Calibration Report of Hioki DT4282 Digital Multimeter

MPC Control #: 001E Serial Number: 160839316

Asset ID: 001E Calibration Date: January 15, 2024

Section 1 - DC Voltage Accuracy Test


Lower Upper Uncertainty
Range Nominal As Found As Left Result
Limit Limit (±)
60 mV 1.000 mV 0.973 mV 1.003 mV 1.003 mV 1.027 mV PASS 0.00041 mV
60 mV 5.000 mV 4.965 mV 5.003 mV 5.003 mV 5.035 mV PASS 0.00044 mV
60 mV 10.000 mV 9.955 mV 10.002 mV 10.002 mV 10.045 mV PASS 0.00048 mV
60 mV 20.000 mV 19.935 mV 20.001 mV 20.001 mV 20.065 mV PASS 0.00055 mV
60 mV 30.000 mV 29.915 mV 30.000 mV 30.000 mV 30.085 mV PASS 0.00063 mV
60 mV 40.000 mV 39.895 mV 39.998 mV 39.998 mV 40.105 mV PASS 0.00070 mV
60 mV 54.000 mV 53.867 mV 53.998 mV 53.998 mV 54.133 mV PASS 0.00081 mV
600 mV 540.00 mV 539.82 mV 539.94 mV 539.94 mV 540.19 mV PASS 0.0034 mV
6V 5.4000 V 5.3985 V 5.3994 V 5.3994 V 5.4016 V PASS 0.000022 V
60 V 54.000 V 53.985 V 53.995 V 53.995 V 54.016 V PASS 0.00031 V
600 V 540.00 V 539.82 V 539.97 V 539.97 V 540.18 V PASS 0.0039 V

Section 2 - AC Voltage Accuracy Test @ 50 Hz


Lower Upper Uncertainty
Range Nominal As Found As Left Result
Limit Limit (±)
60 mV 54.000 mV 53.744 mV 54.019 mV 54.019 mV 54.256 mV PASS 0.011 mV
600 mV 540.00 mV 537.44 mV 540.23 mV 540.23 mV 542.56 mV PASS 0.032 mV
6V 5.4000 V 5.3867 V 5.4004 V 5.4004 V 5.4133 V PASS 0.00029 V
60 V 54.000 V 53.867 V 54.005 V 54.005 V 54.133 V PASS 0.0035 V
600 V 540.00 V 538.67 V 540.03 V 540.03 V 541.33 V PASS 0.18 V
1000 V 900.0 V 895.7 V 900.1 V 900.1 V 904.3 V PASS 0.29 V

Section 3 - DC Current Accuracy Test


Lower Upper Uncertainty
Range Nominal As Found As Left Result
Limit Limit (±)
0.1000 A 0.0973 A 0.1004 A 0.1004 A 0.1027 A PASS 0.000026 A
0.5000 A 0.4965 A 0.4994 A 0.4994 A 0.5035 A PASS 0.00011 A
6A
1.0000 A 0.9955 A 0.9984 A 0.9984 A 1.0045 A PASS 0.00019 A
5.4000 A 5.3867 A 5.3992 A 5.3992 A 5.4133 A PASS 0.0024 A
10 A 9.000 A 8.977 A 8.985 A 8.985 A 9.023 A PASS 0.0038 A

Section 4 - AC Current Accuracy Test @ 50 Hz


Lower Upper Uncertainty
Range Nominal As Found As Left Result
Limit Limit (±)
6A 5.4000 A 5.3548 A 5.3901 A 5.3901 A 5.4452 A PASS 0.0026 A
10 A 9.000 A 8.923 A 8.987 A 8.987 A 9.077 A PASS 0.0043 A

Confidential: Controlled Document. Page 1 of 3


Cannot be reproduced without the approval of the Quality Department. TQDG-FM-002 rev05
Calibration Report of Hioki DT4282 Digital Multimeter
MPC Control #: 001E Serial Number: 160839316

