Pf9800 Series
Pf9800 Series
V 3.01
Foreword
Thank you for purchasing the EVERFINE PF9800 series of digital power meter. This
manual contains useful information regarding the functions and operating procedures
of the instrument as well as precautions that should be observed during use. To ensure
proper use of the instrument, please read this manual thoroughly before operating it.
Keep the manual in a safe place for quick reference whenever a question arises.
Notes:
• The contents of this manual are subject to change without prior notice.
• Every effort has been made in preparation of this manual to ensure the accuracy
of its contents. However, should you have any question or find any errors, please
department of EVERFINE.
"Ensure the quality, insist on continuous improvement and make every customer more
satisfied" is the quality policy of EVERFINE. Therefore, the quality of products and
services provided by EVERFINE should be better than those have been promised. If you
have further suggestions or advices on our products and services, please provide your
Copyright Statement
The copyright of this manual and the related information belongs to EVERFINE,
and it is protected by the copyright law of Peoples Republic of China and other
assume law responsibility and all loss of EVERFINE. Any infringement related
above can be traced back to the responsible user by the unique product number
If EVERFINE has signed a written agreement with user and the contents in the
agreement are in conflict with above terms, the contents in the written agreement
Precautions
To ensure the safety of the persons and instruments, please notice the following
1. Operation cautions
(1) The rated power supply of the instrument is 198VAC-242V AC, 50Hz /60Hz.
(2) Do not dismantle the instrument without permission as its partial high voltage.
Please contact with sale and technical service center if internal inspection or
debug is needed.
(3) Please do not touch the wiring terminal or testing line to prevent the operator
from electric shock during testing. Inserting or pulling out communication port
(4) It is dangerous to touch the input interface when electrical circuit works.
(5) No pressure on the instrument and cables. Keep the instrument far away from
heat source.
(6) Please make sure that the ventilation on both upper and down panel are
instrument.
(7) Please cut off the power supply immediately if bad smell or smoke is
perceived and contact with sale and technical service center of EVERFINE.
(8) It is extremely dangerous for the instrument comes in connect with liquid like
water or oil, or other metal scraps, and it is forbidden to place any items on the
instrument.
(9) Please pull out the power cord plug from the power outlet if the instrument is
(10) Ensure all cables are disconnected especially the power cord before delivery.
Carry the instrument by the handles on both sides of the instrument gently
without collisions.
2. Storage cautions
(1) The instrument shall be deposited in the condition with temperature between
0℃ and 45℃, and relative humidity below 80% with no condensation. The
(2) Please keep the product packaging materials including carton, cushion layer
and plastic bags to protect the instrument from temperature change, shock and
(3) Please do not deposit the apparatus in the environment with dust, smoke or
chemical gas.
3. Calibration cautions
(3) Please connect EVERFINE Test and Calibration center for more information
Contents
Foreword........................................................................................................................ 1
Copyright Statement.......................................................................................................2
Precautions..................................................................................................................... 3
Contents..........................................................................................................................5
Chapter 1 Overview........................................................................................................7
Chapter 2 Specifications.................................................................................................8
Overview...................................................................................................................... 32
Specifications............................................................................................................... 33
Chapter 1 Overview
PF9800 series digital power meter is a compact and economic digital power meter,
which can meet the basic power frequency electrical parameter measurement ,has the
high cost performance and is widely used in the fields of production, test, evaluation
and research.
PF9800 series digital power meter contains three digital power
meter:PF9804,PF9810and PF9811.PF9810 and PF9811 have the same test function:
voltage,current,power,power factor,voltage crest factor ,current crest factor,harmonic
analysis,serial communication and other function.The two digital power meters only
have difference in the current measurement range.
Product Standard: Q/HYG 003-2011 Intelligent power meter.In addition to the basic
power frequency electric paraments measurement,PF9804 can also presuppose
current,the range of power and determine whether a measured value is transfinite
automatically.PF9800 series digital power meter also can mute noise and have latched
function ,It can lock the currently displayed value to facilitate data record and
view,which meet the large and repetitive testing requirements in the production line.
Chapter 2 Specifications
Instantaneous maximum
1000V 4000mA 40A
allowable input for 1s
Continuous maximum
700V 3000mA 30A
allowable input
The measuring value become The measuring value become less than
Range down
less than 30% of the rated range 20% of the rated range
tiv
Item
e Power
Voltage Current Frequency
po factor
Conditions
we
Temperature : 23℃±5℃
Supply voltage: 220V±10V ± (0.4% of rdg + 0.1% of ±(0.004+0.001 ±(0.1%of rdg+1 digit)
wave digit)
.
