Fuji Drichem Nx600

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 268

EN

INSTRUCTION MANUAL
Safe Usage and
Handling Precautions 1

FUJI DRI-CHEM Part Names and


Consumables 2
CLINICAL CHEMISTRY ANALYZER

FUJI DRI-CHEM NX600 Measurement 3

FUJI DRI-CHEM NX600i Applicable Usage


Methods 4

Loading Consumables 5

Maintenance 6

Quality Control 7

Troubleshooting 8

Function Menu 9

Other Functions 10

USM 11

Specifications 12

Glossary 13
This Manual describes details on how to operate
the FUJI DRI-CHEM NX600/NX600i and
cautions to be observed when operating it.
Please read this manual thoroughly before

2nd Edition
actually operating the FUJI DRI-CHEM NX600/
NX600i. After reading this manual, store it
nearby the FUJI DRI-CHEM NX600/NX600i so
that you can see it whenever necessary.
897N203029A January 2022
Introduction
Thank you for purchasing the FUJI DRI-CHEM NX600 or FUJI DRI-CHEM NX600i.
Accompanying documents were originally drafted in the English language.
Installation may only be conducted by authorized service personnel.

<Intended Purpose>
The FUJI DRI-CHEM NX600 and FUJI DRI-CHEM NX600i (henceforth, the FDC NX600) is an in vitro
diagnostic medical device to automatically quantitate the concentration or the activity of the components
in blood or urine by colorimetric end-point, rate and ISE tests, using the FUJI DRI-CHEM SLIDEs. Do
not use the equipment for other purposes.
This equipment must be operated by Healthcare professionals who are familiar with its intended pur-
pose and correct operation. And, the equipment can be used in a near patient setting by Healthcare
professional.

CAUTIONS
1. No part or all of this manual may be reproduced in any form without prior permission.
2. The information contained in this manual may be subject to change without prior
notice.
3. FUJIFILM shall not be liable for malfunctions and damage resulting from installation,
relocation, remodeling, maintenance, and repair done by other than dealers specified
by FUJIFILM.
4. FUJIFILM shall not be liable for malfunctions and damage of FUJIFILM products due
to the use of products of other manufacturers not supplied by FUJIFILM.
5. FUJIFILM shall not be liable for malfunctions and damage resulting from remodeling,
maintenance, and repair using repair parts other than those specified by FUJIFILM.
6. FUJIFILM shall not be liable for malfunctions and damage resulting from negligence of
precautions and operating methods contained in this manual.
7. FUJIFILM shall not be liable for malfunctions and damage resulting from use under
environment conditions outside the range of conditions required for proper use of this
product, such as power supply, installation environment, etc. contained in this manual.
8. FUJIFILM shall not be liable for malfunctions and damage resulting from natural
disasters such as fires, earthquakes, floods, lightning, etc.

Regulation (EU) 2017/746 requires that any serious incident that has occurred in relation to the device
should be reported to the manufacturer or European Authorized Representative, and the competent
authority of your country. In case you become aware of any serious incident, please report immediately
to your local or our official dealer and the competent authority of your country.

Trademark
“DRI-CHEM” is a trademark or a registered trademark of FUJIFILM Corporation.
Other holder’s trademarks
“QR Code” is registered trademark of DENSO WAVE INCORPORATED.
* This product adopts the font of ARPHIC TECHNOLOGY CO., LTD.
Copyright © 2020-2022 FUJIFILM Corporation. All rights reserved.

897N203029A FDC NX600 Instruction Manual 2022.01 (i)


Introduction

The FDC NX600 is a diagnostic medical device, and this manual is available with all the series (FUJI
DRI-CHEM NX600 and FUJI DRI-CHEM NX600i; the former includes the PF (Plasma Filter) unit, and
the latter does not. So for the latter, sections that describe the unit can be skipped).

Name PF Unit

FUJI DRI-CHEM NX600 4

FUJI DRI-CHEM NX600i -

NOTE: 4 : The PF Unit is mounted on the device.


- : The PF Unit is not mounted on the device.

(ii) 897N203029A FDC NX600 Instruction Manual 2022.01


Table of Contents
1 Safe Usage and Handling Precautions
1.1 Precautions (marks) Used in This Manual……………………………………………… 1-1
1.2 Operation Precautions……………………………………………………………………… 1-1
1.3 Biohazards and Disposal…………………………………………………………………… 1-2
1.4 Warnings Regarding Explosives…………………………………………………………… 1-2
1.5 Warnings Regarding Electric Shock……………………………………………………… 1-3
1.6 Electromagnetic Compatibility (EMC)…………………………………………………… 1-3
1.7 Caution Regarding Moving Parts………………………………………………………… 1-4
1.8 Installation Locations……………………………………………………………………… 1-4
1.9 Action for Malfunction……………………………………………………………………… 1-5
1.10 Measurement Results……………………………………………………………………… 1-6
1.11 FUJI DRI-CHEM Slides…………………………………………………………………… 1-6
1.12 CRP Measurement………………………………………………………………………… 1-7
1.13 QC Card System…………………………………………………………………………… 1-7
1.14 Handling Samples…………………………………………………………………………… 1-7
1.15 Sampler Cover……………………………………………………………………………… 1-8
1.16 Diluents and Reference Fluid……………………………………………………………… 1-8
1.17 FUJIFILM PF Plasma Filters……………………………………………………………… 1-8
1.18 FUJI DRI-CHEM AUTO TIPS, FUJI Sample Tubes,
and FUJI DRI-CHEM MIXING CUPS……………………………………………………… 1-8
1.19 Light Source Lamp………………………………………………………………………… 1-9
1.20 Recording Paper…………………………………………………………………………… 1-9
1.21 Labels………………………………………………………………………………………… 1-10
1.22 Symbols……………………………………………………………………………………… 1-11
2 Part Names and Consumables
2.1 Part Names………………………………………………………………………………… 2-1
2.2 Touch Panel Screen Names and Functions……………………………………………… 2-5
2.2.1 Measurement Preparation Screen (Top Screen)……………………………… 2-5
2.2.2 Help Screen………………………………………………………………………… 2-6
2.3 Names and Functions of Software Keyboard Screens………………………………… 2-8
2.4 Bundled Items……………………………………………………………………………… 2-10
2.5 Consumables and Optional Items………………………………………………………… 2-11
2.5.1 Consumables……………………………………………………………………… 2-11
2.5.2 Optional Items……………………………………………………………………… 2-12
3 Measurement
3.1 Operation Overview………………………………………………………………………… 3-1
3.2 Measurement Preparation………………………………………………………………… 3-3
3.3 Basic Measurements……………………………………………………………………… 3-7
3.3.1 Lamp Off and Measurement-start Scheduling………………………………… 3-14
3.4 Measurement Results Screen……………………………………………………………… 3-15
3.5 ISE Tests……………………………………………………………………………………… 3-18
3.6 Dilution Tests………………………………………………………………………………… 3-21
3.7 CRP Tests…………………………………………………………………………………… 3-25
3.8 Shutting-down the FDC NX600…………………………………………………………… 3-27

897N203029A FDC NX600 Instruction Manual 2022.01 i


Table of Contents

3.9 Sample Tubes for the FDC NX600………………………………………………………… 3-30


3.9.1 Blood Collection Tubes…………………………………………………………… 3-30
3.9.2 FUJI Sample Tubes……………………………………………………………… 3-32
3.9.3 Assembling the Sample Rack…………………………………………………… 3-33
3.10 QC Card System…………………………………………………………………………… 3-37
4 Applicable Usage Methods
4.1 CRP Calibration……………………………………………………………………………… 4-1
4.1.1 Calibration Measurement………………………………………………………… 4-1
4.1.2 Configuring Calibrator Concentration…………………………………………… 4-10
4.2 Manual Spotting (Colorimetric Slide)……………………………………………………… 4-13
4.2.1 Measuring by Manual Spotting (Colorimetric Slide)…………………………… 4-13
4.2.2 Using a Pipette…………………………………………………………………… 4-17
4.3 Tests Using a Plasma Filter (PF)………………………………………………………… 4-18
4.3.1 Measurement……………………………………………………………………… 4-20
4.3.2 Blood Collection Tubes for Tests Using a PF…………………………………… 4-23
4.4 Retesting……………………………………………………………………………………… 4-24
4.5 Slide Loading Order When Measuring Different Items Simultaneously……………… 4-27
4.6 Automatic Acquisition of Sample Information (Work Lists)……………………………… 4-28
4.7 Editing Sample Numbers and Sample IDs……………………………………………… 4-30
4.8 Using the Sample Barcode Reader (Sold Separately)………………………………… 4-31
5 Loading Consumables
5.1 Loading Consumables……………………………………………………………………… 5-1
5.2 Setting Tips…………………………………………………………………………………… 5-2
5.3 Loading Mixing Cups……………………………………………………………………… 5-5
5.4 Loading Diluent……………………………………………………………………………… 5-7
5.5 Loading Reference Fluid…………………………………………………………………… 5-9
6 Maintenance
6.1 Periodic Maintenance Items……………………………………………………………… 6-2
6.1.1 Items for Daily and Periodic Inspection by the User…………………………… 6-2
6.1.2 Items for Maintenance and Inspection by Repair Personnel………………… 6-3
6.2 Cleaning the Air Filters……………………………………………………………………… 6-4
6.3 Inspecting and Cleaning the Transfer Bar, Incubator, Sample Application Unit,
and ISE Unit………………………………………………………………………………… 6-6
6.3.1 Reference Plate Level Check…………………………………………………… 6-7
6.3.2 Cleaning the Transfer Bar, Incubator, Sample Application Unit,
and ISE Unit………………………………………………………………………… 6-9
6.3.3 Reassembling Parts and Ending Cleaning……………………………………… 6-17
6.4 Replacing Recording Paper………………………………………………………………… 6-18
6.5 Replacing and Cleaning the Light Source Lamp………………………………………… 6-19
6.6 Maintenance of Sampler O-ring…………………………………………………………… 6-22
6.6.1 Cleaning the Sampler O-ring……………………………………………………… 6-22
6.6.2 Inspecting the Sampler O-ring…………………………………………………… 6-24
6.6.3 Exchanging the Sampler O-ring………………………………………………… 6-27
6.7 Cleaning the Slide Reading Section……………………………………………………… 6-30
6.8 Cleaning and Replacing the Reference Fluid Cap Packing…………………………… 6-31
6.9 PF Unit Maintenance……………………………………………………………………… 6-33
6.9.1 Cleaning the PF Unit (Sensor, Reference Plate, and Suction Pad)………… 6-33
6.9.2 Inspecting the PF Suction Pad…………………………………………………… 6-36
6.9.3 Replacing the PF Suction Pad…………………………………………………… 6-38
6.9.4 Replacing the PF Pump Tube…………………………………………………… 6-41

ii 897N203029A FDC NX600 Instruction Manual 2022.01


Table of Contents

7 Quality Control
7.1 Control Fluids………………………………………………………………………………… 7-1
7.2 Measurement with Control Fluids………………………………………………………… 7-1
7.3 Monitoring Results………………………………………………………………………… 7-1
7.4 Control Mode………………………………………………………………………………… 7-1
7.5 Troubleshooting Quality Control Results………………………………………………… 7-1
7.6 Control Chart………………………………………………………………………………… 7-2
8 Troubleshooting
8.1 Error Display………………………………………………………………………………… 8-1
8.1.1 Error Log…………………………………………………………………………… 8-2
8.1.2 Error Code and Reference Section Table……………………………………… 8-4
8.1.3 Printout Reference Table………………………………………………………… 8-8
8.2 Troubleshooting……………………………………………………………………………… 8-10
8.2.1 Startup Errors……………………………………………………………………… 8-10
8.2.2 Printer or LCD Trouble…………………………………………………………… 8-10
8.2.3 Slide Reading Errors……………………………………………………………… 8-11
8.2.4 Sampler Errors……………………………………………………………………… 8-11
8.2.5 Photometric System Errors……………………………………………………… 8-19
8.2.6 Transfer Errors……………………………………………………………………… 8-21
8.2.7 Temperature Controller Errors…………………………………………………… 8-22
8.2.8 Circuit Board Signal Errors……………………………………………………… 8-24
8.2.9 QC Card, PF Card, and DI Card Errors………………………………………… 8-29
8.2.10 Slide Loading Errors……………………………………………………………… 8-31
8.2.11 ISE Measurement Errors………………………………………………………… 8-33
8.2.12 PF Errors…………………………………………………………………………… 8-36
8.2.13 Errors Related to Data Communication, Sample Barcode Reader…………… 8-37
8.2.14 Calibration Errors………………………………………………………………… 8-39
8.2.15 Other Errors………………………………………………………………………… 8-42
8.3 Slide Code Table…………………………………………………………………………… 8-44
9 Function Menu
9.1 Control………………………………………………………………………………………… 9-3
9.2 Normal Mode………………………………………………………………………………… 9-7
9.2.1 Date & Time Settings……………………………………………………………… 9-8
9.2.2 Brightness • Volume……………………………………………………………… 9-8
9.2.3 Lamp Management………………………………………………………………… 9-9
9.2.4 Lot Information……………………………………………………………………… 9-9
9.3 Mode Function for Administrator…………………………………………………………… 9-10
9.3.1 Language Setting………………………………………………………………… 9-11
9.3.2 Host Connection Setting………………………………………………………… 9-11
9.3.3 Calculated Parameter……………………………………………………………… 9-14
9.3.4 Unit Conversion…………………………………………………………………… 9-16
9.3.5 Switch Measurement Range Display…………………………………………… 9-17
9.3.6 Reference Interval Range Setting……………………………………………… 9-18
9.3.7 Sample No. /ID Settings…………………………………………………………… 9-20
9.3.8 Work List Selection Setting……………………………………………………… 9-20
9.3.9 Correlation Coefficients…………………………………………………………… 9-21
9.3.10 Lot Compensation Coefficients…………………………………………………… 9-25
9.3.11 Standard Curve Coefficients……………………………………………………… 9-27
9.3.12 Lot Qualification…………………………………………………………………… 9-30
9.3.13 PF Calibration Coefficients……………………………………………………… 9-31
9.3.14 Dilution Setting……………………………………………………………………… 9-33
9.3.15 Tube Setting for Diluent & Reference Fluid……………………………………… 9-35
9.3.16 Print Setting for Measured Result………………………………………………… 9-35

897N203029A FDC NX600 Instruction Manual 2022.01 iii


Table of Contents

9.3.17 Control Fluid Setting……………………………………………………………… 9-36


9.3.18 Control Measurement Setting…………………………………………………… 9-37
9.3.19 Operator ID Setting………………………………………………………………… 9-37
9.3.20 Test Item Name Setting…………………………………………………………… 9-38
9.3.21 Default Sample Type Setting……………………………………………………… 9-39
9.3.22 Password Setting………………………………………………………………… 9-39
10 Other Functions
10.1 Data Communication……………………………………………………………………… 10-1
11 USM (User Safety Mode)
11.1 Operator ID Input When Measurement Start and In Mode Function………………… 11-1
11.2 Difference of Operable Functions among Operator ID Level (OPE,ADM,EMG)……… 11-2
11.3 USM Menu…………………………………………………………………………………… 11-5
11.3.1 Operator ID -Import………………………………………………………………… 11-5
11.3.2 Operator ID -Export………………………………………………………………… 11-7
11.3.3 Editing Registered Operator ID…………………………………………………… 11-9
11.3.4 Editing Emergency ID…………………………………………………………… 11-11
12 Specifications

13 Glossary

iv 897N203029A FDC NX600 Instruction Manual 2022.01


1 Safe Usage and Handling Precautions
This section contains the safety precautions that must be followed for safe operation of the FUJI
DRI-CHEM NX600 (For the remainder of the manual, FUJI DRI-CHEM NX600 is shortened to “FDC
NX600”). Before using this analyzer, please read this chapter carefully and follow the precautions given,
so that you can operate it correctly.

1.1 Precautions (marks) Used in This Manual 1


The following section lists the meaning of precautions (marks) used in this manual.

“Warning” indicates hazardous situations that may lead to serious injury, death,
WARNING: or the transmission of infectious agents if the precaution is not followed.

“Caution” indicates hazardous situations that may lead to minor injury,


CAUTION: moderate injury, or physical damage if the precaution is not followed.

“Important” indicates improper handling that could have an adverse effect on


IMPORTANT: the accuracy of the measurement values if the precaution is not followed.

NOTE “Note” indicates procedures requiring special attention, instructions that


must be followed, supplementary explanations, etc.

1.2 Operation Precautions

CAUTION:
Before using this equipment, please read this Instruction Manual carefully so that you can operate
the equipment correctly.

CAUTION:
Whenever you operate this equipment, be sure to observe the precautions described in this manual.
Failure to do so may subject you to injuries, cause property damage, or produce incorrect measure-
ment results.

CAUTION:
The operators must be sufficiently trained for the usage of sampler cover lock key and observe its
usage method and precautions. Store the lock key carefully so that it is not lost.
(Refer to Section 1.15)

CAUTION:
Do not modify the FDC NX600. Otherwise, the safety will not be guaranteed.

CAUTION:
The effective period of service (service life) of this analyzer is 6 years after delivery (as long as us-
age precautions are observed and regular maintenance is performed). If you will use the analyzer
after the service life expires, consult with the manufacturer/seller beforehand. The use of medical
equipment that has exceeded its service life may increase the chance of safety and performance
problems.

IMPORTANT:
Use consumables on or before their listed expiration date.

897N203029A FDC NX600 Instruction Manual 2022.01 1-1


1 Safe Usage and Handling Precautions

1.3 Biohazards and Disposal

WARNING:
Items such as measured slides, tips, mixing cups, and sample tubes coated with samples such as
1 blood and urine, and cotton swabs and cloths used to clean the equipment are infectious industrial
waste. Process this waste in compliance with any applicable regulations in your country, such as
by incineration, melting, sterilization, or disinfection. Furthermore, when entrusting disposal to an-
other party, request processing from a contractor with a license for special management of industrial
waste processing work.

WARNING:
Items that may be contaminated with blood or urine are infectious industrial waste. When discarding
the main unit, follow applicable regulations in your country and dispose of it appropriately.

WARNING:
Process waste correctly, as stipulated by local law or any regulations that apply.
When discarding the FDC NX600 that incorporates the lithium battery, be sure to contact a licensed
waste disposal contractor because it cannot be disposed of as a general waste.
Not doing so may cause environmental pollution.

WARNING:
Before discarding the main unit, you must delete the measurement results (personal information)
stored on the main unit. Contact your dealer before discarding.

WARNING:
When handling samples (blood or urine) and performing maintenance (cleaning the analyzer), al-
ways follow biohazard procedures (wear gloves, a lab coat, and safety goggles) in accordance with
the sample handling rules of your facility. If you accidentally touch a sample, immediately rinse the
contaminated area thoroughly under running water, and then disinfect the area. Seek medical assis-
tance if necessary.

WARNING:
Do not touch used slides and tips with bare hands as this may cause contamination. If you acciden-
tally touch a contaminated consumable, immediately rinse the contaminated area thoroughly under
running water, and then disinfect the area. Seek medical assistance if necessary.

WARNING:
When samples come in contact with the analyzer, immediately clean and disinfect it.

1.4 Warnings Regarding Explosives

WARNING:
As this equipment is not explosion-proof, be sure not to use flammable and explosive gas around
the equipment.

1-2 897N203029A FDC NX600 Instruction Manual 2022.01


1 Safe Usage and Handling Precautions

1.5 Warnings Regarding Electric Shock

WARNING:
The voltage supplied to the equipment is AC 100 - 240 V. To avoid electrical shock, observe the fol-
lowing precautions. 1
• Avoid installation locations in which water may splash on the equipment.
• Make sure that the equipment is properly grounded to a protective earth lead for indoor wiring.
• Make sure that all cables have been properly connected.

WARNING:
Plug the power cable of the equipment into an outlet with a grounding receptacle. Otherwise, there
is risk of electrical shock.

WARNING:
When plugging the power cable in or removing it from an outlet, be sure to hold onto the plug body,
not just the cable. Otherwise, the power cable may be damaged and cause electric shock or fire.

WARNING:
Do not remove covers or other parts that are secured with screws. Doing so may result in electrical
shock from exposure to hazardous voltage or injury from moving parts.

1.6 Electromagnetic Compatibility (EMC)


This equipment conforms to the following emission and immunity requirements:
- IEC 61326-2-6:2012 (Class B)
- EN 61326-2-6:2013 (Class B)
- IEC 61326-2-6:2005 (Class B)
- EN 61326-2-6:2006 (Class B)
• The electromagnetic environment should be evaluated prior to operation of the equipment.
• Do not use this equipment in close proximity to sources of strong electromagnetic radiation (e.g.
unshielded intentional RF sources), as these may interfere with the proper operation.
However, there is no guarantee that interference will not occur in a particular installation. If this equipment
does cause harmful interference to other devices, which can be determined by turning the equipment off
and on, the user is encouraged to try to correct the interference by one or more of the following measures.
• Reorient or relocate the receiving device.
• Increase the separation between the equipment.
• Connect the equipment into an outlet on a circuit different from that to which the other device(s)
are connected.
Consult the manufacturer or field service technician for help.

CAUTION:
Do not use other devices (such as mobile phone) which generate and can radiate radio frequency en-
ergy near the FDC NX600. Otherwise, physical damage or malfunction on the FDC NX600 may occur.
NOTE It is the manufacturer's responsibility to provide equipment electromagnetic compatibility
information to the customer or user. On the other hand, it is the user’s responsibility to
ensure that a compatible electromagnetic environment for the equipment can be main-
tained in order that the equipment will perform as intended.

897N203029A FDC NX600 Instruction Manual 2022.01 1-3


1 Safe Usage and Handling Precautions

1.7 Caution Regarding Moving Parts

WARNING:
Do not place your fingers near moving parts (sampler and consumable disk) when operating the
1 equipment, including during maintenance.
Also be sure not to get your fingers, hair, clothing, or accessories caught in moving parts.

WARNING:
During sample processing, ensure the sampler cover is closed and locked with the sampler cover
lock key to prevent injuries and biohazards. If your finger touches a tip that has sample inside, the
sample may scatter and cause infection.
When the display reads [Ready to test], the sampler cover can be opened.

1.8 Installation Locations

WARNING:
Plug the power cable of the equipment into an outlet with a grounding receptacle. Otherwise, there
is risk of electrical shock.

CAUTION:
Avoid the following installation sites.
• Places where spills or water leakage may occur.
• Places where the equipment is exposed to direct sunlight.
• Places near sources of heat such as heaters.
• Places where the temperature may drastically change.
• Places where the equipment is subject to vibration or its support table is unstable.
(1) Install the equipment in the following environmental conditions.
Installation environment: Indoor use (do not expose to direct sunlight)
Illumination: 6,000 cd/m2 (lux) or less(3,000 cd/m2 (lux) or less when using
the sample barcode reader)
Atmospheric pressure: 810 hPa or more (equivalent to altitude 2,000 m)
Transient overvoltage category: II
Pollution degree: 2
Ingress protection rating: IP00 (No Protection)
Operating temperature: 15 to 32 °C
Operating humidity: 30 to 80% RH (no vapor condensation)
(2) Use the equipment under the following electrical specifications.
Power supply voltage: single phase AC 100 - 240 V ±10%
Frequency: 50 to 60 Hz
Consumption current: 1.9 - 0.8 A
Type of protection against
electrical shock: CLASS 1 EQUIPMENT
(3) Use an independent outlet separate from other equipment.
(4) Make at least 10 cm of space behind and beside the equipment.
(5) Unplug the equipment from the AC outlet if it will not be used for an extended period of time.

1-4 897N203029A FDC NX600 Instruction Manual 2022.01


1 Safe Usage and Handling Precautions

CAUTION:
Hold the handle of the device when lifting or moving it.

Handle

CAUTION:
• When moving the device to another installation location, remove all slides, samples, diluent, refer-
ence fluid, mixing cups, tips, and other consumables beforehand.
• Do not set the device on its side or upside down while moving it.
• Place the device on a flat and stable stand.
• When the device is subject to vibrations, such as when being moved by vehicle, it is necessary to
secure the moving parts within the device. Contact your dealer.
• The weight of 28kg is supported by 4 portions of rubber feet. The load on each portion is approx.
7kg. Make sure that the load bearing of the table at the installation location is sufficient.
Failure to observe these precautions may result in unexpected errors or equipment malfunctions.

IMPORTANT:
When wax was used to clean the floor of the installation location, make sure that wax will not adhere
to slides and the device. If wax has adhered, the measurement results may be affected.

1.9 Action for Malfunction

CAUTION:
When the equipment generates unusual noise, smell, or smoke, immediately unplug the power ca-
ble, and contact your dealer.

CAUTION:
Do not place any objects near the wall outlet so that the power cable can be easily unplugged.

CAUTION:
In case of emergency, clearance at the back of the equipment is required so that the power cable
can be removed from the inlet of the equipment.

897N203029A FDC NX600 Instruction Manual 2022.01 1-5


1 Safe Usage and Handling Precautions

1.10 Measurement Results

IMPORTANT:
Make a diagnosis in a comprehensive manner, considering other relevant test results or clinical symptoms.
1
IMPORTANT:
When an equipment problem (error) appears before or during measurement, or a warning appears
in the measurement results, the measurement results may not be accurate. Refer to Chapter 8,
Troubleshooting, and then rerun the measurement.

IMPORTANT:
To maintain the accuracy of your test results, it is recommended to perform quality control on routine
bases by use of control fluids. (Refer to Chapter 7.)

1.11 FUJI DRI-CHEM Slides


NOTE For the remainder of the manual, FUJI DRI-CHEM slide is shortened to “slide”.

CAUTION:
Please confirm about the usable slide items in this device to your dealer for assistance.

WARNING:
Do not touch used slides with bare hands as this may cause infection. If you accidentally touch an
item, immediately rinse the contaminated area thoroughly under running water, and then disinfect
the area. Seek medical assistance if necessary.

WARNING:
Used slides are infectious industrial waste. Process this waste in compliance with any applicable
regulations in your country, such as by incineration, melting, sterilization, or disinfection. Further-
more, when entrusting disposal to another party, request processing from a contractor with a license
for special management of industrial waste processing work.

CAUTION:
Please read the Instructions for Use of the slides carefully and follow the instructions.

IMPORTANT:
• The slides should be stored in a refrigerator still wrapped to avoid humidity, light, and heat.
• Take out the required number of slides from the refrigerator, and allow them to reach room tem-
perature for 5 minutes or more with the condition of the individual package before unwrapping and
using them. Otherwise, the measurement results and the analyzer operation may be affected.
• Complete the measurement within 30 minutes after unwrapping the individual package.
• Do not touch the center part (both sides) of a colorimetric slide.
• Do not touch the thread bridge part of an electrolyte slide.
• Do not scratch the identification information on the back of a slide.
• Do not reuse a slide.

NOTE Slide packages and containers for liquids such as diluent and reference fluid may change
without notice.

1-6 897N203029A FDC NX600 Instruction Manual 2022.01


1 Safe Usage and Handling Precautions

1.12 CRP Measurement

WARNING:
As calibrator CP is derived from blood, it should be handled in the same way as a patient sample.
When handling calibrator CP, always follow biohazard procedures (wear gloves, a lab coat, and 1
safety goggles) in accordance with the sample handling rules of your facility. If you accidentally
touch a sample, immediately rinse the contaminated area thoroughly under running water, and then
disinfect the area. Seek medical assistance if necessary.

IMPORTANT:

CRP measurements require periodic calibration. For details, refer to the “Instructions for Use” of the
slides and “Section 3.7 CRP Tests” in this document.

1.13 QC Card System


(1) A QC card is packed together with CM (Colorimetric) slides in the same box. Before you start
using a new lot of CM slides, read the QC card using the card reader.
The analyzer can record up to 5 lots per test, so you can mix and use slides of different lots.
Store the read QC card in the box it came in until the accompanying lot of slides has been used
up.
NOTE When QC information exceeds 5 lots, the oldest QC information is deleted. Slides with-
out QC information cannot be measured.

(2) A QC card contains the slide test name and the correction coefficient corresponding to the man-
ufacturing number (lot). Always use slides from the same box.

1.14 Handling Samples


WARNING:
When handling samples (blood or urine), always follow biohazard procedures (wear gloves, a lab
coat, and safety goggles) in accordance with the sample handling rules of your facility. If you acci-
dentally touch a sample, immediately rinse the contaminated area thoroughly under running water,
and then disinfect the area. Seek medical assistance if necessary.

IMPORTANT:

Sample preparations differ depending on the test. Read the “Instructions for Use” of the slides.

IMPORTANT:

When using plasma or serum, check that the samples have not undergone hemolysis, and that they
do not contain any fibrin deposits.

IMPORTANT:

When using whole blood, test the sample as soon as possible after drawing it.
In the case of whole blood, blood cell precipitation may affect the measurement results.

897N203029A FDC NX600 Instruction Manual 2022.01 1-7


1 Safe Usage and Handling Precautions

1.15 Sampler Cover


WARNING:
During sample processing, ensure the sampler cover is closed and locked with the sampler cover
1 lock key to prevent injuries and biohazards. If your finger touches a tip that has sample inside, the
sample may scatter and cause infection. When the display reads [Ready to test], the sampler cover
can be opened.

1.16 Diluents and Reference Fluid


CAUTION:
To ensure safe handling of diluent and reference fluid, refer to the handling methods in the “Instruc-
tions for Use” of each fluid or follow the chemical handling rules of your facility (wear gloves, a lab
coat, and safety goggles). If you accidentally touch an item, immediately rinse the contaminated
body part thoroughly under running water. Seek medical assistance if necessary.

