Comparative Analysis of Cadmium-Zincum-Telluride Cameras Dedicated To Myocardialperfusion SPECT - A Phantom Study

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 9

ORIGINAL ARTICLE

Comparative analysis of cadmium-zincum-


telluride cameras dedicated to myocardial
perfusion SPECT: A phantom study
Orazio Zoccarato, PhD,a Domenico Lizio, MSc,b Annarita Savi, PhD,c
Luca Indovina, PhD,d Camilla Scabbio, MSc,e Lucia Leva, MD,b Angelo del Sole,
MD,e Claudio Marcassa, MD,a Roberta Matheoud, PhD,b Michela Lecchi, PhD,e
and Marco Brambilla, PhDb,f
a
Unit of Nuclear Medicine and Department of Cardiology, S. Maugeri Foundation, IRCCS,
Scientific Institute of Veruno, Veruno, NO, Italy
b
Departments of Medical Physics and Nuclear Medicine, University Hospital ‘Maggiore della
Carità’, Novara, Italy
c
Department of Nuclear Medicine, Hospital San Raffaele, Milan, Italy
d
Department of Medical Physics, Polyclinic Agostino Gemelli, Rome, Italy
e
Department of Health Sciences, University of Milan and Nuclear Medicine Unit, San Paolo
Hospital, Milan, Italy
f
Medical Physics Department, University Hospital of Novara, Novara, Italy

Received Feb 5, 2015; accepted Jun 2, 2015


doi:10.1007/s12350-015-0203-7

Background. This investigation used image data generated by an anthropomorphic


phantom with a cardiac insert for a comparison between two solid state cameras: D-SPECT and
D530c.
Methods. For each camera, two sets (with and without a simulated transmural defect (TD))
of scans were acquired starting from the in vivo standard count statistics in the left ventricle
(LV). Other two acquisitions corresponding to 150% and 50% of the reference count statistics
were acquired. Five performance indices related to spatial resolution, contrast, and contrast-to-
noise ratio (CNR) were analyzed.
Results. D-SPECT showed a lower LV wall thickness and an inferior sharpness than
D530c. No significant differences were found in terms of contrast between LV wall and the
inner cavity, TD contrast or CNR. No significant differences were observed in CNR when
moving from the reference level of count statistics down to 50% or up to 150% of the counts
acquired on the LV.
Conclusions. Our results show that D-SPECT and D530c have different performances. The
lack of differences in the image performance indices along the range of count statistics explored,
indicates that there is the possibility for a further reduction in the injected activity and/or the
acquisition time, for both systems. (J Nucl Cardiol 2015)
Key Words: SPECT Æ Cardiac Æ Cadmium-zincum-telluride

Reprint requests: Marco Brambilla, PhD, Medical Physics Department,


University Hospital of Novara, Novara, Italy; marco.brambilla@
maggioreosp.novara.it
1071-3581/$34.00
Copyright Ó 2015 American Society of Nuclear Cardiology.
Zoccarato et al Journal of Nuclear CardiologyÒ
Performance of CZT Cameras in cardiac SPECT

