Incubator Shaker 50X User Manua
Incubator Shaker 50X User Manua
Warning
This manual is periodically updated. The updates shall be included in the new releases of
the Software.
Information supplied in this manual may be subject to modifications before the products
described become available.
This manual may contain information or references relating to certain bioMérieux S.A.
products, software or services which are not available in the country of release; this shall not
mean that bioMérieux S.A. intends to market such products, software or services in such
country
To request copies of publications or for any technical request, contact bioMérieux S.A. or
your local distributor.
IMPORTANT! Use of the instrument and manual implies acceptance of the clauses below and the
clauses set out in the regulatory booklet. Users are invited to refer to these clauses.
Trademarks
bioMérieux and the blue logo are used, pending and/or registered trademarks belonging to
bioMérieux S.A. or one of its subsidiaries.
Any other name or trademark is the property of its respective owner.
The list of revisions below summarizes replacements or additional pages in your User’s
Manual.
Date of
Version Modifications Pages modified
printing
2 General
Introduction ............................................................................................................................................................... 2-1
Limitations, Warnings and Cautions.......................................................................................................................... 2-1
Limitations.......................................................................................................................................................... 2-1
Control LEDs .............................................................................................................................................. 2-1
Incubator Slot Preheating ........................................................................................................................... 2-1
Module Shake Speed ................................................................................................................................. 2-1
Warnings and Cautions...................................................................................................................................... 2-1
General ....................................................................................................................................................... 2-1
Temperature ............................................................................................................................................... 2-2
Shaking....................................................................................................................................................... 2-2
Operation .................................................................................................................................................... 2-3
Maintenance ............................................................................................................................................... 2-3
Technical Specifications............................................................................................................................................ 2-3
General Specifications ....................................................................................................................................... 2-3
Incubator specifications ..................................................................................................................................... 2-4
Shaker Specifications ........................................................................................................................................ 2-4
Description ................................................................................................................................................................ 2-5
Front View.......................................................................................................................................................... 2-5
Rear View .......................................................................................................................................................... 2-6
Keyboard and display ........................................................................................................................................ 2-7
Thermal Microplate Holder................................................................................................................................. 2-8
3 Installation
Unpacking procedure ................................................................................................................................................ 3-1
Environmental requirements ..................................................................................................................................... 3-1
Voltage Selector........................................................................................................................................................ 3-1
Power cable ....................................................................................................................................................... 3-2
Setup ................................................................................................................................................................. 3-2
6 Device interfaces
Note: Screenshots, figures, and messages are given for information purposes only.
This manual provides the appropriate instructions to install, operate and maintain the
Incubator Shaker 50x.
Warnings
The manufacturer does not assume any responsibility for user modifications. Such
modifications might change the EMC-behavior of the instrument.
Any modification of this instrument by the user and any use of the instrument outside the
description in the User Manual ends the warranty given by the manufacturer. Any repair or
modification to this instrument by another person than an bioMérieux representative ends the
warranty given by the manufacturer.
Introduction
The Incubator Shaker 50x is a compact, modular instrument designed for incubation and
shaking of plates used in microplate-based assays. The Incubator Shaker 50x may consist
of one to three modules, each module having a capacity for 3 microplates at a time.
The Incubator Shaker 50x provides a wide incubation temperature range, from room
temperature up to 40°C, as well as programmable shaking speed and time settings. It
operates as a stand-alone instrument.
Thermal microplate holders are heated electrically via connectors in the plate carrier. An
upper heater within the top of the module prevents condensation from forming within the
microplate holder. Orbital shaking occurs within a module at selectable speed settings. All slot
positions within a module are shaken simultaneously.
Limitations
Control LEDs
The orange control LED (Light Emitting Diode) flashes when a slot is heating up to the
programmed target temperature. Refer to section Slot Status – Control LEDs for information.
The LED turns to a continuous orange color when the slot reaches the target temperature.
However, the continuous orange LED may falsely indicate that a slot is at the target
temperature under the following conditions:
− If the microplate holder is removed and then reinserted into the slot during the initial
preheating of the slot.
− If the programmed target temperature is lower than the actual temperature of the slot.
Incubator Slot Preheating
Heating of the incubator slots does not occur automatically when the instrument is powered
on. The operator must turn the temperature function on (Preheating) after the instrument has
been powered on.
Module Shake Speed
All slots within a module will shake at the same speed. The slots within a module shake at the
speed that is programmed into the most recently started processing protocol. Refer to section
Warnings and Cautions.
General
V i e wi n g P r o g r a m m e d P r o c e s s i n g P a r a m e t e r s
Programmed processing parameters can only be viewed while in the programming mode.
