Professional Documents
Culture Documents
Comparison of Impression Materials For Direct Multi-Implant Impressions
Comparison of Impression Materials For Direct Multi-Implant Impressions
CLINICAL IMPLICATIONS
Use of medium consistency polyether and high consistency addition silicones is recom-
mended to make a direct implant impression, depending on the amount of hard tissue
undercuts present in the arch. In this study, polyether minimized the chance of acci-
dental displacement of the direct impression coping when the abutment replicas were
tightened. Addition silicone in a partial edentulous arch facilitates the removal of the
impression tray when hard tissue undercuts are present, although care must be taken
to avoid accidental rotation of the impression coping. A double impression technique
using a lower consistency addition silicone wash did not appear to present any advan-
tage for use in direct implant impressions.
Fig. 2. Compudriver with attachment and slotted drive. Fig. 3. Slotted drive of Compudriver rotating square impres-
sion coping in implant impression.
Polyethers Impregum F (Premier Dental Products Co, Medium 28621 2%-3% 35-60 –0.24%
Norristown, Pa.) High 28621 3% 40-50 –0.19%
Silicones Low 7-1062 3-6 35 –0.15%
Addition Extrude (Kerr Mfg Co, Romulus, Mich.) Medium 7-1140 2%-5% 30 –0.17%
High 7-1125 2%-3% 60 –0.15%
High/Medium 7-1125/7-1140
High/Low 7-1125/7-1062
Very high 1%-2% 50-75 –0.14%
Condensation Elasticon (Kerr Mfg Co, Romulus, Mich.) Very low 7-1160 4%-9% 15-30 –0.60%
Very high 7-1160 2%-5% 50-65 –0.38%
Polysulfide Permlastic (Kerr Mfg Co, Romulus, Mich.) Medium 7-1259 11%-15% 30 –0.45%
High 9%-12% 35 –0.44%
*Italicized areas highlight impression materials that were tested for rotational torque displacement only. Bold areas highlight impression materials that were tested
for rotational torque displacement and for solid cast accuracy.
domly. Direct impression copings were hand-tightened ed Device Inc, City of Industry, Calif.) was attached to
to the master model by guide pins. Adhesive was paint- an adapter one-quarter drive (Craftsman, Hoffman
ed on the impression trays 24 hours before impressions Estate, Ill.) and a 5-6 bit slotted 3⁄8-in. driver (Crafts-
were made.33 Impressions were made on the master man) (Fig. 2). The 5-6 bit slotted driver was placed on
model and allowed to set for twice as long as the man- the slotted guide pin of the impression coping to be
ufacturers’ recommended setting times.34 tested, then the driver was rotated clockwise until the
impression coping was displaced once in the impression
Torque displacement
complex (Fig. 3). The Compudrive device recorded the
Sixteen impressions were made of the master model, amount of torque required to rotate the coping in the
namely, 2 impressions for each of the 8 impression impression through the slotted guide pin–abutment
materials/combinations evaluated (Table I, italicized analog complex. Torque values for 5 impression cop-
and bold-highlighted materials). Stainless steel abut- ings for each of the 16 impressions were measured.
ment replicas (DCA 174, Nobel Biocare USA Inc) Statistical analyses of torque data were performed
were secured carefully to the direct impression copings with SAS software (version 6.12, SAS Institute Inc,
with guide pins. Impressions were allowed to set for 30 Cary, N.C.).35 Repeated measures analysis of variance
minutes to simulate laboratory pouring time from (ANOVA) (α=.05) was used to evaluate torque values
impression taking. A Compudriver device (Consolidat- within and among impression material groups, as well
15 mm guide pins to the 5 direct impression copings in the steel balls and the reference ball C were calculated
each impression. Impressions were poured 30 minutes (Fig. 6). The corresponding linear distance of the mas-
after removal from the master model to simulate a clin- ter model, namely, CA, CB, and so forth, was then sub-
ical situation. As recommended in a previous study,14 tracted from each mean linear distance of the experi-
impressions were poured in Resin Rock material (Whip mental cast, with the absolute value taken of the result.