IMPORTANT:

For diluents used in measurements that require dilution, refer to the “Instructions for Use” of the
slides.

IMPORTANT:

Use the reference fluid included with the ISE slides for ISE measurements. Refer to the “Instructions
for Use” of the slides.

IMPORTANT:

Use only specified products. Otherwise, errors or malfunctions may occur.

1.17 FUJIFILM PF Plasma Filters


NOTE FUJIFILM plasma filter PF is shortened to "PF" in this document.

IMPORTANT:

Before using a FUJIFILM PF, refer to “Instructions for Use” provided with the PF, paying special at-
tention to the precautions concerning the samples measured and the blood collection tubes used.

1.18 FUJI DRI-CHEM AUTO TIPS, FUJI Sample Tubes, and FUJI
DRI-CHEM MIXING CUPS
NOTE In this manual, FUJI DRI-CHEM AUTO TIPS may be shortened to “tip”, FUJI Sample Tube
to “tube”, and FUJI DRI-CHEM MIXING CUPS to “mixing cup”.

FDC NX600 performs sample application automatically via the sampler. However, it is necessary to
prepare dedicated FUJI DRI-CHEM AUTO TIPS, FUJI Sample Tubes, and specified blood collection
tubes and FUJI DRI-CHEM MIXING CUPS.
The FUJI Sample Tubes include the following types.
• FUJI HEPARIN TUBE 1.5 mL
• FUJI HEPARIN TUBE 0.5 mL
• FUJI PLAIN TUBE 1.5 mL
• FUJI PLAIN TUBE 0.5 mL

1-8 897N203029A FDC NX600 Instruction Manual 2022.01


1 Safe Usage and Handling Precautions

WARNING:
Do not touch used tips, tubes, and mixing cups with your bare hands as this may cause infection. If
you accidentally touch an item, immediately rinse the contaminated area thoroughly under running
water, and then disinfect the area. Seek medical assistance if necessary.
1
WARNING:
Used tips, tubes, and mixing cups are infectious industrial waste. Process this waste in compliance
with any applicable regulations in your country, such as by incineration, melting, sterilization, or dis-
infection. Furthermore, when entrusting disposal to another party, request processing from a con-
tractor with a license for special management of industrial waste processing work.

IMPORTANT:

Always use new FUJI DRI-CHEM AUTO TIPS, FUJI Sample Tubes, blood collection tubes, and
FUJI DRI-CHEM MIXING CUPS for each sample. Do not reuse old ones.

IMPORTANT:

Use only specified products. Otherwise, errors or malfunctions may occur.

IMPORTANT:

When the power is shut off during test processing, such as by a power outage, replenish tips and
replace mixing cups. The analyzer may lose the information of the usage positions of the consuma-
bles and use used mixing cups (holes).

1.19 Light Source Lamp

CAUTION:
The light source lamp gets very hot. Before replacing the lamp, turn the power off, wait at least 5
minutes, and then check that the lamp has cooled.
(1) The light source lamp is a precision halogen lamp. Do not touch the glass surface of the lamp
with your bare hands.
(2) The lamp is a consumable with a lifetime. Always keep a spare lamp on hand in case the lamp
burns out.

1.20 Recording Paper


(1) Use specified recording paper for the FDC NX600.
(2) If other recording paper is used, the printer head section may be damaged.

897N203029A FDC NX600 Instruction Manual 2022.01 1-9


1 Safe Usage and Handling Precautions

1.21 Labels
The following safety labels are affixed to the FDC NX600.

1 Power button

Biohazard label Sampler cover lock


key caution label

 Refer to Section 1.15

Biohazard Moving parts


label caution label

High tempera- Lamp caution PF tube caution


ture caution label label Biohazard label
label
 Refer to Section 6.5  Refer to Section 6.5  Refer to Section 6.9.4

Name-plate

NOTE: The example


above uses an
USB RS-232C LAN
illustration for the Power switch
connector connector terminal
FDC NX600. For | : Depress to power ON
the FDC NX600i,
O : Depress to power OFF
the description of
the plate partially
differs.

1-10 897N203029A FDC NX600 Instruction Manual 2022.01


1 Safe Usage and Handling Precautions

1.22 Symbols

Sign Description

Warning, Caution, Important, consult documents 1

Biological risks

High temperature caution

Power button

Power switch

LAN terminal

RS-232C connector

USB connector

Fragile item, handle carefully

Protect from rain

This way up

2 Stacking limitation

Temperature limitation

Manufacturer

SN Serial number

897N203029A FDC NX600 Instruction Manual 2022.01 1-11


1 Safe Usage and Handling Precautions

Sign Description

Date of manufacture

1
EC REP Authorized representative in the European Community

In vitro diagnostic medical devices

CE mark of conformity

Importer

Consult an electronic Instructions for Use (e-IFU)

Unique device identifier

Near-patient testing

1-12 897N203029A FDC NX600 Instruction Manual 2022.01


2 Part Names and Consumables
2.1 Part Names
NOTE The following is an illustration of the FDC NX600 (full option model). FDC NX600i has slight
difference.

Printer

LCD (touch panel)


Sampler cover
2

Power button

Sampler cover lock

Front bottom cover


Card reader

Disposal box

Sampler

Incubator cable

Plasma filter unit (PF unit)

Incubator cover

Consumable disk

897N203029A FDC NX600 Instruction Manual 2022.01 2-1


2 Part Names and Consumables

Air filter
(Louver)
USB connector
2 (For sample barcode reader only)
*‌Do not connect other USB
devices.

LAN terminal
CAUTION:
RS-232C connector
Do not remove the cover to
look into the QR code read- USB connector
er’s illuminator. It could impair Power inlet
your eyes. Power switch

Air filter
(Louver)

2-2 897N203029A FDC NX600 Instruction Manual 2022.01


2 Part Names and Consumables

Disposal box

Light source lamp

Slide cartridge Reference fluid holder


Diluent holder

Sample rack

RE

Mixing cup

Tip rack

Sample application
unit cover

897N203029A FDC NX600 Instruction Manual 2022.01 2-3


2 Part Names and Consumables

< Details of PF unit >

PF unit

2 PF sensor / suction
pad

PF reference plate

PF sample rack
RE

PF pump

2-4 897N203029A FDC NX600 Instruction Manual 2022.01


2 Part Names and Consumables

2.2 Touch Panel Screen Names and Functions


2.2.1 Measurement Preparation Screen (Top Screen)
If the power is switched on, the Measurement-preparation screen (top screen) is displayed after startup.

1 Help button
1 Display the help screen.

2 è Refer to Section 2.2.2 2


3 2 Function menu button
Display the function menu screen.
4
è Refer to Chapter 9
3 Work list button
Obtain order information from the host PC
5 via a DMS connection.
è Refer to Section 4.6
4 Clear button
6 Clear the entered information.
5 Plasma filter (PF) production num-
7 ber (lot) display
Displays the lot number of the plasma filter
8 (PF) in use. Make sure this number matches
the production number (lot) of the plasma
9 filter (PF) in use.
0 6 Mixing cups
Displays the quantity of mixing cups. Tap to
a replace the mixing cups.
è Refer to Section 5.3
CAUTION:
7 Tips
The touch panel does not have multi-touch input Tap to replenish tips.
function. è Refer to Section 5.2
Do not touch plural points on the panel at the 8 [START] button
same time.
Start a measurement.
9 O perator ID button (displayed ac-
cording to settings)
If tapped, you can enter the operator ID.
è Refer to Section 9.3.19
0 Feed button
Feed the recording paper a fixed length.
[STOP] button (displayed during
sample application operation)
If the [STOP] button is tapped during sample
application, tips are discarded, operation is
stopped, and the device enters a paused state.
a Measurement results button
Display the measurement results.
è Refer to Section 3.4

897N203029A FDC NX600 Instruction Manual 2022.01 2-5


2 Part Names and Consumables

2.2.2 Help Screen


You can tap the button on the Measurement
preparation screen to check the procedures
about measurement, maintenance, consum-
able exchange, and other actions on the FDC
NX600.

2-6 897N203029A FDC NX600 Instruction Manual 2022.01


2 Part Names and Consumables

1 1 button
The Measurement-preparation screen ap-
pears.
2 button
Return to the previous screen.

897N203029A FDC NX600 Instruction Manual 2022.01 2-7


2 Part Names and Consumables

2.3 Names and Functions of Software Keyboard Screens


When data entry for operations and configura-
tions is necessary, the software keyboard ap-
pears on the LCD automatically. Enter alphanu-
meric characters and symbols.

2 LCD
(LCD touch panel)

1 1 Alphabetical characters, numbers,


and symbols buttons
2 Enter alphabetical characters, numbers,
and symbols.

3 2 button
Delete all entered characters, and then
quit keyboard entry.
3 Move cursor buttons
Move the position of the cursor left or
right by one character.
4 [Clear] button
Delete all entered characters.
7 4 5 [OK] button
6 Confirm the entered characters.
5
6 button
Delete the character to the left of the
cursor.
7 [Space] button
Enter a space of one single character
width.

2-8 897N203029A FDC NX600 Instruction Manual 2022.01


2 Part Names and Consumables

Change the entry screen by using the [ABC], [abc], [123], or [-#/], button.

< When the [ABC] button is selected > < When the [abc] button is selected >

2
< When the [123] button is selected >

< When the [-#/] button is selected >

8 9 0 a

8 [ABC] button 0 [123] button


Display the entry screen for uppercase alpha- Display the entry screen for numbers, com-
betical characters and numbers. ma (,), and minus symbol (-).
9 [abc] button a [- # /] button
Display the entry screen for lowercase alpha- Display the entry screen for symbols and
betical characters and numbers. numbers.

897N203029A FDC NX600 Instruction Manual 2022.01 2-9


2 Part Names and Consumables

2.4 Bundled Items


Recording paper 2 rolls

Slide cartridge 2

Slide weight 2

Light source lamp 1

2 O-ring 2

Tip rack 1

Sample rack assembly kit 1 set

Sampler leak check tool 2

PF leak check tool (For FDC NX600) 1

FUJI DRI-CHEM AUTO TIPS 1 case

FUJI PLAIN TUBE 0.5 (0.5 mL type) 1 pack

FUJI DRI-CHEM MIXING CUPS 1 pack

Manual pipette cover 1

European authorized representative label 1

Key (for sampler cover) 2

Instruction manual 1

Installation method 1

NOTE An AC power cable is not supplied as a bundled item.


Please prepare an AC power cable conforming to the specifications below:

<The requirements of the cable specifications for the FDC NX600>

Power voltage: 100V-120V


Requirements for the plug/connector: AC125V 10A
Requirements for the cable: SVT 3/18AWG 60°C

Power voltage: 200V-240V


Requirements for the plug/connector: AC250V 10A
Requirements for the cable: GTCE-3 1.0mm2 70°C

NOTE The AC power cable is required to be compliance with any applicable regulations in your country.

NOTE Do not change AC power cable to inappropriate one.

NOTE Specifications and capabilities are subject to change without notice.

2-10 897N203029A FDC NX600 Instruction Manual 2022.01


2 Part Names and Consumables

2.5 Consumables and Optional Items


To purchase the consumables or optional items listed below, please contact the dealer from whom
you purchased the FDC NX600.

2.5.1 Consumables
Name Package/unit
* FUJI DRI-CHEM AUTO TIPS 6 cases
FUJI Sample Tubes
FUJI HEPARIN TUBE (1.5 mL) 500×1 box
2
FUJI HEPARIN TUBE (0.5 mL) 500×1 box
FUJI PLAIN TUBE (1.5 mL) 500×1 box
* FUJI PLAIN TUBE (0.5 mL) 500×1 box
* Recording paper 6 rolls×1 box
* Light source lamp 1
* O-ring 4
Plasma Filter PF 50×1 bag
PF Suction Pad 1
PF Tube 1
* FUJI DRI-CHEM MIXING CUPS 50×1 box
Reference fluid cap packing 5×1 bag

NOTE Parts names marked with “*” are the same parts packed with the FDC NX600.

Chemicals
FUJI DRI-CHEM CONTROL QP-L
FUJI DRI-CHEM CONTROL QP-H
FUJI DRI-CHEM CONTROL QN
FUJI DRI-CHEM REFERENCE FLUID RE
FUJI DRI-CHEM DILUENT DL (CRP)
FUJI DRI-CHEM CALIBRATOR CP (CRP)

NOTE Types of packages or containers for consumables are subject to change without notice.

897N203029A FDC NX600 Instruction Manual 2022.01 2-11


2 Part Names and Consumables

2.5.2 Optional Items


Name Package/unit
* Sample rack assembly kit 1 set
* Slide cartridge 1
* Slide weight 1
Sample barcode reader 1

2 CAUTION:

The sample barcode reader specified for the FDC NX600 can be used.
Do not connect a barcode reader other than specified for the FDC NX600. Otherwise, physical
damage or danger of fire may result.

NOTE Parts names marked with “*” are the same parts packed with the FDC NX600.

NOTE Specifications and capabilities are subject to change without notice.

2-12 897N203029A FDC NX600 Instruction Manual 2022.01


3 Measurement
3.1 Operation Overview
On the FDC NX600, first load the consumables and other items to prepare for measurement (Step 1),
and then perform the actual measurement (Step 2).

Step 1 - Measurement preparation

Cleaning the disposal box Refer to Section 3.2(1)

3
Checking the recording paper Refer to Section 3.2(2)

Starting the FDC NX600 Refer to Section 3.2(3)

Checking the date and time Refer to Section 3.2(4)

Loading consumables Refer to Chapter 5

For dilution measurements For ISE For dilution


For CRP measurement measurements measurements

Diluent Reference fluid Mixing cups


Diluent for CRP (Refer to (Refer to
(Refer to Section 5.5) Section 5.3)
Section 5.4)

12 Tips
RE
(Refer to Section 5.2)

3-1 897N203029A FDC NX600 Instruction Manual 2022.01


3 Measurement

Step 2 - Measurement
Basic Refer to ISE test measurements  Refer to Section 3.5
Dilution measurements  Refer to Section 3.6
Section 3.3
measurements CRP measurements  Refer to Section 3.7

2 Sample
1 Slide

RE

Check Plasma/Serum Check the sample type


the tips

3
START button

Slide set order Refer to Section 4.5


To test CRP and To test CM To test CRP, CM, To test CM slides
CM (Colorimetric) slide and ISE slide and ISE slide that require dilution

ISE Slide
x5 Dil
CRP ISE Slide CRP
CM Slide CM Slide CM Slide x2 Dil

Checking the measurement results Refer to Section 3.3 (6)

Ending measurement Refer to Section 3.3 (7)

897N203029A FDC NX600 Instruction Manual 2022.01 3-2


3 Measurement

3.2 Measurement Preparation


(1) Cleaning the disposal box

Empty the disposal box.

IMPORTANT:

If the disposal box is full, a slide transfer error or tip


disposal error may occur, which may negatively af-
fect the measurement values.

3 IMPORTANT:

After emptying the disposal box, be sure to load the


disposal box. If the analyzer is used without the box,
the measurement results may be negatively affected.

WARNING:
Do not touch used slides and tips with your bare
hands, as this may cause infection. If you acciden-
tally touch an item, immediately rinse the contami-
nated area thoroughly under running water, and
then disinfect the area. Seek medical assistance if
necessary.

WARNING:
Used slides and tips are infectious industrial
waste. Process this waste in compliance with any
applicable regulations in your country, such as by
incineration, melting, sterilization, or disinfection.

3-3 897N203029A FDC NX600 Instruction Manual 2022.01


3 Measurement

(2) Checking the recording paper

Press the area shown in the illustration to release


Cover the lock (1), and then lift up the cover (2).

Press the
concave
section.
3

Check the remaining recording paper (FUJI thermal


paper). If the recording paper is short for the meas-
Recording urement, replace the old roll with a new one.
paper
è Refer to Section 6.4

NOTE Before closing the cover, pull out the edge


of the recording paper.
Recording NOTE Close the cover securely until it clicks.
paper Otherwise, the paper feed may fail and the
test results will not be printed out.

Cutter
opening

897N203029A FDC NX600 Instruction Manual 2022.01 3-4


3 Measurement

(3) Starting the FDC NX600

Press the power button to start the FDC NX600.

NOTE If the equipment is connected to DMS,


switch on the equipment before starting
DMS. Failure to observe this precaution
could interfere with data transfer.

IMPORTANT:

If the power has been shut off during measurement

3 due to a power outage or other issue, replenish


the tips (refer to Section 5.2) and replace the mix-
Power button ing cups (refer to Section 5.3). The analyzer may
lose the information of the usage positions of the
consumables and use wells (holes) of used mixing
cups.

NOTE When the power switch is not in the ON po-


sition, push the switch to the " | " side.

Power switch

0013
NOTE The listing to the left is displayed during
start-up once a month. Perform the periodic
maintenance listed in Chapter 6.

3-5 897N203029A FDC NX600 Instruction Manual 2022.01


3 Measurement

(4) Checking the date and time

Check that the date and time displayed on the LCD


are correct.
If they are incorrect, set the date and time using
[Date & Time settings] in Normal mode.
è (Refer to Section 9.2.1)

IMPORTANT:

If the date and time are not adjusted correctly, the


analyzer may fail to determine the expiration of the
slides and calibration, and measurement results 3
might be incorrect.

(5) Loading consumables


Load the consumables (tips, mixing cups, diluent, and reference fluid) in accordance with the
type of measurement.
è Refer to Chapter 5

(6) Reading a QC card

The FDC NX600 can compensate for the differ-


ences of the slide production lots by reading the QC
card in the box of slides.
When [Warming up] or [Ready to test] appears on
the LCD, set the QR code on the left side, and then
insert the QC card into the reader fully.
è (Refer to Section 3.10)
QR code
NOTE If the QC card is dirty or folded, the read er-
ror may occur.

897N203029A FDC NX600 Instruction Manual 2022.01 3-6


3 Measurement

3.3 Basic Measurements


NOTE By default, this analyzer switches off the light source lamp to save lamp life when the ana-
lyzer has not been used for the time configured under [Lamp management] in Normal mode.
When the lamp is off, refer to Section 3.3.1 before operating the equipment.

WARNING:

When handling samples (blood or urine), always follow biohazard procedures (wear gloves, a lab
coat, and safety goggles) in accordance with the sample handling rules of your facility. If you acci-
dentally touch a sample, immediately rinse the contaminated area thoroughly under running water,
and then disinfect the area. Seek medical assistance if necessary.
3
(1) Checking the LCD

When [Ready to test] appears on the LCD, make


sure that the analyzer is not applying samples.

NOTE You can stop sample application by press-


ing the [STOP] button.

NOTE During sample application, a new sample


for the next measurement cannot be loaded
on the analyzer.

(2) Setting tips

Open the sampler cover, and then check the tip


rack has been loaded with tips. When there are not
enough tips, tap the tip display on the LCD to dis-
RE play the replacement screen, and then load the tips.
For details of the loading method, refer to Section
5.2.

IMPORTANT:

If the power has been shut off during measure-


ment due to a power outage or other issue, replen-
ish the tips and replace the mixing cups (refer to
Section 5.3). The analyzer may lose the informa-
tion of the usage positions of the consumables
and use wells (holes) of used mixing cups.

Tips

3-7 897N203029A FDC NX600 Instruction Manual 2022.01


3 Measurement

(3) Loading samples and slides

WARNING:

When handling samples (blood or urine), always follow biohazard procedures (wear gloves, a lab
coat, and safety goggles) in accordance with the sample handling rules of your facility. If you acci-
dentally touch a sample, immediately rinse the contaminated area thoroughly under running water,
and then disinfect the area. Seek medical assistance if necessary.

1. Open the packaging of the slide to be


used.

3
IMPORTANT:

Take out the required number of slides from the re-


frigerator, and allow them to reach room tempera-
ture for 5 minutes or more with the condition of the
individual package before unwrapping and using
them. Otherwise, the measurement results and the
analyzer operation may be affected.

IMPORTANT:

After the package has been unwrapped, finish


measurement within 30 minutes.

Slide weight 2. Place slides to be measured in the slide


cartridge with the test name facing up.
Be sure to put the slide weight on top of
the stacked slides.
NOTE The item to be measured is determined
Slide from the slide information on the back of
the slide.

NOTE ISE slides can also be loaded at the


same time as the slide cartridge. Includ-
ing ISE slides, a maximum of 20 slides
can be loaded.

3. Load the slide cartridge while aligning


Slide cartridge its orientation.

RE

897N203029A FDC NX600 Instruction Manual 2022.01 3-8


3 Measurement

4. Load the dedicated sample rack for the


blood collection tube or FUJI sample
tube to be used.
NOTE To avoid a suction error or malfunction,
use a sample tube that matches the la-
bel on the sample rack.

NOTE Refer to Section 3.9 for sample tubes


that can be used.

3 5. Remove the cap of the blood collection


tube or FUJI Sample Tube that contains
sample, and then load it on the dedicat-
ed sample rack.
NOTE Make sure there are no bubbles in the
sample in the blood collection tube or
FUJI Sample Tube. Use a centrifuge to
remove bubbles. If bubbles are present,
the sample may be detected incorrectly.

Blood collection tube


or FUJI Sample Tube

3-9 897N203029A FDC NX600 Instruction Manual 2022.01


3 Measurement

(4) Registering sample information

Enter or select the sample information (sample num-


ber, sample ID, reference standard, and sample type).
1
1 Enter the sample number using the software key-
2
board (Section 2.3).
3
2 Obtain order information from the host PC via a
4 DMS connection. è Refer to Section 4.6
5
3 Clear the entered information.
6
4 Enter the sample ID using the software keyboard
7 (Section 2.3).
5 Select a reference standard from the list.
6 Select a sample type from the list.
3
7 Select a dilution factor from the list.
è Refer to Section 3.6 and Section 9.3.14

NOTE When the analyzer is connected to DMS, sample information can be acquired automati-
cally from the work list.
è Refer to Section 4.6

IMPORTANT:

The sample type of the sample, slide, and the sample type setting of the FDC NX600 must be same.
Otherwise, an incorrect measurement result will be obtained.

IMPORTANT:

Be sure to enter the correct sample number and sample ID.

IMPORTANT:

When measuring a low concentration protein fluid (such as urine), some bubbles may appear dur-
ing measurement in the tip. If bubbles have formed during measurement, perform the measure-
ment again from the beginning.

897N203029A FDC NX600 Instruction Manual 2022.01 3-10


3 Measurement

(5) Starting measurement

Close the sample cover, and then lock it with the


sampler cover lock key.

WARNING:

Ensure the sampler cover is locked during sample


processing to prevent injuries and infection.
When the display reads [Ready to test], the sam-
pler cover can be opened.

Tap the [START] button to start measurement.

IMPORTANT:

Close the disposal box. If it is open, the measure-


ment results will be affected.

3-11 897N203029A FDC NX600 Instruction Manual 2022.01


3 Measurement

When the spotting operation is completed, the


screen on the left is displayed and the time until the
measurement is completed is displayed.

Time until the end


of measurement
Measurement
start time

897N203029A FDC NX600 Instruction Manual 2022.01 3-12


3 Measurement

(6) Checking the measurement results

After measurement, tap the button to check the


test results. The test results are printed out in the
order that the slides were loaded in the cartridge.

To retest (Section 4.4), tap the (retest) button.

3 NOTE For the details of the measurement results


screen, refer to Section 3.4.

(7) Ending measurement


To complete the measurement, unlock the sampler cover and open it, and then remove the
measured sample.

NOTE When a measurement is not being performed, empty the disposal box as appropriate.
The disposal box can hold 100 slides and 50 AUTO TIPS. Always empty the disposal
box before these numbers are exceeded.

3-13 897N203029A FDC NX600 Instruction Manual 2022.01


3 Measurement

3.3.1 Lamp Off and Measurement-start Scheduling


By default, when this analyzer has not been used for some time (a time from 20 through 240 min-
utes can be set) while [Ready to test] is displayed on the LCD, it switches off the light source lamp
to save lamp life. After the lamp is switched on again, the start of measurement can be scheduled
even when [Lamp warming up] is displayed on the LCD.

NOTE Do not use measurement-start scheduling for measurements using PF. (Refer to Section 4.3)
After inversion, whole blood samples start to get separated. PF filtering may not occur normal-
ly, and an error may occur.

NOTE The start of measurement cannot be scheduled in manual sample application mode.

The touch panel screen becomes darker when the 3


lamp is off. If the screen is tapped, the lamp switch-
es on, and [Lamp warming up] is displayed on the
LCD. In the case of the Measurement-preparation
screen, the start of measurement is scheduled if the
[START (Reserve)] button is tapped.

If measurement is possible, it starts automatically.

NOTE To cancel a scheduled measurement-start,


tap the [STOP] button.

897N203029A FDC NX600 Instruction Manual 2022.01 3-14


3 Measurement

3.4 Measurement Results Screen


After measurement has ended, tap button, and
then check the measurement results.

NOTE Results of the last 270 samples are backed


up in data.

Result data display area < Display of measurement results >


3 1
3 Green: Within reference interval
Red: Outside reference interval
Or, when the measurement result
is not displayed due to an error

Lower limit of reference interval


Upper limit of reference interval
1
2
Green: ‌All measurement results are
within reference interval.

Red: ‌The results include values


outside the reference interval.
Or, when an error has oc-
curred.
2
3
Measurement results that include values out-
side the reference interval are indicated by
, and the corresponding test name and value
become red. If tapped, the details screen ap-
pears.
And, will be indicated when an error has oc-
curred.
* Also, in case of the following, will be indi-
cated.
• When a colorimetric/ISE test result is outside
of the measurement range (determination
range) or the qualitative range
• When a result of calculated parameter is dis-
played by [****]

NOTE If measurement items that display H, L, #,


or other characters are tapped, the details
screen appears.

3-15 897N203029A FDC NX600 Instruction Manual 2022.01


3 Measurement

< Buttons of the measurement results screen >

1 1 button
Close the measurement results screen, and
2 then return to the Measurement-preparation
screen.
2 Change sample No. button
Change the sample number for which meas-
urement results are to be displayed.
3 Feed button
3 Feed the recording paper with a fixed
4 amount.
3
5 4 Print button
Print all measurement results of the dis-
6 played sample.
5 Send button
Send the measurement results of the dis-
7 played sample to a PC (via DMS).
6 Retest button
8 Perform the measurement again under the
same conditions for sample number, sample
ID, reference standard, and sample type.
Only the dilution factor can be changed.

9 a 0 9
è Refer to Section 4.4
7 Result data scroll button
NOTE When the past test results are dis- Change the display of the measurement re-
played, the results are displayed just as sult data of the slide.
they were obtained using the settings 8 Sample No. tab
when they were measured, such as Unit
Tap to display the measurement results of
conversion, Correlation coefficients,
the selected sample number.
Reference interval range, Test item
name. The test results sent by the [Send] 9 Sample No. tab scroll button
button have also the same format. Change the display of the Sample No. tab.
This button is available when five or more
samples have been measured.
0 Date scroll button
Change the measurement date.
a Display calendar button
Display the calendar. If a date within the cal-
endar is tapped, the measurement results of
that day are displayed.
è Refer to the next page.

897N203029A FDC NX600 Instruction Manual 2022.01 3-16


3 Measurement

< Calendar display >


1 1 Measurement month scroll buttons
Change the display of the calendar month.
2 Date buttons
Display the measurement results screen of
2 a date. The currently displayed date is dis-
played in blue.

NOTE Dates with no measurements are


displayed in gray and cannot be se-
lected.

3-17 897N203029A FDC NX600 Instruction Manual 2022.01


3 Measurement

3.5 ISE Tests

IMPORTANT:

Analyzer cannot detect expiries of ISE slides if the installed software is version 1.0. Be sure manu-
ally that slides are still valid.
Otherwise, with the software version 1.1 or later is installed, analyzer can automatically detect ex-
piries of the ISE slides that will expire 2023 or later or whose lot numbers are 130000 to 139999.

NOTE When urine (U) is selected as the sample type for measurement, the dilution setting is set to
2x automatically. Therefore, diluent is needed in addition to reference fluid. The data is cal-
culated automatically, so the results can be reported without adjustment. 3
NOTE Refer to the “Instructions for Use” of the slide for usable diluent for urine measurements.

NOTE Two tips are necessary for each slide; one for sample, the other for reference fluid.

1. Load the reference fluid.


For the reference fluid container (FUJI plain
tube 0.5 mL or 1.5 mL), use the container con-
figured for [Tube setting for diluent & reference
fluid] in Mode function for administrator (Section
9.3.15).
For details of the loading method, refer to Sec-
tion 5.5.
2. Check quantity of tips.
• Loading tips è Refer to Section 5.2
Reference fluid
3. Load diluent for urine sample measure-
ment.
Also, check the quantity of mixing cups. If the
number is insufficient, replace with new ones.
• Loading mixing cups è Refer to Section 5.3
• Loading diluent è Refer to Section 5.4

NOTE When performing an ISE measurement


that requires dilution, use at least 3 tips
and 1 well.

897N203029A FDC NX600 Instruction Manual 2022.01 3-18


3 Measurement

4. Check the sample type setting.


Register sample information as necessary.
è Refer to Section 3.3 (4)

Sample
type setting

Remaining
3 number of
mixing cups

Arrows pointing
5. Load a sample and slides.
right
矢印右向き Place the ISE slide in the slide cartridge with
the spotting hole above and the arrows point-
ing right. For details of the loading method, see
Section 3.3 (3).

