Not all Radiofrequency Devices Are Created Equal:
A Thermal Assessment
Erez Dayan, MD*; Spero Theodorou, MD†
O
ver the past 20 years, radiofrequency has emerged 2012 when it received 510K FDA clearance for cutting,
as a leading technology to achieve nonexci- coagulation, and soft tissue ablation.10 This device uses
sional soft tissue contraction.1–3 A variety of deliv- helium gas plasma, fueled by an electrical current to
ery methods of radiofrequency are currently available, treat tissues at high temperatures for short periods of
including monopolar, bipolar, multipolar, fractional, and time. The current delivered is low, resulting in minimal
plasma-driven.1,4–9 Combination technologies (ie, laser, depth of thermal effect and prevention of overheating
vacuum) have also been developed to change the imped- tissue when performing multiple passes. Because plasma-
ance characteristics of tissue and influence thermal deliv- driven radiofrequency treats for short time-intervals, the
ery patterns. surrounding treatment sites remain at relatively cool
Multiple studies have demonstrated that soft tissue temperatures. Also, the volume of un-ionized helium
contraction of 30%–40% can be achieved with radiofre- gas in the treatment space serves as a cooling system to
quency-assisted liposuction; as opposed to approximately avoid prolonged heating. The plasma-driven electrical
10% with mechanical liposuction alone.10 The degree current travels 360 degrees from applicator tip without
of soft tissue contraction after thermal injury has been a focused direction and will preferentially travel through
shown to follow an Arrhenius relationship of tempera- tissues with the least resistance. The benefit of plasma-
ture versus time.10,11 (See figure, Supplemental Digital driven rapid heating and cooling is that there is a rel-
Content 1, which displays the (a) soft tissue contraction: atively low epidermal burn risk. This is supported by
Arrhenius time versus temperature relationship, and (b) surface temperatures rarely exceeding 38°C during treat-
bipolar radiofrequency. http://links.lww.com/PRSGO/ ment.12 However, the limitations of this method of radio-
B911.) frequency delivery are that the low current is unable to
This relationship indicates that a higher temperature penetrate higher impedance tissues, leading to minimal
exposure for a shorter time period provides equivalent depth of effect. Also, as the device passes into proxim-
soft-tissue contraction to a lower temperature exposure ity of previously treated tissue, the energy will follow the
for a longer time period. For example, collagen heated path of least resistance (lower impedance). However, ide-
at 65°C for 120 seconds will provide a significant contrac- ally the structures with high impedance (adipose tissue
tion, equivalent to a temperature of 85°C for approxi- and fibroseptal networks) should be treated for optimal
mately 0.044 seconds.11 soft tissue contraction.
Two radiofrequency delivery methods that differ In contrast, the latest generation bipolar radiofre-
significantly in temperature versus time philosophy are quency with internal/external temperature monitors
plasma-driven radiofrequency (J-Plasma/Renuvion; as well as impedance control was introduced in 2012 by
Apex Medical) and bipolar radiofrequency (Accutite, Invasix (Yokenam, Israel). (See figure, Supplemental
BodyTite, FaceTite; InMode). (See Video 1 [online], Digital Content 1, http://links.lww.com/PRSGO/
which displays the comparison thermal imaging video B911.) The device utilizes a small cannula placed under-
of body and face treatment of monopolar versus bipo- neath the skin to gradually bulk-heat soft tissue between
lar radiofrequency.) Plasma-driven radiofrequency first the two electrodes in a radiant distribution. The con-
became popularized with J-Plasma/Renuvion in early trolled directionality of the radiofrequency energy
(internal electrode to external electrode) focuses ther-
mal energy across tissues with higher impedance (ie, adi-
pose tissue, fibroseptal networks, reticular dermis). This
From the *Avance Plastic Surgery Institute, Reno/Tahoe, Nev.; and volumetric heating is more gradual than plasma-driven
†BodySculpt, New York, N.Y.
Received for publication February 26, 2021; accepted December 2,
2021.
Copyright © 2022 The Authors. Published by Wolters Kluwer Health,
Disclosure: Dr. Erez Dayan receives book royalties from
Inc. on behalf of The American Society of Plastic Surgeons. This
Thieme and Elsevier. He is a consultant at InMode and
is an open-access article distributed under the terms of the Creative
Co-investigator at Allergan, InMode, Galderma, and MTF.
Commons Attribution-Non Commercial-No Derivatives License 4.0
Dr. Spero Theodorou is a consultant at InMode.
(CCBY-NC-ND), where it is permissible to download and share the
work provided it is properly cited. The work cannot be changed in
any way or used commercially without permission from the journal.
Related Digital Media are available in the full-text ver-
Plast Reconstr Surg Glob Open 2022;10:e4077; doi: 10.1097/ sion of the article on www.PRSGlobalOpen.com.