cameras over a wide range of acquisition count statistics,


Abbreviations
TD Transmural defect using an anthropomorphic phantom and selected phys-
LV Left ventricle ical figures of merit for the image evaluation. The
AC Attenuation correction application of attenuation correction was also studied
CNR Contrast-to-noise ratio with the D530c.
CZT Cadmium-zinc-telluride
D530c Discovery 530c
ROI Region of interest MATERIALS AND METHODS
FWHM Full width at half maximum
CIC Percent contrast between LV wall and Gamma cameras
inner cavity Two nuclear medicine departments (referred later as
CTD Percent TD contrast in LV wall ‘centers’) participated to the study allowing the use of the
following gamma cameras: D-SPECT (Center N.1) and D530c
(Center N.2). The scanners characteristics and the acquisition
See related editorial, doi:10.1007/s12350- and reconstruction protocols are detailed in Table 1.
Both D-SPECT and D530c use identical square CZT
015-0216-2. modules. Each module consists of a 16x16 pixelated CZT crystal
backed by a proprietary application-specific integrated circuit.
The D-SPECT system uses 9 rotating detector column
INTRODUCTION blocks of pixelated CZT detectors (pixel size, 2.5 9 2.5 mm)
associated with a wide-angle square-hole tungsten collimator.
Recently two direct-conversion cadmium-zinc-tel- A total of 120 projections are recorded by each block.
luride (CZT) detector cameras have become available: the D-SPECT requires a brief prescan, thereafter the technologist
D-SPECT (Spectrum Dynamics, Caesarea Israel) and draws a region of interest (ROI) to define the left ventricle
Discovery 530c (D530c, GE Healthcare, Haifa Israel). (LV) for cardiocentric detector column movements. Utilizing
These systems enhance the sensitivity of count count rate information from the prescan, D-SPECT adjusts
detection enabling the reduction of scan times and/or scan time in 1 second increments to attain a predetermined
tracer activity using very different technologies: 9 number of counts. A specific algorithm of iterative recon-
mobile columns of CZT detectors with wide parallel- struction is used to compensate for the collimator-related loss
hole collimators for D-SPECT and 19 fixed CZT in spatial resolution.10
detectors with multi-pinhole collimators for the D530c. D530c employs an L-shaped array of 19 stationary
pinhole units, which are placed over an arc of 180°, orientated
Several studies have been published since CZT
to focus on the heart area and arranged in three rows. Each unit
cameras were implemented in the clinical practice. The consists of a tungsten-tipped lead-bodied pinhole collimator
major aim of these studies was to evaluate the physical with an effective aperture of 5.1 mm, projecting onto 4 CZT
characteristics,1,2 the performances in comparison to modules of 32x32 CZT pixels (pixel size, 2.5 9 2.5 mm)
conventional Anger cameras,3,4 their potential to reduce arranged in a 2 9 2 square.2 Nineteen projections are thus
scan times and/or tracer activity,5,6 and/or their effec- recorded, and an iterative reconstruction allows modeling for
tiveness in clinical use.7,8 collimator geometry. Attenuation correction for the D530c
The superior performances reported for these two studies was performed by means of a computed tomography
CZT systems in comparison to conventional Anger scan of the phantom.
cameras, although based on identical detector material, After phantom acquisitions and image reconstruction,
are strictly dependent on the intrinsically different image quality evaluation and statistical analysis were per-
formed at a core laboratory.
technology and acquisition techniques which are, in
clinical conditions, potentially capable of producing
significantly different results. Anthropomorphic Phantom Preparation
The only published study comparing different CZT
cameras showed different performances depending on the An anthropomorphic phantom of the chest, with inserts
figure of merit considered.9 However, it was restricted to simulating lungs, liver, spine, LV wall, LV chamber, and a
only a single point of the count statistic (500 kilocounts transmural defect (TD), was used (Torso PhantomTM and
Cardiac InsertTM, Data Spectrum Corporation, Hillsborough,
over the left ventricle) and used a simplified phantom,
NC,USA) (Figure 1). The ratio of radioactivity concentrations
which is not expected to accurately mimic the conditions for the different inserts in the phantom study was derived from
encountered during clinical acquisitions. a previous publication.11 The TD insert (45° 9 2 cm) was
The aim of this study was to characterize the filled with non-radioactive water and located in mid-septal
relative performances of the above-mentioned CZT position in the LV wall. The lung inserts, filled with
Journal of Nuclear CardiologyÒ Zoccarato et al
Performance of CZT Cameras in cardiac SPECT

Table 1. Details of gamma cameras, acquisition, and reconstruction protocols used for the phantom
studies

D-SPECT Discovery 530c#


Acquisition
Collimator Wide-angle parallel hole Multi-pinhole
Energy window 140 keV ± 10% 140 keV ± 10%
Number of projections 120 (99 blocks) 19
Detector angle between consecutive projections 0.4°-7°* Fixed detector
Reconstruction
Method Iterative 3D Iterative 3D
Number of iterations 9 subsets (4 ? 3) 9 32 60 9 1
Iteration filter Kernel (0.125 mm)
Post reconstruction filter Normalization filter§ Butterworth (order 7,
cutoff 0.37 cm-1)
#
Offline CT was employed to acquire the attenuation map used to correct the emission data.
* Angles are approximately 0.4° for projections passing through cardiac region (as defined by ROI positioning on the left
ventricle on prescan images) and up to 7° for other projections.
§
Proprietary infomations.

