Professional Documents
Culture Documents
Toe 8951 - Toe 8952
Toe 8951 - Toe 8952
8950E-Manual-Rev10.doc
Software V2.61-2.60
SELECT ENTER
LINE VOLTAGE CURRENT MENU SET / LOCAL
TRACK ERROR
REMOTE TRACK SENSE EXT OVP OT ERROR LOCK SHIF T
OUTPUT
CV CC OFF/ON CV CC
CURSOR CURSOR CP CP
POWER
ON
OFF NEXT MODIFY
1. Introduction
The 400 W power supplies from the TOE 8950 series are representatives of a new genera-
tion of DC power supplies. The power supplies are available as single-output or dual-output
devices with two electrically isolated outputs. Resulting from proven microprocessor control
and state-of-the-art circuitry, the power supplies exhibit exceptional functionalities with excel-
lent analog output data.
Stabilized power supplies are required equally in laboratories, production departments,
schools and workshops. It is expected that they deliver exactly adjustable and reproducible
values for voltage and current at all times. The power supplies from the TOE 8950 series are
rugged, good value, and appropriate to practical requirements. They can be manually oper-
ated, or also remote-controlled via an RS-232 computer interface, analog interface or (op-
tionally) USB or GPIB interface. As a result of their simple yet convenient user interface, they
are equally suitable for laboratory use or in systems. All power supplies can be used as
constant voltage and constant current sources.
1.1 Features
1.1.1 General power supply design
The power supplies of the TOE 8950 series deliver a total output power of 400 W, and are
available in many different versions (→→ 1.2 Summary of power supplies). As a result of their
state-of-the-art circuitry, these power supplies operate with a high efficiency of > 80 % at full
load, and feature a low residual ripple with voltage and current regulation.
The power supplies are also resistant to continuous loads and short-circuits. Input and out-
put peaks or transient reactions do not occur. The floating main outputs are provided with
safe electrical isolation from the mains input and the computer interfaces. The safety sockets
of the main outputs are located on the front panel for laboratory operation. A screw terminal
block is present at the rear with main outputs parallel to those on the front panel as well as
sense inputs.
The output voltage and output current are set manually using wear-free knobs. These knobs
deliver pulses to the intelligent input unit which in turn calculates the corresponding voltage
and current values. This type of input eliminates the known problem of wear-dependence
when using potentiometers. The measured values and setpoints for the output variables, as
well as the menu items, are output on a two-line LCD with 20 characters per line.
A further significant feature of the TOE 8950 series is the high-resolution measurement of
the voltage and current output values. In remote control mode, the measured values can be
read by the controller at a rate of up to 20 measurements per second, meaning that it is usu-
ally unnecessary to use additional measuring instruments.
Depending on the set values for voltage and current, and also on the load conditions, the
power supplies can be used either as a voltage or current source or with a previously set
power limit. Special LEDs indicate the actual source mode of the power supply: constant
voltage (CV), constant current (CC) or constant power (CP) mode.
Fast, easy calibration of all output parameters is possible from the outside within a few min-
utes without having to adjust any trimmers or open up the unit. This "electronic self-calibra-
tion" can be carried out using the manual control elements and guided by menus, or in re-
mote control mode within a test system.
All models are equipped with a powerfail-proof memory for 100 complete power supply set-
tings. The setting valid when a power supply is switched off is also saved, and can be se-
lected if desired when switching on again.
A further convenient feature is the direct switching on/off of the outputs, which permits im-
mediate reduction of the voltage and current values to zero in the off state, or the setting to
programmed values in the on state. Switching of the outputs can be carried out either using
a key operation or a remote control command. By using an active downward programming
circuit, a short reduction time for the output voltage is guaranteed even independent of the
load. The automatically monitored sense mode can be selected for regulation of the voltage
drop on the load feeders. With the dual-output power supplies, it is possible to activate
tracking mode to enable advantageous single-key operation of the voltage setpoints for both
outputs.
The power supplies are equipped with various protective functions such as overvoltage pro-
tection (OVP), overtemperature protection and overload protection. In addition, the power
supplies feature convenient menu operation with which numerous further functions can be
set.
1.1.2 Autoranging
The power supplies of the TOE 8950 series have an autoranging function. They can output
their rated power over a large, continuous range of voltage/current combinations.
For example, the TOE 8951-40 model – with maximum values of 40 V and 20 A – permits a
current of 20 A in the voltage range from 0 to 20 V at its rated power of 400 W.
In the voltage range from 20 to 40 V, the available current is 400 W divided by the actual
voltage, e.g.:
With 30 V output voltage: 400 W : 30 V = 13.33 A
With 40 V output voltage: 400 W : 40 V = 10 A.
The autoranging power supplies from the TOE 8950 series therefore have a significantly
larger working range than standard power supplies with the same power.
40V
A ... Power range of a 400 W standard power
B
supply with 30 V output voltage
B ... Additional power range with the autoranging
C
30V power supplies TOE 8951 and TOE 8952
C ... 400 W working point with standard power
supply (30 V x 13.33 A = 400 W)
20V
A B
Options
- System remote control additionally with USB or GPIB interface
- Arbitrary function (sequence control with up to 1000 steps)
- Interlock control (external enabling of main output)
- Inhibit control (external off/on switching of main output)
2. Manual operation
2.1 Startup
Warning Safe operation of these units depends on them being started up by qualified personnel
with observation of the warnings in this Instruction Manual.
The general directives for installation and safety (e.g. DIN/EN and VDE) must be observed
in particular. The failure to observe these directives could result in death, severe personal
injury or substantial damage to property.
These devices comply with protection class I (protective earth connection) according to
DIN EN61010-1. Before switching on, make sure that the operating voltage set on the de-
vice agrees with the local mains voltage (→ rating plate 230 V/115 V ±10 %, 48 - 65 Hz).
Except when used in rooms with particular protective measures, the mains plug must only
be inserted into a socket with PE contact. The protective earth effect must not be can-
celled by using an extension cord without PE conductor. The mains plug must be inserted
into the socket before the device is switched on and before the measuring and control cir-
cuits are connected. Any interruption in the PE conductor inside or outside the device, or
loosening of the PE conductor connection, could make the device dangerous. An intentio-
nal interruption in the PE conductor connection is not permissible. The local directives con-
cerning grounding must be observed.
Make sure that only fuses of the specified type and rated current are used as replace-
ments (→ 10. Technical specifications). The use of repaired fuses or the short-circuiting of
the fuse holder is not permissible.
If it can be assumed that operation free of danger is no longer possible, the devices must
no longer be used, and must be protected against unintentional use.
Interface Interface
RS-232: 9600 alternativ depending on interface option "GPIB: Dev04",
or "USB: ready" resp. "USB: suspend"
Type display with dual-output power supply (same voltages for parts I and II)
Type display with dual-output power supply (different voltages for parts I and II)
SELECT Complete display; the display remains switched on as long as the key is pressed.
OUTPUT Initialization with standard values; confirm using ENTER within approx. 5 s.
SHIFT OUTPUT Initialization with factory settings; confirm using ENTER within approx. 5 s.
SHIFT ENTER Delete calibration security code; confirm using ENTER within approx. 5 s.
SHIFT VOLTAGE Call type display menu; the menu can be left again using MENU or SHIFT RETURN .
With certain key combinations it is additionally necessary to confirm the desired power-on
mode by pressing the ENTER key within approx. 5 s; e.g. the following display appears when
switching on with the key combination SHIFT ENTER (delete calibration security code):
Clear Cal Code Special power-on mode must be confirmed using ENTER
The following display appears following confirmation using the ENTER key:
If the special power-on mode is not confirmed within approx. 5 s, the power supply starts up
in the normal manner without consideration of the action associated with the special power-
on mode.
