Calibration of vernier calliper (mm)
Name Nitin UID c0887966
Gauge
Block QEMT012-GB- Vernier
No. 2 caliper no. QMET003-V-11
Calibration of Vernier caliper
Venier Theoretical Measured Diff
37.50 37.50 37.50 0.00
75.00 75.00 75.00 0.00
112.50 112.50 112.50 0.00
150.00 150.00 150.00 0.00

Figure 1 Vernier calliper (mm)

The measured dimensions exactly same as the Theoretical measurements, therefore indicating
the vernier calliper used is calibrated aa is within tolerance range.
Calibration of Micrometre (mm)
Name Nitin UID c0887966
Gauge
Block QEMT012-GB- Micrometer QMET032-OM-
No. 11 no. 7
Calibration of Micrometer (MM)
Venier Theoretical Measured Diff
6.25 6.25 6.26 0.01
12.50 12.50 12.51 0.01
18.75 18.75 18.76 0.01
25.00 25.00 25.01 0.01

Differance
0.01

0.01
6.26 12.51 18.76 25.01

Figure 2 Micrometre (mm)

As we can see in the graph there is the consistent variation of +0.01. The micrometre is within
acceptable tolerance range of ±0.01.
Calibration of Vernier Calliper (inch)
Name Nitin UID c0887966
Gauge
Block QEMT012-GB- Vernier
No. 11 caliper no. QMET003-V-1
Calibration of Vernier caliper
Venier Theoretical Measured Diff
1.476 1.476 1.476 0.000
2.953 2.953 2.953 0.000
4.429 4.429 4.429 0.000
5.906 5.906 5.906 0.000

Figure 3 Vernier Calliper (inches)

Similarly in this graph also it can be seen that there is no variation in the measure dimensions
through gauge block. The vernier used is calibrated.
Calibration for Micrometre (inch)
Name Nitin UID c0887966
Gauge
Block QEMT012-GB- Micrometer QMET032-
No. 11 no. OM-10
Calibration of Micrometer (inch)
Venier Theoretical Measured Diff
0.246 0.246 0.246 0.000
0.492 0.492 0.493 0.001
0.738 0.738 0.738 0.001
0.984 0.984 0.985 0.001

Figure 4 Micrometre (inch)

The micrometre inches showed no variation in 0.246” dimension and the variation changed in
later dimensions i.e. 0.001. and least count is ±0.001.