Download as pdf or txt
Download as pdf or txt
You are on page 1of 68

HHS

Control and Tube Assembly Tests

46-017561
Revision 37
US English
General service documentation.
© 2023 GE HealthCare. GE is a trademark of General Electric Company used under trademark license.
Reproduction and/or distribution is prohibited.
Control and Tube Assembly Tests

Important...X-Ray Protection
WARNING

X-ray equipment if not properly used may cause injury. Accordingly, the instructions
herein contained should be thoroughly read and understood by everyone who will
use the equipment before you attempt to place this equipment in operation. GE
HealthCare will be glad to assist and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation
other than the useful beam, no practical design of equipment can provide complete
protection. Nor can any practical design compel the operator to take adequate
precautions to prevent the possibility of any persons carelessly exposing themselves
or others to radiation.
It is important that anyone having anything to do with x-radiation be properly
trained and fully acquainted with the recommendations of the National Council on
Radiation Protection and Measurements as published in NCRP Reports available
from NCRP Publications, 7910 Woodmont Avenue, Room 1016, Bethesda, Maryland
20814, and of the International Commission on Radiation Protection, and take
adequate steps to protect against injury.
The equipment is sold with the understanding that GE HealthCare, its agents, and
representatives have no responsibility for injury or damage which may result from
improper use of the equipment.
Various protective materials and devices are available. It is urged that such materials
or devices be used.

46-017561 Revision 37 HHS 3/68


Control and Tube Assembly Tests

Language Policy
DOC0371395 - Global Language Procedure

ПРЕДУПРЕЖ Това упътване за работа е налично само на английски език.


ДЕНИЕ • Ако доставчикът на услугата на клиента изиска друг език, задължение на клиента е да
(BG) осигури превод.
• Не използвайте оборудването, преди да сте се консултирали и разбрали упътването за
работа.
• Неспазването на това предупреждение може да доведе до нараняване на доставчика
на услугата, оператора или пациентa в резултат на токов удар, механична или друга
опасност.
警告 本维修手册仅提供英文版本。
(ZH-CN) • 如果客户的维修服务人员需要非英文版本,则客户需自行提供翻译服务。
• 未详细阅读和完全理解本维修手册之前,不得进行维修。
• 忽略本警告可能对维修服务人员、操作人员或患者造成电击、机械伤害或其他形式的伤
害。
警告 本服務手冊僅提供英文版本。
(ZH-HK) • 倘若客戶的服務供應商需要英文以外之服務手冊,客戶有責任提供翻譯服務。
• 除非已參閱本服務手冊及明白其內容,否則切勿嘗試維修設備。
• 不遵從本警告或會令服務供應商、網絡供應商或病人受到觸電、機械性或其他的危險。
警告 本維修手冊僅有英文版。
(ZH-TW) • 若客戶的維修廠商需要英文版以外的語言,應由客戶自行提供翻譯服務。
• 請勿試圖維修本設備,除非 您已查閱並瞭解本維修手冊。
• 若未留意本警告,可能導致維修廠商、操作員或病患因觸電、機械或其他危險而受傷。
UPOZOR- Ovaj servisni priručnik dostupan je na engleskom jeziku.
ENJE • Ako davatelj usluge klijenta treba neki drugi jezik, klijent je dužan osigurati prijevod.
(HR) • Ne pokušavajte servisirati opremu ako niste u potpunosti pročitali i razumjeli ovaj servisni
priručnik.
• Zanemarite li ovo upozorenje, može doći do ozljede davatelja usluge, operatera ili pacijenta
uslijed strujnog udara, mehaničkih ili drugih rizika.
VÝSTRAHA Tento provozní návod existuje pouze v anglickém jazyce.
(CS) • V případě, že externí služba zákazníkům potřebuje návod v jiném jazyce, je zajištění překladu
do odpovídajícího jazyka úkolem zákazníka.
• Nesnažte se o údržbu tohoto zařízení, aniž byste si přečetli tento provozní návod a pochopili
jeho obsah.
• V případě nedodržování této výstrahy může dojít k poranění pracovníka prodejního servisu,
obslužného personálu nebo pacientů vlivem elektrického proudu, respektive vlivem mechan-
ických či jiných rizik.
ADVARSEL Denne servicemanual findes kun på engelsk.
(DA) • Hvis en kundes tekniker har brug for et andet sprog end engelsk, er det kundens ansvar at
sørge for oversættelse.
• Forsøg ikke at servicere udstyret uden at læse og forstå denne servicemanual.
• Manglende overholdelse af denne advarsel kan medføre skade på grund af elektrisk stød,
mekanisk eller anden fare for teknikeren, operatøren eller patienten.

4/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests

WAAR- Deze onderhoudshandleiding is enkel in het Engels verkrijgbaar.


SCHUWING • Als het onderhoudspersoneel een andere taal vereist, dan is de klant verantwoordelijk voor
(NL) de vertaling ervan.
• Probeer de apparatuur niet te onderhouden alvorens deze onderhoudshandleiding werd
geraadpleegd en begrepen is.
• Indien deze waarschuwing niet wordt opgevolgd, zou het onderhoudspersoneel, de operator
of een patiënt gewond kunnen raken als gevolg van een elektrische schok, mechanische of
andere gevaren.
WARNING This service manual is available in English only.
(EN) • If a customer's service provider requires a language other than English, it is the customer's
responsibility to provide translation services.
• Do not attempt to service the equipment unless this service manual has been consulted and
is understood.
• Failure to heed this warning may result in injury to the service provider, operator or patient
from electric shock, mechanical or other hazards.
HOIATUS See teenindusjuhend on saadaval ainult inglise keeles.
(ET) • Kui klienditeeninduse osutaja nõuab juhendit inglise keelest erinevas keeles, vastutab klient
tõlketeenuse osutamise eest.
• Ärge üritage seadmeid teenindada enne eelnevalt käesoleva teenindusjuhendiga tutvumist ja
sellest aru saamist.
• Käesoleva hoiatuse eiramine võib põhjustada teenuseosutaja, operaatori või patsiendi vigas-
tamist elektrilöögi, mehaanilise või muu ohu tagajärjel.
VAROITUS Tämä huolto-ohje on saatavilla vain englanniksi.
(FI) • Jos asiakkaan huoltohenkilöstö vaatii muuta kuin englanninkielistä materiaalia, tarvittavan
käännöksen hankkiminen on asiakkaan vastuulla.
• Älä yritä korjata laitteistoa ennen kuin olet varmasti lukenut ja ymmärtänyt tämän huolto-
ohjeen.
• Mikäli tätä varoitusta ei noudateta, seurauksena voi olla huoltohenkilöstön, laitteiston käyt-
täjän tai potilaan vahingoittuminen sähköiskun, mekaanisen vian tai muun vaaratilanteen
vuoksi.
ATTENTION Ce manuel d'installation et de maintenance est disponible uniquement en anglais.
(FR) • Si le technicien d'un client a besoin de ce manuel dans une langue autre que l'anglais, il
incombe au client de le faire traduire.
• Ne pas tenter d'intervenir sur les équipements tant que ce manuel d'installation et de mainte-
nance n'a pas été consulté et compris.
• Le non-respect de cet avertissement peut entraîner chez le technicien, l'opérateur ou le
patient des blessures dues à des dangers électriques, mécaniques ou autres.
WARNUNG Diese Serviceanleitung existiert nur in englischer Sprache.
(DE) • Falls ein fremder Kundendienst eine andere Sprache benötigt, ist es Aufgabe des Kunden für
eine entsprechende Übersetzung zu sorgen.
• Versuchen Sie nicht diese Anlage zu warten, ohne diese Serviceanleitung gelesen und verstan-
den zu haben.
• Wird diese Warnung nicht beachtet, so kann es zu Verletzungen des Kundendiensttechnikers,
des Bedieners oder des Patienten durch Stromschläge, mechanische oder sonstige Gefahren
kommen.
ΠΡΟΕΙΔΟΠΟΙ Τοπαρόν εγχειρίδιο σέρβις διατίθεται στα αγγλικά μόνο.
ΗΣΗ • Εάν το άτομο παροχής σέρβις ενός πελάτη απαιτεί το παρόν εγχειρίδιο σε γλώσσα εκτός των
(EL) αγγλικών, αποτελεί ευθύνη του πελάτη να παρέχει υπηρεσίες μετάφρασης.
• Μηνεπιχειρήσετε την εκτέλεση εργασιών σέρβις στον εξοπλισμό εκτός εάν έχετε
συμβουλευτεί και έχετε κατανοήσει το παρόν εγχειρίδιο σέρβις.
• Εάν δεν λάβετε υπόψη την προειδοποίηση αυτή, ενδέχεται να προκληθεί τραυματισμός στο
άτομο παροχής σέρβις, στο χειριστή ή στον ασθενή από ηλεκτροπληξία, μηχανικούς ή άλλους
κινδύνους.

46-017561 Revision 37 HHS 5/68


Control and Tube Assembly Tests

FIGYELMEZ- Ezen karbantartási kézikönyv kizárólag angol nyelven érhető el.


TETÉS • Ha a vevő szolgáltatója angoltól eltérő nyelvre tart igényt, akkor a vevő felelőssége a fordítás
(HU) elkészíttetése.
• Ne próbálja elkezdeni használni a berendezést, amíg a karbantartási kézikönyvben leírtakat
nem értelmezték.
• Ezen figyelmeztetés figyelmen kívül hagyása a szolgáltató, működtető vagy a beteg áramütés,
mechanikai vagy egyéb veszélyhelyzet miatti sérülését eredményezheti.
AÐVÖRUN Þessi þjónustuhandbók er aðeins fáanleg á ensku.
(IS) • Ef að þjónustuveitandi viðskiptamanns þarfnast annas tungumáls en ensku, er það skylda
viðskiptamanns að skaffa tungumálaþjónustu.
• Reynið ekki að afgreiða tækið nema að þessi þjónustuhandbók hefur verið skoðuð og skilin.
• Brot á sinna þessari aðvörun getur leitt til meiðsla á þjónustuveitanda, stjórnanda eða sjú-
klings frá raflosti, vélrænu eða öðrum áhættum.
AVVERTENZA Il presente manuale di manutenzione è disponibile soltanto in lingua inglese.
(IT) • Se un addetto alla manutenzione richiede il manuale in una lingua diversa, il cliente è tenuto
a provvedere direttamente alla traduzione.
• Procedere alla manutenzione dell'apparecchiatura solo dopo aver consultato il presente
manuale ed averne compreso il contenuto.
• Il mancato rispetto della presente avvertenza potrebbe causare lesioni all'addetto alla man-
utenzione, all'operatore o ai pazienti provocate da scosse elettriche, urti meccanici o altri
rischi.
警告 このサービスマニュアルには英語版しかありません。
(JA) • サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業はその業者
の責任で行うものとさせていただきます。
• このサービスマニュアルを熟読し理解せずに、装置のサービスを行わないでくださ
い。
• この警告に従わない場合、サービスを担当される方、操作員あるいは患者 さんが、感
電や機械的又はその他の危険により負傷する可能性があります。
경고 본 서비스 매뉴얼은 영어로만 이용하실 수 있습니다.
(KO) • 고객의 서비스 제공자가 영어 이외의 언어를 요구할 경우, 번역 서비스를 제공하는 것은
고객의 책임입니다.
• 본 서비스 매뉴얼을 참조하여 숙지하지 않은 이상 해당 장비를 수리하려고 시도하지 마십
시오.
• 본 경고 사항에 유의하지 않으면 전기 쇼크, 기계적 위험, 또는 기타 위험으로 인해 서비스
제공자, 사용자 또는 환자에게 부상을 입힐 수 있습니다.
BRĪDINĀ- Šī apkopes rokasgrāmata ir pieejama tikai angļu valodā.
JUMS • Ja klienta apkopes sniedzējam nepieciešama informācija citā valodā, klienta pienākums ir
(LV) nodrošināt tulkojumu.
• Neveiciet aprīkojuma apkopi bez apkopes rokasgrāmatas izlasīšanas un saprašanas.
• Šī brīdinājuma neievērošanas rezultātā var rasties elektriskās strāvas trieciena, mehānisku vai
citu faktoru izraisītu traumu risks apkopes sniedzējam, operatoram vai pacientam.
ĮSPĖJIMAS Šis eksploatavimo vadovas yra tik anglų kalba.
(LT) • Jei kliento paslaugų tiekėjas reikalauja vadovo kita kalba – ne anglų, suteikti vertimo paslau-
gas privalo klientas.
• Nemėginkite atlikti įrangos techninės priežiūros, jei neperskaitėte ar nesupratote šio eksploa-
tavimo vadovo.
• Jei nepaisysite šio įspėjimo, galimi paslaugų tiekėjo, operatoriaus ar paciento sužalojimai dėl
elektros šoko, mechaninių ar kitų pavojų.

6/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests

ADVARSEL Denne servicehåndboken finnes bare på engelsk.


