Professional Documents
Culture Documents
Air Purge Tool Manual 5086900A
Air Purge Tool Manual 5086900A
Air Purge Tool Manual 5086900A
Table of Contents
1.0 SAFETY INFORMATION .......................................................................................... 4
1.1 General ...................................................................................................................... 4
1.2 Radiation Protection.................................................................................................. 5
1.3 Electrical Safety ........................................................................................................ 5
1.4 Disposal..................................................................................................................... 5
1.5 Transportation ........................................................................................................... 5
2.0 INSTALLATION & MAINTENANCE ....................................................................... 6
2.1 Installation................................................................................................................. 6
2.2 Planned Maintenance – General ............................................................................... 6
3.0 Damage & Return ......................................................................................................... 6
3.1 Carrier Damage ......................................................................................................... 6
3.2 Return Process .......................................................................................................... 6
4.0 Contents ........................................................................................................................ 7
5.0 Instructions .................................................................................................................. 10
5.1 Procedure ................................................................................................................ 11
5.1.1 Preparation ....................................................................................................... 11
5.1.2 Checking Pump and Heat Exchanger for Contaminants.................................. 13
5.1.3 Prepare the Purge Tool..................................................................................... 14
5.1.4 Connect Purge Tool to Heat Exchanger and Pump Assemblies ...................... 16
5.1.5 Connect Purge to Tube..................................................................................... 19
5.1.6 Heat Exchanger Compensation ........................................................................ 20
5.1.7 Check Tube Alignments .................................................................................. 25
6.0 Cleanup ....................................................................................................................... 26
It was discovered that the quick disconnects for the oil cooling system have a tendency to
leak oil and/or to admit air into the cooling system. After a few X-ray tube changes (3-5
tubes), there could be enough air in the system to induce image artifacts. The VCT Air
Purge Tool is designed to remove any air within the cooling system and replace any lost
oil. We highly encourage (and even insist) that all tube changes on VCT scanners be
checked for air using the Oil flow Check Tool. The Air Purge Tool includes the Oil Flow
Check Tool. Failure to perform the Purge once air is identified within the system can and
will result in image artifacts and, with enough air, could damage the oil pump. A failed
oil pump results in overheating and possibly destruction of the X-ray tube. The Oil Flow
Check Tool also includes a second hose assembly to verify the oil flow rate.
On occasion, the X-ray tube assembly may experience an anode failure. Such a failure
may result in compromising the insert. The Oil Flow Check Tool will allow you to verify
if anode particles have been released into the oil and cooling system. This is a situation
where the complete cooling system consisting of the X-ray tube, oil pump, and heat
exchanger must be replaced.
For additional information, please contact one of our Technical Service Engineers at 1-
800-238-3780 or +1-630-585-2000.
Copyright
© 2010 DUNLEE
All rights reserved. Reproduction in whole or in part is prohibited without the prior written consent
of the copyright-holder. DUNLEE reserves the right to make changes in specifications or to
discontinue any product, at any time without notice or obligation, and is not liable for any
consequences resulting from the use of this publication.
Potential for Burns. Severe personal injury via burns can occur
from contact with hot oil. Do not service while tube and oil are
Potential for personal injury. Gantry oil pump should not be energized during air
removal process.
Purge procedure.
It is always the user who is responsible for compliance with the regulations applying to installation
and operation of the X-ray unit
• You must never use the X-ray Tube Housing Assembly or any associated diagnostic
imaging equipment if it has any electrical, mechanical or radiological defects. This
particularly applies to faulty indicators, displays, warnings and alarms.
• If the assembler wishes to connect this product to other equipment, components or
assemblies and if it is not apparent from the technical data whether it can be safely
combined with such equipment, components or assemblies, the user must ensure that
the safety of the patient, operating staff, third parties and the environment is not affected
by the planned combination by consulting the manufacturers involved or by making
inquires from an expert.
• Dunlee and Philips are responsible for the safety features of its products only if persons
explicitly authorized to do so by Dunlee or Philips have performed maintenance, repairs
and modifications.
• As with any technical appliance, this equipment requires - correct operation - regular,
competent maintenance - care
1.4 Disposal
• Take-back, proper disposal and recovery of this product must take place in accordance
with the European WEEE Directive (Waste Electrical and Electronic Equipment) and the
requirements of national legislation.
