Terumo Europe N.V. T-ACE II+ Service manual version 1.0
CONTENTS
1. INTRODUCTION............................................................................................ 3
-2-
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
1. INTRODUCTION
The T-ACE II+ Automatic Blood Components separator is intended for the separation of
different blood components that were centrifuged once or twice. The different fractions
(packed red cells, plasma, buffy-coat, platelet concentrate,...) are obtained in an easy,
fast and standardised way.
The T-ACE II+ is designed for use with TERUMO Teruflex®, TERUMO Imuflex®, and
TERUMO Penpol bags with or without additive solution.
Both conventional as top-and-bottom bags can be used.
This manual is written for the T-ACE II+ with the software version 1.x
SYMBOLS:
EMC Group 1 Equipment: ISM (Industrial, Scientific, Medical) equipment in which there is
intentionally generated and/or used conductively coupled radio-frequency energy which
is necessary for the internal functioning of the equipment itself.
-3-
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
If the equipment is not used following the intended use described in this manual, the
manufacturer will not guarantee the results nor the safety of the device and shall not be
liable.
Carefully read the entire content of this manual before using the T-ACE II+.
This manual is for the T-ACE II+ with software version 1.x The version appears on the
display during the self-check of the device after switching it on.
WARNING
Never use the T-ACE II+ on blood bags that are connected to a patient.
CAUTION
• The T-ACE II+ makes use of scales. Be careful when using them. Each hard shock
can damage the scales or may require calibration of the scales.
• Non-properly working scales will negatively influence the quality of components and
processes.
-4-
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
4.1. SPECIFICATIONS
4.2. ACCESSORIES
Standard accessories
- AC power cord
- Allen key
- Bar code reader, support, cable and cd-rom
- Instructions for use - English
- Instructions for use other languages: on cd-rom
- Warranty card
Optional accessories
- Air compressor 230V (ME+05TA08)
110V (ME+05TA72)
- Filter holder (ME-05TA2HF)
- Soft filter holder (ME-05TA2SF)
- Router + T-SOFT software (ME-05TA2P03)
- Wireless router + T-SOFT software (ME-05TA2P04)
- WiFi module (ME+05TA2P02)
-5-
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
After opening the transport box, you will find the following accessories:
Take out all the accessories and take off the upper packaging protection.
To take out the T-ACE II+ itself, please be careful because the machine is very heavy.
Preferably, take the machine out of the box with 2 persons and put it on its location.
Place the device on a flat surface and make sure you can always reach the on/off switch.
WARNING Never take the machine out of the box by pulling the telescopic bag
support.
CAUTION Make sure the T-ACE II+ machine is never exposed to direct sunlight
because this can negatively influence the sensitivity of the B.C. layer
detection.
NOTE When using multiple T-ACE II+ machines, read point 5.2 (Configuration
T-ACE II+ Network) for a correct layout and placement of the T-ACE II+
network.
-6-
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
1. All T-ACE II+ devices must be connected to the same Ethernet network to enable
communication between the T-ACE II+ devices and T-SOFT. This can be done with a
wired or a wireless router.
Typical schematic example layouts of a T-ACE II+ network can be as shown below:
…
Ethernet Air line
port
Wireless router
…
Ethernet Air line
port
-7-
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
1. To work with the machine, the telescopic bag support has to be moved upwards.
Please contact your local Terumo representative to do this and to check proper
functioning of the machine before using with blood products in daily routine.
2. Connect the device to the air supply (connect an outside air tube to the air inlet on
the back of the T-ACE II+). For connection of the T-ACE II+ to an air compressor,
carefully read the instructions in point 5.5 (working with an air compressor).
3. For connection of the T-ACE II+ to a PC, carefully read the instructions in point 5.6
(working with computer connection).
4. Plug the bar code reader cable in the RS-232 communication port on the rear of the
T-ACE II+.
5. Connect the system to the power supply. (Connect the AC power cable to the AC
inlet on the back of the T-ACE II+ and then to the earthed AC socket).
6. Switch on the machine for the calibration of the scales and read carefully the
instructions in point 5.7 (Calibration of the scales).
7. Open the door and place the bags on the T-ACE II+ device.
Depending on the bag configuration and application, different ways exist to place the
bags. The correct bag positioning for the selected program will be indicated on the
graphical display.
• Always make sure that the tubing is correctly placed inside the clamps and in
the flow regulator!
This is very important because the clamps will guide the flow of liquid to the
appropriate bags. If tubing is not correctly placed, it is understandable that the
machine will not operate in a proper way.
