Professional Documents
Culture Documents
Advanced Instruments Osmotech Maintenence
Advanced Instruments Osmotech Maintenence
Chamber cleaning
7. Inspect the solenoid plunger for excessive wear and Fuse replacement
deposits.
Tools required:
• If the plunger shows signs of fouling, clean it with a • Small flat screwdriver
lint-free cloth dampened with isopropyl alcohol or • Static Grounding Wrist Strap
a cleaning solution for protein removal. (Thermo
Scientific Orion Cleaning Solution, part number
900021 is recommended.) WARNING: Hazardous Voltage
deposits
No filings
present
Fuse Holder
Plating is Door (open)
even, bright,
and shiny
5. Insert two replacement fuses into the Fuse Holder. Sampler plunger wire replacement and verification
NOTE: Use replacement fuse P/N 70011: To ensure proper instrument operation, you should
Fuse, 1.0A 250V SB 5 x 20 mm. replace the sampler plunger wire every 100 tests (or
NOTE: Double-check the values marked on the new each time you empty a XT100 test kit).
fuses before you install them. The instrument NOTE: A sampler plunger wire is included with each test
automatically adjusts for voltages between kit.
100 VAC and 240 VAC, but appropriately
rated fuses must be installed. Failure to replace the plunger wire may affect instrument
accuracy and repeatability.
Calibration
gauge and key
Shaft bushing
Shaft
Finger grip
Calibration
set-screw
Body
access
6. Re-install the Fuse Holder into the back of the
instrument.
Tip
Sampler tip
Key