The power supplies of the TOE 8950 series are set to the standard settings in the following
cases:
- When switching on using the OUTPUT key (→ 2.1.4)
- When switching on with power-on mode in the setting "RST" (→ 2.1.5)
- When executing the command "*RST" in the menu or in remote control mode.
The power supplies of the TOE 8950 series are set to the factory settings in the following
case:
- When switching on using the SHIFT OUTPUT key (→ 2.1.4).
[1] POWER Power button for switching the mains supply on/off. The green POWER LED lights up
when the power supply is switched on.
[2] I II
2x 20-digit LCD for display of the measured values or setpoints for voltage, current
and power.
In menu mode, the LCD is used to display the individual menu items.
SELECT ENTER
LINE VOLTAGE CURRENT MENU SET/ LOCAL
TRACK ERROR
[3] SHIFT This key is used to activate the second function of a key (identified by red inscription
underneath key), and must be pressed prior to calling the second function. The SHIFT
LED underneath the display lights up when the shift function has been activated.
LINE This key is used to select the input line. The line in which the parameters can be
changed using the two knobs is identified by a preceding selection arrow "
".
SHIFT LOCK With this second function of the key, all input functions which can influence the output
are locked. The actual settings can be checked but not changed. This state is indica-
ted by lighting up of the LOCK LED underneath the display.
[5] VOLTAGE This key is used to input voltage variables: setpoint, limit, sense and OVP.
[6] CURRENT This key is used to input current variables: setpoint and limit.
[7] METER This key is used to select measuring mode and to configure the meter display.
SHIFT TRACK With this second function of the key, tracking mode is activated or deactivated for the
dual-output power supplies. The respective state of tracking mode is indicated by the
TRACKING LED underneath the display.
[8] MENU This key is used to switch menu mode on and off.
SHIFT LAST This second function of the key is used to access the last menu item.
[9] OUTPUT This key is used to switch over the main output between the off state (voltage 0 V, cur-
rent 0 A) and the on state (previously set voltage and current values). The respective
state of the main output is indicated by the OUTPUT LED above the output sockets.
SHIFT ERROR This second function of the key is used to scan error messages.
[10] ENTER This key is used to acknowledge error messages or – in menu mode – to access the
next lower menu level or to execute menu functions.
SET/LOCAL In manual mode, this key is used to display the set values of voltage, current and/or
power.
In remote control mode, pressing this key sets the power supply back to the local
state, provided the device is not in the local lockout state.
SHIFT RETURN This second function of the key is used to return to the next higher menu level.
[11] CURSOR
knob keys
[12] The key function integrated in each of the two knobs is used to switch on cursor
mode. When the key is pressed, the current digit position starts to flash without chang-
ing position. By rotating the knob with the key kept pressed, the cursor is moved by
one position at a time in the direction of rotation. When the final position has been
reached on the left or right, the cursor jumps to the first or last position again depend-
ing on the direction of rotation.
[15] Output (single-output power supply) or Output-II (dual-output power supply part II)
Guarded main output sockets on front panel for 4 mm banana plugs; single-output
power supply or part II with dual-output power supply.
GPIB
Option "GPIB remote control"; 24-contact micro-ribbon socket for connection to the
GPIB bus. The pin assignments correspond to IEEE 488.1.
[18] Protective earth The housing is connected to protective earth using this screw.
Caution: This connection must never be removed!
[19] ANALOG I/O 15-contact Sub-D socket with control inputs and outputs for analog voltage and cur-
rent control of the power supplies (pin assignments → 8. Analog control).
[21] Features box The features box shows the voltage version and the installed options of the power
supply.
[22] Main outputs and A 4-contact (single-output power supply) or 8-contact (dual-output power supply)
sense inputs screw terminal block contains additional rear panel outputs parallel to the main outputs
present on rear on the front panel and the sense inputs.
panel
[23] Rating plate The rating plate contains the following data: company name, device range, serial
number, mains supply with line fuse and power consumption.
[24] Appliance plug This 3-contact standard appliance plug is used to connect the power cord. The plug
with line fuse contains a compartment for the device fuse and an additional compartment for a
replacement fuse.
CV mode Constant voltage mode (CV mode) is identified by lighting up of the CV LED, and indicates
that the power supply is operating as a voltage source. The voltage setpoint is present at the
output, and the output current is smaller than the current setpoint.
CC mode Constant current mode (CC mode) is identified by lighting up of the CC LED, and indicates
that the power supply is operating as a current source. A current flows at the output with the
magnitude of the current setpoint, and the output voltage is smaller than the voltage setpoint.
CP mode Using the integral autoranging function, the power supplies of the TOE 8950 series are able
to switch to automatic power control (CP mode) when the set power limit is exceeded. The
output power is then limited to the set power value, and the voltage and current are automa-
tically set corresponding to the connected load. This operating state is identified by simulta-
neous lighting up of the CV and CC LEDs.
Note When connecting two sources in parallel at the dual-output power supply the correct func-
tion of the automatic power control is not guaranteed.
2.5 Meter
2.5.1 Meter with single-output power supply
Design of With the single-output power supply, the measured values of the voltage and current are
meter always displayed in the top line. The measured value of the power can be displayed in the
bottom line. If a measured value exceeds the range limit, the text message "OFL" is output in
the corresponding area of the display instead of the measured value.
In measuring mode, all displayed parameters can be directly adjusted using the two
knobs. The two knobs only act on the line in which the selection arrow "
" is posi-
ioned. The selection arrow can be switched between the two lines using the LINE key.
Input of the individual parameters is described in detail further below.
Adaptation of Certain functions of the meter can be adapted for the single-output power supply. On the one
meter hand, the second line can either display the measured power or remain empty. On the other
hand, the measuring mode can be switched between sampling sampling measurement and
average measurement.
METER By pressing the METER key, the selection for the meter can be switched on, switched fur-
ther, or switched off. The ENTER key can be used to directly quit adaptation of the meter
and to return to the measurement. The parameters are changed using MODIFY.
Adaptation of Certain functions of the meter can be adapted for the dual-output power supply. On the one
meter hand, the second line can be switched separately for parts I and II between display of current
or power. On the other hand, the measuring mode can be switched between sampling meas-
urement and average measurement.
METER By pressing the METER key, the selection for the meter can be switched on, switched furt-
her, or switched off. The parameters are changed using I and II. The ENTER key can
be used to directly quit adaptation of the meter and to return to the measurement.
Setpoint V Voltage:Set
Ç 40.00V Adjust using MODIFY.
Limit V Voltage:Limit
Ç 41.00V Adjust using MODIFY.
Setpoint V Voltage:Set
Ç 20.00V 40.00V Adjust using I and II.
Limit V Voltage:Limit
Ç 41.00V 41.00V Adjust using I and II.
Setpoint C Current:Set
Ç 20.000A Adjust using MODIFY.
Limit C Current:Limit
Ç 20.200A Adjust using MODIFY.
Setpoint C Current:Set
Ç 5.000A 10.000A Adjust using I and II.
Limit C Current:Limit
Ç 10.200A 10.200A Adjust using I and II.
3. Menu operations
Summary of control elements used for menu operations
MENU The power supplies have a menu function for adjusting parameters which are seldom used.
Menu operation is switched on or off by pressing the MENU key. All parameter modifications
carried out in the menu are saved in a non-volatile memory and are also valid when the
power supply is switched on again.
SHIFT LAST The last menu item can be selected using the SHIFT LAST keys.
and The individual menu items are selected using the knob NEXT. Adjustments within the
individual menu items are carried out using the knob MODIFY or with I or II and with
the cursor keys .
ENTER Menu items which have submenus are identified by the ">" character at the end of the line. A
submenu is accessed using the ENTER key. In turn, a submenu can be recognized in that
the "<" character is present at the beginning in the menu heading.
SHIFT RETURN Returning to the next higher menu level or completely leaving the menu from the uppermost
level is carried out using the SHIFT RETURN keys. The main menu and the submenus can
also be left directly by pressing the MENU key.