Care should be taken during review of parameters to ensure that unintentional changes are
not made to the programmed values.
Splashing may occur if shaking is used with wells that have too much liquid for a given speed.
To avoid splashing and potential cross-contamination from well to well, DO NOT overfill wells
for a given shaking speed setting.
To minimize the possibility of splashing, aerosols and evaporation during shaking and
incubation, it is recommended that you use microplate sealers.
Spills
All spills of liquids should be wiped up immediately from all surfaces of the instrument to
prevent leakage into the electronics and mechanics of the instrument. The surfaces should be
cleaned and decontaminated if necessary.
Incubation Time
DO NOT remove the microplate holder until the incubation time has elapsed. If the microplate
holder is removed from the slot before the incubation time has elapsed, the incubation timer
does not stop and continues to count down causing tracking of actual incubation time to be
lost.
D O N O T R e a c h I n t o S l o t W h e n P o we r I s T u r n e d O n
To prevent electrical shock, DO NOT reach into a slot and/or touch the power contacts at the
rear of the slot when the power is turned on to the instrument. If access to the slot is required
(e.g. cleaning), turn power off to the instrument and unplug the power cable.
High Temperatures
The microplate holders are designed to stop heating if temperatures reach 53°C ±10°C in the
unlikely event that temperature control to a microplate holder is lost or enters a runaway
condition. Microplate holders at this temperature will be hot when touched. To help prevent
loss of temperature control, ensure that the power contacts on the back of the microplate
holder are clean and free from dirt or corrosion and that the microplate holder is correctly
inserted into the slot before starting the processing protocol.
Temperature
When changing the programmed temperature setting, make sure that the slot temperature is
at the new target setting before starting the processing protocol. The instrument does not
prevent the processing protocol from starting if the slot is not at the target temperature. Actual
slot temperatures can be viewed on the instrument display. Refer to Preheating and Heating
in section Operating the Incubator Shaker 50x.
Allow the slots to reach the programmed target temperature before starting any processing
protocol. The slot has reached target temperature when the orange LED has stopped flashing
and stays a continuous orange color.
Slot Temperature
Ensure that the temperature of the slot is within the acceptable range appropriate for the
application to be carried out before starting the processing protocol. The actual slot
temperature can be viewed on the instrument display.
Shaking
Module Shake Speed Set By the Most Recently Started Processing Protocol
The slots within a module shake at the speed that is programmed into the most recently
started processing protocol. Incorrect shaking speed may result if the operator starts a
protocol within the same module that has a shaking speed different than the other slots
already in use. To minimize the likelihood of this occurring, it is recommended that all
processing protocols within a module be programmed with the same shaking speed.
Interval Times
Interval time is a programmable parameter that allows a user to specify a period of time
during which shaking is alternately switched off and on (e.g. an interval time of 00:00:10
results in 10 seconds of shaking turned on, 10 seconds of shaking turned off, 10 seconds of
shaking turned on, etc.) for the duration of the programmed shaking time.
Operation
Ensure that the microplate holder is correctly inserted into the slot before starting the
processing protocol. When inserting a microplate holder into a slot, push the microplate
holder completely in and down to latch it into its normal operating position. If the microplate
holder is not latched into position, it may lose contact with the temperature controlling
connectors at the rear of the slot.
I n s t r u m e n t R e s e t D u e t o P o we r L o s s
If power to the instrument is lost, the instrument will reset to the READY mode when power is
restored. All shaking and incubation functions, including timing, will be turned off. No LEDs
will be lit and the current time of day will be shown on the display.
Maintenance
T u r n P o we r O f f a n d U n p l u g I n s t r u m e n t D u r i n g M a i n t e n a n c e P r o c e d u r e s
To prevent electrical shock, turn the power off to the instrument and unplug the power cable
during maintenance procedures.
Ensure that the power contacts on the back of the microplate holder are clean and free from
dirt or corrosion. Dirty contacts may affect temperature control of the microplate holder in the
slot.
Corrosive Materials
Technical Specifications
General Specifications
PARAMETERS CHARACTERISTICS
Weight 1-module instrument 15 kg
Overall dimensions 400 mm (W) x 450 mm (D) x 220mm (H)
Mains power supply 100 – 120 Vac, 60 Hz, nominal (operating range 90 to 136 V)
200 − 240 Vac, 50 Hz, nominal (operating range 188 to 256 V)
Power consumption 380 VA max.
2 x T3.15 A/250 V, 5x20 mm
Fuses
IEC 127-3 / III
Operating conditions +15°C to +40°C,
(indoor use) RH 90% max.