(NO) • Hvis kundens serviceleverandør har bruk for et annet språk, er det kundens ansvar å sørge for
oversettelse.
• Ikke forsøk å reparere utstyret uten at denne servicehåndboken er lest og forstått.
• Manglende hensyn til denne advarselen kan føre til at serviceleverandøren, operatøren eller
pasienten skades på grunn av elektrisk støt, mekaniske eller andre farer.
OSTRZEŻE- Niniejszy podręcznik serwisowy dostępny jest jedynie w języku angielskim.
NIE • Jeśli serwisant klienta wymaga języka innego niż angielski, zapewnienie usługi tłumaczenia
(PL) jest obowiązkiem klienta.
• Nie próbować serwisować urządzenia bez zapoznania się z niniejszym podręcznikiem serwi-
sowym i zrozumienia go.
• Niezastosowanie się do tego ostrzeżenia może doprowadzić do obrażeń serwisanta, operatora
lub pacjenta w wyniku porażenia prądem elektrycznym, zagrożenia mechanicznego bądź
innego.
ATENÇÃO Este manual de assistência técnica encontra-se disponível unicamente em inglês.
(PT-BR) • Se outro serviço de assistência técnica solicitar a tradução deste manual, caberá ao cliente
fornecer os serviços de tradução.
• Não tente reparar o equipamento sem ter consultado e compreendido este manual de assis-
tência técnica.
• A não observância deste aviso pode ocasionar ferimentos no técnico, operador ou paciente
decorrentes de choques elétricos, mecânicos ou outros.
ATENÇÃO Este manual de assistência técnica só se encontra disponível em inglês.
(PT-PT) • Se qualquer outro serviço de assistência técnica solicitar este manual noutro idioma, é da
responsabilidade do cliente fornecer os serviços de tradução.
• Não tente reparar o equipamento sem ter consultado e compreendido este manual de assis-
tência técnica.
• O não cumprimento deste aviso pode colocar em perigo a segurança do técnico, do operador
ou do paciente devido a choques eléctricos, mecânicos ou outros.
ATENŢIE Acest manual de service este disponibil doar în limba engleză.
(RO) • Dacă un furnizor de servicii pentru clienţi necesită o altă limbă decât cea engleză, este de
datoria clientului să furnizeze o traducere.
• Nu încercaţi să reparaţi echipamentul decât ulterior consultării şi înţelegerii acestui manual
de service.
• Ignorarea acestui avertisment ar putea duce la rănirea depanatorului, operatorului sau pa-
cientului în urma pericolelor de electrocutare, mecanice sau de altă natură.
ОСТОРОЖНО Данное руководство по техническому обслуживанию представлено только на английском
! языке.
(RU) • Если сервисному персоналу клиента необходимо руководство не на английском, а на
каком-то другом языке, клиенту следует самостоятельно обеспечить перевод.
• Перед техническим обслуживанием оборудования обязательно обратитесь к данному
руководству и поймите изложенные в нем сведения.
• Несоблюдение требований данного предупреждения может привести к тому, что
специалист по техобслуживанию, оператор или пациент получит удар электрическим
током, механическую травму или другое повреждение.
UPOZOR- Ovo servisno uputstvo je dostupno samo na engleskom jeziku.
ENJE • Ako klijentov serviser zahteva neki drugi jezik, klijent je dužan da obezbedi prevodilačke
(SR) usluge.
• Ne pokušavajte da opravite uređaj ako niste pročitali i razumeli ovo servisno uputstvo.
• Zanemarivanje ovog upozorenja može dovesti do povređivanja servisera, rukovaoca ili paci-
jenta usled strujnog udara ili mehaničkih i drugih opasnosti.

46-017561 Revision 37 HHS 7/68


Control and Tube Assembly Tests

UPOZORNE- Tento návod na obsluhu je k dispozícii len v angličtine.


NIE • Ak zákazníkov poskytovateľ služieb vyžaduje iný jazyk ako angličtinu, poskytnutie preklada-
(SK) teľských služieb je zodpovednosťou zákazníka.
• Nepokúšajte sa o obsluhu zariadenia, kým si neprečítate návod na obluhu a neporozumiete
mu.
• Zanedbanie tohto upozornenia môže spôsobiť zranenie poskytovateľa služieb, obsluhujúcej
osoby alebo pacienta elektrickým prúdom, mechanické alebo iné ohrozenie.
ATENCIÓN Este manual de servicio sólo existe en inglés.
(ES) • Si el encargado de mantenimiento de un cliente necesita un idioma que no sea el inglés, el
cliente deberá encargarse de la traducción del manual.
• No se deberá dar servicio técnico al equipo, sin haber consultado y comprendido este manual
de servicio.
• La no observancia del presente aviso puede dar lugar a que el proveedor de servicios, el
operador o el paciente sufran lesiones provocadas por causas eléctricas, mecánicas o de otra
naturaleza.
VARNING Den här servicehandboken finns bara tillgänglig på engelska.
(SV) • Om en kunds servicetekniker har behov av ett annat språk än engelska, ansvarar kunden för
att tillhandahålla översättningstjänster.
• Försök inte utföra service på utrustningen om du inte har läst och förstår den här service-
handboken.
• Om du inte tar hänsyn till den här varningen kan det resultera i skador på serviceteknikern,
operatören eller patienten till följd av elektriska stötar, mekaniska faror eller andra faror.
OPOZORILO Ta servisni priročnik je na voljo samo v angleškem jeziku.
(SL) • Če ponudnik storitve stranke potrebuje priročnik v drugem jeziku, mora stranka zagotoviti
prevod.
• Ne poskušajte servisirati opreme, če tega priročnika niste v celoti prebrali in razumeli.
• Če tega opozorila ne upoštevate, se lahko zaradi električnega udara, mehanskih ali drugih
nevarnosti poškoduje ponudnik storitev, operater ali bolnik.
DİKKAT Bu servis kılavuzunun sadece ingilizcesi mevcuttur.
(TR) • Eğer müşteri teknisyeni bu kılavuzu ingilizce dışında bir başka lisandan talep ederse, bunu
tercüme ettirmek müşteriye düşer.
• Servis kılavuzunu okuyup anlamadan ekipmanlara müdahale etmeyiniz.
• Bu uyarıya uyulmaması, elektrik, mekanik veya diğer tehlikelerden dolayı teknisyen, operatör
veya hastanın yaralanmasına yol açabilir.
ЗАСТЕРЕЖЕ Даний посібник з експлуатації доступний тільки англійською мовою.
ННЯ • Якщо постачальник послуг клієнта спілкується іноземною мовою, тоді клієнт
(UK) зобов'язаний забезпечити переклад.
• Заборонено проводити огляд обладнання без попереднього звертання до даного
посібника з експлуатації і розуміння інформації, поданої у ньому.
• Недотримання цього застереження може завдати шкоди здоров’ю постачальника
послуг, оператора або пацієнта через ураження електричним струмом, механічну
травму або інше ушкодження.

8/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests

Revision History
Rev Date Reason for change
0 Sept. 24, 1993 Initial release.
1 Apr. 30, 1996 Multiple changes to bring test procedures up-to-date.
2 Dec. 1, 1997 Added reference to Keithley Non-invasive Voltage Divider use in
AMX–4 Series (Section 3).
3 Nov. 24, 2000 Added Manual mA accuracy – for INNOVA 2000 only (New Section 4).
4 Nov. 5, 2001 Added modifications in Manual mA accuracy – for INNOVA 2000 only
(New Section 4).
5 March 14, 2003 Updated for Innova 4100 for ME2. SPR BUge86469
6 July 24, 2003 Updated for M3a. SPR BUge87092 and 89570. Added modifications
in Chapter 5 – Manual mA accuracy - for INNOVA 4100 (In Cardiac
on Large Focal Spot, the step 13 table of the procedure has been
updated and the steps 22 to 25 has been added.) In Section 5, ±15%
replaces ±20%. Addition of “TECHNICAL ACCURACY – kV for INNOVA
4100” section. SPR BUCge94810
7 August 14, 2003 Updated for M3a. SPR BUge96138
8 October 17, 2003 Updated for M3b. SPR BUge93442
9 April 23, 2004 Section 5, mA Accuracy updated, BUCge93442
10 June 30, 2004 Section 6, step 3 updated, EURge12233
11 Jan 19, 2005 Section 5, and 6, title modified to include Innova 3100, EURge20430
12 Mar 18, 2005 Section 1, 5, and 6, title modified to include Innova 2100–IQ,
EURge23969
13 February 15, 2006 Updated for Innova 2100/2100–IQ, Innova 3100/3100–IQ and Innova
4100/4100–IQ
14 June 5, 2006 CFR 1020.30/31/32 regulatory compliance.
15 July 20, 2007 Added: Innova 2121–IQ and Innova 3131–IQ information,
EURge48215, EURge66413. Section 2–3, step 2 has been updated
Section 2–3 updated per EURge73953 Sections 6 and 7 updated per
EURge75794, EURge76403 and EURge77294.
16 Dec. 20, 2007 Conversion from Interleaf into e-CLS Platform
17 May 4, 2010 Update of REG0052, REG0056, REG0057 and REG0058
18 August 5, 2010 New update of REG0052, REG0056, REG0057 and REG0058
19 January 26, 2011 New update of REG0058
20 July 27, 2011 Update of REG0052, REG0056, REG0057, which introduces Innova
IGS 520, 530, 540.
21 September 23, 2011 Update of REG0052, REG0056, REG0057 to introduce Discovery IGS
730
22 January 20, 2012 Update of REG0052, REG0056, REG0057 to introduce Innova IGS
620, 630
23 January 27, 2012 Update of REG0052, REG0056, REG0057, REG0058 for Discovery IGS
730
24 May 29, 2012 Update of REG0052, REG0056, REG0057 and REG0058
25 August 20, 2013 Update of REG0052, REG0056 and REG0057 to introduce Discovery
IGS 740 and change product name to DiscoveryTM IGS 730, Discov-
eryTM IGS 740
Update of REG0056 and REG0057 for Jedi tool cover p/n.

46-017561 Revision 37 HHS 9/68


Control and Tube Assembly Tests

Rev Date Reason for change


26 September 4, 2013 Update of REG0052, REG0057 and REG0058 to introduce Optima
IGS320. Also in REG0057, new P/N have been added.
27 September 4, 2014 Update of REG0052, REG0056 and REG0057 to change product
name to “InnovaTM IGS 620” and “InnovaTM IGS 630”.
28 March 4, 2015 Update of REG0052 and REG0057 to introduce “Optima IGS 330”
and change product name to “Optima IGS 320”.
29 June 30, 2015 Update of REG0057 to introduce tool p/n 5507803 for Optima IGS
systems.
30 July 5, 2017 Update of REG0052, REG0056 and REG0057 to introduce “InnovaTM
IGS 5”, “InnovaTM IGS 6”, “DiscoveryTM IGS 7” and “DiscoveryTM IGS
7 OR”.
31 October 15, 2019 Update of REG0056 and REG0057.
32 January 2021 Updated of REG0052, REG0056, REG0057 and REG0058 to to intro-
duce "Allia™ IGS 7", "Allia™ IGS 7 OR", "Allia™ IGS Systems" and
"Optima IGS Plus"
33 November 2021 Update of REG0052, REG0056, REG0057 and REG0058 to introduce
"Allia™ IGS 3" and "Allia™ IGS 5".
34 June 2022 Update of REG0052 and REG0057 to introduce "Optima™ IGS Ultra"
and "Optima™ IGS Mega".
35 December 2022 Update of REG0056 and REG0057 for detector and tube pairing
improvement.
Update of REG0058 for SUIF Calibration option update.
36 April 2023 Added new procedure REG0153 - GAIA mA Internal Accuracy meas-
urement.
37 September 2023 Update of REG0058 and REG0153 for improvement.

10/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests Contents

Contents
1 HHS Control and Tube Assembly Tests .............................................................................. 13
1.1 Safety .................................................................................................................................................... 14
1.2 Introduction.......................................................................................................................................... 16
1.3 REG0052 - Generator Operator Indicators........................................................................................... 17
1.4 REG0054 - Technique Accuracy - Generator Timer ............................................................................. 21
1.5 REG0053 - Technique Accuracy - kV/mA .............................................................................................. 22
1.6 REG0055 - Manual mA Accuracy (Innova 2000 only) ........................................................................... 25
1.7 REG0056 - Tech & Manual mA Accuracy - Tool 2389357...................................................................... 28
1.8 REG0057 - Tech & Manual mA Accuracy - Tool 2389357-2 or 5480040 or 5507803 ............................ 38
1.9 REG0153 - GAIA mA Internal Accuracy measurement......................................................................... 48
1.10 REG0058 - Technique Accuracy - kV (20, 30 and 40 cm Detector)..................................................... 53
1.11 REG0059 - Technique Accuracy - mAs................................................................................................ 57
1.12 REG0060 - Indirect Linearity .............................................................................................................. 58
1.13 REG0061 - AEC Maximum mAs........................................................................................................... 59
1.14 REG0062 - AEC Minimum Exposure Time .......................................................................................... 60
1.15 REG0063 - Reproducibility Of Exposure ............................................................................................ 61
1.16 REG0064 - Direct Linearity Of Exposure Alternate Method............................................................... 64

46-017561 Revision 37 HHS 11/68


Page intentionally left blank
1
1 HHS Control and Tube Assembly Tests

46-017561 Revision 37 HHS 13/68


1.1 Safety Control and Tube Assembly Tests

1.1 Safety
X-Ray Protection

WARNING

X-RAY EQUIPMENT IF NOT PROPERLY USED MAY CAUSE INJURY. ACCORDINGLY,


THE INSTRUCTIONS HEREIN CONTAINED SHOULD BE THOROUGHLY READ AND
UNDERSTOOD BY EVERYONE WHO WILL USE THE EQUIPMENT BEFORE YOU ATTEMPT
TO PLACE THIS EQUIPMENT IN OPERATION. THE GENERAL ELECTRIC COMPANY,
MEDICAL SYSTEMS GROUP, WILL BE GLAD TO ASSIST AND COOPERATE IN PLACING
THIS EQUIPMENT IN USE.
ALTHOUGH THIS APPARATUS INCORPORATES A HIGH DEGREE OF PROTECTION
AGAINST X-RADIATION OTHER THAN THE USEFUL BEAM, NO PRACTICAL DESIGN
OF EQUIPMENT CAN PROVIDE COMPLETE PROTECTION. NOR CAN ANY PRACTICAL
DESIGN COMPEL THE OPERATOR TO TAKE ADEQUATE PRECAUTIONS TO PREVENT
THE POSSIBILITY OF ANY PERSONS CARELESSLY EXPOSING THEMSELVES OR OTHERS
TO RADIATION. IT IS IMPORTANT THAT EVERYONE HAVING ANYTHING TO DO
WITH X-RADIATION BE PROPERLY TRAINED AND FULLY ACQUAINTED WITH THE
RECOMMENDATIONS OF THE NATIONAL COUNCIL ON RADIATION PROTECTION
AND MEASUREMENTS AS PUBLISHED IN NCRP REPORTS AVAILABLE FROM NCRP
PUBLICATIONS, 7910 WOODMONT AVENUE, ROOM 1016, BETHESDA, MARYLAND
20814, AND OF THE INTERNATIONAL COMMISSION ON RADIATION PROTECTION,
AND TAKE ADEQUATE STEPS TO PROTECT AGAINST INJURY. THE EQUIPMENT IS
SOLD WITH THE UNDERSTANDING THAT THE GENERAL ELECTRIC COMPANY, MEDICAL
SYSTEMS GROUP, ITS AGENTS, AND REPRESENTATIVES HAVE NO RESPONSIBILITY
FOR INJURY OR DAMAGE WHICH MAY RESULT FROM IMPROPER USE OF THE
EQUIPMENT. VARIOUS PROTECTIVE MATERIAL AND DEVICES ARE AVAILABLE. IT IS
URGED THAT SUCH MATERIALS OR DEVICES BE USED.