These clamps can also be used as sealing heads to seal the tubing at the end
of the separation so that manual clamping and making seals after the separation
is not longer necessary.
• The bags have to be placed on the hooks of the plasma and/or SAGM scale.
These hooks are connected with scales and will weigh the amount of grams
entering or leaving the bags.
• The primary bag has to be placed with the label side visible (in front position)!
So also no label may be adhered on the backside of the primary bag!
-8-
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
version 1.x
2. Make sure no bags or weights are placed on the scales and close the door.
The T-ACE II+ machine will check the status of its internal components and if
necessary, recalibrate the scales and/or the flow regulator detection. The press
position and the LED voltages are controlled too but not calibrated. If these are out
of specification, an error message will be displayed on the screen. During this self-
check the operation indicator lights up. If some components are not functioning in
between its specifications and cannot be automatically adjusted, an error message
will be displayed.
An extensive list of errors is given at the end of this manual (section Alarm & error
messages). The self-check can be stopped by pressing ‘STOP’ or can be continued
by pressing ‘START’.
After completion of the self-check, the door will open again and the ‘start’ menu will
be shown:
program 01 CONV
place the bags and ‘push start’ or close the door
-9-
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
3. Choose the appropriate program by pressing “menu”. The program “selection menu”
will then be displayed:
SELECTION MENU
program 1 CONV
Use the “scroll” key to go to the correct program number. After pressing “stop” the
program will be selected or after pressing “menu” and entering the password, the
parameter settings of the selected program will be displayed and can be changed if
wanted. Press “stop” to leave the parameter settings menu.
4. Place the bags, close the door, break the cliktips of the blood bag(s) and press
“start” to start the separation process. During the entire process, the operation
indicator is illuminated.
5. When the separation is completed, the operation indicator starts blinking. After
removing the bags, the T-ACE II+ goes back to its initial status and the start menu
is again shown on the display.
NOTE After 10 minutes, the backlight of the screen goes out. Push “stop” to
return to normal operation mode.
CAUTION Do not touch the bags during operation as the T-ACE II+ is
continuously checking the weight of the components.
- 10 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
1. Connect the air compressor with outside air tube to the air inlet on the backside of
the T-ACE II+.
3. If the external Sil-Air low-noise compressor, which can be optionally bought, is used
for the first time please fill the compressor first with 0,5 l of the supplied oil before
switching on. Keep the air outlet valve of the compressor closed until enough
pressure is build up. If not enough pressure is delivered; it is always possible to
adjust with the air regulation valve.
4. At least 4 bar is required for each machine, more pressure is also all right because
the T-ACE II+ has a safety valve inside which cuts the air pressure on 8 bar.
Air outlet
- 11 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
1. All T-ACE II+ devices must be connected to the same Ethernet network to enable
communication between the T-ACE II+ devices and T-SOFT. This can be done with
a wired or a wireless router.
When using a wired router: connect the router with the RJ-45 inlet of the T-ACE II+
via a network cable. Connect the router also with the PC via another network cable.
When using a wireless router: Connect the router with the PC via a network cable.
Place a WiFi module on each T-ACE II+ device that has to be connected to the
Ethernet network. (Contact your local Terumo Representative to install the WiFi
option)
2. The T-ACE II+ has to be activated and configured for PC communication, because
PC communication is in default not activated.
1) Switch on the machine. After the self-check, the start menu appears.
2) In the start menu, press 2 x “MENU” and as standard password 5 x “VALUE” and
1 x “STOP”.
3) Then browse to the menu ‘communication settings’ by using the “menu” key.
COMMUNICATION SETTINGS
PC transmission yes
T-ACE II+ 99
ID bag yes
ID operator auto
data 1 no
data 2 no
data 3 no
Select parameter ‘PC transmission’ by using the “scroll” key and change its
value from “no” to “yes” with the “value” key.
5) After switching ‘ON’ again and passing the self-check, the following menu will
appear.
- 12 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
version 1.x
initialising network
At this time, the T-ACE II+ is making connection with the computer and T-SOFT.
T-SOFT has to run and all the T-ACE II+ devices need to be connected to the
Ethernet network.
If the message “T-ACE II+ not installed / install T-ACE II+ is shown, communication
with T-SOFT was successful.
initialising network
T-ACE II not installed / install T-ACE II+
+
initialising network
install network start: yes stop: no
initialising network
install completed machine number: 01
Automatically, the T-ACE II+ will receive an ID number from the computer. This
number will be given chronically and cannot be changed manually.