Phantom Acquisitions and Reconstruction


In each center, two sets (with and without TD) of scans
were acquired starting from the in vivo standard count
statistics of 1 9 106 counts in the LV. Then, the other two
acquisitions corresponding to 150% and 50% of the reference
count statistics were acquired, changing in each step only the
acquisition time. Images were reconstructed using all the
reconstruction parameters currently recommended for clinical
routine (Table 2). Images were reconstructed with and without
the attenuation correction (AC) for the D530c. No AC is
currently available on the D-SPECT.

Image Quality Evaluation


After reconstruction, all image sets were sent to a core
Figure 1. The anthropomorphic torso phantom used for the laboratory for processing using the following five physical
study. figures of merit, associated with image quality in cardiac SPECT:
(a) LV wall thickness;
TM
(b) Sharpness index;
Styrofoam beads and non-radioactive water, were used to (c) Contrast between LV wall and inner chamber;
simulate lung tissue attenuation density. The other phantom (d) Contrast-to-noise ratio;
chambers were filled with 99mTc solutions of different (e) TD contrast with LV wall;
radioactivity concentrations: 158 ± 5 kBqmL-1 for LV wall,
95 ± 3 kBqmL-1 for liver, and 8.1 ± 0.6 kBqmL-1 for chest *** The first four indexes were evaluated using the image
and LV chamber. Actual activity concentrations in the phan- sets without TD insert in the LV wall. For each figure of merit,
tom’s chambers were planned using activity meters with an each measurement was repeated at least fourfold in order to
accuracy within 5%, as measured during routine quality provide an estimate of the measurement errors. The figures of
controls. All compartment contents were well mixed to avoid merit were evaluated using ImageJ software.12
artifacts caused by inhomogeneous activity. The same phan- LV wall thickness. The LV wall thickness was
tom was used in all centers. The accurate positioning of the evaluated as ‘‘full width at half maximum’’ (FWHM) of the
phantom was secured by a flat portion of the phantom which myocardial signal in the mid-ventricular section of the LV
allowed a reproducible positioning on the imaging bed and by short axis.11 The FWHM was calculated convolving the central
marking the position of the first acquisition. count profile, obtained using ImageJ software, with a
Zoccarato et al Journal of Nuclear CardiologyÒ
Performance of CZT Cameras in cardiac SPECT

Table 2. Results for the five physical indexes used in this study to evaluate image quality in relation to
scanner/software combinations and count statistics

Thicknessof LV Sharpness
Main effect Description wall (mm) index (cm21) CIC (%) CNR CTD (%)
Scanner/ D-SPECT 13.6 ± 0.7 0.56 ± 0.02 80.9 ± 3.2 4.4 ± 0.9 59.3 ± 4.6
software Discovery 530c 19.6 ± 1.3 0.66 ± 0.04 83.9 ± 5.8 4.5 ± 0.7 55.8 ± 5.9
Discovery 530c-AC 19.6 ± 1.1 0.69 ± 0.01 84.5 ± 3.5 4.8 ± 0.8 56.1 ± 2.8
ANOVA: P value <0.0001* <0.0001* 0.25 0.41 0.11
Study count 1.5 Mcounts 17.6 ± 2.9 0.63. ± 0.06 78.5 ± 5.7 5.0 ± 0.7 57.9 ± 4.4
statistics 1.0 Mcounts 17.7 ± 3.1 0.63. ± 0.07 79.6 ± 5.2 4.5 ± 0.7 57.7 ± 4.8
0.5 Mcounts 17.5 ± 3.1 0.64. ± 0.05 80.5 ± 5.1 4.2 ± 0.9 56.1 ± 5.3
ANOVA: P value 0.67 0.62 0.55 0.018* 0.57