Note The off/on state of the main output does not belong to the parameters saved for a device
setting.
Selection of memory location 0 to 99 using MODIFY. Pressing the ENTER key for the first
time prepares the memory recall operation with the following display:
< Menu:Memory:Recall
*ENTER for Recall Trigger memory recall operation using ENTER .
Pressing the ENTER key for the second time triggers the memory recall operation. With the
memory operation carried out correctly, the saved data are immediately imported as the cur-
rent power supply settings, and the OK message is output in the display for approx. 1 s. The
current power supply settings are retained in the event of an error, and an error message is
output in the display.
OK message following recall operation Error message following recall operation
Selection of memory location 0 to 99 using MODIFY. Pressing the ENTER key for the first
time prepares the memory save operation with the following display:
< Menu:Memory:Save
*ENTER for Save Trigger memory save operation using ENTER .
Pressing the ENTER key for the second time triggers the memory save operation and the
current power supply settings are imported into the selected memory location. With the
memory operation carried out correctly, the OK message is output in the display for approx.
1 s. In the event of an error, an error message is output in the display.
< Menu:System
NEXT External Control > Call submenu using ENTER .
< Menu:System
NEXT Voltage Slew Rate > Call submenu using ENTER .
< Menu:System
NEXT Power-On RST Select power-on mode using MODIFY.
< Menu:System
NEXT *RST Command Prepare RST operation using ENTER .
Pressing the ENTER key for the first time prepares RST operation with the following display:
< Menu:System
*ENTER for RST Trigger RST operation using ENTER .
Pressing the ENTER key for the second time triggers RST operation. The standard settings
are immediately imported as the current power supply settings, and a message is displayed
for approx. 1 s to confirm the process.
< Menu:System
!RST done
< Menu:System
NEXT Device IDN > Call submenu using ENTER .
< Menu:System
NEXT T +41ºC Fan 34%
< Menu:Remote
NEXT Interface RS-232 Select interface using MODIFY.
If the option "USB remote control" is installed, and if USB is selected as the remote control
interface, a check is immediately carried out whether a host computer with a suitable driver
is connected. The procedure is output in the display for approx. 1 s.
< Menu:Remote
!Checking USB Checking of USB interface
The end of the check is indicated by an audible signal, and the result is also subsequently
displayed for approx. 1 s.
The "ready" message indicates that the USB interface is connected and ready for operation.
If the "suspend" message is displayed, a connection could not be established to a host
computer.
< Menu:Remote
NEXT RS-232 Setting > Call submenu using ENTER .
In the baud rate menu item, you can select baud rates of 1,200, 2,400, 4,800, 9,600, 19,200,
38,400 and 57,600 bps. The factory setting is 9600.
The following data formats can be selected in the data format menu item:
- 8 Bit 8 data bits, no parity bits (factory setting)
- odd 7 Bit 7 data bits, odd parity
- even 7 Bit 7 data bits, even parity
1 or 2 stop bits can be selected in the stop bits menu item. The factory setting is 1 stop bit.
< Menu:Remote
NEXT GPIB Setting > Call submenu using ENTER .
The GPIB device address can be set between 0 and 30 in this menu item, where the
address 0 is normally reserved for the system controller. The factory setting for the device
address is 4.
< Menu:Remote
NEXT Language COMP Select "Comp" or "SCPI" language using MODIFY.
4. Remote control
The power supplies of the TOE 8950 series are fitted as standard with a serial RS-232 inter-
face. The power supplies can also be optionally fitted with a serial USB interface or a parallel
GPIB interface according to IEEE 488.1. Almost all functions of the power supplies can be
remote-controlled by each of the interfaces.
Note Only one interface can be active at a time. You select the desired interface manually in the
menu (→ 3.6.1). The selected interface is saved in the non-volatile memory, and is not
changed when the device is switched off. The RS-232 interface is the factory setting.
The command structure is designed for both interfaces according to IEEE 488.2. The instruc-
tion set consists of three groups:
1. Standard commands according to IEEE 488.2
2. SCPI commands (Standard Commands for Programmable Instruments)
3. Short commands compatible with older Toellner power supplies.
It is possible to switch between the SCPI commands and the Toellner short commands using
the language selector (→ 3.6.4). The standard commands according to IEEE 488.2 are al-
ways valid, irrespective of the language setting. In the command overview (→ 4.6), all com-
mands are summarized in tabular form. Each individual command is discussed in detail in
the command explanations (→ 4.7).
REMOTE LED The REMOTE LED lights up when the power supply is in remote control mode. In this state,
the power supply is locked for manual operation. The power supply is set to the local state
again by pressing the SET/LOCAL key. The REMOTE LED goes out, and the power supply
can be operated manually again. The SET/LOCAL key has no function if the power supply is
set to the locked remote control mode "Local Lockout". In this case a short warning tone is
generated when the SET/LOCAL key is pressed, and the error message "550, Remote with
Local Lockout" is output on the display.
GPIB setting The bus address can be conveniently entered in the system submenu "GPIB Setting". It is
(→ 3.6.3) saved in the non-volatile memory, and is not changed when the power supply is switched off.
The factory setting for the address is 4.
Devices in a GPIB network are connected together by cables standardized according to
IEEE 488.1 and equipped at both ends with 24-contact connectors. The cable length
between any two devices must not be greater than 2 m, and the total length not greater than
20 m. When operating the power supplies on the GPIB bus in an environment with critical
EMC conditions, the cable should be kept as short as possible and double-shielded.
The pin assignments of the GPIB interface are in accordance with IEEE 488.1, and need not
be explained here.
When operating via the RS-232 interface, the power supply starts to send the reply string as
soon as send mode has been enabled by the computer. Enabling can be carried out by the
software (Xon/Xoff protocol) or by the hardware (RTS/CTS handshake) (setting → 3.6.2
System submenu "RS-232 Setting").
When operating via the USB interface, the reply string is ready in the send buffer of the USB
interface as soon as it has been prepared. The status byte can be polled analogous to the
GPIB interface. The reply string can be read in if the MAV bit is set in the status byte
register.
Several It is possible to transmit several commands in one data string. The individual commands are
commands separated from one another by a semicolon ";". It is permissible to mix normal setting com-
in one string mands and queries. The commands are processed in the sequence in which they are re-
ceived. If a command string contains several queries, the various replies are also separated
by a semicolon ";" in the reply string.
Output queue The output queue saves all reply messages generated as a result of queries in a command
string. The output queue has a FIFO character, i.e. the first requested reply message is also
saved first etc. The messages are separated from one another by a semicolon ";". The MAV
bit is set in the status byte register as soon as information is present in the output queue
(→ 4.7.7 Status register model). The MAV bit is reset as soon as the output queue has been
completely read by the system controller.
RS-232 and When receiving data via the RS-232 or the USB interface, the power supplies are able to
USB delimiter process the following delimiters:
- NL Next Line = Line Feed (ASCII 0x0A)
- CR & NL Carriage Return (ASCII 0x0D) and Next Line = Line Feed (ASCII 0x0A)
When sending data, the power supplies use the delimiter CR & NL for operation via the
RS-232 resp. the USB interface.
GPIB delimiter When receiving data via the GPIB interface, the power supplies are able to automatically
process the following delimiter combinations:
- NL Next Line = Line Feed (ASCII 0x0A)
- NL & EOI Next Line and EOI line active
- DAB & EOI Last character with EOI line active
The delimiter DAB & EOI is not in compliance with IEEE 488.2 which specifies that each
command string must be terminated by the delimiter NL.
When sending data, the power supplies use the delimiter NL & EOI for operation via the GPIB
interface.
Device-specific The maximum length of a command must not exceed 509 characters. The maximum length
limitations of a reply string which can be processed when transmitting several queries in one string is
510 characters. Corresponding errors are generated if the input or output buffer overflows
(→ 6.3 Error list).