Mix Corp, Louisville, Ky.) that was vacuum mixed with The formula (CAexperimental – CAmaster = [A]) provides
distilled water, in accordance with the manufacturer’s an illustration for the derivation of an experimental
instructions. Experimental implant casts were allowed cast’s CA measurement to a measurement of [A], or
to set for 1 hour before the guide pins were unscrewed the absolute micron deviation of the experimental cast
and the impression was removed. One experimental abutment from the master model. The absolute micron
cast was poured from each impression producing 10 deviation for each cast’s 4 linear measurements, that is
casts per group. Any debris remaining on the abutment [A], [B], [D], [E], were then averaged for each cast,
replicas was removed. All experimental implant casts resulting in a cast’s mean absolute cast error. Absolute
were numbered and stored in ambient conditions for at micron deviations of all casts’ linear measurements
least 24 hours.36 were compared statistically using a repeated measures
Fig. 5. Diagram of traveling microscope setup with implant cast and steel ball on abutment
replica.
ANOVA (α=.05) to evaluate within-group, between- Table II. Ryan-Einot-Gabriel-Welsch multiple range test
group differences, and the variables’ interaction effect. (REGWQ) groupings (P>.05) for torque measures
As illustrated, positive and negative linear distances Mean SD REGWQ
were used to calculate [A]; however, only absolute Material N (×10–3 mN) (×10–3 mN) groupings
(positive) values were compared for statistical analysis. Polyether 2 141.3 0.4 A
A 1-way ANOVA (α=.05) was then used to evaluate Addition silicone 2 71 6.4 B
the significance among the material groups’ mean Polysulfide 2 51.5 7 C
absolute cast errors. Finally, a post hoc using REGWQ
was used to rank the material groups’ mean absolute
cast errors.
A measurement error (SD) of 3 µm was computed groups’ torque values (P=.1866), and a significant
by measuring the distance between abutment replicas A interaction effect was not identified between impres-
and E on the master cast 10 times. The SAS statistical sion coping position and impression material
program35 was used to compute the statistical analyses. (P=.4574). Notably, mean torque values differed sig-
nificantly among impression material groups (P=.001),
RESULTS
with the highest overall torque values identified for
Of the impression materials tested (Table I), torque polyether (medium), followed by addition silicone
values were detected for only 3: polyether (medium), (high), and then polysulfide (medium) (Table II).
addition silicones (high), and polysulfides (medium) Similarly, regarding measures of accuracy within-
(Fig. 7). Impression coping position did not produce a group absolute micron deviations ([A], [B], [D], [E])
statistically significant difference within material did not reveal significant variation (P=.6881), and no
square Nobel Biocare impression coping in an implant point of reference from which the distortion is mea-
impression. Results of our arbitrary selection criteria sured.42 Barrett et al15 used absolute distortion analy-
coincide with impression materials that will most likely sis, whereby the point of reference is an external refer-
be used for direct square implant impressions. For ence point, namely, not include the impression copings.
instance, polyether is recommended repeatedly in the In contrast, this study, and many other distortion stud-
dental literature.12,18-22,25,28-31 Addition silicones are ies, used relative distortion analysis, whereby one of the
recommended as well,15,16,32 and are the preferred abutment replicas/impression copings is used as the
fixed prosthodontic impression material as taught in reference to which the distortion of the other replicas/
US dental schools (70% or 42/53 schools).38 Polysul- impression copings are measured.16,17,21,43 It may be
fide, although not mentioned in the implant literature, argued that the use of relative distortion analysis
is the preferred removable prosthodontic impression provides more clinically relevant data than absolute dis-
material as taught in US dental schools (48% or 47/54 tortion analysis. The implant prosthesis connects all
schools).39 abutments together. Therefore, the amount of strain in
From a clinical standpoint, results of this study sup- the implant prosthetic–implant bone system is related
port the use of polyether for completely edentulous to the relative position of the implant abutments to one
multi-implant impressions. The rigidity of polyether another and not to an external reference point.
