Professional Documents
Culture Documents
Template SOP
Template SOP
: TNT/CB/53
RELEASE OF AUTHORIZATION
This document is issued under the authority of Dr. Sreedevi D, Lab Director. Do not share this document either verbally or by any other means with out
the written consent of the Lab Director.
QUANTITATIVE ESTIMATION OF TOTAL PROTEIN IN HUMAN SERUM ON COBAS 6000
SOP NO. TNT/CB/53
DISTRIBUTION LIST
The following are the details of holders of controlled/uncontrolled of this document.
02 Biochemist Controlled
AMENDMENT SHEET
1.0 PURPOSE AND SCOPE: This procedure is intended for the quantitative estimation of total
protein Cobas 6000. Measurement of total protein is used in the diagnosis and treatment of a
variety of diseases involving the liver, kidney or bone marrow as well as metabolic or
nutritional disorder.
Alkaline
Solution
The color intensity is directly proportional to the protein concentration which can be
determined photo metrically.
4.3 Primary Sample Collection:
4.3.1 Collect blood in a yellow or red top vacutainer and allow it to clot and separate the
serum by centrifugation. Centrifuge samples at 2500 – 3500 rpm for 10 – 15 minutes.
4.3.2 To prevent formation of fibrin in serum samples, complete clot formation should take
Internal quality control is run after each new calibration and after every major
equipment breakdown or maintenance.
QC results are entered into Unity software and verified for acceptability before reporting
patient results. The same results are also entered into IQC Record (TNT/CB/IQCR-RB).
If QC outlier is observed, appropriate corrective and preventive action is taken and
recorded in the IQC Corrective Action Record (TNT/CB/IQCCAR-RB).
Controls are tested in the same manner and by the same personnel testing patient
samples.
Patient samples are processed after review of QC results by the Section Head.
The instrument will be operated only by a trained Technician.
The samples are processed directly by programming on the screen or by a bar code
system.
The sample can be processed while the instrument is in standby or add samples while
the system is in running status or the sample details can be directly downloaded from an
LIS.
Program all the samples and load as per the wok order. Ensure all the lids are closed and
press RUN on the key board. If a single sample is processed at a time select PROCESS
SINGLE SAMPLE from the Enter Sample Data screen.
If bar coded samples are processed the patient and test details will be automatically
downloaded from the LIS.
For more details on sample order and processing refer to the Cobas 6000 operator
manual.
In case of breakdown of the instrument, samples shall be processed on the back-up
Cobas 8000 Pro
4.9 Repeat & Recheck Criteria:
When the results are beyond the reference ranges of the laboratory.
Samples having higher concentration are diluted in 1:3 dilution and rerun function are
automatically multiplied by a factor of 3.
Results less than 0.2 g/dL should be reported as “< 0.2 g/dL.”
Repeat values are recorded in Repeat Tests Record (TNT/CB/RTR).
Repeat testing is performed after informing the Section Head.
4.10Biological Reference Intervals: Normal value: 6.6 – 8.7 g/Dl
Expected values according to Josephson
w Adults: 66-87 g/L (6.6-8.7 g/dL)
Expected values according to Tietz
w Umbilical cord: 48-80 g/L (4.8-8.0 g/dL)
w Premature: 36-60 g/L (3.6-6.0 g/dL)
w Newborn: 46-70 g/L (4.6-7.0 g/dL)
w 1 week: 44-76 g/L (4.4-7.6 g/dL)
w 7 months-1 year: 51-73 g/L (5.1-7.3 g/dL)
w 1-2 years: 56-75 g/L (5.6-7.5 g/dL)
w >3 years: 60-80 g/L (6.0-8.0 g/dL)
w Adults (ambulatory): 64-83 g/L (6.4-8.3 g/dL)
4.11 Critical/Panic Values: < 3.4 & > 9.5 g/dL
4.12 Performance Characteristics:
4.12.1 Uncertainty of Measurement: 2.13 % CV
4.12.2 Accuracy:- NA
4.12.3 Limit of Quantification (LOQ):-0.2 g/dL
4.12.4 Limit of Detection (LOD): 0.2 g/dL
4.12.5 Analytical Measurement Range (AMR): 0.2 – 12.0 g/dL
4.12.6 Clinical Reportable Range (CRR): 0.2 – 12.0 g/dL
4.12.7 Sensitivity: 0.2 g/dL