Figure 3.2 The Front Panel of PF9810/PF9811
Functional
Operation Keys Function Description
classification
Window display switch button
FUNCTION(the right of D
Display switching When in the upper and lower limit state of
display window
active power,the key be used to switch the
Table 3.2 Indicative function and Operation Keys on the panel of PF9810/PF9811
Functional
Operation Keys Function Description
classification
FUNCTION(the right of D Window display switch button
Display switching
display window
To hold the display value when the switch
Display Maintenance HOLD
is on.Press the key again to refresh value
In the harmonic analysis state, the
harmonic order or communication address
∧
is added 1.
0 → A → K → U → + →
1 → B → L → V → - →
2 → C → M → W → × →
3 → D → N → X → ÷ →
4 → E → O → Y →
5 → F → P → Z →
6 → G → Q → c →
7 → H → R → h →
8 → I → S → i →
9 → J → T → u →
Please connect the power supply and load according to Figure 4.1 when measuring the
voltage and current.
Note:A load current flows in the thick lines shown in the diagrams; therefore, a wire
2000mA of current
1) The Value of Urms or Irms is more than 110% of the current measurement
range.
2) The value of Upk or Ipk is more than 165% of the current measurement
range.
(2) Range change down (The range will be automatically changed down when
1) The value of Urms is less than 30% of the current measuring range or the
2) The value of Upk or Ipk is less than 165% of the next measuring range.
(3) When the input effective voltage or current is more than 140% of the
Note:Because the range is set to auto, the measuring range switches according to
range up / range down conditions. Therefore the range may vary even if the measured
values remain the same. The instrument is set to maxim range when power on.
There are 4 display windows in this instrument: A, B, C and D, which can display
Figure 5.1 the loop switching display content of normal measurement by PF9811/PF9810
When the PF9811 and PF9810 are used to harmonic measurement ,the displays which
the function keys to switch the window are shown in Figure 5.2.
Figure 5.2 the loop switching display content of harmonic measurement by PF9811/PF9810
When using the normal measurement,the refresh rate of display is about 0.5
seconds,and he refresh rate of display is about 2seconds by using the harmonic
measurement.
Voltage/Current Hz D
Frequency
frequency
Peak Ratio Voltage Peak ratio CF A
(No for PF9804 ) Current Peak ratio CF B
The calculation method of harmonic analysis is IEC. There are two methods IEC and
(1) Two calculation methods of total harmonic distortion and each harmonic
distortion
IEC method():Computes the ratio of the rms value of the 2nd to 50th order
component to that of the fundamental (the 1st order). The computation equations are
as follows:
n n
∑ (U k )2
k =2
∑ (I
k =2
k )2
Uthd = ×100% Ithd = ×100%
U1 or I1
Uk Ik
U k% = × 100% I k% = ×100%
U1 or I1
CSA method():
Computes the ratio of the rms value of the 2nd to 50th order component to that of the
rms value the 1st to 50th component. The computation equation are as follows:
n n
∑ (U k ) 2
k =2
∑ (I
k =2
k )2
Uthd = × 100% or Ithd = × 100%
n n
∑ (U
k =1
k )2 ∑ (I
k =1
k )2
Uk Ik
U k% = × 100% or I k% = × 100%
n n
∑ (U
k =1
k )2 ∑ (I
k =1
k )2
voltage or current;
voltage or current;
k:harmonic order;
k:harmonic order;
Uk:The effective value of single harmonic
voltage;
current;
Note:In calculating the total harmonic voltage,current and power,loss and calculate
from the fundamental to the effective value of the maximum harmonic ,and 0
harmonic value (DC component) and the harmonic value which is more than the
PF9804 has the judgment function of upper and lower limit,the operations are as
shown in Figure 5.3.
Note:
When the measured power is zero, the instrument will not judge the limit of
power. When the measured current is zero, the instrument will not judge the
limit of current.
If you don’t want to judge the limit of power, you can set the upper value of
power as 9999kW and the lower value of power as 0.000W. If you don ’t
want to judge the limit of current, you can set the upper value of current as
9999A and the lower value of current as 0.000A.
Being limited to display digit, when the active power is negative, the minus
is not displayed (negative polarity can be judged depending on the power
factor). The value of power must be over the preset limit because it is
positive polarity, and will result in alarm. You can turn off the signal input
and change the position of the connecting wires of the two terminals in
rear panel(Voltage and Current can not be changed at the same time), which
makes the measured value of power be positive.
Lock the display value,the “LOCK”key light is turn on,Press the LOCK key again to
If you don’t need the sound alarm, you can press MUTE key to disable the bell to ring.
In MUTE state the MUTE indicator will light up. Press the MUTE key again to exit
C:\EVERFINE\PF981X_X.XX.
stored in PC.