• Our state-of-the art X-ray equipment is manufactured according to safety and
environmental protection procedures. Under normal operation, there are no risks to
persons or the environment.
• To comply with regulations, it is necessary to use materials which may be harmful to the
environment and therefore have to be disposed of in a proper manner.
• Refer to the disposal section in this manual for details.
• CrossTrans 206 Oil does NOT contain Dioxin.
• Dunlee/Philips supports you in the recycling of the Air Purge Tool in accordance with
valid regulations and laws.
• Contact your Dunlee/Philips organization for proper handling and disposal.
1.5 Transportation
Extreme care must be taken when transporting the Air Purge Tool. It is recommended to transport
this product in the original packaging material. Transportation using other packaging material or in
any other way, may result in product damage
2.1 Installation
Only trained and qualified service personnel must perform installation and operation of the Air
Purge Tool. The installation must be performed in accordance with the system documentation
provided. Consult the appropriate documentation for the removal and installation procedure.
Drain Hose
with Male
Connector
12 VDC
Pump
Plastic
Funnel
Pump On/Off
rocker switch Oil outlet
pump –
female
disconnect
Oil/Air Return to Illustration 2: Oil Purge Tool
on blue hose
reservoir - male
disconnect on clear
hose
12 VDC
Pump
Reservoir
12 V, 120 W
50/60 Hz,
Single
output Reservoir
fill cap
Compensation
Tool
Clear Hose
Flow Meter
Hose
This procedure is intended for use on GE CT Scanners using the GE Performix Pro 100 or
Dunlee Reevo 240G X-ray tubes.
LightSpeed VCT
LightSpeed VCT Select
LightSpeed TR16
LightSpeed Xtra
Discovery VCT
LightSpeed Pro 16 (100 kW version only)
This document provides the necessary steps to remove air from the CT tube oil cooling system.
The CT tube oil cooling system is a closed system. The system consists of the heat exchanger,
tube oil pump, and the CT tube as its main components. In the event that air is introduced into the
system, Image Artifacts and/or tube over heating issues can occur.
2. Lower the table to the lowest position for front cover removal.
4. Power off the 120 VAC, Axial Drive Enable and HVDC Enable service switches.
5. Remove the Mylar window, covers from top, left side, and front of the gantry.
The front cover will need to be moved to the right side of the table so that the
electrical cables can be reconnected to the gantry. If there is no room for the front
cover to be placed on the right side of the table, the Gantry Display and Gantry
Control Panel will need to be put in the service positions. (See the Gantry Front
Cover Removal and Re-Install procedure located in the OEM service
documentation under Enclosure.)
7. Unbolt the collars that secure the quick disconnects to the tube.
9. Remove the hose bracket that secures the hose in a coiled position on the tube
housing.
3. Power on 120 VAC service switch, press reset on the Service Panel, and monitor
Clear Hose for air bubbles and particles circulating through hose. If no air
bubbles or particles are observed, tilt gantry to 30 degrees and monitor hose for
bubbles or particles.
5. Disconnect the Clear Hose from the heat exchanger and pump assemblies.
6. Disconnect the oil pump power connector, J52.
2. Open the Purge Tool, remove the reservoir cap, and place rags or paper towels
around the opening to capture any oil spilled during the filling process.
3. Place the funnel in the reservoir opening and fill reservoir with oil until at least
half full. (Put the reservoir cap in place to check oil level. Oil level should be at
least half full.)
5. Plug in the Purge Tool’s power cable into the service outlet on the gantry located
on the lower right side of the gantry. Power on the 120 VAC service switch on
the gantry and press the reset button on the Service Panel.
6. Once the gantry completes the boot up process, raise the table to a comfortable
working height. Do not advance the table into the gantry.
8. Once all of the air bubbles have disappeared from the hoses (about 2 minutes),
power off the Purge Tool and uncouple the quick disconnects.
3. With the Purge Tool’s power off, connect the Purge Tool’s hoses to the Pump and
Heat Exchanger assemblies (where the tube was connected).
4. Secure all of the hoses (loosely) to tube housing with cable ties to allow for
manual, unobstructed rotation of the gantry by hand.