NOTE The direction of the arrow on the slide is


opposite to the slide transport direction.

NOTE When you measure an ISE slide and


CM slide at the same time, we recom-
mend loading the ISE slide last in the
cartridge.
è Refer to Section 4.5

6. Close the sample cover, and then lock it


with the sampler cover lock key.

WARNING:

Ensure the sampler cover is locked during sample


processing to prevent injuries and infection.
When the display reads [Ready to test], the sam-
pler cover can be opened.

3-19 897N203029A FDC NX600 Instruction Manual 2022.01


3 Measurement

7. Tap the [START] button to start measure-


ment.

897N203029A FDC NX600 Instruction Manual 2022.01 3-20


3 Measurement

3.6 Dilution Tests


There are 2 types of dilution measurement.
• Dilute and measure automatically by using [Dilution setting] in Mode function for administrator to
preset dilution factors for each test and sample type.
è Refer to Section 9.3.14
• Dilute and measure all items to be measured by setting a dilution factor for each sample at the
sample information registration screen.
The second method at the sample information registration screen is explained in this section.

IMPORTANT:
3 Bubbles may appear in the tip when diluting a plasma/serum sample more than 20 times or dilut-
ing a urine sample. If bubbles have formed during dilution, perform the measurement again from
the beginning.

NOTE The results of measurements that included dilution are calculated automatically, so the re-
sults can be reported without adjustment.

NOTE The maximum number of dilution measurements from one mixing cup well (hole) is three.
When the diluted sample in the well is insufficient, the analyzer automatically dilutes using
the next well.

NOTE Do not dilute whole blood samples.

NOTE The default dilution factors designated by the manufacturer are as follows: ISE urine tests:
2x; CRP tests: 21x.
Excluding ISE tests, the following section explains the relationships among the dilution fac-
tor configured for each sample at the sample information registration screen, the dilution
factor for each test configured by [Dilution setting] in Mode function for administrator, and
the dilution factor designated by the manufacturer. For ISE tests, the analyzer ignores the
dilution factors configured for each sample or each test and using default dilution factor of
the manufacturer.
• Dilution factor when configured for each sample at the sample information registra-
tion screen
Dilution factor configured Default
×
for each sample dilution factor
• Dilution factor when not configured for each sample at the sample information regis-
tration screen
(Configured by [Dilution setting] in Mode function for administrator)
Dilution factor for each Default
×
test dilution factor
(Example) With a CRP measurement with the dilution factor for each sample or test
configured to 3x, the analyzer will automatically dilute 63x.
NOTE Only one type of diluent can be used to measure a sample.

3-21 897N203029A FDC NX600 Instruction Manual 2022.01


3 Measurement

(Example) When diluting Sample 2x

Diluent 1. Check the quantity of the consumables


(tips and mixing cups).
If consumables are insufficient, supplement or
replace them.
• Loading tips è Refer to Section 5.2
• Loading mixing cups è Refer to Section 5.3
2. Load the diluent.
RE • Loading diluent è Refer to Section 5.4

3. Load a sample and slides.


For details of the sample and slide loading
method, refer to Section 3.3 (3).

4. Tap the [Dilution] button.

897N203029A FDC NX600 Instruction Manual 2022.01 3-22


3 Measurement

5. In the list, tap [x2].


- : ‌Dilution is performed according to [Dilution
setting] in Mode function for administra-
tor.
None: No dilution

NOTE If necessary, enter or choose other


sample information.

3-23 897N203029A FDC NX600 Instruction Manual 2022.01


3 Measurement

6. Close the sample cover, and then lock it


with the sampler cover lock key.

WARNING:

Ensure the sampler cover is locked during sample


processing to prevent injuries and infection.
When the display reads [Ready to test], the sam-
pler cover can be opened.

7. Tap the [START] button to start measure-


ment.

8. The dilution factor is added on the test


result.

897N203029A FDC NX600 Instruction Manual 2022.01 3-24


3 Measurement

3.7 CRP Tests


NOTE Each time a new box of CRP slides is used, the new QC card must be read.
è Refer to Section 3.10
NOTE Calibration is necessary for CRP measurements under the following conditions.
• When the slide lot has changed
• When the validity term of calibration has expired (For the validity term, refer to the “In-
structions for Use” of the slides)

NOTE If the validity term has expired, a “#” warning mark is printed along with the measurement
results.
3 NOTE Specified diluent and mixing cups are necessary for CRP measurements.

NOTE When the results of a measurement will exceed the measurement range, dilution can be
performed automatically at 21x or more by configuring the dilution factor at the sample in-
formation registration screen. (When 2x has been set at the sample information registration
screen, the dilution factor is 42x.)

Diluent for 1. Check the quantity of the consumables


CRP (tips and mixing cups).
If consumables are insufficient, supplement or
replace them.
• Loading tips è Refer to Section 5.2
• Loading mixing cups è Refer to Section 5.3

NOTE When performing measurements that


require dilution, at least 2 tips are used.

NOTE When performing CRP tests that require


dilution at 2x or more, 2 wells (holes) of
mixing cups are used.
2. Load the diluent for CRP.
• Loading diluent è Refer to Section 5.4

3. Load a sample and slides.


For details of the sample and slide loading
CRP method, refer to Section 3.3 (3).
一般比色
CM slide
NOTE When measuring a CRP slide and CM
slide at the same time, stack the CRP

図-4.9 slide last or first in the cartridge. If a CM


slide is measured before or after a CRP
slide, extra tips are used.

3-25 897N203029A FDC NX600 Instruction Manual 2022.01


3 Measurement

4. Set the sample type to [Plasma/Serum].


NOTE To further set the dilution factor to 3x for
CRP measurements, tap the [Dilution]
button, and then specify 3x.
è Refer to Section 3.6
NOTE If necessary, enter or choose other
sample information.

5. Close the sample cover, and then lock it


with the sampler cover lock key.

WARNING:

Ensure the sampler cover is locked during sample


processing to prevent injuries and infection.
When the display reads [Ready to test], the sam-
pler cover can be opened.

6. Tap the [START] button to start measure-


ment.

897N203029A FDC NX600 Instruction Manual 2022.01 3-26


3 Measurement

3.8 Shutting-down the FDC NX600

Sample 1. Remove the samples.


2. Remove the tubes of diluent and refer-
Diluent ence fluid from the analyzer.
Reference
fluid

3 RE

Sampler cover 3. Close the sampler cover.

4. Make sure that measurement and mode


operation are not in progress.

CAUTION:

Do not switch off the power during measurement or


mode operation. Otherwise, the analyzer may mal-
function. If the power was shut off during measure-
ment, load tips and other consumables, and then
run the measurement again from the start. Always
replace mixing cups for dilution measurements.

3-27 897N203029A FDC NX600 Instruction Manual 2022.01


3 Measurement

5. Press the power button.

Power button
3
6. Tap [OK].

CAUTION:

Shutdown will be automatically performed after the


process completed. Make sure NOT to turn the
power switch off.

CAUTION:

Were you not able to operate on the screen, press


the power button at front to turn off the power.

NOTE When the analyzer will not be used for a


long time, set the power switch to the [○]
side.

897N203029A FDC NX600 Instruction Manual 2022.01 3-28


3 Measurement

7. Empty the disposal box.

WARNING:

Do not touch used slides and tips with your


bare hands, as this may cause infection. If you
accidentally touch an item, immediately rinse
the contaminated area thoroughly under run-
ning water, and then disinfect the area. Seek
medical assistance if necessary.

WARNING:
3 Used slides and tips are infectious industrial
waste. Process this waste in compliance with
any applicable regulations in your country,
such as by incineration, melting, sterilization,
or disinfection.

NOTE When disinfecting the disposal box, use


ethyl alcohol for disinfection or a 0.5%
sodium hypochlorite solution. When a
sodium hypochlorite solution was used,
wash the disposal box well and dry it
before use.

3-29 897N203029A FDC NX600 Instruction Manual 2022.01


3 Measurement

3.9 Sample Tubes for the FDC NX600


The following sample tubes can be used on the FDC NX600.
• Ø16×100 mm blood collection tube • Ø16×100 mm blood collection tube (when
• Ø13×100 mm blood collection tube using PF)
• Ø13×75 mm blood collection tube • Ø13×100 mm blood collection tube (when
using PF)
• FUJI Sample Tube 1.5 mL
• Ø13×75 mm blood collection tube (when us-
• FUJI Sample Tube 0.5 mL
ing PF)

3.9.1 Blood Collection Tubes


(1) Size and effective liquid surface range of blood collection tubes that can be
3
used
Outer diameter of Ø16 indicates Ø15.0 to 16.6 mm and Ø13 indicates Ø12.0 to 13.3 mm. The
following illustrations show the design suction ranges. If the sample volume is low, use a FUJI
Sample Tube (1.5 mL or 0.5 mL).
〈φ16またはφ13
Ø16 x100mm採血管〉
or Ø13×100 mm blood collection tube 〈φ13
Ø13×75 x75mm採血管〉
mm blood collection tube

Not measurable
測定不能範囲 10mm
10 mm 10mm
10 mm 測定不能範囲
Not measurable

測定可能範囲
Measurable 測定可能範囲
Measurable
75mm
75 mm
100mm
100 mm

30mm
30 mm 測定不能範囲
Not measurable

Not測定不能範囲
measurable 52mm
52 mm

Sample volume confirmation holes (left and right)


Load the blood collection tube with filled sample
in the rack, so that you can see the actual sample
through the holes.

897N203029A FDC NX600 Instruction Manual 2022.01 3-30


3 Measurement

(2) Tube contents and inspection items that can be measured

IMPORTANT:

Refer to the “Instructions for Use” of the slides for usable blood collection tubes and observe the
listed items.

IMPORTANT:

For contents, inspection items, cap colors, and other details, contact the manufacturer of the blood
collection tube to be used.

3 The tube contents of a blood collection tube determine which inspection items can be meas-
ured. The following section provides information about general blood collection tube contents
and the inspection items that can be measured.

Heparin blood collection tubes


• Inspection items: Applicable to all items. For GLU and NH3, measure the samples as soon
as possible after blood collection.
• General color of cap: Green (in Japanese market)
Plain (coagulant included) blood collection tubes
• Inspection items: All other than GLU and NH3
• General color of cap: Red (in Japanese market)
NaF blood collection tubes
• Inspection item: GLU (blood sugar) only
• General color of cap: Gray (in Japanese market)
EDTA blood collection tubes
• Inspection items: NH3 only. After blood collection, keep samples in ice and measure the
samples as soon as possible.
• General color of cap: Pale purple (in Japanese market)

3-31 897N203029A FDC NX600 Instruction Manual 2022.01


3 Measurement

3.9.2 FUJI Sample Tubes


Use these tubes in the place of blood collection tubes when the sample volume is low.

WARNING:

When handling samples (blood or urine), always follow biohazard procedures (wear gloves, a lab
coat, and safety goggles) in accordance with the sample handling rules of your facility. If you acci-
dentally touch a sample, immediately rinse the contaminated area thoroughly under running water,
and then disinfect the area. Seek medical assistance if necessary.

Make sure that the sample volume is sufficient by

1.5 Upper
referring to the illustration to the left. The top sur-
face of the sample fluid must be within the shaded
3
検体上限
limit of
0.5 section of the tubes in the illustration.
sample
1.0 0.3 NOTE If the sample in the tube is below the suc-
Lower
分離用limit
of 検体下限
sample for tion limit mark, the analyzer will indicate an
0.5 separation 0.1 insufficient sample volume error.
Suction
吸引限界 limit
NOTE To aspire only the supernatant fluid (plasma)
of the centrifuged sample (and avoid aspi-
ration of red blood cells), at least 1.0 mL (for

0025
a 1.5 mL tube) or 0.3 mL (for a 0.5 mL tube)
of a whole blood sample for separation is
necessary.

NOTE Only 5 tests or fewer can be performed us-


ing a 0.3 mL whole blood sample with a
FUJI Sample Tube (0.5 mL).

Effective liquid surface range of a FUJI Sample Tube


Upper 検体上限
limit of sample

1.5 検体上限
0.5

1.0 0.3

0.1
0.5
吸引限界

注:検体は、50μl 以上必要です。

吸引限界
Suction limit NOTE: At least 50 µl of sample is needed.

富士チューブ サンプルカップ2ml

897N203029A FDC NX600 Instruction Manual 2022.01 3-32


3 Measurement

3.9.3 Assembling the Sample Rack


Assemble the sample rack according to the type of blood collection tube and tube to be used.

IMPORTANT:

Incorrect assembly may cause bad suction, which may adversely affect the test results or cause a
failure.

<Assembly preparation>
The following parts are included in the bundled items.

3 NOTE The number in parentheses is the number of parts included in the bundled items.

NOTE The spacer (9) is are not used for the FDC NX600i.

2Tube holder (1) 3Tube holder (1)


for φ 16mm tube for φ 13mm tube

1Sample rack body (4) 4Tube holder (1) 5Tube holder (1) 6Calibrator CP
for FUJI tube 1.5mL for FUJI tube 0.5mL holder (1)

0Pin (9)

H H

7Spacer (1) 8Spacer (1) 9Spacer (1)


for 75mm tube for 100mm tube for PF 75mm tube
H = 54.5mm (long) H = 32mm (medium) H = 25mm (short)
aLabel (1)

Note: There are upper- and lower-sides of spacers. Do not turn them upside down.

3-33 897N203029A FDC NX600 Instruction Manual 2022.01


3 Measurement

<Sample rack variations>

(a) Sample rack for φ16mm × 100mm


Affix the label
After setting the 8
spacer, insert the 2
tube holder

Set the 8 spacer (H=32mm)


(hole is down)

Insert the 0 pin


as shown
3
(b) Sample rack for φ13mm × 100mm
Affix the label
After setting the 8
spacer, insert the 3
tube holder

Set the 8 spacer (H=32mm)


(hole is down)

0 Pin

(c) Sample rack for φ13mm × 75mm


Affix the label
After setting the 7
spacer, insert the 3
tube holder

Set the 7 spacer (H=54.5mm)


(hole is down)

0 Pin

(d) Sample rack for FUJI sample tube 1.5mL


Affix the label Insert the 4 tube
holder

0 Pin

0 Pin

897N203029A FDC NX600 Instruction Manual 2022.01 3-34


3 Measurement

(e) Sample rack for FUJI sample tube 0.5mL


Affix the label Insert the 5 tube
holder

0 Pin

3
(f) Sample rack for φ16mm × 100mm when
using plasma filter PF Affix the label

0 Pin
0 Pin

(g) Sample rack for φ13mm × 100mm when


using plasma filter PF Affix the label

0 Pin
0 Pin

(h) Sample rack for φ13mm × 75mm when us-


ing plasma filter PF Affix the label

Set the 9 spacer (H=25mm)


(hole is down)

0 Pin
0 Pin

3-35 897N203029A FDC NX600 Instruction Manual 2022.01


3 Measurement

(i) Sample rack for calibrator CP


Affix the label Insert the 6 calibrator CP
holder

0 Pin

0 Pin

897N203029A FDC NX600 Instruction Manual 2022.01 3-36


3 Measurement

3.10 QC Card System


(1) Overview
The QC card system compensates for variations between lots of FUJI DRI-CHEM slides and
ensures uniform performance.

Y
Y Calculation method
A variety of the reference fluids developed for FUJI
標Reference

yy33 DRI-CHEM slides are measured on each lot of


準液基準値 liquid standard value Y

slides.
The relationship between the actual measurement
yy22 values X (x1, x2, x3) and the standard values Y (y1, y2,
3 Y y3) is expressed by the formula Y=c+dX+eX2. The
lot compensation coefficients (c, d, and e) derived
yy11 from the formula are used to compensate for varia-
tions in X of each lot.
XX
x11 x22 x33

0030 スライド (補正前)


Actual measurement の実測値
values of slideX(before
compensation) X

(2) Inputting lot compensation coefficients using a QC card


The lot compensation coefficients are recorded on QC cards, and one QC card is packed with
each box of slides. When data is input from a new card, the previously stored lot compensation
coefficients are updated. Always read the new QC card when you switch to a new box of slides,
because each QC card also contains slide data essential for measurement, in addition to the lot
compensation coefficients.

NOTE The “e” values recorded on the QC cards are expressed as e×103.

NOTE The analyzer records the QC card information of up to 5 lots for each test. Reading QC
cards of more than 5 lots results in deletion of QC information starting from the oldest.
(3) Using QC cards
A QC card must be read when a new test slide will be used or when changing to a new lot of
slides.
Read the QC card while [Ready to test] or [Warming up] is displayed on the LCD.

3-37 897N203029A FDC NX600 Instruction Manual 2022.01


3 Measurement

Set the QR code on the left side, and then insert


the QC card into the card reader fully.

NOTE If the QC card is dirty or folded, the read


error may occur.

QR code

The slide test and production number (lot) are


displayed on the LCD and output to the printer
as well.

NOTE Do not insert another QC card while the


slide test and lot number are displayed.

NOTE When the analyzer beeps and the test


is not displayed, insert the QC card into
the card reader again.

(4) Using DI cards


When you receive DI cards from FUJIFILM due to a change of the slide’s manufacturing pro-
cess, read the DI cards according to the instructions. The reading method is the same as for QC
cards.

897N203029A FDC NX600 Instruction Manual 2022.01 3-38


4 Applicable Usage Methods
4.1 CRP Calibration
4.1.1 Calibration Measurement
Calibrate when you start a new production number (lot) of CRP slides or when calibration has ex-
pired.

WARNING:

As calibrator CP is derived from blood, it should be handled in the same way as a patient sample.
When handling calibrator CP, always follow biohazard procedures (wear gloves, a lab coat, and
safety goggles) in accordance with the sample handling rules of your facility. If you accidentally
touch a sample, immediately rinse the contaminated area thoroughly under running water, and
then disinfect the area. Seek medical assistance if necessary.

IMPORTANT:

Perform the calibration measurement using Unit (A) (mg/dL).


4
IMPORTANT:

When the calibrator cap is opened, there may be a liquid film on the bottle opening. In this case,
always remove the film before the calibration process.

IMPORTANT:

The cap of the calibrator should not be open more than 5 minutes during a single calibration.

IMPORTANT:

Use a calibrator 10 times or less, even during its validity term.

IMPORTANT:

If the date and time are not adjusted correctly, the analyzer may fail to determine the expiration of
the slides and calibration, and measurement results might be incorrect.

NOTE When performing calibration measurements, the analyzer automatically dilutes the calibrator.

NOTE Let CRP diluent and calibrator stand until they reach room temperature before use.

897N203029A FDC NX600 Instruction Manual 2022.01 4-1


4 Applicable Usage Methods

1. Read the QC card via the analyzer.


Read the QC card of the slide to be calibrated.
è Refer to Section 3.10

2. Tap the button.

3. Tap the [Calibration] button.

4. Check the displayed calibrator concen-


tration settings (CP1, CP2, and CP3).

IMPORTANT:

When the calibrator concentration values are in-


correct, calibration is not performed normally.

NOTE Refer to the “Instructions for Use” of the


calibrator.

NOTE You can change the calibrator concen-


tration settings by tapping the [Settings]
button. In such a case, the administra-
tor password must be entered at the
Administrator password entry screen.
For the calibrator concentration settings,
refer to Section 4.1.2.

4-2 897N203029A FDC NX600 Instruction Manual 2022.01


4 Applicable Usage Methods

5. Tap the [START Calibration] button.


The guide for setting consumables such as cali-
brators is displayed.

4
6. Check the quantity of the consumables
(tips and mixing cups).
If consumables are insufficient, tap the [Set con-
sumables] button to replenish or replace them.
• Setting of tips è Refer to Section 5.2
• Setting of mixing cups è Refer to Section 5.3
7. Load the diluent for CRP.
• Loading diluent è See Section 5.4

IMPORTANT:

Always use diluent for CRP.

897N203029A FDC NX600 Instruction Manual 2022.01 4-3


4 Applicable Usage Methods

8. Prepare for measurement of calibrator


CP1.
NOTE A set of calibration requires 6 CRP
slides in total.
(a) Set the sample rack for the calibrator CP.

NOTE Set the sample rack securely so that it


does not rise up.
(b) Insert 2 CRP slides into the slide cartridge
and set them on the analyzer.

CP sample rack NOTE Set the slide cartridge securely so that it


does not rise up.
Otherwise, the slide reading error may
occur.
4 (c) Open the lid of the CP and place it in the
sample rack.

CP
IMPORTANT:

The CP should be open no more than 5 minutes


during a single calibration.
(d) Close the sampler cover, and then lock it
with the sampler cover lock.

Diluent for CRP


IMPORTANT:

Always lock the sampler cover with a key during


sample processing to prevent injuries and infec-
tion.
When the display reads [Ready to test], the sam-
pler cover can be opened.

4-4 897N203029A FDC NX600 Instruction Manual 2022.01


4 Applicable Usage Methods

9. Tap the [START] button. The CP1 calibra-


tion spotting screen is displayed.

10. After CP1 sampling, open the sampler


cover, take CP1 out from the sample
rack, and then firmly close the lid.
The CP2 calibration measurement start screen
is displayed.

897N203029A FDC NX600 Instruction Manual 2022.01 4-5


4 Applicable Usage Methods

11. Perform measurement of calibrator CP2.


Perform (b) to (d) of Step 8, and then tap
the [START] button.
The CP2 calibration spotting screen is dis-
played.

NOTE If a slide with a different lot number has


been loaded, an error screen is dis-
played.

If the [OK] button is tapped, the original


screen returns. Check the slide and
continue the measurement.

4-6 897N203029A FDC NX600 Instruction Manual 2022.01


4 Applicable Usage Methods

12. After CP2 sampling, open the sampler


cover, take CP2 out from the sample
rack, and then firmly close the lid.
The CP3 calibration start measurement screen
is displayed.
13. Perform measurement of calibrator CP3.
Perform (b) to (d) of Step 8, and then tap
the [START] button.
The CP3 calibration spotting screen is dis-
played.

897N203029A FDC NX600 Instruction Manual 2022.01 4-7


4 Applicable Usage Methods

14. After CP3 sampling, open the sampler


cover, take CP3 out from the sample
rack, and then firmly close the lid.
The calibration measuring screen is displayed.

4
15. When measurement ends, the Calibra-
tion measurement result screen is dis-
played.
NOTE Print the calibration measurement re-
sults and store them for the calibration
validity term. To print, tap the but-
ton.

NOTE The calibration parameters are saved to


the analyzer automatically.

16. Tap the [OK] button to end calibration.

4-8 897N203029A FDC NX600 Instruction Manual 2022.01


4 Applicable Usage Methods

(Supplement) List display area

The lot numbers (6 digits) and validity terms of read


QC cards (CRP) are displayed in a list. Items are
displayed in the order the QC card was read. Items
within their validity term have their dates displayed
in black. The following items are displayed in red:
the date when the calibration validity term has ex-
pired, [Slide expired] when the validity term of the
slide has expired, and [Unexecuted] when calibra-
tion is not been performed.

897N203029A FDC NX600 Instruction Manual 2022.01 4-9


4 Applicable Usage Methods

4.1.2 Configuring Calibrator Concentration


Correct the calibrator concentration when the concentration included with the calibrator CP to be
used and the concentration in the calibrator concentration display area at the Calibration Lot No.
screen are different.

1. Tap the [Settings] button.


The administrator password entry screen ap-
pears.

2. Enter the administrator password, and


then tap the [OK] button.
The Calibrator concentration configuration
screen appears.

NOTE The administrator password is neces-


sary when moving from the Function
menu screen to the Administrator function
screen.

4-10 897N203029A FDC NX600 Instruction Manual 2022.01


4 Applicable Usage Methods

3. Tap the [Input] field of each CP.


The software keyboard screen appears.

4. Enter a value, and then tap the [OK] button.


For the details of the software keyboard screen,
refer to Section 2.3.

IMPORTANT:

Enter the value using Unit (A) (mg/dL).

897N203029A FDC NX600 Instruction Manual 2022.01 4-11


4 Applicable Usage Methods

5. If the button is tapped, the entered


values are registered, and the analyzer
returns to the Calibration Lot No. screen.
NOTE If the [Reset] button is tapped, the en-
tered values are reset. If the button
is tapped, the entered values are regis-
tered and printed.

4-12 897N203029A FDC NX600 Instruction Manual 2022.01


4 Applicable Usage Methods

4.2 Manual Spotting (Colorimetric Slide)


4.2.1 Measuring by Manual Spotting (Colorimetric Slide)

WARNING:

When handling samples (blood or urine), always follow biohazard procedures (wear gloves, a lab
coat, and safety goggles) in accordance with the sample handling rules of your facility. If you acci-
dentally touch a sample, immediately rinse the contaminated area thoroughly under running water,
and then disinfect the area. Seek medical assistance if necessary.

IMPORTANT:

In the case of a CM slide, use a dedicated pipette that matches the sample application volume
specifications in the "Instructions for Use" of the slide. Carefully read the operation method of the
pipette in Section 4.2.2. (Be aware that the sample application volume of GLU-W is 6 µL, which is
different from other slide items.) 4
NOTE Manual sample application is not possible with ISE slides.

NOTE When the [Dilution] button was used to configure the dilution factor in manual sample appli-
cation mode, the measurement value is calculated by a dilution factor.

1. Prepare the pipette, slides, and a sample.

2. Tap the button.

897N203029A FDC NX600 Instruction Manual 2022.01 4-13


4 Applicable Usage Methods

3. Tap the [START Manual pipetting] button.

4. Register the measurement information.


NOTE You can also display the measurement
results screen from the screen on the
left and request a retest.
è Refer to Section 4.4

5. Load slides.
For details of the loading method, see Section
3.3 (3).

Manual pipette cover


6. Place the Manual pipette cover on the
slide cartridge.
NOTE If the cover is not set, the analyzer may
fail to read the slide information and
cannot start the measurements.

RE

4-14 897N203029A FDC NX600 Instruction Manual 2022.01


4 Applicable Usage Methods

7. Aspirate the sample by using the pipette.


NOTE Refer to the “Instructions for Use” of the
slide for the amount of sample to be
aspirated (applied).

8. Prepare sample pipetting.


Hold the pipette and touch the tip mounted on
the pipette to the guide (your side corner) of the
Manual pipette cover.
(This helps position the tip in the center of the
slide.)

Guide the tip using your side corner.

9. Tap the [START] button.


A slide is transferred to the sample application
position.

10. Prepare so that sample application with


the pipette is always possible, and then
tap the [START] button.

897N203029A FDC NX600 Instruction Manual 2022.01 4-15


4 Applicable Usage Methods

11. After the progress bar has finished (all


□ change to ■), apply the sample to the
slide while [Spot now] is displayed.
The spotted slide is automatically transferred
to the measurement section after the display of
[Spot now] disappears.

IMPORTANT:

Apply the sample while the buzzer is sounding


(while [Spot now] is displayed). Otherwise, the
measurement results will be inaccurate.

IMPORTANT:

4 When applying sample to a CM slide, apply the sam-


ple within 1 mm of the center of the slide. Otherwise,
the measurement results will be inaccurate. Run the
measurement again.

参照液
検体
搬送方向
(インキュベータ方向)

スライドの中央に点着する
Apply sample to the center
(比色スライ
of the ド)slide).
slide (CM

NOTE The LCD may not display [Wait] de-


pending on the operation conditions.

12. To apply sample to the next slide, repeat


the procedure from Step 5.
13. Press the [HOME] button to end manual
sample application.
NOTE The sampler unit moves.

Apply sample to the center of the slide


(CM slide)

4-16 897N203029A FDC NX600 Instruction Manual 2022.01


4 Applicable Usage Methods

4.2.2 Using a Pipette

WARNING:

When handling samples (blood or urine), always follow biohazard procedures (wear gloves, a lab
coat, and safety goggles) in accordance with the sample handling rules of your facility. If you acci-
dentally touch a sample, immediately rinse the contaminated area thoroughly under running water,
and then disinfect the area. Seek medical assistance if necessary.

IMPORTANT:

Mount the tip securely into the nozzle of the pipette.


If mounting is poor, the tip may be dislodged, and the sample may scatter.

Applying samples to slides 4


1. Move the tip of the pipette near the slide.
2. Slowly push the push button and make a
drop.

液滴を作ってスライドに近づけます

液滴を作ってスライドに近づけます
Make 液滴を作ってスライ ドに近づけます
a drop and then move it near the slide

3. Place the drop on the center of the slide.

IMPORTANT:

Always make a drop at the top of the tip, and then


液滴をスライドの中央に点着します apply the sample to the slide. If the sample is
discharged to the slide surface without making a
液滴をスライ
Place the drop ドの中央に点着 します
on the center of the slide
drop, the measurement results may be affected.
液滴をスライドの中央に点着します

While the slide is absorbing


スライドが検体を吸い取り ながら the

0057
sample,
内部に拡散 it diffuses
していきinside
ます
スライドが検体を吸い取りながら

0057
内部に拡散
スライ していきます ながら
ドが検体を吸い取り

0057
内部に拡散していきます

897N203029A FDC NX600 Instruction Manual 2022.01 4-17


4 Applicable Usage Methods

4.3 Tests Using a Plasma Filter (PF)


By using a PF, you can filter a plasma sample from a whole blood sample without using a centrifuge.

NOTE The FDC NX600i cannot measure using a PF.

IMPORTANT:

Read the “Instructions for Use” of the PF before measurement.

IMPORTANT:

Be sure to read the PF card when changing to a new production lot of PFs.

