Air Flow Sensor Calibration Report Kemwell 4

You might also like

Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 2

S V SCIENTIFIC PVT LTD

30/A 2nd Floor, Serpentine Road, Kumara Park West


Seshadripuram Post, Bangalore - 560 020

PERFORMANCE QUALIFICATION PROTOCOL


TYPE/MODEL BIO SAFETY CABINET / LA2-4A1 Serial no 2011-57031
DOCPage
No 1 SV/ESCO/BSC/2A2/23-24 Page No.

AIR FLOW SENSOR CALIBRATION REPORT

Instrument m/s External Deference b/w Acceptance Remark Pass/fail


anemometer internal and criteria
external
anemometer
I D I D I D

0.53 0.35 0.53 0.33 0.00 0.02 0.05 m/s Pass

EXTERNAL ANEMOMETER DETAILS:

INSTRUMENT :
MAKE :
MODEL :
INSTRUMENT I. D :
CALIBRATED ON :
CALIBRATION DUE ON :