TB/09/001
New procedure for Optical Gain
Issued by PhD
Adjustment on
Mythic 22 OT/CT
Object:
This technical bulletin describes the new procedure to adjust the optical gain on
Mythic 22 OT/CT.
Who is concerned:
Software version :
G-Cal Solution
This new procedure replaces the optical gain adjustment using WBC latex
particles.
FSC: 80
ALL: 63
The G-Cal doesn’t require adjusting the gains manually using a fresh blood
sample.
The G-Cal is a neutrophile solution. PLEASE KEEP IT AT 2-8°c. Expiry date of
G-CAL is 4 months even opened.
Before to use G-Cal, let it at room temperature.
We deeply recommend to install the new version of the software on your entire
installed base, and to run the G-Cal solution, especially if you receive some
complaints about some parameters too high or too low.
NOTE: You don’t need to recalibrate the instrument at the end of this
procedure.
Philippe Daire
Direct phone number: +41228849083
E-mail address: phdaire@orphee-medical.com