Calibrare TCS B2

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 1

E-102

CHAPTER 4. Calibration
When you calibrate the indicator, the load cell's shield cable must connected with GND,
otherwise the indicator reads will have errors. (Figure 3-3)
Calibration should be operated by measure technician is as follows:
1. Weight calibration
--"-'-""""'~-' -"~''" '

Power on, press "calibration" and " Enter" together when self-checking, loosen when hear
1
buzzer.
"-I. '1
[pass) It will display [pass) after Self checking and ask to
2
input calibration password ~rl /~
Input calibration [ ------1) Press [ to input data, press [Enter) to confirm;
password "01 0201" press [Tare) to modify last bit of password or exit
3
Press [Enter l [d= xxl calibration. Password right to enter division setting,
otherwise exit calibration
[d= xx) Division setting
.rivr-J 4 press [ t l [d= yy) (1, 2, 5, 10, 20, 50, 100, 200); Press [Enter) to confirm
press [Enter) [dot= x) Confirm the division and enter point setting
-"----------

[dot= xl Point setting

[dot= yl I Press [ tl to modify decimal digits, at best 5 digits


5 press [ t l
[ZEroxxl confirm the decimal digits and enter zero range
press [Enter)
setting

[ZEroxxl I Zero range setting

press [ t l [Zeroyy) Press [ tl to modify the zero range, the most zero
6
tracing range will be 3d ;Press [Enter) to confirm
press [Enter) [ Full l onfirm the setting and enter capacity setting
[ Full l

7 Press [ t l [ xxxx l
Press [Enter l [ LoAd)

[LoAd)
[LoAdl
8

L_.