The first command in a string consisting of several commands is always based automatically
on the highest level of the tree, the so-called "Root" level. The following command is then
based on the level of the previous command. In an appropriate case, the command string of
the subsequent command can be based abbreviated on the same level as its predecessor.
MEAS:VOLT?;CURR?
In this example, the measured values of voltage and current are requested without having to
enter the keyword MEAS again for the current measurement. Commands from different sub-
systems in one string are linked together by a preceding colon ":". The colon preceding a
command requests the command interpreter (parser) to return to the "Root" level.
MEAS:VOLT?;:SYST:ERR?
Without the "Root" descriptor ":" preceding the keyword SYST, the power supply would pro-
duce an error.
However, the "Root" descriptor ":" preceding the first command in a string is redundant since
the parser always switches back to the "Root" level at the end of a string. The two following
commands have an identical effect.
OUTP:STAT ON
:OUTP:STAT ON
The standard commands according to IEEE 488.2, identified by a"*" prior to the command
code, can be used without problem together with SCPI commands. Use of the standard com-
mands does not influence the SCPI command level at all. The standard commands can be
inserted in the SCPI command string as desired.
MEAS:VOLT?;*RCL 5;CURR?
Commands without a Toellner short form are identical to the SCPI commands and are there-
fore subject to the SCPI conventions. When using these SCPI commands together with the
Toellner short commands, note that each short command switches back to the highest
"Root" level again. All subsequent SCPI commands in the same string must therefore be
based on the "Root" level.
SYST:LANG?;VERS?;V 17.3
SYST:LANG?;V 17.3;:SYST:VERS?
In the first example, the VERS? command can be based on the previous level. In the second
example, however, :SYST:VERS? must commence again at the highest command level since
the V 17.3 command switches back to the "Root" level.
*IDN? This query scans the device identification of the power supply and writes it as a text string
(with quotation marks) into the output queue. The text string consists of four fields separated
by commas. The first field contains the name of the manufacturer, the second field specifies
the device type with voltage version, the third field contains the serial number of the device,
and the fourth field contains the release version of the software.
The text string has the following format: "TOELLNER,TOE8951-80,20293,2.50-2.50"
*IST? This query scans the individual status of the status byte register (STB) and writes it as a
single-digit decimal number 0 or 1 into the output queue. The value 1 is written when agree-
ment is established when ANDing the status byte register and the mask set by the *PRE
command.
*LRN? This query scans the actual status of the power supply and writes it as a text string into the
output queue. The string contains all commands required for a complete power supply set-
ting.
Output format for Toellner short commands:
TOE 8951: "V nnn.nn;C nn.nnn;P nnnn.n;OVP nnnn.n;PV x;VLIM nnn.nn;CLIM nn.nnn;S x;
EXT x;VSLEW x;AVM x;EX x"
TOE 8952: "SEL 1;V nnn.nn;C nn.nnn;P nnnn.n;OVP nnnn.n;PV x;VLIM nnn.nn;CLIM nn.nnn;
S x;EXT x;VSLEW x;AVM x;SEL 2;V nnn.nn;C nn.nnn;P nnnn.n; OVP nnnn.n;PV x;
VLIM nnn.nn;CLIM nn.nnn;S x;EXT x; VSLEW x;AVM x;TRA x;EX x"
Output format for SCPI commands:
TOE 8951: ":VOLT nnn.nn;:CURR nn.nnn;:POW nnnn.n;:VOLT:PROT:LEV nnnn.n;STAT x;
:VOLT:LIM nnn.nn;:CURR:LIM nn.nnn;:SENS:VOLT:DET INT;:EXT x;:VOLT:SLEW
:STAT x;:SENS:AVER x;:OUTP x"
TOE 8952: ":INST OUT1;:VOLT nnn.nn;:CURR nn.nnn;:POW nnnn.n;:VOLT:PROT:LEV nnnn.n;
STAT x;:VOLT:LIM nnn.nn;:CURR:LIM nn.nnn;:SENS:VOLT:DET INT;:EXT x;
:VOLT:SLEW:STAT x;:SENS:AVER x;:INST OUT2;:VOLT nnn.nn;:CURR nn.nnn;
:POW nnnn.n;:VOLT:PROT:LEV nnnn.n;STAT x;:VOLT:LIM nnn.nn;:CURR
:LIM nn.nnn;:SENS:VOLT:DET INT;:EXT x;:VOLT:SLEW:STAT x;:SENS:AVER x;
:OUTP:TRAC x;:OUTP x"
*OPC This command requests setting of the OPC bit (Operation Complete) in the standard event
status register when all previous commands have been processed.
*OPC? This query requests writing of the decimal number 1 into the output queue when all previous
commands have been processed.
*PRE <NR1> This command sets the enable mask of the parallel poll register in decimal form. The *IST?
scanning is prepared by ANDing of the enable mask with the status byte register.
*PRE? This query scans the actual setting of the enable mask of the parallel poll register and writes
it as a five-digit decimal number into the output queue.
*PSC 0|1 This command defines the power-on status of the enable registers SRE, ESE and PRE.
*PSC 0 results in the enable registers being retained during power-on. *PSC 1 results in the
enable registers being deleted during power-on.
*PSC? This query scans the PSC configuration of the enable registers SRE, ESE and PRE. A deci-
mal number 0 or 1 is written into the output queue (→ *PSC 0|1).
*RCL <NR1> This command results in calling of a complete power supply setting from one of the memory
locations 0 to 99 (→ 3.2.1 Memory operation "Recall"). Memory location 0 contains the stan-
dard values of the power supply (→ 2.1.6) in accordance with the IEEE 488.2 definition. An
error message is generated if an unoccupied memory location is called.
*RST This command sets the power supply to a defined initial status. All masks, the MAV bit in the
status byte register, the output queue, and the errors in the error queue are retained
(→ 2.1.6 Standard settings and factory settings).
*SAV <NR1> This command results in saving of a complete power supply setting in one of the memory lo-
cations 1 to 99 (→ 3.2.2 Memory operation "Save"). Memory location 0 contains the stand-
ard values (→ 2.1.6) and cannot be overwritten.
*SRE <NR1> This command sets the service request enable mask for the status byte register in decimal
form.
*SRE? This query scans the service request enable mask and writes it as a three-digit decimal num-
ber into the output queue.
*STB? This query scans the status of the status byte register and writes it as a three-digit decimal
number into the output queue.
Note In contrast to the read procedure with a serial poll, the status byte register is not destroyed
when read.
*TST? This query carries out a self-test of the power supply. This test corresponds to the automatic
self-test of the power supply when switching on. A 0 is written into the output queue if the
test is without fault. In the event of a fault, a 1 is written into the output queue, and the error
message "−330, Self-test failed" as well as further specific error messages are written into the
error queue.
*WAI This command requests the power supply to wait for the termination of all current operations.
VOLTage <NRf>|MINimum|MAXimum
This command programs the setpoint for internal voltage control. The MIN and MAX para-
meters can be used to set the minimum and maximum values.
VOLTage? [MINimum|MAXimum]
This query scans the actual setting of the voltage setpoint. Scanning with MIN and MAX re-
turns the smallest and largest programmable values.
VOLTage:LIMit <NRf>|MINimum|MAXimum
This command programs the voltage limit for the voltage setpoint. The MIN and MAX para-
meters can be used to set the minimum and maximum values.
VOLTage:LIMit? [MINimum|MAXimum]
This query scans the actual setting of the voltage limit. Scanning with MIN and MAX returns
the smallest and largest programmable values.
VOLTage:PROTection <NRf>|MINimum|MAXimum
This command programs the OVP value of the overvoltage protection unit. The range of
values for the OVP voltage extends from 3 V up to approx. 10 % above the maximum output
voltage. The MIN and MAX parameters can be used to set the minimum and maximum
values.