provides resistance to the accidental displacement of Another methodologic difference requires com-
the impression coping in the implant impression. How- ment. The study by Barrett et al15 evaluated the accu-
ever, use of polyether for an impression of a partially racy of the implant impression itself, as did Phillips et
edentulous arch also presents increased difficulty for al.21 Using this method requires that follow-up studies
intraoral removal of the impression. High consistency be completed to determine the resultant accuracy of
addition silicones and medium consistency polysulfides the implant cast, with the appropriate die material, to
are viable alternative materials of choice for experienced provide data from a clinically relevant end product,
practitioners. Addition silicone with its more favorable such as the solid implant cast. Evaluating the accuracy
modulus of elasticity (rigidity)40 allows easy removal of of the implant impression with reference to a resulting
the set impression. solid implant cast, as completed in this study, eliminates
Other suggested material combinations that were the need for follow-up studies and has become a stan-
not tested in this study include the use of very high dard method across the majority of implant impression
consistency or putty and low consistency wash, as test- accuracy studies.12,16-20,25,43
ed by Barrett et al,15 and the use of adhesive surround- This study also evaluated the accuracy of the relative
ing the impression copings. This concept of using a position between the implant abutments through the use
high and low consistency impression material is bor- of steel balls on the abutment replicas, similar to Carr and
rowed from techniques used in fixed prosthodontics. Master’s method.43 The purpose of this study was not to
The low consistency impression material is syringed evaluate the implant abutment-to-framework relation-
into the sulcus of the prepared tooth, whereas high ship; this study only evaluated the resultant translational
consistency material helps force the low consistency distortion (x-, y-, and z-axes) of the abutment replicas to
material into the sulcus.41 The lack of detectable one another, and not more complex rotational distortion
torque when using high/medium or high/low consis- that occurs in implant prosthodontics.2-6,21,22,45-47 The
tency impression material shows no advantage in using significant difference detected among impression materi-
it for direct implant impression. The design of the al groups indicates that the method of this study was
implant impression coping may also be a relevant factor appropriately sensitive.
to consider. However, the design of most implant To validate this study, a long-term prospective clinical
impression copings is not intricate enough to require study would have to be performed with preliminary data
that a low consistency impression material be syringed available from this study. Patients requiring a particular
around them. type of implant retained prosthesis would be divided
Addition silicones and polyether for direct multi- into 2 groups. The patient’s definitive implant prosthe-
implant impressions for edentulous arches produce sis would be fabricated to fit their master cast accurately
similarly accurate solid implant casts, results consistent using either “laser welding of titanium”7,8 or “electric
with those recorded previously for an indirect implant discharge machining”9-11 method. For 1 group, the
impression technique by Barrett et al.15 The external master cast would be made from a direct implant impres-
validity of this conclusion, although limited to these 2 sion from addition silicone (high) or polyether (medi-
studies, is noteworthy, given the significantly different um), and for the other group, the impression would be
methods of distortion measurement and analysis made from polysulfide (medium). The fit of the pros-
between the 2 studies. thesis would not be adjusted intraorally for every stage
To elaborate, distortion can be defined and mea- of the prosthesis fabrication procedure. The 2 groups
sured as “absolute” or “relative,” depending on the would then be evaluated longitudinally.
CONCLUSION 14. Wee AG, Schneider RL, Aquilino SA, Huff TL, Lindquist TJ, Williamson
DL. Evaluation of the accuracy of solid implant casts. J Prosthodont
The properties of an impression material, including 1998;7:161-9.
rigidity and accuracy, can influence the accuracy of the 15. Barrett MG, de Rijk WG, Burgess JO. The accuracy of six impression tech-
implant impression, the accuracy of the solid implant niques for osseointegrated implants. J Prosthodont 1993;2:75-82.
