Optical gas analyzer Optical gas analyzer

CI-PC20 CI-PC21
Infrared gas analyzer Infrared gas analyzer

CI-PC20 analyzer is an infrared gas detection device The selective absorption of infrared rays by gas
based on the principle of NDIR (non-spectroscopic infrared) molecules is the basis of the design of infrared gas
with micro-controller technology as the core.It has complete analyzer.Polar molecules such as SO2,NOx,CO,CH4 have
self-diagnosis function and monitors the working state of permanent electric dipole torches,and therefore have the
the instrument at any time.The analyzer has the advantag- structure of vibration and rotation.According to quantum
es of intelligence,stability,reliability and long service life.It
mechanics,the energy levels are divided into splitting
is suitable for the concentration measurement of carbon
levels,which can be coupled with the incident characteris-
dioxide,carbon monoxide,methane and other gases in
various industries and environments. tic long infrared radiation to produce absorption.Different
polar molecules have different absorption bands.

Product Features Product Features Technical Specifications


・Using NDIR (non-dispersive infrared measurement),the dimensions: LxW=225mmx180mm (rectangular hole) ・ The specially designed air chamber absorber has strong Maximum load:Provided according to user's requirement,not
sensor is not in contact with the measured gas,and the Fixed hole size:L*W=252mm*170mm (6*4 bolt hole) anti-interference ability and high detection accuracy. affected by load within 500Ω of maximum
poisoning phenomenon of the sensor will not occur. ・ High stability infrared light source,constant tempera- load
・High stability infrared light source and strong anti-inter- Principle ture control of receiver,good stability of instrument; Environmental
ference filter make the analyzer stable for a long time. ・ Measuring chamber mirror gold plating,corrosion temperature: -10~45°C
・With microprocessor as the core,it has the characteris- resistance; Environmental humidity:less than 80% RH
filter
tics of intelligence,stability,high precision and long calibra- ・ Isolated analog signal output,4-20mA/0-20mA/0-1V/ Power supply
tion cycle. 0-5V/0-10V/1-5V analog output mode can be set freely. voltage: 85-264VAC 50/60Hz
・Menu lock switch:prevent misoperation to change the ・ Alarm function:When the gas concentration deviates Shape size: 19-inch profile case
parameters of the analyzer to affect its normal opera- Infrared light source infrared sensor from the alarm value,the alarm value can be set arbitrari- LXWXH=483 mmX300 mmX177 mm
tion. ly in the full range. The measured
Infrared signal

・ RS485 (standard matching)/RS232 (optional matching) gas flow rate: 500-1000ml/min,the best:500 ml/min
Technical Specifications was adopted. Dust: Particles not exceeding1 um,not exceeding
200 ug/Nm3
Detection Mist: none
principle: NDIR,Non-dispersive Infrared (Double Beam) Technical Specifications Pressure: Atmospheric pressure ±10%(outlet should be
Measurement Model: CI-PC21 atmospheric pressure)
wavelength(μm)
Accuracy: ±2%FS Principle of Water content: Saturation (no condensation) not exceeding
Repeatability: ±1%FS Most organic and inorganic polyatomic molecular gases measurement:NDIR Non-Spectrophotometric Infrared Analyzer 5°C
Zero drift: ±1%FS/7d have specific absorption wavelength in the infrared Measuring
Range drift: ±1%FS/7d region.When infrared light passes through,the transmis- components: NOx,SO2,CO,CO2,CH4,H2S can be measured
Response time:T90<15s(500ml/min) sion intensity of these gaseous molecules to specific for any one or two components,and O2 can Applications
Power supply: 85~264VAC,50/60Hz wavelength can be expressed by Lambert-Beer law:I=I0 be added to each component.
Power consump- Range can be Minimum range Maximum range It is widely used in gas analysis of petroleum,chemi-
e-kpl,and the absorption intensity I can be expressed as designed according
tion of analyzer: Less than 35VA cal,smelting,building materials,light industry and other
I=I0-I=I0 (1-e-kpl).In the formula,I0 is the intensity of incident to requirements
Communication: RS485(Standard Matching)/RS232(Optional furnaces,kilns or flues.It is a reliable testing tool for environ-
light,I represent the intensity of transmission light,L is the NOx 0~1000ppm 0~20%
Matching) mental monitoring,bioengineering,medical and health research
thickness of gas medium,P is the concentration of gas,K SO2 0~500ppm 0~100%
Alarm: Alarm value: 2 work.
is the absorption coefficient.In the gas chamber,the infrared CO 0~20ppm 0~100%
Relay Contact Capacity: 24V DC, 0.2A light source emits the infrared band light,and the energy
Analog output CO2 0~10ppm 0~100%
of the light is absorbed by the gas in the gas chamber.The CH4 0~200ppm 0~100%
signal: Group 1,4-20mA/0-20mA/0-1V/0-5V/0-10V/ total energy absorbed corresponds to the concentration Order Information
1-5V,freely set mode H2S 0~1000ppm 0~30%
of all gases in the gas chamber.The radiation detector only
Load: less than 500Ω Accuracy: ±2%FS ・Pneumatic connection mode:
responds to this wavelength because it integrates an
External control:Quantity:5 groups Control signal:9-28VDC Repeatability: ±1%FS A,6 hose connection (standard)
infrared filter for a specific wavelength of the gas being
Environmental Zero drift: ±1%FS/7d B,1/4'-6 Stainless Steel Sleeve Connection
measured.The electrical signal obtained by the detector
temperature: -10 ~+45 ℃ Range drift: ±1%FS/7d C,1/8'-3 Stainless Steel Sleeve Connection
corresponds to the light energy absorbed by the gas under test.
Environmental Humidity:Less than 80% RH Linear error ・Filters:GF2800,AF2000;
Sample limit: ±1%FS ・Gas composition and measurement range under test;
pressure: Atmospheric pressure + 10% (outlet should Instructions Response time:T90≤60s(when sample flow gas volume is ・Measured gas pressure:positive pressure,micro pressure
be atmospheric pressure) ・Measuring Components and Range:See page 91 of this 0.5L/min) or negative pressure;
Cabinet size: L*W*H=310mm*275mm*190mm sample book. Analog signal: 4~20mA/0~20mA/0~1V/0-5V/0-10V/1-5V,- ・The main components,impurities and sulfides of the
Installation ・Application:Please refer to page 92 of this sample book. free to set analog output mode measured gas.