PF9800 SERIES

DIGITAL POWER METER


USER’S MANUAL

V 3.01

EVERFINE Corporation (Stock Code: 300306)


ADD : Bldg.1 #669 Binkang Rd., Binjiang Hi-Tech
Zone, Hangzhou(310053), China
Tel :86-571-86698333
Fax :86-571-86696433
E-mail:Global@everfine.net
Globalservice@everfine.net
http://www.everfine.net

Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF9800 series User’s Manual

Foreword

Thank you for purchasing the EVERFINE PF9800 series of digital power meter. This

manual contains useful information regarding the functions and operating procedures

of the instrument as well as precautions that should be observed during use. To ensure

proper use of the instrument, please read this manual thoroughly before operating it.

Keep the manual in a safe place for quick reference whenever a question arises.

Notes:

• The contents of this manual are subject to change without prior notice.

• Every effort has been made in preparation of this manual to ensure the accuracy

of its contents. However, should you have any question or find any errors, please

contact your dealer or EVERFINE sales office.

• If you have different understanding to this manual, please refer to R&D

department of EVERFINE.

Checking Package Contents


Please check the instrument carefully when you unpack the box for the first time.If the wrong

instrument or accessories have been delivered, or some accessories are missing or

appear abnormal, please contact the dealer immediately.

Warm notice to valued customers of EVERFINE

"Ensure the quality, insist on continuous improvement and make every customer more

satisfied" is the quality policy of EVERFINE. Therefore, the quality of products and

services provided by EVERFINE should be better than those have been promised. If you

have further suggestions or advices on our products and services, please provide your

feedback to our quality supervision department.

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 1


PF9800 Series User's Manual

Copyright Statement

The copyright of this manual and the related information belongs to EVERFINE,

and it is protected by the copyright law of Peoples Republic of China and other

relevant international treaties. Copying, modifying, spreading, excerpting,

backing up or translating the whole or part contents of this manual by any

company or personnel without the written permission of EVERFINE is

prohibited. Otherwise it will be treated as infringement and the infringer will

assume law responsibility and all loss of EVERFINE. Any infringement related

above can be traced back to the responsible user by the unique product number

printed in the manual.

If EVERFINE has signed a written agreement with user and the contents in the

agreement are in conflict with above terms, the contents in the written agreement

have preferential force effect.

2 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF9800 series User’s Manual

Precautions

To ensure the safety of the persons and instruments, please notice the following

general safety precautions:

1. Operation cautions

(1) The rated power supply of the instrument is 198VAC-242V AC, 50Hz /60Hz.

(2) Do not dismantle the instrument without permission as its partial high voltage.

Please contact with sale and technical service center if internal inspection or

debug is needed.

(3) Please do not touch the wiring terminal or testing line to prevent the operator

from electric shock during testing. Inserting or pulling out communication port

with electric is forbidden.

(4) It is dangerous to touch the input interface when electrical circuit works.

(5) No pressure on the instrument and cables. Keep the instrument far away from

heat source.

(6) Please make sure that the ventilation on both upper and down panel are

unblocked. Otherwise, the internal temperature may rise to damage the

instrument.

(7) Please cut off the power supply immediately if bad smell or smoke is

perceived and contact with sale and technical service center of EVERFINE.

(8) It is extremely dangerous for the instrument comes in connect with liquid like

water or oil, or other metal scraps, and it is forbidden to place any items on the

instrument.

(9) Please pull out the power cord plug from the power outlet if the instrument is

not going to be used for some time.

(10) Ensure all cables are disconnected especially the power cord before delivery.

Carry the instrument by the handles on both sides of the instrument gently

without collisions.

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 3


PF9800 Series User's Manual

2. Storage cautions

(1) The instrument shall be deposited in the condition with temperature between

0℃ and 45℃, and relative humidity below 80% with no condensation. The

drying environment at 20℃ is recommended.

(2) Please keep the product packaging materials including carton, cushion layer

and plastic bags to protect the instrument from temperature change, shock and

vibration during transportation.

(3) Please do not deposit the apparatus in the environment with dust, smoke or

chemical gas.

(4) Direct sunlight is forbidden.