16. Humphries RM, Yaman P, Bloem TJ. The accuracy of implant master casts
cast, and ultimately, the accuracy of the cast implant constructed from transfer impressions. Int J Oral Maxillofac Implants
framework. Choosing an impression material for multi- 1990;5:331-6.
implant–retained prosthesis requires consideration of sev- 17. Spector M, Donovan TE, Nicholls JI. An evaluation of impression tech-
niques for osseointegrated implants. J Prosthet Dent 1990;63:444-7.
eral factors, including material accuracy, clinician’s expe- 18. Carr AB. Comparison of impression techniques for a five-implant
rience with material, length of time before the impression mandibular model. Int J Oral Maxillofac Implants 1991;6:448-55.
is poured, and amount of intraoral undercuts. 19. Carr AB. Comparison of impression techniques for a two-implant 15-
degree divergent model. Int J Oral Maxillofac Implants 1992;7:468-75.
Within the limitations of this study, torque required
20. Assif D, Fenton AH, Zarb GA, Schmitt A. Comparative accuracy of
to rotate an impression coping in the impression was implant impression procedures. Int J Periodontics Restorative Dent 1992;
significantly different (in descending order) from each 12:113-21.
other: polyether (medium), addition silicone (high), 21. Phillips KM, Nicholls JI, Ma T, Rubenstein JE. The accuracy of three
implant impression techniques: a 3-dimensional analysis. Int J Oral Max-
and polysulfide (medium). Implant cast made from illofac Implants 1994;9:533-40.
polyether (medium) or addition silicone (high) was sig- 22. Assif D, Marshak B, Schmidt A. Accuracy of implant impression tech-
nificantly different from cast made from polysulfide niques. Int J Oral Maxillofac Implants 1996;11:216-22.
23. Finger IM, Guerra LR. Prosthetic consideration in reconstructive implan-
(medium). New practitioners to the specialty of tology. Dent Clin North Am 1986;30:69-83.
prosthodontics may prefer using polyether (medium), 24. Rasmussen EJ. Alternative prosthodontic technique for tissue-integrated
given its greater overall rigidity. The type of material prostheses. J Prosthet Dent 1987;57:198-204.
25. Inturregui JA, Aquilino SA, Ryther JS, Lund PS. Evaluation of three impres-
used for either the completely or partially edentulous sion techniques for osseointegrated oral implants. J Prosthet Dent 1993;
multi-implant impression will depend on which factors 69:503-9.
are important to the practitioner. 26. Burawi G, Houston F, Byme D, Claffey N. A comparison of the dimen-
sional accuracy of the splinted and unsplinted impression techniques for
I would like to thank Drs Steve Rosenstiel and Alan Carr for their Bone-Lock implant system. J Prosthet Dent 1997;77:68-75.
encouragement and review of the manuscript. Thanks also go to Dr 27. Liou AD, Nicholls JI, Yuodelis RA, Brudvik JS. Accuracy of replacing three
William Johnston for the statistical assistance and Dr Peter tapered transfer impression copings into two elastomeric impression
Monaghan for his suggestions regarding the measuring technique. materials. Int J Prosthodont 1993;6:377-83.
28. Tautin FS. Impression making for osseointegrated dentures. J Prosthet
Dent 1985;54:250-1.
REFERENCES 29. Zarb GA, Jansson TP. Prosthodontic procedures and laboratory proce-
dures and protocol. In: Brånemark PI, Zarb GA, Albrektsson L, editors. Tis-
1. Jemt T, Carlsson L, Boss A, Jorneus L. In vivo load measurements on
sue integrated prostheses. Chicago: Quintessence; 1985. p. 241-82, 293-
osseointegrated implants supporting fixed or removable prostheses: a
315, 233-40, 317-27.
comparative pilot study. Int J Oral Maxillofac Implants 1991;6:413-7.