VOLTage:PROTection? [MINimum|MAXimum]
This query scans the actual setting of the OVP value. Scanning with MIN and MAX returns
the smallest and largest programmable values.
VOLTage:PROTection:STATe 0|1|OFF|ON
This command switches the overvoltage protection unit (OVP) off and on.
VOLTage:PROTection:STATe?
This query scans the status of the overvoltage protection unit (OVP); a value of 0 or 1 is re-
turned.
VOLTage:SLEW:STATe 0|1|SLOW|FAST
This command sets the voltage slew rate with small current setpoints.
Note With the setting FAST, overshoots may occur when increasing the voltage, and thus a
delayed transient response of the power control.
VOLTage:SLEW:STATe?
This query scans the status of the voltage slew rate; a value of 0 (slow) or 1 (fast) is re-
turned.
OUTPut:TRACk 0|1|OFF|ON
This command switches tracking mode off and on with the dual-output power supply.
OUTPut:TRACk? With the dual-output power supply, this query scans the status of tracking mode; a value of 0
or 1 is returned.
MEASure:CURRent?
This query triggers a measurement of the output current, and writes it as a five-digit decimal
number with decimal point into the output queue at the end of the measurement. If the meas-
ured value exceeds the range limit, the value "99999." is output in the data string.
MEASure:POWer?
This query triggers a measurement of the output power, and writes it as a five-digit decimal
number with decimal point into the output queue at the end of the measurement. If the meas-
ured value exceeds the range limit, the value "99999." is output in the data string.
MEASure:VOLTage?
This query triggers a measurement of the output voltage, and writes it as a five-digit decimal
number with decimal point into the output queue at the end of the measurement. If the meas-
ured value exceeds the range limit, the value "99999." is output in the data string.
SENSe:AVERage 0|1|OFF|ON
This command switches over the measuring mode between sampling measurement with
0|OFF and average measurement with 1|ON.
SENSe:AVERage?
This query scans the measuring mode setting. A value of 0 is returned for sampling meas-
urement and 1 for average measurement.
Note It is recommendable to transmit this command in a separate string because switching over
to remote mode means that all commands following in the same string are ignored and
may lead to error messages.
STATus:QUEStionable:INSTrument:ISUMmary1
PTRansition
CONDition NTRansition EVENt ENABle
CV 0
CC 1
not used
CP 3
not used
not used OR
not used
not used
not used STATus:QUEStionable:INSTrument
OVP 9
not used EVENt ENABle
not used not used
not used INST1 1
not used INST2 2
not used not used
not used not used
not used OR
not used
STATus:QUEStionable:INSTrument:ISUMmary2 not used
PTRansition
CONDition NTRansition EVENt ENABle not used
0 not used
CV
1 not used
CC Binary Weight
not used not used
not used
CP 3
not used
20 = 1
not used
not used
21 = 2
not used OR not used
22 = 4
not used 23 = 8
not used 24 = 16
not used 25 = 32
OVP 9 26 = 64
not used 27 = 128
not used 28 = 256
not used 29 = 512
not used 2 10= 1024
not used 2 11= 2048
not used 2 12= 4096
2 13= 8192
2 14= 16384
STATus:QUEStionable Only Dual-Output Power Supplies 2 15= 32768
PTRansition
CONDition NTRansition EVENt ENABle
CV* 0
CC* 1
not used
CP* 3 Output Queue
OT 4
not used OR
not used
not used
not used DAB
OVP* 9 Queue not empty DAB
not used DAB
not used DAB
not used
not used ISUM 13 Parallel Poll Enable
*PRE;* PRE?
not used
not used not used
not used
* Only Single-Output Power Supplies not used
QUES 3
MAV 4
Standard Event Register
Event Enable
ESB 5 OR
* ESR? *ESE;*ESE? MSS 6 IST-Query
OPER 7 * IST?
OPC 0
not used
not used OR not used
QYE 2
3 not used
DDE
4 not used
EXE
5 not used
CME
not used not used
7 not used
PON
not used
The *SRE command sets the service request enable mask of the status byte register. This
mask determines which status bits of the status byte register are to result in setting of the
MSS bit (Master Status Summary) and the RQS bit (Request for Service). When the RQS bit
is set (bit 6), the SRQ line is activated simultaneously. The status byte register can be read
in two different manners:
1. The RQS bit is present in bit 6 when reading the status byte register with the serial poll.
The RQS bit is deleted by the read process, and the SRQ line is thus also switched to
inactive again.
2. The MSS bit is present in bit 6 when reading with the *STB? command. The MSS bit is
not deleted by the read process, but only when the correspondingly masked status bits
which have led to setting of the MSS bit become inactive.
The *PRE command sets the enable mask of the parallel poll register and prepares the *IST?
scan. A 1 is written into the output queue by the *IST? scan if agreement is established du-
ring the bit-serial ANDing of the status byte register and the mask set by the *PRE command.
A 0 is otherwise written.
The service request function cannot be used via the serial interface. However, a comparable
result can be achieved by using the *PRE command in association with the *IST? scan.
The *ESE command sets the enable mask of the standard event status register. This mask
determines which event bits of the standard event status register are to result in setting of
the ESB bits in the status byte register. The following table shows the register configuration
and the meaning of the individual event bits.
The operation and questionable status register model according to the SCPI standard con-
sists of the following registers:
Condition The condition registers indicate the actual status of the power supply. The queries for the
registers condition registers, such as e.g. STAT:OPER:COND? or STAT:QUES:COND?, can be used to
scan the respective status of the operation or questionable condition register. The condition
registers are read-only!
Event The event registers save all events (changes in status) occurring in the condition registers,
registers provided the corresponding bits of the associated edge registers NTR and PTR are enabled.
The queries for the event registers, such as e.g. STAT:OPER[:EVEN]? or STAT:QUES[:EVEN]?,
can be used to scan the respective status of the operation or questionable event register.
The event registers are read-only, and are deleted by the read process! If the corresponding
bits are enabled in the event enable register, the events are displayed as an OPER or QUES
common bit in the status byte register.
Enable The event enable registers can mask the event bits of the corresponding event registers.
registers Setting of a masked bit in the event register results in setting of the OPER or QUES common
bit in the status byte register. The commands for the enable registers, such as e.g.
STAT:OPER:ENAB or STAT:QUES:ENAB, can be used to write and also read the operation or
questionable enable registers.
NTR/PTR The edge registers NTR/PTR can switch through the change of a status bit in the condition
registers register to the event register either with the falling edge (NTR) or with the rising edge (PTR).
The commands for the edge registers, such as e.g. STAT:OPER:NTR or STAT:QUES:PTR,
can be used to write and also read the operation or questionable edge registers.
STATus:OPERation:NTRansition <NRf>
This command sets the negative transition mask (NTR) of the operation condition register in
decimal form. The range of values is from 0 to 65535. If a bit is set in the NTR, and if the cor-
responding bit in the condition register changes from 1 to 0 (negative transition), the corre-
sponding bit is set in the event register.
STATus:OPERation:NTRansition?
This query scans the actual setting of the NTR mask of the operation condition register and
writes it as a five-digit decimal number into the output queue.
STATus:OPERation:PTRansition <NRf>
This command sets the positive transition mask (PTR) of the operation condition register in
decimal form. The range of values is from 0 to 65535. If a bit is set in the PTR, and if the cor-
responding bit in the condition register changes from 0 to 1 (positive transition), the corre-
sponding bit is set in the event register.
STATus:OPERation:PTRansition?
This query scans the actual setting of the PTR mask of the operation condition register and
writes it as a five-digit decimal number into the output queue.
STATus:PRESet This command sets all relevant bits of the positive transition registers in the operation and
questionable status system, and resets all negative transition registers and all enable regi-
sters.
STATus:QUEStionable:CONDition?
