Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Sizing - Pressure Performance

Project :
Customer: C.Project No.:
Contact person: Phone: Fax:
eMail:

TAG : ---
Timestamp: --- Review number: ---
Sales order number:

Sizing sheet
Operating Conditions
Minimum Nominal Maximum Unit
Ambient temperature 25 40 45 °C
Process temperature 50 60 75 °C

Product : Cerabar M PMC51


Sensor 40 bar/ 4000 kPa/ 600 psi gauge
Adjusted span 40 000 mbar
Turn down 1
Accuracy Standard
Output signal 4...20 mA
Calibration unit mbar/ bar
Date 17.02.2021

Extended Order Code


Quantity Item Description Extended Order Code options
1 Transmitter Cerabar PMC51-**1***1SGB

Calculation results

Reference accuracy 0.1 % adj. span

Minimum Nominal Maximum Unit


Total performance 0.1 0.6 0.6 % adj. span
Total error 1 year 0.2 0.7 0.7 % adj. span
Total error 5 years 0.35 0.85 0.85 % adj. span
Total error 10 years 0.5 1 1 % adj. span
Recommended Maintenance <1 <1 <1 years

1 year 5 years 10 years Unit


Long-term stability 0.1 0.25 0.4 % of URL

Warning(s)

Print date: 17.02.2021 08:30 AM -1/2-


Applicator®: 11.10.02 / 20.08.01

Under no circumstances is Endress+Hauser liable for errors, neither in the Software and in its documentation, nor for any errors
and consequential damage which may arise out of their use. The results in Applicator apply to parameters entered by the user. A
change in these parameters could lead to different results. Mandatory data can be found in the corresponding Operating Instruction
(BA) and Technical Information (TI).
Sizing - Pressure Performance
Project :
Customer: C.Project No.:
Contact person: Phone: Fax:
eMail:

TAG : ---
Timestamp: --- Review number: ---
Sales order number:

Chart sheet
Chart

Results (based on nominal values)


Turn down 1:1 2:1 3:1 4:1 5:1 Unit
Total performance 0.60 0.76 0.91 1.07 1.22 % adj. span
Reference 0.10 0.10 0.10 0.10 0.10 % adj. span
accuracy

Print date: 17.02.2021 08:30 AM -2/2-


Applicator®: 11.10.02 / 20.08.01

Under no circumstances is Endress+Hauser liable for errors, neither in the Software and in its documentation, nor for any errors
and consequential damage which may arise out of their use. The results in Applicator apply to parameters entered by the user. A
change in these parameters could lead to different results. Mandatory data can be found in the corresponding Operating Instruction
(BA) and Technical Information (TI).