7. Leave heat exchanger in the 12 o’clock position for about 2 minutes, longer if air
bubbles are observed.
8. Rotate the heat exchanger, clockwise, to the 4 o’clock position for about 2
minutes, longer if air bubbles are observed.
9. Rotate the heat exchanger, clockwise, to the 8 o’clock position for about 2
minutes, longer if air bubbles are observed.
10. Tilt gantry to 30 degrees away from table with heat exchanger in the 8 o’clock
position for about 2 minutes, longer if air bubbles are observed.
11. Rotate the heat exchanger, counter-clockwise (to keep hoses from purge tool
untangled), to the 4 o’clock position for about 2 minutes, longer if air bubbles are
observed.
12. Rotate the heat exchanger to the 12 o’clock position for about 2 minutes, longer if
air bubbles are observed.
13. Tilt gantry to 30 degrees towards the table with heat exchanger in the 12 o’clock
position for about 2 minutes, longer if air bubbles are observed.
14. Rotate the heat exchanger, clockwise, to the 4 o’clock position for about 2
minutes, longer if air bubbles are observed.
15. Rotate the heat exchanger, clockwise, to the 8 o’clock position for about 2
minutes, longer if air bubbles are observed.
16. Power off Purge Tool and disconnect hoses from Heat Exchanger and Pump
assemblies.
1. With the Purge Tool’s power off, connect the Purge Tool’s hoses to the Tube.
2. Check the oil level in the Purge Tool and add if necessary.
3. Secure all of the hoses (loosely) to tube housing with cable ties to allow for
manual, unobstructed rotation of the gantry by hand.
6. Leave tube in the 12 o’clock position for about 2 minutes, longer if air bubbles are
observed.
7. Rotate the tube, clockwise, to the 4 o’clock position for about 2 minutes, longer if
air bubbles are observed.
8. Rotate the tube, clockwise, to the 8 o’clock position for about 2 minutes, longer if
air bubbles are observed.
9. Rotate the tube, counter-clockwise, to the 4 o’clock position for about 2 minutes,
longer if air bubbles are observed.
10. Rotate the tube, counter-clockwise, to the 12 o’clock position for about 2 minutes,
longer if air bubbles are observed.
11. Power off the Purge tool and disconnect the hoses from the tube.
12. Reinstall the bracket to coil the Tube hose on the top rear of the tube housing.
4. Position the Compensation tool so that the valve is higher than the handle. Push
the oil/air back into the oil container. This will help to rid the cylinder of
unwanted air.
5. Repeat step 3, drawing oil into the tool to the mark on the cylinder. If air is still
present in the cylinder, repeat step 4 to minimize the amount of air.
8. Power off the Axial Drive Enable and HVDC Enable service switches.
12. Disconnect the Oil Pump hose from the Tube if not already done.
14. Push the handle on the Compensation tool to force oil into the tube. Once you
feel resistance in the handle, stop.
16. Remove the rubber stopper from the hose on the Compensation tool and place
hose into oil container.
17. Open valve on tool and push handle in to force remaining oil out of tool. Do not
pull handle out at this time.
18. Close valve and wipe away oil from hose. Put rubber stopper back in hose.
gradicles
Compensation Tool
22. Connect the Flow Rate Meter to the Tube and the Oil Pump.
29. Secure all cables and hoses on the gantry and check all plugs and connectors.
If overpressure errors occur during steps 2-5, more oil will need to
be removed from the tube. Repeat steps 1-21 and 28 of the Heat
Exchanger Compensation section and 1-5 of the Check Tube
Alignments section above. If no overpressure errors occur, proceed
with steps below.
6. Bolt the collars that secure the quick disconnects to the Tube.
7. Install covers, power on the 120 VAC, Axial Drive Enable, and HVDC Enable
service switches and press the reset button on the Service Panel.
Be sure to remove the old locking (tamper evident) band from the
container before screwing on the Red Dot Cap.
Once the Red Dot Caps are screwed on and locked on the containers, DO
NOT RE-OPEN them, as this will break the tamper evident band and
possibly lead to oil leakage during shipment.
END
Arlington Facility
2312 Avenue J
Arlington, Texas 76006 USA
Tel: 800.544.9729 (U.S. & Canada)
817.640.7666
Fax: 817.640.6644