CAUTION:

Do not tap the [START] button when the sample has already been filtered by the PF. If the [START]
4 button is tapped, the analyzer may malfunction.

Precautions for samples


NOTE Use whole blood samples that have a hematocrit value (Hct) of 20 to 55%.
Up to about 350 μL of plasma (including 85 μL of dead volume) can be filtered.
The amount of filtered plasma decreases as the hematocrit value increases. In the case of
a hematocrit value of 55%, about 185 μL of plasma (including 85 μL of dead volume) can be
filtered.
The amount of filtered plasma depends on the hematocrit value and viscosity of the blood.

NOTE Hemolyzed samples provide inaccurate results.

NOTE If blood cells in a sample have precipitated, mix the sample, place it in a sample rack, and
then perform measurement immediately. Otherwise, the amount of filtered plasma may be
insufficient.

Precaution for blood collection tubes

IMPORTANT:

Use the blood collection tubes listed in Section 4.3.2.

4-18 897N203029A FDC NX600 Instruction Manual 2022.01


4 Applicable Usage Methods

PF cards
(1) Overview
When the measurement values for filtered plasma are compared to the values for plasma sep-
arated on a centrifuge, systematic differences may be found. Always read the PF card before
using a PF. The PF card contains information to compensate for these differences.
(2) Using PF cards
One PF card is included in one PF box (with 50 filters). Read a PF card in the same way a QC
card is read. Be sure to read the PF card when changing to a new box of PF. The imported pro-
duction number (lot) is displayed on the Measurement-preparation screen. Make sure that this
number matches the production number (lot) of the PF in use. (è Refer to Section 2.2.1)

NOTE A PF card cannot be read during measurement.

897N203029A FDC NX600 Instruction Manual 2022.01 4-19


4 Applicable Usage Methods

4.3.1 Measurement

1. Prepare the slides.


2. Prepare the consumables.
Prepare the tips, mixing cups, diluent, and refer-
ence fluid as required.
• Loading tips è Refer to Section 5.2
• Loading mixing cups è Refer to Section 5.3
• Loading diluent è Refer to Section 5.4
• Loading reference fluid è Refer to Section
5.5

PF sample rack 3. Load the sample rack for PF and the


slide to be measured.
4

4-20 897N203029A FDC NX600 Instruction Manual 2022.01


4 Applicable Usage Methods

4. Load the sample.


(1) Place the specified amount of whole blood
sample in a blood collection tube.
NOTE The PF has about 3 mL of dead vol-
ume (that will be absorbed into the
filter). If the sample volume is low,
the filtered sample volume will be
too low for measurement. Further-
more, sample that was aspirated
into the PF filter cannot be reused.

NOTE If the sample volume is too high,


the sample may overflow from the
blood collection tube when the PF is
loaded. There must be empty space
in the tube. (Refer to Section 4.3.2.)
4
(2) Mix the whole blood sample in the blood
collection tube by turning it over five or six
times.
NOTE Make sure that bubbles do not form.
(3) Remove the cap from the blood collection
tube, and then put it on the PF sample rack.
(4) Insert a PF into the blood collection tube.
Make sure the PF does not rise from the
sample rack.

5. Register the measurement information.

IMPORTANT:

Tests using a PF measure filtered plasma, so al-


ways configure the sample type to [Plasma/Serum].

897N203029A FDC NX600 Instruction Manual 2022.01 4-21


4 Applicable Usage Methods

6. Close the sample cover, and then lock it


with the sampler cover lock key.

WARNING:

Ensure the sampler cover is locked during sample


processing to prevent injuries and infection.
When the display reads [Ready to test], the sam-
pler cover can be opened.

4
7. Tap the [START] button to start measure-
ment.
8. After measurement is ended, discard the
used PF, and then remove the sample
container.

WARNING:

When discarding used PF, always follow biohazard procedures (wear gloves, a lab coat, and safe-
ty goggles) in accordance with the sample handling rules of your facility. If any part of the body
comes in contact with a sample, immediately rinse the contaminated body part thoroughly under
running water and then use ethyl alcohol as a disinfectant. Seek medical assistance if necessary.

WARNING:

Used PF is infectious industrial waste. Process this waste in compliance with any applicable regu-
lations in your country, such as by incineration, melting, sterilization, or disinfection.

4-22 897N203029A FDC NX600 Instruction Manual 2022.01


4 Applicable Usage Methods

4.3.2 Blood Collection Tubes for Tests Using a PF


NOTE The FDC NX600i cannot measure using a PF.

The following blood collection tubes can be used with PF measurements: φ16×100 mm, φ13×100
mm, and φ13×75 mm.
(Outer diameter of Ø16 indicates Ø15.0 to 16.6 mm and Ø13 indicates Ø12.0 to 13.3 mm.)

IMPORTANT:

Use blood collection tubes containing Heparin.

IMPORTANT:

Refer to the “Instructions for Use” of the plasma filter (PF) and always follow the instructions.

4
NOTE Do not use blood collection tubes that may cause the tip of the sample suction nozzle of the
PF to be physically blocked by additives or other materials.

NOTE As shown in the illustration below, the required sample volume is 6.5 mL or more for
Ø16×100 mm tubes, 6 mL or more for Ø13×100 mm tubes, and 3 mL or more for Ø13×75
mm tubes.
PF
PF

10mmの空き
10 mm of empty 1010mmの空き
mm of empty
space is necessary
スペースが必要です space is necessary
スペースが必要です

測定可能範囲
Measurable
測定可能範囲
Measurable 75mm
75 mm
100mm
100 mm

検体は
The required sample
φ13x75mm管で3ml以上
検体は volume is 3 mL or more for
The required sample volume 必要です
Ø13×75 mm tubes.
φ16x100mm管で6.5ml以上
is 6.5 mL or more for Ø16×100

0023
mmφ13x100mm管で6ml以上
tubes, and 6 mL or more
for必要です
Ø13×100 mm tubes.

NOTE Use PF sample racks with the following labels for tests using a PF.

897N203029A FDC NX600 Instruction Manual 2022.01 4-23


4 Applicable Usage Methods

4.4 Retesting
You can test again using the previous sample number, sample ID, sample type setting, and refer-
ence interval setting by tapping the (retest) button at the measurement results screen. Further-
more, when a measurement cannot be performed due to an error, you can skip the re-entry and re-
configuration of the setting information by tapping the (retest) button. When a PF measurement
is run again, the analyzer does not filter using the PF, but instead uses plasma that was previously
filtered by the PF.

IMPORTANT:

The (retest) button is designed only to automatically enter the sample number and sample ID of
the previously measured sample. Do not use this button to retest samples from before the previous
sample.

CAUTION:
4 Do not tap the [START] button when the sample has already been filtered by the PF. If the [START]
button is tapped, the analyzer may malfunction.

Retesting after filtering by the PF


• When you will rerun a measurement using the PF, do not use the [START] button to restart the
measurement. Always use the (retest) button.
• Make sure that the sample volume is sufficient.
• Make sure that the liquid surface of the filtered sample is level.

IMPORTANT:

Do not retest GPT (ALT) of PF filtered plasma because its activity decreases over time.

4-24 897N203029A FDC NX600 Instruction Manual 2022.01


4 Applicable Usage Methods

< How to retest >

1. When [Ready to test] appears on the


LCD, make sure that the analyzer is not
applying samples, and then load the
consumables and sample.
è Refer to Section 3.3
NOTE If the amount of remaining sample is
insufficient, run the test with manual pi-
petting.
è Refer to Section 4.2

2. Tap the (retest) button on the meas-


urement results screen screen.
4

NOTE The sample number or sample ID that


was measured just before is displayed.

NOTE In case of a sample scheduled for re-


testing, is displayed.

NOTE Tap the [Dilution] button to set the dilu-


tion factor.

897N203029A FDC NX600 Instruction Manual 2022.01 4-25


4 Applicable Usage Methods

3. Close the sample cover, and then lock it


with the sampler cover lock key.

WARNING:

Ensure the sampler cover is locked during sample


processing to prevent injuries and infection.
When the display reads [Ready to test], the sam-
pler cover can be opened.

4
4. Tap the [START] button to start retesting.
NOTE When a test is rerun, PF filtering process
is not performed.

4-26 897N203029A FDC NX600 Instruction Manual 2022.01


4 Applicable Usage Methods

4.5 Slide Loading Order When Measuring Different Items Simul-


taneously
NOTE Any order of slide stacking can be measured. However, please follow the recommended
slide stacking rules below. Otherwise, the analyzer may consume excessive amounts of dis-
posal tips and cups.

(1) Measuring CRP and CM (colorimetric) slides

Always put a CRP slide first or last.


If a CRP slide is loaded between CM slides, an ex-
CRP tra tip will be used.
一般比色
CM slide

図-4.5.2-1
(2) Measuring CM and ISE slides 4
We recommend loading an ISE slide last.
If an ISE slide is measured first, it might take extra
電解質
ISE time for measurement.
一般比色
CM slide
If a CM slide is loaded after an ISE slide, an extra
tip will be used.

図-4.5.2-1
(3) Measuring CRP, CM, and ISE slides

Stack CM slides before CRP and ISE slides.


電解質
ISE
CRP

一般比色
CM slide

図-4.5.2-4
(4) Measuring CM slides that require dilution

If slides are configured with different dilution rates,


stack the slides so that slides with same dilution
5倍希釈
5x dilution factor are measured successively. Otherwise, extra
tips and mixing cups will be used.
2倍希釈
2x dilution
NOTE If all CM slides have the same dilution fac-

図-4.5.2-2
tor, there are no restrictions on the stacking
order.
<Not recommended slide loading order>

Only one loading position is prepared for the diluent


電解質
CRP in this analyzer. In the example, diluent (purified
CRP
CM (Dil) water etc.) for the colorimetric slide that requires di-
lution [CM (Dil)] is loaded and the measurement will
一般比色
CM slide start. The next measurement requires the specified
diluent for CRP, so that an error (E0160 Incorrect

図-4.5.2-4
diluent type error) occurs before the measurement,
and the measurement will be completed. Rerun the
CRP measurement from the beginning.

897N203029A FDC NX600 Instruction Manual 2022.01 4-27


4 Applicable Usage Methods

4.6 Automatic Acquisition of Sample Information (Work Lists)


If the analyzer is connected to DMS, you can receive patient IDs and their sample information by
tapping the button at the Measurement-preparation screen before measurement.

NOTE DMS is sold separately.

1. Tap the button. The analyzer receives


a work list (a list of patients for the sam-
ples to be measured) from DMS.

4 The Work list index screen appears.

NOTE If there are no measurement requests,


the buzzer sounds, and the message
[There is no request information] is dis-
played.

NOTE The display of the list is configured by


[Work list selection setting] in Mode
function for administrator.
è Refer to Section 9.3.8
NOTE On DMS, "Sample No." is "Sample ID"
and "Sample ID" is "Patient ID”.

NOTE When there are multiple pages in a


list, you can scroll the pages using the
/ buttons.

NOTE If the [Cancel Work list] button is tapped,


the analyzer returns to the Measure-
ment preparation screen.

2. Tap the sample information you want to


register.

4-28 897N203029A FDC NX600 Instruction Manual 2022.01


4 Applicable Usage Methods

The work list measurement items of the regis-


tered sample information are displayed.

NOTE If the button is tapped, registration


is cleared.
NOTE The work list measurement items can
be printed by tapping the button.

3. Tap the [OK] button to finish registration.


The Measurement preparation screen (top
screen) returns.

NOTE The work list button lights.

897N203029A FDC NX600 Instruction Manual 2022.01 4-29


4 Applicable Usage Methods

4.7 Editing Sample Numbers and Sample IDs


You can edit the sample number and sample ID of measurement results. You can also send edited
data to a host computer.

1. To edit the sample number, tap the [No.]


button at the measurement results screen
(Refer to Section 3.4).

2. Enter the administrator password, and


then tap the [OK] button.
The Sample number entry screen appears.
3. Edit the sample number, and then tap
the [OK] button.
The analyzer returns to the measurement results
screen.

NOTE You can also edit the sample ID using


the same procedure.

IMPORTANT:

Be sure not to make any entry errors.

NOTE The sample ID can also be edited by using


the barcode reader.

4-30 897N203029A FDC NX600 Instruction Manual 2022.01


4 Applicable Usage Methods

4.8 Using the Sample Barcode Reader (Sold Separately)


You can easily enter the sample ID and operator ID by using the optional sample barcode reader.

IMPORTANT:

If the sample barcode reader malfunctions or the print quality of the barcode label is poor, incor-
rect barcode data may be read. Make sure that the barcode data (sample ID) printed in the meas-
urement results are correct.

NOTE Carefully read the instructions included with the sample barcode reader before using it. A
maximum of 13 alphanumerical characters can be input for a sample ID. Always use the
sample barcode reader under lighting at 3,000 lux or lower and avoid direct sunlight.

NOTE Always connect the cable of the sample barcode reader before switching on the FDC
NX600.
4
<When inputting the sample ID>

1. Tap the [Sample ID] field. The software


keyboard appears.
NOTE In the following cases, this operation
is not required. The barcode reading is
possible on the top screen (left):
• When the USM function is enabled.
• When the setting "Use operator ID" is
OFF (USM function is disabled).
è Refer to Section 9.3.19.
NOTE You can also enter the sample ID using
the barcode reader on the measurement
results screen.

897N203029A FDC NX600 Instruction Manual 2022.01 4-31


4 Applicable Usage Methods

2. Read the barcode of the sample using


the barcode reader.
After reading is complete, the analyzer makes a
beep sound. The sample ID is displayed on the
LCD and is registered.

バーコードを読み取ります

NOTE Position the sample barcode reader


properly facing to the barcode label.
Otherwise, an incorrect reading may oc-
cur.

NOTE Make sure the displayed ID is correct.

NOTE When reading a barcode again, tap the


[Sample ID] field, and then re-read the
4 barcode.

4-32 897N203029A FDC NX600 Instruction Manual 2022.01


4 Applicable Usage Methods

<When inputting the operator ID>

1. Tap the operator ID button. The software


keyboard appears.

2. Read the operator ID using the sample


barcode reader.
After reading is complete, the analyzer makes a
beep sound. The operator ID is displayed on the
LCD and is registered.

NOTE Position the sample barcode reader


properly facing to the barcode label.
Otherwise, an incorrect reading may oc-
cur.

NOTE Make sure the displayed ID is correct.

NOTE To re-input a barcode, tap the Operator


ID button, and then re-read the barcode.

897N203029A FDC NX600 Instruction Manual 2022.01 4-33


5 Loading Consumables
5.1 Loading Consumables
To load tips and mixing cups, tap the display of consumables to display the loading screen. When
loading is complete, press the [Finish] button to exit.

NOTE If there are not enough consumables (tips, mixing cups, diluent, reference fluid, etc.), sam-
ple application may be stopped.

NOTE Load consumables while the analyzer is not applying samples.

897N203029A FDC NX600 Instruction Manual 2022.01 5-1


5 Loading Consumables

5.2 Setting Tips


Tips are required for all measurements. Check the number of remaining tips on the LCD before per-
forming measurements. If necessary, add new tips.

WARNING:
During sample application, tap the [STOP] button to stop application, and then load the tips. Load-
ing is unsafe during sample application because the sampler and the consumable disk are mov-
ing.

1. With [Warming up], [Lamp warming up]


or [Ready to test] displayed on the LCD,
make sure that the analyzer is not apply-
ing samples.
2. Tap the consumables display on the
LCD.

Consumable display

3. The Consumables loading screen is dis-


played.

5-2 897N203029A FDC NX600 Instruction Manual 2022.01


5 Loading Consumables

4. Remove the tip rack from the consum-


able disk.
NOTE Always remove the tip rack from the
consumable disk before loading tips.

RE

Tip rack

5. Load tips in all holes of the tip rack.


NOTE Always set tips in all holes of the tip
rack.

6. Load the tip rack on the consumable


disk.

RE

Tip rack loading position

897N203029A FDC NX600 Instruction Manual 2022.01 5-3


5 Loading Consumables

7. Check that tips have been set in all holes


of the tip rack. And then, tap the [Replenish]
button of tips.
8. Tap the [Finish] button to exit.

5-4 897N203029A FDC NX600 Instruction Manual 2022.01


5 Loading Consumables

5.3 Loading Mixing Cups


Mixing cups are necessary for measurements requiring dilution (refer to Section 3.6) and CRP
measurements (refer to Section 3.7). Furthermore, mixing cups are also necessary when urine (U)
is selected for the sample type during an ISE measurement because measurement is performed
automatically at 2x dilution. (Refer to Section 3.5) Check the quantity of mixing cups on the LCD be-
fore measurement. If necessary, load new mixing cups. Each measurement uses one mixing cup. If
no mixing cups remain, a measurement cannot be run.

WARNING:
During sample application, tap the [STOP] button to stop application, and then load the mixing
cups. Loading is unsafe during sample application because the sampler and the consumable disk
are moving.

WARNING:
Always wear gloves when handling used mixing cups to avoid biohazards.

NOTE The quantity of mixing cups that can be


used is displayed on the LCD. When the 5
amount remaining is insufficient, a dilution
measurement cannot be performed.

1. With [Warming up] or [Ready to test] dis-


played on the LCD, make sure that the
analyzer is not applying samples.
2. Tap the consumables display on the
LCD.

Consumable display

3. The Consumables loading screen is dis-


played.

897N203029A FDC NX600 Instruction Manual 2022.01 5-5


5 Loading Consumables

Mixing cups 4. Remove the mixing cups from the con-


sumable disk, and then set new mixing
cups.

WARNING:
Used mixing cups are infectious industrial waste.
Process this waste in compliance with any appli-
RE cable regulations in your country, such as by incin-
eration, melting, sterilization, or disinfection.

5. Tap the [Replace] button of mixing cups.


6. Tap the [Finish] button to exit.

5-6 897N203029A FDC NX600 Instruction Manual 2022.01


5 Loading Consumables

5.4 Loading Diluent


Diluent is necessary for dilution measurements (refer to Section 3.6) and CRP measurements (refer
to Section 3.7). Dilution and CRP measurements each require a specified diluent. Before measure-
ment, refer to the technical documentation for the diluent to be used. Furthermore, mixing cups are
also necessary when urine (U) is selected for the sample type during an ISE measurement because
measurement is performed automatically at 2x dilution. (Refer to Section 3.5)

WARNING:
During sample application, tap the [STOP] button to stop application, and then load diluent. Load-
ing is unsafe during sample application because the sampler and the consumable disk are mov-
ing.

IMPORTANT:

Use specified tube for the diluent. Otherwise, measurement errors or malfunctions may occur.
The specified tubes for diluent are the FUJI PLAIN TUBE (1.5 ml and 0.5 ml).

IMPORTANT:

The diluent for CRP measurements evaporates progressively, and the concentration of the fluid
5
may affect the measurement values.
Replace diluent for each test.

NOTE For each CRP measurement, 0.1 mL of diluent is necessary.

NOTE Configure the type of diluent container to be used via [Tube setting for diluent & reference
fluid] in Mode function for administrator (refer to Section 9.3.15).

FUJI PLAIN TUBE FUJI PLAIN TUBE 1. Prepare the diluent.


(1.5 ml) (0.5 ml)
Pour the following amounts into each type of
FUJI PLAIN TUBE.
For 1.5 mL tube: 1.0 to 1.5 mL
1.5 0.5
Pour diluent up to For 0.5 mL tube: 0.3 to 0.5 mL
this range 0.3 Suction
Suction 1.0 range NOTE The suction range is as shown on the
range 0.1 left.
0.5 When 0.5 mL of diluent was poured into
a FUJI PLAIN TUBE of 1.5 mL, one dilu-
tion is possible for a CRP measurement.

897N203029A FDC NX600 Instruction Manual 2022.01 5-7


5 Loading Consumables

2. With [Warming up] or [Ready to test] dis-


played on the LCD, make sure that the
analyzer is not applying samples.

Diluent 3. Load a tube containing diluent.

RE

5 NOTE Tap the consumables display on the top


screen and select the [Diluent] tab to dis-
play the diluent setting guidance screen.
This operation is not required because only
the guidance screen is displayed.

5-8 897N203029A FDC NX600 Instruction Manual 2022.01


5 Loading Consumables

5.5 Loading Reference Fluid


Reference fluid is necessary for ISE measurements. (Refer to Section 3.5).

WARNING:
When handling reference fluid, follow biohazard procedures (wear gloves, a lab coat, and safety
goggles) in accordance with the medical handling rules of your facility. If you accidentally touch an
item, immediately rinse the contaminated body part thoroughly under running water.

WARNING:
Discard reference fluid according to the medical handling rules of your facility and related laws.

WARNING:
During sample application, tap the [STOP] button to stop application, and then load the reference
fluid. Loading is unsafe during sample application because the sampler and the consumable disk
are moving.

IMPORTANT:
5
Use a FUJI PLAIN TUBE (0.5 or 1.5 mL) for reference fluid. Otherwise, measurement errors or
malfunctions may occur.
Configure the type of container to be used via [Tube setting for diluent & reference fluid] in Mode
function for administrator (Refer to Section 9.3.15).

IMPORTANT:

Replace FUJI PLAIN TUBES (1.5 mL) within 24 hours.


Replace FUJI PLAIN TUBES (0.5 mL) for each measurement.

FUJI PLAIN TUBE FUJI PLAIN TUBE 1. Prepare reference fluid.


(1.5 ml) (0.5 ml)
Pour the following amounts into each type of
FUJI PLAIN TUBE.
For 1.5 mL tube: 1.0-1.5 mL
0.5
Pour reference fluid For 0.5 mL tube: 0.3-0.5 mL
within the range. 0.3
1.0 NOTE Use reference fluid after it reaches room
0.1 temperature.
0.5

2. With [Warming up] or [Ready to test] dis-


played on the LCD, make sure that the
analyzer is not applying samples.

897N203029A FDC NX600 Instruction Manual 2022.01 5-9


5 Loading Consumables

Reference fluid 3. Load the tube containing the reference


fluid.
Press the lever to open the lid and load the con-
tainer.

IMPORTANT:

Clean the cap of the reference fluid every time


you load the reference fluid.
è Refer to Section 6.8

NOTE Make sure the reference fluid cap is


closed properly.

5
RE

Reference fluid

NOTE Tap the consumables display on the top


screen and select the [Reference fluid] tab
to display the reference fluid setting guid-
ance screen. This operation is not required
because only the guidance screen is dis-
played.

5-10 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance 6 Maintenance

To keep the performance of the FDC NX600 at its best, users must perform daily and periodic mainte-
nance, and designated repair personnel must perform maintenance. When repairs are necessary, con-
tact repair personnel.

WARNING:

When you clean the analyzer, always wear gloves, a lab coat, and safety goggles.
If you accidentally touch a contaminant, immediately rinse the contaminated area thoroughly under
running water, and then disinfect the area.
Seek medical assistance if necessary.

WARNING:

Always unplug the power cord before cleaning. (Cleaning of the transfer bar and cleaning around the
PF unit are excluded.)

WARNING:

Process contaminated swabs or cloths used for cleaning the analyzer as infectious waste.

CAUTION:

Before cleaning and maintaining the analyzer, get training from our technical support staff.
6
CAUTION:

Refer to each section for information about usable solvents for cleaning and decontamination.

CAUTION:

Do not use alcohol to clean the sampler cover (translucent). Otherwise, the surface will be damaged.

IMPORTANT:

After cleaning, always connect the power cord, switch on the power, and then check operation.

IMPORTANT:

If you do not perform the maintenance in this manual, the analyzer’s performance and specifications
will not be maintained, which may negatively affect the measurement values.

IMPORTANT:

Be sure to reassemble the parts removed for maintenance and tighten thumbscrews securely. Other-
wise, measurement values may be negatively affected.

NOTE If the LCD (touch panel section) becomes soiled, use a soft cloth or a cloth soaked in a small
amount of ethyl alcohol to lightly wipe it.

NOTE When cleaning the outer covers of the equipment, wipe using a soft cloth moistened with water
or ethyl alcohol for disinfection (excluding the sampler cover).

897N203029A FDC NX600 Instruction Manual 2022.01 6-1


6 Maintenance

6.1 Periodic Maintenance Items


6.1.1 Items for Daily and Periodic Inspection by the User
Users must perform the daily and periodic maintenance and replacement of parts in the following table.
Inspection
Location Cleaning interval Replacement interval Procedure
interval
Refer to
Air filter Once a month - -
Section 6.2.
Once every 3 months,
or as appropriate, such
Refer to
Transfer bar as when inaccurate - -
Section 6.3.
measurement values
occur
Once every 3 months,
or as appropriate, such
Refer to
Incubator as when inaccurate - -
Section 6.3.
measurement values
occur
As appropriate, such as
when a slide transfer Refer to
Sample application unit - -
error or tip ejection error Section 6.3.
occurs
Once every 3 months,
or as appropriate, such
6 ISE unit as when inaccurate
measurement values
- -
Refer to
Section 6.3.
occur
Appearance of red lines on both Refer to
Recording paper - -
sides of paper Section 6.4.
When a lamp replacement error
Refer to
Light source lamp - - occurs, or when the cumulative
Section 6.5.
lighting time exceeds 1,000 hours
Refer to
Sampler O-ring Once a week Once a month Once a year
Section 6.6.
When slide reading Refer to
Slide reading section - -
errors occur frequently Section 6.7.
Reference fluid cap When replacing refer- Refer to
- Once every 3 months
packing ence fluid Section 6.8.
PF sensor As appropriate, such as - -
Refer to
PF reference plate when errors related to - -
Section 6.9.
PF Suction Pad PF extraction occur Once a month Once every 2 years
Once every 2 years or when
Refer to
PF Tube - - the power has been off for one
Section 6.9.4.
month

WARNING:
After inspection, always connect the power cord, switch on the power, and then check operation.

6-2 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance

6.1.2 Items for Maintenance and Inspection by Repair Personnel


Designated repair personnel must perform the maintenance and replacement of parts in the follow-
ing table.
Parts Replacement interval
Interference filter Once every 4 years
Syringe Once every 3 years
Incubator pressure plate Inspection once every 3 years

WARNING:
After inspection, always connect the power cord, switch on the power, and then check operation.

NOTE The standard number of slides used per year for this analyzer is assumed to be 7,200 col-
orimetric slides and 240 ISE slides. (In the case of 240 days of use per year, 30 colorimetric
slides/day and one ISE slide/day)

897N203029A FDC NX600 Instruction Manual 2022.01 6-3


6 Maintenance

6.2 Cleaning the Air Filters


The air filters should be checked and cleaned once a month.

IMPORTANT:

If the air filters are not cleaned, an increase in the temperature inside the analyzer may negatively
affect the measurement values.

NOTE Clean all 4 filters.

1. Switch off the analyzer power.

2. Remove the four louvers and pull out the


air filters located inside of the louvers.
Lift the louver upward 1 and then pull forward
2 to remove.

Louver

Louver
6

6-4 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance

Air filter 3. Remove dust adhering to the filter with a


vacuum cleaner or running water.
NOTE If you wash the filter with running water,
make sure it is dried well before setting
in the analyzer.

4. Insert the filter into the louver and put


the louver back in the analyzer.
Push up the louver 1 and then push it toward
2.

1 IMPORTANT:
6
If the analyzer is used without setting the filters,
measurement values may be negatively affected.
2

897N203029A FDC NX600 Instruction Manual 2022.01 6-5


6 Maintenance

6.3 Inspecting and Cleaning the Transfer Bar, Incubator, Sam-


ple Application Unit, and ISE Unit
Measuring results are affected by stains (from blood samples, etc.) inside of the incubator and ISE
unit. These parts should be checked and cleaned at least once every three months. When a slide
transfer error or a tip ejecting error occurred, clean the sample application unit.

WARNING:
Always wear gloves, a lab coat, and safety glasses while working.
If you accidentally touch a contaminant, immediately rinse the contaminated area thoroughly un-
der running water, and then disinfect the area.
Seek medical assistance if necessary.

WARNING:
Dispose of contaminants used for cleaning as infectious waste.

IMPORTANT:

Staining of the incubator, spotting parts, and electrolyte parts with samples such as blood may af-
fect the measured values.

6-6 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance

6.3.1 Reference Plate Level Check


Perform [Reference plate level check] before cleaning.

1. Tap the button.

2. Tap the [Normal mode] button.

897N203029A FDC NX600 Instruction Manual 2022.01 6-7


6 Maintenance

3. Tap the [Reference plate level check]


button to run a check.
If check ends, the Check results screen appears.
4. Check the results of [Reference plate
level check].
When has appeared in the check results,
cleaning is necessary. Tap the [OK] button. Using
the procedure on the following pages, clean the
transfer bar, incubator, sample application unit,
and ISE unit.

In the case of Level 1 In the case of Level 2 In the case of Level 3


soiling (0-49%) soiling (50-79%) soiling (80% or higher)

6-8 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance

6.3.2 Cleaning the Transfer Bar, Incubator, Sample Application Unit, and
ISE Unit
(1) Cleaning the transfer bar

1. Tap the [Cleaning mode] button in Normal


mode.

2. Remove the slide cartridge and the tip 6


rack from the analyzer.
3. Tap the [Start Cleaning] button.
NOTE Keep your fingers clear as the analyzer
runs.

897N203029A FDC NX600 Instruction Manual 2022.01 6-9


6 Maintenance

4. The screen to the left appears.

5. Loosen the 2 thumbscrews and pull the


sample application unit cover frontward
to remove it.