CAUTION

United States Federal law restricts this device to use by or on the order of a physician.

Certified Electrical Contractor Statement


All electrical installations that are preliminary to positioning of the equipment at the site prepared
for the equipment shall be performed by licensed electrical contractors. In addition, electrical
feeds into the Power Distribution Unit shall be performed by licensed electrical contractors.
Other connections between pieces of electrical equipment, calibrations, and testing shall be
performed by qualified GE Medical personnel. The products involved (and the accompanying
electrical installations) are highly sophisticated, and special engineering competence is required.
In performing all electrical work on these products, GE will use its own specially trained field
engineers. All of GE’s electrical work on these products will comply with the requirements of the
applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers,
personnel of third-party service companies with equivalent training, or licensed electricians) to
perform electrical servicing on the equipment.
Damage in Transportation
All packages should be closely examined at time of delivery. If damage is apparent, have
notation “damage in shipment” written on all copies of the freight or express bill before delivery is
accepted or “signed for” by a General Electric representative or a hospital receiving agent. Whether

14/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.1 Safety

noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in
any event, within 14 days after receipt, and the contents and containers held for inspection by the
carrier. A transportation company will not pay a claim for damage if an inspection is not requested
within this 14 day period.
Call Traffic and Transportation, Milwaukee, Phone: 800-548-3366. Option 6 immediately after 1
damage is found. At this time be ready to supply name of carrier, delivery date, consignee name,

HHS Control and Tube Assembly Tests


freight or express bill number, item damaged and extent of damage.
Complete instructions regarding claim procedure are found in Section “S” of the Policy &
Procedure Bulletins (6/17/94).

46-017561 Revision 37 HHS 15/68


1.2 Introduction Control and Tube Assembly Tests

1.2 Introduction
The tests in this section require making exposures and recording data. Review all the tests of this
section at this time and plan an efficient test program to suit equipment.
Figure 1-1

16/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.3 REG0052 - Generator Operator Indicators

1.3 REG0052 - Generator Operator Indicators


Table 1-1 Personnel requirements
1
Required persons Preliminary require- Procedure Finalization

HHS Control and Tube Assembly Tests


ments
1 Not Applicable 40 minutes Not Applicable

Table 1-2 Required conditions

Condition Reference Effectivity


When two or more X–Ray tubes - Innova IGS 620 and Innova IGS 630
are controlled by the same expo- or Innova IGS 6
sure switch (hand or foot), indicator
lamps shall be mounted on the tube
units to indicate the selection.
For X-Ray controls that incorporate - -
automatic exposure control: The
product of X-Ray tube current and
exposure time shall not exceed 600
mAs per exposure for tube potential
equal to or greater than 50 kVp.
This product shall not exceed 2000
mAs per exposure for tube potential
less than 50 kVp.In the event that
an exposure is terminated by the
600 mAs/2000 mAs backup, manual
reset is required before further au-
tomatically timed exposures can be
made. Visual indication is required
that the 600 mAs/2000 mAs backup
has been actuated.
Audible tones are required for: Ra- - -
diographic exposure termination,
Fluoro 5 minute timer termination.
X-Ray warning label shall be - -
present.

Application
All X-Ray controls at installation and preventive maintenance calls, or when replacing major
components.
Discovery™ IGS 730, Discovery™ IGS 740 systems or Discovery™ IGS 7, Discovery™ IGS 7 OR systems
are also named Discovery™ IGS Systems. The procedure is applicable indifferently to Discovery™
IGS 730 orDiscovery™ IGS 740 or Discovery™ IGS 7 or Discovery™ IGS 7 OR.
Allia™ IGS 5 and Allia™ IGS 7 and Allia™ IGS 7 OR are also named Allia™ IGS Systems. The procedure
is applicable indifferently to Allia™ IGS 5 or Allia™ IGS 7 or Allia™ IGS 7 OR.

Procedure
1. (Not applicable to Discovery™ IGS Systems, Allia™ IGS 5 (from SP5) and Allia™ IGS 7 and Allia™
IGS 7 OR Systems) Check to see that indicator lights are mounted and functional on any tube
units where the same exposure switch actuates either tube (example: RAD 1 and RAD 2 both
controlled from handswitch). Most certified tube units are supplied with lamps. However, if

46-017561 Revision 37 HHS 17/68


1.3 REG0052 - Generator Operator Indicators Control and Tube Assembly Tests

additional indicator lamps are needed, order kit 46-172124G1 from PRO. This is an important
check since this is the only way extra indicator lamps are recognized and procured.
2. (Not applicable to Allia™ IGS 5 (from SP5), Allia™ IGS 7 and Allia™ IGS 7 OR (from CFG8) Systems)
X–Ray Timer for systems delivered before June 2006:
Fluoro Timer: Set to 1 minute and allow to run out with beam blocked. There should be
an audible indication at about 30 seconds remaining time. Depending on control type,
the exposure may or may not terminate when the time runs down. If the exposure is not
terminated by the run out of the timer, there must be a device which keeps track of the
cumulative exposure time. If some of the controls have been placed on a remote panel, be sure
that the audible indicator is active at this panel.
3. X–Ray Timer for Innova 2100-IQ, Innova 3100/3100-Q, Innova 4100/4100-IQ systems, Innova IGS
520, 530, 540, Innova IGS 5, Allia™ IGS 3, Allia™ IGS 5, Discovery™ IGS 730, Discovery™ IGS 740,
Discovery™ IGS 7, Discovery™ IGS 7 OR, Allia™ IGS 7, Allia™ IGS 7 OR, Optima™ CL320i/CL323i,
Optima™ IGS Plus, Optima™ IGS 320, Optima™ IGS 330, Optima™ IGS Ultra and Optima™ IGS
Mega:
3.1. The cumulated elapsed time of Fluoro and Record X–Ray exposure sequences are
counted, and the operator is warned upon five minutes of cumulated time:
3.1.1. Run a fluoro without releasing the pedal until it automatically stops.
3.1.2. (For Innova single plane, Optima and Discovery™ IGS Systems) Check that the
audible and visual warnings raise in control/exam rooms (i.e. on VCIM and TSSC)
after 5 minutes +0/–15 secondes.
3.1.3. (For Allia™ IGS Systems and Allia™ IGS 3) Check that the audible and visual
warnings raise in control/exam rooms (i.e. on VCIM and Touch Panel) after 5
minutes +0/–15 secondes.
3.2. The continuous Fluoro exposure sequence is automatically stopped upon ten minutes
of elapsed time.
3.2.1. Continue the fluoro pressing of the pedal until the fluoro automatically stops.
3.2.2. Check that the fluoro warning is always activated during this period and that the
fluoro timer cut –off the X–ray after a continuous fluoro of 10 minutes +0/–30
secondes
3.2.3. Depress and press again the fluoro pedal. Check that the fluoro starts again.
3.2.4. Reset X-Ray timer for clearing warnings.
4. NOTE
Start the time measurement when image is displayed on monitor for both cases.
X–Ray Timer for Innova 2121-IQ and Innova 3131-IQ systems, Innova™ IGS 620 and Innova™ IGS
630, Innova™ IGS 6:
4.1. To test the cumulated X–Ray Timer Frt/Lat:
4.1.1. Open a new Exam. Click on the X–ray–Timer icon on the monitoring bar on
the top of the console. A pop-up window is opened displaying the Frt and Lat
cumulated timers. Verify Frt and Lat cumulated timers start at 0.0.
4.1.2. Perform several Frt/Lat/Bp acquisitions until the Frt cumulated timer reaches
the warning limit of 5 min. Verify that (a) the X–ray–Timer icon and both Timer
reset buttons on (b) the VCIM and (c) the TSSC start blinking yellow. Verify a
sound is audible from the VCIM and TSSC master. Verify X–rays are not aborted or
inhibited.
4.1.3. Press the Timer reset button (VCIM or TSSC). Verify that in the pop-up window the
Frt timer is reset to 0.0 but the Lat timer is not.

18/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.3 REG0052 - Generator Operator Indicators

4.1.4. Perform several Lat/Bp acquisitions until the Lat cumulated timer reaches the
warning limit of 5 min. Verify that (a) the X–ray–Timer icon and both Timer reset
buttons on (b) the VCIM and (c) the TSSC start blinking yellow. Verify a sound
is audible from the VCIM and TSSC master. Verify X–rays are not aborted or
inhibited. 1
4.1.5. Press the Timer reset button (VCIM or TSSC). Verify that in the pop-up window the

HHS Control and Tube Assembly Tests


Lat timer is reset to 0.0 but the Frt timer is not.
4.2. To test the continuous X–ray Timer Frt/Lat:
4.2.1. Perform a continuous Frontal Fluoro acquisition for 10.0 min without
interruptions and measure the time with a stopwatch until the X–rays are
aborted.
4.2.2. Without interrupting the acquisition, verify that after 5.0 min the icon on the top
of the console and Timer reset buttons on VCIM and TSSC start blinking. Verify the
Timer warning tone is audible.
4.2.3. Verify after 10.0 min (tolerance +0/–5%) the X–rays are aborted, but can be started
again if pedal is released and pressed again.
4.2.4. Press the X–ray Timer reset button and verify that the timers in the pop-up
window are reset to 0.0.
4.2.5. Repeat the above test for a 10 min continuous Lateral Fluoro acquisition and 10
min continuous Biplane Fluoro acquisition.
5. NOTE
Start the time measurement when image is displayed on monitor for both cases.
Verification of automatic exposure control reset and visual indication functions upon
activation of the 600 mAs/2000 mAs backup:
5.1. For Advantx Systems:
The reset and visual indication functions upon activation of the 600 mAs/2000 mAs
backup do not have to befield tested on Advantx systems.
By software design, these functions in applications mode are assured by successful
performance of the Calibration Tool Unit T005 “MAS Integrator Test” in calibration
mode. This Tool Unit T005 is performed in the procedure for maximum mAs for Advantx
systems in Section 6 of this Direction.
See the Advantx operational documentation for information on the operation of the
automatic exposure control reset function.
5.2. For Non-Advantx Systems: the test procedure for the reset and visual indication
functions is covered in the Functional Check chapter of the appropriate generator or
automatic exposure control service manual.
5.2.1. Check for proper operation of the audible tones or buzzers as required by
regulations:
• Radiographic exposure: Make an exposure (no film needed).
• Make a fluoro recording (VTR, Cine, or Photo-spot). Film is not necessary.
5.2.2. Verify the presence of warning label bearing the statement “WARNING: This X–ray
unit may be dangerous to patient and operator unless safe exposure factors and
operating instructions are observed” on the main X–ray control panel.
5.2.3. For Systems manufactured On or After June 10th, 2006:
Verify the presence of a warning label bearing the statement on the main X–
ray control panel. WARNING: This X–Ray unit may be dangerous to patient
and operator, unless safe exposure factors, operating instructions and

46-017561 Revision 37 HHS 19/68


1.3 REG0052 - Generator Operator Indicators Control and Tube Assembly Tests

maintenance schedules are observed. To be used by authorized personnel


only.
6. Record Data
Record the data (apply any corrective factor if required and check final results are within
acceptance criteria).

Finalization
No finalization steps.

20/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.4 REG0054 - Technique Accuracy - Generator Timer

1.4 REG0054 - Technique Accuracy - Generator Timer


Table 1-3 Personnel requirements
1
Required persons Preliminary require- Procedure Finalization

HHS Control and Tube Assembly Tests


ments
1 Not Applicable Not Applicable Not Applicable

Table 1-4 Required conditions

Condition Reference Effectivity


When the equipment is connected - -
to an adequate power supply as
specified, the maximum deviation
of the exposure time from the se-
lected time shall not exceed the lim-
its given in the specifications (equip-
ment operating manual).

Application
All X-Ray units, (except those having only mAs selection capability, for such units, refer to REG0053
- Technique Accuracy - kV/mA on page 22), at installation, periodic maintenance calls, and when
repairing timer.

Procedure
1. This test is performed at the same time as the kV/mA accuracy test, refer to REG0053 -
Technique Accuracy - kV/mA on page 22.
2. Record data on Form F3382.
Figure 1-2 kV/mA Accuracy Chart

Finalization
No finalization steps.