If more machines really need to be connected to T-SOFT, a 2nd network has to be created.
If a machine was already installed previously, the following message will appear:
T-SOFT registers the machines based on the MAC address of the Ethernet module, so if the
machine was installed once, T-SOFT will always recognize this machine and it will always keep
the same T-ACE II+ ID number.
- 13 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
SCALE CALIBRATION
remove all the bags and press ‘start’
SCALE CALIBRATION
put 200 g on the PLASMA scale and press ‘start’
After hanging a calibrated 200 g weight on the plasma (upper left) scale, press “start”.
The measured value will be displayed; these values should both be equal to 200.
SCALE CALIBRATION
measured PLASMA value: 200 g ; press ‘start’
NOTE • If the measured value is not equal to 200, please contact your local
Terumo service organization.
- 14 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
Press “start” and repeat the on-screen instructions for hanging 200 g on the S.A.G.M.
(upper right) scale and B.C. (bottom right) scale.
SCALE CALIBRATION
put 200 g on the SAGM scale and press ‘start’
SCALE CALIBRATION
measured SAGM value: 200 g ; press ‘start’
SCALE CALIBRATION
put 200 g on the BC scale and press ‘start’
SCALE CALIBRATION
measured BC value: 200 g ; press ‘start’
SCALE CALIBRATION
remove the bags and press ‘start’
- 15 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
After calibration, the menu to change the calibration date is shown automatically.
This date can only be changed after a scale calibration.
The item selected (day, month or year) is indicated in capital letters. To switch from
item, press “scroll” ; to change the value, press “+” or “-“.
Press “STOP” to leave calibration mode and to return to normal operation mode.
- 16 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
6. SERVICING MENU
The T-ACE has a special service menu in which additional information for the technical
servicing can be found.
program 01 CONV
place the bags and ‘push start’ or close the door
SELECTION MENU
Program 1 CONV
- 17 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
PASSWORD: **********
After putting in the correct password the different parameter menus will be shown by
simply pressing "Menu"
PARAMETERS 1
PARAMETERS 2 Dedicated parameter menus related to the protocol
PARAMETERS 3
PARAMETERS 4
---------------------------------------------------------------------------------------------------
UPLOAD PARAMETERS
OPTIONS
COMMUNICATION SETTINGS
SEALING OPTIONS
MANUAL MODE Common parameters menus, independent of protocol
GENERAL SETTINGS
ANALOG INPUTS
NUMERIC POTENTIOMETERS
P.S. For detailed explanation of the other menus, see the Users manuals of the T-ACE
- 18 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
UPLOAD PARAMETERS
Mac address 00:40:9D 2A:CB:C0
Upload Free/waiting
Export Free/waiting
Mac address
This is the T-ACE II+ address on the network. Each T-ACE II+ has a different address.
Firmware version
This is the T-ACE II+ software version
Upload
This parameter gives the possibility to the T-ACE II+ to be programmed from a computer.
This can be used to program several T-ACE II+ with the same parameters by use of a
computer.
Waiting: waiting for the parameters
Free: normal use
Upload: communication from the computer to the T-ACE II+.
Export
This parameter gives the possibility to the T-ACE II+ to send parameters to the computer
to be sent to another T-ACE II+.
Free: normal use
Upload: communication from the T-ACE II+ to the computer.
ANALOG INPUTS
bc/rbc scale weight 001
- 19 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
NUMERIC POTENTIOMETERS
bc/rbc scale
Indication of the bc/rbc scale numeric potentiometer. If necessary, it can be
adjusted by pressing “VALUE”. Depending if it has to be increased or decreased,
the “SHIFT” button has to be pressed.
After a readjustment of the scale potentiometers, a software calibration needs
to be done. See chapter 5.7
plasma scale
Indication of the plasma scale numeric potentiometer. If necessary, it can be
adjusted by pressing “VALUE”. Depending if it has to be increased or decreased,
the “SHIFT” button has to be pressed.