LV, left ventricular; IC, inner chamber; TD, transmural defect; AC, attenuation correction.
* Statistically significant.

theoretical Gaussian one. The true wall thickness of the LV endocardial and epicardial borders. Mean background counts
phantom insert is 10 mm. and corresponding SD were determined on a half-moon-shaped
Sharpness index. A sharpness index was deter- ROI, at 15 mm from the lateral wall (Figure 2B). Contrast-to-
mined on the horizontal profile of the mid-ventricular short- noise ratio was estimated using the following formula:
qffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
axis slice. This index was computed as the maximal slope of
CNR ¼ ðMMy  MBG Þ= SD2My þ SD2BG ;
the decrease in myocardial counts (cm-1) on the epicardial
border of the lateral wall and after exclusion of the low-count
where MMy is mean myocardial counts, MBG is mean
part of the profile (activities \ 25% of the maximal value).9
background counts, SDMy is mean myocardial SD, and
Contrast between LV wall and inner
SDBG is mean background SD.9
chamber. Two ROIs were drawn on the mid-ventricular
short-axis images in order to estimate gLV and gIC as the mean TD contrast in LV wall. TD contrast in LV wall
concentration activity per voxel for LV wall and inner cavity, was evaluated both in LV short-axis and in horizontal long-
respectively. The percent contrast between LV wall and inner axis images (Figure 2C). ROIs were drawn on the central
cavity (CIC) was defined as the ratio between measured short-axis and horizontal long-axis images of the phantom
contrast (CM) and the true one (CT) as: acquisitions with TD defect in order to estimate gLV and gIC as
the mean counts per voxel for the LV wall and TD,
CM respectively. The TD contrast in LV wall (CTD) was defined
CIC ð%Þ ¼  100;
CT as the percent difference between gLV and gIC as:
gLV  gTD
where CM and CT were calculated using: CTD ð%Þ ¼  100
g  gIC gLV
CM ¼ LV
gLV þ gIC The ideal CTD (%) is 100% for all central acquisitions.11
CLV  CIC
CT ¼
CLV þ CIC
Statistical Analysis
and CLV and CIC are the true activity concentrations in
The impact of the scanner/software combinations and of
the inserts simulating the LV and the inner chamber,
the different levels of count statistics on each of the image
respectively. The ideal value for CIC is 100% for all quality indexes, was assessed by a two-way factorial
center acquisitions.11 Figure 2A provides a schematic ANOVA. Scanner/software combinations and the relative
representation of the ROI dimensioning and placement level of count statistics were considered as independent
for the calculation of CIC (%). variables (factors) and image quality indexes as dependent
Contrast-to-noise ratio. This ratio was deter- variables.
mined on the mid-ventricular short-axis slice of each SPECT A post hoc test (Scheffe F test) was performed to identify
recording. Mean myocardial counts and corresponding SD the main sources of variability. If a significant F value was
were determined on a ring-shaped ROI encompassing the found for one independent variable, then this was referred as a
Journal of Nuclear CardiologyÒ Zoccarato et al
Performance of CZT Cameras in cardiac SPECT

main sources of variability. Analysis was performed with


Statistica version 6.0 (StatSoft Inc., Tulsa, OK, USA) using a
two-sided type I error rate of P = .05. Values are reported as
observed means ± standard deviations and shown as least
square averages ± standard errors of the means.