Open the PF981X software to show the login interface,as figure 6.1
shown.
Function modules of software interface are as follows:
(1) Menu: New,Open,Save as,Print,Set communication port,Transfer data,Set
(2) Shortcut icon: New,Open,Save as,Transfer data,Show the window and so on.
(5) Status.
6.3.1 ‘File’menu
6.3.2 ‘Communication’menu
“IEC” or “CSA”.
watt;
instrument.
6.3.4 ‘Help’menu
(1)English Edition: Set up the use of language for English.After setting up,you need
(2)Chinese Edition: Set up the use of language for Chinese.After setting up,you need
Note: when using the software which come from PF9810/PF9811,the instrument
Please see the notes in each chapter for other cases for more details.
The precision between 45Hz and 65Hz should better than 0.05%. The uncertainty
of other frequency should be less than 1/3 to 1/5 of the maximum permissible
The precision between 45Hz and 65Hz should better than 0.02%/3min. The
stability of other frequency should be better than 1/5 to 1/10 of the accuracy the
detected instrument.
65Hz)
verification specification.
Test the measuring accuracy of voltage, current, active power and active power in
PF9800 series instrument in normal work. The accuracy of other electric parameters is
(1) The wiring method of test is shown in figure 8.1 . Please connect the
(2) Ensure that the output of voltage and current of signal is in the input range of
the instrument. Read and record the data of voltage, current, active power and
frequency.
The error is calculated according to the formula (1) and the measuring results should
R=BX-BN (1)
where:
Overview
The PF9800 digital power meter series are kinds of portable and economical meter
that can implement the power frequency electrical parameter measurement, which
are cost-effective and suitable for production test and assessment, and scientific
research, etc.
The PF9800 series involve three models, the three models are PF9804, PF9810,
PF9811, repectively. The PF9810 and PF9811 provide the same functionality in
measuring voltage, current power, power factor, frequency, and harmonic analysis and
serial communication, the two models are just different in mesurement ranges. Not
only the PF9804 can implement function like measuring voltage, current power,
power factor and frequency, but also preset current, the upper and lower bound of
power, and sound attenuation, the PF9804 is also able to automatically determine
whether the measurement value exceed the limits.The PF9800 digital power meter
series are also equipped with function of latching current displayed values for data
recording, which can meet the demands on generating large and repeating test.
Specifications
Input
Table 1 Input
Voltage Current
Item
All Model PF9810 PF9811/PF9804
Range 600V/300V/150V/75V 2000mA/800mA/200mA/50mA 20A/8A/2A/0.5A
Measurement Range 3.0V~600V 0.5mA~2000mA 0.005A~20A
Input Resistance More thant 1MΩ Less than 0.2Ω Less than 0.02Ω
Transient maximum
allowable input within 1000V 4000mA 40A
1s
Consecutive maximum
700V 3000mA 30A
allowable input
Range Switching Mode Auto Switch
The measured value is lower The measured value is lower than 20% of the rated
Range Increasing
than 30% of therated range range
Range Decreasing The measured value is over 110% of the rated range
Frequency Range Fundametal 45Hz~65Hz,bandwidth 5kHz
Accuracy
Table 2 Accuracy
Item
Voltage/Current/Active Power Crest Factor Frequency
Condition
Harmonic measurement(PF9810/PF9811)
Table 3 Harmonic measurement
Methods PPL
The Analysis item include: the relative value, effective value, crest factor of
each harmonic component of current or voltage, the relative value of total
Analysis item
harmonic distortion(THD) of voltage or current, the method for THD is IEC
in this instrument or IEC/CSA in upper computer software.
Frequency range The fundamental frequency of PPL source is between 45Hz to 65Hz
PLL source The input signal is equal or greater than 50% of range
Accuracy ±(0.3%range+5%read)
Analysis number 1~50
Refresh rate Refresh every two second
Communication(PF9810/PF9811)
Table 4 Communication
Communication
Standard configuration:RS-232,or RS-485、CAN
interface
Baud rate 2400
Communication
0、1、2……31
address
General Specifications
Table 4 General Specifications
Preheating Time Approx. 30 minutes or above
Operating Temperature: 5℃~40℃
environment Humidity: 20%R.H.~80%R.H. (no condensation)
Insulation >10M Ω ( the insulation resistance among the signal input, power input and
Resistance shell)
AC2000V (the withstand voltage between the signal input and shell, the
signal
Withstand Voltage input and power input within 1 minute),
DC2200V (the withstand voltage between the power
input and shell within 1 minute)
Power Supply AC 198V~242V,50Hz/60Hz
External
252mm×116mm×385.5mm (W×H×D)
Dimensions
PF9810/PF9811: 3.5kg
Weight
PF9804:2.4kg