After a readjustment of the scale potentiometers, a software calibration needs
to be done. See chapter 5.7
- 20 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
- 21 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
7. TROUBLESHOOTING
# Problem Cause/remedy Reference
1 No first detection Check the parameters Instruction for use
Check the flow regulator Problem # 3
Check the clamps Problem # 5
Adjust the press detection board
voltage Press detection board adjustment
Replace the pressdetection board Press detection board replacement
LED press detectors ribbon cable
defect Replace the ribbon cable (ME+05TA61)
2 No second detection Check the parameters Instruction for use
Check the flow regulator Problem # 3
Check the clamps Problem # 5
Adjust the press detection board
voltage Press detection board adjustment
Replace the pressdetection board Press detection board replacement
LED press detectors ribbon cable
defect Replace the ribbon cable (ME+05TA61)
3 No flow reduction Check the parameters Instruction for use
Check the flow regulator manually "MANUAL" menu
Check the detection Problem # 1 and # 2
Adjust the flow regulator Flow regulator - adjustment
Replace the flow regulator Flow regulator - replacement
Always Plasma
4 alarm Check the parameters Instruction for use
Analog inputs and/or numeric
Always SAGM alarm Check scale calibration potentiometers
Scales not ok Replace scale Scales - replacement
5 Sealing head / clamp Check the clamp manually "MANUAL" menu
not working Check the air pressure Air pressure - adjustment
Check the pneumatic valve block Pneumatic valve block
Check the HF generator connections HF generator- replacement
Clean the clamp Sealing head - cleaning/maintenance
Replace the clamp Sealing head - replacement
6 No air pressure Check the air pressure Check the manometer
Check compressor and connections
Adjust air pressure
Replace air pressure
7 Pneumatic valve Check the air pressure Check manometer
not working Check pneumatic valve Pneumatic valve block
8 Display not working Check AC power supply Problem # 9
Clean display Display - cleaning/maintenance
Check display connector Display - replacement
Replace display Display - replacement
9 No power Check AC plug Power entry components
Check fuses Power entry components
Check the power supply Power supply
10 Press does not Check the press manually "MANUAL" menu
close/open Check air pressure Problem # 6
Check pneumatic valve Problem # 7
Adjust the press speed Press piston - adjustment
Replace the press piston Press piston - replacement
11 Door does not close Check the locking mechanism Door locking detection - adjustment
Check the door locking detection Door locking detection - adjustment
- 22 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
12 Door does not lock Lock and open manually "MANUAL" menu
Check air pressure Problem # 6
Check pneumatic valve Problem # 7
Check the door locking detection Door locking detection - adjustment
Adjust door locking piston Door locking detection - replacement
Replace door locking piston Door locking detection - replacement
13 Door does not open Check the locking mechanism Door
Adjust the door Door - adjustment
Replace the door Door - replacement
14 Keypad not working Check connections Keypad - replacement
15 Language not ok Change the language Instruction for use
16 Other problems Check actions in manual mode "MANUAL" menu
Check main board connections Main board - replacement
Change main board Main board - replacement
17 No communication Check T-SOFT program T-SOFT User's Guide
Check all network connections Configuration network
Check connections inside the T-ACE
- 23 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
8. ASSEMBLIES/PARTS
CLEANING/MAINTENANCE
During preventive maintenance, remove the condens and oil residue in the container.
ADJUSTMENT
● remove the back plate
● lift up the wheel on the regulator
● adjust the air pressure to 5 bar by turning the wheel and checking the manometer
● lock the setting by pushing in the wheel
regulator wheel
4 fixing screws
REPLACEMENT
Necessary equipment:
2,5 mm Allen Key
- 24 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
8.2 DISPLAY
This is a 5 inch colour screen. The screen light will switch off after 10 minutes in
stand by mode. Press “Stop” key to switch it on again.
The screen assembly is composed by the screen itself and his driver board.
The screen is connected to the CPU Board by use of a flat cable.
Screen
Driver board
CLEANING/MAINTENANCE
Regularly check that the display is free of dirt and clean when necessary.
ADJUSTMENT
No contrast adjustment possible
REPLACEMENT
Necessary equipment:
● 2 mm Allen key
- 25 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
● fit another display, fix it with the 2 Allen screws. Be careful to centre it.
● fix the ribbon cable to the main board
● close the back plate
2 mm Allen
screws.
- 26 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
8.3 DOOR
CLEANING/MAINTENANCE
Ensure that the door is clean, especially the white part in the door, which is needed for
the detection of the layer.
Do not use any alcohols or abrasive cleaners because they can damage the door. Please
refer to the cleaning section at the end of this manual.
ADJUSTMENT
Necessary equipment:
3,0 ; 4,0 mm Allen Key
If the closing or the locking of the door is not functioning as it should, adjust the position
of the door as follows:
● remove the back plate, remove the HF generator as instructed in the appropriate
section
● loosen the 4 fixation screws of the door hinge on the inside of the machine
● go to the “manual” menu and close the door
MANUAL MODE
clamp 1 off clamp 2 off
clamp 5 off
seal
press off
door close
● fix the 4 fixation screws again when the position is good. When you fix the door
back into the chassis, take care of its positioning. It may not touch the frame
during opening and closing. Use small insert to centre the door.