46-017561 Revision 37 HHS 21/68


1.5 REG0053 - Technique Accuracy - kV/mA Control and Tube Assembly Tests

1.5 REG0053 - Technique Accuracy - kV/mA


Table 1-5 Personnel requirements

Required persons Preliminary require- Procedure Finalization


ments
1 Not Applicable Not Applicable Not Applicable

Application
kVp and mA accuracy: At installation, periodic maintenance, and repair.
kVp and mA accuracy applies to all tubes in system.

Requirements
NOTE
Specified accuracy refers to technic accuracy statements made in the Product
Specification section of generator operating manuals.
1. kVp Accuracy – Actual kVp shall match selected kVp within the “specified accuracy”.
2. mA Accuracy – Actual mA shall match selected mA within the “specified accuracy”.
Each connected tube and generator forms a distinct X-Ray system. The tube usually limits the
range and combination of technic factors. Thus, accuracy of kVp and mA selections applies
only within ratings of the tube and focal spot used, even though selections beyond tube
ratings are available at the control.
mA and kVp limits for tubes are shown by curves on radiographic rating charts. The charts have
curves of mA or kW values. Selected mA must never exceed the largest mA value shown. For
kVp, the selection must never be less than the lowest kVp on the mA curve, whether it ends
at the bottom of the kVp scale or at a definite point above. This minimum kVp limit is necessary
for practical space charge compensation to meet accuracy requirements and to preserve
filament life. For maximum kVp, tube protector circuits are set to limit kVp approximately
where the mA curve crosses the selected time value.
Set up HHS data spreadsheet on laptop computer per 46–017560, HHS TOOLS.

Procedure
1. Be sure that generator panel meter accuracy has been verified. Refer to service manual for
procedure. Measure and record line voltage into generator.
2. Use non-invasive divider (refer to Figure 1-3 Detector Positioning for kVp
Measurement on page 23, Figure 1-4 Detector Positioning (vertical) for kVp
Measurement on page 23 and Figure 1-5 Divider Positioning for kVp Measurement (for under-

22/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.5 REG0053 - Technique Accuracy - kV/mA

table tube) on page 23). Turn unit on and select tube to be checked, leaving both energized
continuously. Wait 1 hour and 15 minutes before making measurements.
Figure 1-3 Detector Positioning for kVp Measurement
1

HHS Control and Tube Assembly Tests


Figure 1-4 Detector Positioning (vertical) for kVp Measurement

Figure 1-5 Divider Positioning for kVp Measurement (for under-table tube)

NOTE
The following applies:

46-017561 Revision 37 HHS 23/68


1.5 REG0053 - Technique Accuracy - kV/mA Control and Tube Assembly Tests

AMX-4 Series mobile X-Ray units require that a Keithley Non-Invasive kVp Divider
(Model 35080A with Deviation 535 or Model 35080B, both using Mobile Filter
Pack Plus 37946C and optional Low Range Filter Pack38237C) be used. No other
substitutions for non-invasive kVp Dividers are approved.
After several hours of no exposures, the first exposures should not be made at full
ratings.
3. Enter data required by technique accuracy charts on Form 3382 on laptop computer.
Recalibrate, if necessary, to obtain satisfactory performance.

Finalization
No finalization steps.

24/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.6 REG0055 - Manual mA Accuracy (Innova 2000 only)

1.6 REG0055 - Manual mA Accuracy (Innova 2000 only)


Table 1-6 Personnel requirements
1
Required persons Preliminary require- Procedure Finalization

HHS Control and Tube Assembly Tests


ments
1 Not Applicable Not Applicable Not Applicable

Application
mA accuracy: at installation, periodic maintenance, and repair.
mA accuracy applies to all tubes in system.
The tests described in this evaluation plan verify the accuracy of the mA in manual mode by
comparing the mA values entered manually into the system to the mA value measured by the VAMP
generator through its internal mAs meter.
Scope
Test accuracy between mA command and mA measured by the VAMP mAs–meter for various kV,
PW, fps, focalspot and acquisition modes.

Procedure
NOTE
Configure the system to perform manual acquisitions.
1. Select the X-220-Fluoro running at 30 fps.
2. Enter the value for kV, pulse–width and mA according to the table below in the range allowed
by the system and enter these values into the system.
NOTE
At low mA feedback, a part of the inaccuracy comes from the quantization step
(0.025) on this signal.
3. Perform a fluoro acquisition with these values for approximately 20 seconds.
4. Read the VAMP average mA feedback displayed on Titan Console just before the end of the
Fluoro acquisition.
5. Log in the table below this VAMP value and the technics feedback displayed in the status bar of
the SUIF. Compare the mA demand values to the corresponding mA feedback. The result of this
test is OK when the % difference is within the range (–20%, +20%).
Table 1-7

fps kVp mAs (mA) PW Average mA demand Average mA feedback % difference


30 65 0.025 (5) 5 0.75 (*)
30 70 0.05 (10) 5 1.5
30 75 0.15 (30) 5 4.5
30 80 0.375 (30) 12.5 11.25
30 100 0.375 (30) 12.5 11.25
30 120 0.41625 (33.3) 12.5 12.4875

NOTE
The real demanded values to the generator are those updated in the status bar
about is after the acquisition has started.
46-017561 Revision 37 HHS 25/68
1.6 REG0055 - Manual mA Accuracy (Innova 2000 only) Control and Tube Assembly Tests

(*) This point is known to be close to the limits when tube is cold, repeat the point if
needed.
6. Select the X-220-Cardiac running at 30 fps.
7. Select the 0.6 FS and enter the values for kV, pulse–width and mA according to the table below
in the range allowed by the system and enter these values into the system.
8. Perform a cardiac acquisition with these values for approximately 10 seconds.
9. Read the VAMP average mA feedback displayed on Titan Console just before the end of the
Cardiac acquisition.
10. Log in the table below the VAMP value and the technics feedback displayed on the status bar of
the SUIF. Compare the mA demand values to the corresponding mA feedback. The result of this
test is Ok when the % difference is within the range (–20%, +20%).
Table 1-8

fps kVp mAs (mA) PW Average mA demand Average mA feedback % difference


30 70 0.2 (40) 5 6
30 75 0.4 (80) 5 12
30 80 1.25 (250) 5 37.5 (*)
30 100 2.4 (240) 10 72
30 120 2 (200) 10 60

NOTE
(*) This point is known to be close to the upper limit on a very hot tube. Repeat the
point if needed on a cooler state of the tube (but not cold).
11. Select the X-220-Cardiac running at 30 fps.
12. Select the 1.2 FS and enter the values for kV, pulse–width and mA according to the table below
in the range allowed by the system and enter these values into the system.
13. Perform a cardiac acquisition with these values for approximately 10 seconds.
14. Read the VAMP average mA feedback displayed on Titan Console just before the end of the
Cardiac acquisition.
15. Log in the table below the VAMP value and the technics feedback displayed on the status bar of
the SUIF. Compare the mA demand values to the corresponding mA feedback. The result of this
test is Ok when the % difference is within the range (–20%, +20%).
Table 1-9

fps kVp mAs (mA) PW Average mA demand Average mA feedback % difference


30 80 1.25 (250) 5 37.5
30 80 2 (400) 5 60
30 90 7 (700) 10 210
30 100 6.4 (640) 10 192
30 120 5.33 (533) 10 159.9

NOTE
The real demanded values to the generator are those updated in the status bar
about is after the acquisition has started.

26/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.6 REG0055 - Manual mA Accuracy (Innova 2000 only)

Finalization
No finalization steps.
1

HHS Control and Tube Assembly Tests

46-017561 Revision 37 HHS 27/68


1.7 REG0056 - Tech & Manual mA Accuracy - Tool 2389357 Control and Tube Assembly Tests

1.7 REG0056 - Tech & Manual mA Accuracy - Tool


2389357
Table 1-10 Personnel requirements

Required persons Preliminary require- Procedure Finalization


ments
1 Not Applicable 90 minutes Not Applicable

Table 1-11 Tools and test equipment

Item Quantity Effectivity Part number Manufacturer


Detector Lead Block- 1 - 2337600 -
er
JEDI mA shunt meas- 1 - 2389357 -
urement external
tool
JEDI tool cover 1 All systems 2404211 -
JEDI tool cover 1 Innova single plane, 2404211-20 -
Discovery™ IGS sys-
tems and Allia™ IGS
Systems and Allia™
IGS 3
Standard Field Engi- 1 - - -
neering Tools with
metric and US
wrenches
Fluke 87 serie 3 mul- 1 - - -
timeter

Table 1-12 Safety

WARNING

RESIDUAL VOLTAGE PRESENT.


BEFORE SERVICING THE CABINET, AFTER HAVING POWERED IT OFF, ENSURE THAT
JEDI HAS NO RESIDUAL VOLTAGE

Table 1-13 Required conditions

Condition Reference Effectivity


The X-Ray tube mA _ Average com- - -
puted value with applied correc-
tion (mA_Average correction 24 m
or mA_Average correction 28.70 m
or mA_Average correction 30 m)
should be within ± 20% of the
mA_Average Demand value.
The actual mA_average value updat- - -
ed at exposure end, available in the
X220 status bar, should be within
± 20% of the mA_Average Demand
value.

28/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.7 REG0056 - Tech & Manual mA Accuracy - Tool 2389357

Application
This procedure is applicable for Innova™ 2100-IQ/2121-IQ, Innova™ 3100/3100-IQ/3131-IQ, Innova™
4100/4100-IQ systems, Innova™ IGS 520, 530, 540, Innova™ IGS 5, Allia™ IGS 3, Allia™ IGS 5, Innova™
IGS 620, Innova™ IGS 630, Innova™ IGS 6, Discovery™ IGS 730, Discovery™ IGS 740, Discovery™ IGS 7, 1
Discovery™ IGS 7 OR, Allia™ IGS 7, Allia™ IGS 7 OR.

HHS Control and Tube Assembly Tests


Discovery™ IGS 730, Discovery™ IGS 740 systems or Discovery™ IGS 7 and Discovery™ IGS 7 OR
systems are also named Discovery™ IGS Systems. The procedure is applicable indifferently to
Discovery™ IGS 730 or Discovery™ IGS 740 or Discovery™ IGS 7 or Discovery™ IGS 7 OR.
Allia™ IGS 5 and Allia™ IGS 7 and Allia™ IGS 7 OR are also named Allia™ IGS Systems. The procedure
is applicable indifferently to Allia™ IGS 5 or Allia™ IGS 7 or Allia™ IGS 7 OR.
This procedure will be performed in all following cases:
• System installation

NOTICE

Based on manufacturing process and component pairing, it is not required to


repeat these mA Accuracy measurements during System installation if the JEDI,
the Ingrid and the X-Ray tube are paired. For details on the Pairing Process and
how to check it, please refer to information provided in the Tube and Detector
Calibration in SUIF Job Card.
• System repair
Purpose
The tests described (see Table 1-14 Test Angio Parameters on page 34 and Table 1-15 Test Cardiac
Parameters on page 34) in this procedure verify the accuracy of the X–Ray tube mA_Average in all
following system acquisition modes for kV_MIN and kV_MAX at mA peak tube current equal to 20%
of the highest tube current.
• Angio mode (on Compressed, Small and Large focus)
• Cardiac mode (on Small and Large focus)
NOTE
The FLUORO acquisition mode is not performed because it is the same as Cardiac on
Small focus
The X–Ray tube mA _ Average is measured with an external tool and a multimeter in
manual mode using the X220 tool unit (located in the SUIF):
For SINGLE PLANE use X220 Generic.
For BIPLANE, use X220 Frontal and X220 Lateral.
NOTE
The accepted tools for this Procedure are: mA tool 2389357 & only with multimeter Fluke
87 Serie 3 (this procedure requires a specific function available only on the multimeter
Fluke 87 Serie 3).

Acceptance Criteria – Rejection Limit


As a result of ± 5% due to inherent inaccuracies of the instrumentation and test method, the
rejection limits are set to ± 15%. Record the data (apply any corrective factor if required and check
final results are within acceptance criteria).
• (For Innova and Allia™ IGS 5 and Allia™ IGS 3 Systems)

46-017561 Revision 37 HHS 29/68


1.7 REG0056 - Tech & Manual mA Accuracy - Tool 2389357 Control and Tube Assembly Tests

If your system is equipped with 24 meters HV cables - Fill in the spreadsheet with correction
related to mA_Average correction 24 m.
• (For Innova and Allia™ IGS 5 and Allia™ IGS 3 Systems)
If your system is equipped with 30 meters HV cables - Fill in the spreadsheet with correction
related to mA_Average correction 30 m.
• (For Discovery™ IGS systems and Allia™ IGS 7 and Allia™ IGS 7 OR Systems)
Fill in the spreadsheet with correction related to mA_Average correction 28.70 m.

Prerequisites
1. Switch OFF the system.
2. For Jedi equipped with a front cover [A], loosen the four captive screws [B] and remove the
front cover. Then go to Step 6.
If the Jedi is not equipped with a front cover [A], perform the Step 3 to Step 5.
Figure 1-6 Remove Jedi Front Cover

3. WARNING

BEFORE REMOVING JEDI INVERTER COVER, WAIT UNTIL ALL LED’S ARE
EXTENGUISH ON JEDI AC/DC BD
(For Jedi without Front Cover) For Single Plane, open C1/C-FRT Cabinet (for Biplane, open C1
FRT or C1 LAT or C-FRT or C-LAT Cabinet) and remove the JEDI inverter cover, refer to:
(For Innova Single Plane and Discovery™ IGS Systems with C1 Cabinet) DR0187 - Jedi KVCTL
Board , steps 1 to 4 of the Disassembly procedure.
(For Innova Single Plane, Discovery™ IGS Systems and Allia™ IGS Systems and Allia™ IGS 3 with
C-FRT Cabinet) DR1161 - Jedi KVCTL Board , steps 1 to 4 of the Disassembly procedure.
(For Innova Biplane IGS Systems with C1 FRT and C1 LAT Cabinet) DR0187 - Jedi KVCTL Board ,
steps 1 to 4 of the Disassembly procedure.
(For Innova Biplane IGS Systems with C-FRT and C-LAT Cabinet) DR1161 - Jedi KVCTL Board , steps
1 to 4 of the Disassembly procedure.