RESULTS

Phantom and Image Quality Evaluation


An example of the reconstructed image sets (best
short-axis slices for TD phantom) for the two cameras
and a relative level of count statistics of 100% and of
50%, is reported in Figure 3. The detailed comparison
between the statistical results for the five physical
indexes used in this study to evaluate image quality in
relation to scanner/software combination and relative
level of count statistics are reported in Table 2, and
summarized below.
LV wall thickness. The scanner/software com-
bination (F = 376; P \ .0001) was a main effect with a
statistically significant impact on the thickness degra-
dation of the LV wall. Post hoc test of the different
scanner/software combinations showed a significant
increase in FWHM values from D-SPECT to D530c
(13.6 ± 0.7 vs 19.6 ± 1.3 mm; P \ .0001). No signifi-
cant differences were observed between D530c and
D530c_AC (19.2 ± 1.1 mm; P = .99). The relative
level of count statistics was not a main effect
(F = 0.4; P = .67) (Figure 4).

Figure 2. Example of the ROI dimension and position used


for the evaluation of the image quality indexes: A CIC; B CNR;
C CTD. Figure 3. The transaxial slices of the TD phantom acquired
with the D_SPECT (left column), the D530c (central column),
and D530C_AC (right column) for count statistics of 1.5,
main effect. When a main effect was found, then a post hoc test (upper row), 1 (central row), and 0.5 Mcounts (lower row), are
(Scheffé test) was performed to compare the dependent shown. The display image zoom has been adjusted to take in
variable with the levels of the factor, thus identifying the account for the different pixel size.
Zoccarato et al Journal of Nuclear CardiologyÒ
Performance of CZT Cameras in cardiac SPECT

significant impact on the LV wall/inner chamber con-


trast (Figure 6).
Contrast-to-noise ratio. The scanner/software
combination was not a main effect (F = 0.92; P = .41).
The count statistics (F = 4.4; P = .018) was a main
effect with a statistically significant impact on the CNR.
Post hoc count tests showed a significant increase in the
CNR values from 0.5 to 1.5 Mcounts (4.2 ± 0.9 vs
5.0 ± 0.7; P = .020). No significant differences were
observed in CNR between 0.5 and 1 Mcounts (4.2 ± 0.9
vs 4.5 ± 0.7; P = .19) or between 1 and 1.5 Mcounts
(4.5 ± 0.7 vs 5.0 ± 0.7; P = .57) (Figure 7).
TD contrast in LV wall. Neither the scan-
ner/software combination (F = 2.29; P = .11;
Figure 7A) nor the count statistics (F = 0.58;
Figure 4. LV wall thickness, expressed as FWHM (mm), as a P = .57) were main effects with a statistically signifi-
function of scanner/software combination and count statistics. cant impact on the TD contrast in LV wall (Figure 8).
Points represent least square averages; vertical bars represent
95% confidence intervals.
DISCUSSION
Sharpness index. The scanner/software combi- In the present study, head-to head comparisons were
nation (F = 64; P \ .0001) was a main effect with a conducted between the D-SPECT and D530c cameras,
statistically significant impact on the sharpness index. including D530c acquisitions in conjunction with an
Post hoc test of the different scanner/software combi- offline AC, by analyzing a number of physical figures of
nations showed a significant increase in sharpness values merit related to the quality of myocardial SPECT
from D-SPECT to D530c (0.56 ± 0.02 vs images, namely LV wall thickness, sharpness of the
0.66 ± 0.04 cm-1; P \ .0001). No significant differ- myocardial contour, contrast between the LV wall and
ences were observed between D530c and D530c_AC the inner chamber, contrast-to-noise ratio and transmural
(0.69 ± 0.01 cm-1; P = .13). The relative level of count defect contrast in the LV wall. This is the first study, to
statistics was not a main effect (F = 0.5; P = .621) the best of our knowledge, investigating the behavior of
(Figure 5). these performance parameters over a wide interval of
Contrast between LV wall and inner count statistics, related to clinical studies.
chamber. Neither the scanner/software combination Different from previous studies, that used a simpli-
(F = 1.5; P = .25) nor the count statistics (F = 0.55; fied cardiac insert, in the present study the parameters
P = 0.59) were main effects with a statistically