- 27 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
Be careful of the door closing sensor position to be sure it will not be damaged by
the door.
● replace the HF generator as instructed in the appropriate section
● if the door will not close automatically, check paragraph “door locking mechanism”
REPLACEMENT
Necessary equipment:
3,0 ; 4,0 mm Allen Key
- 28 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
CLEANING/MAINTENANCE
No cleaning procedures are necessary for this item.
ADJUSTMENT
Necessary equipment:
● small “-” screw driver
● 4 mm Allen Key
If the door does not close or if the “locking alarm” always appears, check the position of
the door closing detection. The inductive door detection and the closed door may not
touch each other. The screw that fixes the detector housing can be loosened and the
detector can be repositioned.
On the inside of the machine, the correct operation of the door closing detection can be
checked with the LED indicator on the detector. The LED indicator will be illuminated
when the detection is working.
door closing
detection
If the door closes either too slowly or too violently, the air pressure can be adjusted on
the locking piston.
4 screws attaching
door piston to case
- 29 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
REPLACEMENT
Necessary equipment:
● 3,0; 4,0 mm Allen Key
block fixed
to the case
piston
screw fixing
piston to locking bars
screws fixing
block to piston
1 of the 2 door
locking bars
● unscrew the 2 screws fixing the piston to the block fixed to the case
● unscrew the screw fixing the piston to the locking bars assembly
● replace the piston
● perform all the actions of removing the piston in the reverse order
● adjust opening and closiong speed as specified in adjustment
● reposition the back plate
door closing
detection
fixation nut
- 30 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
CLEANING/MAINTENANCE
No cleaning procedures are necessary for this item.
REPLACEMENT
Necessary equipment:
● 2.5 and 4 mm Allen key
- 31 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
● When you fix the door back into the chassis, take care of its positioning. It may
not touch the frame during opening and closing. Use small insert to centre the
door.
- 32 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
CLEANING/MAINTENANCE
Regularly check that there is no dirt in the flow regulator and clean if necessary.
ADJUSTMENT
● go to the menu "PARAMETERS 2" and set ‘flow regulator closed’ to 0 and ‘flow
regulator open’ to 50
PARAMETERS 2
sagm 100 g
% sagm 0%
add type 2
air/plasma transfer 3g
security detection 20 %
MANUAL MODE
clamp 1 off clamp 2 off
clamp 5 off
seal
press off
door close
- 33 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
REPLACEMENT
Necessary equipment:
● 2,0 ; 2,5 ; 4,0 mm Allen key
● soldering iron; solder; shrinking tube
screws attaching
motor to the case
guide
- 34 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
PARAMETERS 2
sagm 100 g
% sagm 0%
add type 2
air/plasma transfer 3g
security detection 20 %
MANUAL MODE
clamp 1 off clamp 2 off
clamp 5 off
seal
press off
door close
▫ make sure that the shaft of the motor is at the stop, high up in the head
opening
▫ repeat the adjustments if necessary
- 35 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
CLEANING/MAINTENANCE
Regularly check that the inside of the regulator unit is free of dirt.
ADJUSTMENT
Normally, no adjustment is necessary: the numeric potentiometers will calibrate it selves.
If during the self-check, the T-ACE II+ is giving an error message it is possible to
manually adjust the numeric potentiometers.
In the ANALOG INPUTS menu it is possible to see the voltage. This voltage should be
1.30 ± 0.10 V (1.20 ≈ 1.40 V) without black paper in between (full light) and 5.00 V with
the black paper in between (no light). The 5 V can not be adjusted, it must be 5 V or the
detector is broken.
ANALOG INPUTS
bc/rbc scale weight 001
If the value is out of range and the T-ACE II+ gives an error message during self check, it
is possible to adjust the voltage in the NUMERIC POTENTIOMETERS menu.
With the “SCROLL” highlight the “flow regulator detection” line and with the “Value” you
can increase or decrease the voltage. If the “Shift” is on you will decrease and if it is off
you will increase the voltage.