Thumbscrew

6. Remove the two slide transfer weights.


NOTE There are 2 types of slide transfer
weight, one each for CM tests and ISE
tests.
The slide transfer weight for CM tests has hooks
on the left, so lift up from the right before remov-
ing it. When installing it, insert the hooks before
loading it.
For CM

Slide transfer weight For ISE

7. Use gauze or a cotton swab that has


been soaked in lukewarm or cold water
to wipe the staining of the removed slide
transfer weight.

6-10 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance

Transfer bar 8. Use gauze or a cotton swab that has


been soaked in lukewarm or cold water
to wipe the upper surface of the transfer
bar visible in the sample application unit.

9. Next, clean the incubator, sample appli-


cation unit, and ISE test unit.
NOTE To quit the cleaning, attach the removed
slide transfer weight and sample ap-
plication unit cover at their original po-
sitions, and then tap the button to
return to the previous screen.

CAUTION:
6
Keep your fingers clear as the analyzer runs.

10. Tap the button to switch off the


power.

CAUTION:

Keep your fingers clear as the analyzer runs.

897N203029A FDC NX600 Instruction Manual 2022.01 6-11


6 Maintenance

WARNING:
Always unplug the power cord before the following procedure.

(2) Incubator

1. Open the sampler cover and panel unit.


Remove the panel unit pedestal.

Panel unit
pedestal

6 Connector 2. Unplug the connector of the incubator


cable (low end) by turning it to the left.

Incubator cover 3. Remove the incubator cover.

6-12 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance

2 4. Loosen the 5 screws on the upper sur-


face of the incubator, and then remove
5
4 the incubator.

3
1

Thumbscrew
5. Moisten the cloth with lukewarm water.

IMPORTANT:

Wring the cloth when it is soaked too much. If wa-


ter gets inside the machine, it will cause rust.

6. Place the removed incubator on a flat


surface with the pressure plates facing
up. 6
IMPORTANT:

Put on gloves before beginning wiping.

Incubator pressure plate 7. If dirt is noticeable or a transport error oc-


curs within the cleaning cycle, shorten the
cleaning cycle (to at least once a week).

897N203029A FDC NX600 Instruction Manual 2022.01 6-13


6 Maintenance

8. Wipe the tapered part of each pres-


Tapered part sure plate (the metallic luster taper part
around the black part).
NOTE Wipe the entire circumference as blood
stains may be difficult to see by eye.

9. Wipe the surface (black part) of each


Surface pressure plate.
NOTE Wipe gently to avoid scratching, as it
will affect the measurement values.

IMPORTANT:

6 When wiping, change the cloth surface for each


cell. Otherwise, dirt may be transferred.

IMPORTANT:

Do not use a cotton swab.


The components of the cotton swab may nega-
tively affect the measurement values.

IMPORTANT:

After cleaning, do not touch the pressure plate sur-


face with your fingers.
This may affect the measurement values.

10. Wipe the lower part (measurement cells)


in the same way.

Measurement cell

6-14 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance

11. Loosen the 2 screws fixing the reference


plate and remove the plate from the ana-
lyzer.

Thumbscrew
Reference
plate

Reference Reference 12. Turn over the reference plate and clean
black plate white plate the white plate and black plate with a dry
cotton swab.

Photometer 13. Clean the photometer head with a dry 6


cotton swab.

14. Load the reference plate.


Tighten the 2 thumbscrews evenly.

IMPORTANT:

Tighten the two thumbscrews securely. Otherwise,


the measurement results may be negatively af-
fected.

Thumbscrew
Reference
plate

897N203029A FDC NX600 Instruction Manual 2022.01 6-15


6 Maintenance

(3) Sample application unit

1. Make sure that the sample application


unit cover and two slide transfer weights
were removed for cleaning of the trans-
fer bar.

Sample application unit 2. Wipe the sample application unit with


gauze or a cotton swab moistened with
lukewarm or cold water.

(4) ISE unit probe


6 Make sure there is no dust on the six ISE probes.
When you clean the ISE probes, wipe off the dust
using a dry cotton swab.

NOTE Do not use solvents, such as methyl alcohol


or isopropyl alcohol.

Probes
CAUTION:

ISE probes are needle-shaped parts. Do not touch


them directly by hand. Be careful not to bend the
probes.

6-16 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance

6.3.3 Reassembling Parts and Ending Cleaning

2 1. Reassemble the incubator.


5 Align the arrow marks on the incubator and the
4 reference plate, and then tighten the 5 thumb-
screws evenly by tightening opposing thumb-
screws in numerical order (1 to 5).
3
1

IMPORTANT:

Tighten the five thumbscrews securely. Otherwise,


the measurement results may be negatively af-
fected.
Thumbscrew

4 Incubator cable 2. Reassemble each part as shown in the


3 Incubator cover illustration on the left.
1 Load the slide transfer weights.
2 Load the sample application unit cover
and tighten the thumbscrews.
3 Load the incubator cover.
4 While aligned with the groove, plug the
1 ‌Slide transfer connector of the incubator cable into the 6
5P‌ anel unit weight incubator and then lock it by turning it to
pedestal 2 Sample application unit the right.
Incubator cable 5 Load the panel unit pedestal.
Incubator cover
IMPORTANT:

Tighten the thumbscrews securely. Otherwise, the


measurement results may be negatively affected.

IMPORTANT:

Be sure to attach the incubator cover. Otherwise,


the measurement results may be negatively af-
fected.

NOTE When you insert the connector, make


sure not to twist or apply too much pres-
sure to the cable. If the cable is twisted
more than once, it may break.
3. Close the panel section to end cleaning.
Run [Reference plate level check] again, and
then check the results.
è Refer to Section 6.3.1
NOTE Wait until the temperature of the incuba-
tor is steady.

897N203029A FDC NX600 Instruction Manual 2022.01 6-17


6 Maintenance

6.4 Replacing Recording Paper


When red lines appear along the sides of the recording paper, the printer is nearly out of paper, so
replace the recording paper roll with a new one.

NOTE Use the recording paper specified for the FDC NX600.

1. Press the section indicated in the illus-


tration (1) to release the lock, and then
open the cover (2).
1

Press the
concave
section.
2. Remove the remaining recording paper.

6 3. Load the new recording paper.


NOTE If the paper replacement was performed
during a measurement, the measure-
ment results might be printed out.

NOTE Set the paper roll so the paper is pulled


from under the paper roll as shown below.
OK Incorrect

NOTE If the paper is loaded in reverse, no text


is printed on the thermal paper.

4. Pull the top end of the recording paper


out slightly, and then close the cover.
NOTE When setting a new recording paper,
pull the edge of paper out to where the
tape for fixing edge is affixed. Failure
to observe this precaution may cause
printing malfunction.

NOTE Close the cover securely until it clicks.


Otherwise, the paper feed may fail and
the test results will not be printed out.

6-18 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance

6.5 Replacing and Cleaning the Light Source Lamp


The average life of the lamp is about 1,000 hours. Replace the lamp if a lamp replacement error oc-
curs or the cumulative lighting time of the lamp approaches 1,000 hours.

CAUTION:

The light source lamp gets very hot. Before cleaning the light source unit or replacing the light
source lamp, turn off the power, wait at least 5 minutes, and then check the lamp has cooled.

1. Pull out the disposal box, and then re-


move the front bottom cover.

Front bottom cover


6
2. Loosen the thumbscrew as shown in the
left figure and remove the lamp unit.
3. Unplug the connector of the light source
1 lamp and remove the lamp unit.
Thumb-
3
screw
2

Lamp unit
Connector

897N203029A FDC NX600 Instruction Manual 2022.01 6-19


6 Maintenance

4. Wipe the light source unit lens using a


dry cotton swab.

Lens

5. Turn the thumbscrew of the lamp unit to


the left, remove the lamp, and then re-
place it with a new one.

Lamp base IMPORTANT:

Tighten the thumbscrews securely. Otherwise, the


measurement results may be negatively affected.
6 Thumbscrew
NOTE When you replace the lamp, hold it by
its base. Do not touch the glass surface
with bare hands.
IMPORTANT:

Secure so that metal


hooks hold from above
the lamp housing.

6-20 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance

Thumbscrew B 6. Insert the connector of the light source


lamp.
7. Insert the lamp unit into the analyzer
2
with thumbscrew B facing up. Tighten
3
thumbscrew A securely.
1

IMPORTANT:

Tighten the thumbscrew securely. Otherwise, the


Thumbscrew A measurement results may be negatively affected.

Connector

8. Attach the front bottom cover, load the


disposal box, and then switch on the
power of the FDC NX600.
NOTE Put the front bottom cover back horizon-
tally as it was.

IMPORTANT:

Be sure to load the disposal box. If the analyzer 6


is used without the box, the measurement results
may be negatively affected.

9. After starting the analyzer, go to [Lamp


management] in Normal mode, and then
reset [Lamp ON time].
è Refer to Section 9.2.3

897N203029A FDC NX600 Instruction Manual 2022.01 6-21


6 Maintenance

6.6 Maintenance of Sampler O-ring


The sampler nozzle O-ring wears over time. Perform weekly cleaning, monthly inspection, and an-
nual replacement.

WARNING:
When inspecting and replacing O-rings, always wear gloves, a lab coat, and safety goggles.
If you accidentally touch a contaminant, immediately rinse the contaminated area thoroughly un-
der running water, and then disinfect the area.
Seek medical assistance if necessary.

IMPORTANT:

If you do not inspect and replace the sampler O-ring, the sample application volume will be inac-
curate, which may negatively affect the measurement values.

6.6.1 Cleaning the Sampler O-ring


Clean the sampler O-ring once a week as follows.

1. Prepare a dry / damp gauze or cloth (eye-


6 glass wipes, lint-free cloth, etc.).

2. Switch off the analyzer before cleaning.

3. Rotate the sampler part toward you.

6-22 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance

4. While holding the arm with one hand,


wipe the O-ring while rotating the damp
cloth 3 to 4 turns around the circumfer-
ence. Next, wipe down from the top to
cover the entire circumference. Similarly,
wipe the other O-ring around the circum-
ference and down from the top.

5. Use a dry cloth to lightly wipe the mois-


ture from the O-ring portion. Remove
any foreign objects remaining on the
O-ring.

6. Perform a measurement and confirm that 6


no errors occur.

IMPORTANT:

Do not store the AUTO TIPS with the storage box


lid open. If AUTO TIPS contaminated with dust or
dirt are mounted to the nozzles, the dirt or dust
may adhere to the O-rings, causing dispensing/
dilution clogging errors or leak check failure.

IMPORTANT:

If dirt is noticeable or an error occurs, shorten the


cleaning cycle (once a week or more).
7. If an error occurs, inspect the sampler
O-ring (Refer to Section 6.6.2).

897N203029A FDC NX600 Instruction Manual 2022.01 6-23


6 Maintenance

6.6.2 Inspecting the Sampler O-ring

1. Tap the button.

2. Tap the [Normal mode] button.

3. Tap [Sampler leak check].

6-24 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance

4. Follow the on-screen instructions to pre-


pare for the leak check.

5. Tap the nozzle to be checked to select it,


and then run the leak check.
Once check ends, the Check results screen ap-
pears.

CAUTION:
6
Keep your fingers clear as the sampler unit
and consumable disk move.

6. Check the results of the [Sampler leak


check].
If the check results are [OK], it is not necessary
to replace the O-ring.

NOTE The check results can be printed by


tapping the button.

897N203029A FDC NX600 Instruction Manual 2022.01 6-25


6 Maintenance

If the check results are [NG], tap the [Replace


O-ring] button, and then replace the O-ring
according to the following procedure.

NOTE The check results can be printed by


tapping the button.

7. Once the check ends, open the sampler


cover, and then remove the sampler leak
check tool mounted on the sampler noz-
zle by hand.

6-26 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance

6.6.3 Exchanging the Sampler O-ring


The sampler nozzle O-ring wears over time. Replace it once a year.

WARNING:
Process used O-rings as infectious waste.

IMPORTANT:

If you do not replace the O-ring periodically, the sample application volume will be inaccurate,
which may negatively affect the measurement values.

IMPORTANT:

Be sure to use the O-ring designed for the FDC NX600. Otherwise, the sample application volume
will be inaccurate, and the measurement results may be negatively affected.

1. Remove the sample rack and slide car-


tridge.

2. Tap the button.


6

3. Tap the [Normal mode] button.

897N203029A FDC NX600 Instruction Manual 2022.01 6-27


6 Maintenance

4. Tap [Replace O-ring].

5. Tap the [START] button.


The sampler nozzles turn to the replacement
position.

6 CAUTION:

Keep your fingers away as the sampler unit


moves.

6. Follow the on-screen instructions to


replace the O-ring.
If this button is tapped, the following expla-
nation appears.

6-28 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance

7. Tap [Leak check] to check for leaks.


è Refer to Section 6.6.2

897N203029A FDC NX600 Instruction Manual 2022.01 6-29


6 Maintenance

6.7 Cleaning the Slide Reading Section


When a slide-detection error has occurred, dust and other soiling may be the cause. Clean the slide
reading section to avoid these errors.

WARNING:
Always wear gloves, a lab coat, and safety glasses while working.
If you accidentally touch a contaminant, immediately rinse the contaminated area thoroughly un-
der running water, and then disinfect the area.
Seek medical assistance if necessary.

WARNING:
Dispose of contaminants used for cleaning as infectious waste.

1. Switch off the analyzer power.


2. Open the sampler cover and panel unit.
3. Remove the slide cartridge.

6
RE

4. Using your finger, move the transfer bar


under the right cover.

RE

5. Air blow from the gap where the slide


cartridge is placed, as shown in the left
figure.

RE

6-30 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance

6.8 Cleaning and Replacing the Reference Fluid Cap Packing


Clean the reference fluid cap packing each time the reference fluid is replaced.
Replace the reference fluid cap packing once every three months.

IMPORTANT:

If you do not clean and replace the reference fluid cap packing, the reference fluid will be concen-
trated, which may negatively affect the measurement values.

1. If reference fluid is loaded, remove the


reference fluid tube and discard it.

2. Remove the reference fluid cap packing


as shown in the illustration on the left.

6
Reference
fluid cap
packing

3. Wipe off the reference fluid cap packing


using a cotton swab or gauze moistened
with lukewarm or cold water as shown in
the illustration on the left, and then allow
it to dry.
Cotton swab
NOTE Replace it with a new cap if stains can-
not be wiped off.

Reference fluid cap packing

897N203029A FDC NX600 Instruction Manual 2022.01 6-31


6 Maintenance

4. Set the reference fluid cap packing prop-


erly.
NOTE Do not touch the bottom side (reference
fluid side) of the reference fluid cap
packing.

Reference
fluid cap
packing

6-32 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance

6.9 PF Unit Maintenance

WARNING:
During maintenance, always wear gloves, a lab coat, and safety glasses.
If you accidentally touch a contaminant, immediately rinse the contaminated area thoroughly un-
der running water, and then disinfect the area.
Seek medical assistance if necessary.

WARNING:
Dispose of contaminants used for cleaning as infectious waste.

6.9.1 Cleaning the PF Unit (Sensor, Reference Plate, and Suction Pad)
Clean when errors occurred during PF filtering.

1. Tap the button.

2. Tap the [Normal mode] button.

897N203029A FDC NX600 Instruction Manual 2022.01 6-33


6 Maintenance

3. Tap [PF maintenance].

4. Tap the [Cleaning] button.

5. Remove the sample rack, mixing cup,


and tip rack.
6. Tap the [OK] button.

CAUTION:

Keep your fingers clear as the PF part moves.

6-34 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance

7. Clean according to the on-screen in-


structions.

8. Tap the [OK] button to end cleaning.

Surface of PF sensor PF Suction Pad 6


(glass surface) (black)

PF reference plate

9. Run the leak check.


è Refer to Section 6.9.2
NOTE If the check was [NG], exchange the PF
suction pad.
è Refer to Section 6.9.3

897N203029A FDC NX600 Instruction Manual 2022.01 6-35


6 Maintenance

6.9.2 Inspecting the PF Suction Pad


The PF suction pad must be inspected monthly and replaced every two years.

1. Touch the [PF Maintenance] button as in


the previous section.

2. Tap the [Leak check] button.


6

6-36 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance

3. Follow the on-screen instructions to load


the PF sample rack and PF leak check
tool to the sample position, and then tap
the [START] button to run the leak check.
If check ends, the Check results screen appears.

CAUTION:

Keep your fingers clear as the PF part moves.

4. Check the results of the [Leak check].


If the check results are [OK], it is not necessary
to clean or replace the pad.

NOTE The check results can be printed by tap-


ping the button. 6

If the check results are [NG], tap the [Cleaning]


button, clean the PF unit, and then run the leak
test again.
è Refer to Section 6.9.1
If this does not resolve the problem, tap the
[Procedure] button, and then check the assem-
blies of the PF pump tube or replace the tube,
referring to the displayed instructions.

NOTE If the [Result details] button is tapped,


the Advanced leak check results display
screen appears.

NOTE The check results can be printed by tap-


ping the button.

5. Once the check ends, remove the PF


sample rack and PF leak check tool.

897N203029A FDC NX600 Instruction Manual 2022.01 6-37


6 Maintenance

6.9.3 Replacing the PF Suction Pad


Replace a PF suction pad once every 2 years.

WARNING:
Process used PF Suction Pad as infectious waste in compliance with any applicable regulations in
your country.

1. Touch the [PF Maintenance] button as in


the previous section.

2. Tap the [Suction pad] button.


The PF sampler rotates forward and stops.

6-38 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance

3. Remove the sample rack, mixing cups,


and tip rack.

RE

Sample rack

Tip rack Mixing cups


6
4. Tap the button to switch off the
power.

897N203029A FDC NX600 Instruction Manual 2022.01 6-39


6 Maintenance

5. Turn the PF Suction Pad in the direction


shown and remove it while loosening.
6. Install the new PF Suction Pad by turn-
ing it in the opposite direction.
NOTE Be sure to tighten the PF Suction Pad
securely.
7. Switch on the analyzer.
8. Run the leak check.
If the check results are [NG], perform the re-
RE placement again. è Refer to Section 6.9.2

6-40 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance

6.9.4 Replacing the PF Pump Tube

WARNING:
Before the following procedure, always switch off the power of the analyzer, and unplug the power
cord.

Replace the PF pump tube once every 2 years. Replacement is also necessary if the PF has not
been used for one month or longer.

1. Remove the disposal box, and remove


the front bottom cover.

Front bottom cover


6
Cover (B) Screw (A) 2. Loosen the screw (A) and remove the
cover (B).

Holder (C) Tube (D) 3. Slide the holder (C) to the left and re-
move the tube (D) from the rotor.

897N203029A FDC NX600 Instruction Manual 2022.01 6-41


6 Maintenance

PF pump 4. While pushing the head of joint (G), pull


Joint (G) the joint (G) to the right to remove the
tube (D).

Red tube (D)

Pull off joint (G) to the right Joint (G)


while pushing this part to
the right.

Holder (C) Tube (H) 5. Slide the holder (C) to the left and load
the new tube (H) on to the rotor.
NOTE As shown in the diagram to the left,
align the base of the joint (E) with the
cover hole (F).

NOTE Set so that the two front marks face the


front.
6

Cover hole (F) Joint (E) 6. Reattach the cover (B) using the screw
(A).

Front Marks

Joint (E)
Cover hole (F)
When attaching the cover (B), make
sure that the hooks are hooked and
interlocked.

NOTE If attachment is incorrect, a “W0134 PF


sampling error” may occur.

7. Insert the tip of the red tube (H) in the


Red tube (H) Joint (G) joint (G) until the line.

Line

6-42 897N203029A FDC NX600 Instruction Manual 2022.01


6 Maintenance

8. Attach the front bottom cover.


NOTE Put the front bottom cover back horizon-
tally as it was.
9. Attach the disposal box.
10. After the replacement, run the leak
check.
If the check results are [NG], perform the re-
placement again. è Refer to Section 6.9.2

897N203029A FDC NX600 Instruction Manual 2022.01 6-43


7 Quality Control

IMPORTANT:

To maintain the accuracy of your test results, it is recommended to perform quality control on routine
bases by use of control fluids. Perform quality control in compliance with local regulation and proce-
dure which you have to obey. Also, it is also recommended to do after major repairs are performed on
the analyzer.

7.1 Control Fluids


Use the controls which are prepared for FUJI DRI-CHEM or which are recommended to use.

NOTE Follow the descriptions in the “Instructions for Use” of control fluids.

7.2 Measurement with Control Fluids


Measure with control fluids by following the same procedure as for measurement of patient samples.
Load the control fluid in the sample loading position of the sample rack, load a slide, and then meas-
ure.

7.3 Monitoring Results


Configure the target values and acceptable ranges for measurements from control fluid results, and
monitor for changes over time.

7.4 Control Mode


You can measure control liquid by using [Control] of the Function menu to configure the entered cor-
relation coefficients “a” and “b” to a = 1 and b = 0 temporarily. This state is called Control mode. 7
è Refer to Section 9.1
After using this mode, always cancel control mode by tapping [HOME] on the LCD.

7.5 Troubleshooting Quality Control Results


If there is a discrepancy in the control fluid measurement results, confirm the following points, and
then measure the control fluid again.
• Make sure that the periodic cleaning and replacement of consumables listed in this manual are
being performed.
• Make sure the handling methods, storage conditions, usage limitations, and other details listed in
the “Instructions for Use” of slides have been observed.
If the control results are still a problem, please contact your dealer.

897N203029A FDC NX600 Instruction Manual 2022.01 7-1


7 Quality Control

7.6 Control Chart

1. Tap the button.

2. Tap the [Control] button.

3. The measurement-preparation screen of


control mode appears.
4. Tap the (graph display) button.

7-2 897N203029A FDC NX600 Instruction Manual 2022.01


7 Quality Control

5. Tap the button of the measurement item


to be viewed.

6. The control chart appears.


NOTE If the control liquid button is tapped, you
can change the control liquid to be dis-
played.

NOTE Measurement results of outside of


measurement range (@), disposal box
7
open (*), and expired slide (#) are not
reflected in the control chart.

NOTE If the measurement date is touched and


selected, and the show/hide button is
touched, the plot of these measurement
results can be hidden (or shown).

Show/hide Acceptable Acceptable


button range range
lower limit upper limit
Measurement
result average

Outside of display
area

Control lot was


changed
Control lot was
changed
Outside of accept-
able range

897N203029A FDC NX600 Instruction Manual 2022.01 7-3


7 Quality Control

7. If a measurement date is selected and


then the [Detail] button is tapped, the de-
tail screen appears.

< Display of measurement results >


Green: when within the acceptable range of control
Red: when outside of the acceptable range of control

7
Acceptable Acceptable
range range
lower limit upper limit

NOTE The acceptable range of control displays


the value entered for Range setting for
control (Refer to Section 9.3.17).

NOTE In case of ALPi and LDHi slides, the


display of “Test name” will not follow the
“Test Item Name Setting”, but “ALPi-PS”
or “LDHi-PS”.
But, the test name of each measure-
ment result will follow the “Test Item
Name Setting”.
è Refer to Section 9.3.20.

7-4 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

WARNING:

When you troubleshoot, always wear gloves, a lab coat, and safety goggles.
If you accidentally touch a contaminant, immediately rinse the contaminated area thoroughly under
running water, and then disinfect the area.
Seek medical assistance if necessary.

IMPORTANT:

If an analyzer malfunction (error) appears before or during measurement or warnings appear in the
measurement results, the measurement results may be inaccurate. Refer to the related troubleshoot-
ing pages and rerun the measurement.

8.1 Error Display


If an error occurs, an error message appears on the LCD.

< Example of display >

Error code

Content of error message

Additional information
8
NOTE Some error codes may show additional infor-
mation.

(Example)
1050 (GLU-P slide code)
120001 (Lot number)
*Refer to Section 8.3 for slide codes.

NOTE The error messages listed in this document and the descriptions displayed on the LCD may
differ.

897N203029A FDC NX600 Instruction Manual 2022.01 8-1


8 Troubleshooting

8.1.1 Error Log


You can view the history of errors that occurred by using [Error log] in the function menu.

1. Tap the button.

2. Tap the [Error log] button.


An error log list is displayed.

8
3. Tap an error.
The error details are displayed.

8-2 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

897N203029A FDC NX600 Instruction Manual 2022.01 8-3


8 Troubleshooting

8.1.2 Error Code and Reference Section Table


Error code Error description Reference Reference
section page
E0021 The disposal box was open during measurement 8.2.15 8-42
E0035 Photometric control system error 8.2.8 8-24
E0050 Light source lamp is burned out 8.2.5 (2) 8-20
E0080 Tip detection sensor error 8.2.15 8-42
E0082-E0085 Tip mounting error 8.2.4 (1) 8-11
E0086, E0087 Tip disposal error 8.2.4 (2) 8-12
E0111, E0112 Clogging error during sampling of sample 8.2.4 (5) 8-14
E0121, E0122 Clogging error during sampling of diluent 8.2.4 (8) 8-17
E0124 Clogging error during sampling of diluted sample 8.2.4 (5) 8-14
E0125 Clogging error during sampling of diluted sample 8.2.4 (5) 8-14
E0127, E0128 Clogging error during sampling of reference fluid 8.2.4 (8) 8-17
E0132, E0133, PF errors 8.2.12 8-36
E0150
E0154 Mixing cup error 8.2.4 (3) 8-13
E0160 Incorrect diluent type error 8.2.10 8-32
E0200-E0203, Slide transfer error 8.2.6 8-21
E0210-E0213,
E0220-E0227
E0300, E0301, Reference white plate measurement error during 8.2.8 8-24
E0302 initial operation
E0509, E0510 Signal system error inside circuit board 8.2.8 8-24
E0511 Light intensity error 8.2.5 (1) 8-19
8 E0515, E0516,
E0517
Signal system error inside circuit board 8.2.8 8-24

E0530, E0531 ISE measurement unit (self-test) error 8.2.11 (1) 8-33
E0532, E0533, Signal system error inside circuit board 8.2.8 8-24
E0537, E0538
E0534, E0535, Signal system error inside circuit board 8.2.8 8-24
E0536, E0539,
E0540
E0565 ISE temperature control error 8.2.7 8-22
E0600-E0604 Incubator motor error 8.2.6 8-21
E0700-E0703 ISE measurement unit motor error 8.2.6 8-21
E0900-E0903 Interference filter motor error 8.2.5 (3) 8-20
E1000-E1003 Sampler vertical operation error 8.2.4 (9) 8-18
E1100-E1103 Sampler rotation error 8.2.4 (9) 8-18
E1200-E1203 Syringe operation error 8.2.4 (9) 8-18
E1300-E1304 Errors related to PF 8.2.12 8-36
E1400-E1403 Sample application unit feed error 8.2.6 8-21
E1500, E1501, Temperature control error 8.2.7 8-22
E1505

8-4 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

Error code Error description Reference Reference


section page
E1502, E1503, Temperature control error 8.2.7 8-23
E1504
E1510, E1511, ISE temperature control error 8.2.7 8-22
E1512, E1514,
E1515
E1520, E1521 Installation environment temperature error 8.2.7 8-22
E1530, E1531, Signal system error inside circuit board 8.2.8 8-25
E1532
E1600-E1603 Consumable disk operation error 8.2.6 8-21
E1750-E1755, Slide reading control system error 8.2.8 8-25
E1757
E3900 Signal system error inside circuit board 8.2.8 8-25
E4200, E4201 Calibration error 8.2.14 (3), 8-39
(4)
E4202, E4203 Calibration error 8.2.14 (5) 8-40
E4204 Calibration error 8.2.14 (6) 8-41
E4205 Calibration error 8.2.14 (5) 8-40
E4401, E4403 Signal system error inside circuit board 8.2.8 8-25
E4410 Measurement result reference error 8.2.8 8-25
E4411 Control chart reference error 8.2.8 8-25
E4454, E4475, Signal system error inside circuit board 8.2.8 8-26
E4478-E4481,
E4490, E4600-
E4613
E5101, E5102, Printer error 8.2.2 (3) 8-10
E5103 8
E5500 Signal system error inside circuit board 8.2.8 8-28
E5510 Slide position offset during measurement 8.2.5 (4) 8-20
E5901 Signal system error inside circuit board 8.2.8 8-28
E6200, E6201, Communication error 8.2.13 (1) 8-37
E6210

897N203029A FDC NX600 Instruction Manual 2022.01 8-5


8 Troubleshooting

Error code Error description Reference Reference


section page
W0020 The disposal box is open 8.2.15 8-42
W0022 The disposal box is full 8.2.15 8-42
W0030 Insufficient light intensity 8.2.5 (1) 8-19
W0036 Reference black plate soiling 8.2.5 (1) 8-19
W0040, W0041 The sample type configuration and sample type 8.2.10 8-31
of the slide do not match
W0042, W0043 Sample rack configuration error 8.2.10 8-31
W0060 Insufficient sample or calibrator volume 8.2.4 (4) 8-13
W0070 Sampler rotation error 8.2.4 (9) 8-18
W0088, W0089, Sample tip detection error 8.2.4 (1) 8-11
W0096
W0090 The ISE slide is loaded in reverse 8.2.10 8-31
W0091, W0092 ISE measurement error 8.2.11 (3) 8-35
W0120 Sample surface cannot be detected 8.2.4 (4) 8-13
W0123, W0129 Liquid volume error 8.2.4 (7) 8-16
W0134, W0138 PF errors 8.2.12 8-36
W0137 PF card has not been read 8.2.9 (3) 8-29
W0140 Liquid volume error 8.2.4 (4) 8-13
W0141-W0144 Liquid volume error 8.2.4 (7) 8-16
W0151, W0152 Mixing cup error 8.2.4 (3) 8-13
W0170 QC card has not been read 8.2.9 (2) 8-29
W0173 The slide information has not been read 8.2.3 8-11
W0175 No slides have been set 8.2.10 8-31
8 W0180 Calibration has not been performed with an im-
muno-chemical test
8.2.14 (1) 8-39

W0190, W0191 Calibration error 8.2.14 (2) 8-39


W0214 Slide transfer error 8.2.6 8-21
W0500-W0503 Malfunction of the internal fan 8.2.15 8-42
W0704 ISE measurement unit motor error 8.2.6 8-21
W0904 Interference filter motor error 8.2.5 (3) 8-20
W1004 Sampler vertical operation error 8.2.4 (9) 8-18
W1104 Sampler rotation error 8.2.4 (9) 8-18
W1204 Syringe operation error 8.2.4 (9) 8-18
W1308 PF errors 8.2.12 8-36
W1404 Sample application unit feed error 8.2.6 8-21
W1604 Consumable disk operation error 8.2.6 8-21
W1700 Malfunction of the QC card reader 8.2.9 (5) 8-30
W1703 QC card read ERR 8.2.9 (5) 8-30
W4010 QC type has not been read 8.2.9 (2) 8-29
W4011 Panel card error 8.2.9 (5) 8-30
W4012 DI card error 8.2.9 (4) 8-30

8-6 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

Error code Error description Reference Reference


section page
W4013, W4014, QC card error 8.2.9 (5) 8-30
W4015
W4440 Setting value reset warning 8.2.15 8-43
W4460-W4475, Signal system error inside circuit board 8.2.8 8-26
W4477, W4483
W4500, W4501, Serial communication error 8.2.13 (1) 8-37
W4502
W4510 LAN communication error 8.2.13 (1) 8-37
W4511 Type 4 LAN communication error 8.2.13 (1) 8-37
W4520 Error related to work list 8.2.13 (1) 8-37
W4720 Clock time error 8.2.8 8-27
W4721, W4733 Signal system error inside circuit board 8.2.8 8-28
W5000, W5020 Signal system error inside circuit board 8.2.8 8-28
W5002-W5019 USB memory read error 8.2.8 8-28
W5050 USB memory has not been inserted 8.2.8 8-28
W5100 Printer error 8.2.2 (3) 8-10
W5110 No recording paper 8.2.2 (3) 8-10
W5330-W5332 Sample barcode reader error 8.2.13 (2) 8-38
W5400 Expired slide 8.2.10 8-32
W66xx Input value waring 8.2.15 8-43
I5920 Reboot warning 8.2.15 8-42

897N203029A FDC NX600 Instruction Manual 2022.01 8-7


8 Troubleshooting

8.1.3 Printout Reference Table


Printout Meaning and conditions Troubleshooting
H Measurement value is higher than the upper limit of the
-
preset reference interval.
L Measurement value is lower than the lower limit of the
-
preset reference interval.
> Measurement value is higher the upper limit of the meas-
-
urement range.
< Measurement value is lower than the lower limit of the
-
measurement range.
@ The measurement values are reference values outside
the measurement range.
-
IMPORTANT:

Measurement values might be incorrect.