3. Calibration cautions

(1) In order to ensure measurement accuracy, regular calibration is recommended.

(2) Calibration period depends on the practical use of the instrument.

Recommended calibration period is at least once a year.

(3) Please connect EVERFINE Test and Calibration center for more information

about calibration. The email is lab@everfine.net.

4 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF9800 series User’s Manual

Contents
Foreword........................................................................................................................ 1

Copyright Statement.......................................................................................................2

Precautions..................................................................................................................... 3

Contents..........................................................................................................................5

Chapter 1 Overview........................................................................................................7

Chapter 2 Specifications.................................................................................................8

2.1 Main specifications............................................................................................8

2.2 General technical requirements......................................................................... 9

Chapter 3 Description of Instrument panel.................................................................. 11

3.1 Front Panel.......................................................................................................11

3.2 Rear Panel........................................................................................................13

Chapter 4 Measurement Condition Settings.................................................................15

4.1 The Wiring Mode of the Measurement Circuit............................................... 15

4.2 Measurement Range........................................................................................ 15

Chapter 5 Measurement Function................................................................................ 17

5.1 Display window...............................................................................................17

5.2 Function of Normal measurement...................................................................18

5.3 Function of harmonic measurement(PF9810/PF9811)................................... 19

5.4 Judgment Function of Limit(PF9804).............................................................21

5.5 Function of key lock........................................................................................22

5.6 Function of Mute(PF9804)..............................................................................22

5.7 Communication Setting(PF9810/PF9811)...................................................... 22

Chapter 6 Software Instruction(PF9810/PF9811)........................................................ 24

6.1 System requirements....................................................................................... 24

6.2 Software installation........................................................................................24

6.3 Software Use................................................................................................... 24

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 5


PF9800 Series User's Manual

Chapter 7 Common Failures and Resolution............................................................... 28

7.1 Common failures and shooting........................................................................28

Chapter 8 Calibration and Inspection...........................................................................29

8.1 Conditions for calibration and inspection....................................................... 29

8.2 Technical documents of calibration and test based......................................... 29

8.3 Equipments for calibration and test.................................................................29

8.4 Inspection items and methods......................................................................... 30

Overview...................................................................................................................... 32

Specifications............................................................................................................... 33

6 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF9800 series User’s Manual

Chapter 1 Overview
PF9800 series digital power meter is a compact and economic digital power meter,
which can meet the basic power frequency electrical parameter measurement ,has the
high cost performance and is widely used in the fields of production, test, evaluation
and research.
PF9800 series digital power meter contains three digital power
meter:PF9804,PF9810and PF9811.PF9810 and PF9811 have the same test function:
voltage,current,power,power factor,voltage crest factor ,current crest factor,harmonic
analysis,serial communication and other function.The two digital power meters only
have difference in the current measurement range.
Product Standard: Q/HYG 003-2011 Intelligent power meter.In addition to the basic
power frequency electric paraments measurement,PF9804 can also presuppose
current,the range of power and determine whether a measured value is transfinite
automatically.PF9800 series digital power meter also can mute noise and have latched
function ,It can lock the currently displayed value to facilitate data record and
view,which meet the large and repetitive testing requirements in the production line.

Product standard:Q/HYG 003-2014 digital power meter

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 7


PF9800 Series User's Manual

Chapter 2 Specifications

2.1 Main specifications


(1) Input
Table 2.1 Input specifications
Voltage Current
Item
All PF9810 Rest

600V/300V 2000mA/800mA/ 20A/8A


Range
/150V/75V 200mA/50mA /2A/0.5A

Measuring range 3.0V~600V 0.5mA~2000mA 0.005A~20A

Input resistor > 1MΩ <0.2Ω <0.02Ω

Instantaneous maximum
1000V 4000mA 40A
allowable input for 1s

Continuous maximum
700V 3000mA 30A
allowable input

Range mode Auto range

Range up The reading value exceeds 110% of the rated range

The measuring value become The measuring value become less than
Range down
less than 30% of the rated range 20% of the rated range

Frequency range fundamental frequency is from 45Hz to 65Hz,brand width is 5kHz.