This query scans the status of the questionable condition register and writes it as a five-digit
decimal number into the output queue. The condition register is read-only. (Questionable sta-
tus bits → 4.7.7 Status register model)
STATus:QUEStionable:ENABle <NRf>
This command sets the enable mask of the questionable event register in decimal form. The
range of values is from 0 to 65535. This mask can be used to mask the bits which are to re-
sult in setting of the QUES bit in the status byte register.
STATus:QUEStionable:ENABle?
This query scans the actual setting of the enable mask of the questionable event register
and writes it as a five-digit decimal number into the output queue.
STATus:QUEStionable?
This query scans the status of the questionable event register and writes it as a five-digit
decimal number into the output queue. This event register saves all events which have
occurred in the condition register and have been enabled by the corresponding bits of the
associated edge setting NTR or PTR.
Note The questionable event register is read-only, and is deleted by the read process!
STATus:QUEStionable:NTRansition <NRf>
This command sets the negative transition mask (NTR) of the questionable condition register
in decimal form. The range of values is from 0 to 65535. If a bit is set in the NTR, and if the
corresponding bit in the condition register changes from 1 to 0 (negative transition), the cor-
responding bit is set in the event register.
STATus:QUEStionable:NTRansition?
This query scans the actual setting of the NTR mask of the questionable condition register
and writes it as a five-digit decimal number into the output queue.
STATus:QUEStionable:PTRansition <NRf>
This command sets the positive transition mask (PTR) of the questionable condition register
in decimal form. The range of values is from 0 to 65535. If a bit is set in the PTR, and if the
corresponding bit in the condition register changes from 0 to 1 (positive transition), the corre-
sponding bit is set in the event register.
STATus:QUEStionable:PTRansition?
This query scans the actual setting of the PTR mask of the questionable condition register
and writes it as a five-digit decimal number into the output queue.
STATus:QUEStionable:INSTrument:ISUMmary<n>:CONDition?
This query scans the status of the questionable isummary condition register of part I (<n>=1)
or II (<n>=2) and write it as a five-digit decimal number into the output queue. The condition
register is read-only. (Questionable status bits → 4.7.7 Status register model)
STATus:QUEStionable:INSTrument:ISUMmary<n>:ENABle <NRf>
This command sets the enable mask of the questionable isummary event register of part I
(<n>=1) or II (<n>=2) in decimal form. The range of values is from 0 to 65535. This mask can
be used to mask the bits which are to result in setting of the INST bit in the questionable in-
strument register.
STATus:QUEStionable:INSTrument:ISUMmary<n>:ENABle?
This query scans the actual setting of the enable mask of the questionable isummary event
register of part I (<n>=1) or II (<n>=2) and write it as a five-digit decimal number into the out-
put queue.
STATus:QUEStionable:INSTrument:ISUMmary<n>?
This query scans the status of the questionable isummary event register of part I (<n>=1) or
II (<n>=2) and write it as a five-digit decimal number into the output queue. This register
save all events which have occurred in the condition register and have been enabled by the
corresponding bits of the associated edge setting NTR or PTR.
Note The questionable isummary event register is read-only, and is deleted by the read
process!
STATus:QUEStionable:INSTrument:ISUMmary<n>:NTRansition <NRf>
This command sets the negative transition mask (NTR) of the questionable isummary condi-
tion register of part I (<n>=1) or II (<n>=2) in decimal form. The range of values is from 0 to
65535. If a bit is set in the NTR, and if the corresponding bit in the condition register changes
from 1 to 0 (negative transition), the corresponding bit is set in the event register.
STATus:QUEStionable:INSTrument:ISUMmary<n>:NTRansition?
This query scans the actual setting of the NTR mask of the questionable isummary condition
register of part I (<n>=1) or II (<n>=2) and writes it as a five-digit decimal number into the
output queue.
STATus:QUEStionable:INSTrument:ISUMmary<n>:PTRansition <NRf>
This command sets the positive transition mask (PTR) of the questionable isummary condi-
tion register of part I (<n>=1) or II (<n>=2) in decimal form. The range of values is from 0 to
65535. If a bit is set in the PTR, and if the corresponding bit in the condition register changes
from 0 to 1 (positive transition), the corresponding bit is set in the event register.
STATus:QUEStionable:INSTrument:ISUMmary<n>:PTRansition?
This query scans the actual setting of the PTR mask of the questionable isummary condition
register of part I (<n>=1) or II (<n>=2) and writes it as a five-digit decimal number into the
output queue.
5. Calibration
The power supplies of the TOE 8950 series are equipped with electronic self-calibration
which can be carried out by users without having to adjust any trimmers either manually in
the menu item "Calibration" or by using remote control commands.
Calibration To protect the calibrated settings from misuse, users can assign the power supply with a
security code security code for the calibration. The security code can only be deleted again by a special
power-on procedure (→ 2.1.4).
In the event of a faulty calibration, you should always check first whether the power supply
has been correctly connected for the calibration item. If the fault occurs again despite a cor-
rect connection, a device-internal fault is probably present and the power supply must be re-
turned to the factory for repair.
Note The power supply must be in operation for at least 30 minutes before starting the
calibration.
and The individual calibration items are selected using the knob NEXT. Adjustments within the
individual calibration items are carried out using the knob MODIFY and the cursor keys
.
ENTER The ENTER key is used to access the submenus of calibration items, to complete inputs,
and to trigger calibration procedures. Calibration items possessing a submenu are identified
by the ">" character at the end of the line. In turn, a submenu can be recognized in that the
"<" character is present at the beginning in the menu heading.
SHIFT RETURN The input and the calibration item associated with it can be aborted using the SHIFT
RETURN keys, and a return is made to the next higher menu level.
First switch on menu made using the MENU key. As described in the menu operation (→ 3.),
first select the main menu item "Calibration" using NEXT. The following is then output on
the display:
Single-output power supply Dual-output power supply
The OK tick "É" in the second line in front of the calibration menu indicates that the complete
calibration of a source has been terminated successfully and that the calibration values have
been saved. The OK tick "É" is also used in the same manner for the individual calibration
items within the calibration menu, such as e.g. in the calibration item "Cal:Voltage". In this
case, the OK tick "É" shows that the voltage calibration has been carried out.
Single-output power supply Dual-output power supply
< Menu:Cal
Code 0000 Enter security code using MODIFY and .
If a security code is not used, the following display is initially output, and the calibration un-
locked after approx. 2 s with an automatic switch to the calibration menu.
< Menu:Cal
Code ---- Automatic unlocking after approx. 2 s.
Complete input of the security code using ENTER . If the code is entered correctly, the un-
locked display is output for approx. 2 s. An automatic switch is then made to the calibration
menu.
A fault display is output if the entered code is incorrect, and a new code input blocked for ap-
prox. 5 s. A new attempt to enter the correct security code can only be carried out at the end
of this blocking time.
Unlocked display Fault display
< Menu:Cal
NEXT ÉCal:Voltage > Access the voltage calibration using ENTER .
By pressing ENTER the voltage calibration is accessed at the low calibration value, shown
here for a TOE 8951-40.
To calibrate the low calibration value, enter the value measured on the voltmeter using
MODIFY and and confirm using ENTER . Confirmation of the input is output on the
display for approx. 2 s, and calibration of the low value is carried out.
< Menu:Cal:Voltage
Volt:Low entered Calibration procedure active for approx. 2 s.
To calibrate the high calibration value, enter the value measured on the voltmeter using
MODIFY and and confirm using ENTER . Confirmation of the input is output on the dis-
play for approx. 5 s, and calibration of the high value is carried out together with calibration
of the voltage measurement.
< Menu:Cal:Voltage
Volt:High entered Calibration procedure active for approx. 5 s.
At the end of the calibration process the display outputs either the OK message for approx.
2 s or one of several error messages which may provide information on the possible cause
of the fault. A return is then made automatically to the calibration menu as shown here.
