Professional Documents
Culture Documents
9588 PDF C06
9588 PDF C06
regarding these monitoring techniques constitute the first important steps in design
of a monitoring network.
• Type of pollutants
• Averaging time specified by air quality criteria or standards
• Expected pollutant levels
• Available resources
• Availability of trained personnel
• Presence in the air of interfering materials
In order to evaluate the current air quality and the effect of air pollution control
measures, most larger communities maintain monitoring networks. The EPA, as well
as most of the states, maintains its own surveillance networks. The ideal objective
when installing a monitoring network is to be able to obtain continuous real time
data. Table 6.1 lists standard methods of measurement. Only 3 of these standard
methods employ continuous or semicontinuous monitors.
No satisfactory device exists as yet for determining suspended particulate on a
continuous basis. However, Table 6.2 lists continuous methods which are now com-
monly employed for the measurement of the five regulated air pollutants.
TABLE 6.1
Federal Reference Methods
Pollutant Collection Method Analysis
TABLE 6.2
Continuous Monitor Methods for Federally Regulated Pollutants
Pollutant Continuous Monitor Method
OP = (1 − I I 0 ) × 100 (6.1)
where
OP = percent opacity
I = light flux leaving the plume
I0 = incident light flux
Particulate Matter
Appendix B
Reference Method for the Determination of Suspended Particulate in the Atmosphere — High Volume
Method (Total suspended particulate has been replaced by PM10 and PM2.5.)
Carbon Monoxide
Appendix C
Measurement Principle and Calibration Procedure for the Continuous Measurement of Carbon Monoxide
in the Atmosphere (Non-Dispersive Infrared Spectrometry)
Ozone
Appendix D
Measurement Principle and Calibration Procedure for the Measurement of Photochemical Oxidants
Corrected for Interferences Due to Nitrogen Oxides and Sulfur Dioxide
Hydrocarbons
Appendix E (Corrected for Methane)
Reference Method for Determination of Hydrocarbons Corrected for Methane
Nitrogen Dioxide
Appendix F
Measurement Principle and Calibration Procedure for the Measurement of Nitrogen Dioxide in the
Atmosphere (Gas Phase Chemiluminescence)
Lead
Appendix G
Reference Method for the Determination of Lead in Suspended Particulate Matter Collected from Ambient
Air
Ozone
Appendix H
Interpretation of the National Ambient Air Quality Standards for Ozone
PM10
Appendix J
Reference Method for the Determination of Particulate Matter as PM10 in the Atmosphere
Particulate Matter
Appendix K
Interpretation of the National Ambient Air Quality Standards for Particulate Matter (Includes PM2.5)
free of electronic drift over these long-time periods, and would possess suitable
dynamic response.
Many automated environmental surveillance systems employing continuous
monitors exist in the U.S. and throughout the world. None are quite as sophisticated
as would be implied by the system of Figure 6.1.
1. Gravimetric
2. Volumetric
3. Microscopy
4. Instrumental
a. Spectrophotometric
i. Ultraviolet
ii. Visible (Colorimetry)
iii. Infra-red
b. Electrical
i. Conductometric
ii. Coulometric
iii. Titrimetric
c. Emission Spectroscopy
d. Mass Spectroscopy
e. Chromatography
The EPA Emission Measurement Center (EMC) at Research Triangle Park and
the Midwest Research Institute have created a data base on CEM. You may use
browse techniques or search the data base by HAP or analyzer type. The data base
is found on the EPA home page at: www.epa.gov/ttnami1/siteindx.html.
Figure 6.3 is a schematic flow diagram of a general continuous air-quality mon-
itor. The device contains:
1. Primary air-moving device, usually a vacuum pump, to pull the air sample
through the instrument
2. Flow-control and -monitoring device, usually a constant pressure regulator
and rotameter
3. Pollutant detection by various primary sensing techniques
4. Automatic reagent addition where needed
5. Electronic circuitry for transducing the primary signal to a signal suitable
for recording and telemetering
6. Provisions for automatic calibration, usually several solenoid valves which
can be operated remotely to connect the inlet gas to a scrubbing train for
removal of all pollutants and establishing a chemical zero, or, alternatively,
to one or more span gases for setting the chemical range of the instruments.