Asset ID: 001E Calibration Date: January 15, 2024

Section 5 - Resistance Accuracy Test


Lower Upper Uncertainty
Range Nominal As Found As Left Result
Limit Limit (±)
60 Ω 54.000 Ω 53.818 Ω 54.068 Ω 54.068 Ω 54.182 Ω PASS 0.0024 Ω
600 Ω 540.00 Ω 539.74 Ω 540.07 Ω 540.07 Ω 540.26 Ω PASS 0.014 Ω
6 kΩ 5.4000 kΩ 5.3982 kΩ 5.3998 kΩ 5.3998 kΩ 5.4018 kΩ PASS 0.00015 kΩ
60 kΩ 54.000 kΩ 53.982 kΩ 53.997 kΩ 53.997 kΩ 54.018 kΩ PASS 0.0023 kΩ
600 kΩ 540.00 kΩ 539.82 kΩ 540.04 kΩ 540.04 kΩ 540.18 kΩ PASS 0.016 kΩ
6 MΩ 5.4000 MΩ 5.3915 MΩ 5.4001 MΩ 5.4001 MΩ 5.4085 MΩ PASS 0.00066 MΩ
60 MΩ 54.00 MΩ 53.09 MΩ 54.04 MΩ 54.04 MΩ 54.91 MΩ PASS 0.38 MΩ
600 MΩ 300.0 MΩ 289.0 MΩ 301.9 MΩ 301.9 MΩ 311.0 MΩ PASS 1.7 MΩ

Section 6 - Continuity Check


Lower Upper Uncertainty
Range Nominal As Found As Left Result
Limit Limit (±)
20.0 Ω 19.4 Ω 20.1 Ω 20.1 Ω 20.6 Ω PASS 0.0017 Ω
600 Ω
250.0 Ω 248.3 Ω 250.1 Ω 250.1 Ω 251.8 Ω PASS 0.0071 Ω

Section 7 - Capacitance Accuracy Test


Lower Upper Uncertainty
Range Nominal As Found As Left Result
Limit Limit (±)
1 nF 0.900 nF 0.871 nF 0.909 nF 0.909 nF 0.929 nF PASS 0.017 nF
10 nF 9.00 nF 8.86 nF 9.07 nF 9.07 nF 9.14 nF PASS 0.030 nF
100 nF 90.0 nF 88.6 nF 90.1 nF 90.1 nF 91.4 nF PASS 0.30 nF
1 µF 0.900 µF 0.886 µF 0.901 µF 0.901 µF 0.914 µF PASS 0.0030 µF
10 µF 9.00 µF 8.77 µF 9.08 µF 9.08 µF 9.23 µF PASS 0.030 µF
100 µF 90.0 µF 87.7 µF 90.7 µF 90.7 µF 92.3 µF PASS 0.32 µF
1 mF 0.900 mF 0.877 mF 0.910 mF 0.910 mF 0.923 mF PASS 0.0045 mF

Section 8 - Frequency Accuracy Test


Lower Upper Uncertainty
Range Nominal As Found As Left Result
Limit Limit (±)
99.999 Hz 90.000 Hz 89.979 Hz 90.000 Hz 90.000 Hz 90.021 Hz PASS 0.00058 Hz
999.99 Hz 900.00 Hz 899.79 Hz 900.00 Hz 900.00 Hz 900.21 Hz PASS 0.0058 Hz
9.9999 kHz 9.0000 kHz 8.9979 kHz 9.0000 kHz 9.0000 kHz 9.0021 kHz PASS 0.000058 kHz
99.999 kHz 90.000 kHz 89.979 kHz 90.000 kHz 90.000 kHz 90.021 kHz PASS 0.00058 kHz
500.00 kHz 450.00 kHz 449.88 kHz 450.00 kHz 450.00 kHz 450.12 kHz PASS 0.0058 kHz

Confidential: Controlled Document. Page 2 of 3


Cannot be reproduced without the approval of the Quality Department. TQDG-FM-002 rev05
Calibration Report of Hioki DT4282 Digital Multimeter
MPC Control #: 001E Serial Number: 160839316

Asset ID: 001E Calibration Date: January 15, 2024

Section 9 - Temperature Accuracy Test


* Type K
Lower Upper Uncertainty
Range Nominal As Found As Left Result
Limit Limit (±)
-30.0 °C -32.9 °C -29.8 °C -29.8 °C -27.2 °C PASS 0.19 °C
( -40 ̴ 800 ) °C 0.0 °C -3.0 °C 0.1 °C 0.1 °C 3.0 °C PASS 0.18 °C
750.0 °C 743.3 °C 750.4 °C 750.4 °C 756.8 °C PASS 0.24 °C

Section 10 - Diode Accuracy Test


Lower Upper Uncertainty
Range Nominal As Found As Left Result
Limit Limit (±)
3.600 V 3.000 V 2.992 V 3.000 V 3.000 V 3.008 V PASS 0.00019 V

Note:The tolerance used for this calibration segment according to BS EN 50342-1:2015+A2:2021 / TCVN 7916-1 : 2008.

- End of Calibration Report -