-40°C to +70°C,
Transportation conditions
packed in transport packaging
-25°C to +50°C,
Storage conditions
packed in transport packaging
Altitude Up to 2.000 m
Installation category
II
(overvoltage category):
Pollution degree: 2
Incubator specifications
PARAMETERS CHARACTERISTICS
Programmable +14°C to +40°C
temperature range
Controlled incubation Ambient plus 3°C to +40.0°C
range
Resolution 0.1°C
Incubation time Up to 48 hours in steps of 1 second
Slot warmup time ≤ 20 minutes to reach a 37°C ± 1.0°C target temperature in the empty plate carrier
when starting from ambient temperature (approximately 24°C)
Inaccuracy ± 0.3°C (calculated as the mean of the minimum and maximum well temperatures of
wells A1, A12, H1, H12 and E7, after 1 hour at 37°C, with an ambient temperature of
24°C, with 200 µL water/well, without a plate sealer, and without shaking)
Uniformity ≤ 0.3°C (calculated as the difference between the maximum and minimum well
temperatures of wells A1, A12, H1, H12 and E7, after 1 hour at 37°C, with an ambient
temperature of 24°C, with 200 µL water/well, without a plate sealer, and without
shaking)
Evaporation ≤ 300 mg/plate after 1 hour (at 37°C, without plate sealer)
Shaker Specifications
PARAMETERS CHARACTERISTICS
Frequency 400 to 1400 revolutions per minute (rpm) in steps of 250 rpm
Speed settings:
0 = 0 rpm 3 = 900 rpm
1 = 400 rpm 4 = 1150 rpm
2 = 650 rpm 5 = 1400 rpm
Tolerance ± 50 rpm
Amplitude 1 mm (radius 0.5 mm)
Shaking time Up to 48 hours in steps of 1 second
Interval time Up to 48 hours in steps of 1 second
Description
F r o n t V i ew
1. Incubation slot
2. Incubation module with three slots
3. Control LED (Light Emitting Diode)
4. Display
5. Keyboard
6. Power input socket
7. Power switch
8. Voltage selector and fuses
R e a r V i ew
1. LCD display
2. Keyboard keys
The alphanumeric LCD display has two lines with up to 20 characters per line.
Keys Functions
stop Stops the ongoing operation.
temp Sets the incubation temperature ON or OFF. Used for
starting preincubation (preheating) of the thermal microplate
holders to the temperature set in processing protocol
parameters.
shake Starts shaking with speed and time set in processing protocol
parameters.
start Starts operation defined by processing protocol parameters.
C Clears an incorrectly keyed numeric value.
Enters a keyed value or continues the program.
The thermal microplate holder is designed for holding the microplate during the incubation
process, ensuring accurate temperature control.
IMPORTANT ! Adjust the screws (5) with a screwdriver so that the microplate is held firmly in place
during shaking.
• Place the microplate into the thermal microplate holder so that well A1 is located in the
upper left corner as shown in Figure 2-4.
CAUTION! Care should be taken not to spill the contents of a microplate when loading it into the
holder and when inserting the holder into the slot. Carefully insert the plate into the
holder and close the cover. The covers of the microplate holders are not spring-loaded,
but care should be taken not to trap your fingers when closing the cover of the holder.
Avoid bumping the microplate holder while inserting it into the slot.
DO NOT handle the thermal microplate holder from its cover since it will come off.
Note: When inserting a thermal microplate holder into a slot, push the microplate holder completely
in and down to latch it into its normal operating position. If the microplate holder is not latched
into position, it may lose contact with the temperature controlling connectors at the rear of the
slot.
Unpacking procedure
• Unpack the Incubator Shaker 50x carefully and check for any damage that may have
occurred during transportation
CAUTION! If you observe any damage on the packaging or equipement, contact your bioMérieux
representative before installing the instrument.
Environmental requirements
• Locate the Incubator Shaker 50x to avoid exposure to excess dust, vibrations, strong
magnetic fields, direct sunlight, draft, excessive moisture or large temperature
fluctuations.
• Install the Incubator Shaker 50x on a sturdy, level surface.
• Leave sufficient clearance (10 cm) behind and at the right side of the instrument for
adequate air circulation.
Voltage Selector
The voltage selector/fuse block assembly is located on the left-hand side of the instrument
(see (8), Figure 2-1).
1. Fuse holder
2. Voltage selector
3. Fuses
WARNING! Disconnect the power supply before changing the fuse or voltage selector and before
removing any access panel.
For protection against fire hazard, replace only with the same type and rating of fuse.
Not operator serviceable parts inside. Refer servicing to qualified personnel.