Many monitors of the general type described above have been developed for all
of the federally regulated gaseous pollutants and others as well. The remainder of
this section will provide details of these devices. This list is not exhaustive. Further-
more, although the devices described below are indicated for a particular pollutant,
they can be used for other types of pollutants as well.
H 2O + SO2 → H 2SO3
H 2SO3 + 1 2 O2 → H 2SO 4
2I − → I 2 + 2e −
The SO2 is oxidized by the I2. Unreacted iodine is reduced to iodide, I–, at the
cathode.
I 2 + 2e − → 2I −
Due to the loss of I2 by reaction with SO2, the cathode current is less than the
anode current. The difference is proportional to the SO2 concentration through a
Faradaic expression. The reference electrode serves as a reference to both poles and
registers this imbalance in anodic input current and cathodic output current,
ir = ia = ic. A selective scrubber removes interferents such as O3, mercaptans, and
H2S, improving specificity of instrument. A zero gas is provided by using an activated
carbon filter to remove impurities in air, including SO2. The gas flow is regulated
by pressure control and a capillary tube which provides a constant pressure and
pressure drop, and thus a constant flow. A pulsation damper adds volume to the
system to provide stability of flow. The water supply is constantly replenished when
evaporation occurs.
Two infrared sources are used, one for the sample energy beam, the other for
the reference energy beam. The beams are blocked simultaneously ten times per
second by the chopper, a two-segmented blade rotating at five revolutions per second.
In the unblocked condition, each beam passes through the associated cell and into
the detector. The sample cell is a flow-through tube that receives a continuous stream
of sample. The reference cell is a sealed tube filled with a reference gas. The reference
gas is selected for minimal absorption of infrared energy of those wavelengths
absorbed by the sample component of interest.
The detector consists of two sealed compartments separated by a flexible metal
diaphragm. Each compartment has an infrared transmitting window, to permit entry
of the corresponding energy beam. Both chambers are filled, to the same subatmo-
spheric pressure, with the vapor of the component of interest. Use of this substance
as the gas charge in the detector causes the instrument to respond only to that portion
of net difference in energy due to the presence of the measured component.
In operation, the presence of the infrared-absorbing component of interest in the
sample stream causes a difference in energy levels between the sample and reference
sides of the system. This differential energy increment undergoes the following
sequence of transformation:
a. Radiant energy: In the sample cell, part of the original energy of the
sample beam is absorbed by the component of interest. In the reference
cell, however, absorption of energy from the reference beam is negligible.
b. Temperature: Inside the detector, each beam heats the gas in the corre-
sponding chamber. However, since energy of the reference beam is greater,
gas in the reference chamber is heated more.
c. Pressure: Higher temperature of gas in the reference chamber raises the
pressure of this compartment above that of the sample chamber.
d. Mechanical energy: Gas pressure in the reference chamber distends the
diaphragm toward the sample chamber. The energy increment is thus
expended in flexing the diaphragm.
e. Capacitance: The diaphragm and the adjacent stationary metal button
constitute a two-plate variable capacitor. Distention of the diaphragm
away from the button decreases the capacitance.
When the chopper blocks the beams, pressures in the two chambers equalize,
and the diaphragm returns to the undistended condition. As the chopper alternately
blocks and unblocks the beams, therefore, the diaphragm pulses, thus changing
detector capacitance cyclically. The detector is part of an amplitude modulation
circuit that impresses the 10 Hz information signal on a 10 MHz carrier wave
provided by a crystal-controlled radio-frequency oscillator. Additional electronic
circuitry in the oscillator unit demodulates and filters the resultant signal, yielding
a 10 Hz signal. The 10 Hz signal is routed to the amplifier/control section for
amplification and phase inversion, then back into the analyzer section for synchro-
nous rectification. The resulting fullwave-rectified signal is returned to the ampli-
fier/control section for filtering and additional conditioning, as required, to drive the
meter and recorder.