+ OR Measurement value of the ISE test is higher than the up-
Refer to Section 8.2.11
per limit of the measurement range.
- OR Measurement value of the ISE test is lower than the lower
Refer to Section 8.2.11
limit of the measurement range.
+, -, $ Temperature control error during measurement. Refer to Section 8.2.7
*, ? Light intensity is fluctuating during measurement. Refer to Section 8.2.5
E Unable to calculate the measurement result because Check whether other errors
measurement was interrupted due to a transfer problem have occurred, and then
or other error. perform the necessary trou-
The measurement results are printed as [****]. bleshooting.
& Abnormally high measurement value. The glucose value Rerun the measurement.
of the sample is high in a CRP measurement. In the case Perform a dilution meas-
8 of a Ca-P slide, the sample has not been applied. urement by following the
“Instructions for Use” of
the slides.

8-8 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

# The valid term of the slide has expired.


For immuno-chemical tests, the validity term of the slide or
the validity term of calibration has expired.

IMPORTANT:

The measurement value might be incorrect.

IMPORTANT:

Expired slides cannot be used. Always use valid


slides.
Use slides within their
IMPORTANT: validity terms.
Analyzer will not show # whose means that ISE
slides have been expired if the installed software is
version 1.0.
Otherwise, with the software version 1.1 or later
is installed, analyzer will show # for the ISE slides
that expire 2023 or later or of lot number 130000 to
139999.
NOTE Depending on the device settings, there may be
cases in which # is not displayed, and an expired
slide cannot be measured.
¥ Problem with sample application on slide (not applied) Rerun the measurement.

(Supplement) Measurement range


The measurement range (determination range) and the reference interval differ by measure-
ment item. Their relation is shown below. For details, refer to the “Instructions for Use” of the
slides. 8
NOTE You can set the values of the reference interval by using [Reference interval range set-
ting] in Mode function for administrator. (Refer to Section 9.3.6)
High

Reference Measurement range


Measured value
interval (Determination range)

Low

897N203029A FDC NX600 Instruction Manual 2022.01 8-9


8 Troubleshooting

8.2 Troubleshooting
8.2.1 Startup Errors
(1) The analyzer does not start after the power button is pressed.
Check that the power cable is properly connected. And, turn the power switch OFF and ON
again, and press the power button.
If the analyzer still does not start, please contact your dealer for assistance.

8.2.2 Printer or LCD Trouble


(1) Printed characters are too faint or dark
Please contact your dealer for assistance.

(2) White lines appear on printed characters or certain parts of characters are
missing.
The printer head has malfunctioned. Please contact your dealer for assistance.

(3) Printer errors

Error
Error message Error condition Troubleshooting
code

W5110 Printer error No recording paper. Load recording paper onto the printer.
No paper. è Refer to Section 6.4
Set paper.

W5100, Printer error The printer is malfunc- Please contact your dealer for assis-
E5101, Contact your dealer in tioning. tance.
E5102, case the ERR repeats.
8 E5103

(4) LCD is too faint or dark


Adjust the brightness using [Brightness • Volume] in Normal mode.
è Refer to Section 9.2.2

8-10 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

8.2.3 Slide Reading Errors

Error
Error message Error condition Troubleshooting
code

W0173 Slide identification The slide information Clean the slide reader, and then rerun
ERR on backside of the slide the measurement.
No slide information. could not be read cor- è Refer to Section 6.7
Check the slide. rectly. NOTE If the error occurs again, please
contact your dealer for assis-
tance.

8.2.4 Sampler Errors


(1) When tip cannot be mounted

Error
Error message Error condition Troubleshooting
code

E0082, Sample tip detection No tip can be detected The tip is not loaded correctly. Load
W0088 ERR on the sample nozzle a tip correctly, and then rerun the test
No tip detected. after mounting. from the beginning.
Set tips and rerun the è Refer to Section 5.2
tests from the begin- NOTE Set tips into all holes of the tip
ning. rack after this error has oc-
curred.
E0083, Reference fluid tip No tip has been de-
W0089 detection ERR tected on the refer-
No tip detected. ence fluid nozzle after
mounting.
Set tips and rerun the
tests from the begin- 8
ning.

E0084 Sample tip clogging A clog was detected If the error occurs again, please contact
ERR during mounting of the your dealer for assistance.
Rerun the tests from sample nozzle tip.
the beginning.
Contact your dealer in
case the ERR repeats.

E0085 Reference fluid tip A clog was detected


clogging ERR during mounting of the
Rerun the tests from reference fluid nozzle
the beginning. tip.
Contact your dealer in
case the ERR repeats.

W0096 No tip WARNING The number of tips Load tips in all holes of the tip rack.
Tip is insufficient. needed for measure-
Set tips and rerun the ment is insufficient.
test from the begin-
ning.

897N203029A FDC NX600 Instruction Manual 2022.01 8-11


8 Troubleshooting

(2) When tip cannot be ejected

Error
Error message Error condition Troubleshooting
code

E0086 Sample tip eject ERR Tips cannot be dis- After analyzer shutdown, empty dis-
After analyzer shut- carded. posal box, and then turn the power ON.
down, please check
the disposal box.
Turn the power ON.

E0087 Reference fluid tip


eject ERR
After analyzer shut-
down, please check
the disposal box.
Turn the power ON.

8-12 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

(3) When there are insufficient number of or no mixing cups

Error
Error message Error condition Troubleshooting
code

W0151 Lack of mixing cup There are not enough After you replace the mixing cups, re-
Lack of mixing cups. mixing cups. run from the measurement request.
Set mixing cup and re- è Refer to Section 5.3
run the tests from the
beginning.

W0152 No mixing cup


Set mixing cup and re-
run the tests from the
beginning.

E0154 Mixing cup detection Error of the mixing cup If the error still occurs after switching
sensor ERR detection sensor. the analyzer off and on, please contact
Contact your dealer. your dealer for assistance.

(4) During a sample volume error

Error
Error message Error condition Troubleshooting
code

W0120 Sample shortage Either the sample is Make sure that the sample volume in
Sample is insufficient. not loaded or the sam- the tube is within the measurable range
Check the sample. ple volume is insuffi- and an appropriate sample tube is used
cient. in the sample rack, and then rerun from
the measurement request.
W0060 Sample shortage There is not enough
Check the sample vol- sample volume.
NOTE If the sample volume is low, add
sample to the tube or perform
8
ume.
manual sample application.
Confirm that a sam-
ple rack matching the
sample tubes is used.

W0140 S u r f a c e d e t e c t i o n There is too much


ERR (sample) sample volume.
Sample volume is too
much.
Check the sample vol-
ume.

897N203029A FDC NX600 Instruction Manual 2022.01 8-13


8 Troubleshooting

(5) During a sample aspiration or application error

Error
Error message Error condition Troubleshooting
code

E0111 Aspiration clogging Clogging was detected The sample may contain fibrin deposits.
ERR (sample) during sample aspira- After checking the sample, rerun from
Clogging was detected tion. the measurement request.
during sampling. NOTE If the error occurs again, there
Check the sample. might be a piping system prob-
lem. Please contact your dealer
E0124 Aspiration clogging
for assistance.
ERR (diluted sample)
Clogging was detected
during sampling.
Check the sample and
diluent.

E0112 S p o t t i n g c l o g g i n g A clog was detected


ERR (sample) during sample applica-
Clogging was detected tion (dispensing).
during sampling.
Check the sample.

E0125 S p o t t i n g c l o g g i n g A clog was detected The sample may contain fibrin deposits.
ERR (diluted sample) during sample applica- After checking the sample, rerun from
Clogging was detected tion (dispensing). the measurement request.
during sampling. NOTE If the error occurs again, there
Check the sample and might be a piping system prob-
diluent. lem. Please contact your dealer
for assistance.
8

8-14 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

(6) When application is poor

Error
Error message Error condition Troubleshooting
code

Sample application problem Because the sample is Check the following points.
not applied to the slide, (a) Make sure that FUJI DRI-CHEM
the error mark [<] or [@] AUTO TIPS are being used.
appears continuously. (b) Make sure that tips are not being
reused.
IMPORTANT: (c) Make sure that there are no bub-
bles on the sample surface in the
Measurement values sample tube.
with an [@] mark
(d) Make sure that a sample rack that
might be incorrect.
matches the blood collection tube
Run the measure-
or FUJI tube is being used.
ment again.
(e) Inspect the sampler O-ring. (Refer
to Section 6.6.2)

IMPORTANT:

FUJI DRI-CHEM AUTO TIPS cannot


be reused.
NOTE If errors still occur often, please
contact your dealer for assis-
tance.

897N203029A FDC NX600 Instruction Manual 2022.01 8-15


8 Troubleshooting

(7) During a diluent/reference fluid volume error

Error
Error message Error condition Troubleshooting
code

W0141 Surface detection ERR There is too much dilu- Follow the instructions of the error code
(diluent) ent volume. to replace the diluent or reference fluid,
Diluent volume is too and then rerun from measurement re-
much. quest.
Check the diluent vol- • Diluent è Refer to Section 5.4
ume. • Reference fluid è Refer to Section
5.5
W0142 Surface detection ERR There is too much ref-
NOTE Replace the diluent / reference
(reference fluid) erence fluid volume.
fluid container with a new one.
Reference fluid is too
Do not add diluent or reference
much.
fluid.
Check the reference
fluid volume.

W0123 Diluent is insufficient. There is not enough


Diluent is insufficient. diluent.
Check the diluent.

W0143 Diluent is insufficient.


Diluent is insufficient.
Check the diluent vol-
ume.

W0129 Insufficient reference There is not enough


fluid reference fluid.
Reference fluid is in-
sufficient.
8 Check the reference
fluid.

W0144 Insufficient reference


fluid
Reference fluid is in-
sufficient.
Check the reference
fluid volume.

8-16 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

(8) During a diluent / reference fluid error

Error
Error message Error condition Troubleshooting
code

E0121 Aspiration clogging Clogging was detected Be sure that FUJI tubes configured in
ERR (diluent) during diluent aspira- Tube setting for diluent and reference
Clogging was detected tion. fluid (è Refer to Section 9.3.15) are
during sampling. used. If there was a loading problem,
Check the diluent. replace the diluent and reference fluid,
and then rerun from the measurement
E0122 Spotting clogging ERR Clogging was detected request.
(diluent) during diluent applica- NOTE If the error occurs again, there
Clogging was detected tion. might be a piping system prob-
during sampling. lem. Please contact your dealer
Check the diluent. for assistance.
E0127 Aspiration clogging Clogging was detected
ERR (reference fluid) during reference fluid
Clogging was detected aspiration.
during sampling.
Check the reference
fluid.

E0128 Spotting clogging ERR Clogging was detected


(reference fluid) during reference fluid
Clogging was detected application.
during sampling.
Check the reference
fluid.

897N203029A FDC NX600 Instruction Manual 2022.01 8-17


8 Troubleshooting

(9) During a sampler drive system error

Error
Error message Error condition Troubleshooting
code

W0070 S p o t t i n g p o s i t i o n Rotation error of the Turn the analyzer off after the meas-
ERR sampler has occurred. urement, and make sure that foreign
Sampler rotation error. matters such as tips and slides are not
stuck or left inside of the sampler.
E1000, S a m p l e r u p / d o w n Up-and-down move- NOTE If the error still occurs after
E1001, motor ERR ment error of the motor
switching the analyzer off and
E1003, Problem moving the
on, please contact your dealer
W1004 sampler up and down.
for assistance.
Perform inspection.
Contact your dealer.

E1002 Sampler up/down


motor operation ERR
Problem moving the
sampler up and down.
Perform inspection.
Contact your dealer.

E1100- Sampler movement Rotation error of the


E1103, motor ERR sampler
W1104 Problem with sampler
movement.
Perform inspection.
Contact your dealer.

E1200, Syringe motor ERR Movement error of the


E1201, Problem with the sy- syringe motor
8 E1203, ringe.
W1204 Contact your dealer.

E1202 Syringe motor opera-


tion ERR
Problem with the sy-
ringe.
Contact your dealer.

8-18 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

8.2.5 Photometric System Errors

IMPORTANT:

If a photometric system error has occurred, perform each troubleshooting step listed below and
rerun the measurement.

(1) During a fluctuation of light source intensity

Error
Error message Error condition Troubleshooting
code

W0036 Soiled black plate The reference plate is Switch off the analyzer and clean each
WARNING dirty. part listed below.
Reference black plate • Clean the light source unit lens
soiled.
Clean the black plate. è Refer to Section 6.5
• Clean the reference plate
W0030 Lamp replacement The light intensity of
WARNING the light source lamp is è Refer to Section 6.3
Low light intensity. insufficient. • Clean the photometer
Clean photometer.
è Refer to Section 6.3
After cleaning, switch on the power.
If the errors still occur after cleaning,
replace the light source lamp.
è Refer to Section 6.5
NOTE If the error occurs again, please
contact your dealer for assis-
tance. 8
E0511 Reference white plate The light intensity is
ERR changing during meas- CAUTION:
Contact your dealer. urement. The light source lamp gets very hot.
Before cleaning the light source unit
or replacing the light source lamp, turn
off the power, wait at least 5 minutes,
and then check the lamp has cooled.

Error marks “ * ” or “ ? ” are printed


on the results

897N203029A FDC NX600 Instruction Manual 2022.01 8-19


8 Troubleshooting

(2) When lamp is burned out

Error
Error message Error condition Troubleshooting
code

E0050 Lamp replacement The light source lamp Switch off the power and replace the
Low light intensity. is burned out. light source lamp. Switch the analyzer
Replace the lamp. on again after replacement.
è Section 6.5

CAUTION:

The light source lamp gets very hot.


Before cleaning the light source unit
or replacing the light source lamp, turn
off the power, wait at least 5 minutes,
and then check the lamp has cooled.

(3) During an interference filter motor error

Error
Error message Error condition Troubleshooting
code

E0900, Filter motor ERR Movement error of the Switch the analyzer off and on.
E0901, Problem with the inter- filter motor NOTE When the error is not resolved,
E0903, ference filter. please contact your dealer for
W0904 Contact your dealer. assistance.
E0902 Filter motor opera-
tion ERR
8 Problem with the inter-
ference filter.
Contact your dealer.

(4) Other errors

Error
Error message Error condition Troubleshooting
code

E5510 Slide slip ERR The slide stop position Contact your dealer and have the incu-
Contact your dealer. shifted during meas- bator pressure plate inspected.
urement.

8-20 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

8.2.6 Transfer Errors

Error
Error message Error condition Troubleshooting
code

E0200- Feed motor ERR Movement error of the The sample application unit or the incu-
E0203, Problem with the slide slide supply motor bator may be dirty or the disposal box
E0220- transfer system. may be full of slides and tips. Switch off
E0227 Perform inspection. the analyzer, and perform the following.
Contact your dealer. • Cleaning the disposal box
• Cleaning the incubator
E0210- Eject motor ERR è Refer to Section 6.3
E0213, Problem with the slide
W0214 transfer system. • Cleaning the sample application unit
Perform inspection. è Refer to Section 6.3
Contact your dealer. NOTE If the error occurs again, please
E0600- Incubator motor ERR Movement error of the contact your dealer for assis-
E0604 Problem with the incu- incubator motor of the tance.
bator. incubator
Perform inspection.
Contact your dealer.

E0700, Probe motor ERR Movement error of the


E0701, Problem with ISE. probe motor of the ISE
E0703, Perform inspection. unit
W0704 Contact your dealer.

E0702 Probe motor opera-


tion ERR
Problem with ISE.
Perform inspection. 8
Contact your dealer.

E1400- Transfer change motor Movement error of the


E1403, ERR ISE transfer motor of
W1404 Problem with the slide the sample application
transfer system. unit
Contact your dealer.

E1600- C o n s u m a b l e d i s k The disk motor move- Turn the analyzer off after the meas-
E1603, motor ERR ment error in consum- urement, and make sure that foreign
W1604 Problem with the con- able disk motor matters such as tips and slides are not
sumable disk. stuck or left around the consumable
Contact your dealer. disk.
NOTE If the error occurs again, please
contact your dealer for assis-
tance.

897N203029A FDC NX600 Instruction Manual 2022.01 8-21


8 Troubleshooting

8.2.7 Temperature Controller Errors

Error
Error message Error condition Troubleshooting
code

E0565 ISE temperature con- When the measure- Please contact your dealer for assis-
trol ERR ment results are ap- tance.
Problem with ISE. pended with a “+” or “-”
Contact your dealer. mark or an error listed
to the left occurs, the
E1500 Temperature control temperature inside the Perform the following troubleshooting.
ERR incubator is outside the (a) Check the room temperature is
Check the ambient specified range. within 15 to 32 °C. If the tempera-
temperature. ture is out of the range, control the
Clean the air filter. room temperature.
Reboot the analyzer. (b) Switch off the power.
(c) Clean the air filter.
E1501, Temperature control
è Refer to Section 6.2
E1505 ERR
(d) Make sure that thumbscrews of the
Check the ambient incubator are tightened properly
temperature. and the incubator cable is connect-
Reboot the analyzer. ed to the connector properly.
E1510 ISE temperature con- èRefer to Section 6.3.3
trol ERR (e) Make sure that the sample ap-
Check the ambient plication unit cover is attached to
temperature. the analyzer properly. Tighten the
thumbscrews securely.
Clean the air filter.
(f) Leave the analyzer for about 10
Reboot the analyzer.
minutes to allow the interior to cool,
E1511, Temperature control and then switch on power.
8 E1515 ERR NOTE If the error occurs again, please
Check the ambient contact your dealer for assis-
temperature. tance.
Reboot the analyzer.

E1520, Room temperature


E1521 WARNING
Check the ambient
temperature.
Reboot the analyzer.

8-22 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

Error
Error message Error condition Troubleshooting
code

E1502 Temperature control When the measure- Switch off the power and connect the
ERR ment results are ap- incubator cable again.
Incubator temperature pended with a “$” mark NOTE After inserting the connector of
sensor disconnected. or an error listed to the the incubator cable, you must
Check the incubator left occurs, the incuba- turn it to the right to lock it.
cable. Contact your tor cable either is not
è Refer to Section 6.3.3
dealer. connected or has been
severed. NOTE If the error occurs again, please
E1503 Temperature control contact your dealer for assis-
ERR tance.
Incubator temperature
sensor damaged.
Contact your dealer.

E1504 Temperature control


ERR
Incubator heater dis-
connected.
Check the incubator
cable. Contact your
dealer.

E1512 ISE temperature con- Disconnection error of Please contact your dealer for assis-
trol ERR the ISE incubator tance.
ISE temperature con-
trol element damaged.
Contact your dealer.

E1514 ISE temperature con-


trol ERR 8
ISE heater disconnect-
ed.
Contact your dealer.

897N203029A FDC NX600 Instruction Manual 2022.01 8-23


8 Troubleshooting

8.2.8 Circuit Board Signal Errors

Error
Error message Error condition Troubleshooting
code

E0035 Light too intense A circuit board signal If the error still occurs after switching
Light too intense. error has occurred. the analyzer off and on, please contact
Contact your dealer in your dealer for assistance.
case the ERR repeats.

E0300, Gain setting ERR


E0301, Contact your dealer in
E0302 case the ERR repeats.

E0509 AD conversion cali- Please contact your dealer for assis-


bration ERR tance.
Contact your dealer.

E0510 Faulty A/D controller


Contact your dealer.

E0515 DA conversion ini-


tialization ERR
Contact your dealer.

E0516 AD conversion reset


ERR
Contact your dealer.

E0517 AD conversion ini-


tialization ERR
Contact your dealer.
8 E0532, ISE self test AD con- When an error occurs after turning the
E0533, version ERR power on and off, please contact your
E0537, For ISE measurement, dealer for assistance. Tests other than
E0538 turn the power SW ISE do not require a restart.
OFF and ON.
Contact your dealer in
case the ERR repeats.
Other parameters
besides ISE can be
measured.

E0534, ISE AD conversion


E0535, ERR
E0536, For ISE measurement,
E0539 turn the power SW
OFF and ON.
Contact your dealer in
case the ERR repeats.
Other parameters
besides ISE can be
measured.

8-24 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

Error
Error message Error condition Troubleshooting
code

E0540 ISE AD conversion A circuit board signal When an error occurs after turning the
initialization ERR error has occurred. power on and off, please contact your
For ISE measurement, dealer for assistance. Tests other than
turn the power SW ISE do not require a restart.
OFF and ON.
Contact your dealer in
case the ERR repeats.
Other parameters
besides ISE can be
measured.

E1530, Temperature AD con- Please contact your dealer for assis-


E1531, version ERR tance.
E1532 Contact your dealer.

E1750- Slide reading unit If the error still occurs after switching
E1755, failure the analyzer off and on, please contact
E1757 Contact your dealer. your dealer for assistance.

E3900 Internal communica-


tion ERR
Contact your dealer.

E4401, Save ERR


E4403 Contact your dealer.

E4410 Measurement result This measurement result cannot be


reference error referred.
The number of meas-
urement results, with 8
the new result, current-
ly exceeds capacity.
The selected measure-
ment result was de-
leted.

E4411 Control chart refer- Please go back to the parameter selec-


ence error tion screen and display the chart again.
The selected chart is
not up-to-date with the
new measurement.
Please go back to the
parameter selection
screen and display the
chart again.

897N203029A FDC NX600 Instruction Manual 2022.01 8-25


8 Troubleshooting

Error
Error message Error condition Troubleshooting
code

E4454 Memory loading ERR A circuit board signal Please contact your dealer for assis-
Contact your dealer. error has occurred. tance.

W4460 Memory recovery


W4461 WARNING (Example:
W4464 setting info.)
W4469 Recovery performed
W4470 from backup memory.
W4473 Contact your dealer.
W4474

W4462 Memory recovery Check whether the setting for reference


W4463 WARNING intervals is correct.
Recovery performed
from backup memory.
Check whether the
setting for reference
intervals is correct.

W4465 Memory recovery Check the lot information whether the


W4466 WARNING QC card has been read. If the card has
W4468 Recovery performed not been read, read the QC card again.
from backup memory.
Check whether the QC
card is read by con-
firming the lot number.
If not, re-read the QC
card.
8 W4467 Memory recovery Check the lot information whether the
WARNING QC card has been read. If the card has
Recovery performed not been read, contact your dealer for
from backup memory. assistance.
Check whether the QC
card is read by con-
firming the lot number.
If not, contact your
dealer.

W4471 Memory recovery Check the setting for control fluids.


W4472 WARNING
Recovery performed
from backup memory.
Check the control fluid
setting.

8-26 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

Error
Error message Error condition Troubleshooting
code

E4475, Memory initial-value A circuit board signal Please contact your dealer for assis-
E4478- recovery ERR error has occurred. tance.
E4481 Recovery performed
from backup memory.
Contact your dealer.

W4477 Memory recovery Check the operator ID setting.


WARNING
Recovery performed
from backup memory.
Check the Operator ID
setting.

W4483 Memory recovery If the error occurs again, please contact


WARNING (Password) your dealer for assistance.
The administrator
password has been
returned to the default.
Contact your dealer in
case the ERR repeats.

E4490 File deletion ERR Please contact your dealer for assis-
Contact your dealer in tance.
case the ERR repeats.

E4600- Calculation ERR


E4613 Contact your dealer.

W4720 Clock setting WARN- The date and time Set the date and time using [Date &
ING have been reset. Time settings] in Normal mode. 8
Set date & time. è Refer to Section 9.2.1
Contact your dealer in If the error occurs again, please contact
case the ERR repeats. your dealer for assistance.

IMPORTANT:

If the date and time are not adjusted


correctly, the analyzer may fail to de-
termine the expiration of the slides and
the calibration, and the measurement
results may not be accurate.

897N203029A FDC NX600 Instruction Manual 2022.01 8-27


8 Troubleshooting

Error
Error message Error condition Troubleshooting
code

W4721 Clock control failure A circuit board signal Set the date and time using [Date &
WARNING error has occurred. Time settings] in Normal mode.
Set date & time. è Refer to Section 9.2.1
Contact your dealer in
case the ERR repeats.

W4733 Power switch control Please contact your dealer for assis-
WARNING tance.
Contact your dealer.

W5000 USB memory read


ERR
Contact your dealer.

W5002- USB memory read USB memory read er- Check the USB memory and retry ac-
W5019 ERR ror has occurred. cording to the error message.
Lack of necessary files If the error occurs again, please contact
in the USB memory. your dealer for assistance.
Check the USB memo-
ry, and then reselect a
menu.
Contact your dealer in
case the ERR repeats.
(Failed to copy files
from the USB memory.)

W5020 USB memory write A circuit board signal Please contact your dealer for assis-
ERR error has occurred. tance.

8 Contact your dealer.

W5050 No USB memory The USB memory is Insert USB memory.


USB memory can not not inserted. If the error occurs again, please contact
be recognized. your dealer for assistance.

E5500 A b n o r m a l s a m p l e Signal system error When an error occurs after turning the
spreading on slide inside circuit board. power on and off, please contact your
Contact your dealer. dealer for assistance.

E5901 Software logic ERR A software error has Please contact your dealer for assis-
Contact your dealer. occurred. tance.

8-28 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

8.2.9 QC Card, PF Card, and DI Card Errors


(1) When data is not displayed after a QC card is read

Error
Error message Error condition Troubleshooting
code

Problem reading QC Read the QC card again after making


card information sure its correct orientation.
If it does not improve, insert the air
blow nozzle into the QR code reader
slot and blow out the dust inside.

― ―

Spray by
pushing in
the end of the
nozzle about
50 mm.

(2) When QC information has not been read

Error
code
Error message Error condition Troubleshooting
8
W0170 No QC card info The QC card of the Read the QC card of the printed slide
No QC information. slides loaded in the code.
Read QC card. slide cartridge has not
been read.

W4010 No QC type info. There is no type infor- Read the corresponding QC card.
No type information mation for the corre-
corresponding to the sponding item.
slide.
Read QC card corre-
sponding to the slide.

(3) When a PF card has not been read

Error
Error message Error condition Troubleshooting
code

W0137 No PF card info. The PF sample rack Read the PF card included with the PF.
No PF card info. has been identified,
Read PF card. but no PF card has not
been read.

897N203029A FDC NX600 Instruction Manual 2022.01 8-29


8 Troubleshooting

(4) When a DI card has not been read

Error
Error message Error condition Troubleshooting
code

W4012 No DI info. There is no DI card Please contact your dealer for assis-
No DI info found. Can- information. tance.
not read this card.
Contact your dealer.