OK message Error message
Note Following a successful voltage calibration, it is essential to the recalibrate the OVP
voltage.
< Menu:Cal
NEXT Cal:OVP > Access the OVP calibration using ENTER .
Pressing the ENTER key for the first time prepares the automatic OVP calibration with the
following display:
< Menu:Cal:OVP
*ENTER for Auto Cal Start the automatic OVP calibration using ENTER .
Pressing the ENTER key again starts the automatic OVP calibration, and the display for the
active calibration process is output for approx. 2 s:
< Menu:Cal:OVP
Auto Cal started Automatic OVP calibration procedure active for approx. 2 s.
At the end of the calibration process the display outputs either the OK message for approx.
2 s or one of several error messages which may provide information on the possible cause
of the fault. A return is then made automatically to the calibration menu as shown here.
OK message Error message
< Menu:Cal
NEXT ÉCal:Current > Access the current calibration using ENTER .
By pressing ENTER the current calibration is accessed at the low calibration value, shown
here for a TOE 8951-40.
To calibrate the low calibration value, enter the value measured on the voltmeter divided by
the shunt resistance value using MODIFY and and confirm using ENTER . Confirma-
tion of the input is output on the display for approx. 2 s, and calibration of the low value is
carried out.
< Menu:Cal:Current
Curr:Low entered Calibration procedure active for approx. 2 s.
To calibrate the high calibration value, enter the value measured on the voltmeter divided by
the shunt resistance value using MODIFY and and confirm using ENTER . Confirma-
tion of the input is output on the display for approx. 3 s, and calibration of the high value is
carried out together with calibration of the current measurement.
< Menu:Cal:Current
Curr:High entered Calibration procedure active for approx. 3 s.
At the end of the calibration process the display outputs either the OK message for approx.
2 s or one of several error messages which may provide information on the possible cause
of the fault. A return is then made automatically to the calibration menu as shown here.
OK message Error message
< Menu:Cal
NEXT Cal:Save > Start save process using ENTER .
The save process is started by pressing the ENTER key. If the calibration has not been car-
ried out completely (one or more calibration items are without the OK tick "É"), the following
warning is displayed for approx. 2 s:
< Menu:Cal
!Cal not complete Warning if calibration is incomplete.
If the calibration has been carried out completely, the save process is prepared with the
following display:
< Menu:Cal
*ENTER for Cal Save Trigger save process using ENTER .
Pressing the ENTER key again triggers the save process. If carried out correctly, the OK
message is displayed for approx. 2 s, or the error message if faulty. A return is the automati-
cally made to the main menu or the calibration menu.
OK message Error message
< Menu:Cal
NEXT Set:Code 0000 Enter security code using MODIFY and .
Input of the security code is prepared using ENTER and the following is displayed:
< Menu:Cal
*ENTER for Code Save Save security code using ENTER .
Pressing the ENTER key again saves the security code, and the confirmation is displayed. A
return is the subsequently made automatically to the calibration item "Set:Code".
< Menu:Cal
!New Code saved Saving of new security code is confirmed.
< Menu:Cal
Set:Code 0000
CALibration:CURRent:LEVel MINimum|MAXimum
This command selects the low current calibration item (1/100 CMAX) with MIN and the high
current calibration item (CMAX) with MAX, and switches the power supply output to the on
state.
CALibration:DATA <NRf>
This command transmits the measured values of voltage and current to the power supply. It
is first necessary to select the desired calibration item using the CAL:CURR:LEV or
CAL:VOLT:LEV command.
CALibration:PASSword <NRf>
This command transmits the 4-digit security code for unlocking the calibration.
CALibration:PASSword:NEW <NRf>
This command transmits a new 4-digit security code for the calibration. The calibration must
first be unlocked using the CAL:PASS command and the old security code.
CALibration:PASSword:STATe?
This query scans the locking state of the calibration; a value of 0 (locked) or 1 (unlocked) is
returned.
CALibration:SAVE
This command carries out the powerfail-proof saving of the calibration values following the
complete calibration. Until the new calibration values have been saved using this command,
the calibration process can be aborted at any time using the CAL:STAT OFF command with-
out destroying the old calibration values.
CALibration:STATe 0|1|OFF|ON
This command switches the calibration off or on. Only the calibration commands, status
commands and error scanning commands are permissible when calibration mode is active.
CALibration:STATe?
This query scans the calibration state; a value of 0 or 1 is returned.
CALibration:VOLTage:LEVel MINimum|MAXimum
This command selects the low voltage calibration item (1/100 VMAX) with MIN and the high
voltage calibration item (VMAX) with MAX, and switches the power supply output to the on
state.
CALibration:VOLTage:PROTection
This command accesses the OVP calibration item and automatically carries out the OVP
calibration.
6. Error messages
6.1 General
A short warning tone is generated when an error occurs, and an error message displayed for
approx. 2 s when in manual mode. In addition, the error message is saved in the error queue
and can be read and acknowledged at any time as described further below.
The error queue is organized as a FIFO, i.e. the first error which has occurred is at the first
position, the second error at the second position etc. The ERROR LED underneath the dis-
play lights up as soon as an error is present in the error queue. If more than 20 errors have
occurred, the last error code is replaced by the code "−350, Queue overflow". No further error
codes are then accepted until the error queue has been partially or completely read out and
space has therefore been produced again.
The error queue can be viewed in manual mode using the SHIFT ERROR keys. Example of
a displayed error message:
As described in 6.1, the power-on errors detected when the power supply is switched on can
be viewed in the error queue and also deleted.
Fatal device If a fatal device error within the controller occurs when switching on, the power supply is
errors blocked and cannot be operated. It must be returned to the factory for repair. The ERROR
LED flashes, and the following error message is displayed:
Device errors
501 Thermal overload
The thermal protection circuit has been triggered and has switched the main output to the off
state.
502 Thermal sensor failed
The temperature sensor is delivering incorrect temperature values, or none at all.
503 High voltage at output
An output voltage > 60 V has been detected at the main output while in the off state.
504 Overvoltage detection at output
The OVP circuit has been triggered and has switched the main output to the off state.
510 General overload at output
An impermissible primary current overload has occurred for more than 2 s in the switched-
mode power supply.
511 Power overload at output
The power limit of the main output has been exceeded illegally for more than 4 s.
512 Current overload at output
The maximum permissible current at the main output has been exceeded for more than 10 s.
513 Voltage overload at output
The maximum permissible voltage at the main output has been exceeded for more than 1 s.
520 Sense error at output
With sense mode switched on, a sense error has been detected at the main output.
521 Input buffer overrun
The receive buffer with max. 509 characters for command data in remote control mode is
full. A command string must not exceed this length.
522 Output buffer overrun
The output queue with max. 510 characters for reply data in remote control mode is full. A
single query cannot exceed this limit, but several queries sent in one string can make the
output queue overflow.
550 Remote with Local Lockout
An attempt has been made to switch the power supply to local mode although it is in locked
remote control mode.
551 Manual Operation locked
An attempt has been made to change settings on the power supply although operation of the
device has been locked by SHIFT LOCK .
552 Setting limited
An attempt has been made to set the voltage or current outside the limits.
Power-on errors
600 Device error; system code
A special device error with a three-digit decimal error code for diagnosis were detected
during power-on.
601 Identification failed control processor
The specified control processor was not recognized during power-on.
602 Different program versions control processor
Different software versions were detected for the control processors of the dual-output power
supply during power-on.
603 Main program memory failed
The program memory of the input processor is faulty.
603 Program memory failed control processor
The program memory of the specified control processor is faulty.
604 Main data memory failed
The data memory of the input processor is faulty.
604 Data memory failed control processor
The data memory of the specified control processor is faulty.