(2) Measurement accuracy


Table 2.2 Measurement accuracy of voltage, current and active power
Ac

tiv
Item
e Power
Voltage Current Frequency
po factor
Conditions
we

8 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF9800 series User’s Manual

12 months after calibration

Temperature : 23℃±5℃

Humidity: 30% to 75% R.H.

Supply voltage: 220V±10V ± (0.4% of rdg + 0.1% of ±(0.004+0.001 ±(0.1%of rdg+1 digit)

Input waveform: 45Hz~65Hz sine rng +1 digit) of rdg + 1

wave digit)

Input range: (10%~100%)Range

Common mode voltage: 0VDC

Power and PF are measured under

the voltage of 110/220/380V

(3) Harmonic measurement(PF9810/PF9811)


Table 2.3 Function of harmonic measurement
Method Phase-locked loop synchronization method(PLL)
The effective value of each harmonic component and relative value of
voltage, current and active power
Analysis of The relative values of total harmonic distortion of the voltage and current
Projects The calculation method of harmonic distortion is IEC
the total calculation method of harmonic distortion can be set IEC or CSA in
computer.
Frequency range The fundamental frequency of PLL source is between45Hz and 65Hz.
PLL Source The input signal o should be more than 50% of the range.
Accuracy ±(0.3% of rng +5% of rdg)
Analysis Order 1~50
Update Interval of
Approx.2 seconds once time
harmonic display
(4) Communication(PF9810/PF9811)
Table 2.4 Communication settings
Communication interface Standard Configuration:RS-232,Optional:RS-485 、LAN
Baud rate 2400
Communication address 0, 1, 2……31

2.2 General technical requirements

Table 2.5 General technical requirements


Preheating time ≥30 minutes
Operating Temperature: 5℃~40℃

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 9


PF9800 Series User's Manual

environment Humidity: 20%R.H.~80%R.H.(without condensation)


Insulation Signal input end, shell and power supplying input end should be insulated
resistance from each other, with insulation resistance over 10M.
AC2000V should last 1 minute between signal input end and the either of
Withstand voltage shell and power supply input end. DC2200V should last 1 minute between
shell and power supply input end.
Power supply 198VAC~242VAC,50Hz/60Hz
Box size 252mm×116mm×385.5mm(W×H×D)
PF9810/PF9811: 3.5kg
Weight
PF9804: 2.4kg

10 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF9800 series User’s Manual

Chapter 3 Description of Instrument panel

3.1 Front Panel


The front panel of PF9804 is shown in Figure 3.1.

Figure 3.1 The Front Panel of PF9804


The front panel of PF9810/PF9811 is shown in Figure 3.2

.
Figure 3.2 The Front Panel of PF9810/PF9811

3.1.1 Indicative Function and Operation Keys on the panel


Table 3.1 Indicative function and Operation Keys on the panel of PF9804

Functional
Operation Keys Function Description
classification
Window display switch button
FUNCTION(the right of D
Display switching When in the upper and lower limit state of
display window
active power,the key be used to switch the

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 11


PF9800 Series User's Manual

active power unit W and kW


Set the upper and lower limit value of
Limit Setting SETUP
power and current
To hold the display value when the switch
Display Maintenance HOLD
is on.Press the key again to refresh value
When the current or power is out of the
preset limit, the bell is inhibited to ring.
Noise Mute MUTE The indicator will light up. When it is
pressed again, the MUTE can be
cancelled.
Move the decimal point from left to right
· when setting the value of upper and lower
limit of power
Move the cursor from left to right when
Common settings > setting the value of upper and lower limit
of power
The number in cursor is added 1 when
∧ setting the value of upper and lower limit
of power

Table 3.2 Indicative function and Operation Keys on the panel of PF9810/PF9811

Functional
Operation Keys Function Description
classification
FUNCTION(the right of D Window display switch button
Display switching
display window
To hold the display value when the switch
Display Maintenance HOLD
is on.Press the key again to refresh value
In the harmonic analysis state, the
harmonic order or communication address

is added 1.

Common settings In the harmonic analysis state, the


harmonic order or communication address

is subtracted 1.