NUMERIC POTENTIOMETERS
- 36 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
REPLACEMENT
Necessary equipment:
2,5 mm Allen key
soldering iron
If the value displayed on the “ANALOG INPUTS” menu is not in line with those described
in the adjustment section and the adjustment of the numeric potentiometers does not
cause any variation, check whether the leads are properly connected to the terminal on
the Interface Board. If properly connected, change the detector:
● unscrew the 2 screws of the detector at the top of the flow regulator
● remove the detector
● position a new detector, with the white point downwards. Be sure not to forget to
place the spacers
● replace the whole unit
● reconnect the flow regulator sensor connector to the Interface Board
You can solder the detector, use crimping tube to protect the wires.
After replacement, switch off the T-ACE II+ and let it make the automatic calibration. It is
possible that the self-check will take longer time because the new sensor has to be
calibrated.
The Flow regulator detector has a better tube holder, the tube will not come easily out of
the holder.
- 37 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
8.8 HF GENERATOR
CLEANING/MAINTENANCE
No cleaning necessary
ADJUSTMENT
No adjustment necessary
REPLACEMENT
Necessary equipment:
● 3,0 mm Allen key
Fibre connections
sealing head 4
sealing head 3
sealing head 2
sealing head 1
sealing head 5
110 – 230 V
AC power inlet
24 V DC power outlet
- 38 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
If the sealing heads do not work (either one or all of them) make sure:
● that the correct LSB, MSB and/or 5th LED’s are illuminated when an address is
selected
● that the busy signal is illuminated on the HF generator and going back to the main
board when the HF generator is sealing. So when the ‘busy’ cable is unplugged on
the main board side, a red light should be seen when a sealing head should be
activated.
If this is not the case, check the fibre optic connections and the direction.
If all of this is ok, change the HF generator. Proceed as follows:
- 39 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
CLEANING/MAINTENANCE
No cleaning/maintenance necessary
ADJUSTMENT
No adjustment necessary
REPLACEMENT
This board should not be replaced in normal conditions.
Connections should be made the following way:
This board is used to centralise all connections coming from the power supplies, scales,
detection board, etc.
Optical fibers
To generator
To CPU
To barcode wires
Board
To BC scale
To keyboard
buttons
To com. Wires
To keyboard
LEDS To Plasma scale
To inductive To FR motor
door closing switch ini
detector
To indication
Board
Power supply 5 V.
- 40 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
8.10 KEYPAD
The keyboard is connected to the Interface Board by 2 electrical ribbon cables. One for
the LED and one for the buttons.
CLEANING/MAINTENANCE
No cleaning/maintenance necessary
ADJUSTMENT
No adjustment necessary
REPLACEMENT
- 41 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
8.11 MAINBOARD
This is a SMT technology CPU Board. This board is connected to the Interface Board by
use of one unique connector and fixed to the chassis by a screw.
One connector is used to load or to upgrade the software by use of a special
communication box.
Display connector
CLEANING/MAINTENANCE
No cleaning/maintenance necessary
ADJUSTMENT
No adjustment necessary
REPLACEMENT
- 42 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
CLEANING/MAINTENANCE
No action required.
ADJUSTMENT
No adjustment necessary
REPLACEMENT
Necessary equipment:
● 3,0 mm Allen key
If this is not the case, disconnect the SUBD 25 pin connector from the pneumatic valve
block, remove it and use a multi-meter to measure the change of state when the
connector pins are activated.
Pin 1: Clamp 4 Pin 5: Door lock
Pin 2: Clamp 3 Pin 6: Press
Pin 3: Clamp 2 Pin 7: Clamp 5
Pin 4: Clamp 1
The voltage is 24V at rest, and it falls when the pneumatic valve is activated. If the
voltage does not change, replace the motherboard. Otherwise change the pneumatic
valve unit.
● remove the back plate
● unplug the air connections
● unscrew the fixings on the pneumatic valve and remove it
● fit a new pneumatic valve block and reassemble the unit
● reconnect the air tubes according to the following picture
silencer
air regulator
input
clamp 1 clamp 5
clamp 2 clamp 4
clamp 3
- 43 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
CLEANING/MAINTENANCE
No cleaning/maintenance necessary
ADJUSTMENT
No adjustment necessary
REPLACEMENT
No replacement necessary in normal conditions, only reconnection is likely.
AC cable (N)
Power switch
AC cable input
AC (GND) to case
AC (GND) outlet to supply
AC (L) to comm. board
AC (L) outlet to supply
AC (N) outlet to supply
AC (N) to comm. board
AC (N) to HF gen.
AC (N) inlet
AC (L) to HF gen.
AC (L) inlet
AC (GND) to HF gen.
