SWITZERLAND
d z ELECTRICAL DATA
Euro plug-in card pin assignment
2 z2 / K1 Measuring circuit 1st measuring probe
d2 / K2 Connection to MTI clamps (K1/2) 24 V-Version
4 z4 / K1 Measuring circuit 2st measuring probe
d4 / K2 Connection to MTI clamps (K1/2)
6 d/z6 Earth
8
Error message programmable in
10 0.1 mA-steps;
0.5 – 3.9 / 20.1 – 22 mA
12 z12 + Analog output 1 MC1: 4-20mA active
d14 - (Burden max. 750 Ohm potential free)
14
Optokoppler
Opto electronic coupler transistor
NPN
16 z16 E- output (NPN) C+
d16
d16 C+ Switch point 2 measuring circuit 2
24 V DC (max. 50 mA SPS/PLS E-
18 z16
Switch point 1 for measuring circuit 1 or 2 FSL (Fail Safe Low) Lo-Alarm
Opto-electronic coupler transistor output NPN 1 disabled (measured value < limit switch value)
Switch point 1 for measuring circuit 1 or 2 FSH (Fail Safe High) Hi-Alarm
Opto-electronic coupler transistor output NPN 1 disabled (measured value > limit switch value)