ENTER To confirm the parameters setting


Start the harmonic analysis. When the key
Harmonic Analysis ANALYZE is pressed again, the instrument exits from
the harmonic analysis state
To set communication Mode,Can choose
Communication
COM MODE multi-machine communication mode or
Mode
normal communication mode

12 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF9800 series User’s Manual

3.1.2 Digital and character display

Table 3.3 Comparison table of digital and character display

0 →  A →  K →  U →  + → 
1 →  B →  L →  V →  - → 
2 →  C →  M →  W →  × → 
3 →  D →  N →  X →  ÷ → 
4 →  E →  O →  Y → 
5 →  F →  P →  Z → 
6 →  G →  Q →  c → 
7 →  H →  R →  h → 
8 →  I →  S →  i → 
9 →  J →  T →  u → 

3.2 Rear Panel


The rear panel is shown in Figure 3.3

Figure 3.3 The Rear Panel of PF9800 series

Table 3.4 The Rear Components

Numbers Component Function description


binding post of voltage
1 ± terminal is black and V terminal is red.
input

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 13


PF9800 Series User's Manual

binding post of current


2 ± terminal is black and A terminal is red.
input
PF9810/pf9811:Standard Configuration: RS-232,
3 Communication interface Optional:RS-485 、LAN
PF9804: This port is not a communication interface
To ensure safety, please make sure that the
4 Ground terminal
instrument is reliable grounding in use.
Input interface of instrument power, including 0.5A
5 Power supply socket
fuse

14 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF9800 series User’s Manual

Chapter 4 Measurement Condition Settings

4.1 The Wiring Mode of the Measurement Circuit

Please connect the power supply and load according to Figure 4.1 when measuring the
voltage and current.

Figure 4.1 The Wiring Mode

Note:A load current flows in the thick lines shown in the diagrams; therefore, a wire

with sufficient current capacity must be used for these lines.

4.2 Measurement Range

4.2.1 Range of voltage ,current and Power


Table 4.1 The range of voltage ,current and power
Range of voltage Range of current Range of Active Power

All Models PF9804/PF9811 PF9810 All Models

75V/150V/300V/600V 0.5A/2A/8A/20A 50mA/200mA/800mA/ Range of voltage × Range

2000mA of current

4.2.2 Condition for automatic range switch


(1) Range change up (The range will be automatically changed up when meeting

any of the following conditions):

1) The Value of Urms or Irms is more than 110% of the current measurement

range.

2) The value of Upk or Ipk is more than 165% of the current measurement

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 15


PF9800 Series User's Manual

range.

(2) Range change down (The range will be automatically changed down when

meeting any of the following conditions):

1) The value of Urms is less than 30% of the current measuring range or the

value of Irms is less than 20% of the current measuring range.

2) The value of Upk or Ipk is less than 165% of the next measuring range.

(3) When the input effective voltage or current is more than 140% of the

maximum range, overrange dispose will occurs,the corresponding value of

voltage / current and power display “- - - -”.

Note:Because the range is set to auto, the measuring range switches according to

range up / range down conditions. Therefore the range may vary even if the measured

values remain the same. The instrument is set to maxim range when power on.

16 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF9800 series User’s Manual

Chapter 5 Measurement Function

5.1 Display window

There are 4 display windows in this instrument: A, B, C and D, which can display

different measurement results simultaneously. The content of display window can be

selected by the corresponding FUNCTION key on the window D.


Table 5.1 The display content of PF9804
Window Display Content Measurement Range
A V U:3.0V~600V
B A I:0.005A~20A
C (k)W P:1W~12.00kW
D PF,Hz PF:-1.000~+1.000
Hz:45Hz~65Hz
Table 5.2 The display content of PF9810/PF9811
Window Display Content Measurement Range
A V U:3.0V~600V
B A(PF9811) I:0.005A~20A(PF9811)
mA(PF9810) 0.5mA~2000mA(PF9810)
C (k)W(PF9811) P:1W~12.00kW
W(PF9810) PF:-1.000~+1.000
D PF,Hz or.**/tHd Hz:45Hz~65Hz
When the PF9811 and PF9810 are used to normal measurement ,the displays which
the function keys to switch the window are shown in Figure 5.1.

Figure 5.1 the loop switching display content of normal measurement by PF9811/PF9810

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 17


PF9800 Series User's Manual

When the PF9811 and PF9810 are used to harmonic measurement ,the displays which
the function keys to switch the window are shown in Figure 5.2.

