Document number: 841060_H_applic_AST 2, Date: 02/04, s.w. 6.

0
Test definition AST 2 Calibr. parameters
Test type Photometric Test in use YES Calibration type NONE Factor -1481 Bias 0
Full name AST IFCC, 2 Low High Units
reag. system
Online name Test limit 6 1800 U/l Bias corr. in use NO
Result Unit U/l low( Ö ), high

Number of decimals 0 Initial absorbance 1 2.5 A

Acceptance AUTOMATIC

Dilution 1 + 0 Dilution limit * 300 U/l


Sample type
Secondary dil. 1+ 0 5
ýSerum ýPlasma ¨Urine
¨C.S.F ¨Other Correction factor 1
More
Correction bias 0

Test flow Reagent definition


Blank NO Normal cuvette
Antigen excess NO

Syringe BCR Id. Stable


speed on board
Name (days)
Curve type
Incubation Incubation
Reagent Volume (µl) Time (sec.) Reagent Time (sec.) λ 1 (nm) LINEARCUT
AST1 15 300 AST2 120 340 Nonlinearity
Resp. (mA/min)

Volume (µl) Disp. with Volume (µl) λ 2 (nm) 20 AST1 Normal 015 7
100 EXTRA 25 380

Disp. with Volume (µl) Disp. with Res. Net. Abs. AST2 Normal 016 7
EXTRA 15 EXTRA 0.4
Time (sec)
240
Volume (µl) Dilution with Volume (µl)
20 WATER 8
Points & Inter.
Konelab
Wash reagent Wash reagent Wash reagent 60/30/20XT 9/27
Konelab 20 6/42
NONE NONE NONE

Konelab™ AST (IFCC, two reagents) - code: 981363 (7 x 50 ml), code: 981771 (8 x 20 ml)

Reagent: Code 981363: Add 400 µl of Pyridoxal phosphate 981790 into one bottle of reagent A. Reagent B is ready to use.
Code 981771: Add 160 µl of Pyridoxal phosphate 981790 into one bottle of reagent A. Reagent B is ready to use.
Ö measuring range low

Q.C. sample: Nortrol 981043 & Abtrol 981044. Reconstitute with 5.00 ml of distilled water.