(5) Other precautions

Error
Error message Error condition Troubleshooting
code

W1700 QC card reader failure The card reader has Please contact your dealer for assis-
QC card reader dam- malfunctioned. tance.
aged.
Measurement is still
possible.
Contact your dealer.

W1703 QC card read ERR QC card read error Make sure the card is inserted at cor-
Check the QC card rect orientation.
direction and re-read it.
(Insert the QR code so
that it faces the far left
side.)
If the error occurs re-
peatedly, contact your
8 dealer.

W4011 QC card ERR Panel card read error Please contact your dealer for assis-
Cannot save data. tance.

W4013 QC card ERR A QC card data error Check whether the QR code of the QC
Check whether this is has occurred. card is dirty.
the correct QC card. NOTE When the error is not resolved,
please contact your dealer for
assistance.

W4014 QC card ERR The QC card cannot After mode function or sample applica-
Cannot read QC card be read at this time. tion ends, read the card again.
for “Mode Function
Underway” or “Spotting
Underway”.
Read again after mode
function ends, or after
spotting ends.

W4015 Expired QC card QC card has been Use QC cards that are before their ex-
QC card has expired expired. piration dates.
and cannot be used.

8-30 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

8.2.10 Slide Loading Errors

Error
Error message Error condition Troubleshooting
code

W0040 Incorrect sample The sample type se- Match the configuration by the [Sample
WARNING lected by the [Sample type] button and the sample type to
The sample type set type] button and the measure the slide to be used, and then
in analyzer does not sample type of the rerun the measurement from the begin-
correspond to the read slide do not match. ning.
sample type. (Example)
Set the sample type • Whole blood (W)
by SAMPLE key and slides were loaded
the slide’s sample type when either [P/S] or
corresponding to the [U] had been se-
sample. Rerun the lected by the [Sample
tests from the begin- type] button.
ning. • Plasma/serum (P)
slides were loaded
when [W] had been
selected by the
[Sample type] button.

W0041 Incorrect sample type The sample type set- Set the sample type to P/S.
WARNING ting of PF measure-
Although a PF sam- ment does not match.
ple rack is loaded, the
sample type setting is
not P/S.

W0042 Incorrect sample rack A re-run test for a Load a PF sample rack in which a
WARNING
Correct sample rack
measurement by us-
ing PF is attempted,
sample extracted by PF has been
loaded. 8
not set in place. but a rack other than
Check the sample a PF sample rack was
rack. loaded.

A PF sample rack has Load a sample rack other than a PF


been loaded for a con- sample rack.
trol measurement.

A sample rack other Load a sample rack for calibrator CP.


than for CP has been
loaded for a calibration
measurement.

W0043 N o s a m p l e r a c k No sample rack is Set a sample rack.


WARNING loaded.
Sample rack is not set.
Set a sample rack.

W0090 ISE slide direction The ISE slide is loaded Load the ISE slide in the slide cartridge
ERR in reverse. in the proper direction, and then meas-
Incorrect loading direc- ure again.
tion of ISE slide.
Check the ISE slide.

897N203029A FDC NX600 Instruction Manual 2022.01 8-31


8 Troubleshooting

Error
Error message Error condition Troubleshooting
code

E0160 Incorrect diluent type Diluent type does not Use specified diluent for the test name,
error match the slide to be and rerun the measurement from the
Diluent type is incor- used. beginning.
rect.
Use specified diluent
for the test name.

W0175 No slide No slides are loaded in Load slides to be tested, and rerun the
Slides have not been the slide cartridge. tests from the beginning.
set.
Rerun the tests from
the beginning.

W5400 Expired slide The validity term of the Always use valid slides.
Slide is expired and slide has expired.
cannot be used.
IMPORTANT:

Slide validity may not be detected in


ISE tests.
è Refer to Section 8.1.3
NOTE Depending on the device set-
tings, this warning may not be
generated.

8-32 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

8.2.11 ISE Measurement Errors

IMPORTANT:

ISE measurement values to which error codes have been appended (E1 - E8) may not be correct.
Perform the following troubleshooting and rerun the measurement.

(1) When a self-test error occurred

Error
Error message Error condition Troubleshooting
code

E0530 ISE self test gain ERR An error was detected The analyzer must be restarted for an
For ISE measurement, during a self-test per- ISE measurement. If the error occurs
turn the power SW formed before the start again, please contact your dealer for
OFF and ON. of an ISE measure- assistance.
Contact your dealer in ment. Tests other than ISE do not require a
case the ERR repeats. restart.
Other parameters
besides ISE can be
measured.

E0531 ISE self test offset


ERR
For ISE measurement,
turn the power SW
OFF and ON.
Contact your dealer in
case the ERR repeats.
Other parameters
besides ISE can be
measured. 8

897N203029A FDC NX600 Instruction Manual 2022.01 8-33


8 Troubleshooting

(2) Errors are appended to measurement results

During an ISE measurement, the checks listed


Check menu Description in the table to the left are performed.
Checking abnormal time course If there is a problem with a measurement value,
Drift check an error code from E1 through E8 is appended
(voltage) during testing
to the measurement result.
Checking the slide impedance
Impedance check
after completing a test
Checking test result is within
Over range check
the measurement range

Code: Description

E1 Drift error

E2 Impedance error

E3 Outside of the measurement range (This is not an analyzer malfunction.)

E4 Drift error and impedance error

E5 Impedance error and outside of the measurement range

E6 Drift error and outside of the measurement range

E7 Drift error, impedance error, and outside of the measurement range

E8 Impossible measurement

< When the error code is not E3 >


The sample or reference fluid may have not been sufficiently absorbed into the slide. Perform the fol-
8 lowing troubleshooting steps.
1. Inspect the sampler O-ring. è Refer to Section 6.6.2
2. Clean the spotting part and ISE probe unit. è Refer to Section 6.3.2
3. Make sure that the analyzer can measure correctly. è Refer to Chapter 7

NOTE If the error occurs again, please contact your dealer for assistance.

< When the error code is E3 >


There is no problem with the analyzer. The concentration of the sample is outside of the measure-
ment range.

NOTE Refer to the “Instructions for Use” of the slide for the measurement range.

8-34 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

(3) Other Errors

Error
Error message Error condition Troubleshooting
code

W0091 Unacceptable ISE An ISE measurement Check ISE measurement errors that oc-
test error has occurred, curred before this error, and then carry
ISE tests cannot run. and ISE measurement out the corresponding troubleshooting.
Turn the power SW cannot be started. After troubleshooting, turn the power
OFF and ON. Rerun of the device on and off, and then start
the tests from the be- measurement again from the begin-
ginning. ning.

W0092 Unacceptable ISE There was an attempt Measure an ISE slide using automatic
manual pipetting to measure an ISE sample application.
ISE test cannot be slide using manual
done in this test mode. sample application
Set a measurable slide measurement.
and redo the test from
the beginning.

897N203029A FDC NX600 Instruction Manual 2022.01 8-35


8 Troubleshooting

8.2.12 PF Errors

Error
Error message Error condition Troubleshooting
code

E0132 PF pressure ERR A pressure error oc- Replace the PF Tube.


Replace the PF pump curred during the filter- è Refer to Section 6.9.4
tube. ing process.

E0133, PF sensor ERR A PF sensor error has Please contact your dealer for assis-
E0150 Contact your dealer. occurred. tance.

W0134 PF sampling ERR An error occurred dur- Clean the PF Suction Pad.
Error occurred during ing PF filtering. è Refer to Section 6.9.1
PF sampling.
Check for a PF leak.

W0138 Unacceptable PF A PF measurement Check PF measurement errors that oc-


sampling WARNING error has occurred, curred before this error, and then carry
For PF measurement, and PF measurement out the corresponding troubleshooting.
turn the power SW cannot be started. After troubleshooting, turn the analyzer
OFF and then ON. on and off, and then start measurement
Contact your dealer in again from the beginning.
case the ERR repeats.
Measurements other
than PF measurements
can be measured with-
out turning the power
SW OFF and then ON.

E1300, PF motor ERR PF motor error move- Turn the power on and off. If the error
E1301, Problem with the plas- ment error occurs again, please contact your dealer
8 E1303, ma filter. for assistance.
E1304 Contact your dealer.

W1308 PF sensor ERR A PF sensor error has Please contact your dealer for assis-
Problem with the plas- occurred. tance.
ma filter.
Contact your dealer.

E1302 PF motor operation PF motor error move- Turn the power on and off. If the error
ERR ment error occurs again, please contact your dealer
Problem with the plas- for assistance.
ma filter.
Contact your dealer.

8-36 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

8.2.13 Errors Related to Data Communication, Sample Barcode Reader


(1) When a communication error occurs

Error
Error message Error condition Troubleshooting
code

W4500, Serial communica- Serial communication Check the following: communication


W4501, tion ERR error cable, host computer, and communica-
W4502 Check the specifica- tion destination settings.
tions and connection è Refer to Section 9.3.2
of the communication NOTE If the error occurs again, please
cable. Contact your contact your dealer for assis-
dealer in case the ERR tance.
repeats.

W4510 LAN communication LAN communication


ERR error
Check the specification
and the connection
of the communication
cable.
Contact your dealer in
case the ERR repeats.

W4511 Type 4 LAN commu- Type 4 LAN communi- Check the host computer and type 4
nication ERR cation ERR settings.
Check whether the set- è Refer to Section 9.3.2
ting for type 4 or host NOTE If the error occurs again, please
computer is correct. contact your dealer for assis-
Contact your dealer in tance.
case the ERR repeats.
8
W4520 Incorrect work list in- The received work list Check the work list information.
formation information does not NOTE If the error occurs again, please
Check whether work list match specifications. contact your dealer for assis-
information for the host tance.
computer is correct.
Contact your dealer in
case the ERR repeats.

E6200, Communication ERR A communication error Please contact your dealer for assis-
E6201, Contact your dealer. has occurred. tance.
E6210

897N203029A FDC NX600 Instruction Manual 2022.01 8-37


8 Troubleshooting

(2) Sample barcode reader error

Error
Error message Error condition Troubleshooting
code

W5330 Barcode reader not The sample barcode Connect the sample barcode reader to
connected reader is not connected. the USB connector, and then turn the
Check the barcode analyzer off and on.
reader connection.

W5331 Exceeded number of Barcode label error Check the barcode label.
characters from bar-
code reader
The number of input
characters from the bar
code reader exceeded
the capacity.

W5332 Barcode WARNING Barcode data error Check the input value and re-enter.
Check input value and
re-enter.

8-38 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

8.2.14 Calibration Errors


(1) An immuno-chemical test has been performed without calibration.

Error
Error message Error condition Troubleshooting
code

W0180 Calibration not per- An immuno-chemical Perform the calibration.


formed test has been per- è Refer to Section 4.1
Perform calibration. formed without calibra-
tion.

(2) A slide not suited for calibration has been loaded.

Error
Error message Error condition Troubleshooting
code

W0190 Wrong slide for cali- A slide not suited for Check the slide. Rerun the calibration
bration calibration has been process again.
A slide that does not loaded during the cali-
require calibration is bration process.
loaded. Check the
slide and perform the
calibration from the
beginning.

W0191 Wrong slide lot for A slide different from Load a slide of the specified lot, and
calibration the lot specified during then calibrate again.
A different lot of slide calibration has been
other than designated loaded.
in the calibration mode
is loaded. Load a slide 8
corresponding to the
designated lot and
perform the calibration
from the beginning.

(3) Insufficient calibration data

Error
Error message Error condition Troubleshooting
code

E4200 Calibration ERR This error may occur Rerun 2 measurements for the printed
Invalid number of due to a lack of calibra- level.
slides tion data or no applica-
The number of meas- tion on the slide.
ured slides for the fol-
lowing calibration data
is insufficient or the
sample is unspotted.

897N203029A FDC NX600 Instruction Manual 2022.01 8-39


8 Troubleshooting

(4) Calibration result problem

Error
Error message Error condition Troubleshooting
code

E4201 Calibration variation The variation of meas- Rerun 2 measurements for the printed
error urement values during level.
Variations in measured calibration is large.
value for the following
calibration are large.

(5) Standard curve configuration error

Error
Error message Error condition Troubleshooting
code

E4202 Calibration curve set- The wrong calibrator Check the expiration date and the set-
ting ERR (span error) may have been used. ting position of the calibrator, and then
Misplacement of cali- calibrate again from the beginning.
brator.
Check the level and
expiration date of the
calibrator and perform
the calibration from the
beginning.

E4203 Calibration curve set-


ting ERR (range error)
Misplacement of cali-
brator.
8 Check the level and
expiration date of the
calibrator and perform
the calibration from the
beginning.

E4205 Calibration calcula- Calibration calculation Calibrate again from the beginning.
tion ERR problem
Failed in calibration
calculation.
Do the calibration
again.

8-40 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

(6) Measurement errors during calibration

Error
Error message Error condition Troubleshooting
code

E4204 Calibration measure- Error marks such as Troubleshoot the printed errors and
ment ERR “ * ”, “ ? ”, and “$” are then calibrate again from the beginning.
Error in measured printed on the meas-
result. urement results.
After coping with the
printed error, perform
the calibration from the
beginning.

897N203029A FDC NX600 Instruction Manual 2022.01 8-41


8 Troubleshooting

8.2.15 Other Errors

Error
Error message Error condition Troubleshooting
code

W0020 No disposal box Measurement was Insert the disposal box completely, and
Put Disposal box to started with the dis- then start measurement.
the end and then press posal box pulled out.
START key.

E0021 Disposal box open The disposal box was Close the disposal box, and then start
ERR open during measure- measurement again from the begin-
Disposal box opened ment. ning.
during measurement.
Close Disposal box
and run the tests from
the beginning.

W0022 Disposal box full The disposal box is full While a measurement is not being per-
Disposal box became of slides and tips. formed, discard the slides and tips in
full of slides and tips. the disposal box.
Discard slides and tips NOTE When there are 150 slides or
to empty the box when 75 tips in the disposal box, this
measurement is not in error occurs. If the disposal box
progress. is pulled out, it is determined
that the slides and tips were
discarded, and the disposal
count is reset. Always empty
the disposal box after pulling it
out.

8 E0080 Sample tip sensor


ERR
Malfunction of the
sample tip sensor.
Please contact your dealer for assis-
tance.
Contact your dealer.

W0500 Rear fan 1 stop A fan error has oc-


Contact your dealer. curred.

W0501 Rear fan 2 stop


Contact your dealer.

W0502 Intake fan stop


Contact your dealer.

W0503 Exhaust fan stop


Contact your dealer.

I5920 Turn the power ON A long time has passed Switch the analyzer off and on.
30 days or more have since start-up.
passed from when the
analyzer was started.
For stable system
performance, turn the
power ON.

8-42 897N203029A FDC NX600 Instruction Manual 2022.01


8 Troubleshooting

Error
Error message Error condition Troubleshooting
code

W4440 Setting value reset The settings have Check settings for the reference in-
WARNING been reset. terval range and the control fluid, and
The reference interval check correlation coefficients.
range setting, control
fluid setting, and corre-
lation coefficients have
been reset. Check the
settings.

W66xx Input value ERR The entered value is Check the entered value and enter it
Check input value and incorrect. again.
re-enter.

897N203029A FDC NX600 Instruction Manual 2022.01 8-43


8 Troubleshooting

8.3 Slide Code Table


Slide code Slide code
Test Test
Test code Sample code Test code Sample code
GLU -P 10 50 GGT -P 30 50
BUN -P 11 50 GOT/AST -P 31 50
UA -P 13 50 GPT/ALT -P 32 50
TCHO -P 14 50 CPK -P 33 50
NH3 -P 15 50 LDH -P 34 50
TG -P 16 50 ALP -P 35 50
CRE -P 17 50 LAP -P 37 50
TP -P 18 50 CKMB -P 38 50
ALB -P 20 50 CHE -P 39 50
TBIL -P 21 50 AMYL -P 43 50
Ca -P 23 50 LIP -P 44 50
IP -P 24 50 TCO2 -P 45 50
DBIL -P 25 50 CRP -S 51 50
HDLC -P 26 50 Na 91 00
Mg -P 28 50 K 92 00
Cl 93 00
GLU -W 10 00
NH3 -W 15 00

NOTE Slide codes are subject to change without notice.


Check the slide code printed on the slide box.

8-44 897N203029A FDC NX600 Instruction Manual 2022.01


9 Function Menu
If the (Function menu) button is tapped, the Function menu screen appears.

Function menu list


Function menu items Settings Advanced settings Reference
section
Calibration — — 4.1.1
Control — — 9.1
START — — 4.2
Manual pipetting
Date & Time settings — 9.2.1
Brightness • Volume Display brightness 9.2.2
Error sound
Other general sound
Lamp management Lamp ON time 9.2.3 9
Lamp off timer setting
Lot information — 9.2.4
Normal mode Sampler leak check Sample nozzle 6.6.2
Reference fluid nozzle
Replace O-ring — 6.6.3
PF maintenance Cleaning 6.9.1
Leak check 6.9.2
Suction pad 6.9.3
Reference plate level check — 6.3.1
Cleaning mode — 6.3.2

897N203029A FDC NX600 Instruction Manual 2022.01 9-1


9 Function Menu

Function menu items Settings Advanced settings Reference


section
Language setting — 9.3.1
Host connection setting None 9.3.2
Serial connection
LAN connection
Calculated Parameter — 9.3.3
Unit conversion — 9.3.4
Switch measurement range dis- 9.3.5

play
Reference interval range setting — 9.3.6
Sample No. /ID settings — 9.3.7
Work list selection setting Sample ID / Patient name 9.3.8
Patient ID / Patient name
Correlation coefficients — 9.3.9
Lot compensation coefficients — 9.3.10
Standard curve coefficients — 9.3.11
Mode function for Lot Qualification — 9.3.12
administrator
PF calibration coefficients — 9.3.13
Dilution setting — 9.3.14
Tube setting for diluent & refer- Diluent 9.3.15
ence fluid Reference fluid
Print setting for measured result Number of printing sheets 9.3.16
Print order
Print with the ranges
Control fluid setting — 9.3.17
Control measurement setting — 9.3.18
Operator ID setting — 9.3.19
Test item name setting Display and printout 9.3.20
9 Data transmission
UREA setting
Default sample type setting — 9.3.21
Password setting — 9.3.22
Error log — — 8.1.1

9-2 897N203029A FDC NX600 Instruction Manual 2022.01


9 Function Menu

9.1 Control
This mode is used for measurement of control serum (FUJI DRI-CHEM CONTROL QP) and for per-
forming control surveys. When control mode is set, measurements are performed under the condi-
tions listed below.
• Correlation coefficients: follow [Control measurement setting] (Initial values disabled (a=1, b=0))
of the function menu.
• Measurement units: follow [Unit settings] of the function menu.
• Dilution conditions: dilution is not normally performed. (This can be configured at the Sample
information registration screen.)

NOTE When the analyzer is switched on again, control mode is canceled.

1. Prepare the slide to be measured, and


then load it on the analyzer.
2. Place the control fluid to be measured in
the tube, and then load it in the sample
position.
è Refer to Section 3.3 (3)
3. Tap the button.

4. Tap the [Control] button.

897N203029A FDC NX600 Instruction Manual 2022.01 9-3


9 Function Menu

5. Enter or choose the sample information.


1 Enter the sample number using the software key-
1
board (Section 2.3).
2
2 Enter the sample ID using the software keyboard
3 (Section 2.3).
4 3 Select the control liquid from the list (when nec-
essary).
5
4 Select the control liquid lot from the list (when
6 necessary).

IMPORTANT:

9 When you configure the control liquid lot, make


sure it matches the lot of the control liquid to be
used. If it does not match, the measurement may
not be valid.
5 For the sample type, generally (in the case of
QP-L, QP-H, and QE) select [Plasma/Serum].
6 For dilution, generally select [None].

9-4 897N203029A FDC NX600 Instruction Manual 2022.01


9 Function Menu

6. Close the sample cover, and then lock it


with the sampler cover lock key.

WARNING:

Ensure the sampler cover is locked during sample


processing to prevent injuries and infection.
When the display reads [Ready to test], the sam-
pler cover can be opened.

7. Tap the [START] button to start control


measurement.

897N203029A FDC NX600 Instruction Manual 2022.01 9-5


9 Function Menu

8. After measurement, tap the button to


check the measurement results.

IMPORTANT:

After using control mode, always tap the [HOME]


button to clear control mode.

<Display of measurement results>


Green: when within the acceptable range of control
Red: when outside of the acceptable range of control

Acceptable Acceptable
lower limit upper limit

NOTE The acceptable range of control displays the


value entered for [Control fluid setting] (see
Section 9.3.17).

9-6 897N203029A FDC NX600 Instruction Manual 2022.01


9 Function Menu

9.2 Normal Mode


General operators use this function menu to, for example, set the time and configure the brightness
of the LCD.

Settings Advanced settings Reference


section
Date & Time settings — 9.2.1
Brightness • Volume Display brightness 9.2.2
Error sound
Other general sound
Lamp management Lamp ON time 9.2.3
Lamp off timer setting
Lot information — 9.2.4
Sampler leak check Sample nozzle 6.6.2
Reference fluid nozzle
Replace O-ring — 6.6.3 9
PF maintenance Cleaning 6.9.1
Leak check 6.9.2
Suction pad 6.9.3
Reference plate level check — 6.3.1
Cleaning mode — 6.3.2

897N203029A FDC NX600 Instruction Manual 2022.01 9-7


9 Function Menu

9.2.1 Date & Time Settings


This mode is used to set the date and time settings stored in the memory.

• Tap “ ” or “ ” to change the date and time


values.
• Configure the order of the year, month, and date
by using [Date format setting].

NOTE The analyzer must be restarted to enable


a change of [Date format setting].
The confirmation dialog for the power-
off will appear when is tapped to
exit Normal mode. After the shutdown,
reboot the analyzer.

9.2.2 Brightness • Volume


You can configure the brightness of the LCD and buzzer sounds (error warning, measurement com-
9 pleted).

• Configure the brightness of the LCD using


[Display brightness]. Use and to set the
minimum and maximum brightness.
• Configure the buzzer sound when an error oc-
curs by using [Error sound]. Use and to
set the minimum and maximum.
• Configure the buzzer sound when measurement
ends by using [Other general sound]. Use
and to set the minimum and maximum. If
is configured, the buzzer is muted.

9-8 897N203029A FDC NX600 Instruction Manual 2022.01


9 Function Menu

9.2.3 Lamp Management


Display and reset the cumulative illumination time of the lamp currently installed in the analyzer. Fur-
thermore, choose whether to turn off the lamp automatically when the analyzer has not been used
for some time.

• You can check the cumulative illumination time


by using [Lamp ON time]. If the [Reset] button is
tapped, the cumulative illumination time is reset.

NOTE Reset the cumulative illumination time


of the lamp after replacing the lamp.
• If the [Lamp off timer setting] checkbox is select-
ed, the lamp is switched off automatically after
some time. You can configure the time until the
lamp is switched off.

NOTE The default setting is 20 minutes.

9.2.4 Lot Information


The screen displays the lot information of currently registered QC cards. You can also print all cur-
rently registered lot information.

• When there are two or more pages that can be


displayed, tap the and buttons to scroll
the pages.
• If the button is tapped, all currently regis-
tered lot information is printed.
9
NOTE Up to five lots are registered for one
test.

897N203029A FDC NX600 Instruction Manual 2022.01 9-9


9 Function Menu

9.3 Mode Function for Administrator


NOTE To display Mode function for administrator, open the Function menu page, tap the [Mode
function for administrator] button, and then enter the password.

NOTE In order to prevent other people from oper-


ating the analyzer, the input dialog for the
password will appear when no operation is
performed for a fixed period of time during
the operation of [Mode function for adminis-
trator].

Settings Advanced settings Reference section


Language setting — 9.3.1
Host connection setting None 9.3.2
Serial connection
LAN connection
Calculated Parameter — 9.3.3
Unit conversion — 9.3.4
Switch measurement range display — 9.3.5
Reference interval range setting — 9.3.6
Sample No. /ID settings — 9.3.7
Work list selection setting Sample ID / Patient name 9.3.8
9 Patient ID / Patient name
Correlation coefficients — 9.3.9
Lot compensation coefficients — 9.3.10
Standard curve coefficients — 9.3.11
Lot Qualification — 9.3.12
PF calibration coefficients — 9.3.13
Dilution setting — 9.3.14
Tube setting for diluent & reference fluid Diluent 9.3.15
Reference fluid
Print setting for measured result Number of printing sheets 9.3.16
Print order
Print with the ranges
Control fluid setting — 9.3.17
Control measurement setting — 9.3.18
Operator ID setting — 9.3.19
Test item name setting Display and printout 9.3.20
Data transmission
UREA setting
Default sample type setting — 9.3.21
Password setting — 9.3.22

9-10 897N203029A FDC NX600 Instruction Manual 2022.01


9 Function Menu

9.3.1 Language Setting


Configure the language to be used.

Tap the button of language to be configured.

NOTE Rebooting the device is necessary for the


setting to take effect. The confirmation di-
alog for the power-off will appear when
is tapped to exit Mode function for adminis-
trator. After the shutdown, reboot the ana-
lyzer.

9.3.2 Host Connection Setting


Configure the connection terminal and communication parameters to connect to the host computer
(PC). Furthermore, switch the built-in printer on or off.

• When you will not connect to a host computer,


set [None].

897N203029A FDC NX600 Instruction Manual 2022.01 9-11


9 Function Menu

• If [Serial connection] is set, configure the com-


munication type and whether to output from the
printer. You can choose from 3 communication
types.
- Type 1: Bidirectional connection
- Type 2: Unidirectional connection (bidirectional
format)
- Type 3: Unidirectional connection (standard
format)

NOTE Rebooting the device is necessary for


the setting to take effect. The confirma-
tion dialog for the power-off will appear
when is tapped to exit Mode function
for administrator. After the shutdown, re-
boot the analyzer.

• If [LAN connection] is set, configure the com-


munication type and whether to output from the
printer. You can choose from 4 communication
types.
- Type 1: Bidirectional connection
- Type 2: Unidirectional connection (bidirectional
format)
- Type 3: Unidirectional connection (standard
format)
- Type 4: Unidirectional connection (HL7 for-
matted)

• When was tapped, the LAN configuration


screen appears. If an entry field is tapped, the
software keyboard (refer to Section 2.3) appears
9 so you can edit the value.

NOTE The following section shows the default


LAN settings.
FDC NX600 settings
IP address:192.168.100.1
Subnet mask:255.255.255.0
Port number:65535
Gateway: 0.0.0.0
PC settings
IP address:192.168.100.2
Port number:65535
NOTE Rebooting the device is necessary for
the setting to take effect. The confirma-
tion dialog for the power-off will appear
when is tapped to exit Mode function
for administrator. After the shutdown, re-
boot the analyzer.

9-12 897N203029A FDC NX600 Instruction Manual 2022.01


9 Function Menu

If the button is also tapped when Type 4 was


selected, the settings screen of Type 4 of page 3
becomes active.

• Tap [Examination code] to select the standard for


examination item classification code to be used
in communication.
• Tap [MLLP setting] to set MLLP (Minimal Lower
Layer Protocol) in communication.
Full: MLLP is applied to full message in com-
munication.
Custom: MLLP is applied to only at the end of
message in communication.
• Tap [Sample type] to select the sample type used
in communication.

NOTE Rebooting the device is necessary for the


setting to take effect. The confirmation di-
alog for the power-off will appear when
is tapped to exit Mode function for adminis-
trator. After the shutdown, reboot the ana-
lyzer.

897N203029A FDC NX600 Instruction Manual 2022.01 9-13


9 Function Menu

9.3.3 Calculated Parameter


You can configure the settings to calculate the following items from simultaneously measured values
and then display the results. Items registered for calculation have their measurement results dis-
played and printed.
Test name Calculation Unit Calculation formula
test name
LDL-C = TCHO value - (HDL-C value + TG value/5) [mg/dL]
mg/dL LDL-C = TCHO value - (HDL-C value + TG value/2.2) [mmol/L]
LDL Cholesterol (LDL-C) LDL or
mmol/L NOTE LDL-C is NOT calculated when the measured value of
TG exceeds 400 mg/dL (4.52 mmol/L).
mg/dL non-HDL = TCHO value - HDL-C value
non-HDL Cholesterol
non-HDL or
(non HDL-C)
mmol/L
g/dL or GLOB = TP value - ALB value
Globulin GLOB
g/L
Albumin/Globulin ratio ALB/GLOB — ALB/GLOB = ALB value / (TP value - ALB value)
BUN/CRE = BUN value/CRE value [mg/mg]
BUN/CRE = BUN value ×1,000/CRE value[(SI)]
mg/mg
BUN/Creatinine ratio BUN/CRE or NOTE BUN measured value in SI unit (mmol/L) is
(SI) shown as molar concentration of Urea. BUN/
CRE ratio in SI unit (molar ratio) is shown Urea
concentration/creatinine concentration ratio.
GOT/GPT = GOT value / GPT value
(AST/ALT = AST value / ALT value)
GOT/GPT ratio GOT/GPT
—NOTE The calculation test name can be changed by
(AST/ALT ratio) (AST/ALT)
[Test item name setting] in the Mode function for
administrator. è Refer to Section 9.3.20.
Sodium/Potassium ratio Na/K — Na/K = Na value/K value
mEq/L Anion Gap = Na value - (Cl value + TCO2 value)
Anion Gap Anion Gap or
9 mmol/L

IMPORTANT:

Calculation items display a value calculated from simultaneously measured values. Therefore, cal-
culation results are impacted by errors included in measurement values and the effects of interfer-
ing materials. Please understand that when you use the results for a diagnosis, you must make an
overall determination in combination with the measurement values is used in the calculation.

