UV aging test chamber
• Overview: • Standard:
• UV weathering chambers simulate the hazards caused • Popular: 、、ASTM G151、
by sunlight, rain, and dew. ASTM G154、 JIS D 0205、 SAE J2020
• UV lamps are used to simulate the effects of sunlight, • Coating:ASTM D3794、 ASTM D4587、 FED-
and condensed moisture is used to simulate rain and STD-141B、 GM 9125P、 JIS K 5600-7-8、 ISO
dew. 11507、 ISO 20340、 M598-1990、 NACE TM-
• The material under test is placed in a cyclic 01-84、 NISSAN M0007、 PrEN 927-6
programme of alternating light and moisture at a • Adhesive:ASTM C1184、 ASTM C1442、 ASTM
certain temperature. D904、 ASTM D5215、 UNE 104-281-88
• UV can reproduce the hazards of months to years • Plastic:ANSI C57.12.28、 ANSI, A14.5、 ASTM
outdoors in a matter of days or weeks. D4329、 ASTM D4674、 ASTM D5208、 ASTM
• Types of hazards include: fading, discolouration, loss of D6662、 DIN 53 384、 GB_T 16422.3-2014/ISO
light, chalking, cracking, clouding, bubbling, brittleness, 4892-3、 JIS K 3750、 UNE 53.104
loss of strength and oxidation. • Roofing materials:ANSI/RMA IPR-1-1990、
• The test data can help customers select new materials, ASTM D4799、 ASTM D4811、 ASTM D3105、
as well as evaluate how changes in formulation affect ASTM D4434
the durability of new products.
1. UV Aging: Fluorescent UV lamps are used to simulate the aging of materials by sunlight. Although ultraviolet (UV) energy accounts for only 5 per cent
of the total energy of sunlight, it is a major factor in the deterioration of outdoor products, so it is possible to test the degradation and ageing of polymer
materials by UV light alone.
2.1 range simulates the sun's UV rays very well, so the UVA-340 test correlates well with outdoor
exposure.
2.2 UV light that is stronger than that of normal sunlight on the earth's surface, and it maximises the speed of the
test. Customers should choose the appropriate lamp according to the different conditions of use of the product.
The choice
of lamp
Domestic
Huaqiang lamp Q-LAB Altas
Studies have shown that the main contributor to this outdoor humidity is dew, not rain.
UV accelerated weathering chambers simulate outdoor moisture erosion by means of a condensation function. In the condensation cycle,
water vapour is generated by heating the water tank at the bottom of the tester. The hot water vapour keeps the chamber at 100% relative
humidity at higher temperatures.
H O T ! : How many hours of testing in a UV aging chamber is equivalent to one year of outdoor
exposure for a test sample?
Theoretically, it's hard to really answer this question. Even State-of-the-art equipment is also hard to come by
Why? Too much factors as following:
1. Geographical latitude of outdoor exposure field (the closer to the equator, the stronger the UV)
2. Altitude (Higher altitude means higher UV)
3. Local geography, such as wind speed, can affect how dry the test sample is or proximity to water sources can promote dew formation.
4. Because the climate varies from year to year, the influence of the weather in the same area on the sample in the next year may be twice that of the
previous year.
5.Seasonal effects, such as exposure to sunlight in winter, are only one-seventh as damaging as in summer.
6.Orientation of the sample (5 degrees south is very different from true north).
7.Whether the sample is insulated or not (outdoor samples on an insulating carrier usually age 50% faster than those not on an insulating carrier)
8.Test cycle of the aging chamber (irradiation time and moisture time)。
9.Operating temperature of the aging box (the higher the temperature, the faster the aging)
10.Testing of special materials。
11.Spectral Distribution of Laboratory Light Sources (SPD)
So in short, weather data is relative data.
However, you can still get excellent burn-in data from artificial accelerated burn-in tests. But you must recognize that this data is relative data, not
absolute data.
If the customer must need data, then 1 day in Chamber(60°C, 0.68w/㎡ 340 nm) eaqul 27 days outside
2. Key spec we need to check for proposal
Tilting tower type can only do standard samples. Box type UV can do non-standard samples.
1 . Determine sample size and
shape with customer If the sample size is non-standard, determine if the conventional box type meets the space
requirements;
I r ra d i a t i o n e n e r g y ra n g e .
3.Lamps type N o r m a l i m p o r t e d l a m p s 0 . 4 - 0 . 9 W / M 2 . ( Q - l a b , At l a s )
Domestic lamp 0.3-0.7W/M2
U VA - 3 4 0 a n d U V B - 3 1 3 a r e c o m m o n l y u s e d f o r s e l e c t i o n ;
4、Determine test light source U V - A 3 4 0 l a m p s e m i t s p e c t ra l e n e r g y m a i n l y c o n c e n t ra t e d i n t h e v i c i n i t y o f 3 4 0 n m w a v e l e n g t h .
with customer U V - B 3 1 3 l a m p l i g h t - e m i t t i n g s p e c t ra l e n e r g y i s m a i n l y c o n c e n t ra t e d i n t h e 3 1 3 n m w a v e l e n g t h o f
the neighbourhood
斜面式与箱体式差异
VS
Tower type UV Box type UV
Inner size
D×W×H:450×1140×500:mm D×W×H :650×1140×500:mm
75×150*2(mm) W×H
Sample form Within 1100*600*500(mm)
size
Pre condition: Our machine can meet their test requirement (In the range 0.65~1.0W/m2)
You pay more money for cameras with lots of features, but most of them you can't use. This would be wasteful on the cost.
• Irradiance meter: Measurement of yellow and green lines, DC 0.3-0.5V, no voltage, proof of irradiance meter failure
• Spray pump: In spray mode, measure whether there is 220V voltage, there is a voltage pump does not run, prove that the pump is bad, no
voltage, then check the relay.
• Dimming module: When all the lights are off, check whether there is voltage at the control input end of the module, DC0-10V. If there is voltage, check
whether the AC input end is 220V and the output end is DC200-400V. If there is input but no output, the module is bad. If there is no input or output, check the
AC contactor
• Ballast: Symptom, one or two lamps are not bright, unplug the light tube plug, plug the light tube wire into the light
tube, such as the lamp can be lit, it proves that the lamp ballast fault