PC: Yes
Dimensions: Depth:740mm;Height:560mm;Lenght:860mm
Weight: 145 kg
UPS: ≥2000VA
Drainage:
• The length of the waste tube is less than 5m and greater than 1m.
• The waste outlet should be no less than 8mm wide.
• The waste outlet should be lower than sewer for at least 0.5m.
•
Chemistry Analyzer
PC: Yes
Dimensions: Depth:660mm;Height:550mm;Lenght:860mm
Weight: 130 kg
UPS: ≥3000VA
Drainage:
• The analyzer discharges high-concentration waste and low-concentration waste separately with
high concentration waste discharged into a waste tank and low-concentration waste to the
sewer or the low concentration waste tank. Follow the local regulations for waste discharge.
• When Low-concentration waste is discharged to a sewer, the diameter of the discharge outlet
must be no less than 50mm and the height of the discharge outlet must be no higher than
100mm.
Hematology Analyzer
PC: Yes
Dimensions: Depth:840mm;Height:600mm;Lenght:400mm
Weight: 70.6 kg
UPS: ≥1500VA
Drainage:
Coagulation Analyzer
PC: No
Dimensions: Depth:410mm;Height:130mm;Lenght:380mm
Weight: 4.95 kg
UPS: ≥500VA
Drainage: NA
Urine Analyzer
PC: No
Dimensions: Depth:330mm;Height:210mm;Lenght:390mm
Weight: ~6 kg
UPS: ≥500VA
Drainage: NA
SUGGESTED LOCATION
Hole
- The purpose of this hole is for the reagent tubing, waste, and water.
- Reagents, waste, and water is placed below the analyzer.
- Below is a sample hole (just remove the cover).