30. Taylor R, Bergman G. Laboratory techniques for the Brånemark system.
2. Jemt T. Three-dimensional distortion of gold-alloy castings and welded
Chicago: Quintessence; 1990.
titanium frameworks: measurements of the precision of fit between com-
31. Goll GE. Production of accurately fitting full-arch implant frameworks:
pleted implant prostheses and the master cast in routine edentulous situ-
part I—clinical procedures. J Prosthet Dent 1991;66:377-84.
ations. J Oral Rehabil 1995;22:557-64.
3. Jemt T. In vivo measurements of precision of fit involving implant-sup- 32. Schneider RL, Wee AG. Fabricating low-fusing metal casts for more accu-
ported prostheses in the edentulous jaw. Int J Oral Maxillfac Implants rate implant prosthodontics. J Prosthodont 1996;5:301-3.
1996;11:151-8. 33. Dixon DL, Breeding LC, Brown JS. The effect of custom tray material type
4. Tan KB. The clinical significance of distortion in implant prosthodontics: and adhesive drying time on the tensile bond strength of an impression
is there such a thing as passive fit? Ann Acad Med Singapore 1995;24: material/adhesive system. Int J Prosthodont 1994;7:129-33.
138-57. 34. American Dental Association. Revised American Dental Association
5. Wee AG, Aquilino SA, Schneider RL. Strategies to achieve fit in implant specification No: 19 for non-aqueous, elastomeric dental impression
prosthodontics: a review of the literature. Int J Prosthodont 1999;12:167- materials. J Am Dent Assoc 1977;94:733-41.
78. 35. SAS Institute 1. SAS User’s Guide: Basic. 5 ed. Cary, NC: SAS Institute Inc;
6. Jemt T, Rubenstein JE, Carlsson L, Lang BR. Measuring fit at the implant 1985.
prosthodontic interface. J Prosthet Dent 1996;75:314-24. 36. Arlo K. Evaluation of the Vident die and model system. Report No.: Pro-
7. Riedy SJ, Lang BF, Lang BE. Fit of implant framework fabricated by differ- ject 90-63. San Antonio: USAF Dental Investigation Services; 1991.
ent techniques. J Prosthet Dent 1997;78:596-604. 37. Nicholls JI. The measurement of distortion: concluding remarks. J Prosthet
8. Jemt T, Linden B. Fixed implant supported prosthesis with welded titani- Dent 1980;43:218-23.
um frameworks. Int J Periodontics Restorative Dent 1992;12:177-83. 38. Petropoulos VC, Weintraub A, Weintraub GS. Predoctoral fixed prostho-
9. Linehan AD, Windeler AS. Passive fit of implant-retained prosthetic frame- dontics curriculum survey. J Prosthodont 1998;7:183-91.
work improved by electric discharge machining. J Prosthodont 1994;3: 39. Arbree NS, Fleck F, Askinas SW. The results of a brief survey of complete
88-95. denture prosthodontic techniques in predoctoral programs in North
10. Schmitt SM, Chance DA. Fabrication of titanium implant-retained restora- American dental schools. J Prosthodont 1996;5:219-25.
tions with nontraditional machining techniques. Int J Prosthodont 1995;8: 40. Chai J, Takahashi Y, Lautenschlager EP. Clinically relevant mechanical
332-6. properties of elastomeric impression materials. Int J Prosthodont 1998;11:
11. LaBarge KW. Electrical discharge machining. J Dent Technol 1997;14:19- 219-23.
22. 41. Rosenstiel SF, Land MF, Fujimoto J. Tissue management and impression
12. Hsu CC, Millstein PL, Stein RS. A comparative analysis of the accuracy of making. In: Rosenstiel SF, Land MF, Fujimoto J, editors. Contemporary
implant transfer techniques. J Prosthet Dent 1993;69:588-93. fixed prosthodontics. 2nd ed. St Louis: Mosby; 1995. p. 314-7.