Figure 5.2 the loop switching display content of harmonic measurement by PF9811/PF9810
When using the normal measurement,the refresh rate of display is about 0.5
seconds,and he refresh rate of display is about 2seconds by using the harmonic
measurement.

5.2 Function of Normal measurement

The measurement parameters calculated in normal measuring state is shown in Table


5.3.
Table 5.3 Measurement parameters in normal measuring state

Parameter Unit/Symbol Display window


Voltage true effective value V A
Current true effective value A B
Power Active power W C
Power Factor Power factor PF

18 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF9800 series User’s Manual

Voltage/Current Hz D
Frequency
frequency
Peak Ratio Voltage Peak ratio CF A
(No for PF9804 ) Current Peak ratio CF B

5.3 Function of harmonic measurement(PF9810/PF9811)

Harmonic component of voltage and current is analyzed in harmonic measurement.

The calculated parameters can be measured in the harmonic measuring state , as

shown in table 5.4.


Table 5.4 Display content of each window in harmonic measuring state
Parameter Unit/Symbol Display window
Voltage true effective value V A
A/mA B
Current true effective value
(PF9811/PF9810)
Power Active power W C
Power Factor Power factor PF D
Frequency Voltage(Current )frequency Hz
Peak Ratio Voltage Peak ratio CF A
(No for PF9804 ) Current Peak ratio CF B
Total/Single Total/each harmonic voltage V A
harmonic Total/each harmonic current A/mA B
measurement (PF9811/PF9810)
Total harmonic Total harmonic voltage V% A
distortion/Relativ distortion/Relative value of
e value of single harmonic voltage
harmonic Total harmonic current A% B
measurement distortion/Relative value of
harmonic current

The measurement function fully meets the requirements of IEC61000-3-2 standard in

harmonic measuring state.

The calculation method of harmonic analysis is IEC. There are two methods IEC and

CSA to select in computer communication software .

(1) Two calculation methods of total harmonic distortion and each harmonic

distortion
IEC method():Computes the ratio of the rms value of the 2nd to 50th order

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 19


PF9800 Series User's Manual

component to that of the fundamental (the 1st order). The computation equations are
as follows:
n n

∑ (U k )2
k =2
∑ (I
k =2
k )2
Uthd = ×100% Ithd = ×100%
U1 or I1

Uk Ik
U k% = × 100% I k% = ×100%
U1 or I1

CSA method():
Computes the ratio of the rms value of the 2nd to 50th order component to that of the
rms value the 1st to 50th component. The computation equation are as follows:
n n

∑ (U k ) 2
k =2
∑ (I
k =2
k )2
Uthd = × 100% or Ithd = × 100%
n n

∑ (U
k =1
k )2 ∑ (I
k =1
k )2

Uk Ik
U k% = × 100% or I k% = × 100%
n n

∑ (U
k =1
k )2 ∑ (I
k =1
k )2

Where Uthd or Ithd: the total harmonic distortion of voltage or current

U1 or I1:the effective value of fundamental frequency (the first

harmonic) of voltage or current;

Uk or Ik: the effective value of kth harmonic of

voltage or current;

Uk% or Ik%:the relative value of kth harmonic of

voltage or current;

k:harmonic order;

n:the maximum harmonic order set by user,n=50.

(2) Calculation method of total harmonic voltage, current and power


n
ΣU = ∑ (U
k =1
k )2
Total harmonic voltage:
n
ΣI = ∑ (I
k =1
k )2
Total harmonic current:

20 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF9800 series User’s Manual

In the formula,where n:the maximum harmonic order set by user;

k:harmonic order;
Uk:The effective value of single harmonic
voltage;

Ik: The effective value of single harmonic

current;

Note:In calculating the total harmonic voltage,current and power,loss and calculate

from the fundamental to the effective value of the maximum harmonic ,and 0

harmonic value (DC component) and the harmonic value which is more than the

maximum are not calculated.

5.4 Judgment Function of Limit(PF9804)

PF9804 has the judgment function of upper and lower limit,the operations are as
shown in Figure 5.3.