AC (GND) inlet
- 44 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
CLEANING/MAINTENANCE
No action required.
ADJUSTMENT
Necessary equipment:
● 3,0 mm Allen Key
● screw driver
● multimeter
Proceed as follows:
● remove the back plate
● connect a multi-meter to the black and white wires of the connector
● adjust the voltage on the power supply to 5,40 V by turning the potentiometer
5 V adjustment screw
black & red wire
5 V output
REPLACEMENT
Necessary equipment:
● 3,0 mm Allen Key
● small ‘-‘ screw driver
Proceed as follows:
● remove the back plate
● unscrew the AC input wires on the front side of the supply
● unscrew the DC output connector on the backside of the power supply
DC output
AC input connector
connections
● unscrew the power supply fixation screws from the chassis and remove the card
● put the new card in position and screw it to the chassis
● reconnect the power connections
- 45 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
CLEANING/MAINTENANCE
The Plexiglas plate in front of the detectors and the door should be cleaned regularly, to
ensure that the surface is completely free of dirt.
ADJUSTMENT
If during the self check, the T-ACE II+ is giving an error message it is possible to
manually adjust the numeric potentiometers.
The voltage of the selected LED is indicated after “press detection LED x” (x = 1 to 10)
Press “SCROLL” to change the number of the LED ‘x’
Scroll through the LED’s and select the lowest value. The lowest LED value should be
above 2.60 V.
ANALOG INPUTS
bc/rbc scale weight 001
NUMERIC POTENTIOMETERS
- 46 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
With the “SCROLL” highlight the “press detection LED 2” line and with the “Value” you
can increase or decrease the voltage. If the “Shift” is on you will decrease and if it is off
you will increase the voltage.
Increase or decrease and check in the “ANALOG INPUTS” the voltage modification.
ANALOG INPUTS
bc/rbc scale weight 001
P.S. The higher the difference between light and dark, the more sensitive the detection
will be.
- 47 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
REPLACEMENT
Necessary equipment:
● 2,0 mm Allen key
● 2,5 mm Allen Key
● 5,5 mm spanner
If the LED voltage can not be adjusted and the T-ACE is giving an error message during
selfcheck check that:
● the ribbon cable is firmly connected both on Interface and press detection board
● the ribbon cable is not damaged.
4 protection
plate screws
● dis-attach the press plate from the press piston by lifting the clip and pulling out
the pin
clip
pin
flat cable
● lift the press up and remove the PCB from the press by unscrewing the 4 screws
while holding the nuts with the 5,5 mm spanner
- 48 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
press detection
board screws
● reposition a new PCB board with washers, screws and nuts in reverse disassembly
order
plastic washer
nut
screw spacer
- 49 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
CLEANING/MAINTENANCE
No action required
ADJUSTMENT
Necessary equipment:
● short “-” screw driver
To adjust opening and closing of the press, adjust the piston as follows:
● remove the back plate
● go to the "MANUAL" menu
MANUAL MODE
clamp 1 off clamp 2 off
clamp 5 off
seal
press on/off
door close
- 50 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
adjustment press
opening
adjustment press
closing
REPLACEMENT
Necessary equipment:
● 3 mm Allen Key
● 10 mm spanner
clip
clip
pin
pin
- 51 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
● remove the plate by unscrewing the connection to the door on the plunger
● dis-attach the press position detection from the piston by unscrewing the 4 screws
● perform all the actions of removing the piston in the reverse order
● adjust the press opening and closing speed as described in the adjustment section
● reposition the back plate
- 52 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
CLEANING/MAINTENANCE
No action required
ADJUSTMENT
During the self-check the voltage limits are checked. If they are out of specification, an
error message will come on the screen. See here under how to adjust.
In the “ANALOG INPUTS” menu you can check the voltage of the detector under the line
“press position detection”.
ANALOG INPUTS
bc/rbc scale weight 001
- 53 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
REPLACEMENT
Necessary equipment:
● pincers
● spanner 10 mm
● Allen key 2,5; 5,0 mm
If it is not possible to adjust to the right voltage or if you move the press forwards and
backwards without voltage modification, replace the press position detection as follows:
clip
clip
pin
pin
plunger
shrinkage ring
● unscrew the connection piece to the press plate by fixing the plunger with a
spanner
● remove the plate which holds the press position detection plunger
● loosen the setscrew and pull out the press position detection
● replace the press position detection and perform the above actions in reverse
order
● please check after replacement whether the analog value in “ ANALOG INPUTS”
will change when the press is moved. The voltage with a closed door and press
should be 4.40 ± 0.05 V.