Figure 6. LV wall/inner chamber contrast (CIC (%)) as a


Figure 5. Sharpness index as a function of scanner/software function of scanner/software combination and count statistics.
combination and count statistics. Points represent least square Points represent least square averages; vertical bars represent
averages; vertical bars represent 95% confidence intervals. 95% confidence intervals.
Journal of Nuclear CardiologyÒ Zoccarato et al
Performance of CZT Cameras in cardiac SPECT

and/or noise suppression demonstrated values of wall


thickness (16-19 mm) even lower. Liu et al14 also
reported thicker myocardial walls using the D530c and
an anthropomorphic phantom (21-26 mm) in compar-
ison to a conventional Anger camera (16-18 mm). The
use of AC on D530c studies did not result in significant
differences in wall thickness; this was an anticipated
result, since AC is not expected to impact on spatial
resolution. At odds with the previous data, lower values
of wall thickness (13.6 mm) were documented with
D-SPECT, in comparison to both D530c and conven-
tional cameras. This is clearly visible from Figure 3.
Other authors9 have previously demonstrated that the
central spatial resolution is slightly better with the
D530c (6.7 mm) than with D-SPECT (8.6 mm). Thus,
Figure 7. Contrast-to-noise ratio (CNR) as a function of the lower LV wall thickness exhibited by D-SPECT can
scanner/software combination and count statistics. Points be explained only by the reconstruction protocol. The
represent least square averages; vertical bars represent 95% default algorithm D-SPECT uses is entirely based on
confidence intervals. OSEM reconstruction. An initial standard OSEM image
(generally 3 iterations) is first reconstructed on which
the myocardial contours are then estimated to generate a
uniform model of the myocardium. This model is then
used as initial guess (prior) for additional iterations
(generally 4 iterations). Thus, the D-SPECT image is the
optimal result reached by the OSEM iterative recon-
struction using the myocardial contours as initial guess.
Other manufacturers use a FBP image as initial guess for
the iterative reconstructions. It is worth saying that
D-SPECT is neither projecting counts, nor placing the
reconstructed pixels of the myocardium into the best
position. In a similar way, the ‘‘motion-frozen’’
approach proposed by Slomka et al, based on three-
dimensional LV contours identified on the images of the
individual time phases, has demonstrated to improve LV
wall thickness.15
Figure 8. Transmural defect contrast (CTD) as a function of
scanner/software combination and count statistics. Points Sharpness Index
represent least square averages; vertical bars represent 95%
confidence intervals. The sharpness profile is a parameter influenced by
spatial resolution.16 Imbert et al9 demonstrated values
were determined on an anthropomorphic torso phantom for the sharpness index of myocardial contours on
with a cardiac insert, in order to mimic the conditions of human images using the D530c (1.02 cm-1) higher than
attenuation, scatter, and partial volume effects that can those obtained with the D-SPECT (0.92 cm-1). Our
be encountered in clinical practice. results support this trend, showing significantly superior
values for the sharpness index measured on phantom
images for the D530c (0.66 cm-1) with respect to
Wall Thickness
D-SPECT (0.56 cm-1) and are in quantitative agree-
According to our results, the wall thickness mea- ment, as for the D530c, with the results reported by Liu
sured on the D530c was of the same order of magnitude et al using the same phantom.14 This can be accounted
(20 mm) of that measured on conventional gamma mainly because of the difference in voxel size used in
cameras using OSEM reconstruction11 or equipped with image reconstruction, since we can expect higher
confocal collimators with dedicated reconstruction soft- degradation of the sharpness index with higher voxel
ware.13 Moreover, the same conventional gamma size. As a matter of fact, the voxel size was 4.92 mm in
cameras equipped with software for resolution recovery D-SPECT studies and 4.0 mm in D530c studies.
Zoccarato et al Journal of Nuclear CardiologyÒ
Performance of CZT Cameras in cardiac SPECT