In 1889, Arrhenius first proposed that an energy of activation was required in order
for a chemical reaction to take place. By acquiring this activation energy, a reacting
molecule or atom could combine with other molecules or atoms in a chemical
reaction. Around 1914 to 1930, the collision theory of reaction prevailed in which
the molecules were postulated to acquire the necessary energy of activation by
colliding with each other. Molecules were then conceived to be somewhat like hard
spheres, and the rate of reaction was determined to be proportional to the number
2 NO + O2 → ( NO − O2 ) Activated
NO → ( NO)
‡
(NO)‡ → (NO − O3 )
‡
(NO − O3 ) → ( NO2 ) + O2
‡ ‡
5. NO2 drops to a lower energy state with the emission of a quanta of light, hv.
(NO2 )
‡
→ NO2 + hv
usefulness. Even more limited in specifying instruments is the rise time and fall
time, which provides no information about the shape of the transient curve. The bad
news is the difficulty to which the experimenter is put in order to determine the
transient characteristics or the transfer function. The good news is that the need for
the detailed transient information provided by the transfer function is not usually
required.
Additional instrument specifications are concerned with the effect of the ambient
conditions imposed on the instrument including interfering substances. A common
glossary of terms follows. These terms have been classified according to definition,
steady-state specification, dynamic specification, or ambient specification.
One of the most difficult problems encountered in calibrating air pollution instru-
mentation is the production of low-level concentrations which are in the range
encountered in the atmosphere. There are several suppliers of gas cylinders with
prepared and certified low-level gas concentrations.
Most usually these gases will have to be diluted before use, which poses a
problem in maintaining strictly controlled temperatures, pressures, and flow rates.
This writer discovered a paradox when using a cylinder of carbon monoxide prepared
to a specified low concentration. A small but significant difference between two low
levels of concentration was found when calibrating a nondispersive infrared analyzer
(NDIR). The companies supplying the gases were contacted. One company reported
that they were sure of their analysis because they used the best instrument for low-
level carbon monoxide concentration determination — a nondispersive infrared
instrument. The net result was that the company was comparing NDIR instruments
and not conducting a calibration with a true standard. From the author’s viewpoint
this situation is to be avoided.
The permeation tube is a device which can
provide low-level concentrations. These tubes TABLE 6.3
are made of a material such as Teflon and con- Available Permeation Tubes
tain a compound whose concentration is
desired, sealed into the tube as a liquid. The Sulfur dioxide Propane
Nitrogen dioxide Propylene
vapor will then permeate through the walls of
Hydrogen sulfide Butane
the tube at a constant rate dependent on the
Chlorine Butene-1
surface area exposed and the temperatures to Hydrogen fluoride Cis-butene-2
which the tube is subjected. The basic determi- Dimethyl sulfide Trans butene-2
nation of permeation rate can be made gravi- Dimethyl disulfide Ethylene oxide
metrically; thus, a permeation tube can become Methyl mercaptan Others
a true standard. Table 6.3 lists some of the
gases for which permeation tubes are available.
Tubes of proper length can be used to calibrate gas analyzers at a steady-state
flow condition. An apparatus must be prepared to hold the tube in a pure-air or
nitrogen stream of known flow rate and at a constant temperature. A tube holder
may be made from a standard vacuum trap with the tube in the center well and the
outer section filled with glass beads to promote heat exchange. The tube holder is
CF
L= (6.2)
K Pt
where
L = length in cm
C = concentration in ppm (vol)
F = flowrate of carrier gas in cm3/min
K = constant dependent upon gas at 1 atm, 25°C
0.382 for SO2; 0.532 for NO2; 0.556 for propane
Pt = permeation rate for stated operating temperature
REFERENCES
1. Lodge, J. P., Jr., Ed., Methods of Air Sampling and Analysis, Interscience Committee
of A&WMA, ACS, AIChE, APWA, ASME, AOAC, HPS, and ISA; Lewis Publishers
Inc., Chelsea, MI, 1989.
2. Glasastone, S., Laidler, K. J., and Eyring, H., The Theory of Rate Processes, McGraw-
Hill Book Co., New York, 1941.