30/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.7 REG0056 - Tech & Manual mA Accuracy - Tool 2389357

(For Optima Systems) DR0871 - Jedi KVCTL Board , steps 1 to 4 of the Disassembly procedure.
4. (For Jedi without Front Cover) Install the JEDI cover tool (2404211): put in place the plastic cover
tool for JEDI inverter protection [1].
Figure 1-7 Connection of the JEDI mA Shunt Measurement Tool 1

HHS Control and Tube Assembly Tests


5. Remove the flat cable in J1 of KVCTL board ([2], Figure 1-7 Connection of the JEDI mA Shunt
Measurement Tool on page 31).

46-017561 Revision 37 HHS 31/68


1.7 REG0056 - Tech & Manual mA Accuracy - Tool 2389357 Control and Tube Assembly Tests

6. Plug the J1 of the JEDI mA shunt measurement tool [3] (2389357) in the J1 connector of the
KVCTL board.
Figure 1-8 mA Measurement Tool (2389357) and Multimeter Connection

7. Plug the flat cable in the JEDI mA shunt measurement tool (male side, connetor J3).
8. Connect the cables ([4], Figure 1-8 mA Measurement Tool (2389357) and Multimeter
Connection on page 32) of the multimeter ([5], Figure 1-8 mA Measurement Tool (2389357) and
Multimeter Connection on page 32) to the connector J4 of the JEDI mA shunt measurement
tool.

Test Procedure
mA Shunt Resistor Measurement with JEDI OFF
1. Select Ohmmeter function on FLUKE 87 serie 3.
2. Select maximum display accuracy function on FLUKE 87 serie 3 – Press the 41/2 DIGITS button
for 1 second (the 41/2 DIGITS display is switched ON).
3. Connect the FLUKE 87 serie 3 multimeter between Ground (Black pin) and Shunt A (Red pin) on
the external tool.
4. Record the Shunt A resistor value with 2 decimal places (measured around 5 ohms).
5. Connect the FLUKE 87 serie 3 multimeter between Ground (Black pin) and Shunt B (Blue pin)
on the external tool.
6. Record the Shunt B resistor value with 2 decimal places (measured around 0 ohm).

mA Shunt Resistor Accuracy Verification


1. Record and compile values according to Table 1-14 Test Angio Parameters on page 34 and
Table 1-15 Test Cardiac Parameters on page 34.

32/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.7 REG0056 - Tech & Manual mA Accuracy - Tool 2389357

2. The rejection limit is 5 ohms ± 2%.

Power ON the System


1. Verify connection of the FLUKE 87 serie 3 multimeter between Shunt A (Red pin) and Shunt B
(Blue pin) on the mA tool.
1
2. For security, be sure the inverter cover tool is installed and secured with screws.

HHS Control and Tube Assembly Tests


3. Power ON the system to carry out the following tests.

Measurement procedure
1. Put in place the DLB (Detector Lead Blocker).
2. Select the FLUKE 87 serie 3 multimeter in Vdc voltage function with 4 V Range. Select the Mode
Record MAX display with a response time of 1 second. To achieve this mode carry out the
following actions:
2.1. Press the 4 1/2 DIGITS button for 1 second (the 41/2 DIGITS display is switched ON).
2.2. Press briefly the MIN MAX button 1st time - a long time Bip is emitted (the Mode 1s REC
is switched ON).
2.3. Press briefly the MIN MAX button 2nd time - (the Mode 1s REC MAX is switched OFF).
3. (For Allia IGS systems) Set the system in manual mode using the X220 unit in the SUIF (refer to
Figure 1-9 X220 Unit on page 33).
3.1. Enter in the SUIF (on the DL browser, click on the knife tool icon).
Figure 1-9 X220 Unit

NOTE
mA average value is displayed in the status bar.
3.2. Click on Calibrations button.
3.3. Select Regulatory and Dose Measurement.
3.4. Select X220 (for Single Plane) or Select X220 Frontal or X220 Lateral (for Biplane).
4. (For Innova systems, Discovery Systems and Optima systems) Set the system in manual mode
using the X220 unit in the SUIF (refer to Figure 1-10 X220 Unit on page 34).

46-017561 Revision 37 HHS 33/68


1.7 REG0056 - Tech & Manual mA Accuracy - Tool 2389357 Control and Tube Assembly Tests

4.1. Enter in the SUIF (on the DL browser, click on the knife tool icon).
Figure 1-10 X220 Unit

4.2. Click on Calibration button.


4.3. Select HHS and Dose Measurement.
4.4. Select X220 Generic (for Single Plane) or Select X220 Frontal or X220 Lateral (for
Biplane).
5. Choose exposure technics parameters according to the Tables listed below.
Table 1-14 Test Angio Parameters

Fluke 87 serie 3 Range Measurement on 4 Vdc Test 1 Test 2 Test 3 Test 4 Test 5 Test 6
Exp Mode DSA DSA DSA DSA DSA DSA
Frame Rate 3.75 fps 3.75 fps 3.75 fps 3.75 fps 3.75 fps 3.75 fps
Focal Spot 0.3 0.3 0.6 0.6 1 1
kVPeak 50 125 50 125 50 125
mAPeak 40 40 80 80 200 200
Pulse Width 100 ms 100 ms 100 ms 100 ms 100 ms 100 ms
Spectral Filter None None None None None None
DLB Presence Yes Yes Yes Yes Yes Yes
Collimator LONG Leave Empty
Collimator LAT Leave Empty

Table 1-15 Test Cardiac Parameters

Fluke 87 serie 3 Range Measurement on 4 Vdc Test 7 Test 8 Test 9 Test 10


Exp Mode Cardiac Cardiac Cardiac Cardiac
Frame Rate 30 fps 30 fps 30 fps 30 fps
Focal Spot 0.6 0.6 1 1

34/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.7 REG0056 - Tech & Manual mA Accuracy - Tool 2389357

Table 1-15 Test Cardiac Parameters (Table continued)

Fluke 87 serie 3 Range Measurement on 4 Vdc Test 7 Test 8 Test 9 Test 10


kVPeak 60 120 60 120
mAPeak 80 80 200 200
1
Pulse Width 9 ms 9 ms 9 ms 9 ms

HHS Control and Tube Assembly Tests


Spectral Filter None None None None
DLB Presence Yes Yes Yes Yes
Collimator LONG Leave Empty
Collimator LAT Leave Empty

6. Select Spectral filter: None, DLB presence = Yes, Collimator Long / Lat: Leave empty.
7. Select SUBMIT to set all parameters.
8. Depress the hand switch to initiate the X–Ray exposure during approximately 10 to 20 seconds
to allow multimeter acquisition stabilization (Maintain the X–Ray exposure until FLUKE 87 serie
3 multimeter stop to bip for maximum value recorded).
9. Record the multimeter voltage value displayed.
10. Record the mA_average value updated at exposure end displayed in the X220 status bar (refer
to Figure 1-10 X220 Unit on page 34).
11. Record and compile values according to Table 1-14 Test Angio Parameters on page 34 and Table
1-15 Test Cardiac Parameters on page 34.
12. After each measurement, RESET the FLUKE 87 serie 3 displayed value:
• Press a long time (1 s) the MIN MAX button 1st time - (the Mode 1s REC is switched OFF).
• Press briefly the MIN MAX button 2nd time - (the Mode 1s REC is switched ON).
• Press briefly the MIN MAX button 3rd time - (the Mode 1s REC MAX is switched ON).
13. Repeat all previous actions to perform all 10 tests.
14. When all tests are performed, click on ABORT to exit.

SpreadSheet fill in procedure


1. Fill in data
1.1. Before data fill in, choose the right spreadsheet in accordance to your system as SP and
Frontal for Single Plane or Frontal Biplane, and Lateral for Lateral Biplane.
1.2. Furthermore, in each spreadsheet you need to select the High Voltage cable length
equipment (24 meters or 28.7 meters or 30 meters) and the mA measurement tool
procedure, after, you will be able to enter the following values:
• The measured resistor values (in Ohm, with 2 decimal places) displayed on
multimeter.
• The measured voltage (in mVdc) displayed on multimeter.
• The mA_avg value (in mA) displayed on X220 status bar.
2. Computation Details
2.1. For mA Shunt resistor verification test:
A calculated mA Shunt resistor value is automatically computed as following formula
–––> (Shunt A resistor - Shunt B resistor); the rejection limits are 4.90 Ohm (LSL) and
5.10 Ohm (USL) for PASS / FAIL test.

46-017561 Revision 37 HHS 35/68


1.7 REG0056 - Tech & Manual mA Accuracy - Tool 2389357 Control and Tube Assembly Tests

2.2. For each mA accuracy station test:


• A calculated mA_avg value (without correction) is automatically computed as
following formula –––> (mVdc / 5Ω).
• A corrected mA_avg value (with correction) is automatically computed as following
formula –––>[(mVdc / 5Ω) – mA_avg correction value @ HV cable length] see
following Table for mA_avg correction values.
• This corrected mA_avg value is automatically compared to the rejection limit @
selected mA measurement tool procedure (see following Table for rejection limit
values) for PASS / FAIL test.
• The mA_avg value returned in X220 status bar is automatically compared to the
rejection limit (see following Table for rejection limit values) for PASS / FAIL test.
NOTE
For Discovery™ IGS Systems and Allia™ IGS Systems and Allia™ IGS 3 refer
to column “mA_avg correction value @ 28.70 m HV cable (mA)”.

TEST No. mA_avg mA_avg mA_avg LSL ejection mA_avg De- USL rejec-
correction correction correction limit (–15% mand value tion limit
value @ 24 value @ value @ 30 of mA_avg (mA) (+15% of
m HV cable 28.70 m HV m HV cable Demand) mA_avg De-
(mA) cable (mA) (mA) (mA) mand) (mA)
1 0.46 0.51 0.53 12.75 15 17.25
2 1.16 1.28 1.31 12.75 15 17.25
3 0.48 0.53 0.55 25.5 30 34.5
4 1.21 1.33 1.37 25.5 30 34.5
5 0.55 0.60 0.61 63.75 75 86.25
6 1.37 1.49 1.53 63.75 75 86.25
7 0.15 0.15 0.15 18.36 21.6 24.84
8 0.30 0.30 0.30 18.36 21.6 24.84
9 4.83 5.31 5.44 45.9 54 62.1
10 9.67 10.62 10.88 45.9 54 62.1

3. Removal the mA Shunt Tool and JEDI Cover Tool


3.1. Switch OFF the system.
3.2. WARNING

BEFORE REMOVING JEDI INVERTER COVER TOOL, WAIT UNTIL ALL LED’S
ARE EXTINGUISHED ON JEDI AC/DC BD.
Remove the multimeter.
3.3. (For Jedi without Front Cover) Remove the inverter cover tool (2404211).
3.4. Remove the mA shunt tool (2389357).
3.5. Reinstall the flat cable in the J1 connector.
3.6. (For Jedi with Front Cover) Reinstall the front cover and attach it with the four captive
screws.
3.7. (For Jedi without Front Cover) Reinstall the Inverter metallic cover, refer to:
(For Innova Single Plane and Discovery™ IGS Systems with C1 Cabinet) DR0187 - Jedi KVCTL
Board , Reassembly procedure.

36/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.7 REG0056 - Tech & Manual mA Accuracy - Tool 2389357

(For Innova Single Plane, Discovery™ IGS Systems and Allia™ IGS Systems and Allia™ IGS 3
with C-FRT Cabinet) DR1161 - Jedi KVCTL Board , Reassembly procedure.
(For Innova Biplane IGS Systems with C1 FRT and C1 LAT Cabine) DR0187 - Jedi KVCTL
Board , Reassembly procedure.
1
(For Innova Biplane IGS Systems with C-FRT and C-LAT Cabinet) DR1161 - Jedi KVCTL
Board , Reassembly procedure.

HHS Control and Tube Assembly Tests


(For Optima Systems) DR0871 - Jedi KVCTL Board , Reassembly procedure.
3.8. Close Auxiliary ACDC assembly in the C1/C-FRT cabinet.
3.9. Remove the Lead Blocker.
3.10. Switch ON the system.
3.11. Carry out an X–ray expose to control the system operation.

Finalization
No finalization steps.

