Professional Documents
Culture Documents
05 Ise PDF
05 Ise PDF
05 Ise PDF
ISE
5-0
5. ISE
Item Specification
Measurable samples Serum, urine, blood
plasma
Measuring temperature 35 ± 2°C
Throughput Continuous measurement 100 samples/hr max.
Measurement system Dilution to 1/50 Sample volume: 10 µL
Dilution volume: 490 µL
Data accuracy Data reproducibility Simultaneous reproducibility:
Serum CV ≤ 1%
(N = 30) Urine CV ≤ 2%
Data accuracy Normal standard serum:
Na, Cl ±1.5 mmol/L
K ±0.1 mmol/L
Measuring concentration Serum Na : 80 to 180 mmol/L
range K : 1.5 to 10 mmol/L
Cl : 60 to 120 mmol/L
Urine Na, Cl : 10 to 250 mmol/L
K : 1 to 100 mmol/L
Maintenance Rinsing operation Kind : Daily
Method : Automatic, pipetting from sample
cup
Reagent priming Internal standard, reference electrode
solution
Potential measurement Measurement of electromotive force of
Na/K/Cl in stand-by status
Sampling Upon simultaneous Priority given to ISE for calibrator, routine
request for photometry sample and control sample
and ISE
Printout Upon independent request Data printed out immediately after
for ISE calculation
5-1
5.2 Outline of ISE Unit
Sample disk
Sample/rinse solution
Sample diluent
Internal Diluent
standard
Reagent disk
Sipper syringe
5-2
(1) Movement of Reaction Disk (Model 902)
Cell rinse
Sample/rinse solution
The reaction sequence consists of processes carried out while the reaction disk provided
with 48 cells turns one revolution on 48 cycles.
The processes are classified into operation process and measurement process.
Process Work Photometry ISE Cell Position
Operation Sample pipetting • • 1
Reagent pipetting/stirring • • 18
(R1, R2, R3)
Electrode aspiration • 7
Cell rinse • • 38 to 47
Measurement Cell blank measurement • 26
Reaction monitor photometry • 26
For ISE analysis, two cells are required for each sample (one cell for sample and the
other for internal standard solution).
ISE
No. Operation Sequence Photometry Internal
Sample
Standard
1 S pipetting into cell 1 and R3 pipetting into cell • •
18
2 Stirring of cell 18 at same position •
3 12-step feed and R2 pipetting into cell 6 •
4 Stirring of cell 6 at same position •
5 5-step feed and R1 pipetting into cell 1 • • •
6 Stirring of cell 1 at same position • •
7 Measurement at cell position 26 which passes •
optical axis at intervals of 18 sec and data
acquisition
8 Aspiration of solution at cell position 7 into • •
electrode after 6-cycle turning
9 Rinsing of cell after lapse of 10-minute reaction • • •
time at cell position 38 to 47
10 Feed of cell 48 to S pipetting position • • •
11 Repetition of no. 1 to 10 on 18-second cycle •
5-3
(2) Mechanical Operation when Sample is Short
For details of (A) and (B), refer to the specifications of the sample probe.
(Sample probe alarm occurrence conditions and mechanical operation)
5-4
5.3 Configuration of ISE Hardware
AMP AMP
(G = 1) (G = 20) ADC control Software
AMP AMP
Na electrode (Vn) MPX (G = 1) (G = 1)
K electrode (Vk) Vout ADC DPRAM Driver Application
Cl electrode (Vc)
Ref electrode (Vr)
Reference voltage (V2)
Vos
Reference voltage (V8)
Reference voltage (V0) ADC controller
(1) The ADC controller converts the analog values of the ISE electrodes and reference
voltages (V2/V8/V0) to digital ones upon receiving an instruction from the application via
the driver and stores the digital values into the DPRAM (dual port RAM). The values are
taken into the application via the driver.
(2) The application accesses the driver, but does not control the ADC directly.
(a) When MPX selects Na electrode Vout = -20 × (Vn - Vr) + Vos [mV]
(b) When MPX selects K electrode Vout = -20 × (Vk - Vr) + Vos [mV]
(c) When MPX selects Cl electrode Vout = 20 × (Vcl - Vr) + Vos [mV]
(d) When MPX selects Ref electrode Vout = -Vr + Vos [mV]
(e) When MPX selects reference voltage (V2) Vout = 2,000 ± 5 [mV]
(f) When MPX selects reference voltage (V8) Vout = 8,000 ± 5 [mV]
(g) When MPX selects reference voltage (V0) Vout =Vos [mV]
5-5
5.4 ISE Measurement Sequence
For the dilution bath in ISE measurement, be sure to use two connected reaction cuvettes.
SMP.
Sample pipetting volume: Fixed at 10 µL
Cuvette no. n+2 n+1 n
Cycle no. 11 12 13
Stop no. 1 1 1
DIL
SMP. Diluent pipetting volume : Fixed at 308 µL
Water : Fixed at 182 µL
Cuvette no. n+2 n+1 n Total 490 µL
Cycle no. 11 12 13
Stop no. 3 3 3
DIL
IS SMP. Internal standard
solution pipetting volum : Fixed at 318 µL
Cuvette no. n+2 n+1 n Water : Fixed at 182 µL
Cycle no. 12 13 14 Total 500 µL
Stop no. 3 3 3
5-6
(5) Internal Standard Solution Aspiration
ISE sipper
DIL
SMP. Internal standard solution measurement
∗ Electrode conditioning is carried out when 2 hours or 24 hours have elapsed after ISE
measurement. Two cells are used for the conditioning and internal standard solution is
aspirated.
5-7
5.5 ISE A/D Conversion
10 msec
1 2 5
(a) As shown in the above figure, measure the electromotive force 5 times each for Na,
K and Cl in this order.
10 msec
2V 8V OFF
(c) At the time of initialization and T/M resetting, the RESET command is issued for the
ADC controller.
5-8
5-9
5 - 10
5 - 11
5 - 12
5 - 13
5 - 14
5 - 15
5 - 16
5 - 17
5 - 18
5 - 19
5 - 20
5 - 21
5 - 22
5 - 23