NOTE In the case of additional testing and retesting, items subject to calculation must be meas-
ured at the same time.

NOTE When there are two or more slides of the same item subject to calculation in the slides
measured at the same time, the initial measurement value is used.

NOTE When the measurement results of items subject to calculation exceed the measurement
range (determination range), the calculation results of this calculation item are displayed
and printed as "****".

NOTE When an item subject to calculation has expired, an (expiry) warning symbol (#) is displayed
and printed for the measurement results of this item and the calculation results.

9-14 897N203029A FDC NX600 Instruction Manual 2022.01


9 Function Menu

If the [Calculated Parameter] button is tapped in


Mode function for administrator, precautions for use
of calculation items are displayed. To use calcula-
tion items after confirming the precautions, tap the
[OK] button.

Configure the calculation items to be used.

NOTE If the QC card of an item to be used for cal-


culation has not been read, the calculation
item is not displayed.

897N203029A FDC NX600 Instruction Manual 2022.01 9-15


9 Function Menu

9.3.4 Unit Conversion


Configure the main unit and secondary unit for each measurement item. The following table shows
the units that can be used for each configuration item and the conversion coefficient
Test Test Unit (A) Unit (B) Conversion Test Test Unit (A) Unit (B) Conversion
code coefficient code coefficient
GLU 10 mg/dL mmol/L 0.05551 GGT 30 U/L μkat/L 0.0167
BUN 11 mg/dL mmol/L 0.357 GOT/ 31 U/L μkat/L 0.0167
AST
UA 13 mg/dL μmol/L 59.48 GPT/ALT 32 U/L μkat/L 0.0167
TCHO 14 mg/dL mmol/L 0.02586 CPK 33 U/L μkat/L 0.0167
NH3 15 μg/dL μmol/L 0.7139 LDH 34 U/L μkat/L 0.0167
TG 16 mg/dL mmol/L 0.01129 ALP 35 U/L μkat/L 0.0167
CRE 17 mg/dL μmol/L 88.4 LAP 37 U/L μkat/L 0.0167
TP 18 g/dL g/L 10 CKMB 38 U/L μkat/L 0.0167
ALB 20 g/dL g/L 10 CHE 39 U/L μkat/L 0.0167
TBIL 21 mg/dL μmol/L 17.1 AMYL 43 U/L μkat/L 0.0167
Ca 23 mg/dL mmol/L 0.25 LIP 44 U/L μkat/L 0.0167
IP 24 mg/dL mmol/L 0.3228 TCO2 45 mmol/L mmol/L 1
DBIL 25 mg/dL μmol/L 17.1 CRP 51 mg/dL mg/L 10
HDL-C 26 mg/dL mmol/L 0.02586 Na 91 mEq/L mmol/L 1
Mg 28 mg/dL mmol/L 0.4114 K 92 mEq/L mmol/L 1
Cl 93 mEq/L mmol/L 1

NOTE Unit (B)= Unit (A)× Conversion coefficient

NOTE BUN measured value in Unit (B) (mmol/l) is shown as molar concentration of Urea.

• Tap the button of the unit to be used for each


measurement item.

Next-page buttons

9-16 897N203029A FDC NX600 Instruction Manual 2022.01


9 Function Menu

9.3.5 Switch Measurement Range Display


Switch the display method for data outside the measurement range (passing the upper/lower limit of
the determination range) for all tests.

• If [@] is configured, an [@] symbol is added to


indicate the displayed measurement value is
outside the measurement range.
• If [< / >] is configured, a value outside the meas-
urement range is displayed by a [<] or [>] sym-
bol.

897N203029A FDC NX600 Instruction Manual 2022.01 9-17


9 Function Menu

9.3.6 Reference Interval Range Setting


You can add, edit, and change the order of the reference interval names that can be selected when
registering sample information. (Refer to Section 3.3 (4)) Furthermore, you can set a reference inter-
val for each sample type ([P/S, W] or [U]) of a reference interval name.

Adding and editing of reference interval names

Tap the button of the item you want to edit or


add.

Next-page buttons

If an entry field is tapped, the software keyboard


(refer to Section 2.3) appears so you can edit the
test name and other details. To change the sample
9 type when configuring the reference interval, tap the
[P/S, W] button or [U] button.

9-18 897N203029A FDC NX600 Instruction Manual 2022.01


9 Function Menu

Changing the display order

Tap the button of the test for which you want to


change the display order.

Next-page buttons

Tap the field of the test for which you want to con-
figure the transfer destination, and then tap the [OK]
button.

897N203029A FDC NX600 Instruction Manual 2022.01 9-19


9 Function Menu

9.3.7 Sample No. /ID Settings

• Configure whether to display the sample number


in the measurement results under [Edit and print
No.].
• Configure whether to record the sequence num-
ber of the analyzer under [Store No.].

9.3.8 Work List Selection Setting

Configure whether to display the sample ID or pa-


tient ID at the Work list index screen.

9-20 897N203029A FDC NX600 Instruction Manual 2022.01


9 Function Menu

9.3.9 Correlation Coefficients


Enter, reset, or print the correlation coefficients for each sample type ([P/S,W] or [U]). For details,
refer to “Description of the correlation function” at the end of the document.

IMPORTANT:

Incorrect entry of “a” and “b” causes incorrect measurement results. Always make sure that the
entries (a, b) are correct.

Next-page buttons Tap the [Data input] button.

NOTE If the button is tapped, all correlation


coefficient settings are printed.

NOTE If the [Reset all tests] button is tapped, a


reset confirmation screen appears. To reset
the correlation coefficients of all measure-
ment items, tap the [OK] button. To cancel
the reset, tap the [Cancel] button.

897N203029A FDC NX600 Instruction Manual 2022.01 9-21


9 Function Menu

Tap the button of the measurement item you want


to edit.

NOTE Only measurement items for which QC


cards have been read are displayed.

If the [Input] field of an item is tapped, the software


keyboard (refer to Section 2.3) appears so you can
enter the value.
To change the sample type, tap the [P/S], [W], or [U]
button.

NOTE If the [Reset] button is tapped, a reset con-


firmation screen appears. To reset the dis-
played correlation coefficients for the sam-
ple type of the selected measurement item,
tap the [OK] button. To cancel the reset, tap
the [Cancel] button.

NOTE If values are entered for "a" and "b" when


9 values have already been entered, the
current values are not discarded, and cor-
rections are also applied. For details, refer
to following “Description of the correlation
function”.

9-22 897N203029A FDC NX600 Instruction Manual 2022.01


9 Function Menu

IMPORTANT:

Description of the correlation function

This function is designed to determine the correlation


between the measured data obtained using the FDC
Y NX600 and the data obtained using the conventional
measuring method with your own instruments.
Y=aX+b
The X-axis shows the measured data obtained us-
ing your instruments, and the Y- axis shows the
FDC measured data obtained using the FDC NX600.
NX600 The correlation regression equation in this case is:
Y = aX + b
After values for the two coefficients (a, b) are reg-
istered in the analyzer memory, the FDC NX600
performs compensation calculations internally us-
ing the formula: X = (Y-b)/a. In this way, measured
data obtained using the FDC NX600 are adjusted to
X match measured data that would be obtained using
Conventional measuring method
your instruments.
using your instruments

Method
Y
Y=X In order to obtain a better correlation, it is es-
(mg/dl)
Y=0.98X+5
sential to exercise caution regarding the amount
600 of data and the sample.
1. You can configure a and b values by sam-
ple type.
FDC
reference 2. The correlation function is independent of
other compensation functions.
3. If values for (a, b) have already been input,
old values will not be canceled as follows. If
the old values are represented as (a1, b1),
9
0
X and the new values as (a2, b2), the result-
0 A 10 600 B
(mg/dl)
ing values for (a, b) will be determined ac-
Measurement using your instruments cording to the following formulas:
a=a1×a2
b=a1×b2+b1

897N203029A FDC NX600 Instruction Manual 2022.01 9-23


9 Function Menu

4. Depending on the values input for (a, b), the


apparent limits of the measurement range
will shift as follows.
(Example) GLU measurements
If a=1 and b=0 (Y=X), the measure-
ment range is 10 to 600 mg/dl. But by
inputting regression coefficients as
shown, the lower limit changes to A
and the upper limit to B.

Printout example If the regression formula is Y = 0.98X


+ 5:
Value
under the A = 5 mg/dl
meas- B = 607 mg/dl
urement
range
Value
above the
meas-
urement
range

If the regression formula is Y = 1.1X - 5:


Y
Y=1.1X-5
A = 14 mg/dl
(mg/dl)
Y=X B = 550 mg/dl
600

FDC
reference

9 0
X
0 10 A B 600
(mg/dl)
Measurement using your instruments

Printout example If the measured data is below the


Value measurement range, an “@” mark is
under the added as shown in the example print-
meas- out.
urement If the measured data is above the
range measurement range, a “>” sign is print-
ed as shown in the example printout.
Value
above the
meas- IMPORTANT:
urement
range Reported values with “@” marks
may not be accurate. Run the meas-
urement again.

9-24 897N203029A FDC NX600 Instruction Manual 2022.01


9 Function Menu

9.3.10 Lot Compensation Coefficients


Enter the lot compensation coefficients (c, d, and e) printed on the QC cards of slides. This mode is
needed if it is not possible to read the data directly because the QC card was lost or spoiled.

IMPORTANT:

Incorrect inputs for (c, d, and e) will cause incorrect measurement results.
Make sure that the inputs for (c, d, and e) are correct using this mode.

IMPORTANT:

When the slide type number was changed, even if the new values (c, d, and e) are entered by us-
ing this setting, the new standard curve information will not be updated. It is necessary to read the
QC card when the slide type number was changed.
The slide type number is the most significant digit of a lot number.
(Example) The slide type number of “Lot No.123456” is 1.

Tap the button of the “Measurement item - sample


type” you want to edit.

NOTE Only “Measurement item - sample type”


items that have been read from QC cards
are displayed as buttons.

NOTE Immuno-chemical items are not displayed.

NOTE Measurement items are displayed in as-


cending order of test code, and sample type
is displayed in the following order: [PS] and
[W].

897N203029A FDC NX600 Instruction Manual 2022.01 9-25


9 Function Menu

Tap the button of the Lot No. for which you want to
enter a lot compensation coefficient.

NOTE If the button is tapped, all lot compen-


sation coefficient settings are printed.

If the [Input] field of an item is tapped, the software


keyboard (refer to Section 2.3) appears so you can
enter the value.

NOTE The most significant digit of a [Lot] number


(the type number) is fixed. Enter from the
second digit.

NOTE Incorrect entries will cause incorrect meas-


urement results. Always make sure that the
displayed values are correct.

NOTE If the button is tapped, all lot compen-


sation coefficient settings are printed.
9 NOTE For the expiration date for use (EXP), enter
xxxx-xx (four digits for the year and two dig-
its for the month). Also enter the hyphen (-).

9-26 897N203029A FDC NX600 Instruction Manual 2022.01


9 Function Menu

9.3.11 Standard Curve Coefficients


The calibration curve coefficients (A, B, C, D, E, and F) of the QC card information can be entered.

IMPORTANT:

Incorrect entry of calibration curve coefficients causes incorrect measurement results.


Always make sure that the entered coefficients are correct.

NOTE The calibration curve coefficients of this setting are entered automatically by QC card read-
ing and calibration.

<Configuring the lot number>

1. Tap the button of the measurement item


for which the calibration curve is to be
edited.
NOTE Only CRP items for which QC cards
have been read are displayed.

2. From the list display area, tap the lot


number for which the calibration curve
is to be edited. 9

List display area

897N203029A FDC NX600 Instruction Manual 2022.01 9-27


9 Function Menu

3. If the [Input] field of an item is tapped,


the software keyboard appears so you
can enter the value.
NOTE If the button is tapped, the calibra-
tion curve coefficients are printed.
4. First tap the [Input] field of [Lot].
The Lot number entry screen (software key-
board screen) appears.

5. Enter a lot number, and then tap the [OK]


button.
NOTE The most significant digit (the type
number) is fixed. Enter from the second
digit.
The Calibration curve coefficient (A:) entry
screen (software keyboard screen) appears.

9-28 897N203029A FDC NX600 Instruction Manual 2022.01


9 Function Menu

<Configuration of calibration curve coefficients (A, B, C, D, E, and F)>


1. Enter the calibration curve coefficient
(A:), and then tap the [OK] button.
The Calibration curve coefficient (B:) entry
screen (software keyboard screen) appears.
2. Enter B through F using the same proce-
dure.

<Configuring the expiration date for use>


For the expiration date for use (EXP), enter xxxx-
xx (four digits for the year and two digits for the
month).

NOTE Also enter the hyphen (-).

IMPORTANT:

Incorrect entries will cause incorrect measurement results.


Always make sure that the displayed values are correct.

NOTE When the button is tapped, values of the calibration curve coefficient configuration re-
sults are printed.

897N203029A FDC NX600 Instruction Manual 2022.01 9-29


9 Function Menu

9.3.12 Lot Qualification


This function button appears (enabled) when USM function is enabled. It is possible to (dis-)qualify
slide LOTs which are registered in analyzer.

1. Select a slide name which needs to


change the setting.

2. (Dis-)qualification for selected slide LOT


can be changed by selecting [ON] or
[OFF] button.
NOTE When the disqualified slide lot is meas-
ured, the test result will be indicated as
“CHK QC” on screen, print and in com-
munication.

9-30 897N203029A FDC NX600 Instruction Manual 2022.01


9 Function Menu

9.3.13 PF Calibration Coefficients


Configure the PF calculation correction coefficients.

Next-page buttons Tap the [Data input] button.

NOTE I f t h e button is tapped, all PF


calculation correction coefficients are
printed.

NOTE If the [Reset all tests] button is tapped, a


reset confirmation screen appears. To reset
the PF calculation correction coefficients of
all measurement items, tap the [OK] button.
To cancel the reset, tap the [Cancel] button.

Tap the button of the measurement item you want


to edit.

NOTE Only measurement items for which QC


cards have been read are displayed as but-
tons.

897N203029A FDC NX600 Instruction Manual 2022.01 9-31


9 Function Menu

If the [Input] field of an item is tapped, the software


keyboard (refer to Section 2.3) appears so you can
enter the value.

NOTE If the [Reset] button is tapped, a reset con-


firmation screen appears. To reset the PF
calculation correction coefficients of select-
ed items, tap the [OK] button. To cancel the
reset, tap the [Cancel] button.

NOTE To cancel the configuration, tap the button.

9-32 897N203029A FDC NX600 Instruction Manual 2022.01


9 Function Menu

9.3.14 Dilution Setting


You can configure the dilution factor and position of the diluent to be used for each measurement
item and sample type (PS and U). When a dilution factor is not configured for each sample on the
sample information registration screen, measurement is performed using the dilution factor config-
ured here.

Next-page buttons 1. Tap the [Data input] button.


NOTE When the button is tapped, all dilution
factor settings will be printed.

2. Tap the button of the measurement item


to be edited.

897N203029A FDC NX600 Instruction Manual 2022.01 9-33


9 Function Menu

3. For the dilution factor, tap a button, and


then select from [None], [2], [3], [4], [5],
or [10].

NOTE In the case of CRP, the selection menu


to the left appears.

9-34 897N203029A FDC NX600 Instruction Manual 2022.01


9 Function Menu

9.3.15 Tube Setting for Diluent & Reference Fluid


Configure the test tube type to be used for diluent and reference fluid.

NOTE For the types of containers that can be used for diluent and reference fluid, refer to Section 5.4
and Section 5.5.

Tap the button of the container you want to use.

9.3.16 Print Setting for Measured Result


Configure the print quantity (1 to 3 sheets) and print order of measurement results.

• Tap the print quantity you want to configure.


• For [Print order], you can select [Slide stacking
order] or [Fixing].

NOTE If [Fixing] is selected, printing occurs in the


following order. 9
<Plasma/serum and urine>
TP → ALB → TBIL → DBIL → GOT/AST
→ G P T / A LT → A L P → L D H → C H E
→ GGT → LAP → CPK → AMYL → TCHO
→ HDLC → TG → UA → BUN → CRE
→ Na-K-Cl → Ca → IP → Mg → GLU
→ CRP → NH3 → CKMB → LIP → TCO2

<Whole blood>
Na-K-Cl → GLU → NH3

• For [Print with the ranges], to print out the ranges


along with test results, check the checkbox.

NOTE By default, the checkbox is checked (ena-


bled).

897N203029A FDC NX600 Instruction Manual 2022.01 9-35


9 Function Menu

9.3.17 Control Fluid Setting


Configure the control liquid name / lot number and acceptable range.

1. Tap the button of the control liquid to be


edited.
NOTE [-] (control liquid not registered) of Step
1 cannot be edited.

2. Tap a [Lot] button to choose a lot.


3. If you touch a display section (lower limit
or upper limit) of the acceptable range to
be edited, a software keyboard (Refer to
Section 2.3) appears so you can edit the
value.

IMPORTANT:

If the entry of the acceptable range is incorrect,


the judgement of the control measurement results
9 is incorrect. Always make sure that the entered
values are correct.

NOTE The units configured by the [Unit conver-


sion] of the function menu are displayed.

NOTE If the acceptable range is changed, in-


formation from before the change is not
saved.

4. To edit a lot number, touch the [Lot edit-


ing] button, and then enter the number
using the software keyboard that ap-
pears.
NOTE To edit or add a control liquid name,
tap the button of a control liquid name,
and then enter the information using the
software keyboard that appears.

9-36 897N203029A FDC NX600 Instruction Manual 2022.01


9 Function Menu

9.3.18 Control Measurement Setting


Configure whether to enable the correlation coefficient when performing a control measurement.

To enable the correlation coefficient for a control


measurement, tap the checkbox to select it.

NOTE By default, the checkbox is cleared (not


enabled).

9.3.19 Operator ID Setting


NOTE When the USM function is disabled, the following setting will appear. In case that the USM is
enabled, refer to Section 11.3.
Configure whether to display the Operator ID at the Measurement-preparation screen and measure-
ment results screen.

To display the operator ID, tap the checkbox.

NOTE By default, the checkbox is cleared (not


enabled).
9

897N203029A FDC NX600 Instruction Manual 2022.01 9-37


9 Function Menu

9.3.20 Test Item Name Setting


You can configure the display of item names. The names appear in the screen display and printed
paper of measurement results.

1 Choose the item names to be used for display


and printing.
2 Choose the names to be used for external com-
munication output to the host computer.
The selection of [ALPi, LDHi]/[ALP, LDH] is the
1 display setting of the test name ALPi, LDHi.
This is not the display setting of the test name
ALP, LDH.
(ALPi and LDHi are slides compatible with the
2 IFCC method. Availability depends on the re-
gion of sale.)

NOTE If [ALP, LDH] is selected, ALPi, LDHi ap-


3 pears the same as ALP, LDH.
Check this setting again before testing.

3 Choose the item names to be used. Tap either


the [BUN] or [UREA] button.

NOTE When the name of BUN is changed to


UREA, or vice versa, set again the fol-
lowing items.
ITEMS
Reference interval
Correlation coefficient

9 Control fluid settings

NOTE When BUN/UREA is changed, the change


is reflected in the following items:
• Test Result
• Unit conversion
• Reference Interval setting
• Calculated Parameter
• Dilution factor Coefficients

NOTE Regarding the BUN ↔ UREA conversion


coefficient
UREA = BUN×2.14 [mg/dL]
UREA = BUN [mmol/L]

NOTE Rebooting the device is necessary for the setting to take effect. The confirmation dialog for
the power-off will appear when is tapped to exit Mode function for administrator. After the
shutdown, reboot the analyzer.

9-38 897N203029A FDC NX600 Instruction Manual 2022.01


9 Function Menu

9.3.21 Default Sample Type Setting


Set the default (initial setting) sample type.

Tap a sample type to select it.

9.3.22 Password Setting


When tapping the [Mode function for administrator] button on the Function menu screen, the pass-
word input may be required. The password can be changed here.

If the input field is tapped, the software keyboard


(refer to Section 2.3) appears so you can edit the
password.

NOTE A password can be 4 or 5 alphanumeric


characters long.

NOTE In the event of a password being lost,


please contact your dealer for assistance. 9

897N203029A FDC NX600 Instruction Manual 2022.01 9-39


10 Other Functions
10.1 Data Communication
The analyzer can transmit measurement results to a host computer or PC that has been approved
by IEC/UL60950-1 or IEC/UL62368-1.
To communicate, it is necessary to prepare the software and a cable to receive data. Before you use
this function for the first time, please contact your dealer for assistance.

NOTE Use a shielded LAN cable. Otherwise, electromagnetic interference may occur.

NOTE When a LAN or RS-232C cable is not connected, measurement results are not sent. Make
sure a cable is connected.

NOTE When connecting this analyzer to a network, make sure it is placed inside of the firewall to
avoid cybersecurity risks.

10

897N203029A FDC NX600 Instruction Manual 2022.01 10-1


11 USM (User Safety Mode)

11.1 Operator ID Input When Measurement Start and In Mode


Function
When measurement start button or function menu button is tapped, an Operator ID input display is
shown. Then, input Operator ID. If input operator ID is registered and also within expiry, the meas-
urement starts or the function menu becomes active.

NOTE When not registered ID/expired ID/Emer-


gency ID counted up to the maximum us-
age limit are input, measurement does not
start, nor mode function screen become ac-
tive.

11

897N203029A FDC NX600 Instruction Manual 2022.01 11-1


11 USM (User Safety Mode)

11.2 Difference of Operable Functions among Operator ID Level


(OPE,ADM,EMG)
Operable functions are different depending on operator ID level.

NOTE OPE: Operator


ADM: Administrator
EMG: Emergency

Measurement Result
OPE/ADM: When expired slide is used, EXP is indi-
cated instead of measurement result.
EXP is used not only on screen indica-
tion, but also on print and in communi-
cation.
This is also the case for CRP measure-
ment with expired calibration.

EMG: When expired slide is used in EMG,


measurement result with # mark is indi-
cated.

11

11-2 897N203029A FDC NX600 Instruction Manual 2022.01


11 USM (User Safety Mode)

Mode function
OPE/EMG: Operable function is limited to [Control],
[Calibration], [START Manual pipetting]
and some maintenance functions.

11

897N203029A FDC NX600 Instruction Manual 2022.01 11-3


11 USM (User Safety Mode)

ADM: All functions are operable.

11

11-4 897N203029A FDC NX600 Instruction Manual 2022.01


11 USM (User Safety Mode)

11.3 USM Menu


11.3.1 Operator ID -Import
1. Create Operator ID data and save it by use of Operator ID Editor.
* File name: dcn_uid_yyyymmddhhmmss.dat Ex) dcn_uid_20220425201612.dat
* For the detail, see the manual of the Operator ID Editor.

2. Create the following folder structure in your USB.


Rename the file which is created by the tool to [dcn_uid.dat],
Store it in the USB,
WData
└ GUI
└ dcn_uid.dat
*Please use FUJIFILM recommended USB only.

3. Select [Operator ID setting] in ADM


mode.
4. Select [Operator ID import].

11

897N203029A FDC NX600 Instruction Manual 2022.01 11-5


11 USM (User Safety Mode)

5. The left pop-up is shown. Then set USB,


and after a few seconds, select [OK].

6. The left pop-up will be shown when im-


port is complete. Take off the USB and
then tap [Shutdown].

11

11-6 897N203029A FDC NX600 Instruction Manual 2022.01


11 USM (User Safety Mode)

11.3.2 Operator ID -Export


1. Select [Operator ID setting] in ADM
mode.
2. Select [Operator ID export].

3. The left pop-up is shown. Then set USB,


and after a few seconds, select [OK].

11

897N203029A FDC NX600 Instruction Manual 2022.01 11-7


11 USM (User Safety Mode)

4. The left pop-up will be shown when ex-


port is complete. Then tap [OK], and take
off the USB.

5. Operator ID data, [dcn_uid.dat], is stored in below folder structure below in the


USB.
WData
└ GUI
└ dcn_uid.dat
* Browsing and editing of dcn_uid.dat is possible in Operator ID Editor.
For the detail, see the manual of the Operator ID Editor.

11

11-8 897N203029A FDC NX600 Instruction Manual 2022.01


11 USM (User Safety Mode)

11.3.3 Editing Registered Operator ID


Registered Operator ID can be edited as the following procedure.

NOTE Editable items are the expiry of ID and the level.

1. Select [Operator ID setting] in Mode


function for administrator screen.
2. Select [Operator ID setting].

3. Select an Operator ID which needs edit-


ing.

11

897N203029A FDC NX600 Instruction Manual 2022.01 11-9


11 USM (User Safety Mode)

4. To change the expiry, touch the expiry


field and edit it using the software key-
board.
5. To change the level, select either [ADM/
OPE].

11

11-10 897N203029A FDC NX600 Instruction Manual 2022.01


11 USM (User Safety Mode)

11.3.4 Editing Emergency ID


(Dis-)qualification of registered Emergency ID can be set.
Also measurement limit count by use of Emergency ID can be set.

1. Select [Operator ID setting] in Mode


function for administrator screen.
2. Select [Edit Emergency ID].

3. To (dis-)qualify the Emergency ID, tap


the checkbox.

11

897N203029A FDC NX600 Instruction Manual 2022.01 11-11


11 USM (User Safety Mode)

4. To edit the total limit count, select [Edit


emergency ID count].

5. To reset the current count, tap [Reset]


button.
To set a new limit count, select either [0],
[50] or [100].
In case of 0, emergency ID does not
function.

NOTE Measurement count by use of Emergency ID cannot exceed the limit count.

11 When the current count is up to the limit, the measurement by use of Emergency ID is not
possible.
In this case, reset the current count.

11-12 897N203029A FDC NX600 Instruction Manual 2022.01


12 Specifications
Processing speed 128 tests per hour (simultaneous CM and ISE measurements)
Number of incubator cells CM: 13; ISE: 1
Incubation temperature 37°C (CM) / 30°C (ISE)
Incubation time 1 to 6 minutes
Capacity for simultaneously 20
loaded slides
Slide lot information Each item: maximum 5 lots
Sampler
(1) Sample application fluid 50 µl or less (automatic switching)
volume
(2) Dilution Automatic dilution (maximum 21x dilution)
(3) PF filtering time Max. 90 sec.
Slide ejection Automatic ejection (max. 100 slides into the disposal box)
Used tip ejection Automatic ejection (max. 50 tips into the disposal box)
Measurement wavelengths 400 nm, 415 nm, 505 nm, 540 nm, 577 nm, 600 nm, 625 nm, 650 nm
Light source Halogen lamp (6 V, 10 W)
Measurement precision σ<
= 0.0004 OD / 5 minutes
Measurement accuracy Within ±0.005 OD (600 nm)
Display 7-inch color touch panel
Backup data of test results For 270 samples
Printer Thermal type (paper size: 58 mm x 25 m)
Warming up time Approx. 10 minutes / 25°C, approx. 20 minutes / 15°C
Environmental conditions Indoor use (6,000 cd/m2 (lux) or lower. 3,000 cd/m2 (lux) or lower
when using the sample barcode reader.)
Altitude: Up to 2,000m (810hPa)
Transient overvoltage category: II
Pollution degree: 2
Ingress protection rating: IP00 (No Protection)
Operating temperature: 15 to 32°C
Operating humidity: 30 to 80% RH (no vapor condensation)
Storage and transport conditions Temperature: -10 to 50°C
Humidity: 10 to 90% RH (no vapor condensation)
Voltage used Single phase AC; 100 - 240 V ±10%; 50 to 60 Hz
Consumption current 1.9 - 0.8A
Type of protection against electri- CLASS 1 EQUIPMENT
cal shock
Volume level Within 70 dB (including buzzer sound) 12
Data transfer RS-232C (1 port); USB (2 port); LAN (1 port)
External dimensions (W) 470× (D) 360× (H) 370 mm
Weight FUJI DRI-CHEM NX600: 28Kg
FUJI DRI-CHEM NX600i: 26Kg

897N203029A FDC NX600 Instruction Manual 2022.01 12-1


12 Specifications

Effective period of service 6 years after installation


(service life) (This period is valid as long as the precautions for use are followed
and regular periodic maintenance is performed correctly.)

12

12-2 897N203029A FDC NX600 Instruction Manual 2022.01


13 Glossary
Glossary of the display & printout messages

NOTE Because the display spaces and printouts are limited, some abbreviations are used.

NOTE Periods will not be used after abbreviations on the display and printout messages.

Abbreviation Meaning
AD Analog/Digital
ADM administrator
ave average
Bef comp Before compress
cali calibration
charac character
comm communication
ctrl control
DA Digital/Analog
Decomp Decompress
DL FUJI DRI-CHEM DILUENT DL (CRP)
EMG emergency
ERR error
EXP expiration
info information
NG No Good
OD Optical Density
OPE operator
Prob problem
USM User Safety Mode

13

897N203029A FDC NX600 Instruction Manual 2022.01 13-1


26-30, NISHIAZABU 2-CHOME, MINATO-KU, TOKYO 106-8620, JAPAN
https://www.fujifilm.com/

Heesenstrasse 31, 40549 Duesseldorf, Germany


EU

Oudenstaart 1, 5047 TK Tilburg, The Netherlands

You might also like