46-017561 Revision 37 HHS 37/68


1.8 REG0057 - Tech & Manual mA Accuracy - Tool 2389357-2 or 5480040 Control and Tube Assembly Tests
or 5507803

1.8 REG0057 - Tech & Manual mA Accuracy - Tool


2389357-2 or 5480040 or 5507803
Table 1-16 Personnel requirements

Required persons Preliminary require- Procedure Finalization


ments
1 Not Applicable 90 minutes Not Applicable

Table 1-17 Tools and test equipment

Item Quantity Effectivity Part number Manufacturer


Detector Lead Blocker 1 Innova, Discovery™ 2337600 -
IGS Systems and Allia™
IGS Systems, 40 cm
detector
Detector Lead Blocker 1 Innova, Discovery™ 2396564 -
IGS Systems and Allia™
IGS Systems and Allia™
IGS 3, 30 cm detector
Detector Lead Blocker 1 Innova Systems and 5125039 -
Allia™ IGS 5 Systems,
20 cm detector
Detector Lead Blocker 1 Optima™ IGS Ultra, 2353163 -
40 cm detector
Detector Lead Blocker 1 Optima™ CL320i/ 5413024 -
CL323i, Optima™ IGS
Plus, Optima™ IGS 330,
Optima™ IGS Ultra,
30 cm detector and
Optima™ IGS Mega
Detector Lead Blocker 1 Optima™ IGS Plus, Op- 5488772 -
tima™ IGS 320 and Op-
tima™ IGS Ultra, 20 cm
detector
JEDI mA measure- 1 Innova, Discovery™ 2389357-2 -
ment external tool IGS Systems and Allia™

IGS Systems and Allia™


IGS 3
JEDI mA measure- 1 Optima Systems 5480040 -
ment external tool
JEDI mA measure- 1 Optima Systems 5507803 -
ment external tool
JEDI tool cover 1 All systems 2404211 -
JEDI tool cover 1 Innova single plane, 2404211-20 -

Discovery IGS Sys-
tems and Allia™ IGS
Systems and Allia™ IGS
3
JEDI tool cover 1 Optima IGS systems 5480038 -

38/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.8 REG0057 - Tech & Manual mA Accuracy - Tool 2389357-2 or 5480040
or 5507803

Table 1-17 Tools and test equipment (Table continued)

Item Quantity Effectivity Part number Manufacturer


Standard Field Engi- 1 - - -
neering Tools with 1
metric and US
wrenches

HHS Control and Tube Assembly Tests


Multimeter with volt- 1 - - -
age accuracy ≤ 0.1
%. A multimeter with
min/max function is
preferred for easy use.

Table 1-18 Safety

WARNING

RESIDUAL VOLTAGE PRESENT.


BEFORE SERVICING THE CABINET, AFTER HAVING POWERED IT OFF, ENSURE THAT
JEDI HAS NO RESIDUAL VOLTAGE

Table 1-19 Required conditions

Condition Reference Effectivity


The X–Ray tube mA _ Average com- - -
puted value with applied correc-
tion (mA_Average correction 24 m
or mA_Average correction 28.70 m
or mA_Average correction 30 m)
should be within ± 20% of the
mA_Average Demand value.
The actual mA_average value updat- - -
ed at exposure end, available in the
X220 status bar, should be within
± 20% of the mA_Average Demand
value.

Application
This procedure is applicable for Innova™ 2100-IQ/2121-IQ, Innova™ 3100/3100-IQ/3131-IQ, Innova™
4100/4100-IQ systems,Innova™ IGS 520, 530, 540, Innova™ IGS 5, Allia™ IGS 3, Allia™ IGS 5, Innova™
IGS 620, Innova™ IGS 630, Innova™ IGS 6, Discovery™ IGS 730, Discovery™ IGS 740, Discovery™ IGS
7, Discovery™ IGS 7 OR, Allia™ IGS 7, Allia™ IGS 7 OR, Optima™ CL320i/CL323i, Optima™ IGS Plus,
Optima™ IGS 320, Optima™ IGS 330, Optima™ IGS Mega and Optima™ IGS Ultra.
Discovery™ IGS 730, Discovery™ IGS 740 systems orDiscovery™ IGS 7 and Discovery™ IGS 7 OR
systems are also named DiscoveryTM IGS Systems. The procedure is applicable indifferently to
Discovery™ IGS 730 or Discovery™ IGS 740 or Discovery™ IGS 7 or Discovery™ IGS 7 OR.
Allia™ IGS 5 and Allia™ IGS 7 and Allia™ IGS 7 OR are also named Allia™ IGS Systems. The procedure
is applicable indifferently to Allia™ IGS 5 or Allia™ IGS 7 or Allia™ IGS 7 OR.
This procedure will be performed in all following cases:
• System installation

46-017561 Revision 37 HHS 39/68


1.8 REG0057 - Tech & Manual mA Accuracy - Tool 2389357-2 or 5480040 Control and Tube Assembly Tests
or 5507803

NOTICE

Based on manufacturing process and component pairing, it is not required to


repeat these mA Accuracy measurements during System installation if the JEDI,
the Ingrid and the X-Ray tube are paired. For details on the Pairing Process and
how to check it, please refer to information provided in the Tube and Detector
Calibration in SUIF Job Card.
• System repair
Purpose
The tests described (see Table 1-20 Test Angio Parameters on page 45 and Table 1-21 Test Cardiac
Parameters on page 45) in this procedure verify the accuracy of the X–Ray tube mA_Average in all
following system acquisition modes for kV_MIN and kV_MAX at mA peak tube current equal to 20%
of the highest tube current.
• Angio mode (on Compressed, Small and Large focus)
• Cardiac mode (on Small and Large focus)
NOTE
The FLUORO acquisition mode is not performed because it is the same as Cardiac on
Small focus
The X–Ray tube mA _ Average is measured with an external tool and a multimeter in
manual mode using the X220 tool unit (located in the SUIF):
For SINGLE PLANE use X220 Generic.
For BIPLANE, use X220 Frontal and X220 Lateral.
NOTE
The accepted tools for this Procedure are: mA tool 2389357-2 & a multimeter with
voltage accuracy ≤ 0.1 %, input impedance ≥10 Mohm, the ohmmeter function with
accuracy ≥ 0.2 % and resolution < 0.1 ohm. A multimeter with min/max function is
preferred for easy use.

Acceptance Criteria – Rejection Limit


As a result of ± 8% due to inherent inaccuracies of the instrumentation and test method, the
rejection limits are set to ± 12%. Record the data (apply any corrective factor if required and check
final results are within acceptance criteria).
• (For Innova, Optima Systems and Allia™ IGS 5 Systems and Allia™ IGS 3)
If your system is equipped with 24 meters HV cables - Fill in the spreadsheet with correction
related to mA_Average correction 24 m.
• (For Innova, Optima Systems and Allia™ IGS 5 Systems and Allia™ IGS 3)
If your system is equipped with 30 meters HV cables - Fill in the spreadsheet with correction
related to mA_Average correction 30 m.
• (For Discovery™ IGS Systems and Allia™ IGS 7 and Allia™ IGS 7 OR Systems)
Fill in the spreadsheet with correction related to mA_Average correction 28.70 m.

Prerequisites
1. Switch OFF the system.

40/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.8 REG0057 - Tech & Manual mA Accuracy - Tool 2389357-2 or 5480040
or 5507803

2. For Jedi equipped with a front cover [A], loosen the four captive screws [B] and remove the
front cover. Then go to Step 6.
If the Jedi is not equipped with a front cover [A], perform the Step 3 to Step 5.
Figure 1-11 Remove Jedi Front Cover 1

HHS Control and Tube Assembly Tests


3. WARNING

BEFORE REMOVING JEDI INVERTER COVER, WAIT UNTIL ALL LED’S ARE
EXTENGUISH ON JEDI AC/DC BD
(For Jedi without Front Cover) For Single Plane, open C1/C-FRT Cabinet (for Optima IGS systems,
open System control cabinet; for Biplane, open C1 FRT or C1 LAT or C-FRT or C-LAT Cabinet)
and remove the JEDI inverter cover, refer to:
(For Innova Single Plane and Discovery™ IGS Systems with C1 Cabinet) DR0187 - Jedi KVCTL
Board , steps 1 to 4 of the Disassembly procedure.
(For Innova Single Plane, Discovery™ IGS Systems and Allia™ IGS Systems and Allia™ IGS 3 with
C-FRT Cabinet) DR1161 - Jedi KVCTL Board , steps 1 to 4 of the Disassembly procedure.
(For Innova Biplane IGS Systems with C1 FRT and C1 LAT Cabinet) DR0187 - Jedi KVCTL Board ,
steps 1 to 4 of the Disassembly procedure.
(For Innova Biplane IGS Systems with C-FRT and C-LAT Cabinet) DR1161 - Jedi KVCTL Board , steps
1 to 4 of the Disassembly procedure.
(For Optima Systems) DR0871 - Jedi KVCTL Board , steps 1 to 4 of the Disassembly procedure.

46-017561 Revision 37 HHS 41/68


1.8 REG0057 - Tech & Manual mA Accuracy - Tool 2389357-2 or 5480040 Control and Tube Assembly Tests
or 5507803

4. (For Jedi without Front Cover) Install the JEDI cover tool (2404211 or 2404211-20 or 5480038):
put in place the plastic cover tool for JEDI inverter protection [1].
Figure 1-12

5. Remove the flat cable in J1 of KVCTL board ([2], Figure 1-12 on page 42).
6. Plug the J1 [3] of the JEDI mA shunt measurement tool [4] (2389357-2 or 5480040/5507803) in
the J1 connector of the KVCTL board.
Figure 1-13

7. Plug the flat cable in the JEDI mA shunt measurement tool (male side, connector J3) ([5],
Figure 1-13 on page 42).

42/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.8 REG0057 - Tech & Manual mA Accuracy - Tool 2389357-2 or 5480040
or 5507803

8. Connect the cables ([6], Figure 1-13 on page 42) of the multimeter ([7], Figure
1-13 on page 42) to the banana plugs (mA and shunt–B) of the JEDI mA shunt measurement
tool.

Test Procedure 1

HHS Control and Tube Assembly Tests


mA shunt resistor measurement with JEDI OFF
1. Select Ohmmeter function on multimeter.
2. Select maximum display accuracy function on multimeter.
3. Connect the multimeter between Ground (Black pin) and Shunt A (Red pin) on the mA tool.
4. Record the Shunt A resistor value with 2 decimal places (measured around 5 ohms).
5. Connect the multimeter between Ground (Black pin) and Shunt B (Blue pin) on the mA tool.
6. Record the Shunt B resistor value with 2 decimal places (measured around 0 ohm).

mA Shunt resistor accuracy verification


1. Record and compile values according to Table 1-20 Test Angio Parameters on page 45 and
Table 1-21 Test Cardiac Parameters on page 45.
2. The rejection limit is 5 ohms ± 2%.

Power ON the System


1. Verify connection of the multimeter between ”mA” (Green pin) and Shunt B (Blue pin) on the
mA tool.
2. For security, be sure the inverter cover tool is installed and secured with screws.
3. Power ON the system to carry out the following tests.

Measurement procedure
1. Put in place the DLB (Detector Lead Blocker).
2. Select the multimeter in Vdc voltage function.
3. Select the Record MAX display Mode (if not available on the multimeter, the operator must read
the maximum value displayed on the multimeter during each measurement).
4. (For Allia IGS systems) Set the system in manual mode using the X220 unit in the SUIF (refer to
Figure 1-14 on page 44).

46-017561 Revision 37 HHS 43/68


1.8 REG0057 - Tech & Manual mA Accuracy - Tool 2389357-2 or 5480040 Control and Tube Assembly Tests
or 5507803

4.1. Enter in the SUIF (on the DL browser, click on the knife tool icon).
Figure 1-14

NOTE
mA average value is displayed in the status bar.
4.2. Click on Calibrations button.
4.3. Select Regulatory and Dose Measurement.
4.4. Select X220 (for Single Plane) or Select X220 Frontal or X220 Lateral (for Biplane).
5. (For Innova systems, Discovery Systems and Optima systems) Set the system in manual mode
using the X220 unit in the SUIF (refer to Figure 1-15 on page 44).
5.1. Enter in the SUIF (on the DL browser, click on the knife tool icon).
Figure 1-15

44/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.8 REG0057 - Tech & Manual mA Accuracy - Tool 2389357-2 or 5480040
or 5507803

5.2. Click on Calibration button.


5.3. Select HHS and Dose Measurement.
5.4. Select X220 Generic (for Single Plane) or Select X220 Frontal or X220 Lateral (for
Biplane). 1
6. Choose exposure technics parameters according to the Tables listed below.

HHS Control and Tube Assembly Tests


Table 1-20 Test Angio Parameters

Multimeter Range Measurement on Vdc Test 1 Test 2 Test 3 Test 4 Test 5 Test 6
Exp Mode DSA DSA DSA DSA DSA DSA
Frame Rate 3.75 fps 3.75 fps 3.75 fps 3.75 fps 3.75 fps 3.75 fps
Focal Spot 0.3 0.3 0.6 0.6 1 1
kVPeak 50 125 50 125 50 125
mAPeak 40 40 80 80 200 200
Pulse Width 100 ms 100 ms 100 ms 100 ms 100 ms 100 ms
Spectral Filter None None None None None None
DLB Presence Yes Yes Yes Yes Yes Yes
Collimator LONG Leave Empty
Collimator LAT Leave Empty

Table 1-21 Test Cardiac Parameters

Multimeter Range Measurement on Vdc Test 7 Test 8 Test 9 Test 10


Exp Mode Cardiac Cardiac Cardiac Cardiac
Frame Rate 30 fps 30 fps 30 fps 30 fps
Focal Spot 0.6 0.6 1 1
kVPeak 60 120 60 120
mAPeak 80 80 200 200
Pulse Width 9 ms 9 ms 9 ms 9 ms
Spectral Filter None None None None
DLB Presence Yes Yes Yes Yes
Collimator LONG Leave Empty
Collimator LAT Leave Empty

7. Select Spectral filter: None, DLB presence = Yes, Collimator Long / Lat: Leave empty.
8. Select SUBMIT to set all parameters.
9. Depress the hand switch to initiate the X–Ray exposure during approximately 20 seconds to
allow multimeter acquisition stabilization (Maintain the X–Ray exposure until maximum value
is recorded by multimeter).
10. Record the multimeter voltage value displayed.
11. Record the mA_average value updated at exposure end displayed in the X220 status bar.
12. Record and compile values according to Table 1-20 Test Angio Parameters on page 45 and Table
1-21 Test Cardiac Parameters on page 45.
13. After each measurement, if the max record mode is used, reinitialize the multimeter for a new
record.
14. Repeat all previous actions to perform all 10 tests.