Image Contrast NEW KNOWLEDGE GAINED


No statistically significant differences were found in Some differences between the D-SPECT and D530c
the contrast between the myocardial wall and the LV camera performances are documented, mainly in the LV
inner cavity among the D-SPECT (81%), the D530c wall thickness; however, perfusion defect detection was
(84%), and the D530c_AC (85%). Noteworthy, the similar using both cameras. The most important finding
values of CIC (%) using CZT cameras were higher than is that similar performances of both cameras were
those observed using conventional gamma cameras obtained with lower and higher acquired counts. Thus,
(46%-48%) even when the latter were equipped with increased tracer doses are not needed for improved
software for resolution recovery (64%) and scatter image quality. Additional decrease of tracer dose or
corrections were also applied (73%).11 additional decrease of imaging time may be possible
A slight trend toward better values for CTD (%) for using CZT cameras.
D-SPECT (59.3%) than for D530c (55.8%) can be
observed, as reported in Figure 8, although statisti-
CONCLUSIONS
cally not significant. Comparable values of CTD (%)
(58%-60%) were previously reported using a conven- The two analyzed SPECT systems, although based on
tional gamma camera with confocal collimators and the same detectors (semiconductor CZT detectors), are
without AC in Caobelli et al.13 inherently different in both the geometry of data acqui-
sition as well as in the image reconstruction algorithms.
In clinical conditions and using the reconstruction
Contrast-to-Noise Ratio
parameters currently recommended for the clinical
The CNR values obtained were 4.4, 4.5, and 4.8 for routine, our results show that these different character-
the D-SPECT, D530c, and D530_AC, respectively; istics lead to significantly different results, mainly in the
these values are in substantial agreement with those LV wall thickness. No significant differences were
reported by Imbert et al for the D-SPECT (4.1) and for found in the contrast between the myocardial wall and
the D530c (4.6).9 However, CNR values in our study are the inner cavity or the contrast on the transmural defect.
higher than those reported by Liu et al for the D530c The application of the AC to the D530c system, did
(*3.8).14 not improve any of the figures of merit considered.
However, these results were obtained with a phantom
whose dimensions reproduce a patient with normal
Effect of Count Statistics
weight, and do not necessarily apply to overweight or
For none of the above performance parameters, the obese patients.
level of the count statistics resulted in a significant factor The lack of significant differences in the considered
(Figures 4, 5, 6, and 8). Hence, in the range of 50%- parameters along the range of count statistics explored,
150% of the reference count statistic of 1 Million counts combined with the increased sensitivity of these systems
over the LV, the values of wall thickness, sharpness, CIC (fourfold for the D530c to sevenfold for the D-SPECT)
(%), and CTD (%) were not affected by the level of in comparison to a conventional Anger camera,9,14
counts acquired. On one side, this suggests that increas- seems to indicate that there is still a possibility for a
ing the count statistics above the level recommended by further reduction in the injected activity and/or the
the manufactures and currently used in the two partic- acquisition time, in order to improve patients’ comfort
ipating in the present study does not bring any additional and contain both patients’ and operators’ radiation
benefits in terms of resolution and/or contrast. On the exposure.
other side, there is still a possibility for the optimization
of acquisition protocols in order to further reduce the
Disclosure
emission scan duration and/or the injected activity of
radiopharmaceutical. The authors have indicated that they have no financial
The only performance parameter which seems conflict of interest.
somewhat influenced by the level of count statistics is
the CNR (Figure 7). However, the only comparison that
reached a statistical significance was that between 0.5 References
Mc and 1.5 Mc, while no statistical significant differ-
ences were observed moving from the reference level of 1. Erlandsson K, Kacperski K, Van Gramberg D, Hutton BF. Per-
count statistics down to 50% or up to 150% of the counts formance evaluation of D-SPECT: A novel SPECT system for
acquired on the LV. nuclear cardiology. Phys Med Biol 2009;54:2635-49.
Journal of Nuclear CardiologyÒ Zoccarato et al
Performance of CZT Cameras in cardiac SPECT