P ow e r c a b l e
• Connect the mains supply cable to the mains power socket. If a suitable mains supply
cable is not provided, contact your local bioMérieux representative.
Use only cables approved by the local authorities.
• Connect the instrument to a correctly installed line power outlet, which has a protective
conductor that is grounded.
CAUTION! Never operate your instrument from a power outlet that has no ground connection.
Setup
Principle of operation
POWER ON
READY
READY READY
Programming protocols
Operation Display
Operation Display
If no shaking is desired,
• select the value 0.
In this incubation module, shaking is disabled in every PLACE 2
slot. PROHIBITED OPERATION
Note: If you try to start a slot with a protocol with shaking, an
error message will appear on the display:
READY 09:25:15
9. On the display the READY state will appear after the
programming phase.
Each slot of the Incubator Shaker 50x has a control LED that displays the status of the slot.
The LED will change to a specific color based on the status of the slot.
The colors of the LED are defined below.
Normal operation
Operation Display
2. The protocol will start and the LED will light up with a red READY 09:25:15
color.
Operation Display
4. Heating:
• Press START to turn the heating on.
• Press STOP to turn the heating off.·
• Press a number key or ENTER to move to another slot.
Before the temperature has reached the set value, the
LED flashes with an orange color. After the temperature
has reached the set value, the orange LED lights up
continuously.
• Press TEMP 0 to display the heating status of all slots.
PLACES 1-9
5. If the heating is ON in every slot, the display will show: HEAT ON
PLACES 1-9
6. If the heating is OFF in every slot, the display will show: HEAT OFF
7. If some slots are ON and some are OFF, the display will
PLACES 1-9
show: HEAT ON/OFF
8. At this point,
• press START to turn the heating ON in all slots READY 09:25:15
or
• press STOP to turn the heating OFF in all slots.
• Press C to return to the READY state.
Shaking
Operation Display
READY 09:25:15
4. After the shaking protocol has finished, the control LED turns
to green and a beeping sound is heard.
The parameters programmed for each slot can be displayed. To display the programmed
parameters, press the appropriate slot number and then press ENTER. Continue to press
ENTER to display the programmed parameters for temperature, incubation time, shake time,
interval time and shake speed.
CAUTION! Programmed processing parameters can only be viewed while in the programming
mode. Care should be taken during review of parameters to ensure that unintentional
changes are not made to the programmed values.
For reliable daily operation, keep the instrument free of dust and spills from liquids. It is also
advised to cover the instrument with the dust cover when not in use.
In the event of any damage, contact your local bioMérieux representative for service.
Cleaning Procedures
DANGER! PRIOR TO CLEANING, DISCONNECT THE Incubator Shaker 50x FROM THE MAIN
POWER SUPPLY TO AVOID ANY RISK OF FIRE AND EXPLOSION.
WARNING! The cleaning procedures and the disinfectants should comply with the local applicable
regulations.
In the event of spills of infectious agents on the surfaces of the instrument, the following
disinfectant procedure should be performed:
• Switch the power off and unplug the instrument.
In the event of spills of infectious agents on the surfaces of the thermal microplate holder, the
following disinfectant procedure should be performed:
• Use disposable gloves.
• Absorb the spill with disposable towels.
If the above mentioned cleaning procedure is insufficient because of the extensiveness of the
spill of infectious agent, the following disinfectant procedure should be performed:
• Place the thermal microplate holder into a container containing disinfectant solution with
2% glutaraldehyde for 30 minutes (or according to the directions provided by the
manufacturer of the disinfectant solution).
• Rinse the thermal microplate holder with water.
Periodically inspect the connectors on the microplate holders for dirt or corrosion. If dirty or
corroded, wipe the connectors with a soft cloth slightly dampened with water.
Refer to the instructions for use in the "General safety and regulatory information" booklet and
follow the instructions described in the section "Cleaning Procedures" on page 5-1 of this
manual.
Replacing Fuses
WARNING! Disconnect the power supply before changing the fuse or voltage selector and before
removing any access panel.
For protection against fire hazard, replace only with the same type and rating of fuse.
Not operator serviceable parts inside. -Refer servicing to qualified personnel.
The fuses are located on the left side of the instrument (see (8), Figure 2-1).
• Switch the instrument off.
• Unplug the instrument and disconnect the power cable from the power input socket.
• Loosen the fuse holder (see (1), Figure 5-1).
• Replace the blown fuses with the same type (3.15 A).
• Place the fuse holder into the slot and push it in until a click is heard.
• Connect the power cable to the instrument and switch it on.
1. Fuse holder
2. Voltage selector
3. Fuses