7. Programming examples
Adjustment of This section describes two typical application programs for both the TOE 8951 single-output
the interface in power supply and the TOE 8952 dual-output power supply. The following examples are
the menu (→ 3.6) based on the SCPI commands whose language syntax is described in Section 4.5.1. These
examples can be used for the serial RS-232 and USB interfaces as well as for the GPIB
interface. Prerequisites are appropriate transmission software, adjustment of the interface
with all required parameters, and switching of the power supply to remote mode. This is
carried out with the RS-232 and USB by means of the SYSTem:REMote command, and with
the GPIB by activating the remote line.
To improve understanding, commands only effective on part I or part II of the dual-output power supply
have been indented. Several commands can be sent in one string in this case too, e.g.:
INST OUT1;:CURR 7;:VOLT 12
INST OUT2;:CURR 8;:VOLT 14.4;PROT 16;STAT ON
INST OUT1;:MEAS:CURR?;VOLT?
INST OUT2;:MEAS:CURR?;VOLT?
8. Analog control
The power supplies of the TOE 8950 series can be controlled in analog mode, i.e. the set-
points for the voltage and current can be defined independent of one another using external-
ly applied bipolar control voltages of ±5 V. The relationship between control voltage and
output variable is linear. A control voltage of 0 - 5 V always results in an output voltage of
0 - VMAX and an output current of 0 - CMAX. The setpoints generated by the external analog
control voltages are added to the internal setpoints in the case of a positive control voltage,
and subtracted from them in the case of a negative control voltage.
The analog control is switched on and off in the menu item " Menu:Syst:Ext" (→ 3.4.1
System submenu "External Control"). The activated state is indicated by the EXT LED. With
the dual-output power supply, each part can be controlled separately in analog mode. The
control voltage for the parameters to be controlled externally must be connected to the
power supply via the 15-contact Sub-D socket "ANALOG I/O" on the rear panel.
Note Note that the specifications for the residual ripple of voltage and current may become
slightly worse with the external analog control.
The same relationships apply to the analog control current as to the analog control volt-
age.
Analog control in Tracking mode with the dual-output power supply operates exclusively with the internal volt-
tracking mode age settings. The external analog control of all voltage and current parameters is carried out
in the manner described above independent of the tracking mode.
Connection The control inputs VEXTV and VEXTC are non-floating. The respective reference potential is
information connected internally to the negative pole of the associated power output. The inverting
control inputs VEXTV− and VEXTC− must therefore be connected to the analog reference
potential of the associated power output. This can be carried out in two ways:
1. If the control voltage with its reference potential is connected to the negative pole of the
power output on the load side, the inverting control inputs VEXTV− and VEXTC− must be
connected to these nodes.
2. If the control voltage is floating with respect to the output, the inverting control inputs
VEXTV− and VEXTC− must be connected to the associated auxiliary reference potential
AUX-GND.
Warning The auxiliary reference potentials AUX-GND of the control inputs must not be used as
power outputs and connected to loads!
In order to retain electrical isolation of the two sources with the dual-output power supply,
the control voltages of the sources must also be electrically isolated from each other.
Electrical isolation between the control voltages for voltage and current for one source is
unnecessary.
The failure to observe this warning information may result in malfunctions and pos-
sibly in damage to the power supply.
B B
D
C ENTER
ENU SET / LOCAL
ST RETURN
ER OUTPUT
CK ERROR
REMOTE
CC
CP
POWER
A O
O
A
E
B
B B
C ENTER ENTER
ENU SET/ LOCAL ENU SET / LOCAL
ST RETURN ST RETURN
ER OUTPUT ER OUTPUT
CK ERROR CK ERROR
REMOTE REMOTE
CC CC
CP CP
POWER POWER
A O
A O
O
D O
C
A A
5. Remove the inner mounting screw (A) of the housing from the rear panels of the two
power supplies using the Allan key.
6. Screw the two power supplies together by means of the rear connection bracket (E) using
the previously removed mounting screws (A).
ANALOG I / O ANALOG I / O
SENSING
SENSING
SENSING
40V 40V
60V 60V
80V OUTPUT 80V OUTPUT
TOE 8950 TOE 8950
OPTIONS OPTIONS
015 se: T8L 015 se: T8L
se: T4L se: T4L
FUSE FUSE
0VA max. 0VA max.
lner.de lner.de
ermany ermany
A A
E
Remote control Interface Standard with RS-232 interface or optionally with USB or
GPIB interface; electrically isolated from main outputs,
adjustable using menu
Software Command syntax acc. to IEEE 488.2 with command set
switchable between compatible Toellner commands and
standardized SCPI commands
RS-232 interface
Connector 9-pin Sub-D plug
Transmission mode Half-duplex, asynchronous
Transmission rate 1,200 - 57,600 bps adjustable using menu
Setting rate Typically 20 settings/s
Measuring rate Typically 15 measurements/s
Data format 7 data bits without parity or 8 data bits with even or odd parity;
1 or 2 stop bits
Protocol Software protocol with Xon/Xoff mode or
hardware protocol with RTS/CTS handshake
USB interface (option)
USB standard USB 2.0 full speed (max. 12 MBit/s), compatible to USB 1.1
Connector 9-pin USB socket type "B"
Setting rate Typically 25 settings/s
Measuring rate Typically 20 measurements/s
GPIB interface (option)
Connector 24-contact micro-ribbon socket acc. to IEEE 488.1
Setting rate Typically 25 settings/s
Measuring rate Typically 20 measurements/s
Functions AH1, SH1, L4, T6, SR1, PP1, RL1, DC1, DT0, E1, C0
Device address Decimal from 0 - 30, adjustable using menu
SELECT ENTER
V A/W LINE VOLTAGE CURRENT MENU SET/ LOCAL
ERROR
REMOTE SENSE EXT OVP OT ERROR LOCK SHIF T
OUTPUT
CV CC
V CURSOR A/W CURSOR
OFF/ON
CP
POWER
ON
OFF NEXT MODIFY
[1] Power switch with green POWER LED [9] Main output on/off key / error message scan
[2] LCD display [10] Key for confimation / display of set values / switching
[3] Key for second key function to local state / return to previous menu level
[4] Selection key / operation lock key [11] Knob V (NEXT) with integrated cursor key
[5] Voltage menu [12] Knob A/W (MODIFY) with integrated cursor key
[6] Current menu [13] General-purpose LEDs
[7] Selection of measuring mode [15] Main output on front panel
[8] Menu on/off key / last menu item key [16] LEDs referring to output
RS-232
Remote Option
ANALOG I / O
TOE 8951 -
SENSING
SENSING
40V
60V
80V OUTPUT
Series TOE 8950
OPTIONS SN
FUSE
48-65Hz 680VA max.
www.toellner.de
Made in Germany
SELECT ENTER
LINE VOLTAGE CURRENT MENU SET/ LOCAL
TRACK ERROR
REMOTE TRACK SENSE EXT OVP OT ERROR LOCK SHIF T
OUTPUT
CV CC OFF/ON CV CC
CURSOR CURSOR CP CP
POWER
ON
OFF NEXT MODIFY
[1] Power switch with green POWER LED [9] Main outputs on/off key / error message scan
[2] LCD display [10] Key for confimation / display of set values / switching
[3] Key for second key function to local state / return to previous menu level
[4] Selection key / operation lock key [11] Knob part I (NEXT) with integrated cursor key
[5] Voltage menu [12] Knob part II (MODIFY) with integrated cursor key
[6] Current menu [13] General-purpose LEDs
[7] Selection of measuring mode / tracking on/off key [14] Main output on front panel part I
[8] Menu on/off key / last menu item key [15] Main output on front panel part II
[16] LEDs referring to outputs
RS-232
Remote Option
ANALOG I / O
TOE 8952 -
SENSING
SENSING
SENSING
SENSING
40V
60V
80V OUTPUT OUTPUT
Series TOE 8950
OPTIONS SN
FUSE
48-65Hz 680VA max.
www.toellner.de
Made in Germany