2. Bocher M, Blevis IM, Tsukerman L, Shrem Y, Kovalski G, to myocardial perfusion SPECT: A comprehensive analysis of
Volokh L. A fast cardiac gamma camera with dynamic SPECT phantom and human images. J Nucl Med 2012;53:1898-903.
capabilities: Design, system validation and future potential. Eur J 10. Patton JA, Slomka PJ, Germano G, Berman DS. Recent technologic
Nucl Med Mol Imaging 2010;37:1887-902. advances in nuclear cardiology. J Nucl Cardiol 2007;14:501-13.
3. Sharir T, Ben-Haim S, Merzon K, Prochorov V, Dickman D, Ben- 11. Zoccarato O, Scabbio C, De Ponti E, Matheoud R, Leva L,
Haim S, et al. High-speed myocardial perfusion imaging initial Morzenti S, et al. Comparative analysis of iterative reconstruction
clinical comparison with conventional dual detector anger camera algorithms with resolution recovery for cardiac SPECT studies. A
imaging. JACC Cardiovasc Imaging 2008;1:156-63. multi-center phantom study. J Nucl Cardiol 2014;21:135-48.
4. Esteves FP, Raggi P, Folks RD, Keidar Z, Askew JW, Rispler S, 12. Rasband WS. ImageJ. Bethesda, MD: U.S. National Institutes of
et al. Novel solid-state detector dedicated cardiac camera for fast Health; 1997-2012. http://imagej.nih.gov/ij/. Accessed 23 Jan
myocardial perfusion imaging: Multicenter comparison with 2015.
standard dual detector cameras. J Nucl Cardiol 2009;16:927-34. 13. Caobelli F, Kaiser SR, Thackeray JT, Bengel FM, Chieregato M,
5. Gambhir SS, Berman DS, Ziffer J, Nagler M, Sandler M, Patton J, Soffientini A, et al. IQ SPECT allows a significant reduction in
et al. A novel high-sensitivity rapid acquisition single-photon administered dose and acquisition time for myocardial perfusion
cardiac imaging camera. J Nucl Med 2009;50:635-43. imaging: Evidence from a phantom study. J Nucl Med
6. Herzog BA, Buechel RR, Katz R, Brueckner M, Husmann L, 2014;55:2064-70.
Burger IA, et al. Nuclear myocardial perfusion imaging with a 14. Liu CJ, Cheng JS, Chen YC, Huang YH, Yen RF. A performance
cadmium-zinc-telluride detector technique: Optimized protocol for comparison of novel cadmium-zinc-telluride camera and conven-
scan time reduction. J Nucl Med 2010;51:46-51. tional SPECT/CT using anthropomorphic torso phantom and water
7. Duvall WL, Sweeny JM, Croft LB, Barghash MH, Kulkarni NK, bags to simulate soft tissue and breast attenuation. Ann Nucl Med
Guma KA, et al. Comparison of high efficiency CZT SPECT MPI 2015. doi:10.1007/s12149-015-0952-z.
to coronary angiography. J Nucl Cardiol 2011;18:595-604. 15. Slomka PJ, Nishina H, Berman DS, Kang X, Akincioglu C,
8. Buechel RR, Herzog BA, Husmann L, Burger IA, Pazhenkottil Friedman JD, et al. ‘‘Motion-Frozen’’ display and quantification
AP, Treyer V, et al. Ultrafast nuclear myocardial perfusion of myocardial perfusion. J Nucl Med 2004;45:1128-34.
imaging on a new gamma camera with semiconductor detector 16. Shea SM, Kroeker RM, Deshpande V, Laub G, Zheng J, Finn JP,
technique: First clinical validation. Eur J Nucl Med Mol Imaging et al. Coronary artery imaging: 3D segmented k-space data
2010;37:773-8. acquisition with multiple breath-holds and real-time slab follow-
9. Imbert L, Poussier S, Franken PR, Songy B, Verger A, Morel O, ing. J Magn Reson Imaging 2001;13:301-7.
et al. Compared performances of high-sensitivity cameras dedicated

You might also like