46-017561 Revision 37 HHS 45/68


1.8 REG0057 - Tech & Manual mA Accuracy - Tool 2389357-2 or 5480040 Control and Tube Assembly Tests
or 5507803

15. When all tests are performed, click on ABORT to exit.

SpreadSheet fill in procedure


1. Fill in data
1.1. Before data fill in, choose the right spreadsheet in accordance to your system as SP and
Frontal for Single Plane or Frontal Biplane, and Lateral for Lateral Biplane.
1.2. Furthermore, in each spreadsheet you need to select the High Voltage cable length
equipment (24 meters or 28.7 meters or 30 meters) and the mA measurement tool
procedure, after, you will be enter the following values:
• The measured resistor values (in Ohm, with 2 decimal places) displayed on
multimeter.
• The measured voltage (in mVdc) displayed on multimeter.
• The mA_avg value (in mA) displayed on X220 status bar.
2. Computation Details
2.1. For mA Shunt resistor verification test:
A calculated mA Shunt resistor value is automatically computed as following formula
–––>(Shunt A resistor - Shunt B resistor); the rejection limits are 4.90 Ohm (LSL) and 5.10
Ohm (USL) for PASS / FAIL test.
2.2. For each mA accuracy station test:
• A calculated mA_avg value (without correction) is automatically computed as
following formula–––>(mVdc / 5Ω).
• A corrected mA_avg value (with correction) is automatically computed as following
formula –––>[(mVdc / 5Ω) – mA_avg correction value @ HV cable length] see
following Table for mA_avg correction values.
• This corrected mA_avg value is automatically compared to the rejection limit @
selected mAmeasurement tool procedure (see following Table for rejection limit
values) for PASS / FAIL test.
• The mA_avg value returned in X220 status bar is automatically compared to the
rejection limit (the Table for rejection limit values is the same as Procedure A (±
15%)) for PASS / FAIL test.
NOTE
For DiscoveryTM IGS systems refer to column “mA_avg correction value @
28.70 m HV cable (mA)”.

TEST No. mA_avg mA_avg mA_avg LSL rejec- mA_avg De- USL rejec-
correction correction correction tion limit (– mand value tion limit
value @ 24 value @ value @ 30 12% of (mA) (+12% of
m HV cable 28.70 m HV m HV cable mA_avg De- mA_avg De-
(mA) cable (mA) (mA) mand) (mA) mand) (mA)
1 0.46 0.51 0.53 13.2 15 16.8
2 1.16 1.28 1.31 13.2 15 16.8
3 0.48 0.53 0.55 26.4 30 33.6
4 1.21 1.33 1.37 26.4 30 33.6
5 0.55 0.60 0.61 66 75 84
6 1.37 1.49 1.53 66 75 84
7 0.15 0.15 0.15 19.01 21.6 24.19

46/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.8 REG0057 - Tech & Manual mA Accuracy - Tool 2389357-2 or 5480040
or 5507803

TEST No. mA_avg mA_avg mA_avg LSL rejec- mA_avg De- USL rejec-
correction correction correction tion limit (– mand value tion limit
value @ 24 value @ value @ 30 12% of (mA) (+12% of
m HV cable 28.70 m HV m HV cable mA_avg De- mA_avg De-
(mA) cable (mA) (mA) mand) (mA) mand) (mA)
1
8 0.30 0.30 0.30 19.01 21.6 24.19

HHS Control and Tube Assembly Tests


9 4.83 5.31 5.44 47.52 54 60.48
10 9.67 10.62 10.88 47.52 54 60.48

3. Removal the mA Shunt Tool and JEDI Cover Tool


3.1. Switch OFF the system.
3.2. WARNING

BEFORE REMOVING JEDI INVERTER COVER TOOL, WAIT UNTIL ALL LED’S
ARE EXTINGUISHED ON JEDI AC/DC BD.
Remove the multimeter.
3.3. (For Jedi without Front Cover) Remove the inverter cover tool (2404211 or 2404211-20 or
5480038).
3.4. Remove the mA shunt tool (2389357-2 or 5480040 or 5507803).
3.5. Reinstall the flat cable in the J1 connector.
3.6. (For Jedi with Front Cover) Reinstall the front cover and attach it with the four captive
screws.
3.7. (For Jedi without Front Cover) Reinstall the Inverter metallic cover, refer to:
(For Innova Single Plane and Discovery™ IGS Systems with C1 Cabinet) DR0187 - Jedi KVCTL
Board , Reassembly procedure.
(For Innova Single Plane and Discovery™ IGS Systems and Allia™ IGS Systems and Allia™ IGS
3 with C-FRT Cabinet) DR1161 - Jedi KVCTL Board , Reassembly procedure.
(For Innova Biplane IGS Systems with C1 FRT and C1 LAT Cabinet) DR0187 - Jedi KVCTL
Board , Reassembly procedure.
(For Innova Biplane IGS Systems with C-FRT and C-LAT Cabinet) DR1161 - Jedi KVCTL
Board , Reassembly procedure.
(For Optima Systems) DR0871 - Jedi KVCTL Board , Reassembly procedure.
3.8. Close Auxiliary ACDC assembly in the C1/C-FRT cabinet (or System control cabinet for
Optima IGS systems).
3.9. Remove the Lead Blocker.
3.10. Switch ON the system.
3.11. Carry out an X–Ray expose to control the system operation.

Finalization
No finalization steps.

46-017561 Revision 37 HHS 47/68


1.9 REG0153 - GAIA mA Internal Accuracy measurement Control and Tube Assembly Tests

1.9 REG0153 - GAIA mA Internal Accuracy measurement


Table 1-22 Personnel requirements

Required persons Preliminary require- Procedure Finalization


ments
1 Field engineer Defined at system level 5 minutes 10 minutes

Table 1-23 Tools and test equipment

Item Quantity Effectivity Part number Manufacturer


Amperemeter 1 - - -
Current DC Accuracy:
1 % maximum
Gaia mA measure- 1 - 5933222 -
ment Tool

Table 1-24 Safety

WARNING

RESIDUAL VOLTAGE PRESENT


BEFORE SERVICING THE CABINET, AFTER HAVING POWERED IT OFF, ENSURE THAT
THE GENERATOR HAS NO RESIDUAL VOLTAGE.

Figure 1-16 Fluke 175 model with Engineering leads

mA measurement
Test Purpose
This test verifies the mA measurement circuit accuracy of GAIA Generator.

48/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.9 REG0153 - GAIA mA Internal Accuracy measurement

Test setup
1. Turn Off the system and verify the Tank power source is OFF  The 3 LEDs on the DC Bus door
(lit on the picture below) shall be OFF:
1

HHS Control and Tube Assembly Tests

46-017561 Revision 37 HHS 49/68


1.9 REG0153 - GAIA mA Internal Accuracy measurement Control and Tube Assembly Tests

2. Remove the generator Jumper located on the top of the Tank (see item circled in red on picture
below):

3. Connect the mA measurement tool on the Tank measure board: the snipped corners side of the
black connector shall be facing the back of the cabinet.

WARNING

The tool and Jumper shall be plugged and removed when the system is
completely powered Off. No exposure shall be made with the connector on
measure board empty.

50/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.9 REG0153 - GAIA mA Internal Accuracy measurement

HHS Control and Tube Assembly Tests


3.1. Connect the mA measurement tool leads provided by the engineering team on the
Amperemeter.
3.2. Select the 10A caliber on the Amperemeter (or highest caliber available).
3.3. Ensure that the Amperemeter is in DC current measurement mode.
NOTE
Don’t use the Min/Max, Average nor Memory mode, use continue RMS measure
mode.

Test procedure
1. Open the SUIF and launch the X220 Calibration; install the Detector Lead Blocker.
2. Perform the following exposures by pressing the pedal for 5 seconds and record the current
measured by the Amperemeter.

Settings Measure 1 Measure 2


Exp Mode Cardiac Cardiac
Frame Rate 30 fps 30 fps
Focal Spot 0.8 0.8
kVPeak 60 120
mAPeak 200 200
Pulse Width 10 ms 10 ms

Table 1-25 Acceptance criteria

TP LSL (mA) USL (mA)


1 54.1 66.1
2 54.2 66.2

NOTE
Record the lowest value displayed if not stable after 5s.
In the case the measured values are out of range, contact your Service Support to confirm the
Power Unit needs to be replaced.

46-017561 Revision 37 HHS 51/68


1.9 REG0153 - GAIA mA Internal Accuracy measurement Control and Tube Assembly Tests

Finalization
1. Turn OFF the system and make sure there is no more power applied on the Tank (see info at
beginning of Test setup on page 49.
2. Then removed the mA measurement tool and put back in place the Jumper on the Tank
measure board.

WARNING

The tool and Jumper shall be plugged and removed when the system is
completely powered Off. No exposure shall be made with the connector on
measure board empty.
3. Remove the Detector Lead Blocker and exit the X220 application.
4. Turn ON the system and verify acquisitions are allowed.

52/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.10 REG0058 - Technique Accuracy - kV (20, 30 and 40 cm Detector)

1.10 REG0058 - Technique Accuracy - kV (20, 30 and 40


cm Detector) 1
Table 1-26 Personnel requirements

HHS Control and Tube Assembly Tests


Required persons Preliminary require- Procedure Finalization
ments
1 10 minutes 45 minutes 10 minutes

Table 1-27 Tools and test equipment

Item Quantity Effectivity Part number Manufacturer


Non-invasive divider 1 - - -
with an accuracy ≤ ±
3% as recommended
in Illustrations of this
procedure

Table 1-28 Safety

WARNING

RISK OF ELECTRIC SHOCK.


HAZARDOUS VOLTAGE PRESENT.

WARNING

BEFORE ANY INTERVENTION POWER OFF THE SYSTEM (REFER TO PROTECTION


AGAINST ELECTRICAL HAZARDS IN SAFETY AND REGULATORY).

WARNING

RESIDUAL VOLTAGE PRESENT.


BEFORE SERVICING THE CABINET, AFTER HAVING POWERED IT OFF, ENSURE THAT
JEDI HAS NO RESIDUAL VOLTAGE

• (For Allia™ IGS 5 System (from SP5), Allia™ IGS 7 and Allia™ IGS 7 OR Systems
(from CFG8)) Application: kVp accuracy test must be performed at installation, planned
maintenance, and repair (FRU Gaia power unit).
• (For all systems except Allia™ IGS 5 System (from SP5), Allia™ IGS 7 and Allia™ IGS 7 OR
Systems (from CFG8)) Application: kVp accuracy test must be performed at installation and
repair (Generator FRU, HV tank, PPC kV CTL Board).
• Requirements: “Specified Accuracy” refers to technical accuracy statements written in Product
Specification section of operator manuals.

kVp Accuracy: Displayed kVp shall match the measured value within the specified accuracies.

NOTE
Allia™ IGS 5 and Allia™ IGS 7 and Allia™ IGS 7 OR are also named Allia™ IGS Systems. The
procedure is applicable indifferently to Allia™ IGS 5 or Allia™ IGS 7 or Allia™ IGS 7 OR.

46-017561 Revision 37 HHS 53/68


1.10 REG0058 - Technique Accuracy - kV (20, 30 and 40 cm Detector) Control and Tube Assembly Tests

Procedure
1. Use the non–invasive divider with an accuracy =< ± 3% as recommended in Illustrations below.
Figure 1-17 Detector Positioning for kVp Measurement

Figure 1-18 Detector Positioning (vertical) for kVp Measurement

2. In SUIF, set the system in manual mode using the X220 unit in Regulatory and Dose
Measurements from Calibration.
3. Select:
3.1. Expose Mode: DSA
3.2. Frame Rate: 3.75 fps
3.3. Focal Spot:
• (For all systems except Allia™ IGS 5 System (from SP5), Allia™ IGS 7 and Allia™ IGS 7 OR
Systems (from CFG8)) 1

54/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.10 REG0058 - Technique Accuracy - kV (20, 30 and 40 cm Detector)

• (For Allia™ IGS 5 System (from SP5), Allia™ IGS 7 and Allia™ IGS 7 OR Systems (from
CFG8)) 0.8
3.4. kVpeak, mApeak, Pulse Width: Refer to Table 1-29 on page 55.
3.5. Spectral filter: None 1
3.6. DLB presence = Yes

HHS Control and Tube Assembly Tests


3.7. Collimator Long / Lat: Let empty.
4. Select SUBMIT to set all parameters.
5. Perform the exposures under the parameters specified in the table below and at each end of
exposures fill in the table below with both the kVp displayed (check in X220 display status bar)
and the kVp measured.
6. If the kVp measures are out the specifications limits :
• (For all systems except Allia™ IGS 5 System (from SP5), Allia™ IGS 7 and Allia™ IGS 7
OR Systems (from CFG8)): change the PPC kV CTL Board, refer to:
◦ (For Innova and Discovery™ IGS Systems) DR0187 - Jedi KVCTL Board.
◦ (For Allia™ IGS Systems and Allia™ IGS 3) DR1161 - Jedi KVCTL Board.
◦ (For Optima Systems) DR0871 - Jedi KVCTL Board.
or (and) the HV tank, refer to:
◦ (For Innova and Discovery™ IGS Systems) DR0191 - Jedi Inverter or HV Tank.
◦ (For Allia™ IGS Systems and Allia™ IGS 3) DR1163 - Jedi Inverter or HV Tank.
◦ (For Optima Systems) DR0874 - Jedi Inverter or HV Tank.
• (For Allia™ IGS 5 System (from SP5), Allia™ IGS 7 and Allia™ IGS 7 OR Systems (from
CFG8)): change the Power Unit (refer to DR1352 - Gaia Power Unit).
Table 1-29

Mode: DSA on large Focus spot (Cine HP generator mode)


Test # kV mA Pulse Width fps Displayed Lower Spec. Measured Upper Spec.
(ms) kVp Limit (kVp) = kVp (actual Limit (kVp) =
Displayed kVp x kVp) Displayed kVp x
0.93 1.07
Test 1 75 150 100 3.75
Test 2 100 150 100 3.75

Figure 1-19

46-017561 Revision 37 HHS 55/68


1.10 REG0058 - Technique Accuracy - kV (20, 30 and 40 cm Detector) Control and Tube Assembly Tests

7. Record the data (apply any corrective factor if required and check final results are within
acceptance criteria).

Finalization
No finalization steps.

