Professional Documents
Culture Documents
Silicate
Silicate
Silicate
ANALYSIS: SILICATE
RANGE: 1 - 250 µg Si/liter
SAMPLE: DRINKING, SURFACE, FRESH,
WASTE & SEA WATER
SAN++
Catnr. 563-051 issue 011912/MH/99270483
PRINCIPLE
The automated procedure for the determination of Silicate is based on the following reaction; the sample is
acidified and mixed with an ammonium heptamolybdate solution forming molybdosilicic acid. This acid is
reduced with L(+)ascorbic acid to a blue dye, which is measured at 810 nm. Oxalic acid is added to avoid
phosphate interference.
LABORATORY FACILITIES
1. Maximum power consumption depending on the analyser configuration, 2000 VA. Check voltage at the
back of the instrument before installation.
2. Facilities for chemical wastes. Check environmental regulations for proper disposal of waste.
Field of application
Sample preparation for the determination of total alkalinity, aluminium, amino acids, ammonia, anionic
surfactants (MBAS), bromide, calcium, C.O.D., chromium, conductivity, D.O.C., ethanol, hardness,
magnesium, manganese, methanol, nitrate + nitrite, nitrite, nonionics, ortho-phosphate, potassium
permanganate, potassium, total-N (UV digestion), protein, silicate, sodium, volatile acids and urea in water.
Principle
The sample is not preserved and analysed as soon as possible, at least within 24 hours after collection.
Store the sample at 4°C till analysis.
Note : The sample preparation in this method is only meant to be an example procedure. It requires
validation on its correct procedure for all types of samples to be analyzed, which is the responsibility of the
user.
catnr. 563-051 1 of 6
REAGENTS
Remark: All reagents and standards must be prepared in polyethylene and stored in polyethylene
bottles.
Required chemicals: Sulfuric acid ............... 22.5 ml. Preparation: Dilute the sulfuric acid in ± 800 ml
H2SO4 (95-97%) distilled water. Fill up to 1 liter with
Distilled water * ...... 977.5 ml. distilled water and mix.
H2O
Note: Solution is stable for 1 week. Store at
4°C when the solution is not used.
Required chemicals: Oxalic acid ...................... 44 g. Preparation: Dissolve the oxalic acid in ± 800 ml
C2H2O4.2H2O distilled water. Fill up to 1liter with
Distilled water * ....... 1000 ml. distilled water and mix.
H2O
Note: Store in a polyethylene bottle. Solution is
stable for 1 month. Store at 4°C when the
solution is not used.
Required chemicals: L(+)Ascorbic acid ............ 40 g. Preparation: Dissolve the L(+)ascorbic acid in ± 800 ml
C6H8O6 distilled water. Fill up to 1 liter with
Distilled water * ....... 1000 ml. distilled water and mix.
H2O
Note: Solution is stable for 1 week. Store at 4°C
when the solution is not used.
catnr. 563-051 2 of 6
STANDARDS
Working standards
250 µg Si/liter: Dilute 2.5 ml stock solution 10 mg Si/liter to 100 ml with rinsing liquid sampler.
200 µg Si/liter: Dilute 2.0 ml stock solution 10 mg Si/liter to 100 ml with rinsing liquid sampler.
150 µg Si/liter: Dilute 1.5 ml stock solution 10 mg Si/liter to 100 ml with rinsing liquid sampler.
100 µg Si/liter: Dilute 1.0 ml stock solution 10 mg Si/liter to 100 ml with rinsing liquid sampler.
50 µg Si/liter: Dilute 0.5 ml stock solution 10 mg Si/liter to 100 ml with rinsing liquid sampler.
Note:
1. Prepare the working standards fresh daily and store in polyethylene measuring flasks.
2. The water used for the reagents and standards must be degassed properly before making up the
reagents and standards. Especially water produced by reverse osmosis or ion exchange equipment
contains a lot of gasses, which must be removed by degassing (degassing procedure see operational
remarks and trouble shooting).
1. System sample time: 100 sec., wash time: 100 sec., air time: 0 sec.
2. Module sample time: 60 sec., wash time: 60 sec., air time: 0 sec.
3. Calibration order: 1st order (acc.ISO8466-1)
catnr. 563-051 3 of 6
OPERATIONAL REMARKS AND TROUBLESHOOTING
GENERAL REMARKS
1. In most analyses, the first peak coming from the baseline is lower than it should be. The first peak
coming from the baseline is rejected and will not be included in the calculation of the CV-value.
2. If the sample take up volume is less than 1.00 ml/min, a bypass is required to increase the sample
stream to approximately 1.00 ml/min. For SA 5210 debubbler, when the sample is 1.00 ml/min and
higher use 0.23 ml/min debubble tube.
MAINTENANCE
1. Daily; When daily analysis is finished rinse all reagentlines with distilled water for 30 minutes.
2. Weekly; To decontaminate the system, rinse 30 minutes with 0.5M sodium hydroxide solution and
another 30 minutes with distilled water.
3. Monthly; Replace all pump tubing (or shift pump tube one colour bridge on pump)
MODULE CONSUMABLES
catnr. 563-051 4 of 6
MODULE COMPONENTS
REFERENCES
1. Environmental Protection Agency, methods for chemical analysis of water and wastes, 1983.
2. Standard Methods for the determination of water and waste water, 17th edition, 1989.
3. ASTM, 1990.
4. International Organisation for Standardisation, ISO-5667-3.
catnr. 563-051 5 of 6
FLOW DIAGRAM
waste
ml/min
Flowcell 50mm
Filter 810nm
Cor. Filter 1010nm
completelybuilt up in glass
1.20 waste
Sampler
catnr. 563-051 6 of 6