- 54 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
8.18 SCALES
CLEANING/MAINTENANCE
No maintenance necessary
ADJUSTMENT
Necessary equipment:
● calibrated 200 gr
Normally there is no adjustment necessary: both scales are calibrating themselves during
the self-check of the T-ACE. If during the self-check an error message appears, it is
possible to adjust the scales by use of the numerical potentiometers.
In the “ANALOG INPUTS” menu you can see both scales weight in gr.
ANALOG INPUTS
bc/rbc scale weight 001
If the weight is wrong, the scale voltage can be adjusted by use of the numeric
potentiometer:
- 55 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
NUMERIC POTENTIOMETERS
REPLACEMENT
Necessary equipment:
multimeter
Allen Key (1,5 mm; 2,5 mm; 3,0 mm; 4,0 mm)
spanner 7 mm
Changing the scales is not necessary unless the leads have come loose from the strain
gauge or is visually damaged. To test the functioning of the scales, check the impedance
between the 4 leads and the scales, using a resistance (Ω) meter:
When placing a new strain gauge, take great care during the entire operation not to pull
the wires. If the wires on the strain gauge come loose even slightly, irreversible damage
will be caused and the scale can’t be used anymore!
a) Plasma scale
● take away the back plate of the scale support by unscrewing the 4 screws with a
3,0 mm Allen key and taking off the plate
- 56 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
● unplug the operation indicator connector and remove the board by unscrewing the
2 screws
Operator indicator connector
Amplifier board
- 57 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
Bag support
Strain gauge
Support
Stop nut
b) B.C. scale
● take away the bag stop by unscrewing the 2 screws and removing it
Bag stop
Screws
- 58 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
bag hanger
strain gauge
support
● unscrew the 2 screws fixing the bag hanger to the strain gauge, with 1 of them,
first remove the stop nut
● unscrew the 2 screws fixing the strain gauge to the support
● remove the strain gauge and replace with another one
● perform the dismantling actions in the reverse way and replace the stop nut.
● Make sure that the stop nut is placed approximately 1 mm from the holder to
allow enough scale movement.
- 59 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
- 60 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
CLEANING/MAINTENANCE
Be sure that the sealing heads are clean and dry. When blood or other dirt is inside the
heads clean them before reusing.
Cleaning can be done by unscrewing the two Allen screws on the top of the head and
removing the top. Remove the dirt and make sure that the sealing heads are dry.
Close the heads again after cleaning.
ADJUSTMENT
No adjustment necessary.
REPLACEMENT
Necessary equipment:
● 2,5; 3,0 mm Allen Keys
If a welding head does not shut properly, or if it does not weld, make sure:
● that there is no dirt in the clamp
● that the air supply is correct
● that the BNC cable is connected to the generator
● that the generator is working correctly
2 screws attaching
the top cover to the clamp
4 screws attaching
the clamp to the case
P.S. It is possible to place a sealing head in another direction if needed. There are 4
directions possible.
- 61 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
● The back panel ground cable. ● The Plasma scale flat cable
● The barcode reader connector cable ● The screen to CPU Board flat cable
- 62 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
WARNING Before cleaning the unit, switch off the power and disconnect the power
cord.
General cleaning
Always keep the T-ACE II+ clean.
Clean the surface of the T-ACE II+ with soft clots saturated with mild detergent or
a chlorhexidine gluconate solution of less than 0.5%.
When cleaning the T-ACE II+, be careful not to spill any cleaning solutions inside
the instrument. If this occurs, do not operate the instrument.
Professional cleaning and service is required if any liquid (blood, blood product,
cleaning solution or other fluid) is spilled into the internal components of the
machine.
(Preventive) maintenance
− To assure safe, trouble free operation, please have your T-ACE II+ serviced at
least every 12 months, depending on the frequency of use.
− If you experience a system failure, please call Terumo or your local Terumo
representative for servicing. When calling for servicing, please be able to
describe the problem and have the serial number available so that faster
handling of the repair will be possible.
− Do not attempt to dismantle or repair the unit. Only qualified personnel are
authorized to service the machine.
- 63 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
If you have a technical question or you need support, please feel free to call one of the
toll free numbers in the list below or send a fax to +32 16 381420.
- 64 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
- 65 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
For other settings, please refer to the bar code reader manual.
- 66 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
- 67 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
Code Description
- 68 -
Terumo Europe N.V. T-ACE II+ Service manual version 1.0
- 69 -