Figure 5.3 The setup of PF9804 limit


The Judgment of Upper and Lower Limit of Active Power and Current
OK When the measured power and current are within the preset
upper and lower value (PL≤PM≤PH and AL≤AM≤AH), the
green OK indicator will light up.
LOWER When the measured power or current is lower than the preset
lower value (PM < PL or A M<AL), the LOWER red
indicator will turn on and the green OK indicator will
extinguish , and the corresponding unit indicator W or A will
twinkle, and the bell will ring.
UPPER When the measured power or current is upper than the preset
upper value (PM > PL or A M>AL), the UPPER red indicator
will light up ;The green OK indicator will extinguish, and
the corresponding unit indicator W or A will twinkle, and the
bell will ring.

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 21


PF9800 Series User's Manual

Note:
When the measured power is zero, the instrument will not judge the limit of
power. When the measured current is zero, the instrument will not judge the
limit of current.
If you don’t want to judge the limit of power, you can set the upper value of
power as 9999kW and the lower value of power as 0.000W. If you don ’t
want to judge the limit of current, you can set the upper value of current as
9999A and the lower value of current as 0.000A.
Being limited to display digit, when the active power is negative, the minus
is not displayed (negative polarity can be judged depending on the power
factor). The value of power must be over the preset limit because it is
positive polarity, and will result in alarm. You can turn off the signal input
and change the position of the connecting wires of the two terminals in
rear panel(Voltage and Current can not be changed at the same time), which
makes the measured value of power be positive.

5.5 Function of key lock

Lock the display value,the “LOCK”key light is turn on,Press the LOCK key again to

release the lock state.

5.6 Function of Mute(PF9804)

If you don’t need the sound alarm, you can press MUTE key to disable the bell to ring.

In MUTE state the MUTE indicator will light up. Press the MUTE key again to exit

from MUTE state.

5.7 Communication Setting(PF9810/PF9811)

In the Normal mode measurement,Press the “COM MODE”key to the communication


mode setting, you can select communication “SPEC”or “NORM” by “ ∧ ”key and
“∨”key.
SPEC:Multi-machine communication mode.When using this mode,the instrument
address can be selected and saved,the setting range of address :0~31.
NORM: Normal communication mode,in this mode,you can not use multi-machine
communication.

22 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF9800 series User’s Manual

Figure 5.4 PF9810/PF9811 communication mode switching

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 23


PF9800 Series User's Manual

Chapter 6 Software Instruction(PF9810/PF9811)

6.1 System requirements

(1)The operation system must be Windows XP or Win7 (Chinese or English);


(2)More than 2M dis space.
(3)The display resolution is 800×600 or above. 800×600is advised.
(4)One serial port of RS-232.

6.2 Software installation

Please inserts the diskette labeled‘EVERFINE’into floppy drive,GO

into the latest Chinese version of PF9810/PF9811 in the diskette,and run

SETUP.EXE,that is ,it is going into the installation process. The

installation can be finished according to the prompts.

After the installation is completed,the file is placed on the target path

C:\EVERFINE\PF981X_X.XX.

6.3 Software Use

The PC application software of PF981X is used to obtain the measured

value and the data of harmonic wave, which is transmitted to the PC

through the interface of RS-232. These data are displayed on PC and

stored in PC.
Open the PF981X software to show the login interface,as figure 6.1
shown.
Function modules of software interface are as follows:
(1) Menu: New,Open,Save as,Print,Set communication port,Transfer data,Set

total harmonic calculation method,Switch the software interface version of

Chinese and English and so on.

(2) Shortcut icon: New,Open,Save as,Transfer data,Show the window and so on.

(3) Main window display:normal measurement display area, Signal waveform

display area,Harmonic frequency spectr display area,as shown in Figure 6.1.

24 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF9800 series User’s Manual

(4) Harmonic window display: measurement data display area,each harmonic

measurement value, as shown in Figure 6.2.

(5) Status.

Figure 6.1 The main interface software

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 25


PF9800 Series User's Manual

Figure 6.2 the harmonic window

6.3.1 ‘File’menu

(1)New:Create a new data File;

(2)Open:Load the saved file of PF9811/PF9810;

(3)Save as:Save the current test data to the specified file;

(4)Print:Output current display to the print;

(5)Print preview:Can preview the paper after printing;

(6)Exit: Sign out the software.