56/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.11 REG0059 - Technique Accuracy - mAs

1.11 REG0059 - Technique Accuracy - mAs


Table 1-30 Personnel requirements
1
Required persons Preliminary require- Procedure Finalization

HHS Control and Tube Assembly Tests


ments
1 Not Applicable Not Applicable Not Applicable

Table 1-31 Required conditions

Condition Reference Effectivity


When the equipment is connected - -
to an adequate power supply as
specified, the deviation of the tube
current/exposure time product shall
not exceed the limits given in speci-
fications.

Application
AMX, VMX, MPX/SPX, Advantx, and any other controls that have mAs settings; at installation,
preventive maintenance calls, after repair or replacement of mAs or timer circuits, and
replacement of tube unit. If the control allows selections of mAs in addition to the normal
selection of mA and time, then you must perform this test in addition to tests for mA and time.

Procedure
1. Procedures for making this test are given in the individual service manual functional check.
Also refer to REG0054 - Technique Accuracy - Generator Timer on page 21.
2. Record data: enter required data on Form 3382 on laptop computer.

Finalization
No finalization steps.

46-017561 Revision 37 HHS 57/68


1.12 REG0060 - Indirect Linearity Control and Tube Assembly Tests

1.12 REG0060 - Indirect Linearity


Table 1-32 Personnel requirements

Required persons Preliminary require- Procedure Finalization


ments
1 Not Applicable Not Applicable Not Applicable

Table 1-33 Required conditions

Condition Reference Effectivity


For convenience, this test uses kVp - -
and mA accuracy data for an indirect
test of linearity. If system passes test
of REG0054 - Technique Accuracy -
Generator Timer on page 21, this
section is also satisfied and no fur-
ther testing is necessary.

Application
Linearity: Radiographic and Spot-Film tube units supplied by the same generator. At installation,
periodic maintenance, and repair. The requirements apply even across focal spots, e.g., highest mA
on small spot compared to low mA on large spot. An exception is if the small spot is .45 mm or less.
For equipment manufactured after May 3, 1994, the requirements will be extended to units that
have only mAs selections.
Linearity requirements apply to all tube units used directly for filming and which can be operated
at more than one mA station. Consequently, linearity requirements do not apply to:
• Spot-film tube operated at one mA value.
• Tube used only with an image intensifier, whether for fluoro or for recording.

Procedure
NOTE
If a system passes the Indirect Linearity test, there’s no need to perform the Direct
Linearity test, unless some of the indirect test data is marginal. If System fails Indirect
Linearity you must either verify compliance via the Direct Linearity Test or readjust the
system and perform the Indirect Test again.
1. Perform test of REG0054 - Technique Accuracy - Generator Timer on page 21.
2. Ensure the test is passed.

Finalization
No finalization steps.

58/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.13 REG0061 - AEC Maximum mAs

1.13 REG0061 - AEC Maximum mAs


Table 1-34 Personnel requirements
1
Required persons Preliminary require- Procedure Finalization

HHS Control and Tube Assembly Tests


ments
1 Not Applicable Not Applicable Not Applicable

Table 1-35 Required conditions

Condition Reference Effectivity


At tube potentials of 50 kV or more, - -
exposures may not exceed 600 mAs
per exposure: below 50 kV, expo-
sures may not exceed 2,000 mAs.
Requirements do not apply under
any type of equipment failure mode.

Application
X-Ray units with AEC at installation, and periodic maintenance. Does not apply to image
intensifiers.

For Advantx Systems


Use the Calibration Tool Unit T005 MAS Integrator Test, from the appropriate Advantx Calibration
User’s Guide Direction, to functionally test the 600 mAs backup mAs integrator.

For Non-Advantx Systems


Procedure is covered in the Functional Check chapter of the appropriate generator or automatic
exposure control service manual.

Record Data
Record that requirements have been met on laptop computer Form 3382.

Finalization
No finalization steps.

46-017561 Revision 37 HHS 59/68


1.14 REG0062 - AEC Minimum Exposure Time Control and Tube Assembly Tests

1.14 REG0062 - AEC Minimum Exposure Time


Table 1-36 Personnel requirements

Required persons Preliminary require- Procedure Finalization


ments
1 Not Applicable Not Applicable Not Applicable

Table 1-37 Required conditions

Condition Reference Effectivity


At 50 kV or more, minimum expo- - -
sures must be less than 16 ms or
less than the time to deliver 5 mAs,
whichever time is greater.

Application
X-Ray units with AEC, at installation and periodic maintenance. Does not apply to image
intensifiers.

Procedure
1. Set 50 kV, normal AEC density, max. mA, .1 s (or 10 mAs) backup.
2. Align X-Ray beam to detector per standard technique.
3. Lower tube to 40 SID if possible, otherwise set for minimum.
4. Remove all phantoms.
5. Make an exposure in AEC mode.
6. The exposure must either be less than 16 ms, or less than 5 mAs.
7. Repeat the above procedure for each detector segment and for all detectors in the room. Do
not re-align the X-Ray beam to individual detector segments.
8. Record data: record that requirements have been met on laptop computer Form 3382.

Finalization
No finalization steps.

60/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.15 REG0063 - Reproducibility Of Exposure

1.15 REG0063 - Reproducibility Of Exposure


Table 1-38 Personnel requirements
1
Required persons Preliminary require- Procedure Finalization

HHS Control and Tube Assembly Tests


ments
1 Not Applicable Not Applicable Not Applicable

Table 1-39 Tools and test equipment

Item Quantity Effectivity Part number Manufacturer


Radiation meter with 1 - - -
integrate mode.
Probe test stand. 1 - - -

Table 1-40 Consumables

Item Quantity Effectivity Part number Manufacturer


1–1/2 (3.8 cm) thick- AR - - -
ness of aluminum
(from HHS kit), for
use with A.E.C. sys-
tems only.

Table 1-41 Required conditions

Condition Reference Effectivity


For any specific combination of se- - -
lected technique factors, the esti-
mated coefficient of variation, C.V.,
of radiation exposures shall be no
greater than 0.05. Rejection limit
0.045. This requirement holds for
exposures made with and without
use of A.E.C. This does not apply to
image intensified applications, cine,
photospot or digital.

Application
At installation, preventive maintenance, and repair.

Procedure 1 - For exposures made with an A.E.C


If system passes Procedure 1 - For exposures made with an A.E.C on page 61, there is no need
to perform Procedure 2 - For exposures made without the use of A.E.C on page 62. Follow the
instructions that pop up on the laptop computer Form 3382 for this test.
1. For overhead X-Ray tubes, place 1–1/2 (3.8 cm) aluminum on the table. Center the radiation
probe 30 cm (12 in.) above the table. Adjust the SID to about 100 cm. Align the bucky. For
undertable X-Ray tubes, assemble the attenuation block assembly from the HHS kit and place
it on the table. Center the probe beneath the assembly. Position the spot-filmer about 30 cm
above the table.
2. Collimate to a field about 5 inches square (13 cm square).
3. Select 80 kVp and mA such that the exposures will be at least 0.1 second long.

46-017561 Revision 37 HHS 61/68


1.15 REG0063 - Reproducibility Of Exposure Control and Tube Assembly Tests

4. NOTE
Test is applicable at any kVp and mA setting.
Measure the exposure in mR for each of 10 exposures (data sheet provided). After each
exposure, recycle the rotor and the technique switches; that is, set the switches to other values
and then return them to the original values. Be sure to vary the time between spot-film transfer
and exposure start to record any filament heating variations.
5. Enter the data onto the laptop computer Form 3382. If system fails this test, there may be a
defective ion chamber.

Procedure 2 - For exposures made without the use of A.E.C


1. For either overhead or undertable X-Ray tubes, place the probe at approximately 100 cm from
the focal spot. Avoid probe saturation (refer to 46–017560, HHS TOOLS).
2. Select 80 kVp, 0.1 second, and 400 mA to yield 100 mR to 300 mR per exposure.
3. NOTE
Test is applicable at any kVp and mA setting.
Collimate to a field approximately 5 inches square, (13 cm square).
4. Measure the mR for each of 10 exposures. After each exposure, recycle the rotor and the
technic switches; that is, set the switches to other values and then return them to the original
values. Be sure to vary the time between spot–film transfer and exposure start to take into
account any filament heating variations.
5. Enter the data onto the laptop computer Form 3382. If system fails this test, there may be a
system defect. Repair as necessary.

Procedure 3 - For tomographic systems


NOTE
If the system is operable in non-tomo mode, the previous two procedures are also
applicable.
1. Place the probe on a stand 10 cm above the table. Adjust the focal plane height to 10 cm. On
transaxial systems, use the collimator light field to determine the appropriate height for the
probe.
2. Select common tomographic technique factors, e.g., 80 kVp, 100 mA, and adjust for a large
sweep angle at medium speed.
3. Make a sweep without radiation, and, using the collimator light field, ensure that the probe is
contained within the field during the entire sweep.
4. Select backup time longer than sweep time and measure the exposure in mR for each of 10
sweeps. After each exposure, recycle the rotor and the technique switches; that is, set the
switches to other values and then return them to the original values. Observe tomo motor
cooling times.

62/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.15 REG0063 - Reproducibility Of Exposure

5. Enter the data onto the laptop computer Form 3382. If system fails this test, there may be a
defective ion chamber.
Table 1-42 Mini-Troubleshooting Guide

Problem Possible Cause 1


Coefficient of • Poor kVp reproducibility (deadband). In AEC mode this affects kVp compensation on

HHS Control and Tube Assembly Tests


variation too some systems.
high • Improper rotor acceleration delay, improper spotfilm transfer delay, improper fila-
ment boost. All of these will determine if filament is up to temperature when expo-
sure occurs. On AEC systems, a variable filament temperature results in kVp variations
which lead to kVp compensation errors.
• On Tomographic Systems, variations in sweep speed or angle will affect reproducibili-
ty. Always test the non-tomo mode first.
• Line voltage regulation out of specification.

Finalization
No finalization steps.

46-017561 Revision 37 HHS 63/68


1.16 REG0064 - Direct Linearity Of Exposure Alternate Method Control and Tube Assembly Tests

1.16 REG0064 - Direct Linearity Of Exposure Alternate


Method
Table 1-43 Personnel requirements

Required persons Preliminary require- Procedure Finalization


ments
1 Not Applicable Not Applicable Not Applicable

Table 1-44 Tools and test equipment

Item Quantity Effectivity Part number Manufacturer


Radiation meter with 1 - - -
integrated mode.
Probe test stand 1 - - -

Application - For radiographic modes without AEC:


At installation, preventive maintenance, and repair. Use this method whenever indirect linearity
test fails.

Requirements
1. The average ratios of exposure to the selected mAs (mR/mAs) obtained at any two consecutive
tube current settings, shall not differ by more than 0.10 times their sum (rejection limit = 0.09).

That is:

where and are the averages of 10 mR/mAs values obtained at each of two consecutive
tube current settings.
NOTE

If the above relationship is rewritten as: then the


term on the right can be referred to as the coefficient of linearity. So defined, we
can state the requirement as the coefficient of linearity (C.L.)” shall not exceed
0.10 (rejection limit = 0.09).
2. You must make 10 exposures per mA station and not 4 as is common for FDA inspection.
Remember that the FDA is merely looking for non-compliance, but you are trying to prove
compliance (passing a test with 4 exposures does not guarantee compliance).

Procedure
1. For overhead X–Ray tubes, place the probe approximately 100 cm (39.37 in.) from the focal
spot. For undertable X-Ray tubes, attach the probe to a stand and place it about 30 cm (12 in.)
above the table.
2. Select kVp and mA equal to the values where the system failed or was marginal in the Indirect
Linearity Test. Adjust timer to at least .01 seconds but not less than 20 mAs.
3. Collimate to a field about 5 inches square (13 cm square).
4. Set the rate meter to integrate mode and measure the total exposure for 10 exposures. Recycle
the rotor but not the technique switches after each exposure.

64/68 HHS 46-017561 Revision 37


Control and Tube Assembly Tests 1.16 REG0064 - Direct Linearity Of Exposure Alternate Method

5. Enter the data onto the laptop computer form. If system fails this test, refer to Table 1-45 Mini–
Troubleshooting Guide on page 65.
Table 1-45 Mini–Troubleshooting Guide

Problem Possible Cause 1


C.L. too high • mA accuracy out of specification.

HHS Control and Tube Assembly Tests


• Timer accuracy out of specification.
• Probe saturation (refer to REG0052 - Generator Operator Indicators on page 17, Probe
Saturation Effects).
• Excessive cable capacitance due to very long cables combined with low mA selection –
long decay time.
• Line voltage regulation out of specification.
• Test not completed within one-hour limit.

Figure 1-20 Direct Linearity Chart

Finalization
No finalization steps.

46-017561 Revision 37 HHS 65/68


Page intentionally left blank
General Electric Company
283, rue de la Minière
Buc 78530 FRANCE

www.gehealthcare.com

You might also like