, and I.-L. Ren, "Simulation and design of micro
existing MEMS Piezoresistive [CP Pressure Sensor. Model 2 pressure sensors applied to measure the intracranial pressure," in
is considered to be an improved version for small diaphragm NanollvIicro Engineered and Molecular Systems (NElv!S). 2013 8th
although [9] gives the highest sensitivity performance and [10] IEEE International Conference on, 2013, pp. 120-123.
has the smallest diaphragm size. [9] Z. Wu, N. Bhatlacharjee, C. Li, J. Hartings, R. Narayan, and C. H. Ahn,
"A new intracranial pressure sensor on polyimide lab-on-a-tube using
exchanged polysilicon piezoresistors," in Solid-State Sensors. Actuators
TABLE IV. COMPARISON WITH EXISTING MEMS PIEZORESISTIVE ICP
and lvlicrosystems (JRANSDUCERS & EUROSENSORS XXVII). 2013
PRESSURE SENSORS
Transducers & Eurosensors XXVII: The 17th International Conference
Parameter Mode12 [7] [9] [10] on. 2013, pp. 1779-1782.
200 J.!m x 900 J.!m x radius [10] Y. Zhang, Z. Zhang, B. Pang, I.. Yuan, and T. Ren, "Tiny MEMS-based
150 J.!m x
pressure sensors in the measurement of Intracranial Pressure," Tsinghua
Diaphragm size 200 J.!m 900 J.!m 500 J.!m 150 J.!m
Science and Technology, vol. 19, pp. 161-167,2014.
(square) (square) (circle) (square)
Sensitivity [II] Y.-H. Zhang, C. Yang, Z.-H. Zhang, H.-W. Lin, I..-T. Liu, and I.-I..
10J.!V/V 2 J.!V/v
0.l25 0.17 Ren, "A novel pressure microsensor with 30-J.!m-thick diaphragm and
(mV/v/mmHg) ImmHg ImmHg
meander-shaped piezoresistors partially distributed on high-stress bulk
Non-linearity (%) 8.82E-04 <1.2 0.4 0.145 silicon region," Sensors Journal, IEEE, vol. 7, pp. 1742-1748,2007.
[12] S. S. Kumar, A. K. Ojha, R. Nambisan, A. K. Sharma, and B. Pant,
"Design and simulation of MEMS silicon piezoresistive pressure sensor
V. CONCLUSION
for barometric applications," Proceedings of dle ARTCom&AR1EE
Overall, all the models achieved the targeted sensitivity PEIE&itSfP and POE Elsevier. ISBN978-81-910691-8-J. pp. 339-345,
2013.
and linearity of ICP pressure sensor, i.e. above 0.01
[13] Gaeltec, "Product Guide," ed. UK, 2011, p. 19.
mV/V/mmHg and 90% respectively. This sensitivity is higher
[14] A. Nisanth, K. Suja, and R. Komaragiri, "Sensitivity enhancement of a
than the one achieved by the similar sensor size. Model 2
silicon Based MEMS pressure sensor by optimization of size and
which consolidates the O-turn and 3-turns meander shaped position of piezoresistor," in Electronics and Communication Systems
piezoresistors gives the highest sensitivity for pressure range (ICECS). 2014 International Conference on, 2014, pp. 1-5.

You might also like