6.3.2 ‘Communication’menu

(1)Transfer data:Click to start the instrument communication;

(2)Total harmonic calculation:Set total harmonic calculation method,can choose

“IEC” or “CSA”.

(3)Display harmonic current/watt:Set whether to display the harmonic current per

watt;

26 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF9800 series User’s Manual

(4) Set communication port:Set the communication port used in communication of

instrument.

6.3.3 ‘Tool’ menu

(1) Record product features: Set up the instrument type,serial number,date,etc.

(2) Observation:Choose to display ‘the main window’or ‘the harmonic window’.

6.3.4 ‘Help’menu

(1)English Edition: Set up the use of language for English.After setting up,you need

to restart the software to take effect.

(2)Chinese Edition: Set up the use of language for Chinese.After setting up,you need

to restart the software to take effect.

(3)About: Display related information about software.

Note: when using the software which come from PF9810/PF9811,the instrument

choose ‘NORM’communication mode.

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 27


PF9800 Series User's Manual

Chapter 7 Common Failures and Resolution

7.1 Common failures and shooting


Serial
Phenomenon Solutions
number
(1) Ensure the instrument power line connected
The instrument window
correctly.
1 display nothing after
(2) Ensure that the power supply is in the allowed
operation.
range.
(1) Ensure that the working environment temperature
and humidity is in the allowed range.
(2) Ensure that the display is not affected by noise
interference.
The measured value of
2 (3) Check the test line connected correctly or not.
display is not accurate.
(4) Check the connection mode settings correct or
not.
(5) Check the date displayed in the lock or not.
(6) reboot.
3 The key operation is invalid (1) Check there are other keys died or not.
Harmonic measurement is (1) Check voltage and current signals are greater than
4
irregular. the range of 30% or not.
(1) Check whether the communication line is
connected properly or not.
5 Communication is fail.
(2) Check the instrument address and communication
mode are corresponding to the computer or not.

Please see the notes in each chapter for other cases for more details.

28 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF9800 series User’s Manual

Chapter 8 Calibration and Inspection

8.1 Conditions for calibration and inspection

(1) Ambient temperature: 23℃± 2℃

(2) Ambient humidity: 35% ~ 55%

(3) Voltage of power supply: 220V±11V

(4) Frequency of power supply: 50Hz ±1Hz

(5) Preheating time: ≥30 minutes

(6) No mechanical vibration and electromagnetic interference around to affect

the normal work.

8.2 Technical documents of calibration and test based

(1) JJG(Zhe)90-2007 Digital multimeter calibrator

(2) JJG 780-1992 AC digital power meter

(3) JJG(Zhe)89-2007 Electric parameters test meter

8.3 Equipments for calibration and test

(1) Standard meter(0V~600V,0A~20A)

The precision between 45Hz and 65Hz should better than 0.05%. The uncertainty

of other frequency should be less than 1/3 to 1/5 of the maximum permissible

error of the detected instrument.

Recommend: PF2010 digital power meter from EVERFINE.

(2) Signal source(0V~600V,0A~20A)

The precision between 45Hz and 65Hz should better than 0.02%/3min. The

stability of other frequency should be better than 1/5 to 1/10 of the accuracy the

detected instrument.

Recommend: YS6100 standard power from EVERFINE(test between 45Hz and

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 29


PF9800 Series User's Manual

65Hz)

Note:the test equipments should be qualified regular metrological

verification specification.

8.4 Inspection items and methods

8.4.1 Items of calibration and test

Test the measuring accuracy of voltage, current, active power and active power in

PF9800 series instrument in normal work. The accuracy of other electric parameters is

calculated such as wave crest ratio and power factor.

8.4.2 Method of calibration and test

(1) The wiring method of test is shown in figure 8.1 . Please connect the

instrument, signal source and standard meter correctly.

(2) Ensure that the output of voltage and current of signal is in the input range of

the instrument. Read and record the data of voltage, current, active power and

frequency.

The error is calculated according to the formula (1) and the measuring results should

meets the requirements of measurement accuracy in chapter 2.2.