13. Vigolo P, Millstein PL. Evaluation of master cast technique for multiple 42. Nicholls JI. The measurement of distortion: theoretical considerations. J
abutment implant prostheses. Int J Oral Maxillofac Implants 1993;8:439- Prosthet Dent 1977;37:578-86.
45. 43. Carr AB, Master J. The accuracy of implant verification casts compared
with casts produced from a rigid transfer coping technique. J Prosthodont Reprint requests to:
1996;5:248-52. DR ALVIN G. WEE
44. Skalak R. Biomechanical considerations in osseointegrated prostheses. J SECTION OF RESTORATIVE DENTISTRY, PROSTHODONTICS AND
Prosthet Dent 1983;49:843-8. ENDODONTICS
45. Tan KB, Rubenstein JE, Nicholls JI, Yuodelis RA. Three dimensional analy- COLLEGE OF DENTISTRY
sis of the casting accuracy of one-piece, osseointegrated implant retained THE OHIO STATE UNIVERSITY
prosthesis. Int J Prosthodont 1993;6:346-63. 305 W 12TH AVE
46. Ness EM, Nicholls JI, Rubenstein JE, Smith DE. Accuracy of the acrylic COLUMBUS, OH 43210-1241
pattern resin for the implant-retained prosthesis. Int J Prosthodont 1992; FAX: 614-292-9422
5:542-9. E-MAIL: Wee.12@osu.edu
47. Jemt T, Lie A. Accuracy of implant-supported prostheses in the edentulous
jaw: analysis of precision of fit between cast gold-alloy frameworks and Copyright © 2000 by The Editorial Council of The Journal of Prosthetic
master casts by means of a three-dimensional photogrammetric tech- Dentistry.
nique. Clin Oral Implants Res 1995;6:172-80. 0022-3913/2000/$12.00 + 0. 10/1/104915
48. Craig RG. Restorative dental materials. 10th ed. St Louis: Mosby; 1997. doi:10.1067/mpr.2000.104915
Purpose. This article describes the results of the treatment of 50 persons with a “Dahl”-type
appliance at the Eastman Dental Hospital, London, U.K., from 1981 to 1994. Studies by Dahl
have shown that the placement of a “partial bite raising appliance” on the maxillary anterior teeth
may be used to create localized occlusal space before restoration of teeth. The space being pro-
duced by the teeth in contact with the appliance being intruded and those out of contact extrud-
ing. The aim of this retrospective clinical audit was to assess the outcome of using similar appli-
ances placed in all regions of the mouth and the factors that may influence the outcome of treat-
ment.
Material and methods. Fifty appliances were used in 45 patients whose age ranged from 20 to
70 years (median 37 years of age); 54% of patients were women. Appliances were used in the fol-
lowing situations: attrition (19%-38%); erosion (11%-22%); total tooth wear (30%-60%); extrusion
(17%-34%); iatrogenic (2%-4%); and lack of space after adult orthodontic treatment (1%-2%). Sev-
enty-eight percent were cemented to teeth and 22% were removable. In addition, 64% were
placed in the anterior arch (32% in the posterior) and 76% were placed on the maxillary arch (24%
on the mandibular arch). Duration of treatment ranged from 1 month to 24 months, with a mean
treatment of 5.9 months. After treatment, teeth were restored with definitive restorations, includ-
ing conventional crown, fixed partial dentures, resin-retained fixed partial dentures, or hybrid
fixed prostheses.
Results. Success rate of treatment was high (96%), with only 1 appliance in place that produced
no useful tooth movement. All appliances produced enough space for restorations to be placed
conservatively when given enough time to act.
Conclusion. This type of treatment provided a conservative, predictable, and effective method to
general localized occlusal space before prosthetic restoration. 11 References. —RP Renner