R=BX-BN (1)

where:

BN:reading data of standard meter

BX:reading data of detected instrument

R: instrument measurement error.

30 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF9800 series User’s Manual

Figure 8.1 Wiring diagram of calibration and test of PF9800 series

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 31


PF9800 Series User's Manual

Overview

The PF9800 digital power meter series are kinds of portable and economical meter
that can implement the power frequency electrical parameter measurement, which
are cost-effective and suitable for production test and assessment, and scientific
research, etc.
The PF9800 series involve three models, the three models are PF9804, PF9810,
PF9811, repectively. The PF9810 and PF9811 provide the same functionality in
measuring voltage, current power, power factor, frequency, and harmonic analysis and
serial communication, the two models are just different in mesurement ranges. Not
only the PF9804 can implement function like measuring voltage, current power,
power factor and frequency, but also preset current, the upper and lower bound of
power, and sound attenuation, the PF9804 is also able to automatically determine
whether the measurement value exceed the limits.The PF9800 digital power meter
series are also equipped with function of latching current displayed values for data
recording, which can meet the demands on generating large and repeating test.

32 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF9800 series User’s Manual

Specifications

Input
Table 1 Input
Voltage Current
Item
All Model PF9810 PF9811/PF9804
Range 600V/300V/150V/75V 2000mA/800mA/200mA/50mA 20A/8A/2A/0.5A
Measurement Range 3.0V~600V 0.5mA~2000mA 0.005A~20A
Input Resistance More thant 1MΩ Less than 0.2Ω Less than 0.02Ω
Transient maximum
allowable input within 1000V 4000mA 40A
1s
Consecutive maximum
700V 3000mA 30A
allowable input
Range Switching Mode Auto Switch
The measured value is lower The measured value is lower than 20% of the rated
Range Increasing
than 30% of therated range range
Range Decreasing The measured value is over 110% of the rated range
Frequency Range Fundametal 45Hz~65Hz,bandwidth 5kHz

Accuracy
Table 2 Accuracy

Item
Voltage/Current/Active Power Crest Factor Frequency
Condition

Within a year after calibration


Temperature: 23±5℃
Humidity: 30%~75%R.H.
Supply Voltage:220±10V
Input Waveform: 45Hz~65Hz ±(0.4%read + 0.1%range+1dig) ±(0.004 + 0.001/read+1dig) ±(0.1%read+1dig)
sine wave
Input range :( 10%~100% )
range
Common-mode voltage: 0V DC
Active power or PF mesaured
under the voltage of
110V/220V/380V

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 33


PF9800 Series User's Manual

Harmonic measurement(PF9810/PF9811)
Table 3 Harmonic measurement
Methods PPL
The Analysis item include: the relative value, effective value, crest factor of
each harmonic component of current or voltage, the relative value of total
Analysis item
harmonic distortion(THD) of voltage or current, the method for THD is IEC
in this instrument or IEC/CSA in upper computer software.
Frequency range The fundamental frequency of PPL source is between 45Hz to 65Hz
PLL source The input signal is equal or greater than 50% of range
Accuracy ±(0.3%range+5%read)
Analysis number 1~50
Refresh rate Refresh every two second

Communication(PF9810/PF9811)
Table 4 Communication
Communication
Standard configuration:RS-232,or RS-485、CAN
interface
Baud rate 2400
Communication
0、1、2……31
address

General Specifications
Table 4 General Specifications
Preheating Time Approx. 30 minutes or above
Operating Temperature: 5℃~40℃
environment Humidity: 20%R.H.~80%R.H. (no condensation)
Insulation >10M Ω ( the insulation resistance among the signal input, power input and
Resistance shell)
AC2000V (the withstand voltage between the signal input and shell, the
signal
Withstand Voltage input and power input within 1 minute),
DC2200V (the withstand voltage between the power
input and shell within 1 minute)
Power Supply AC 198V~242V,50Hz/60Hz
External
252mm×116mm×385.5mm (W×H×D)
Dimensions
PF9810/PF9811: 3.5kg
Weight
PF9804:2.4kg

34 Copyright © EVERFINE, Copy or spread without authorization is prohibited.